Ultrasound Transducer Selection in Clinical Imaging Practice

Similar documents
Lesson 07: Ultrasound Transducers. This lesson contains 73 slides plus 16 multiple-choice questions.

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

Physical Principles of Ultrasound

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS

Introduction to Biomedical Imaging

ACUSON X500 Ultrasound System. Transducers

Preamble (disclaimer)

GE Healthcare. LOGIQ E9 XDclear 2.0. Complete Ultrasound HEAD TO TOE OBESE TO THIN NEONATE TO GERIATRIC

Dr Emma Chung. Safety first - Physical principles for excellent imaging

SAMSUNG. Mobile excellence. Excellence on the Move ULTRASOUND H60. Ultrasound system HM70A with Plus

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

GE Healthcare. Transducer Guide. VIVID Series

GE Healthcare. Logiq P5 Premium

GE Healthcare. LOGIQ E9 XDclear 2.0. Complete Ultrasound HEAD TO TOE OBESE TO THIN NEONATE TO GERIATRIC

GE Healthcare. Transducer Guide. LOGIQ Series

Terminology Tissue Appearance

ACUSON X300 Ultrasound System, Premium Edition (PE)

The complete, portable patient care solution

Vivid iq Cardiovascular Ultrasound

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

ACUSON X300 Ultrasound System, Premium Edition (PE)

Delivering Top Performance. Outstanding Value. Exceptional Reliability. ACUSON X150 Ultrasound System. Answers for life.

High performance ultrasound

On the feasibility of speckle reduction in echocardiography using strain compounding

ACUSON X300 Ultrasound System, Premium Edition. Transducers Release

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

ACUSON Antares Ultrasound System. Transducers.

Ultrasound Principles cycle Frequency Wavelength Period Velocity

Daily inspiration. Ultrasound system HS70A SAMSUNG MEDISON CO., LTD. Scan code or visit to learn more

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary

CONTENTS. Test Number cpd Tanya Reynolds (Nat. Dip. Diag. Rad., B. Tech. Diag. Rad., B. Tech. Ultrasound)

CHARACTERIZATION OF ANNULAR ARRAY TRANSDUCER

What is Ultrasound? What is Ultrasound? B A. Basic Principles of Ultrasound. Basic Principles of Ultrasound. Basic Principles of Ultrasound

ACUSON Sequoia C512 Echocardiography System. Transducers.

Focus on your needs. Ultrasound system HS60 SAMSUNG MEDISON CO., LTD. CT-HS60 V1.0-GI-FT EN

Linear Ultrasonic Wave Propagation in Biological Tissues

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Point-of-Care Ultrasound: An Introduction

ACUSON S2000 Ultrasound System. Transducers Release 1.0.

Pipeline Technology Conference 2007

The Value. Ultrasound System. Professional Concept for. Women s. Health

Transducers. ACUSON SC2000 Ultrasound System. Precision at the Speed of Life Release 5.0

ACUSON Sequoia C512 Echocardiography System. Transducers.

Confident Diagnosis, Confident Decisions

Transducers. ACUSON SC2000 Ultrasound System. Release 4.0. Answers for life.

The Evolution and Benefits of Phased Array Technology for the Every Day Inspector

Ultrasound Physics & Terminology

Basic Ultrasound Physics Board Review Questions

ACUSON X150 Ultrasound System. Transducer Options Release 1.0

Vision of a new dimension

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426)

Pushing the boundaries

Transducers. ACUSON NX3 Elite Ultrasound System

Versatile, affordable ultrasound. Philips HD6 ultrasound system. The print quality of this copy is not an accurate representation of the original.

3/20/2017. Disclosures. Ultrasound Fundamentals. Ultrasound Fundamentals. Bone Anatomy. Tissue Characteristics

Design Your Performance

ACUSON S2000 Ultrasound System. Transducers Release 1.6.

Ultrasound Physics & Doppler

High resolution ultrasound scanner for skin imaging

ACUSON P500 Ultrasound System

Underwater Acoustic Measurements in Megahertz Frequency Range.

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

What is Ultrasound? Resolution Image production Attenuation Imaging modes Ultrasound artifacts... 7

Supplement (videos)

Confident Diagnosis, Confident Decisions

ACUSON Antares Ultrasound System. Transducers.

FOR IMMEDIATE RELEASE

Application of ultrasonic phased array in acoustic logging

GE Healthcare. Accomplish more. Voluson 730

ACUSON P500. Ultrasound Anytime, Anywhere. ACUSON P500. siemens.com/acusonp500. siemens.com/acusonp500 1

Introduction to Ultrasound Guided Region Anesthesia

GE Healthcare. Vivid i Cardiovascular ultrasound system

Flaw Assessment Using Shear wave Phased array Ultrasonic Transducer

Background & Indications Probe Selection

Confident Diagnosis, Confident Decisions

Reliable versatility. Philips HD5 ultrasound system

Hitachi Aloka Medical manufactured one of the world s first diagnostic ultrasound platforms, and today this imaging modality has become the first

WELCOME! Introduction to Bedside Ultrasound

Augmentation of the In Vivo Elastic Properties Measurement System to include Bulk Properties

Ultrasonic arrays are now widely used in underwater sonar

Basic Physics of Ultrasound and Knobology

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

ACUSON Sequoia C256 Echocardiography System. Transducers.

