TM TM Surgical Technique

Similar documents
nva Anterior Lumbar Interbody Fusion System

nvt Transforaminal Lumbar Interbody Fusion System

nvp Posterior Lumbar Interbody Fusion System

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10

Zimmer Anterior Buttress Plate System. Surgical Technique

Zimmer Facet Screw System Surgical Technique

Table of Contents.

A U X I L I A R Y C O N N E C T O R S Surgical Technique

UniLink Interspinous Fusion System UniLink 5/1 Interspinous Fusion System UniLink Graft. Surgical Technique

INTELLIGENT SPINAL SYSTEM

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE

VTI INTERLINK PEDICLE SCREW SYSTEM

TABLE OF CONTENTS. ShurFit Anterior Cervical Interbody Fusion (ACIF) System Overview 2. Implant Specifications 3. Instrument Features 4

SURGICAL TECHNIQUE GUIDE TRESTLE. Anterior Cervical Plating System

Veyron -C Anterior Cervical System Surgical Technique

OPERATIVE TECHNIQUE. anterior cervical plating system

Royal Oak Cervical Plate System

OPERATIVE TECHNIQUE. CONSTRUX Mini PTC. Mini PTC Spacer System

Alamo C. Cervical Interbody System Surgical Technique. An Alliance Partners Company

Apache Cervical Interbody Fusion Device. Surgical Technique. Page of 13. LC-005 Rev F

Alamo T Transforaminal Lumbar Interbody System Surgical Technique

Y o u r Id e a s En g i n e e r e d t o Li f e

HawkeyeTM Peek. surgical technique

Pinit Plate Small Bone Fusion System Bone Plate & Screw System

LBL-014. Rev

Lapidus Arthrodesis System Instructions for Use

ACP1 CERVICAL PLATE SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE II.

X-spine Systems, Inc. Axle Interspinous Fusion System

Technique Guide KISSloc Suture System

Thunderbolt. surgical technique. MIS Pedicle Screw System. Where Nimble and Secure Intersect

EXACTECH SPINE. Operative Technique. Cervical Spacer System. Surgeon focused. Patient driven. TM

Aesculap CeSpace TM PEEK and CeSpace TM XP Spinal Implant System Instructions for Use

Surgical Technique. Apache Anterior Lumbar Interbody Fusion

TiLock 2 Spinal System. Surgical Technique

X-spine Aranax Anterior Cervical Plating System Instructions for Use / Package Insert

Technique Guide Small Bone Fusion System

Valencia Pedicle Screw Surgical Technique

TiLock XT Minimally Invasive Surgery (MIS) Pedicle Screw System

HydraLok. Operative Technique. Polyaxial Pedicle Screw System

Cervical Spacer System surgical technique

3 Operative Technique U.S. EDITION

Ascent. Surgical Technique. Breakthrough Thinking. Posterior Occipital Cervico-Thoracic (POCT) System. Germany U.S.A.

Royal Oak IBFD System Surgical Technique Posterior Lumbar Interbody Fusion (PLIF)

InFix. Anterior Lumbar Device. Surgical Technique Guide

Surgical Technique. Apache Posterior Lumbar Interbody Fusion Apache Transforaminal Lumbar Interbody Fusion

BONE GRAFT WASHER _Rev A IMPORTANT INFORMATION ON THE BONE GRAFT WASHER

TiLock XT Minimally Invasive Surgery (MIS) Pedicle Screw System

EXCELLA ll. Spinal System


X-spine Systems, Inc. Silex Sacroiliac Joint Fusion System

X-spine Surgical Technique

Hinged Laminoplasty System surgical technique

Gallery Laminoplasty Spine System

Cervical Screw System

Thoracolumbar Solutions. Sacroiliac Joint Fusion System. Surgical Technique Guide

100 Interpace Parkway Parsippany, NJ

SURGICAL TECHNIQUE GUIDE SOLANAS. Posterior Cervico-Thoracic Fixation System Adjustable Bridge System

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

Threshold Pedicular Fixation System Surgical Technique

product overview Implant heights range from 8mm-20mm in 2mm increments, with two lordocic angle options of 6 and 12.

