Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2000i system

Similar documents
USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

SCIEX Vitamin D 200M Assay for the Topaz System

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

ACCESS hstni SCIENTIFIC LITERATURE

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

METHOD VALIDATION: WHY, HOW AND WHEN?

Design Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...

Thyroid Function. Thyroid Antibodies. Analyte Information

Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser

1,25 Dihydroxyvitamin D

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Thyroid Function. Thyroglobulin Analyte Information

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer.

METHOD VALIDATION CASE

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

This supplementary material has been provided by the authors to give readers

Strange thyroid function tests: biological pitfall or real pathology? Fleur Wolff, LHUB-ULB, September 2017

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Grave Clinicopathologic Correlation: A Case of Hyperthyroxinemia

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Thyroid Plus. Central Thyroid Regulation & Activity. Peripheral Thyroid Function. Thyroid Auto Immunity. Key Guide. Patient: DOB: Sex: F MRN:

Free hormone estimates. Never ending story.

Thyroid Screen (Serum)

Sample Type - Serum Result Reference Range Units. Central Thyroid Regulation Surrey & Activity KT3 4Q. Peripheral Thyroid D Function mark

liquicolor (AMP Buffer, IFCC) Design Verification

Total Thyroxine ELISA (T4)

GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS

CATHERINE FILLÉE, JEAN CUMPS, JEAN-MARIE KETELSLEGERS. Cliniques Universitaires St-Luc, Université Catholique de Louvain, Brussels, Belgium SUMMARY

Limits of Liability/Disclaimer of Warranty

Mass Spectrometry Made Simple for Clinical Diagnostic Labs

Performance Characteristics of Eight Estradiol Immunoassays

Method Comparison Report Semi-Annual 1/5/2018

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

None. Thyroid Potpourri for the Primary Care Physician. Evaluating Thyroid Function. Disclosures. Learning Objectives

METHOD VALIDATION CASE

SYSTEM ACCELERATING DECISION-MAKING

Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Cardiac Assessment Controls

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

S4. Summary of the GALNS assay validation. Intra-assay variation (within-run precision)

Four rules to remember: 1) Addition, subtraction, multiplication and division of SD and/or CV.

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Endogenous antibody interference in immunoassays. Ellen Anckaert, M.D., Ph.D. Dienst Klinische Chemie en Radio-immunologie UZ Brussel

LIPASE liquicolor. Design Verification. Multipurpose Reagent

Bone and Mineral. Comprehensive Menu for the Management of Bone and Mineral Related Diseases

AIA-600II Assay Specifications TgAb Test Code 048

Immunoprotective Steroids and SHBG in Non-Treated. Hypothyroidism and their Relationship to Autoimmune Thyroid

Guide to Fulfillment of Measurement Uncertainty

Thyroglobulin Interference in the Determination of Thyroglobulin Antibody in Wash-Out Fluid from Fine Needle Aspiration Biopsy of Lymph Node

Requesting and Management of abnormal TFTs.

Thyroid Function TSH Analyte Information

Thyroid Disorders Towards a Healthy Endocrine System

Instructions for Use. Tg Antibody ELISA

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Technical Bulletin No. 161

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Technical Bulletin No. 162

Thyroid. Dr Jessica Triay November 2018

Decoding Your Thyroid Tests and Results

Status of internal quality control for thyroid hormones immunoassays from 2011 to 2016 in China

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Determination of hemoglobin is one of the most commonly

Standardization of Thyroid Function Tests

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Immunoprotective Steroids and SHBG in Non-Treated Hypothyroidism and their Relationship to Autoimmune Thyroid Disorders

Reference Intervals for Children and Adults

TSH ELISA Kit Medical Device Licence No.: 16419

ImmunoCAP Systems. Optimized for allergy testing - Precision - Accuracy - Consistency

DBC s panel of thyroid hormone immunoassays is now more comprehensive with the new Reverse T3 ELISA kit. Test Your Hormones

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

LABORATORY TESTS FOR EVALUATION OF THYROID DISORDERS

Screening for thyroid dysfunction in adults

Comparison of Five Automated Serum and Whole Blood Folate Assays

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD

The Virtues and Pitfalls of Implementing a New Test

What information on measurement uncertainty should be communicated to clinicians, and how? Mario Plebani

Abstract. Introduction. imedpub Journals

liquicolor Design Verification

Multiple studies are available concerning the use of

Fluorescence Immunoassay (FIA)

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

(Leven and Tomer, 3002). González et al, 3002). Reffubat et al, 7002). (ISPAD) 3000

