Comparison of Methodological Data Measurement Limits in CD4 1 T Lymphocyte Flow Cytometric Enumeration and Their Clinical Impact on HIV Management

Similar documents
Quality Management 333 Liam Whitby 1.2 Liam Whitby 24-Jul Jul-2016 unknown

ACS GUIDELINE FOR LYMPHOCYTE SUBSET IMMUNOPHENOTYPING. Second Edition Australasian Cytometry Society Guideline Document

Current Laboratory Practices in Flow Cytometry for the Enumeration of CD4 + T-Lymphocyte. Subsets

Evaluation of a Single-Platform Technology for Lymphocyte Immunophenotyping

Original Article Time and temperature stability of T-cell subsets evaluated by a dual-platform method

How can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping

D. BARNETT, K. GOODFELLOW, J. GINNEVER, V. GRANGER, L. WHITBY, J.T. REILLY

CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 2000, p Vol. 7, No. 3

UK NEQAS Low Level Leucocyte Enumeration Programme. Mr Matthew Fletcher Advanced Biomedical Scientist UK NEQAS for Leucocyte Immunophenotyping

Affordable CD4 -T-Cell Counting by Flow Cytometry: CD45 Gating for Volumetric Analysis

Received 6 November 2006/Returned for modification 4 January 2007/Accepted 9 January 2007

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets

Technical Brief on CD4 Technologies (May 2010)

When to start: guidelines comparison

FINAL REPORT HBV Serology External Quality Assessment Scheme

External validation for CD4 and viral load assays: South African Experience

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Technical Bulletin No. 157

THE IMPACT OF A CD4 EXTERNAL QUALITY ASSESSMENT PROGRAMME FOR SOUTHERN AFRICA AND AFRICA. Hazel Aggett

PointCare NOW TM Technical Note

Il ruolo delle VEQ per la sicurezza trasfusionale

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2017Oct27

Evaluation of UK NEQAS (H) Hb A 2 and related performance data

EXPERT COMMITTEE ON BIOLOGICAL STANDARDIZATION Geneva, 29 October to 2 November 2018

NIAID DAIDS FLOW CYTOMETRY LABORATORY CERTIFICATION

FINAL REPORT HBV Serology External Quality Assessment Scheme

Workshop on Proficiency Testing for Medical Laboratories. Daniel W. Tholen, M.S. March 13, 2013

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

Lysosomal Enzymes in fibroblasts

Kovit Pattanapanyasat, Ph.D

Recommendations from the Working Group of Senior Scottish Clinical Biochemists on Parathyroid Hormone (PTH) Targets in the Management of Renal Failure

by Automated Cell Counters

Anti-Infective Clinical Trials

Lyophilised Cell Standards for Flow Cytometry. Richard Stebbings, National Institute for Biological Standards and Control, UK

United Kingdom National External Quality Assessment Service for Microbiology [Established 1971]

Application Information Bulletin: Navios Multi-center Study

Key terms: flow cytometry; lymphocyte immunophenotyping; multicenter study; variation; standardization

PanLeucogating Akin Abayomi

Original Article. Noparat Oniem, M.D., Somnuek Sungkanuparph, M.D.

What I will talk about

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2013

Title: Prevalence and epidemiological correlates and treatment outcome of HCV infection in an Italian prison setting

Cost Savings by Reagent Reduction in Flow Cytometry-Based CD4+ T Cell Counts: An Approach to Improve Accessibility for HIV Management

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2014

Latent tuberculosis infection

Molecular Diagnostics Centre

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2015

Sysmex Educational Enhancement and Development No

Guideline on the use of minimal residual disease as a clinical endpoint in multiple myeloma studies

Problems with Standardisation of Automated Titration Local Level Implementation - ISO 15189

Recent developments for combining evidence within evidence streams: bias-adjusted meta-analysis

The challenges of quality assessment for point-of-care testing. Anne Stavelin. EQALM Symposium, Bergen 8 October General challenges

SUPPLEMENTARY INFORMATION In format provided by Stevens et al. (NOVEMBER 2008)