C-THRU Anterior Spinal System

Cervical Solutions. Optio-C Anterior Cervical PEEK. Interbody System. Surgical Technique Guide

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

Asnis. Micro Cannulated screw system. Xpress operative technique

Thoracolumbar Solutions. Aspen. MIS Fusion System. Surgical Technique Guide

ASFORA ANTERIOR CERVICAL PLATE SYSTEM (AACP) SURGICAL PROCEDURE MANUAL

SURGICAL TECHNIQUE. SECURIS Pedicle Screw System for Minimally Invasive Surgery. 2 I SECURIS Pedicle Screw System

Trinica and Trinica Select Anterior Cervical Plate System

X-spine Systems, Inc. Certex Spinal Fixation System

O PE RATIV E TE C HN IQ U E. Minimally Invasive Posterior Fixation

X-spine Systems, Inc. Spider Cervical Plating System

BAK/C Cervical Anterior Interbody Fusion System

Surgical Technique Guide

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine.

GIZA Surgical Technique

Cervical Solutions. Optio-C Anterior Cervical Plate. with Allograft/Autograft. Surgical Technique Guide

EFSPINE CERVICAL COMBINED SET DISC PROTHESIS ORGANIZER BOX

2. Active systemic infection or infection localized to the site of the proposed implantation is contraindications to implantation.

I. II. SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine.

Instructions for Use Reform POCT System Caution: Federal (USA) law restricts this device to sale by or on the order of a physician

PITBULL. Interspinous Process Decompressiong Device

Tempus Anterior Cervical Plate System

MetaFix Ludloff Plate

DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

3.5 mm LCP Olecranon Plates

Solitaire Anterior Spinal System

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique. Hip Knee Spine Navigation

O PE RATIV E TE C HN IQ U E. ProView. Expandable Retractor System U.S. EDITION

Dymaxeon Spine System. Simple, Streamlined, Smart. Surgical Procedure

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L.

EasyStep. Operative technique

A Fusion of Versatility, Performance and Ease.

X-spine Systems, Inc. Irix-A Lumbar Integrated Fusion System

Olecranon Locking Plate II

Advantage ALIF. Keith Shevlin Managing Director

L-VARLOCK. Posterior Lumbar Cage with adjustable lordosis. S urgical T echnique

Transcription:

TM TM Surgical Technique

TABLE OF CONTENTS Reli SP Spinous Plating System Overview Device Description Implant Features Indications Instruments Access Instruments Preparation Instruments Insertion Instruments Locking Instruments Surgical Technique 1. Implant Insertion Overview 2. Patient Positioning 3. Site Identification & Incision 4. Insertion Technique 5. Insertion Technique 6. Compression 7. Set Screw Insertion 8. Final Tightening 9. Implant Removal Implant & Instrument Information Indications 3 3 3 3 4 4 5 5 6 7 7 7 7 8 9 10 10 10 10 11 12 2

RELI SP SYSTEM OVERVIEW Device Description: The Reli SP Spinous Plating System of Precision Spine, Inc. is a posterior, non-pedicle supplemental fixation device to facilitate fusion. Various sizes of these implants are available so that adaptations can be made to take into account pathology and individual patients. All implants are manufactured from titanium alloy per ASTM F-136. All instrument components are made from medical grade stainless steel, titanium or titanium alloy, and aluminum, which comply with such standards as ASTM F-138, ASTM F-136, ASTM B-209, ISO5832-1 or ISO5832-3. All components are supplied clean and NON STERILE All implants are intended for single use only and should not be reused under any circumstances. Implant Features: Two-piece design minimizes disruption and keeps the spinous ligament intact Each plate has 6 Pyramidal Cleats to increase spinous process fixation Square Thread Locking Cap reduces potential for crossthreading 28mm, 35mm, 45mm and 55mm plate sizes Indications: The Reli SP Spinous Plating System of Precision Spine, Inc is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1) of skeletally mature patients. It is intended for single level plate fixation/ attachment to spinous process for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (i.e. fracture or dislocation), spondylolisthesis, and/or tumor. It is not intended for standalone use. Please refer to package insert (LBL-IFU-019) for complete system description, indications and warnings. 3

INSTRUMENTS ACCESS INSTRUMENTS Tissue Awl Used to create hole through anterior region of the interspinous ligament Part Number: 29-0007 Caspar Pin Driver Used to insert appropriate size Distraction Pins (ACP-P1X) in the appropriate spinous processes Part Number: 29-0801 Caspar Pin Sleeves Slide over the Distraction Pins (ACP-P1X) to permit application of the Caspar Retractor (29-0800) Part Numbers: 29-0960 (60mm, 2x); 29-0975 (75mm, 2x); 29-0990 (90mm, 2x) Distraction Pins Inserted into the spinous processes Part Numbers: ACP-P12 (12mm, 4x); ACP-P16 (16mm, 4x) Caspar Retractor Slides over the Caspar Sleeves (29-09XX) and opened to the desired level of distraction Part Number: 29-0800 Tissue Spreader Inserted into the hole created by the Tissue Awl (29-0007) and used to spread the spinous processes Part Number: 29-0001 4

