Influence of body position, PEEP and intraabdominal pressure on the catheter positioning for neurally adjusted ventilatory assist

Similar documents
Using NAVA titration to determine optimal ventilatory support in neonates

SERVO EDUCATION NAVA in neonatal settings Study Guide

Monitoring Neural Output

Weaning: Neuro Ventilatory Efficiency

Faculty Disclosure. Off-Label Product Use

Electrocardiographic Guidance for the Placement of Gastric Feeding Tubes: A Pediatric Case Series

Monitoring Respiratory Drive and Respiratory Muscle Unloading during Mechanical Ventilation

Neurally-adjusted Ventilatory Assist

The calcium sensitizer levosimendan improves human diaphragm function

NAVA-korzyści dla noworodka

Proportional Assist Ventilation (PAV) (NAVA) Younes ARRD 1992;145:114. Ventilator output :Triggering, Cycling Control of flow, rise time and pressure

Potential Conflicts of Interest

Comparison of Pressure-, Flow-, and NAVA-Triggering in Pediatric and Neonatal Ventilatory Care

Comparison of automated and static pulse respiratory mechanics during supported ventilation

Journal Club American Journal of Respiratory and Critical Care Medicine. Zhang Junyi

New Modes to Enhance Synchrony & Dietrich Henzler MD, PhD, FRCPC Division of Critical Care

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

Electrical activity of the diaphragm (EAdi) as a monitoring parameter in difficult weaning from respirator: a pilot study

Monitoring diaphragm function in a patient with myasthenia gravis: electrical activity of the diaphragm vs. maximal inspiratory pressure

Keywords: Non-invasive mechanical ventilation, Respiratory Failure, Respiratory muscles, Hypercapnia, Breathing pattern.

Neurally Adjusted Ventilatory Assist Mode in Pediatric Intensive Care Unit and Pediatric Cardiac Care Unit

SERVO-i ventilator One system, multiple options

Daniel Hadfield Critical Care Nurse NIHR / HEE Clinical Doctoral Research Fellow King s College Hospital

Landmark articles on ventilation

NAVA. In Neonates. Howard Stein, M.D. Director Neonatology. Neurally Adjusted Ventilatory Assist. Toledo Children s Hospital Toledo, Ohio

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

Effect of flumazenil on diaphragm electrical activation during weaning from mechanical ventilation after acute respiratory distress syndrome

Servo-i Ventilator. One system, multiple options. This document is intended to provide information to an international audience outside of the US.

Tracking lung recruitment and regional tidal volume at the bedside. Antonio Pesenti

Ventilator Dyssynchrony - Recognition, implications, and management

Physiologic response to varying levels of pressure support and neurally adjusted ventilatory assist in patients with acute respiratory failure

PRACTICE GUIDELINES: INTRA-ABDOMINAL HYPERTENSION/ABDOMINAL COMPARTMENT SYNDROME

Effect of lung volume on the oesophageal diaphragm EMG assessed by magnetic phrenic nerve stimulation

Evaluation of the Effects of Elevated Intraabdominal. on the Respiratory Mechanics in Mechanically Ventilated Patients

Min Hur, Eun-Hee Kim, In-Kyung Song, Ji-Hyun Lee, Hee-Soo Kim, and Jin Tae Kim INTRODUCTION. Clinical Research

Volume 16 - Issue 3, Cover Story

Subject ventilator synchrony during neural versus pneumatically triggered non-invasive helmet ventilation

Single-Lead DDD System: A Comparative Evaluation of Unipolar, Bipolar and Overlapping Biphasic Stimulation

CURRENT TRENDS IN NON-INVASIVE VENTILATION. Disclosures. Why not invasive ventilation? Objectives. Currently available modes

Innovations in Neonatal Ventilation

Patient Ventilator Interactions. Patient-Ventilator Interactions. Assisted vs Controlled MV. Ventilatory Muscle Fatigue Recovery

Feasibility of neurally adjusted positive end-expiratory pressure in rabbits with early experimental lung injury