Therapeutic Apheresis Services Annual Review 2016/17

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

INTER AND INTRA-LABORATORY VARIABILITY OF CD4: A PRAGMATIC ANALYSIS. Submitted in partial fulfillment of the requirements for the degree

Guidelines for Gamma Irradiation of Blood Components

PARTICULARS, SCHEDULE 2 THE SERVICES, A SERVICE SPECIFICATION. A03/S(HSS)/a Pancreas transplantation service (Adult)

Investigation of CD4+ T cell numbers in HIV-infected patients among smokers and non-smokers in Thailand

High-Sensitivity Troponin: Star Player but No Lone Hero

LOGO. Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol University & Thai Society of Clinical Pathology

Policy: Client Involvement and Empowerment

30 th Anniversary International Scientific Congress

A Cellular Antigen Stabilisation Reagent

An update on the analysis of agreement for orthodontic indices

Summary of Significant Changes. Policy

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ

Integrated Research Application System (IRAS)

EDMA HIV-AIDS TEAM Fact Sheet November 2007

Date of study period: April 12 May 7, 2010 and August September, 2010

Introduction: Table/Figure Descriptions:

Water Microbiology Proficiency Test Scheme. Overview & Description

[Submitted Electronically]

Methods for Determining Random Sample Size

PATIENT SELECTION FOR DECEASED DONOR KIDNEY ONLY TRANSPLANTATION

Single donator specimens as advantage in external quality control assessments of infectious diseases

Thorax Online First, published on December 8, 2009 as /thx

Manchester Royal Infirmary. Antibody Deficiency. Information For Patients

VERBAL CONSENT TO PARTICIPATE IN A RESEARCH STUDY AND HIPPA AUTHORIZATION

New guidelines for HIV testing and areas where wider HIV testing policies should be considered.

Osteoporosis: fragility fracture risk. Costing report. Implementing NICE guidance

Flow cytometric CD4 enumeration of four different HIV-Infected blood samples at the cost of one monoclonal antibody reagent

Hearing aid dispenser approval process review Introduction Hearing aid dispenser data transfer... 6

10.2 Summary of the Votes and Considerations for Policy

FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION CLINICAL BIOLOGY SECTION

An Immunomagnetic Single-Platform Image Cytometer for Cell Enumeration Based on Antibody Specificity

Reference Laboratory Agreement on Multi-Analyte Pneumococcal Antibody Results: An

Guide to Fulfillment of Measurement Uncertainty

Q&A on HIV/AIDS estimates

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture

Day One: After you ve tested positive

An abuse of surrogate markers for AIDS. By David Rasnick Published online in British Medical Journal, March 8, 2003

Clinical Infectious Diseases Advance Access published June 16, Age-Old Questions: When to Start Antiretroviral Therapy and in Whom?

The role of cancer networks in the new NHS

Results of Proficiency Test PFOA/PFOS in Textile March 2017

Prevalent opportunistic infections associated with HIV-positive children 0-5 years in Benin city, Nigeria

Hepatitis C Virus (RNA)A

Transcription:

Cytometry Part B (Clinical Cytometry) 84B:248 254 (2013) Original Article Comparison of Methodological Data Measurement Limits in CD4 1 T Lymphocyte Flow Cytometric Enumeration and Their Clinical Impact on HIV Management Liam Whitby, 1 * Alison Whitby, 1 Matthew Fletcher, 1 Matthew Helbert, 2 John T. Reilly, 1 and David Barnett 1 1 UK NEQAS for Leucocyte Immunophenotyping, 4th Floor, Pegasus House, 463a Glossop Road, Sheffield S10 2QD, United Kingdom 2 Department of Immunology, Manchester Royal Infirmary, Manchester M13 9WL, United Kingdom UK NEQAS for Leucocyte Immunophenotyping, an ILAC G13:2000 accredited External Quality Assessment (EQA) organization, with over 3000 international laboratories participating in 14 programmes, issues 2 proficiency testing samples of stabilized whole blood to 824 participants in the Immune Monitoring (lymphocyte subset) programme every two months. We have undertaken a study of 58,626 flow cytometric absolute CD4 1 T lymphocyte count data sets from these laboratories over a 12-year-period (2001 2012) to determine counting method variation in data measurement limits and how this could influence the clinical management of HIV patients. Comparison of relative error and 99.9% confidence limits for absolute CD4 1 T lymphocyte values was undertaken using dual platform (DP) and single platform (SP) data and showed that the SP consistently outperformed DP, giving lower relative errors and confidence limits at clinically significant absolute CD4 1 T lymphocyte counts. Our data shows that absolute CD4 1 T lymphocyte counts should be obtained using single platform technology to reduce the variability at clinically relevant levels. On data where results (irrespective of platform) were below the international treatment threshold of 350 cells/ll, there was no significant misclassification between either SP or DP techniques meaning most patients would receive the correct treatment at the correct time. However, results that were above the treatment level of 350 cells/ll had a significant difference (P 5 0.04) between DP and SP platforms, suggesting patients monitored using DP technology were 20% more likely to start therapy prematurely than those monitored with SP technology. VC 2013 International Clinical Cytometry Society Key terms: flow cytometry; CD4; T lymphocytes; HIV; quality control; quality assessment; reproducibility How to cite this article: Whitby L, Whitby A, Fletcher M, Helbert M, Reilly JT, Barnett D. Comparison of methodological data measurement limits in CD4 þ T lymphocyte flow cytometric enumeration and their clinical impact on HIV management. Cytometry Part B 2013; 84B: 248 254. In individuals infected with human immunodeficiency virus (HIV), the enumeration of absolute CD4 þ Tlymphocytes is an essential parameter for monitoring and reducing the risk of disease progression (1). In order to effectively manage the patient, it is important to accurately define absolute CD4þ T lymphocyte levels that have clinical and therapeutic relevance (2 5). This is particularly important in the decision to begin therapeutic interventions and to provide prognosis (1). Guidelines for *Correspondence to: Liam Whitby, UK NEQAS for Leucocyte Immunophenotyping, 4th Floor, Pegasus House, 463a Glossop Road, Sheffield S10 2QD, United Kingdom. E-mail: liam.whitby@ukneqasli.co.uk Received 18 December 2012; Revision 11 March 2013; Accepted 26 March 2013 Published online 23 April 2013 in Wiley Online Library (wileyonlinelibrary.com). DOI: 10.1002/cyto.b.21094 VC 2013 International Clinical Cytometry Society