TSH Receptor Autoantibody ELISA

Caris 200 Fully Automated Chemiluminescence Immunoassay (CLIA) System

Chapter 2. High-sensitivity C-reactive protein methods examined

Thyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests)

Chapter 3. Autoimmunity and Hypothyroidism: Anti-TPO antibodies in Hypothyroid Patients in Gujarat Population

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

EliKine Free Thyroxine (ft4) ELISA Kit

Human Thyroid Stimulating Hormone CLIA kit

Thyroid and Antithyroid Drugs. Munir Gharaibeh, MD, PhD, MHPE Faculty of Medicine April 2014

See external label 96 tests ULTRASENSITIVE THYROID STIMULATING HORMONE (u-tsh) TSH Ultra Sensitive

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

Transcription:

Original Article Page 1 of 7 Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2i system Andrea Padoan, Chiara Cosma, Mario Plebani Department of Laboratory Medicine, University Hospital of Padova, Padova, Italy Contributions: (I) Conception and design: All authors; (II) Administrative support: All authors; (III) Provision of study materials or patients: All authors; (IV) Collection and assembly of data: All authors; (V) Data analysis and interpretation: All authors; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors. Correspondence to: Prof. Mario Plebani. Department of Laboratory Medicine, University Hospital of Padova, Padova 35128, Italy. Email: mario.plebani@unipd.it. Background: Diagnosis and monitoring of thyroid diseases are based on laboratory testing of thyroid stimulating hormone (TSH), free triiodothyronine (FT3) and free thyroxine (FT4). Other thyroid markers, such as thyroglobulin (Tg), antibody against Tg (TgAb) and thyroperoxidase (TPOAb), are also widely utilized by physicians. The present study aimed to evaluate the analytical performances of the Mindray CL- 2i compared to other immunoassay systems. Methods: Analytical precision of CL-2i was evaluated by using both third-part and manufacturer s internal quality controls (IQC). One hundred fifty specimens were used for comparing TSH, FT3, FT4, Tg, TgAb, TPOAb in Mindray CL-2i, Roche Cobas e61, Beckman Coulter Access Immunoassay system and LIAISON Diasorin systems. Method limit of detection (LOD) and linearity, as well as matrix exchange effects, were also evaluated. Results: At the most clinically relevant levels, Mindray CL-2i repeatability results obtained by IQC were 3.52%, 1.29%, 3.62%, 2.7%, 4.31% and 4.34% for TSH, FT3, FT4, Tg, TPOAb and TgAb, respectively. Method comparison showed a full comparability between the Mindray CL-2i and the Roche Cobas e61 for TSH and FT3, while a slight to moderate bias was present for the other measurands. Matrix effect was negligible for TSH and linearity assays demonstrated a good linearity for TSH and Tg. LOD were.5 mui/l (CV% =9.46) and.936 (CV% =1.76) for TSH and Tg, respectively. Conclusions: Considering the current state-of-the-art of thyroid testing, the present study results support the comparability of the Mindray CL-2i automated system with respect to other immunoassay methods. Additional features of CL-2i were the good assays linearity and the very low LOD. Keywords: Thyroid testing; methods comparison; Mindray CL-2i system; limit of detection (LOD); method linearity Received: 1 October 218; Accepted: 17 October 218; Published: 3 October 218. doi: 1.2137/jlpm.218.1.3 View this article at: http://dx.doi.org/1.2137/jlpm.218.1.3 Introduction Thyroid stimulating hormone (TSH), or thyrotropin, is a hormone of about 28 kda, produced by the thyrotrope cells in the anterior pituitary glands. The production and the delivery in blood of TSH stimulates, in turn, the thyroid gland to produce the free thyroxine (FT4) and the free triiodothyronine (FT3), two tyrosine-based hormones, that are primarily responsible for the regulation of metabolism. The precursor of FT4 and FT3 is thyroglobulin (Tg), a huge 66 kda dimeric protein that acts as a substrate for the synthesis of thyroxine and triiodothyronine as well as the storage of the inactive forms of thyroid hormone and iodine. The latter is oxidized by thyroperoxidase (TPO)

