FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

Similar documents
CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Comparison of Stent Designs using Computational Fluid Dynamics

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Developing Pulsatile Flow in a Deployed Coronary Stent

Computational Structural Modelling of Coronary Stent Deployment: A Review

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Computational Analysis on Commercially Available Stent Designs

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

FOR many decades, cardiovascular disease has been one of

The influence of strut-connectors in stented vessels: A comparison of pulsatile flow through

Post-conditioning. P a g e 1. To my Thesis Committee,

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Medical device design using Computational Fluid Dynamics (CFD)

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Developed pulsatile flow in a deployed coronary stent

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Analysis of the effects of plaque deposits on the blood flow through human artery

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Blood flow in vessels with artificial or pathological geometrical changes

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration

Computational design of Intracranial Stent using 3D visualization system

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

Blood Flow Simulation toward Actual Application at Hospital

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

CPM Specifications Document Aortic Coarctation: Exercise

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

Modeling of stent implantation in a human stenotic artery

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Investigation of the Hemodynamics of Coronary Arteries - Effect of Stenting. Naresh Kumar Coimbatore Selvarasu

CHALLENGES IN FEMORO-POPLITEAL STENTING

Dynamic simulation of stent deployment effects of design, material and coating

Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Edinburgh Imaging Academy online distance learning courses

An Agent-Based Model of the Response to Angioplasty and Bare-Metal Stent Deployment in an Atherosclerotic Blood Vessel

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

Left main coronary artery (LMCA): The proximal segment

Computational analysis of mechanical stress-strain interaction of a bioresorbable scaffold with blood vessel

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery

Fractional flow reserve and the influence of bifurcation angle and degree of stenosis

Vascular reconstruction: CFD predictions of bypass graft haemodynamics

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

Senior Honors Thesis Prospectus

CYPHER (Polymer-based Sirolimus-eluting DES) New Stent Platforms. Campbell Rogers, M.D. Chief Technology Officer

Flow Analysis on Fusiform Aneurysm during Exercise Condition

In-stent Restenosis: the Achille's Heel of SFA Stenting

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

5-YEAR SUPERIOR RESULTS. Compared to PTA and Zilver BMS 1. NOW WITH 140 mm LENGTH STENTS

Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Numerical analysis of blood flow in human abdominal aorta

ON FLOW DYNAMICS IN TYPE B AORTIC DISSECTION (TBAD)

Effects Of Red Blood Cell Aggregation On Wall Shear Stress In µ-tube System

restenosis; neointimal hyperplasia; scaffolding; prolapse; three-dimensional

CPM Specifications Document Aortofemoral Normal:

Marquette University Sung Kwon Marquette University Recommended Citation

Planar Titanium Stent Design

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

Treatment of lesions at bifurcations: General perspective

Flow features in and around a (stented) coarctation

Mathematical and Computational study of blood flow through diseased artery

Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations

Investigation of metallic surface area of coronary stents

Dynamic Role of the Cardiac Jelly

4D model of hemodynamics in the abdominal aorta

Flow in re-stenosed artery after angioplasty

When Outcomes Matter, Design Matters

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

THE COMPREHENSIVE FINITE ELEMENT MODEL FOR STENTING: THE INFLUENCE OF STENT DESIGN ON THE OUTCOME AFTER CORONARY STENT PLACEMENT

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Evolution In Interventional Cardiology. Jawed Polad Jeroen Bosch Hospital s-hertogenbosch The Netherlands

Resolute Integrity. Independent Conformability Assessment

Between Coronary Angiography and Fractional Flow Reserve

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation

IDEALIZED ABDOMINAL AORTIC ANEURYSM (AAA) GEOMETRY AS PREDICTOR OF HEMODYNAMICS STRESSES

NUMERICAL SIMULATIONS OF BLOOD FLOW IN ARTERIES USING FLUID- STRUCTURE INTERACTIONS. A Dissertation by. Eleyas Shaik

3D Modeling of Plaque Progression in the Human Coronary Artery

SHREYAS S. HEGDE et al: COMPUTATIONAL FLUID DYNAMIC APPROACH TO UNDERSTAND THE EFFECT

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE

Transcription:

LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca, Gabriele Dubini LaBS, Department of Structural Engineering, Politecnico di Milano, ITALY

INTRODUCTION 2

INTRODUCTION 3 A vascular stent is a small metal tube, which is inserted into an artery at the site of a narrowing to act as an internal scaffolding or a support to the blood vessel.