INSTRUMENTS PREPARATION INSTRUMENTS Trials Used along the lateral aspect of the spinous processes, along with lateral fluoroscopy to determine the proper size of the implant Part Numbers: 29-0128 (28mm, 2x); 29-0135 (35mm, 2x); 29-0145 (45mm, 2x); 29-0155 (55mm, 2x) INSERTION INSTRUMENTS Plate Holder Can be used on the Interspinous Female Plate (29-00XX-02) and/or the Interspinous Male Plate (29-00XX-01) for assembly Part Number: 29-0900 (x2) Straight Plate Holder Threaded to the Interspinous Female Plate (29-00XX-02) so that it can be mated onto the cross bar of the Interspinous Male Plate Part Number: 29-0901 Compressor Inserted onto plates by aligning distal tips to the holes of the Interspinous Plate Part Number: 29-0200 (x2) 5

INSTRUMENTS LOCKING INSTRUMENTS Torque Limiting Handle Attaches to the Set Screw Final Tightener (04-9097) to assist in tightening the Plate Set Screw to 30 in-lbs Part Number: 04-9023 Counter Torque Handle Used as a guide to insert the Plate Set Screw (04-1000) and as a counter torque for final tightening Part Number: 29-0006 Set Screw Inserter Used to thread the Set Screw (04-1000) into the Interspinous Female Plate Part Number: 04-9094 Set Screw Final Tightener Attaches to the Torque Limiting Handle (04-9023) to tighten the Plate Set Screw until handle clicks at 30 in-lbs Part Number: 04-9097 (x2) 6

SURGICAL TECHNIQUE 1. Implant Insertion: The technique for using the Reli SP plate implant includes these basic steps: 1. Loading the implant onto the instrument 2. Inserting it into the surgical site at the appropriate level 3. Compressing 4. Provisional locking 5. Final locking Two minimally-disruptive instrumentation options are available for tailoring the surgical technique to surgeon preference and the unique characteristics of the patient s anatomy. Insertion Technique 1 Keeps the interspinous ligament intact Inserts and assembles the implant through the interspinous ligament Insertion Technique 2 Removes the interspinous ligament Inserts a preassembled plate construct with an instrument 2. Patient Positioning: Figure 1 The patient should be placed in the prone position on the operating table (Fig. 1). 3. Site Identification and Incision: Identify the spinous processes at the level to be instrumented using manual palpation and intraoperative imaging. Make a midline incision (3-5cm in length) to expose the spinous process at the correct level (Fig. 2). Figure 2 7

INSTRUMENTS 4. Insertion Technique 1 1. Ligament Puncture Using the Tissue Awl (29-0007), create a hole though the anterior region of the interspinous ligament. Make sure the hooked awl is placed at the midpoint between the adjacent spinous processes (Fig. 3). Figure 3 2. Distraction Two different types of spinous process distraction instruments are offered. Option 1: Insert the Tissue Spreader (29-0001) into the hole created by the hooked awl and spread the spinous processes. Begin to distract adjacent spinous processes (Fig. 4). Figure 4 Option 2: Insert the appropriate size Distraction Pins (ACP-P1X) in the appropriate spinous processes using the Caspar Pin Driver (29-0801) (Fig. 5). Slide the appropriate length Caspar Sleeves (29-09XX) over the Distraction Pins. Slide the Caspar Retractor (29-0800) over the Caspar Sleeves and then open to the desired level of distraction (Fig. 6). Figure 5 3. Trialing Use the appropriate size Trial (29-01XX) along the lateral aspect of the spinous processes and use lateral flouroscopy to determine the proper size (Fig. 7). Plate size should be based on maximum surface area coverage of both spinous processes (Fig. 8). It is important to NOT oversize the plate avoid plate extension beyond the superior margin of the superior spinous process and the inferior margin of the inferior spinous process. Figure 6 Figure 7 Figure 8 8