Page 2 of 7 Journal of Laboratory and Precision Medicine, 218 before being additioned to Tg tyrosine residues. Thyroid gland disorders are among the most common endocrine conditions evaluated and treated by clinicians. Indeed, thyroid dysfunction, initially mostly asymptomatic, may evolve into biochemical physiological changes to clinically symptomatic diseases (1). Because of the lack of specificity of the typical clinical manifestations, the diagnosis of hypothyroidism is based primarily upon laboratory testing (1,2). Therefore, TSH determination has become a quite frequent test requested by physicians prescription, especially for assessing subclinical hypothyroidism (3). In autoimmune thyroid disease (AITD), the measurement of thyroid autoantibodies, such as anti-tpo (TPOAb) and anti-tg (TgAb), are utilized for diagnosis, being the main thyroid autoantigens involved in Hashimoto s thyroiditis (chronic lymphocytic thyroiditis) and Graves disease (4). The measurement of TSH circulating levels is routinely performed in clinical laboratories by using automated platforms. The full automation confers many analytical advantages, such as good precision, rapid measurements and good sensitivity. Ideally, data on performances of an instrument should be made available before the implementation into the laboratory. Indeed, manufacturers, collect a lot of evidences regarding the instrumental performances for precision, linearity, trueness, etc. usually estimated by rigorous protocols such as that suggested by the Clinical and Laboratory Standards Institute (CLSI). Nevertheless, the implementation of a quality system complying with the ISO 15189:212 calls for the verification or validation of analytical methods before the utilization of instruments for routine analyses (Sections 5.5.1.2 and 5.5.1.3) as the means of guaranteeing that their characteristics meet the specifications obtained during the manufacturers validation (5,6). Another aspect to consider is that all immunoassays remain sensitive to endogenous interferences that may cause analytical errors, threatening clinical decision making and patient safety. As example, in biotin-streptavidin immunoassays, elevated endogenous biotin due to therapeutic treatment of food supplementation can result in falsely elevated or depressed hormone levels in competitive or non-competitive sandwich assays, respectively (7). The aims of this study were to evaluate and verify the analytical performances of the new Chemiluminescence Immunoassay (CLIA) System CL-2i from Mindray, that include the analyses of TSH, FT4, FT3, Tg, TPOAb and TgAb for the diagnosis of thyroid disorder. Methods Assays description The Mindray CL-2i is a chemiluminescent system, featured by: (I) high throughput, up to 24 tests/hour; (II) flexible connections for lab automation; (III) a large operational capacity, up to 3 samples in one batch and supporting continuous loading; (IV) an intuitive and easy software interface. CL-2i system utilizes micron superparamagnetic particles platform with alkaline phosphatase (ALP) labeled reagents and AMPPD substrates. CL-2i allows performing several tests, such as tumour marker, bone marker, cardiac marker, hormones and tests for fertility, thyroid and infectious disease. TSH represent a third-generation assay with a functional sensitivity of.1.2 mu/l. TSH and Tg are two-site sandwich assay. FT3 and FT4 are biotin-streptavidin competitive CLIA, while TPOAb and TgAb are biotin-streptavidin two-steps non-competitive assays. CL-2i presents a traceable method for TSH (WHO 81/565), for Tg and TgAb (BCR CRM 457). Precision study Precision was evaluated by utilizing the Mindray Thyroid function multi internal controls materials (IQC) (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) for TSH, FT3, FT4, Tg, TPOAb and TgAb. The Bio-Rad Multiqual 1,2,3 Unassayed (Bio-Rad, Hercules, CA, USA) was used as third part QC material for evaluating precision of THS, FT3, FT4, Tg. Precision estimation was performed by evaluating triplicate measurements of aliquots of the same samples, performed for a total of five non-consecutive days. Analysis of variance (ANOVA) was used to estimate both repeatability (intra-day precision) and within-lab (total) precision, by using the CLSI EP15-A3 protocols (8). The manufactures claims for precision were compared against precision results obtained for the IQC materials and verified by using the method suggested by EP15-A3. Method comparison study and matrix effect evaluation A total of 15 (serum and plasma) specimens, covering the most clinical relevant range of TSH, FT3 and FT4, Tg, TPOAb and TgAb were collected. The following analytical systems were compared with respect to Mindray CL-2i system: (I) Roche Cobas 6 (e61) (Roche Diagnostics,

