3Replacement Transducer Cables

Similar documents
The Facts about Biphasic Defibrillation

Troubleshooting Guide Your introduction to cochlear implants

Geschrieben von: Yadira Jones

Chapter 5: Checking and Maintaining Ultrasound Equipment

Convex Array Transducer

High Frequency Linear Array Transducer

Linear Array Transducer

Convex Array Transducer

Getting started with Wireless Fitting

QUICK SETUP GUIDE SECULIFE DFBASE

TWIST STEPPER NO. 045

Table of Contents. Introduction Indications For Use Contraindications Warnings Precautions...5

Geschrieben von: Mark

TWISTING STAIR STEPPER WITH BANDS NO. 068 USER MANUAL

Geschrieben von: Valentina Cerruty

A-dec 372L or 572L Dental Light on an A-dec Performer Dental Chair INSTALLATION GUIDE

Lily 说明书 尺寸 56*118mm 封面 250G白卡白珠光 烫银 内页 128g双铜纸四色印刷 装订方式 胶装 Tas i, L i! Lily 说明书

MacBook Pro 15" Unibody Mid 2010 Subwoofer & Right Speaker Replacement

Fitting Guide BEHIND-THE-EAR HEARING INSTRUMENTS. Pico RITE Hearing System

ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes

QUICK START GUIDE TYMPANOMETRY AND AUDIOMETRY FOR SENTIERO DESKTOP DEVICES

Jarvik 2000 VAS LAVENDER LAVENDER LAVENDER LAVENDER LAVENDER. 1. Can I do external CPR? Yes, only as a last resort.

Induction loop EZT Instruction manual

TIP CARD. Model Temporary External Pacemaker

Canon Powershot A70 Button Flex Assembly Replacement

GT350 Red Starter Button Kit M-10B776-MR

Bio-logic AuDX PRO FLEX Screening tympanometry and audiometry Quick start guide

Welcome to the S410. Blood Pressure System. Instruction Manual

SMART Biphasic. A Proven, Fixed, Low-Energy Defibrillation Waveform

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T

SP450 OWNER S MANUAL

Welch Allyn Connex Vital Signs Monitor 6000 Power Supply Replacement

Helping you save more lives. Philips HeartStart Resuscitation Solutions Leading the way in quality resuscitation

OWNER S MANUAL. BODYCRAFT 7699 Green Meadows Dr., Lewis Center, Ohio Tel:

MARCY Foldable Bike NS-652. Model NS-652. Retain This Manual for Reference OWNER'S MANUAL IMPEX INC.

Mac mini (PowerPC) Storage Interface Board Replacement

Fitting Guide BEHIND-THE-EAR HEARING INSTRUMENTS. Pico RITE Hearing System

Xbox 360 S Hard Drive Bracket Replacement

Early defibrillation programs for the workplace. Give your workplace the power to save a life

HealthCare International, Inc. PO Box 1509, Langley, WA P: (360) or (800)

Biomedical Engineering in Root Cause Analysis Example: Assessing Infant Apnea-Related Deaths

Bands. Model : No For Customer Service, please contact:

LET OUR IMAGE ENHANCE YOUR IMAGE. Operating Instructions MedRx Video Portable OtoScope System

I and T Shaped Intraoperative Transducer

Contents of the Prodigy Pocket. Meter. PRODIGY Owner s Manual PRODIGY Owner s Manual

TWISTING STAIR STEPPER WITH BANDS

E L E C T R O N I C VA P O R I Z E R S. User Manual

Geschrieben von: Phillip Takahashi

Compatible with the following products TENS. Replacement Pads. User Manual. Last revised: V Last revised: V

ESSENTIAL GUIDE. Companion Microphone. How-to videos on hihealthinnovations.com

Lesson 1: Types of ECG s

A m e r i c a s D e n t a l E q u i p m e n t C o m p a n y. Rev 2/15. D a n s e r e a u

Multipower. User Guide E S S E N T I A L S T R E N G T H

Electrocardiography for Healthcare Professionals

Convex Array Transducer User Guide. Types 8567-S and 8667

Prostate Biplane Transducer

ER75 Electro-Acoustic Ear Simulator. Operating Manual

Instructions For Use. SOLUTION FOR INJECTION IN PRE-FILLED PEN 250 micrograms

N4i Installation Manual. (picture of N4i screen)

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS

Blood Glucose Monitoring System. User Guide

Biomet SpinalPak Non-Invasive Spine Fusion Stimulator System

MacBook Pro 15" Core 2 Duo Models A1226 and A1260 ExpressCard Cage Replacement

Instruction Manual for minivap Vaporizer. By HERMES Medical Engineering SL.