Manual Ultrasonic Inspection of Thin Metal Welds

INTRODUCTION. Getting the best scan. Choosing a probe. Choosing the frequency

Accuracy Versatility Mobility

Pushing the boundaries

Vivid S6. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Mobile excellence SAMSUNG. ULTRASOUND H60 Ultrasound system HM70A with Plus. Excellence on the Move SAMSUNG MEDISON CO., LTD.

Concepts of Imaging and Knobology

Surgical Probes अౠ.6,7,1 / ?K

Ultrasound in Anesthesia: Applying Scientific Principles to Clinical Practice

Full ultrasound breast volumes. Faster scans. Streamlined workflow. ACUSON S2000 Automated Breast Volume Scanner. Answers for life.

Optimization of Phased-Array Transducers for Ultrasonic Inspection in Composite Materials Using Sliding Probes

Confident and Conclusive Diagnosis with Flexible Solutions

Transcription:

SPECIAL COMMUNICATION Ultrasound Transducer Selection in Clinical Imaging Practice Thomas L. Szabo, PhD, Peter A. Lewin, MSc, PhD Invited paper Article includes CME test Many types of medical ultrasound transducers are used in clinical practice. They operate at different center frequencies, have different physical dimensions, footprints, and shapes, and provide different image formats. However, little information is available about which transducers are most appropriate for a given application, and the purpose of this article is to address this deficiency. Specifically, the relationship between the transducer, imaging format, and clinical applications is discussed, and systematic selection criteria that allow matching of transducers to specific clinical needs are presented. These criteria include access to and coverage of the region of interest, maximum scan depth, and coverage of essential diagnostic modes required to optimize a patient s diagnosis. Three comprehensive figures organize and summarize the imaging planes, scanning modes, and types of diagnostic transducers to facilitate their selection in clinical diagnosis. Key Words ultrasound imaging; ultrasound transducer arrays; ultrasound transducers Received July 11, 2012, from the Department of Biomedical Engineering, Boston University, Boston, Massachusetts USA (T.L.S.); and School of Biomedical Engineering, Science, and Health Systems, Drexel University, Philadelphia, Pennsylvania USA (P.A.L.). Revision requested July 30, 2012. Revised manuscript accepted for publication August 17, 2012. We thank GE Healthcare (Milwaukee, WI), Philips Healthcare (Bothell, WA), and Boston Scientific (Natick, MA) for the use of their images to illustrate general transducer types. These images are not intended to endorse any manufacturer. We also thank Kai E. Thomenius, PhD (GE Global Research, Niskayuna, NY), for collaboration and Sverre Holm, MSc, PhD (Department of Informatics, University of Oslo, Oslo, Norway), for permission to use Figure 2. Address correspondence to Thomas L. Szabo, PhD, Department of Biomedical Engineering, Boston University, 44 Cummington St, Boston, MA 02215 USA. E- mail: tlszabo@bu.edu Abbreviations 4D, 4-dimensional; FOV, field of view; 1D, 1-dimensional; 3D, 3-dimensional; 2D, 2- dimensional F or more than 50 years, many kinds of transducers have evolved for medical ultrasound imaging. Transducers operate at different center frequencies, have different physical dimensions, footprints, and shapes, and provide different image formats. For example, the phased array has a small (typically 20 15 mm) footprint (or contact area) to fit between the ribs and has the capability to produce sector images with wide coverage and depth at high (>100 frames/s) frame rates. However, little information is available about why certain transducers are most appropriate for a given clinical application, 1 and the purpose of this article is to address this deficiency. Specifically, the relationship between the transducer, imaging format, and clinical applications is discussed. Systematic selection criteria that allow matching of transducers to specific clinical needs are presented in a new framework that both explains why certain types of transducers are used in specific clinical applications and also provides a basis for selecting transducers for new applications. The criteria include access to and coverage of the region of interest, maximum scan depth and image extent, and coverage of essential diagnostic modes required to optimize a patient s diagnosis. For completeness, single-element transducers, primarily used in intraluminal or catheter applications, are also included in the considerations presented below. Whenever appropriate, a historical perspective for transducer selection is pointed out, but, in general, new trends are emphasized. 2013 by the American Institute of Ultrasound in Medicine J Ultrasound Med 2013; 32:573 582 0278-4297 www.aium.org

