CT SCAN PROTOCOL. Shoulder

Similar documents
TORNIER BLUEPRINT. 3D Planning + PSI SCAN PROTOCOL

TORNIER. scan protocol V2.1. Tornier Upper Extremities

CT Scanning Protocol For V2R Guided Surgery Solutions

Cranial idtm. Cranial restoration. id SolutionsTM. Individually designed. Personalized care.

HI-Res Extremity Sensation 16

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers

Personalized Solutions. MRI Protocol for PSI and Signature Guides

X-Ray & CT Physics / Clinical CT

B. CT protocols for the spine

KNEE ALIGNMENT SYSTEM (KAS) MRI Protocol

PROPHECY. Preoperative Navigation Guides ANKLE CT SCAN PROTOCOL

Shoulder Position: Supine arm in the neutral position. Collateral arm above head Indication: fracture humerus, fracture scapula

Computed tomography Acceptance testing and dose measurements

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Morita 3D Accuitomo Scanning Protocol

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

NewTom 3G & VG Scanning Protocol

SPECIFIC PRINCIPLES FOR DOSE REDUCTION IN HEAD CT IMAGING. Rajiv Gupta, MD, PhD Neuroradiology, Massachusetts General Hospital Harvard Medical School

TrueSight Personalized Planning & Targeting System. Operative technique

True Low Dose. Exact time to display image on screen may vary upon computer and network configuration.

CT Imaging at the Point-of-Care

Dental Cone Beam CT. What is Dental Cone Beam CT?

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

CS 9300 Family. The power of flexibility

Metal Artifact Reduction by Dual Energy CT

Cardiac CTA Prospective Gating Broad Beam

Bone Densitometry Equipment Operator

MSK CT Extremities: Positioning and Reformations

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Full ultrasound breast volumes. Faster scans. Streamlined workflow. ACUSON S2000 Automated Breast Volume Scanner. Answers for life.

Breast Tomosynthesis. What is breast tomosynthesis?

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

Clinical Image Gallery Next Generation Volume 1

EWOO Picasso Trio Scanning Protocol

The Quality Leader in 3D Cone Beam CT

Translating Protocols Across Patient Size: Babies to Bariatric

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Bringing the power of imaging to your patient. Portable full body 32-slice CT scanner

Customizing Contrast Injection for Body MDCT: Algorithmic Approach

ProMax 3D Scanning Protocol

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

Upper Limb Imaging Requirements

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Patient preparation and coronary CTA techniques. Gregory Kicska, M.D. Ph.D. University of Washington, Thoracic Imaging

The Computed Tomography Examination

FDG-18 PET/CT - radiation dose and dose-reduction strategy

Pediatric chest HRCT using the idose 4 Hybrid Iterative Reconstruction Algorithm: Which idose level to choose?

The value of weight-bearing functional CT scans

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

COMPUTED TOMOGRAPHY COURSE

Head to new heights with your imaging SCANORA 3D

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

A more accurate method to estimate patient dose during body CT examinations with tube current modulation

2

Combined Anatomical and Functional Imaging with Revolution * CT

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care.

Whole Body CT Protocol Update 2018

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Managing Radiation Risk in Pediatric CT Imaging

DIAGNOSTIC IMAGING. OPTIMIZED.

Hands PA; Obl. Lat.; Norgaard s Thumb AP; Lat. PA. PA; Lat.: Obls.; Elongated PA with ulnar deviation

Institutional review board approval was obtained prior to the start of this study.

X-ray (Radiography) - Bone

Image Registration for Radiation Therapy Applications: Part 2: In-room Volumetric Imaging

Blair Radiology Exam Examination Packet

Typical PET Image. Elevated uptake of FDG (related to metabolism) Lung cancer example: But where exactly is it located?

For a legacy that lives large: Think Big. Philips CT Big Bore. Computed tomography

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Image Diagnostic Technology Ltd IDT Ireland, 15 Market Street, Kinsale, Co. Cork, Ireland P17 XN65 Tel: Mob: IRL:

Cone Beam CT-arthrography of the wrist: high resolution images at low radiation dose

VistaVox S 3D from Dürr Dental

Low Dose Era in Cardiac CT

Scanora 3D Scanning Protocol

Gender differences in CT calcium scoring: A phantom study

BLUEPRINT. 3D Planning + PSI SURGIC AL TECHNIQUE V 2.1 T I TA N I U M

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

Radiographic Positioning for Dogs

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

ABAS Atlas-based Autosegmentation

BLUEPRINT. 3D Planning + PSI SURGIC AL TECHNIQUE

Metal Artefact Reduction in CT

Ask EuroSafe Imaging. Tips & Tricks. CT Working Group. Optimization of scan length to reduce CT radiation dose