8 Performing Medical Procedures

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

PedAlert 100. PedAlert 120

Glucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only

Activa RC Neurostimulator MEDTRONIC DBS CHARGING EXPERIENCE

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease

TORNADO AIR BIKE SF-B2706 USER MANUAL

TJF-160F/VF Cleaning and Disinfection Checklist

CP lead resting electrocardiograph

IC031 Magnetic Belt Drive Performance Indoor Cycle Bike

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x

LABORATORY INVESTIGATION

My CoughAssist. A patient guide to CoughAssist E70

Operating Instructions SaHoMa -II mobile ultrasonic nebulizer

Asus Transformer T100 Motherboard

Geschrieben von: Andrew Optimus Goldberg

MAGNETIC ELLIPTICAL TRAINER SF-E3608 USER MANUAL

ESSENTIAL GUIDE. Companion Microphone. How-to videos on hihealthinnovations.com

Escalade International Limited

GB Computer i-run se

Versatile, affordable ultrasound. Philips HD6 ultrasound system. The print quality of this copy is not an accurate representation of the original.

Installation Instructions

Assessing Arrhythmia Performance ST/AR Algorithm

1.0 SAFETY WARNINGS 2.0 CONTENTS OF KIT AND RE-ORDER CODES 3.0 SYSTEM SETUP 4.0 PARTS IDENTIFICATION - PHOTOGRAPHIC 5.0 GUIDE TO FITTING 5.

AWAY Model # RTLAGF-900 REV

Low Row. User manual E S S E N T I A L S T R E N G T H

4-20 ma Pressure Transducer

SP-300 OWNER S MANUAL

AMBCO 1000+P AUDIOMETER

CIRCULATORY MASSAGER OPERA TING INSTRUCTION HM01-08QI

SelectTech 3.1 Bench Assembly / Owner s Manual

ABR Screener Excellence in Newborn Hearing Screening

Activated Transponder & Activator Instruction Manual

Laerdal-SonoSim Procedure Trainer

ULTR A SERIES G7-S51

Transcription:

1Service Supplement 2M1355-61652, M1356-61661 and M1358-61681 3Replacement Transducer Cables Use the replacement transducer cables to repair BROWN transducers ONLY. M1355-61652 is the replacement cable for the brown Toco transducer (M1355A). M1356-61661 is the replacement cable for the brown Ultrasound transducer (M1356A). M1358-61681 is the replacement cable for the brown US/MECG Combi transducer (M1358A). CAUTION All blue M1355A, M1356A and M1358A transducers are sealed, watertight units and are NOT repairable. Replacing the cable of a blue transducer, or modifying the transducer in any way, will void any warranty, including the guarantee of watertightness. Dismantling the Ultrasound Transducer (Brown) To dismantle the Brown ultrasound transducer, proceed as follows: CAUTION Take care when removing the wires as they are thin and can easily be damaged. 1 Disconnect the transducer from the monitor. 2 Remove screw covers 1 and 2: press a flat screwdriver into the covers and prise them out. NOTE When reassembling the transducer, you will need new screw covers as the old ones cannot be re-used. 3 Partially undo screws 3. The two halves of the transducer can now be separated by pressing down on the screws. 4 Remove the screws 3 and pull the two halves of the cover apart. Printed in Germany March 2004 1990-2004 Koninklijke Philips Electronics N.V. All Rights Reserved. *M1350-9300A* Part number M1350-9300A 4535 640 05271

5 Carefully disconnect the transducer cable from the crystal board: use a soldering iron to separate the three cable wires from the board. Re-assembly is a reversal of the above procedure. Parts for the Ultrasound Transducer (M1356A) Ultrasound Transducer: Parts List Item Description Part Number 12NC Quantity 1 Case Top (US, Brown) M1356-44562 453563277221 1 2 Cover Screw (Red) M1356-44106 453563277211 1 3 Screw 0515-2121 453563024231 3 4 Cover Screw (Black) 5041-4274 453563099961 2 5 Cable Assembly (US) 2.5m / 8.2ft M1356-61661 453563277261 1 - Transducer Knob Adapters (not shown) M1356-43201 453563277191 3 Ultrasound Transducer: Cable Connections Pin Cable Color 4 Brown 8 Red 1+6 Bridged by Code Resistor 665 Ω 1 Shield 2

Dismantling the Toco Transducer (Brown) CAUTION Take care when removing the wires as they are thin and can easily be damaged. NEVER pull the strain gauge button. 1 Disconnect the transducer from the monitor. 2 Remove screw covers 1 and 2. Press a flat screwdriver into the covers and prise them out. NOTE CAUTION When reassembling the transducer, you will need new screw covers as the old ones cannot be re-used. 3 Partially undo screws 3. The two halves of the transducer can now be separated by pressing down on the screws. 4 Remove the screws 3 and pull the two halves of the cover apart. 5 Disconnect the transducer cable from the strain gauge 4. Use a soldering iron to carefully separate the three cable wires from the gauge. NEVER pull the strain gauge button. 6 Cut the clips holding the strain gauge and remove it. 7 Re-assembly is a reversal of the above procedure. The strain gauge assembly is a press fit into the transducer cover. 3