DATA MEASUREMENT LIMITS IN CD4 þ T LYMPHOCYTE FLOW CYTOMETRIC ENUMERATION 249 HIV infected adults advise that highly active anti-retroviral therapy (HAART) be commenced once the absolute CD4 þ T lymphocyte count falls below 350 cells per microlitre (3 5) (UK and EU guidelines) or below 500 (2) (USA). Other CD4 þ T lymphocyte counts are used to predict higher risk of specific opportunistic infections and suggest the need for antibiotic prophylaxis (4). For example, the risk of Pneumocystis jirovecii pneumonia is increased when the absolute CD4þ Tlymphocytecount falls below 200 cells per microlitre. Therefore, identifying the limitations and inherent variations of a technique used in measuring such levels is important to avoid incorrect treatment intervention and ensure that international studies are comparable when dealing with cases of adult HIV where the use of absolute numbers of CD4 þ lymphocytes is essential. Historically, laboratories enumerating absolute CD4 þ T lymphocytes have used the dual platform (DP) (6). However, UK NEQAS and others have highlighted the limitations of this technique and have recommended that the single platform (SP) approach is the method of choice (7 11). The difference between the DP and SP techniques has been summarized in detail elsewhere (7). A study undertaken by Kunkl and colleagues identified and reported measurement error for 18 Italian laboratories using the DP approach (12). However, as the SP approach has long been established as the method of choice (7,10,11), a similar evaluation needs to be undertaken. We have previously reported the External Quality Assessment (EQA) results from laboratories that undertake absolute CD4 þ T lymphocyte enumeration in the UK NEQAS Immune Monitoring programme. Whilst the Immune Monitoring programme has been operational since 1988, the present study is restricted to data collected over a 12-year period (2001 2012), as this gave the highest mix of DP versus SP data to enable the robust calculation of confidence limits, relative error and to define the boundaries of acceptability for each method. Furthermore, we have examined in detail the use of SP Technology. In addition, we have also studied how these variations will the impact upon the clinical decisions at the therapeutic level of 350 CD4 þ T lymphocytes per microlitre. Currently the British HIV Association (BHIVA (UK)), National Institute for Health (NIH (USA)), European AIDS Clinical Society and World Health Organization (WHO) recommend commencing HAART at CD4 þ T lymphocyte levels of less than 350 cells/ll (2 5). For each sample analyzed during this period the CD4 þ T lymphocyte count was categorized as <350 cells/ll (treat) or >350 cells/ll (don t treat) based on the overall mean of results returned during the EQA send out. METHODS AND MATERIALS Two stabilized blood samples, prepared in the manner previously described (13), were dispatched bimonthly to between 260 (January 2001) and 824 (October 2012) centres worldwide participating in the UK NEQAS Immune Monitoring programme over the 12-year period 2001 to 2012 (2006 2012 for FACSCount data). The absolute CD4 þ T lymphocyte count of the samples ranged from 60 to 1759 cells/ll. All samples were supplied by the National Blood Service and consisted of surplus donated material following informed consent. Laboratories were required to determine the absolute CD4 þ T lymphocyte count (cells per microlitre) for each sample and submit results to UK NEQAS for Leucocyte Immunophenotyping via a dedicated website (www. ukneqasli.org/sampleentry). Results were analyzed using previously reported methods (14). Each centre was monitored for performance; satisfactory performance was based upon the closeness of a participant s result to the trimmed mean for the sample, as defined by the range of the standard deviation. The generation of trimmed mean and standard deviations were derived using the method developed by Healy (15), and was designed to eliminate outliers in EQA exercises. For this study, 58,626 individual data sets from a total of 143 individual send out samples issued over the 12- year period were analyzed. To ensure statistical robustness each data set contained a minimum of 30 participants in each platform subgroup. From the data, we calculated and compared the 99.9% confidence limits and the relative error at all levels of absolute CD4 þ T lymphocyte count for each technology. Following the comparison of relative errors we undertook an additional analysis of the data to assess how the identified variations would impact upon the clinical decision making processes. For each sample analyzed during this period the CD4 þ T lymphocyte count was categorized as <350 cells/ll (treat) or >350 cells/ll (don t treat) based on the overall mean of results returned during the EQA send out. RESULTS The results were stratified in relation to the trimmed mean CD4 þ T lymphocyte count (obtained from the SP user group within each EQA send out), and separated into groups depending on whether DP or SP methodology was used with the SP approach further subcategorized into the specific technology based on whether TruCount, Flow-Count, or FACSCount approaches were used. These were analyzed to calculate the relative errors and 99.9% confidence limits for absolute CD4 þ T lymphocyte counts both as approach and technology based. A comparison of 99.9% confidence limits for dual and the main single platform technologies over the 12-year period at CD4 þ T lymphocyte levels that correspond to HAART initiation is shown in Figure 1 (note FACSCount not shown on graph due to having a shorter analysis period). All three SP technologies studied have significantly smaller 99.9% confidence limits when compared to the dual platform technique (P < 0.0001) indicating that a SP derived result is more precise than one obtained using the DP method. Furthermore, it would appear that the technique with the least amount of sample