Scanning for Images It is well known that piezoelectric transducers can, when placed on or in the body, transmit ultrasound pulses and receive echoes from within the tissue and organs. To produce clinically useful images, an additional ingredient is needed, namely, scanning. 1 Usually the acoustic beam generated by an individual transducer is moved in a prescribed direction either electronically or mechanically to produce a series of pulse echo lines that define an image plane. For orientation, Figure 1A shows a reference system useful for explaining linear scanning in the xz plane. For 2-dimensional (2D) scanning, the image plane is the xz plane. A simple scanning method is to move the acoustic beam incrementally, a little bit (defined as Δx) at a time, along the x-axis. At each position, a pulse echo line is created, and the set of lines is interpolated to produce a rectangularly shaped image in which lateral translation is depicted from line a to line b. An alternative approach to translation is to shift the angle of the acoustic beam in an arc, in small angles, Δθ, at a time, to define an image in the xz plane, as illustrated by Figure 1B. Here the rotation from line c is shown to line d. Note that each line represents an acoustic beam, as illustrated more graphically in Figure 2A. As before, after a complete set of lines is acquired, it is interpolated into a sector-shaped image. A variant of linear shifting is the curved geometry shown in Figure 1C. In this case, the array is situated on a curve formed by a radius of curvature (R), and the line increment, Δs, is along the curved surface rather than a straight line. What is interesting about this geometry is that this increment along a curve going from line e to line f is the equivalent of an angular shift through the relation Δs = R Δθ. As a consequence of this type of scanning along an arc, the lines fan out radially. Similarly, scanning can be defined for the yz plane. In this case, the translation scanning is along the y-axis in increments of Δy, and angular scanning is done in increments of Δθ in the yz plane. To achieve 3-dimensional (3D) scanning, or scanning anywhere in the positive half space defined by the positive x-, y-, and z-axes, scanning in both xz and yz planes can be combined to form a pyramidalshaped volume scan, as illustrated by Figure 2B. that can be controlled in groups or clusters to create pulse echo lines. 2 For a linear array, groups of in-line elements are incrementally switched on and off, effectively sliding an active group of elements laterally by Δx at a time to create the individual pulse echo lines that compose the image plane. Pulse echo lines are interpolated to form the resulting rectangular image format and corresponding transducer Figure 1. A, Cartesian reference system for linear scanning in the xz plane from acoustic line a to line b by increment Δx. B, Coordinate system for sector scans in the xz plane where an acoustic line c is moved to position d by an angular increment Δθ. C, Coordinate system for a curved linear or convex array for sector scans in the xz plane where an acoustic line e is moved to position f by an angular increment Δθ. A B C Imaging Formats Even though early (single-element) transducers were mechanically scanned in 2D planes for ultrasound imaging, by the early 1980s, transducer arrays were routinely used for scanning. 2 An ultrasound array consists of an ensemble of arranged individual single transducers, or elements 574 J Ultrasound Med 2013; 32:573 582

shape, which are shown in format 1 of Figure 3 and the associated linear array transducer of Figure 4A, respectively. Focusing can be accomplished either mechanically or electronically. For the linear format of Figure 4A, electronic focusing is achieved for each scanned image line by controlling the delay time at which voltage excitation of the individual elements is delivered to a group of active elements. In the elevation or yz plane (ie, the plane perpendicular to the image plane, often referred to as the slice thickness), fixed focusing, using a mechanical lens, is achieved. Figure 2. A, Sketch of a focused acoustic beam corresponding to an acoustic line for a 2D image from a 1D array. B, Sketch of an acoustic beam corresponding to an acoustic line electronically focused to a point in the xz and yz planes in a 3D image from a 2D or matrix array (courtesy of Sverre Holm, MSc, PhD). A B To somewhat alleviate the fixed focus limitation, some imaging system manufacturers offer arrays with multiple rows in the elevation direction. However, fully controlled focusing in the elevation plane requires 2D array transducers, which are capable of providing not only improved elevation focusing but also 3D and 4-dimensional (4D) images. In Figure 2B, simultaneous electronic focusing of a 2D array is depicted for both the elevation and azimuth planes, xz and yz. To facilitate matching an image format with a specific ultrasound transducer, Figures 3 and 4 can be used. For example, formats 1 and 4 of Figure 3 are associated with the linear array transducer type A of Figure 4. For an example of sector scanning or angular scanning, the image format is in the shape of a pie slice, as depicted by format 2 of Figure 3 and the corresponding phased array transducer, shown in Figure 4B. Typecasting Transducers With the help of Figures 3 and 4, it is possible to create a systematic way of organizing imaging formats and relating them to transducer types with the realization that scanning types, modes, and planes need to be taken into account. To categorize formats and transducers, abbreviations can be combined to describe specific image-transducer relationships. Specifically, to denote the scanning type, M signifies mechanical scanning; E, electronic scanning; and F, (fixed), no scanning. The scanning direction is either linear (L) along the x-axis or angular (<), or curved (C), or a combination (a more detailed description of these is given below). The scanning plane can also be specified. For 2D imaging, the xz plane is used. Figure 3. Types of image formats: 1, 2D rectangular; 2, 2D sector; 3, 2D convex or curved where the maximum angular extent is the FOV; 4, 2D trapezoidal; 5, 3D parallelepiped; 6, 3D fan or broom shape; 7, 3D truncated prism; 8, 2D donut; and 9, 3D tube. J Ultrasound Med 2013; 32:573 582 575

