Numerical Analysis of Coronary Stent for Diverse Materials

Similar documents
Dynamic simulation of stent deployment effects of design, material and coating

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Modeling of stent implantation in a human stenotic artery

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Evaluation of Stent Performances using FEA considering a Realistic Balloon Expansion

[ICESTM-2018] ISSN Impact Factor

Computational analysis of mechanical stress-strain interaction of a bioresorbable scaffold with blood vessel

Abstract. Introduction. Journal of Biomedical Engineering and Research

THE COMPREHENSIVE FINITE ELEMENT MODEL FOR STENTING: THE INFLUENCE OF STENT DESIGN ON THE OUTCOME AFTER CORONARY STENT PLACEMENT

Computational Structural Modelling of Coronary Stent Deployment: A Review

Computational Analysis on Commercially Available Stent Designs

Planar Titanium Stent Design

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Finite Element Methods to Analyze Helical Stent Expansion

A computational study of stent performance by considering vessel anisotropy and residual stresses

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

A Study of the Mechanical Properties of a Real Coronary Stent Dynamical Behaviour of an Optimal Stent

Investigation of metallic surface area of coronary stents

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Senior Honors Thesis Prospectus

Nitinol biomedical devices: Design analysis

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

Your heart is a muscular pump about the size of your fist, located

Facing Coronary Artery Bypass Surgery? Learn about minimally invasive da Vinci Surgery

UNDERSTANDING ATHEROSCLEROSIS

The Development and Analysis of an FEA Simulation on an Artificial Knee

FINITE ELEMENT ANALYSIS OF SHAPE MEMORY ALLOY NiTi STENT IN CORONARY ARTERY. Amatulraheem Al-abassi. A thesis presented to University of Guelph

TESTING METHOD OF STENT S RADIAL FORCE

Modelling of Atherosclerotic Plaque for Use in a Computational Test-Bed for Stent Angioplasty

Transactions on Biomedicine and Health vol 3, 1996 WIT Press, ISSN

Biomechanical Analysis of Different Knee Prosthesis Biomaterials Using Fem

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p

2017 Cardiology Survival Guide

Topology optimisation of hip prosthesis to reduce stress shielding

Finally, the Control You Need to Deliver Accurate Treatment

EXAMINATION OF MECHANICAL AND MEDICAL APPLICATION PROPERTIES OF CORONARY STENTS

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

Summer Workshop of Applied Mechanics. Theoretical Analysis of Total Hip Joint Replacement of Sandwich design by FEM

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

A PATIENT`S GUIDE TO CORONARY ARTERY DISEASE. AND YOUR NIRxcell CoCr CORONARY Stent on RX System

Computational Analysis on Commercially Available Stent Designs

Is a Standardized Measurement of the Elastic Recoil of Coronary Stents under Vascular Conditions Necessary and Meaningful?

Understanding Peripheral

Introduction 3. What is Peripheral Vascular Disease? 5. What Are Some of the Symptoms of Peripheral Vascular Disease? 6

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

ELUTING NOVEL BIODEGRADABLE VASCULAR STENTS FOR IMPLANTATION

Analysis of Plate Bone Construct Failure Following Tibial Tuberosity Advancment

Numerical Study of the Behavior of a Zirconia Dental Prosthesis with Prior Defect

Introduction What Causes Peripheral Vascular Disease? How Do Doctors Treat Peripheral Vascular Disease?... 9

A COMPREHENSIVE ITERATIVE DESIGN, SIMULATION AND FINITE ELEMENT ANALYSIS OF BARE METAL STENT AND DISEASED ARTERY. A Thesis By.

CORONARY ARTERY BYPASS GRAFT

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS

Experimental investigation of the stent artery interaction

Peripheral Vascular Disease Patient Awareness

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT

Lower Extremity Arterial Disease

Current DES designs and expansion capacity testing

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Carotid Stenosis (carotid artery disease)

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

Modelling of temporomandibular joint and FEM analysis

Measurement of Mechanical Properties of Coronary Stents according to the European Standard pren

Introduction to soft tissues

REBEL. Platinum Chromium Coronary Stent System. Patient Information Guide

Abdominal Exam: The examination of the abdomen used by physicians to detect an abdominal aortic aneurysm.