250 WHITBY ET AL. FIG. 1. 99.9% Confidence limits for CD4 þ T lymphocyte counts by single and dual platform technologies. Table 1 99.9% Confidence Limits for CD4 þ T Lymphocyte Counts at a Level of 350 Cells per Microlitre Method Lower 99.9% confidence limit Upper 99.9% confidence limit Dual platform 302 398 96 Flow-count 304 363 59 (single platform) TruCount 341 378 37 (single platform) 330 370 40 FACSCount (single platform) Confidence limit range manipulation (FACSCount) had a consistent relative error of 7% across the full range of CD4 þ T lymphocyte levels studied for the period 2006 onwards. Results at the clinically significant absolute CD4 þ T lymphocyte level of 350 cells per microlitre (defined by results of all participants), below which guidelines say antiretroviral therapy commencement is recommended (2 5), were examined in more detail. The confidence limit range for each technique were 96 cells per microlitre for dual platform, 59 cells per microlitre for Flow- Count, 40 cells per microlitre for FACSCount and 37 cells per microlitre for TruCount. This means that for a given result of 350 cells per microlitre when using dual platform the true result would lie between 302 and 398 cells per microlitre, 304 363 cells per microlitre when using Flow-Count, and 330 370 and 341 378 cells per microlitre for FACSCount and TruCount system respectively. The 99.9% confidence limits for each of the techniques at this level are shown in Table 1. We then compared our data with a previously published study that reported the relative error for CD4 þ T lymphocyte enumeration for dual platform methods, albeit using a significantly smaller data set (18 sites and 24 samples) (12). The relative errors for the data returned to UK NEQAS and this previous study (12) are shown in Figure 2. Both studies are comparable over the entire range of CD4 þ T lymphocyte counts examined, having a Spearman correlation value (r) of 0.95. However, this comparison also highlighted that single platform techniques have a lower relative error compared to the dual platform approach. Following the comparison of relative errors, an additional analysis of the data was performed to determine the effect that the identified variations would have on clinical decision making processes. The results of this analysis are shown in Figure 3. Below the level of 350 cells/ll it was found that there was no significant difference (P ¼ 0.2) in cases where treatment would be given, irrespective of whether the analysis was performed using dual platform or single platform analysis. However, for samples above the level of 350 cells/ll, laboratories using dual platform techniques would classify 20% of samples incorrectly as being below the 350 cells/ll level, significantly more than laboratories using single platform (P ¼ 0.04) and therefore could result in patients receiving treatment too early. DISCUSSION Previous studies have repeatedly shown that single platform technologies consistently outperform dual platform technology in the enumeration of CD4 þ T lymphocytes (7,9,10,11). However, what is not known is the effect these technologies have upon the clinical and therapeutic use of CD4 þ T lymphocytes and how misclassification of the results impacts upon its use. A previous study by Kunkl et al. (12) reported the relative error from a cohort of 18 laboratories using dual platform in the determination of CD4 þ T lymphocytes, however there is no data assessing this with single platform technologies. To address this we have reviewed 12 years of UK NEQAS data collected from up to 824 international laboratories that yielded 58,626 data points. We then

DATA MEASUREMENT LIMITS IN CD4 þ T LYMPHOCYTE FLOW CYTOMETRIC ENUMERATION 251 FIG. 2. Relative errors for CD4 þ T lymphocyte counts using differing technologies. FIG. 3. Treatment administration for different absolute counting techniques at differing levels of CD4 þ T lymphocytes. compared single platform with dual platform technology and examined the effect upon: relative error, technology confidence limits and the impact results obtained by these different technologies have upon the misclassification of the 350 CD4 þ T lymphocytes/ll clinical therapeutic threshold. We used the same 99.9% confidence limits as used in the earlier work by Kunkl et al. (12) to enable direct comparison to this previous study. This current study shows, for the first time, that the 99.9% confidence limits obtained using single platform technologies are superior to those obtained with dual platform methods. All of the single platform technologies examined gave a smaller range of confidence limits across all CD4 þ T lymphocyte counts when compared to the dual platform method. In addition, we have shown that the data from this study shows a high degree of concordance with the previous study of Kunkl et al. (12) for dual platform approaches, despite the fact that they used fresh samples and our study used stabilized whole blood. This confirms that stabilized blood behaves in an identical manner (but without the degradation effects) as fresh blood (note the study conducted by Kunkl et al. was performed within 24 hours of sample draw). However, and of significance, the difference