According to the above description, each transducer can be associated with scanning types and planes. For example, the linear array, L from Figure 4A is associated with electronic linear scanning, E in the xz plane and fixed focusing, and F in the yz plane; therefore, the resulting designations are abbreviated as ELxz and Fyz, and the associated formats are 1 and 4 from Figure 3. The combined representation is the first example shown as Figure 4A. The trapezoidal format, denoted as 4 in Figure 2, can be thought of as a rectangular format with two partial sectors on each end, for the linear array of Figure 4A. Similarly, the phased array of Figure 4B is linked with sector format 2 of Figure 3 and the same planes as the previous examples. Other transducers and formats are also summarized in Figures 3 and 4. A diverse sampling of transducer types is shown in Figure 5. The curved or convex array (Figure 4C) is similar to a linear array except that the elements are on a curved rather than a flat surface, as described by the scanning method C of Figure 1C, resulting in format 3 of Figure 3. This format, similar in shape to a sector or pie slice with a bite taken out of its top, often is described by a field of view (FOV) angle specifying its lateral angular extent. This example employs electronic linear scanning, E, in the xz plane and fixed focusing, F, in the yz plane; therefore, the resulting designations are abbreviated as ECxz and Fyz, and the associated format is 3, as illustrated by Figure 4B. As the importance of 3D imaging is steadily growing, it is appropriate to discuss it in more detail. For 3D imaging, a volume is scanned instead of a plane, as demonstrated by the outer volume depicted in Figure 2B. For a 2D or matrix array (Figure 4F), scanning can be electronic and typically angular in both directions so that the scanned volume is pyramidal in shape (denoted as format 7 of Figure 3). In this case, electronic focusing is achieved in both planes with angular scanning so the appropriate designations and format are E<xz, E<yz, and format 7, as represented by Figure 4E. Alternatively, to achieve cost-effective 3D imaging, linear or convex arrays can be scanned mechanically Figure 4. Transducer types and associated scanning and image formats: A, linear array; B, phased array; C, convex or curved array; D, endo array; E, 2D or matrix array; F, mechanically scanned linear array; G, mechanically scanned convex array; and H, intravascular ultrasound transducer (courtesy of Boston Scientific, GE Healthcare, and Philips Healthcare). 576 J Ultrasound Med 2013; 32:573 582

Figure 5. Family of transducers. Top left quadrant: top three are transesophageal probes; bottom two are endovaginal arrays. Top right quadrant: a microconvex array in the middle with two phased arrays on each side. Bottom right quadrant, left to right: convex array, three linear arrays, a curved linear array, and a phased array. Bottom left quadrant, left to right: two surgical probes and two intraoperative probes (courtesy of Philips Healthcare). around the x-axis in the yz plane. In these cases, the arrays are moved in fluid-filled acoustically transparent chambers. For example, a linear array (usually type A) is rotated about the z-axis to produce a series of image planes (usually format 1 or 4) so that the result is a mechanically scanned transducer of the type F of Figure 4 and the scanned volume format 5 of Figure 3. Similarly, a curved or convex array (usually type C) is rotated about the axis to produce a series of image planes (usually format 3) so that the result is a mechanically scanned transducer of the type G of Figure 4 and the scanned volume format 6 of Figure 3. In addition to electronically controlled movement, these 1-dimensional (1D) arrays (array type A, B, or C) can also be moved mechanically by hand, in freehand 3D imaging, in which the acquired images are organized usually into 3D volumes. Here it is worth noting that image reconstruction for freehand 3D imaging involves either assumptions of regular spacing or additional spatial information for each spatial imaging plane, which may be achieved by use of position sensors. Finally, for completeness, the images produced by single-element transducers, primarily used in intraluminal or catheter applications (such as intravascular ultrasound or intracardiac echo) are also shown in formats 8 and 9 of Figure 3. A single transducer, shown in Figure 4H, can be scanned mechanically to produce 2D or 3D images, as represented by formats 8 and 9. For format 8, a single transducer (Figure 4H) is swept in an angle circumferentially to produce a donut-shaped image. Here it is appropriate to note that there is also an array version of this intravascular ultrasound device. If this mechanical transducer is rotated and translated along the y-axis, a cylindrical volume image, format 9 of Figure 3, results. In conclusion, the transducer types, depicted in Figure 4, can be associated with the image formats of Figure 3 through the use of format and scan designations given below the transducer shapes in Figure 4. Transducer Characteristics for Imaging This section discusses the criteria for linking which properties of ultrasound imaging transducers and their formats need to be identified for various clinical applications. 3 What follows is primarily applicable to clinically used imaging transducers that operate in the frequency range of 1 to 20 MHz. Transducers operating above this frequency are used for special applications such as intravascular imaging (see Figure 4, F and G) or preclinical imaging of small animals but are also included in the discussion wherever possible. Acoustic Windows How well is the type of transducer suited to the acoustic window or location where it makes contact with the body to visualize the organs or tissues of interest? Standard acoustic windows provide an unobstructed view of an organ or region; many, by convention, have specific names, such as transabdominal or parasternal long-axis, so that images can be compared and described consistently. 3 Typical windows are located in or on the following general regions of the body: head, chest, abdomen, pelvis, limbs, vessels, and various orifices of the body. Transducers can be associated with certain regions through Latin prefixes: trans, through or across; intra, into or inside; and endo, within, etc. An example is transthoracic, a category that includes transducers that image through the chest. Transcranial describes a transducer that images the head through the skull. As already mentioned, for the transthoracic window, the phased array would be the most appropriate if the imaging task requires the transducer to be placed between the ribs; it is designed to fit into intercostal spaces and maximize the scanned area (format 2 of Figure 3). For most contact surfaces that are relatively flat and/or slightly deformable (eg, ones used for small parts or vascular imaging), the most general and frequently used transducer type is the linear array designed to make contact with flat surfaces, with the footprint decreasing in size with increasing frequency. Here rectangular and trapezoidal formats J Ultrasound Med 2013; 32:573 582 577