SURGICAL TECHNIQUE 4. Male Plate Insertion Assemble Plate Holder (29-0900) onto the Interspinous Male Plate (29-00XX-01) (Fig. 9). Using the plate holder, insert the cross bar of the Interspinous Male Plate through the interspinous ligament (Fig. 10). Figure 9 Figure 10 5. Female Plate Insertion Thread Straight Plate Holder (29-0901) to the Interspinous Female Plate (29-00XX-02) (Fig. 11). Mate the Interspinous Female Plate onto the cross bar of the Interspinous Male Plate (Fig. 12). After plates are in the desired position, remove the plate holders. Figure 11 Alternately, a second Plate Holder (29-0900) can be used on the Interspinous Female Plate to assemble it to the Interspinous Male Plate. Figure 12 5. Insertion Technique 2 1. Pre-assembled Plate Assemble appropriate size Interspinous Male and Female Plate and lock into desired position with Straight Plate Holder (29-0901). With interspinous ligament removed, insert plate assembly into desired position (Fig. 13). Distance between the plates can be adjusted by loosening the plate holder and sliding the plates away from each other. Once proper placement is achieved, remove plate holder. Figure 13 9

SURGICAL TECHNIQUE 6. Compression After implant is in desired position, insert Compressors (29-0200) onto plates by aligning distal tips of the Compressor to the holes of the Interspinous Plate (Fig. 14). Begin compression by either squeezing both Compressors simultaneously or alternating back and forth, to ensure the spikes seat properly in both the inferior and superior spinous processes. Remove Compressors once desired position and compression is achieved. Figure 14 7. Set Screw Insertion Using the Counter-Torque handle (29-0006) as a guide, insert the Plate Set Screw (04-1000) (Fig. 15) using the Set Screw Inserter (04-9094) (Fig. 16) and thread into Interspinous Female Plate. Hand tighten once desired position is confirmed. Figure 15 Figure 16 8. Final Tightening With the Counter-Torque handle still in place, use the Set Screw Final Tightener (04-9097) attached to the 30 in-lb Torque Limiting Handle (04-9023) to tighten the Plate Set Screw until handle clicks (Fig. 17). For additional counter torque, the Counter-Torque handle may be used alone or with the Compressors still in place. Or, as an alternative to using the Counter Torque handle, the compressors may be used alone to provide counter torque during final tightening. Figure 17 Figure 18 Figure 19 9. Implant Removal Using the Counter-Torque handle as a guide, insert the Set Screw Final Tightener, coupled with the Torque Limiting Handle (Fig. 18), into the head of the set screw (Fig. 19). Turn counterclockwise to remove the set screw from the plate. Figure 20 Thread Straight Plate Holder into Interspinous Female Plate (Fig. 20). Grab Interspinous Male Plate with Plate Holder. Move Plate Holders laterally until the male plate rod has disengaged from the female plate (Fig. 21). Remove plate from patient. Figure 21 10

IMPLANTS & INSTRUMENTS Implants Instruments Item No. Description Item No. Description 04-1000 Plate Set Screw 29-0028-01 28mm Interspinous Male Plate * 29-0035-01 35mm Interspinous Male Plate 29-0045-01 45mm Interspinous Male Plate 29-0055-01 55mm Interspinous Male Plate 29-0028-02 28mm Interspinous Female Plate * 29-0035-02 35mm Interspinous Female Plate 29-0045-02 45mm Interspinous Female Plate 29-0055-02 55mm Interspinous Female Plate Standard Size Plates 29-6000-CA IMPLANT & INSTRUMENT KIT 04-9023 Torque Limiting Handle 04-9094 Set Screw Inserter 04-9097 Set Screw Final Tightener 29-0001 Tissue Spreader 29-0006 Counter Torque 29-0007 Tissue Awl 29-0128 28mm Trial * 29-0135 35mm Trial 29-0145 45mm Trial 29-0155 55mm Trial 29-0200 Compressor 29-0900 Plate Holder 29-0901 Straight Plate Holder Item No. Description 29-6001-CA RETRACTOR KIT 29-0800 Caspar Retractor 29-0801 Caspar Pin Driver 29-0960 60mm Caspar Sleeve 29-0975 75mm Caspar Sleeve 29-0990 90mm Caspar Sleeve 35mm 45mm 55mm ACP-P12 ACP-P16 12mm Distraction Pin 16mm Distraction Pin * Special Order 11