Assessment of patient-ventilator breath contribution during neurally adjusted ventilatory assist in patients with acute respiratory failure

Closed Loop Ventilation

(12) United States Patent (10) Patent No.: US 8.485,980 B2

AORTIC STENOSIS HENRY FORD HOSPITAL CENTER FOR STRUCTURAL HEART DISEASE

Relax and Learn At the Farm 2012

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)

CSIM annual meeting Acute respiratory failure. Dr. John Ronald, FRCPC Int Med, Resp, CCM. October 10, 2018

Diaphragm Activation during Exercise in Chronic Obstructive Pulmonary Disease

ECG INTERPRETATION MANUAL

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh

Year in Review. NAVA and Edi monitoring

The Impact of Patient-Ventilator. Karen J Bosma, MD, FRCPC Critical Care Medicine and Respirology

Removing ECG Artifact from the Surface EMG Signal Using Adaptive Subtraction Technique

INTRA-ABDOMINAL HYPERTENSION AND SECONDARY ABDOMINAL COMPARTMENT SYNDROME IN MEDICAL PATIENTS COMPLICATION WITH A HIGH MORTALITY

Understanding Atrial Fibrillation

The Correlation Between Esophageal and Abdominal Pressures in. Mechanically Ventilated Patients Undergoing Laparoscopic Surgery

Simulation of Late Inspiratory Rise in Airway Pressure During Pressure Support Ventilation

POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO

The use of proning in the management of Acute Respiratory Distress Syndrome

Neurally Adjusted Ventilatory Assist After Pediatric Cardiac Surgery: Clinical Experience and Impact on Ventilation Pressures

The activity of the respiratory muscles. Neural respiratory drive during apnoeic events in obstructive sleep apnoea

PRESSURE support ventilation (PSV) is the most widely

Clinical Data Summary: Avoid FFS Study

SUPPLEMENTAL MATERIAL

Recognizing and Correcting Patient-Ventilator Dysynchrony

Mechanical Ventilation in COPD patients

DAILY SCREENING FORM

Francois Beloncle 1,2,3, Lise Piquilloud 3,4, Nuttapol Rittayamai 1,2,5, Christer Sinderby 2, Hadrien Rozé 6 and Laurent Brochard 1,2*

Ling Liu 1, Feiping Xia 1, Yi Yang 1, Federico Longhini 1,2, Paolo Navalesi 2,3,4, Jennifer Beck 5,6,8, Christer Sinderby 5,7,8 and Haibo Qiu 1*

Ventilator Waveforms: Interpretation

Lamouret et al. Critical Care (2019) 23:2

The Frequency of Gastroesophageal Reflux Disease in Nutcracker Esophagus and the Effect of Acid-Reduction Therapy on the Motor Abnormality

Mechanical Ventilation in COPD patients

Expiratory muscle pressure and breathing mechanics in chronic obstructive pulmonary disease

The respiratory pressure abdominal volume curve in a porcine model

Charisma High-flow CPAP solution

Interfacility Protocol Protocol Title:

Impedance Cardiography (ICG) Method, Technology and Validity

INCIDENCE OF FAILED VENTILATION AMONGST CRITICAL CARE PATIENTS GOING FOR URGENT SURGERY

Extubation Failure & Delay in Brain-Injured Patients

UCH WEANING FROM MECHANICAL VENTILATION PATHWAY

ABSENCE OF INSPIRATORY LARYNGEAL CONSTRICTOR MUSCLE ACTIVITY DURING NASAL NEURALLY ADJUSTED VENTILATORY ASSIST IN NEWBORN LAMBS

Nasal CPAP, Abdominal muscles, Posture, Diagnostic ultrasound, Electromyogram

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

1.1.2 CPAP therapy is used for patients who are suffering from an acute type 1 respiratory failure (Pa02 <8kPa with a normal or low Pac02).