(1 and 4 of Figure 3) provide appropriate viewing areas. With abdominal imaging, to increase the viewing area with minimal increases of the contact area, convex arrays (Figure 4C) produce format 3 (shown in Figure 3) and are designed to make surface contact in deformable soft areas of the body. Specialized Transducers Specialized transducers are designed to operate inside the body. These include transesophageal probes that are phased arrays suitable for manual manipulation within the esophagus, format 2 and transducer type B in Figure 4. A number of other specialty probes have also been developed for interventional or surgical use such as laparoscopic arrays and intracardiac arrays. These probes can be either linear or phased arrays, depending on the application and access windows. Several endo probes: such as endovaginal, endorectal, and endocavity (type D shapes), are functionally like end-fire phased arrays (format 2 and Figure 4B) or convex arrays (format 3 and Figure 4C) at the end of a small-diameter cylindrically shaped handle to fit in orifices and yet maximize the FOVs. Another example is the intravascular ultrasound transducer (Figure 4H), which is inserted in veins to produce an image plane, format 8 or scanned volume, format 9. Resolution and Penetration The selected scan depth allows viewing over the range of interest. Factors involved in imaging capability include the size of the active aperture (occult to the user, but typically a low F number [F#; focal depth/active aperture width] value of 1 2 is used), the transmit focal depth, and time-gain control settings available. Penetration is the minimum scan depth at which electronic noise is visible, despite optimization of available controls (usually at the deepest transmit focal setting and maximum gain), and electronic noise stays at a fixed depth even when the array is moved laterally. Penetration is primarily determined by the center frequency of the transducer: the higher the frequency, the shallower the penetration because the absorption of the ultrasound wave traveling through tissue increases with frequency. A useful first approximation for estimating a depth of penetration (dp) for a given frequency is dp = 60/f cm- MHz, where f is given in megahertz. Thus, one might expect a 6-cm penetration from a 10-MHz center frequency transducer. As noted earlier, the absorption coefficient (acoustic power loss per unit depth) is a function of frequency and varies from tissue to tissue (values for soft tissues range from 0.6 to 1.0 db/cm-mhz 4 ). A more general term describing acoustic loss is the attenuation coefficient, which includes additional losses due to scattering and diffusion and hence is always greater than the absorption coefficient. The attenuation coefficient is highly patient and acoustic path dependent. To optimize image resolution, users and manufacturers have worked on increasing the imaging frequencies for the various examination types. For example, some 30 years ago, people might have imaged the abdomen with a frequency of 2.25 MHz, whereas today the number is more often 3.5 MHz with some obstetric and gynecologic imaging reaching up to 5 MHz. 5 Similarly, the last decade has seen a steady increase in breast imaging reaching the low teens in megahertz. Transducer Properties and Imaging Other criteria to be included in the above-discussed selection process are the transducer efficiency, 2 transducer-system design, system signal-to-noise ratio, and, as already noted, absorption of the tissues being imaged. A major factor is absorption, the compositions and relative positions of different tissue types in the acoustic path. For example, a thick layer of adipose tissue will reduce penetration due to refractive or aberration errors in the acoustic path to the site of interest. Similarly increased amounts of amniotic fluid with fetal imaging enhance penetration and may permit the use of frequencies higher than those ordinarily used at the given scanning site. The frequency range, or bandwidth, 1,2,6 of the transducer will determine whether it can support B-mode imaging at different center frequencies and also operate in Doppler, harmonic, and color flow modes. With Dopplerbased imaging modes, we often need to operate with lower frequencies than the B-mode frequency to minimize aliasing. With harmonic imaging, by definition, one uses a receive frequency that is a multiple (usually 2) of the transmitted frequency; hence the need for the wide bandwidth. The bandwidth and focusing properties will also influence image resolution. In clinical practice, it is essential to ensure that the image obtained can discern the smallest possible dimensions in both the lateral and axial directions. Finally, the number of individual transducer elements is of interest because the number of active elements (with the exception of phased arrays or angularly scanned 2D arrays) determines the lateral extent or width of the image. For phased arrays, an increasing number of elements is associated with improved resolution and penetration depth. For 2D arrays (usually symmetric), the number of elements along the x and y directions determines the extent of the volume for linearly scanned arrays. For a 2D phased array, resolution and penetration increase with a greater 578 J Ultrasound Med 2013; 32:573 582