INDICATIONS Contraindications: The Reli SP Spinous Plating System contraindications include, but are not limited to: 1. Morbid obesity 2. Mental Illness 3. Alcoholism or drug abuse 4. Fever or leukocytes 5. Pregnancy 6. Severe osteopenia 7. Metal sensitivity/allergies 8. Patients unwilling or unable to follow post-operative care instructions 9. Active infectious process or significant risk of infection 10. Any circumstances not listed in the indication of the device Potential Adverse Effects: All of the possible adverse effects associated with spinal fusion surgery without instrumentation are possible. With instrumentation, a listing of potential adverse events includes, but is not limited to: 1. Non-union 2. Fracture of the vertebra 3. Neurological injury 4. Vascular or visceral injury 5. Early or late loosening of any or all of the components 6. Loss of fixation 7. Device component fracture 8. Foreign body (allergic) reaction to implants, debris, corrosion products, graft material, including metallosis, straining, tumor formation, and/or autoimmune disease 9. Disassembly and/or bending of any or all of the components 10. Infection 11. Hemorrhage 12. Change in mental status 13. Pressure on the skin from component parts in patients with inadequate tissue coverage over the implant possibly causing skin penetration, irritation, and/or pain 14. Pain, discomfort, or abnormal sensations due to the presence of the device 15. Post-operative change in spinal curvature, loss of correction, height, and/or reduction 16. Cessation of any potential growth of the operated portion of the spine 17. Loss of or increase spinal mobility or function 18. Death Note: Additional surgery may be required to correct some of these potential adverse events. Precautions & Warnings: The Reli SP Spinous Plating System should be implanted only by surgeons who are fully experienced in the use of such implants and the required specialized spinal surgery techniques. Knowledge of surgical techniques, proper reduction, selection and placement of implants, and pre- and postoperative patient management are considerations essential to a successful surgical outcome. Appropriate selection, placement and fixation of the spinal system components are critical factors which affect implant service life. Accordingly, strict adherence to the indications, contraindications, precautions, and warnings for this product is essential to potentially maximize service life. (Note: While proper implant selection can minimize risks, the size and shape of human bones present limitations on the size, shape, and strength of the implants). Patients who smoke have been shown to have an increased incidence of pseudoarthrosis. Such patients should be advised of this fact and warned of the potential consequences. Patients with previous spinal surgery at the level to be treated may have different clinical outcomes compared to those without a previous surgery. Based on the fatigue testing results, the physician/ surgeon should consider the level of implantation, patient weight, patient activity level, and other patient conditions, etc. which may have an impact on the performance of the system. If the patient is involved in an occupation or activity which applies inordinate stress upon the implant (e.g. substantial walking, running, lifting, or muscle strain) resultant forces can cause failure of the device. In some cases, progression of degenerative diseases may be so advanced at the time of implantation that the expected useful life of the appliance may be substantially decreased. In such cases, orthopedic devices may be considered only as a delaying technique or to provide temporary relief. Patients should be instructed in detail about limitations of the implant, including, but not limited to, the implant of excessive loading through patient weight or activity, and be taught to govern their activities accordingly. The patient should understand that a metallic implant is not as strong as normal, healthy bone and will bend, loosen, or fracture if excessive demands are placed on it. An active, debilitated, or demented patient who cannot properly use weight supporting devices may be particularly at risk during postoperative rehabilitation. Care must be taken to protect the components from being marred, nicked or notched as a result of contact with metal or abrasive objects. Alterations will produce defects in surface finish and internal stresses which may become the focal point for eventual breakage of the implant. As with all orthopedic and neurosurgical implants, none of the Reli SP Spinous Plating System components should ever be reused under any circumstances. Risks associated with reuse include infection, non-union (pseudarthrosis), serious patient injury or death. (continued on next page) 12

Precautions & Warnings: Due to the presence of implants, interference with roentgenographic, CT and/or MR imaging may result. The Reli SP Spinous Plating System has not been evaluated for safety and compatibility in the MR environment. The Reli SP Spinous Plating System has not been tested for heating or migration in the MR environment. It must be noted that there are several different manufacturers and generations of MRI systems available, and Precision Spine cannot make any claims regarding the safety of Precision Spine implants and devices with any specific MR system. Physician Note: The physician is the learned intermediary between the company and the patient. The indications, contraindications, warnings, and precautions given in this document must be conveyed to the patient. If requested, additional information, including surgical technique manuals, may be obtained through corporate sales representatives. 13

TM Precision Spine, Inc. 2050 Executive Drive Pearl, MS 39208 Customer Service: 1.888.241.4773 Phone: 601.420.4244 Toll Free: 877.780.4370 Fax: 601.420.5501 www.precisionspineinc.com Caution: Federal (USA) law restricts these devices to sale by or on the order of a physician. Reli and Precision Spine are trademarks of Precision Spine, Inc. Copyright 2015 Precision Spine, Inc. All rights reserved. P/N LBL-STG-026 Rev. B 09/2015