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Supplementary appendix

Point-of-Care Ultrasound Closer look at the Inferior Vena Cavae &

Clinical Evidence Summary ACCURYN ADVANCED CRITICAL CARE MONITORING

Effects of Pericardial Effusates of Various Conductivities on Body-Surface Potentials in Dogs

Electroacupuncture decreases the urinary bladder pressure in patients with acute gastrointestinal injury

Patient Resources: Arrhythmias and Congenital Heart Disease

Alexander A Schult, M.D., FCCP. October 21, 2017 Revised 1/10/18

Patient-Ventilator Synchrony and Impact on Outcome

Intra-Arterial Blood Pressure Monitoring

All rights, in particular the proprietary rights of use and exploitation relating to the following contribution as well as parts thereof (including

Patient Asynchrony and Its Impact on Patient Outcome

Transcription:

Intensive Care Med (2011) 37:2041 2045 DOI 10.1007/s00134-011-2373-3 PHYSIOLOGICAL AND TECHNICAL NOTES Jürgen Barwing Cristina Pedroni Michael Quintel Onnen Moerer Influence of body position, PEEP and intraabdominal pressure on the catheter positioning for neurally adjusted ventilatory assist Received: 21 March 2011 Accepted: 11 July 2011 Published online: 14 October 2011 Ó The Author(s) 2011. This article is published with open access at Springerlink.com J. Barwing C. Pedroni M. Quintel O. Moerer ()) Department of Anaesthesiology, Emergency and Intensive Care Medicine, Georg-August University of Göttingen, Robert-Koch-Str. 40, 37075 Göttingen, Germany e-mail: omoerer@gwdg.de Tel.:?49-551-399561 Abstract Purpose: Neurally adjusted ventilatory assist (NAVA) relies on the patient s electrical activity of the diaphragm (EAdi) for actuating the ventilator. Thus a reliable positioning of the oesophageal EAdi catheter is mandatory. We aimed to evaluate the effects of body position (BP), positive end-expiratory pressure (PEEP) and intra-abdominal pressure (IAP) on catheter positioning. Methods: Twenty-one patients were enrolled in this study. In six different situations [supine or 45 head of bed elevation (HBE) at PEEP 5 and 15 cmh 2 O; left lateral antidecubitus at PEEP 5 cmh 2 O; supine at PEEP 5 cmh 2 O with abdominal surgical belt (ASB)] the catheter position was evaluated for the stability of the EAdi signal and information provided by a catheter positioning tool (highlighted electrical activity in central leads, absence of p waves in the distal lead). Results: With an optimal catheter position EAdi signals were stable for all tested situations. During 45 PEEP 15 and supine PEEP 15 absence of p waves in the distal lead revealed a difference compared with supine PEEP 5 (p = 0.03), suggesting a caudal shift of the diaphragm relative to the oesophagus. The analysis of the highlighted electrical activity in the central leads supports this finding, revealing an influence of PEEP, BP and IAP on EAdi catheter position (p \ 0.01). Conclusion: PEEP, BP and IAP may affect the EAdi catheter position, although not compromising a stable signal. Additional information as provided by the catheter positioning tool is needed to ensure an optimal EAdi catheter position. Keywords Neurally adjusted ventilatory assist (NAVA) Body position PEEP Intra-abdominal pressure Electrical activation of the diaphragm (EAdi) Introduction During neurally adjusted ventilatory assist (NAVA) the ventilator is triggered by the electrical activity of the diaphragm (EAdi), assessed by a special gastric tube (EAdi catheter) [1 5]. Pressure assistance is applied in proportion to the amplitude of the EAdi, which represents the respiratory effort of the patient controlled by his respiratory centre [6, 7]. The EAdi catheter positioning is crucial because NAVA relies on a representative signal from the diaphragm. Recently we evaluated a method to determine the optimal position of the EAdi catheter based on the quality and amplitude of the EAdi signal and the trans-oesophageal electrocardiogram (ECG) [8]. Positive end-expiratory pressure (PEEP), body position (BP) and intra-abdominal pressure (IAP) were not considered at that time, although these are factors known to influence the position of the diaphragm [9 11]. Therefore we aimed