number of elements along the x and y directions, but the angular shape or FOV remains the same, independent of the number of active elements used. The focusing in a fixed direction can indirectly affect imaging because focusing is positioned at only one depth and is poorer elsewhere. For 3D imaging, mechanically scanned 2D arrays suffer from the same fixed elevation focal depth limitation encountered in 2D imaging. In contrast, all the elements of fully populated 3D imaging or matrix arrays focus electronically at one point in both azimuth and elevation planes to provide far better resolution. At the deepest depths, it is the maximum number of available active channels in the system that determines the resolution (along with strength of focusing and system noise). Spatial resolution is generally poorer (typically by a factor of 2) than the temporal resolution along scan lines; in the discussion presented here, resolution refers to spatial resolution, unless noted otherwise. For phased arrays, the number of channels usually corresponds to the maximum number of elements. As a general rule of thumb, because elements are typically on half-wavelength spacing, the more elements, the better the spatial resolution, which is inversely proportional to the active aperture in wavelengths. For example, a 64-element array, 32-wavelength aperture will have maximum spatial resolution 2-fold lower (wider beam) than that of a 128-element, 64-wavelength array. In the case of a linear array, which may have several hundred elements, the number of elements determines the lateral extent of the image, but it is the number of active channels that governs the resolution. For these 1D arrays, the resolution out of the imaging plane (also known as slice thickness) is poor except near the fixed elevation focal length. For 2D arrays, the spatial resolution is inversely proportional to the active apertures that form the sides of the 2D array. Two-dimensional arrays have superior resolution compared to 1D array focusing with fixed elevation focusing because true-point focusing can be achieved simultaneously in azimuth and elevation for 3D imaging. Another way of looking at resolution is F#. The smaller the F#, the better the resolution. 1 A simple estimate of the full (beam) width in millimeters, a common measure of resolution, neglecting absorption, is approximately F# λ, where λ is wavelength (1.5 mm/μs/f [MHz]). For example, resolution would be 0.3 mm at 5 MHz for an F# = 1. Focal depths are also active aperture dependent. For example, for a 128-element 64-wavelength array, the deepest focal depth achieved at maximum aperture and an F# = 1 is F = F# L = 64 wavelengths. The actual penetration or useable scan depth would, of course, be deeper than the maximum focal depth. Matching Transducers to Clinical Applications Now that transducer types and properties have been related to imaging and acoustic windows, they can contribute to the selection of transducers for specific clinical applications. 3 The appropriateness of certain transducers to specific applications evolved historically and through specific tailored designs. The primary considerations are the target region of interest and its extent and the available acoustic windows needed for access. Abdominal Imaging When transducer arrays were initially introduced commercially for abdominal imaging (including obstetrics and gynecology) in the 1970s, they were of the linear array type (type A in Figure 4 with image format 1 of Figure 3). In most cases, the contact area with the patient was not a critical issue, and some of these linear arrays were quite long (eg, 8 cm) to cover, say, the third-trimester fetal head. 3,5 However, it was soon realized that one could achieve sufficiently large coverage by the use of curved or convex arrays (type C in Figure 4) without incurring the penalty of having to manipulate the rather unwieldy linear array transducers. The convex arrays (Figure 4C) are the tools of choice for most general 2D imaging applications involving the abdomen. The general form factor, related to ergonomic factors and the suitability of the transducer shape and FOV to the application, for abdominal 3D imaging is still evolving. Three key descriptors for these arrays are the footprint (overall aperture size), FOV, and radius of curvature (Figure 1C). The footprint describes the contact area usually in the form of a rectangle, circle or ellipse. Even though for abdominal imaging, access is not usually a concern, when these types of transducers are considered for new applications, window access is of primary importance. The radius of curvature and FOV (expressed in degrees of maximum angular coverage) are related to image extent and coverage. Advanced signal processing has been added to some systems to increase penetration; however, this feature is usually only available on certain probes. For the mechanical 3D probes, the currently preferred form factor is a mechanically swept convex array (Figure 4G and format 6 in Figure 3); however, fully electronic convex 2D arrays are now becoming available. For these cases, two FOVs are given for the orthogonal scan directions. Alternatively, phased arrays, because of their small footprint and wide sector image format, are also used for abdominal imaging. Finally, 2D or matrix arrays are becoming increasingly prevalent for these applications because of their superior image quality, resolution, and ease of use. J Ultrasound Med 2013; 32:573 582 579