MOTIVATION 4 IN-STENT RISTENOSIS Intimal thickening following a stent implantation with progressive lumen reduction [Mehran R., 2002] Three phases (Edelman e Rogers, 1998): INFLAMMATION during implantation + PROLIFERATION first 3 weeks + REMODELING 10/12 months HYPOTHESIS: non physiological stress state field responsible for restenosis.

STATE OF THE ART 5 QUANTITIES OF INTEREST Effect of wire spacing, wire diameter, vessel diameter and flow conditions Stent design: number, thickness and width of the strut Deployment ratio Comparison of resting or maximal vasodilatation condition Foreshortening Changes in vascular geometry after stent deployment [Moore et al.,2002] [LaDisa et al.,2003-2004-2005] Effect of vessel curvature [Seo et al., 2005] Non-Newtonian condition [Soulis et al.,2002; Seo et al.,2005; Bernard et al.,2004] QUANTITATIVELY OBSERVED PARAMETERS wall shear stress (WSS) distribution velocity vectors recirculation length velocity profiles

THE PROBLEM 6 FROM SIMPLIFIED MODELS TO PLAQUE MODEL Migliavacca et al., Proceedings of 2005 Summer Bioengineernig ASME Conference Healthy artery Artery with plaque Expansion under displacement control until a diameter of 3 mm was reached The stent geometry was modelled as shell elements Cordis BX Velocity (Johnson & Johnson Interventional System, Warren, NJ, USA)

METHODS 7 1. Preliminary step: structural analysis This step is necessary to obtain the correct configuration for the fluid dynamics simulations: fluid domain

METHODS 8 First step: creation of the fluid domain 1 Point cloud of the deformed configuration 2 Creation of the curves and surfaces 3 Creation of each volume 4 Substraction and creation of the final fluid domain

METHODS 9 Second step: Boundary conditions OUTLET Constant fixed pressure ASSUMPTION: - rigid vessel wall - Newtonian fluid: Viscosity = 0.0035 kg/(m s) Density = 1060 kg/m 3 0.2 0.16 0.12 LaDisa et al. (2005) WALL [m/s] 0.08 0.04 No slip condition 0 0 0.1 0.2 0.3 0.4 0.5 0.6 Time [s] 4 cardiac cycles pulse period = 0.54 s INLET Velocity profile: parabolic and transient Fluent (Fluent Inc., Lebanon, NH, USA)

OBSERVATIONS 10 STENTED REGION 0.16 s dynes/cm 2 HEALTHY MODEL 50 PLAQUE MODEL 25 0 The highest WSS magnitude can be noticed on the stent

OBSERVATIONS 11 ARTERIAL REGION INSIDE STENT STRUTS 0.16 s dynes/cm 2 18 9 0 HEALTHY MODEL PLAQUE MODEL high WSS in the regions between the stent struts low WSS were localized around stent struts

AIM OF THE STUDY 12?? Is it correct to ignore the presence of an atherosclerotic plaque four different stent designs previously expanded against the same stented artery Cordis BX Velocity stent like (Johnson & Johnson Interventional System, Warren, NJ, USA) Jostent Flex stent like (JOMED AB, Helsingborg, Sweden) Sorin Carbostent stent like (Sorin Biomedica S.p.A., Saluggia (VC), Italy) Palmaz-Schatz stent like (Johnson & Johnson Interventional System, Warren, NJ, USA) transient simulation for each model comparison of the WSS magnitude distribution during time