2042 to determine their effect on the EAdi signal and the optimal EAdi catheter position. Methods After approval from the local ethics committee (University of Goettingen, Germany) NAVA was initiated on the basis of the clinical decision of the consultant in charge of the intensive care unit (ICU) with the intention to facilitate respirator weaning. We used a ventilator capable of NAVA (Servo-i, Maquet Critical Care, Solna, Sweden). EAdi catheter placement (16 Fr diameter, 125 cm long; Maquet Critical Care, Solna, Sweden) was performed by the study group. Patients were supine with the head of the bed elevated at 30 and pressure support ventilation with preset PEEP was used during EAdi catheter positioning. The catheter was inserted nasally to a maximum distance of 80 cm. For EAdi measurement it is mounted with an electrode array of one reference electrode and eight measuring electrode rings on the distal part of the catheter at intervals of 16 mm, starting at 120 mm from the tip. Catheter positioning was monitored by a special tool implemented in the ventilator that displays four raw leads of electrical activity representing cranial to caudal arranged electrode pairs. The catheter positioning tool identifies the electromyographic (EMG) signal with the highest amplitude and highlights it in blue [12]. The software highlights two adjacent leads unless only the outermost pair of electrodes detects a diaphragmatic EMG signal. In addition the position of the electrodes in relation to the heart can be estimated by evaluating the presence/absence of p wave (if present) and QRS complex in different leads. Further processing of the signal (e.g. low pass filter to remove ECG) results in the EAdi signal in microvolts (lv, analogue to potential difference between electrodes) used to control the ventilator during NAVA. During the placement procedure the catheter was extracted in steps of 1 cm. At each step data were recorded over 60 s using special software (NAVA tracker, Maquet Critical Care, Solna, Sweden). The maximum EAdi value of each breathing cycle during tidal breathing was taken to calculate the median EAdi max for each step. The catheter position was considered optimal (OPT) when the highlighted EAdi appeared in the central leads, the p wave was absent in the distal lead and a stable EAdi max was found [8]. For each patient the catheter placement procedure was randomly repeated for the following six situations without completely removing/reinserting it: Supine or 45 head of bed elevation (HBE) at PEEP 5 and 15 cmh 2 O Supine, PEEP 5 cmh 2 O, with abdominal surgical belt (ASB) Left lateral anti-decubitus, PEEP 5 cmh 2 O The measures for catheter positioning (highlighted EAdi in the central leads, absent p wave in the distal lead, stable EAdi max ) were evaluated over a range of ±5 cm from OPT. The intervention-related catheter positions were referenced to OPT. During data analysis atrial activity was taken into account. Urinary bladder pressure was measured after retrograde filling with 50 ml of normal saline in supine position at PEEP 5, with and without ASB. Results are displayed as median with 25/75 percent quartiles (Q). Wilcoxon matched pairs and analysis of variance (ANOVA) were used to compare between groups. Results EAdi catheter placement was possible in all 21 patients (see Table 1 for patient characteristics). Over all six situations we found a stable EAdi max from -5 cm to OPT to 5 cm with a trend to weaker signals at -5 cm (p = 0.12). The central leads of the catheter positioning tool were highlighted in more than 95% of patients at OPT in all tested situations, but a shift of the highlighted central leads could be detected in all 21 patients (Fig. 1). The tested situations influenced the optimal catheter position (p \ 0.01). P waves could be evaluated in 15 patients (six patients with atrial fibrillation). The OPT condition absence of p wave in distal lead was fulfilled at OPT during left lateral PEEP 5 and supine PEEP 5 in 93%, during supine PEEP 5 ASB in 87%, 45 PEEP 5 in 80% and 45 PEEP 15 and supine PEEP 15 in 53% of the patients (p = 0.03 compared to supine PEEP 5 ). With the EAdi catheter at OPT? 1 cm, p waves were absent in 93% ( 45 PEEP 15 ) and 87% ( supine PEEP 15 ) of patients. Comparing patients with sinus rhythm and atrial fibrillation neither a main effect of heart rhythm (p = 0.41) nor an interaction effect with the tested situations (p = 0.57) could be found (Fig. 2). Considering EAdi max in patients with atrial fibrillation, we could not determine a difference between catheter positions in the evaluated range (±5 cm to OPT; p = 0.44). In patients with sinus rhythm the EAdi max changed (p = 0.04) with smaller values at -5 cm. Discussion NAVA relies on a stable EAdi signal for respirator control. Therefore correct placement of the EAdi catheter is mandatory. The EAdi catheter derives the signal from the part of the diaphragm close to the oesophagus. The position of the diaphragm depends on the application of PEEP, BP and IAP [9 11]. Six different situations with