Journal of Laboratory and Precision Medicine, 218 Page 3 of 7 8 Scatter plot with fit Linear fit (.1389+.9594x) verification limit (UVL). Mean of plasma 7 6 5 4 3 2 1 2 4 6 8 Mean of serum 95% CI 95% prediction interval Results Matrix effect When comparing plasma and serum samples, a statistical significant agreement and interchangeability between the two matrices was obtained, as shown in Figure 1 for TSH assay. In particular, the mean ± standard deviations (SD) for serum and plasma samples were equal to 5.88±16.63 and 5.78±15.96 (t-test for comparing groups, t=.2 P=.978), respectively. Figure 1 Comparison of serum vs. plasma TSH measurements obtained for the Mindray CL-2i system. Basel, Switzerland) for TSH, FT3 and FT4; (II) Beckman Coulter Access Immunoassay system (Beckman Coulter Diagnostics Division Headquarters, Brea, CA, USA) for Tg, LIAISON Diasorin (Saluggia, Vercelli, Italy) for TgAb and TPOAb. Paired results were evaluated for outliers by the Grubbs test and following the Passing-Bablok regression was used to estimate constant and/or proportional bias. Further, for a total of 4 serum specimens, a corresponding lithium-heparin plasma aliquot was available. This allowed to test whether TSH Mindray CL-2i assay presented a matrix measurement bias. Limits of detection and linearity The linearity test was performed using a series of serial dilutions (1/2, 1/4, 1/8, 1/16, 1/32, 1/64, 1/128, 1/256) with dilution sample by manufacture. The detection limit was calculated by repeating a series of pools at known concentrations (for Tg from.4 to. µg/l and for TSH from.1 to. mui/l) for 1 consecutive days and 2 times in the same series. Statistical analyses Statistical analyses were performed by Analyze-it (v2.7), MedCalc Statistical Software version 18.2.1 and by R software for statistical computing v3.4.3 (https://www. R-project.org/). Variance decomposition by ANOVA was performed by ANOVA using an in-house script developed for estimating all the steps suggested by EP15-A3 and for verifying precision data and by calculating also the upper Precision study Table 1 reports the results for the precision study performed using the manufacturers IQC materials. The Repeatability and the within-lab precision were compared with those claimed by the manufacturer. The comparison was performed according to the EP15-A3 guideline and included the calculation of the UVL when necessary. Table 2 shows the precision study results obtained using the Biorad 1,2,3 Multiqual Unassay QC material. Method comparison study and matrix effect evaluation Results of the evaluation of comparability of TSH, FT3, FT4, Tg, TPOAb and TgAb, including the Passing-Bablok regressions and the Bland Altman analyses were reported in Table 3. The results for TSH and FT3 showed a good comparability between the Mindray CL-2i and the Roche Cobas 6 e61, with an absence of constant and proportional bias (Figure 2). Limits of detection (LOD) and linearity The linearity test demonstrated a regression coefficient (R 2 ) equivalent to.9996 for Tg and.9985 for TSH and Spearman s correlation coefficient (R) was 1. (Figure 3 and Table 4) for both tests considered. The low detection limits were found to be.5 mui/ L (CV% =9.46) and.936 (CV% =1.76) for TSH and Tg, respectively. Discussion In the last few years, a discussion has been raised on the

Page 4 of 7 Journal of Laboratory and Precision Medicine, 218 Table 1 Precision results of TSH, FT3, FT4, Tg, TPOAb, TgAb obtained by using the manufacturer IQC materials Measurand Level Mean Repeatability (CV%) Within-lab precision (CV%) TSH (miu/l) Level 1.667 3.52* 3.52 Level 2 39.67 2.5 4.9* FT3 (pmol/l) Level 1 4.8 1.29 2.76 Level 2 16.15 1.49 2.36 FT4 (pmol/l) Level 1 1.26 3.62 3.95 Level 2 38.56.84 1.33 Tg (µg/l) Level 1 9.38 2.7 2.32 Level 2 69.34 2.24 1.65 Anti TPO (IU/mL) Level 1 4.31 3.21* 5.2 Level 2 89.28 3.86* 4.78 Anti Tg (IU/mL) Level 1 18.62 4.34 4.6 Level 2 249.9 2.74 3.4 *, indicates that imprecision value was higher than that declared by manufacturers, also after the calculation of UVL as suggested by the CLSI EP15-A3. IQC, internal quality control; UVL, upper verification limit. Table 2 Precision results of TSH, FT3, FT4, Tg, obtained by using the Biorad 1,2,3 Multiqual Unassay Measurand Level Mean Repeatability (CV%) Within-lab precision (CV%) TSH (miu/l) Level 1.3 3.5 5.51 Level 2 3.51 3.65 4.6 Level 3 22.96 2.29 4.2 FT3 (pmol/l) Level 1 4.15 3.61 3.61 Level 2 1.2 2.23 2.33 Level 3 16.53 1.39 2.11 FT4 (pmol/l) Level 1 9.35 3.55 3.55 Level 2 28.86 1.46 1.46 Level 3 44.65 1.5 1.5 Tg (ug/l) Level 1.61 3.97 3.97 Level 2 1.7 2.4 2.24 Level 3 1.55 3.51 3.51 reproducibility crisis in clinical research, generated from a lack of confidence of results among clinicians due to inappropriate statistical methods or poor experimental competences (9). Thyroid dysfunction and treatment follow-up require accurate measurement of thyroid hormones. Most thyroid disease is treated on an outpatient basis; thus, assays have to be rapid and cost-effective for optimal patient care. Immunoassays are most commonly used because of their ease and availability, even if liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays are available as reference methods. LC-MS/MS assays are, in fact, much more specific, but are laborious with a high machine cost. Projects for improving laboratory results of thyroid testing