Detectability of unruptured intracranial aneurysms on thinslice non-contrast-enhanced CT

MRI to fit your planning. Philips Panorama HFO Oncology Configuration

PATIENT SPECIFIC PLATES FOR MANDIBLE Custom made plates for trauma and reconstruction

Ultrasound - Musculoskeletal

GE Healthcare. Rad Rx. White Paper

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

TRUMATCH PERSONALIZED SOLUTIONS with the SIGMA High Performance Instruments

D. The following factors are of extreme importance to the eventual success of the procedure.

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

American College of Radiology CT Accreditation Program. Testing Instructions

Introduction of FIREFLY Technology

Transcription:

CT SCAN PROTOCOL Shoulder

Purpose and Summary CT images made with this protocol are used to provide the orthopedic surgeon with a detailed 3D anatomical reconstruction of the patient s scapula and proximal humerus. This virtual 3D model is intended for the creation of a virtual patient specific pre-surgical plan, design of patient-specific instrumentation and/or a patient-specific implant for shoulder replacement surgery. This protocol consists of a localizer and a detailed axial scan of the shoulder. A clear visualization of bone structures is needed. Image quality should reach a level required for radiological evaluations of the bone. Deviations from this protocol may result in unusable images for our medical device. When using this protocol, apply dose reduction techniques and optimize scan parameters within the provided ranges to limit the dose delivered to the patient. NOTE CT scan quality can directly affect the design of instruments and implants. Please ensure that all protocol steps are followed for optimum scan quality. If there is a recent CT scan (< 4 months old) available, check whether this scan matches the requirements outlined below to avoid an unnecessary scan. Read the following instructions carefully before scanning. Please contact Materialise s support team if you require further clarification. For questions, please contact Materialise Customer Services: US +1-734-259-6669 Europe +32-1698-0488 Asia-Pacific +603-7724-1418 This CT scan protocol has been designed to provide scan centers personnel with easy-to-use instructions to obtain correct images with optimal quality that will be further used in the process of surgical planning and design of patient specific surgical instruments and implants that support on bone. Materialise cannot be held liable for other possible subsequent uses (i.e. diagnostic uses.) Note that it is highly recommended to perform the surgery within 6 months of the CT scan date to ensure anatomic changes are minimized. If the patient s anatomy has changed significantly since the time of the CT scan, the patient specific models, instruments, and implants should not be used, even if the time period of 6 months has not expired. 2 2018, Materialise NV, All rights reserved. L-100148

SCAN PREPARATIONS Patient Preparation - Discuss the procedure with the patient and instruct not to move during scanning. Patient movement will prevent the accurate production of the patient-specific model. - Remove any non-fixed metal prosthesis, jewelry or zippers that might interfere with the region to be scanned. - Position the patient head first, supine. Arms at sides of the body and with the shoulder in neutral rotation. Arms should be extended with palms facing the body sides, thumbs pointing to the front of the body. Center the patient in the isocenter of the gantry. - Indicate the body side scanned (right or left), for example in the Series description. Patient positioning in the case of metal implants: - If an implant is present in the shoulder (surgical side or contralateral), raise the arm above the head to mitigate artifacts. - If this metal implant includes a glenoid component on the surgical side, do not scan the patient. Please contact Materialise Customer Services (except for patient specific implants selected for 2-stage approach). Scan Requirements - Use true axial slices. Oblique slices are not accepted. No reformatted images. - Gantry tilt is not accepted. - Region of interest: Include the complete scapula and proximal humerus until the distal end of the scapula. Only the bony regions are of interest. Capturing the surrounding soft tissue is not necessary. - Scan range: Scan the entire scapula and proximal humerus on the surgical side. From just above the acromioclavicular joint to just below the inferior angle of the scapula. - If bilateral shoulders are ordered, reconstruct them separately. Acquire at 500 mm FOV and reconstruct individual shoulders at 250 mm FOV. 3 2018, Materialise NV, All rights reserved. L-100148