Parts for the Toco Transducer Toco Transducer: Parts List Item Description Part Number 12NC Quantity 1 Case Top (Toco, Brown) M1355-44552 453563277111 1 2 Cover Screw (Brown) M1355-44105 453563277091 1 3 Screw 0515-2121 453563024231 1 4 Cover Screw (Black) 5041-4274 453563099961 3 5 Cable Assembly (Toco) 2.5m / 8.2ft M1355-61652 453563277161 2 6 Case Bottom (Toco, Brown) M1355-44551 453563277101 1 7 Strain Gauge 0960-0810 453563045191 1 Toco Transducer: Cable Connections Connector Transducer Pin Cable Color Pin 2 Yellow 2 3 Green 4 4 Blue 1 8 Orange 3 1+6 Bridged by Code Resistor 4.53 kω - 4

Dismantling the US/MECG Combi Transducer (Brown) Follow the instructions for the Ultrasound transducer described on page 1. Parts for the US/MECG Combi Transducer 3 1 2 Ultrasound Transducer: Parts List Item Description Part Number 12NC Quantity 1 Combi Transducer (US) M1358-60001 453563277311 1 2 Combi Transducer (MECG) M1358-60002 453563277321 1 3 Combi Cable Assy (US) M1358-61681 453563277341 1 Ultrasound Transducer: Cable Connections Pin Cable Color 4 Brown 8 Red 5 Green 2 Yellow 6 Red 1+6 Bridged by Code Resistor 665 Ω 1 Shield 5

Testing After Repair After repair, carry out the test and inspection activities described in this section. NOTE Test the US Combi transducer as described for the Ultrasound transducer on page 7. See Testing MECG Transducers on page 9 for how to test the MECG Combi transducer. Performance Assurance Tests Self Test Carry out the monitor self-test and the parameter test. The monitor automatically performs a basic-level self test when you switch it on. There are two possible types of error that you might see. A fatal error prevents the monitor from functioning. A non-fatal error allows you to continue to work but warns you of a problem that must be resolved swiftly. If a non-fatal error occurs (for example, if the batteries are low): An error message is displayed for ten seconds. Err xxx, time and date are printed on the paper after ten seconds, and then every ten minutes. ( xxx is the number of the error message.) Switch the monitor off and then on. If the error occurs again, try to solve the problem or, if you cannot, contact your Philips Service Engineer or Response Center. (If the recorder is not on when the monitor is switched on, Err xxx time and date are printed when it is switched on subsequently.) If a fatal error occurs (for example, if a board is defective): An error message is displayed for ten seconds After ten seconds, the monitor tries to restart. If the error occurs again contact your Philips Service Engineer or Response Center. Parameter Test The parameter test tests the processing of the signal to and from the transducer, but not the transducers themselves. To perform the parameter test: Switch on the monitor and the recorder. Connect the transducers for the channels to be tested to the correct sockets. Press. The monitor produces an artificial signal for each transducer connected and the signals are processed. You will see that the test signal is displayed and the mode symbols light. You will also hear a sound specific for the type of transducer connected. The following table shows the values recorded when the different transducers are correctly connected. Ensure that the recorder is switched on. If an error occurs, it is displayed for ten seconds and then Err is printed by the recorder together with the time/date annotation. After this time, Err is printed every ten minutes together with the time/date annotation. 6

Parameter Test Signal US (Cardio 1/Combi) using M1356A: US (Cardio 2) using M1356A: TOCO using M1355A: US/MECG (Cardio 1/Combi) using M1358A: Monitor Response 190 is displayed and printed. Signal quality indicator is green. Fetal heartbeat is heard from loudspeaker. 170 is displayed and printed. Signal quality indicator is green. Fetal heartbeat is heard from loudspeaker. A signal alternating between 10 and 60 (for periods of 2 secs) for as long as the key is pressed is displayed and printed. 190 is displayed. 190 and 120 are printed. Signal quality indicator is green. MECG is on. Fetal and maternal heartbeats are heard form the loudspeaker. Testing Ultrasound Transducers Carry out a visual and electrical check of the repaired transducers. CAUTION Use of ultrasound gel that is not approved by Philips may reduce signal quality and may damage the transducer. This type of damage is not covered by warranty. Visual Check Ensure there are no cracks in the transducer dome, that the cable is not cracked or broken, and that there are no cracks on the connector plug. Electrical Check 1 Connect the transducer to either the Cardio 1/Combi or Cardio 2 socket. (Both the connector and socket are red, and keyed so that they mate in only one position.) Ensure that: The signal quality indicator is red The FHR numerical display is blank When the recorder is switched on, the date, time, mode and paper speed are printed on the recorder trace. 2 Turn the loudspeaker volume up to an audible level. 3 The ultrasound transducer contains seven piezoelectric crystals. Basic functioning of each can be verified by holding a flat bottomed pencil or similar above each crystal and moving it up and down as shown. 7