STENT MODELS 13 CORDIS JOSTENT RADIUS a fter exp a nsio n LENGTH a fter exp a nsio n THICKNESS CORDIS 1.5 3.53 0.1 JOSTENT 1.5 2.30 0.1 PALMAZ 1.5 2.97 0.1 SORIN 1.55 3.50 0.1 SORIN PALMAZ Length: 11.68 mm Internal diameter: 2.15 mm Thickness: 0.5 mm Length: 3.68 mm Internal diameter: 1.25 mm Thickness: 0.45 mm

% of cells RESULTS: WALL SHEAR STRESSES 14 WSS < 5 dynes/cm 2 correlated with sites of intima thickening and smooth muscle cells migration locations where stagnation of blood occurs prone to thrombus formation and platelet accumulation 100 95 0.16 s 0.32 s 0.44 s 90 0 s 0.4 s 85 0 s 0.16 s 0.32 s 0.4 s 0.44 s

% of cells RESULTS: LOW WSS 15 CORDIS WSS < 5 dynes/cm 2 SORIN 0 s 100 95 90 85 0 s 0.16 s 0.32 s 0.4 s 0.44 s JOSTENT [dynes/cm 2 ] 5 2.5 0 PALMAZ

% of cells RESULTS: LOW WSS 16 CORDIS WSS < 5 dynes/cm 2 0.16 s SORIN 100 95 90 85 0 s 0.16 s 0.32 s 0.4 s 0.44 s JOSTENT [dynes/cm 2 ] 5 2.5 0 PALMAZ

RESULTS: MAXIMUM WSS ON STENT 17 [dynes/cm 2 ] 50 40 30 20 10 0 0 s 0.16 s 0.32 s 0.4 s 0.44 s CORDIS PALMAZ JOSTENT SORIN

RESULTS: MAXIMUM WSS 18 [dynes/cm 2 ] 50 40 30 20 10 0 0.16 s

RESULTS: MAXIMUM WSS ON THE ARTERIAL WALL 19 [dynes/cm 2 ] 20 15 10 5 0 0 s 0.16 s 0.32 s 0.4 s 0.44 s CORDIS PALMAZ JOSTENT SORIN

LIMITATIONS AND ASSUMPTIONS 20 Rigid wall: valid in the stented region Newtonian fluid Straight vessel: neglecting the curvature of the coronary artery Post implant condition Single strut Symmetric and hyperelastic plaque

WORKS IN PROGRESS 21 Rigid wall: valid in the stented region Newtonian fluid Straight vessel: neglecting the curvature of the coronary artery Post implant condition Single strut Symmetric and hyperelastic plaque 2 2 n Carreau model: 1 [Seo et al., 2005] 0 1 S

WORKS IN PROGRESS 22 Rigid wall: valid in the stented region Newtonian fluid Straight vessel: neglecting the curvature of the coronary artery Post implant condition Single strut Symmetric and hyperelastic plaque Influence of the stent length:

WORKS IN PROGRESS 23 Influence of the strut thickness: comparison of different stent design with same thickness 0.15 mm 0.1 mm CORDIS JOSTENT SORIN PALMAZ

CONCLUSIONS 24 In each stent model the WSS distribution is similar: the maximum values are located over the stent strut the arterial wall portion delimited by the links and the stent strut showed an increasing WSS value from the zones near the stent to the centre WSS values change during the cardiac cycle, showing an oscillatory behaviour The comparison among the four stent models indicates that: Jostent shows the lowest WSS value during the whole cardiac cycle the best model in terms of minimal neointima thickening is the Cordis stent the maximum WSS on the stent and the arterial wall occurs in the Cordis stent at the systolic peak CFD techniques have the advantages of producing accurate information on local flow variables very close to the arterial wall CFD can thus provide a research tool by complementing experimental studies, especially where experimental measurements are difficult to perform and affected by uncertainties.

25 THANK YOU