Intercostal Imaging The primary applications of this imaging grouping are cardiac scanning and examination of the liver from between the ribs. Simply because of the restrictive anatomy and the limited acoustic windows caused by the ribs and the often encroaching lungs, the transducer choice here is limited to phased arrays (Figure 4B). Even in this area, initial attempts were made to use linear arrays; however, these were rapidly dropped due to the rib shadowing and the superiority of phased array transducer format 2. For cardiac applications, the probes tend to have array dimensions on the order of 20 14 mm depending on the manufacturer. The patient contact area will be slightly larger. These numbers have evolved over the last 40 years and depend on a number of things, such as the general size of the patient population. Age is another consideration; rib spacing and the depth penetration needed vary as children mature into adults. For noncardiac intercostal applications, the dimensions of the arrays are somewhat larger. As noted earlier, the existence of these anatomic limitations creates an upper performance limit for spatial resolution since resolution performance is inversely related to the size of the aperture, as explained above. In both cardiac and general intercostal imaging applications, the imaging depth is deep (depending on the patient size, it may be as deep as 24 cm), forcing the use of lower (1 3.5 MHz) frequencies and resulting in some further loss of imaging performance. There is an interesting aspect of cardiac imaging that has had a profound effect on the nature of the probes. Due to the presence of the ribs and other acoustically hostile tissue in the ray path, echocardiography suffers from imaging artifacts due to reverberant noise. The introduction of harmonic imaging has proven to be highly successful in reducing this noise. As a consequence, the importance of transducer bandwidth has become critical in cardiac transducer design. Today, most cardiac systems transmit at frequencies between 1.5 and 2.0 MHz and, of course, receive signals at frequencies twice that range. A major development in cardiac imaging was the implementation of fully populated 2D or matrix arrays (type E) containing thousands (typically 50 50 or so) of elements. These make possible real-time (4D) depiction of pyramidal volumes (format 7, Figure 3), visualization of arbitrary cut planes, and 4D cardiac imaging and color flow imaging. In addition, true electronic focusing in both the xz and yz planes provides superior resolution in comparison with all other 1D array transducers. Superficial and Breast Imaging This category refers to superficial imaging of carotids, leg veins, breasts, thyroids, testicles, etc and includes the categories of small parts, musculoskeletal, and peripheral vascular imaging. It is the last bastion of the application of linear arrays (type A), which formed the starting design type for the applications discussed earlier. In this clinical category, access is usually not an issue, and the sizes of the probes themselves can be small (because of the use of high 7- to 15-MHz frequencies and the resultant small element sizes). Musculoskeletal applications for imaging muscles, ligaments, and tendons also use arrays of this type. In the last 10 years, breast imaging has gone to very high frequencies (eg, 14 MHz), while imaging of the peripheral vasculature has remained at lower (about 3 11 MHz) values due to the need to include deeper leg veins and successful Doppler performance. Usually the capability of the array to add trapezoidal imaging (format 4) is a considerable advantage. As in abdominal imaging, 3D imaging with mechanically swept probes or electronic 2D arrays is now available for superficial and breast applications, greatly improving the coverage available and image quality. For applications involving imaging vasculature, some probes have advantages of including modes that enhance flow visualization. Obstetrics and Gynecology At the present time, mechanically scanned convex or linear arrays (types G and F) are used widely to provide 3D and 4D imaging of fetuses in vivo (formats 5 7). Matrix or fully populated 2D arrays (type E) are also available for this application (typically format 7). For gynecology, specialized endo-array probe shapes are used (type D). Typically, the arrays are at the end of the probe (end-fire arrays) and are convex or curved arrays with wide FOVs (format 3); however, phased arrays in an endo-array package (type D) can also be used (format 2). Frequencies used are typically 5 MHz and higher. As in other applications, 2D arrays have been designed for 3D imaging of these cases. Neonatal and Pediatric Pediatric transducers tend to have smaller footprints than transducers used for adults, applications A C, and operate at higher-frequency varieties ( 7 MHz) of those that are used for adults. Depending on the body region, types of transducers similar to those for adults are applicable. Phased arrays (type B) and 3D transducers (types E and G) are appropriate for cardiac imaging. Other arrays that are also useful for these clinical needs include static (2D) and, for 3D, mechanically swept linear arrays and convex arrays. 580 J Ultrasound Med 2013; 32:573 582