Figure 1 Testing an Ultrasound Transducer using a Pen A sound should be heard for each crystal tested. The pen should be held 2 to 3 cm from the transducer surface when the test is carried out. Figure 2 Position of Crystals in an Ultrasound Transducer 4 A sound should also be heard when the transducer is moved back and forth over a solid surface, or the hand as shown below. Figure 3 Checking an Ultrasound Transducer If the tests are not as outlined above, repeat the tests with another transducer. 8

Testing TOCO Transducers Carry out a visual and electrical check of the repaired transducers. Visual Check Ensure that the transducer housing is sound, that the cable is not cracked or broken, and that there are no cracks on the connector plug. Electrical Check 1 Connect the Toco transducer to the Toco socket. (Both the connector and socket are brown, and keyed so that they mate in only one position.) Ensure that: the Toco display shows 20. when the recorder is switched on, the date, time, mode and paper speed are printed on the recorder trace. 2 Press the transducer button firmly and look for a deflection on the display and recorder. The external Toco display maximum is 100 units. 3 Lay the transducer face up on a flat surface for a few seconds. 4 Press the Toco Baseline Key to re-adjust the Toco display to 20. 5 Turn the transducer over so that the button is face down on the flat surface. Hold the cable at a point 25 cm from the transducer and ensure that the transducer touches the flat surface only with the button and that the transducer is parallel to the flat surface. 6 The Toco display should read between 40 to 50 units. Toco display = 20 Toco display = 40-50 paitoc30.hpg NOTE The illustration does not show the cable. The appearance of the transducer may differ from the illustration. If the test results are not as outlined above, repeat the test with another transducer. The external Toco recorder display can be between 0 and 127 units. If the test fails, repeat using another transducer. After the test, you must zero the system by pressing the Toco Baseline Key. Testing MECG Transducers 1 Connect the two electrode cables (M1531B) from the maternal ECG electrodes (40493D) to the two press fit type connectors on the MECG Combi transducer. 2 EITHER a. Attach electrodes to the M1531B electrode cables, and apply the electrodes to the skin (for example on the wrists). OR b. Attach the M1531B electrode cables to a patient simulator. 9

NOTE We do not recommend the use of a specific patient simulator. The use of a patient simulator does not allow checking the specification of the ECG-Functionality; it allows only a check of general function. Disposal Result: You should see MECG values displayed on the maternal LCD display or annotated on the recorder trace. If the test results are not as outlined above, repeat the test with other M1531B electrode cables and/or another MECG Combi transducer. Dispose of the equipment in accordance with your country s laws. WARNING To avoid contaminating or infecting personnel, the service environment or other equipment, make sure the equipment has been appropriately disinfected and decontaminated before disposal at the end of its useful life. 10

11

Philips Medical Systems is part of Royal Philips Electronics INTERESTED? Would you like to know more about our imaginative products? Please do not hesitate to contact us. We would be happy to provide specific information about our products and services, or put you on our mailing list for news about new product developments, upcoming events or for our clinical journal, MedicaMundi. We would be glad to hear from you. On the web Contact us through our web site: www.medical.philips.com Via e-mail Our e-mail for all remarks and requests is: medical@philips.com By fax We can be reached at the following fax number: +31 40 27 64 887 By postal service Please write to us at the following address: Philips Medical Systems Global Information Center I.B.R.S. / C.C.R.I. Numéro 11088 5600 VC Eindhoven Pays-Bas / The Netherlands (no stamp required) United States: Philips Medical Systems Cardiac and Monitoring Systems 3000 Minuteman Road Andover, MA 01810 (800) 934-7372 Canada: Philips Medical Systems Canada 281 Hillmount Road Markham, ON L6C 2S3 (800) 291-6743 Europe, Middle East and Africa: Philips Medizin Systeme Böblingen GmbH Cardiac and Monitoring Systems Hewlett-Packard Str. 2 71034 Böblingen Germany Fax: (+49) 7031 463 1552 Latin America Headquarters: Philips Medical Systems 1550 Sawgrass Corporate Parkway #300 Sunrise, Fl 33323 Tel: (954) 835-2600 Fax: (954) 835-2626 Asia Pacific Headquarters: Philips Medical Systems 30/F Hopewell Centre 17 Kennedy Road Wanchai Hong Kong Tel: (852) 2821 5888 Fax: (852) 2527 6727 1990-2004 Koninklijke Philips Electronics N.V. All Rights Reserved. 0366 S