Journal of Laboratory and Precision Medicine, 218 Page 5 of 7 Table 3 Bland Altman analyses and Passing-Bablok regression results for all the evaluated measurands Measurand Slope (95% CI) Passing-Bablok regression results Intercept (95% CI) Bland Altman results, Bias (95% CI) TSH 1.1 (.98 to 1.4).13 (.19 to.9) #.47 (.29 to.197) FT3.93 (.89 to.98) #.27 (.8 to.5) #.37 (.136 to.62) FT4.8 (.77 to.84) # 1.76 ( 2.36 to 1.15) # 4.83 ( 5.123 to 4.537) # Tg 1.29 (1.25 to 1.33) #.12 (.39 to.1) 7.352 (4.221 to 1.484) # TPOAb.73 (.57 to.82) # 1.25 ( 1.76 to.68) # 13.519 ( 23.686 to 3.356) # TgAb.7 (.1 to.22) # 4.12 (3.42 to 4.5) # 53.391 ( 68.378 to 38.395) # #, indicates a significant variation with respect to the expected results under the hypothesis of test equality. Bland Altman analyses were performed by calculating the differences between the Mindray CL2i and the comparative method in the appropriate measurement units. Passing-Bablok regressions were obtained by calculating the results for the following equation: Mindray CL2i = slope comparative method + intercept. A 8 B 3 Mindray CL2i TSH-Roche Cobas e61 TSH (mlu/l) 6 4 2 2 4 6 +1.96 SD 1.9 Mean.1 1.96 SD 2.1 Mindray CL2i FT3-Roche Cobas e61 FT3 (pmol/l) 2 1 1 2 +1.96 SD 1.3 Mean. 1.96 SD 1.3 Figure 2 Bland Altman analyses of TSH (A) and FT3 (B) obtained for the Mindray CL-2i system for TSH and the comparative method (Roche Cobas e61). 8 3 1 2 3 4 5 5 1 15 2 Mean of Mindray CL2i TSH and Roche Mean of Mindray CL2i FT3 and Roche Cobas e61 TSH (mlu/l) Cobas e61 FT3 (pmol/l) A TSH value 1 Linearity plot 9 8 7 6 5 4 3 2 1 14 Linearity plot 12 1 8 6 4 2 5 1 5 1 Expected value for TSH Expected values for Tg B Tg values Figure 3 Verification of linearity for TSH (A) and Tg (B) for the Mindray CL-2i system.