SCAN PARAMETERS Scanner type Multi-detector row CT with number of detector rows 16 1 Scan mode Helical kvp 100 140 (automatic voltage selection, if available) ma(s) Automatic tube current modulation Pitch 1 Detector Configuration Slice Thickness Slice Increment Single collimation slice thickness 1.25 mm or smaller (Do not acquire a thicker slice and retrospectively reconstruct to noted thickness) 50% overlap Matrix 512 x 512 Field of View Reconstruction algorithm 250 mm or smaller (use smallest FOV that includes the complete bony anatomy of interest) Use the following reconstruction algorithms and provide axial images: 1) Use a standard or soft tissue algorithm without edge enhancement (No bone algorithm). Always provide this reconstruction. 2) When scanning for patient-specific implants provide an additional reconstruction with a bone algorithm. 3) If metal is present, provide additional reconstruction(s) with metal artifact reduction applied if available. Always also provide a reconstruction without metal artifact reduction applied. Reconstructions should be obtained from one single acquisition. HU Scale If metal implants are present, use a HU scale of 16-bit. 1 Scanners with > 64 detector rows are sometimes referred to as Volume-CT and can be used in helical scan mode. DO NOT use cone-beam CT. 4 2018, Materialise NV, All rights reserved. L-100148

Scan parameter optimization Scan parameters can be optimized within the given ranges according to best practices in CT imaging. Adapt the scan parameters taking image quality, patient-specific factors, presence of metal, scanner specific factors, and dose considerations into account. 2 IN THE PRESENCE OF METAL - Check whether strategies of optimizing scan parameters to reduce metal artifacts seem beneficial, such as using thin slice collimation and reconstructing to slices of 1.25 mm, lowering pitch, and increasing kvp. - Use a Metal Artifact Reducing algorithm/ filter, if available. Submit this along with the standard scan. - Increase the HU scan range by using a 16 bit or extended CT scale, if available. - If metal is present in the humerus, raise arm above head to reduce amount of artifact in the field of view. WITH REGARD TO DOSE OPTIMIZATION - Adjust parameters depending on patient body habitus (e.g. kvp, mas). - Dose information displayed at your scanner (such as CTDI vol) can be used to optimize scan parameters. - Apply dose reduction techniques such as automatic tube current modulation and automatic voltage selection whenever possible and applicable (e.g. only apply automatic tube current modulation when your system can apply it correctly in the presence of metal in the scan region). - For patients of standard body size without metal implants it is often possible to use a low-dose protocol for bone imaging and 3D applications. - Tip: On some scanners, prospective selection of thin reconstructed slice thickness (e.g. 1mm) can lead to higher doses. Consider a retrospective reconstruction from thin acquisitions according to scan protocol parameters (Image Type needs to be ORIGINAL). - Consult www.imagewisely.org and www.fda.gov/radiation-emittingproducts/ RadiationEmittingProductsandProcedures/MedicalImaging/MedicalX-Rays/ucm115317.htm for additional information about radiation safety. 2 These are recommendations. Please also take your institution s guidelines into account when optimizing scan parameters. In case of questions, contact Materialise Customer Service. 5 2018, Materialise NV, All rights reserved. L-100148

DATA TRANSMISSION File format - Submit DICOM format only. - Uncompressed DICOM data is necessary for processing. Lossy and other forms of compression are NOT allowed (ISO_10918_1, ISO_14495_1, ISO_15444_1 or ISO_13818_1). - The scanner should be set to DICOM format raw image, with no compression. If loading from PACS, import and export the scan as DICOM files with the uncompressed option. - The complete data set of primary DICOM images must be provided. Data anonymization - Do not erase patient name and ID. Patient date of birth and name need to be included. - Ensure necessary rights are obtained for transfer of data to Materialise. - Data will be anonymized by Materialise on receipt of the data, after cross-check with prescription of the surgeon to ensure the images of the right patient are provided. Transfer scan data to Materialise Instructions for image submission can be found in the SurgiCase Online User Guide for Scan Centers Only send the following images: - The requested CT images at the given parameters - The accompanying scout view - An accompanying 3D reconstruction (if available) in case of metal implants. IMPORTANT Your site should keep an archive (PACS) copy of the CT exams, in uncompressed DICOM format and the original scanning parameters. Legal Disclaimer Materialise and the Materialise logo are trademarks of Materialise NV. This brochure is provided by Materialise and may be used for informational purposes only. Materialise uses reasonable efforts to include accurate and current information at the date of publication of this brochure. Materialise makes no warranties or representations of any kind as to its accuracy, currency or completeness. Materialise, nor any party involved in creating, producing or delivering this brochure shall be liable for any damages, including without limitation, direct, incidental, consequential, indirect or punitive damages, arising out of access to, use of or inability to use this brochure, or any errors or omissions in the content thereof. Any legal action or proceeding related to this brochure shall be brought exclusively to the Courts of Brussels (Dutch speaking division). 6 2018, Materialise NV, All rights reserved. L-100148