2043 Table 1 Basic patient characteristics including the catheter position considering data from the positioning tool (OPT), intra-abdominal pressure measured with the patient supine and PEEP set to 5 cmh 2 O (IAP) and intra-abdominal pressure with the abdominal surgical belt in place (IAP ASB) No. Gender Diagnosis Rhythm Age (years) Height (cm) BMI OPT (cm) IAP IAP ASB PEEP EAdi max (lv) 1 M IAH (CHE) SR 67 160 25 61 25 MD 8 2.7 2 F SAH SR 69 160 16 62 6 10 5 3.3 3 M Sepsis Afib 71 178 39 67 14 18 8 8.3 4 M CHD (ACB) SR 78 175 29 66 13 16 5 2.1 5 M Aflut Afib 70 178 25 68 11 14 6 11.8 6 F CHD (CAB) SR 76 152 22 61 8 18 5 3.9 7 M AS (AVR) Afib 73 174 27 69 10 18 5 4.4 8 M CHD (CAB) SR 57 170 26 70 5 18 5 1.2 9 F ACS (CAB) SR 77 164 22 63 9 18 8 7.3 10 F AS (AVR) Afib 74 168 27 69 8 18 6 4.2 11 F Pneumonia Afib 82 165 37 67 10 14 10 5.6 12 F TBI SR 81 170 28 63 13 18 7 7.3 13 F SAH SR 39 170 22 57 9 15 6 3.2 14 M CHD (CAB) SR 60 178 25 66 9 14 6 13.1 15 M AS (AVR) SR 82 165 33 68 14 19 10 3.3 16 M AS (AVR) Afib 82 170 22 68 5 18 5 19.8 17 M CHD (CAB) SR 66 172 31 67 12 14 8 2.0 18 F ACS (CAB) SR 70 160 21 66 7 18 5 1.5 19 M pneumonia SR 84 180 25 67 13 19 10 3.7 20 M ACS (CAB) SR 69 174 30 69 17 21 5 1.6 21 M Pneumonia SR 65 175 27 69 8 18 8 6.7 Median 71 170 26 67 10* 18* 6 3.9 Q25/Q75 67/78 165/175 22/29 63/68 8/13 15/18 5/8 2.7/7.3 No. number of patient, BMI body mass index, NEX mod catheter position calculated according to the modified nose ear xiphoid formula, OPT optimal catheter position, IAP intra-abdominal pressure, ASB abdominal surgical belt, M male, F female, IAH intra-abdominal hypertension, CHE cholecystectomy, SAH subarachnoid haemorrhage, CHD coronary heart disease, CAB coronary artery bypass, Aflut atrial flutter, AS aortic valve stenosis, AVR aortic valve replacement, ACS acute coronary syndrome, TBI traumatic brain injury, SR sinus rhythm, Afib atrial fibrillation, MD missing data (no ASB applied due to IAH) *p \ 0.001 (IAP compared to IAP ASB) Fig. 1 Optimal catheter positions determined by highlighted central leads displayed as mean (filled squares) with 95% confidence interval (whiskers) for six different situations; y-axis catheter position in cm from OPT. Significant differences in post hoc paired comparison (Tukey Kramer) for 45 PEEP 5 vs supine PEEP 15 (*p = 0.002), left lateral PEEP 5 vs supine PEEP 15 ( # p = 0.006) and supine PEEP 15 vs supine PEEP 5 ASB ( p = 0.0005) (n = 21) Fig. 2 Optimal catheter positions determined by highlighted central leads for patients with sinus rhythm (filled squares and dashed lines; n = 15) and atrial fibrillation (open squares and solid lines; n = 6) displayed as mean (squares) with 95% confidence interval (whiskers) for six different situations; y-axis catheter position in cm from OPT

