Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Similar documents
SARASOTA MEMORIAL HOSPITAL

Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts

Figure 2. Normal ECG tracing. Table 1.

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

RN-BC, MS, CCRN, FAHA

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Peri-operative management of pacemakers and implantable cardiac defibrillators

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Pacing Lead Implant Testing. Document Identifier

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Model 5392 EPG Temporary Pacer

Panorama. Arrhythmia Analysis Frequently Asked Questions

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Project Title Temporary Pacemaker Training Simulator

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Guidelines/British Heart Rhythm Society Guidelines Contributes to CQC Regulation 12. Consulted With Post/Committee/Group Date Michelle Colton-Goff

ICD Implantation Patient Information

Electrocardiography for Healthcare Professionals

You have a what, inside you?

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

Temporary pacemaker : Programming &Trouble shooting

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER)

You have a what, inside you?

Temporary Pacemakers

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

Review Packet EKG Competency This packet is a review of the information you will need to know for the proctored EKG competency test.

Cardiac Resynchronisation Therapy Patient Information

ICD: Basics, Programming and Trouble-shooting

TIP CARD. Model Temporary External Pacemaker

Is This Thing Working?

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Pediatric pacemakers & ICDs:

PEDIATRIC CARDIAC RHYTHM DISTURBANCES. -Jason Haag, CCEMT-P

Implantation of an Implantable Cardioverter Defibrillator (ICD)

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.

Chapter 80 Implantable Cardiac Devices

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital

Supplemental Material

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

Defibrillator. BiomedGuy

Implantation of a Cardiac Resynchronisation Therapy-Defibrillator (CRT-D)

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

Kadlec Regional Medical Center Cardiac Electrophysiology

1. What additional information needs to be collected to properly treat this client?

Pacemakers 12/04. Pacemakers. 1. What is a pacemaker?

Pacing after acute myocardial infarction

BEDSIDE ECG INTERPRETATION

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Pacemaker and AV Node Ablation Patient Information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network

Title of Guideline (must include the word Guideline (not. Guidelines. Contact Name and Job Title (author)

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

Cardiac Resynchronisation Therapy Defibrillator (CRT-D)

Cardiac implantable electronic devices (CIEDs) in children include pacemakers and implantable cardioverter defibrillators (ICDs).

Implantable Cardioverter Defibrillator (ICD) and Cardiac Resynchronisation Therapy Defibrillator (CRT-D)

CORONARY ARTERIES HEART

Cardiac arrhythmias. Janusz Witowski. Department of Pathophysiology Poznan University of Medical Sciences. J. Witowski

Dual Chamber Temporary External Pacemaker

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

Information for patients, parents and guardians. Your child s doctor has recommended that your child has a procedure called an ablation.

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

The Electrocardiogram

Arrhythmias. Pulmonary Artery

Quality Standards for the Implantation of Cardiac Rhythm Management Devices. Pan- London Arrhythmia Project Group. Version 3 (18 th July 2011)

There are different types of ICDs:

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)

STANDARD OPERATING PROCEDURE (SOP)

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Different indications for pacemaker implantation are the following:

What to expect when having a pacemaker implantation

Diploma in Electrocardiography

This leaflet is intended for patients with permanent pacemakers and their parents.

The Nuts and Bolts of ICD Therapy

Advanced ICD Concepts

About atrial fibrillation (AFib) Atrial Fibrillation (AFib) What is AFib? What s the danger? Who gets AFib?

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Transvenous Pacemaker Procedures

NHA Certified EKG Technician (CET) Test Plan for the CET Exam

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C

EXTERNAL TEMPORARY PACEMAKERS

Evaluation of pacing for heart block

The Results of Demand Pacing

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8)

Supplementary Online Content

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Guide to Cardiology Care at Scripps

Transcription:

Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division: Division of Salford Health Care Unique ID:TWCG31(12) Issue number: 2.2 Expiry Date: January 2021 Contents Section Page Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2 Guideline 3 1 Indication & Procedure 3 2 Personnel and Equipment 3 2.1 Personnel 3 2.2 Equipment 4 3 Transfer of Patients 4 3.1 Transfer of patients to and from the designated area 4 3.2 On return to the critical care environment 4, 5 3.3 Equipment required at the bedside/available on ward 5 4 Daily Management 5 4.1 Pacing threshold Testing Procedure 6 5 Troubleshooting 6-9 6 Desirbale chest X-Ray appearance 10 7 Roles and Responsibilities 11 8 Explanation of Terms and Definitions 11 9 Appendix - Pathway for patients with temporary pacing wires 12 10 Document control information (Published as separate document) 13 Document Control 13 Policy Implementation Plan 13 Monitoring and Review 14 Endorsement 14 Equality analysis 15-17 Page 1 of 12