Intracavity Probes Intracavity probes constitute a large group of specialty transducers that are designed to image inside the body cavity. Transesophageal transducers are used to enable imaging of internal organs, especially the heart, from inside the esophagus (see Figure 5). They use higher frequencies ( 5 MHz) and are implemented as phased arrays with manipulators and motors to adjust the orientation of the transducer. Miniature transesophageal 2D arrays offer electronic scanning for 3D and 4D imaging. Transducers can be highly specialized for viewing usually within body openings or vessels. Intracardiac phased arrays are inserted through a vessel to gain access to the inner chambers of the heart. Surgical specialty probes include laparoscopic arrays inserted through small incisions to image and aid in laparoscopic surgery (similar to endo probes); these are remarkable for their FOV despite small diameters. Intraoperative arrays are specially shaped to be placed on vessels, organs, and regions made accessible during open surgery (see Figure 5). Others in this class are surgical and interventional probes with unique shapes (see Figure 5). As already noted, the probes that are inserted into the body are designed to fit through small openings and have a wide FOV (90 150 ). These probes include transrectal (or endorectal) for imaging of the pelvic region using the anus for access and the already described endovaginal (also called transvaginal) for imaging the female pelvis and reproductive organs using the vagina as entry for gynecologic and obstetric applications. These endo probes, described earlier, are cylindrical to fit into small orifices and have convex arrays (typically 3 9 MHz) at their ends with large fields of view, biplanes, or mechanically swept convex arrays. Probes for urologic applications include the biplane type. A unique transducer is the biplane probe, which consists of two orthogonal arrays producing images in planes xz and yz. Typically the arrays are small (8 12 mm) and of the convex type. Each format and transducer would correspond to those of a single imaging plane transducer such as format 3 of Figure 3 and the convex array of Figure 4C. However, sector or linear array formats are also possible, depending on the transducer construction, so that several combinations can be used in practice. Alternatively, a subset of the imaging capability of a 2D array is the simultaneous presentation of two orthogonal 2D images. Intravascular transducers are inserted into blood vessels to image the vessel walls for various pathologic conditions (type H and formats 8 and 9). They are most often mechanically rotated single transducers with frequencies greater than 20 MHz and dedicated imaging systems, although there are also tiny (about 2-mm-diameter) arrays designed for this purpose. Head Probes Transcranial imaging of the brain and its vasculature is conducted through limited acoustic windows through the skull such as the temples or eyes. Transorbital arrays are highfrequency (typically >20 MHz) ophthalmologic transducers and are used to image the eye or use the eye as an acoustic window. Transcranial probes are usually lowerfrequency (1 4 MHz) phased arrays used to image blood vessels within the skull through the temples as windows. Conclusions Many ultrasound imaging transducers are designed to operate in certain regions of the body for specific applications. A primary objective of this article is to provide a systematic approach that would aid in matching a transducer to a clinical application, starting with the acoustic window and the region and depth to be imaged. To this end, a checklist for selecting a transducer is given in Table 1. As indicated earlier, the first consideration for imaging a target region or organs is the access: the intended acoustic window. The transducer type must provide access through the selected acoustic window. The transducer type is linked with the image format, and common selections previously discussed include the linear, phased, convex, and 2D arrays. The size or transducer footprint must fit within the window, and in extreme cases in which the transducer window is an orifice, the transducer shape must conform to the available opening. As noted above, in some applications, specialty probes, such as endorectal transducers, are needed that are small enough in diameter (size) and have the elongated shape suitable for entering a body orifice. Second, the size or FOV and image format are selected to obtain the desired coverage over the region of interest. Here both the scan depth and image width or FOV are important. For linear arrays, the availability of trape- Table 1. Transducer Selection Checklist 1. Access to region of interest Acoustic window, footprint, appropriate transducer shape 2. Coverage of region of interest Scan depth, image width, FOV, scan volume, slice thickness 3. Maximum scan depth Frequency, absorption, resolution, signal processing 4. Coverage of essential Bandwidth, modes supported diagnostic modes J Ultrasound Med 2013; 32:573 582 581

zoidal imaging may be necessary for adequate coverage. For 3D or volumetric imaging, the extent of the image may be given as a set of maximum scan angles in orthogonal directions or a FOV and an angle. A somewhat more hidden parameter for 2D imaging for determining coverage for the region of interest is the elevation focal depth that describes the region with the thinnest slice thickness. Third, the maximum scan depth selected determines the highest achievable frequency through the penetration relation given in the Resolution and Penetration section above. For example, if the scan depth is 10 cm, then, as already discussed in the Resolution and Penetration section, the frequency from the penetration depth d is equal to 60/d = 60/10 = 6 MHz. This frequency provides an estimate of the best lateral resolution of about 1 wavelength for an F# = 1, or, for this example, the resolution is λ = c/f = 0.25 mm (from the Transducer Properties and Imaging section). Exceptions to this rule are systems that use advanced signal processing to enhance sensitivity and increase penetration. In addition, the use of piezoelectric materials such as piezo composites or domain-engineered single crystals can increase sensitivity or, correspondingly, penetration depth. 6 Fourth, the coverage of essential diagnostic imaging modes can be determined. From the manufacturerprovided data, the effective bandwidth needed to support different modes of interest may be extracted, or for the system considered, the actual modes of interest may be listed, such as pulsed wave Doppler, multiple imaging frequencies available, or elastographic mode. Transducers with piezoelectric materials such as piezo composites or domain- engineered single crystals can increase bandwidth substantially. 6 In conclusion, transducers and image formats have evolved to better suit specific clinical applications. The classification and organization given in this article provide the background for the selection of a transducer for a particular purpose. In addition, the understanding provided can aid in determining transducer characteristics needed for new cases, thereby extending the range of transducer use. 3. Rumack CM, Wilson SR, Charboneau JW, Levine D. Diagnostic Ultrasound. 4th ed. Maryland Heights, MO: Mosby; 2011. 4. Duck FA. Physical Properties of Tissue. New York, NY: Academic Press; 1990. 5. Woo J. Obstetric Ultrasound website; 1998 2002. http://www.obultrasound.net/. 6. Szabo TL, Lewin PA. Piezoelectric materials for imaging. J Ultrasound Med 2007; 26:283 288. References 1. Szabo TL. Diagnostic Ultrasound Imaging: Inside Out. Burlington, MA: Elsevier; 2004. 2. Reid JM, Lewin PA, Ultrasound imaging transducers. In: Encyclopedia of Electrical and Electronics Engineering. Vol 22. New York, NY: John Wiley & Sons; 1999:664 672. 582 J Ultrasound Med 2013; 32:573 582