2044 regard to PEEP, BP and IAP were evaluated and their effects on the catheter position relative to the diaphragm via the information provided by the catheter positioning tool (highlighted central leads and absence of p waves in the distal lead) were found. The changes of the EAdi signal were not significant comparing catheter positions from -5 to OPT to?5 cm and would not have impaired running the NAVA mode. The electrode array on the EAdi catheter is 14.4 cm wide and the ventilator software always uses the diaphragmatic EMG with the highest amplitude found for calculating the EAdi signal used for NAVA. Thus even with a catheter that is displaced ±5 cm from OPT there are still electrodes close to the diaphragm, thereby ensuring a constant signal quality. This result is in accordance with our previous study and demonstrates the limits of safety implemented in the EAdi signal acquisition [8]. EAdi max therefore is only suitable to detect large changes in the EAdi catheter position relative to the diaphragm. Further limitations have to be considered when looking at our results. Our measurement technique relies on the detection of the electromyographic activity of the diaphragmatic tunnel covering 2 3 cm of the oesophagus [13] and the assumption that the signal originates from its centre [12]. The accuracy of measurement is limited by the electrode distance of 16 mm. Nevertheless we were able to demonstrate changes. The evaluation of the p waves revealed a difference between supine PEEP 5 and 45 PEEP 15 and supine PEEP 15. The fact that the distal lead still had proximity to the right atrium in both PEEP 15 situations may be explained by the fact that the increase in PEEP displaces the diaphragm and thoracic content, including the heart, to a more caudal position. Froese and Byran [10] reported that a PEEP level of 10 cmh 2 O results in a 2-cm caudal shift of the middle part of the diaphragm compared with zero PEEP. Comparing patients with and without p waves we did not find a statistically significant difference although Fig. 2 suggests a shift to deeper catheter positions in patients with sinus rhythm. Figure 2 considers the highlighted central leads only and not the additional information provided by the p waves. Therefore the figure allows the interpretation that OPT was determined at a more cranial position in patients with sinus rhythm compared with patients with atrial fibrillation. This observation is confirmed by the smaller EAdi max at -5 cm from OPT in patients with sinus rhythm but not in patients with atrial fibrillation. A possible explanation lies in the catheter positioning procedure that defined OPT (pulling the catheter out stepwise) and the different prioritization of the variables (first, EAdi highlighted in central leads; second, p waves absent in the distal lead; third, stable EAdi max ). Although we consider only a small collective it might be assumed by the evaluation of the EAdi max data that catheter positioning was more accurate in patients in whom p waves were unavailable. The analysis of the highlighted central leads reveals effects of PEEP, BP and IAP on the diaphragm, with a high PEEP displacing it caudally and an increased IAP displacing it to a more cranial position (Fig. 1). The changes are consistent over all tested situations and similar to the results described above regarding p waves. A 30 HBE increases IAP by 3.7 mmhg [14] and a 45 HBE increases IAP by 9mmHg[15]. This might explain why the caudal shift in supine PEEP 15 seems to be more pronounced than in 45 PEEP 15. Furthermore it explains a caudal shift in supine PEEP 5 as OPT was established at 30 HBE at a median PEEP of 6 cmh 2 O. Conclusion PEEP, BP and IAP affect the optimal EAdi catheter position. NAVA ventilation is not affected due to the wide electrode array compensating for the small diaphragmatic shift seen with such changes. However this only holds true if an optimal catheter position is ensured in advance. A verification of the optimal catheter position should be performed after major changes in ventilator settings, patient positioning or clinical condition e.g. increases of abdominal pressure. Catheter positioning is not impaired by the lack of p waves. Acknowledgments We thank Katharina Lange, Dep. Medical Statistics, Georg-AugustUniversity of Göttingen, Germany for her support. Open Access This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited. References 1. Sinderby C, Navalesi P, Beck J, Skrobik Y, Comtois N, Friberg S, Gottfried SB, Lindstrom L (1999) Neural control of mechanical ventilation in respiratory failure. Nat Med 5:1433 1436 2. Colombo D, Cammarota G, Bergamaschi V, De Lucia M, Corte FD, Navalesi P (2008) Physiologic response to varying levels of pressure support and neurally adjusted ventilatory assist in patients with acute respiratory failure. Intensive Care Med 34:2010 2018 3. Brander L, Leong-Poi H, Beck J, Brunet F, Hutchison SJ, Slutsky AS, Sinderby C (2009) Titration and implementation of neurally adjusted ventilatory assist in critically ill patients. Chest 135:695 703