Who should read this document? This guideline is for use by all medical or nursing staff caring for patients with a temporary pacing wire. This could apply to patients on HCU/ICU/NHDU/MHDU/SHDU or any other ward with a patient with temporary pacing wire. Key Practice Points Temporary transvenous pacing may be required for the urgent management of bradyarrhythmias and of some tachyarrhythmias Temporary transvenous pacing is a specialist cardiology procedure and cardiology advice and assistance should be sought in all cases Scope of the Guideline 1. Indications for temporary pacing 2. Personnel and Eqiuipment 3. Transfer of patients with or requiring temporary pacing 4. Daily management of temporary pacing wire 5. Troubleshooting Background Temporary cardiac transvenous pacing may be required for the management of bradyarrhythmias, and also of some tachyarrhymias causing, or potentially causing, haemodymamic compromise. The requirement may be urgent or more elective depending on the circumstances. Temporary cardiac transvenous pacing is a specialist cardiology procedure and if it is required, the cardiology team should be contacted directy. Cardiology advice and availability for temporary pacing is provided on a 24/7 on call basis. What is new in this version? Change of approval date Page 2 of 12

Guideline 1. Indications and Procedure Temporary pacing wire (TPW) insertion should be considered for: bradycardia with haemodynamic compromise ventricular arrhythmia suppression termination of tachyarrhythmia by overdrive pacing prophylactic temporary pacing preoperatively The procedure will be carried out by cardiology middle-grade or consultant staff. Patients in Ladywell block will have TPW performed in the HCU pacing room and thereafter remain on HCU until arrhythmias have been adequately resolved. Patients in intensive care will have TPW performed in Theatre and then return to ICU. Patients on other wards will either be transferred to HCU or have TPW in Theatre depending on which is decided to be more appropriate. Patients from MHDU, SHDU, NHDU may return to these wards post TPW insertion, patients from other wards should be transferred to HCU. Patients with prophylactic preoperative TPW may be returned to base ward post procedure. 2 Personnel and Equipment 2.1 Personnel: Cardiology consultant or cardiology middle-grade (1 st operator) Cardiology nurse trained in TPW insertion (for HCU insertions) Critical care nurse/theatre staff trained in level 2/3 (for theatre insertions) Anaesthetist if appropriate Radiographer Medical & nursing personnel should be ALS trained Page 3 of 12

2.2 Equipment: C-arm image-intensifier Temporary pacing box 5F temporary pacing wire 5F or 6F introducer sheath External pacemaker / defibrillator Cardiac monitor with non invasive / invasive monitoring facilities Oxygen and air inlet Wall-mounted suction TPW insertion will be performed via internal jugular or femoral or subclavian vein as deemed most appropriate by the team. Insertion will be performed with full sterile technique according to Trust protocol. 3 Tranfer of Patients 3.1 Transfer of patients to and from the designated area All patients requiring TPW insertion in HCU or Theatre will be transferred with external pacemaker/defibrillator immediately available and with an ALS provider. Transferring staff will follow the Trust transfer policy. 3.2 On return to the critical care environment: Bedside handover will be given to the nursing and medical staff by the cardiology staff / nurse trained in TPW. Documentation completed by the cardiology team on the pathway Appendix 1. Pacing safety checks will be completed by the receiving nurse and the cardiology staff and documented on the pathway. Patients will remain on continuous cardiac monitoring and baseline EWS completed and documented on EWS chart Page 4 of 12

Monitor alarm limits to be set according to TPW settings. Observation will be maintained as per EWS policy Post-implant CXR will be performed to document TPW position and exclude pneumothorax. 3.3 Equipment required at the bedside/available on ward: Temporary pacing box with spare set of batteries Continuous cardiac monitoring (excluding patients with prophylactic TPW) External defibrillator/pacer (excluding patients with prophylactic TPW) Immediate access to crash trolley and emergency drug box 4. Daily Management HCU will keep a record of pacing box use across the trust. Cardiology team will review TPW Mon - Fri, complete the safety checks with the medical and nursing staff from ITU/MHDU/NHDU/SHDU, and be available for advice/troubleshooting Sat/Sun HCU nursing staff to contact ITU/MHDU/NHDU/SHDU on a shift basis, to coordinate the checking of the TPW threshold. TPW safety checks to be completed by nursing staff on a shift basis (or every 8 hours) and documented on the pathway. Central line pathways should be completed every shift on EPR The TPW wire will be removed when no longer required. Cardiology will arrange an appropriate date and time for implantation of a permanent pacemaker if required. 4.1 Pacing threshold Testing Procedure 1. Set RATE at least 10 bpm above patient s intrinsic rate. Green PACE light flashing 2. Decrease OUTPUT: Slowly turn Output dial anticlockwise until ECG shows loss of ventricular capture. Green PACE light stops flashing. Page 5 of 12