Page 6 of 7 Journal of Laboratory and Precision Medicine, 218 Table 4 Linearity test for Tg and TSH Test unit Concentration range Linear equation r 2 R Spearman; P value Tg (µg/l).115 235.68 y=1.37x+1.154.9996 R=1.; P<.1 TSH (mui/l) 1.5 96.33 y=.9915x.3258.9985 R=1.; P<.1 received, therefore, an increasingly attention in the last few years, particularly as regards TSH measurement, and some major concerns have been identified on the need to improve the harmonization/standardization of the tests results across all the analytical methods (1-12). The first step in this harmonization process is the accurate evaluation of the performance characteristics of the assays, particularly precision and trueness, and this was the aim of this study. Precision has been differently defined, depending on the guidelines referenced. One definition based on a robust protocol is described in the CLSI EP15-A3. In this guideline, precision may be decomposed in its components, repeatability (also known as intra-day precision), betweendays and within-lab (also known as total precision), even if only the first and the last have valuable interest. In this study, both repeatability and within-lab precisions were evaluated by using manufacturers IQC and third part quality controls (Bio-Rad Multiqual 1,2,3 Unassayed). The results obtained underlined that precision goals claimed by manufacturers were verifiable for most of the measurands, especially for the within-laboratory precision. Considering the results for the third-part quality controls, results were similar to manufacturer s IQC, especially for withinlaboratory precision; indeed, differences were more relevant for repeatability. Other important aspects evaluated in this study were (I) the matrix effect; (II) methods comparability; (III) linearity; and (IV) LOD. Considering the first point, the obtained results showed that completely comparable results for serum and plasma, supporting the workable implementation of the Mindray CL-2i system irrespectively of the sample matrix used by laboratories for thyroid testing. The method comparison evaluation calls for the notion that thyroid tests results, even when performed by different laboratories, should be reliably and successfully interpreted (4). The results obtained for TSH and FT3 showed an optimal comparability between the Mindray CL-2i system and comparative method (Roche Cobas e61), with the absence of proportional biases. Differently, a slight (FT4 and Tg) to a moderate degree (TgAb) of proportional and/or constant bias was found for the other measurands and similar findings were already published in other studies comparing analytical performances of different analytical systems for thyroid testing (13,14). Finally, linearity evaluation underlined excellent results, while LOD evaluation demonstrates a better performance than the comparative method, being the Roche Cobas e61 LOD.5 mui/l for TSH and the Beckman Coulter.1 ng/ml for Tg, respectively. In summary, with respect to the current state-of-the-art of the thyroid testing by immunoassay systems, the results obtained in the present study support the comparability of the Mindray CL-2i automated system with respect to other immunoassay methods for the thyroid testing. Additional features of Mindray CL-2i were the good assays linearity and the low LOD obtainable for TSH and Tg. Acknowledgements None. Footnote Conflicts of Interest: The authors have no conflicts of interest to declare. References 1. LeFevre ML, U.S. Preventive Services Task Force. Screening for thyroid dysfunction: U.S. Preventive Services Task Force recommendation statement. Ann Intern Med 215;162:641-5. 2. Vanderpump MP. The epidemiology of thyroid disease. Br Med Bull 211;99:39-51. 3. Sarkar R. TSH Comparison Between Chemiluminescence (Architect) and Electrochemiluminescence (Cobas) Immunoassays: An Indian Population Perspective. Indian J Clin Biochem 214;29:189-95. 4. Tozzoli R, Bizzaro N. Harmonization in autoimmune thyroid disease diagnostics. Clin Chem Lab Med 218;56:1778-82. 5. Antonelli G, Padoan A, Aita A, et al. Verification of examination procedures in clinical laboratory for

Journal of Laboratory and Precision Medicine, 218 Page 7 of 7 imprecision, trueness and diagnostic accuracy according to ISO 15189:212: a pragmatic approach. Clin Chem Lab Med 217;55:151-8. 6. ISO 15189. Medical Laboratories -- Requirements for quality and competence. Geneva, International Organization for Standardization (ISO), 212. 7. Piketty ML, Polak M, Flechtner I, et al. False biochemical diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences. Clin Chem Lab Med 217;55:78-8. 8. Clinical and Laboratory Standards Institute (CLSI). User Verification of Precision and Estimation of Bias; Approved Guideline - Third Edition. 214. CLSI EP15-A3, Wayne, PA, USA. 9. Ioannidis JPA. The Reproducibility Wars: Successful, Unsuccessful, Uninterpretable, Exact, Conceptual, Triangulated, Contested Replication. Clin Chem 217;63:943-5. 1. Plebani M. Harmonization in laboratory medicine: the complete picture. Clin Chem Lab Med 213;51:741-51. 11. Clerico A, Ripoli A, Fortunato A, et al. Harmonization protocols for TSH immunoassays: a multicenter study in Italy. Clin Chem Lab Med 217;55:1722-33. 12. Plebani M. Towards a new paradigm in laboratory medicine: the five rights. Clin Chem Lab Med 216;54:1881-91. 13. Fillée C, Cumps J, Ketelslegers JM. Comparison of three free T4 (FT4) and free T3 (FT3) immunoassays in healthy subjects and patients with thyroid diseases and severe nonthyroidal illnesses. Clin Lab 212;58:725-36. 14. Diana T, Wüster C, Olivo PD, et al. Performance and Specificity of 6 Immunoassays for TSH Receptor Antibodies: A Multicenter Study. Eur Thyroid J 217;6:243-9. doi: 1.2137/jlpm.218.1.3 Cite this article as: Padoan A, Cosma C, Plebani M. Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2i system. J Lab Precis Med 218;3:93.