2045 4. Allo JC, Beck JC, Brander L, Brunet F, Slutsky AS, Sinderby CA (2006) Influence of neurally adjusted ventilatory assist and positive endexpiratory pressure on breathing pattern in rabbits with acute lung injury. Crit Care Med 34:2997 3004 5. Beck J, Campoccia F, Allo JC, Brander L, Brunet F, Slutsky AS, Sinderby C (2007) Improved synchrony and respiratory unloading by neurally adjusted ventilatory assist (NAVA) in lung-injured rabbits. Pediatr Res 61:289 294 6. Moerer O, Beck J, Brander L, Costa R, Quintel M, Slutsky AS, Brunet F, Sinderby C (2008) Subject-ventilator synchrony during neural versus pneumatically triggered non-invasive helmet ventilation. Intensive Care Med 34:1615 1623 7. Moerer O, Barwing J, Quintel M (2008) Neurally adjusted ventilatory assist (NAVA). A new mode of assisted mechanical ventilation. Anaesthesist 57:998 1005 8. Barwing J, Ambold M, Linden N, Quintel M, Moerer O (2009) Evaluation of the catheter positioning for neurally adjusted ventilatory assist. Intensive Care Med 35:1809 1814 9. Malbrain ML (2004) Is it wise not to think about intra-abdominal hypertension in the ICU? Curr Opin Crti Care 10:132 145 10. Froese AB, Bryan AC (1974) Effects of anesthesia and paralysis on diaphragmatic mechanics in man. Anesthesiology 41:242 255 11. Mutoh T, Lamm WJ, Embree LJ, Hildebrandt J, Albert RK (1991) Abdominal distension alters regional pleural pressures and chest wall mechanics in pigs in vivo. J Appl Physiol 70:2611 2618 12. Beck J, Sinderby C, Lindström L, Grassino A (1996) Influence of bipolar esophageal electrode positioning on measurements of human crural diaphragm electromyogram. J Appl Physiol 81:1434 1449 13. Delattre JF, Palot JP, Ducasse A, Flament JB, Hureau J (1985) The crura of the diaphragm and diaphragmatic passage. Applications to gastroesophageal reflux, its investigation and treatment. Anat Clin 7:271 283 14. Cheatham ML, De Waele JJ, De Laet I, De Keulenaer B, Widder S, Kirkpatrick AW, Cresswell AB, Malbrain M, Bodnar Z, Mejia-Mantilla JH, Reis R, Parr M, Schulze R, Puig S (2009) The impact of body position on intraabdominal pressure measurement: a multicenter analysis. Crit Care Med 37:2187 2190 15. De Keulenaer BL, De Waele JJ, Powell B, Malbrain M (2009) What is normal intra-abdominal pressure and how is it affected by positioning, body mass and positive end-expiratory pressure? Intensive Care Med 35:969 976