252 WHITBY ET AL. in confidence limits at clinically significant levels was such that results derived using dual platform methods were more likely to adversely influence patient treatment compared to results generated by single platform technology by potentially causing an earlier initiation of HAART. A divergence of confidence limits was observed for the Flow-Count system and was especially noticeable above the 1000 cells/ll CD4 þ T lymphocyte level (not shown) but these levels are above those used in clinical management of HIV þ individuals and thus will not be discussed further. However at clinically relevant levels of CD4 þ T lymphocytes the confidence limits of Flow- Count are within those generated by the dual platform users. It is most likely that the explanation lies in the difference between the two single platform techniques. TruCount features a lyophilized pellet in the tube, to which a set volume of blood is added. This method has only two manipulation steps of the sample adequate mixing and accurate blood aliquoting. However, Flow- Count uses a vial of beads that must be premixed and then a set volume of these must be added to a set volume of blood. There are four manipulation steps of the sample using this method sample mixing, accurate blood aliquoting, bead mixing, and accurate bead aliquoting. In any analytical method every step has an inherent error that cannot be avoided, and the error of every step in a process combines to create the compound error for the method. It is, therefore, logical to keep the manipulation steps to a minimum to avoid generating a larger compound error. It is interesting to note that whilst the relative error of TruCount is lower than that seen with FACSCount, at the 350 CD4 þ T lymphocyte cells/ll level, this could be explained by the fact that the FACSCount requires no operator intervention compared to the TruCount and that the FACSCount instrument is probably being used as a point of care instrument by individuals who have a lower skill level. Thus, TruCount operators are able to intervene to reanalyze any suspect data, particularly important at critical decision making points. With four manipulation steps Flow- Count has a larger potential compound error than FACS- Count and TruCount which have only two manipulation steps. It seems likely that this is the cause for the slightly higher confidence limits observed in Flow-Count users and the divergence of Flow-Count results from FACSCount and TruCount users at higher CD4 þ T lymphocyte levels. However, where targeted training is used Flow-Count has been shown to reduce this variability (particularly in CD34 þ Stem Cell counting) and give results comparable to those sites using TruCount approach (16). The observed differences in the confidence limits highlight that there is significant variability of the absolute CD4 þ T lymphocyte counts generated using different technologies and that this may then have an impact on patient treatment. Therefore, centres should tailor their analytical acceptable limits on the techniques in use, a factor that should also be taken into account when performing comparison studies between centres using different technologies to generate absolute counts. In addition centres taking part in clinical trials may have to adjust their target ranges appropriately to correct for the effect that their technique variation is having on results. External quality assurance agencies will have to be aware of the differences in results that can be produced by different technologies, and any performance monitoring performed by EQA organizations will have to take this into account. When comparing the relative errors inherent to dual and single platform technologies an additional comparison was made to a previous study, where the relative errors for dual platform users generating CD4 þ T lymphocyte counts in an EQA programme were calculated (12). Data obtained from both of the single platform methods studied and from dual platform users was compared to this previously published data. From the data comparison the initial finding was that there was no significant difference between the relative errors of dual platform users, obtained using the results returned by UK NEQAS participants, and the relative errors obtained previously for dual platform users (12). As our studies used stabilized blood and the earlier study used fresh blood this data shows that there is no significant difference (P ¼ 0.0214) in the choice of material used in the EQA scheme operation. However the use of stabilized material does have significant benefits over fresh material such as increased antigen stability over time and making logistics of sample shipment easier (13). In addition the correlation of the dual platform users in this study to the previous study also shows that since 2002, when the initial study was published, there has been no improvement in the generation of absolute CD4 þ T lymphocyte counts by dual platform technologies. This is an expected finding because, although new haematology analyzers have been launched, the problems inherent in dual platform technologies are still in place whatever the age of the haematology analyzer. It should be noted that the samples produced by UK NEQAS are optimized for flow cytometry but not for use in routine haematology analyzers. As such there is the potential for bias in this aspect as a sub optimal performance of the samples on a DP methodology could be the reason for the apparent difference between this and SP techniques. However, given that the results (as expressed as confidence limits) for the UK NEQAS DP data have no significant difference to the previous study by Kunkl et al. s DP data (12) it would suggest that in this study the use of a stabilized material has made no difference to the outcomes and findings. The third major finding from the comparison was that much lower relative errors are obtained by users of SP methods when compared to DP users. This was irrespective of the level of the absolute CD4 þ T lymphocyte cell count; this study shows that single platform consistently outperformed dual platform methods, returning relative errors over half those seen in dual platform. The lowest relative errors seen were obtained with TruCount users and although the relative error was higher for