3. Increase OUTPUT: Slowly turn Output dial clockwise until ECG shows consistent capture. Green PACE light flashing. This value is the stimulation threshold (Volts) 4. Set OUTPUT to a value 2 to 3 times greater than the stimulation threshold value. This provides x2-3 safety margin. 5. Restore RATE to previous value. Ideally, the threshold should be <1.0V The threshold will gradually rise day by day ensure output is x2-3 greater than the threshold. Page 6 of 12

5. Troubleshooting Normal ventricular pacing Ventricular pacing spikes followed by wide QRS complexes 5.1 Ventricular non-capture Ventricular pacing spikes are not followed by QRS complexes Causes of Failure to Capture: Insufficient output Low pacemaker battery Displaced or fractured lead Electrolyte abnormalities: acidosis; hypoxia; hypokalaemia Management: Increase output on Output dial Seek cardiology advice lead may need to be repositioned or replaced Check electrolytes View rhythm in different leads Change electrodes Check connections Change battery, cables, pacemaker box Sensitivity threshold Minimum level of intrinsic electric activity generated by the heart detectable by the pacemaker Page 7 of 12

5.2 Ventricular undersensing Ventricular pacing spikes occur regardless of QRS complexes Pacemaker is not seeing intrinsic activity potential for paced ventricular beat to land on T wave with potential for inducing ventricular fibrillation: Management: Increase sensitivity (reduce mv on Sense dial) Seek cardiology advice lead may need to be repositioned or replaced View rhythm in different leads Change electrodes Check connections Change cables, battery, pacemaker box Check electrolytes Page 8 of 12

5.3 Ventricular oversensing Pacing does not occur when intrinsic rhythm is inadequate Causes: Pacemaker inhibited due to sensing of QRS complexes that do not exist Pacemaker too sensitive Possible lead displacement Pacemaker failure Danger of heart block, asystole Management: Reduce sensitivity (increase mv on Sense dial) Seek cardiology advice lead may need to be repositioned or replaced View rhythm in different leads Change electrodes Check connections Change cables, battery, pacemaker box Check electrolytes 3.4 Sensitivity threshold testing procedure: 1. Set Rate 10 bpm slower than patient s HR (Green SENSE light flashing). 2. Increase sensitivity to chamber being tested to minimum level (0.2mV) 3. Decrease sensitivity of the pacer (turn Sense dial anticlockwise, mv) until pacer stops sensing patient (green SENSE light stops flashing) 4. Increase sensitivity of the pacer (turn Sense dial clockwise, mv) until the pacer senses the patient (green SENSE light begins flashing). This is the threshold for sensitivity. 5. Set the sensitivity at ½ the sensitivity threshold value. 6. Example: Set sensitivity at 1mV if the sensitivity threshold was 2mV Page 9 of 12

6. Desirable chest X-ray appearance Pacing lead appearance from R subclavian vein approach The pacing lead tip (arrow) is at the right ventricular apex Pacing lead appearance from femoral vein approach. The pacing lead tip (arrow) is at the right ventricular apex. Page 10 of 12

7. Roles and responsibilities All staff involved in the management of patients with temporary pacing wires should adhere to this guideline The Cardiology Department is responsible for ensuring this guideline is up to date. 8. Explanation of terms & Definitions Temporary Pacing Wire (TPW): A temporary wire positioned within the right ventricle of the heart for the purpose of assisting in the control of the cardiac rhythm. It is positioned percutaneously under fluoroscopic guidance with venous access via the femoral, internal jugular or subclavian vein. Pacing threshold: The minimum voltage delivered by the temporary pacing wire with which the ventricle is consistently paced. Ideally this should be <1.0V, although this not always possible. Sensitivity threshold: Minimum level of intrinsic electric activity generated by the heart detectable by the pacemaker Ventricular non-capture: Explained in document Ventricular undersensing: Explained in document Ventricular oversensing: Explained in document Page 11 of 12

9. Appendix Pathway for patients with temporary pacing wires Name: Date of birth: Hospital Number: Ward/Unit: Date/Time of insertion Screening time (mins) Threshold (mv) Output set @ mv Demand rate Underlying rhythm Complete the SRFT Central Line Flowsheet 1. Continuous cardiac monitoring & record EWS 2. Set cardiac monitor alarm limits 10 beats below pacing box rate and 10 beats above the pacing box rate. 3. Check threshold of TPW every 24 hours and document (cardiology staff will also perform daily). 4. Ensure safety of pacing box at all times. 5. Check TPW connections are securely attached. 6. Observe insertion site for any signs of infection. 7. Ensure site is dry and clean change dressing in line with SRFT central line policy 8. Assess pain control and administer analgesia as prescribed. 9. Complete all central line pathways on daily shift on isoft 10. Liaise with cardiology team regarding management plan. Pacing Record Date/time Threshold Output Demand rate Underlying rhythm Signature Date/time Threshold Output Demand rate Underlying rhythm Signature Date/time Threshold Output Demand rate Underlying rhythm Signature Page 12 of 12