DATA MEASUREMENT LIMITS IN CD4 þ T LYMPHOCYTE FLOW CYTOMETRIC ENUMERATION 253 users of Flow-Count, they were still lower than those seen by users of dual platform techniques. Interestingly, the relative error for FACSCount was a consistent 7% across the entire range of CD4 þ T lymphocyte levels examined for the six years of the FACSCount data sets (60 cells/ll to 1368 cells/ll). When analysis was performed on the data where consensus results (irrespective of platform) were below the international treatment threshold of 350 cells/ll, we found that there was no significant misclassification between either SP or DP technologies meaning most patients would receive the correct treatment at the correct time. However, when we examined results that were above the treatment level of 350 cells/ll a significant difference in misclassifications (P ¼ 0.04) was seen between the DP and SP results. This meant that patients being monitored using DP technology were more likely to start HAART therapy prematurely than those being monitored using SP technology. Whilst several international organizations (3 5) have recommended the use of a 350 cells/ll cut-off for the initiation of HAART, others have recommended beginning earlier at a level of 500 cells/ll (2). Also limitations in the availability of drugs due to cost factors has led several resource poor countries routinely use a limit of less than 350 cells/ll, thus delaying initiation of therapy. Furthermore, where the possibility of coinfection exists with either hepatitis B or hepatitis C then treatment before the 350 cells/ll limit is reached has been recommended (4). Thus, whilst the focus of this study has been on the main therapeutic cut off limit of 350 cells/ ll, it should be noted that the factors relating to confidence limits in DP and SP methods exist across the entire range of CD4 þ T lymphocyte levels examined (Fig. 1). As such the findings of this study can be applied to any CD4 þ T lymphocyte level experienced in routine testing of patients. Whilst, the degree of variability will alter, the underlying message of SP methods having better confidence limits than the DP methods remains unchanged. One interesting point to note here is that if a laboratory/clinician is using a DP approach to monitor patient CD4 þ T lymphocyte levels then it is feasibly possible that the rate of progression will remain undetected using DP compared to SP approach because the variability is wider and therefore less sensitive to change. This would then have the effect of making the monitoring of disease progression more difficult for clinicians and may have adverse outcomes for any patients so affected. Alternatively the wider confidence limits of DP technologies could show changes in CD4 þ T lymphocyte levels when no such alteration was occurring (e.g. a false fall in CD4þ T lymphocyte values), leading to unnecessary clinical interventions. As such, centres performing CD4 þ T lymphocyte monitoring using DP methods will experience the limitations of this technique throughout the course of monitoring patients. Finally, this study has focussed upon the use of absolute values and the uncertainty for commencement of treatment based upon the therapeutic decision making points of such values. It is acknowledged that in certain instances, for example paediatrics, it is often advisable to use percentage values. Whilst this study has not examined such variability of percentage values and this is the subject of a further study, early anecdotal evidence from our EQA program (unpublished observations and data not shown) would suggest that there is very little difference between percentage values irrespective of which platform (SP or DP) is used. This is due to the fact that determination of percentage values is undertaken only on the flow cytometer and does not have any external influence, other than the gating strategy used. We have also previously shown that the variability in absolute counts derived by SP versus DP is due to the use of the total leucocyte count used by DP approaches to generate the absolute count (7) and not the determination of the percentage value as this is an independent variable. In summary it is recommended that laboratories consider changing their technologies to utilize a single platform approach to obtain the benefits outlined and decrease the impact that dual platform will have on treatment modalities. ACKNOWLEDGMENTS The authors would like to acknowledge the participants in the UK NEQAS Immune Monitoring programme, as without their continued support and EQA results this paper would not have been possible. LITERATURE CITED 1. Andrew NP, Jens DL. The CD4 lymphocyte count and risk of clinical progression. Curr Opin HIV AIDS 2006;1:43 49. 2. Panel on Antiretroviral Guidelines for Adults and Adolescents. Guidelines for the use of antiretroviral agents in HIV-1-infected adults and adolescents. Department of Health and Human Services. Available at http://aidsinfo.nih.gov/contentfiles/lvguidelines/adultandadolescentgl.pdf Section accessed 01/10/2012. 3. World Health Organization. Antiretroviral therapy for HIV infection in adults and adolescents. Geneva: WHO, 2010. 4. Nelson M, Dockrell DH, Edwards S on behalf of the BHIVA Guidelines Subcommittee. British HIV Association and British Infection Association guidelines for the treatment of opportunistic infection in HIV-seropositive individuals 2011. HIV Med 2011;12 (Suppl 2): 1 5. 5. European AIDS Clinical Society. Guidelines. Version 6- October 2011. 6. Higgins J, Hill V, Lau K, Simpson V, Roayaei J, Klabansky R, Stevens RA, Metcalf JA, Baseler M. Evaluation of a single platform technology for lymphocyte immunophenotyping. Clin Vaccine Immunol 2007;14:1342 1348. 7. Barnett D, Granger V, Whitby L, Storie I, Reilly JT. Absolute CD4þ T-Lymphocyte and CD34þ stem cell counts by single platform flow cytometry: The way forward. BJH 1999; 106:1059 1062. 8. Robinson G, Morgan L, Evans M, McDermott S, Pereira S, Wansbrough-Jones M, Griffin G. Effect of type of haematology analyser on CD4 count. Lancet 1992;340:485. 9. Connelly MC, Knight M, Giorgi JV, Kagan J, Landay AL, Parker JW, Page E, Spino C, Wilkening C, Mercolino TJ. Standardization of absolute CD4þ lymphocyte counts across laboratories: An evaluation of the ortho CytoronAbsolute flow cytometry system on normal donors. Cytometry 1995;22:200 210. 10. Schnizlein-Bick CT, Spritzler J, Wilkening CL, Nicholson JKA, O Gorman MRG. Evaluation of TruCount absolute-count tubes for determining CD4 and CD8 cell numbers in human immunodeficiency virus positive adults. Clin Diagn Lab Immunol 2000;7:336 343. 11. Reimann KA, O Gorman MRG, Spritzler J, Wilkening CL, Sabath DE, Helm K, Campbell DE. Multisite comparison of CD4 and CD8 T-lymphocyte counting by single- versus multiple-platform methodologies: Evaluation of Beckman Coulter flow-count fluorosperes and the tetraone system. Clin Diagn Lab Immunol 2000;7:344 351.

254 WHITBY ET AL. 12. Kunkl A, Risso D, Terranova MP, Girotto M, Brando B, Mortara L, Lantieri PB. Grading of laboratories on CD4þ T-lymphocyte evaluations based on acceptable data boundaries defined by the measurement error. Cytometry 2002;50:117 126. 13. Barnett D, Granger V, Mayr P, Storie I, Wilson GA, Reilly JT. Evaluation of a novel stable whole blood quality control material for lymphocyte subset analysis: Results from the UK NEQAS immune monitoring scheme. Cytomtery 1996;26:216 222. 14. Whitby L, Granger V, Storie I, Goodfellow K, Sawle A, Reilly JT, Barnett D. Quality control of CD4þ T-lymphocyte enumeration: Results from the last 9 years of the United Kingdom national external quality assessment scheme for immune monitoring (1993 2001). Cytometry 2002;50:102 110. 15. Healy MJR. Outliers in clinical chemistry quality control schemes. Clin Chem 1979;25:675 677. 16. Barnett D, Granger V, Kraan J, Whitby L, Reilly JT, Papa S, Gratama JW. Reduction of intra- and interlaboratory variation in CD34 stem cell enumeration by the use of stable test material, standard protocols and targeted training. Br J Haematol 2000; 108:784 792.