CHALLENGES IN FEMORO-POPLITEAL STENTING

Lessons learned From The National PCI Registry

Oil Transmission Pipelines Condition Monitoring Using Wavelet Analysis and Ultrasonic Techniques

MECHANICAL PROPERTIES OF CORONARY VEINS

Interventions in Congenital & Structural Heart Disease: Who Drives New Techniques and Devices?

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Design Optimisation for Stent Manufacture

Link between effectiveness and cost data Costing was conducted prospectively on the same patient sample as that used in the effectiveness analysis.

ASX Small to Mid Caps New York Conference. March 1, 2012

Pathology of Cardiovascular Interventions. Body and Disease 2011

Stents selection and optimal implantation: sizes, design, deployment Abbott Vascular. All rights reserved.

DON T LET LEG PAIN BECOME A REAL THREAT.

A Finite Element Study of the Stress Redistribution of the Lumbar Spine after Posterior Lumbar Interbody Fusion Surgery

EDUCATION. Peripheral Artery Disease

Prediction of stent endflare, arterial stresses and flow patterns in a stenotic artery

Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 12, June 2013

Cover Page. Author: Wang, Ancong Title: Automatic quantification of intravascular optical coherence tomography Issue Date:

III. United States Patent (19) Sheiban 5,226,889. Jul. 13, and at least a pair of inflatable balloons carried on the

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

Patient Brochure. Clearstream Technologies, Ltd. Moyne Upper Enniscorthy Co. Wexford, Ireland. PK Rev. 0 05/17

UNDERSTANDING TREATMENT OPTIONS FOR HEART DISEASE. Visit

BLOOD PRESSURE. Unit 3: Transportation and Respiration

Sites of Atherosclerosis In order of Frequency

Effect of Plaque Composition on Biomechanical Performance of a Carotid Stent: Computational Study

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Wellness: Concepts and Applications 8 th Edition Anspaugh, Hamrick, Rosato

Cost effectiveness of drug eluting coronary artery stenting in a UK setting: cost-utility study Bagust A, Grayson A D, Palmer N D, Perry R A, Walley T

TAXUS Paclitaxel-Eluting Coronary Stent System

Failure criteria for Adhesives. Sainath Kadam, 3mE 19 oktober 2014

EVALUATION OF CAROTID STENT SCAFFOLDING THROUGH PATIENT-SPECIFIC FINITE ELEMENT ANALYSIS

A Novel Low Pressure Self Expanding Nitinol Coronary Stent (vprotect): Device Design and FIH Experience

Transcription:

Numerical Analysis of Coronary Stent for Diverse Materials Vasantha Kumar Assistant Professor Department of Mechanical Engineering Bearys Institute of Technology Mangalore, Karnataka, India Abstract Coronary stent implantation is one of the most important procedures to treating coronary artery disease such atherosclerosis. Due to its efficiency, flexibility and simplicity, the use of coronary stents procedures has increased rapidly. In order to have better output of stent implantation, it is needed to analyze the biomechanical behavior of this device before manufacturing and utilizing. Finite element method is one of the most effective methods to investigate biomechanical behavior of the stent. The objective of this research work is to investigate expansion characteristics of coronary stent for diverse materials, such as PEEK 450G, Magnesium AZ31 and Cobalt- Chromium L605. The results demonstrate the radial displacement, dogbone ratio, foreshortening, and flexibility of diverse materials. Keywords: Atherosclerosis, Stent, Cytotoxicity, Biocompatibility, Finite element method I. INTRODUCTION Coronary Artery Disease (CAD) is the most common cause of death in the world, accounting for 48% of all deaths in the world. Coronary artery disease, also known as atherosclerosis occurs when excess cholesterol attaches itself to the walls of blood vessels as shown in fig1. Embedded cholesterol also attracts cellular waste products, calcium and fibrin. This leads to a thickening of the vessel wall by complex interaction with constituents of the artery. The resulting pasty build-up known as plaque can narrow or even block the arteries that obstruct the flow of oxygen rich blood [1]. CADs are now a days the leading cause of death in the Western countries: a recent report of American Heart Association states that, on the basis of 2006 mortality rate, nearly 2300 Americans die of CAD each day, an average of 1 death every 38 seconds. This data explains well the high incidence of such pathologies which lead to high social and economical costs; in fact, the estimated direct and indirect cost of CVD for 2010 is $ 503.2 billion [2]. Several procedures are available to revascularise a blocked artery, including balloon angioplasty and stenting, bypass surgery and atherectomy [3]. Fourty years ago, coronary artery bypass surgery (CABG) was the popular revascularization treatment used to treat obstructive coronary artery diseases. However, it was claimed that frequent coronary closures occurred and hence emergency surgical revascularization was necessary [4, 5]. Stenting shows some advantages compared to other possible treatments, as it does not any surgical operation and has less complication, pain and a more rapid recovery. So the use coronary stents in interventional procedures has rapidly increased from 10% in 1994 to over 80% in current practice C.M Ramesha Associate Professor Department of Mechanical Engineering M.S Ramaiah Institute of Technology Bangalore, Karnataka, India [6]. In order to have the better output of stent implantation, it is needed to analyze the mechanical behavior of the stent before manufacturing and utilizing. One of the most effective methods to investigate the mechanical behavior of stent is finite element method. In comparison with expensive experiments carried out in hospitals and laboratories, numerical simulations accomplished by computers have advantages in both flexibility and cost [7]. A stent is an expandable tube like device that is inserted in to a natural conduit of the human body to restore a disease induced localized stenosis. The material used for human cardiovascular stent implants are conventional materials, with heavier density, not economical and unstable. It is observed that implant materials are affecting the coronary vessel where the failure of implant is maximum due to wear, corrosion, and fatigue; the above problems are due to improper material selection. From the literature review PEEK450G polymer shows an excellent wear resistance, chemical resistance, mechanical properties, low density material, physically stable, biocompatible and also machinable[8]. Therefore in this intended investigation, low density and stable material such as PEEK450G can be selected for cardiovascular stent implant. And also comparing this material with existing stent materials such as Magnesium AZ31 and Cobalt-Chromium L605. Fig.1 Coronary Stent with Artery II. MATERIALS AND METHODS In the present study materials used for coronary stents are PEEK 450G (Polyether-ether ketone) Magnesium AZ31 and Cobalt-Chromium L605 alloy. Primary model of stent was produced by using commercially available software. The model was constructed on the basis of images from literature [9]. The stent has an outer diameter of 1.915mm, length of 15mm and a thickness of 0.05mm. Fig.2 shows the linear array of 5 instances stent model in its unexpanded configuration. 387

Fig. 2 Linear array of 5 instances stent C. Material Properties The mechanical properties for Co-Cr L605 and Mg AZ31 are found in reference [10]. Table 7.1 gives the mechanical properties of the two alloys. The mechanical properties of PEEK 450G are found in VICTREX PEEK450G medical document data is considered as per room temperature [11]. TABLE.1 BIO-INERT MATERIAL PROPERTIES [10, 11] III. FINITE ELEMENT ANALYSIS OF STENT A. Coronary stent The model shown in the fig.2 is exported as IGES file. This file is imported into the HYPERMESH V11 software where the implant model is opened. The 3D finite element model required for analysis is created by discretizing the geometric model as shown in the fig.3; the discretization was performed in HYPERMESH software. The stent modeled by linear 8-noded solid element which is defines as solid C3D8R in abaqus. The finite element model of the stent consisted of 108080 nodes and 40852 elements. Material Young s Modulus E in D. Loads and Boundary Conditions Material Properties Density ρ in (kg/mm3) Poisson s Ration υ Yield σy in PEEK 450G 3700 1.30E-06 0.36 100 Magnesium (MgAZ31) 43000 1.70E-06 0.35 164 Cobalt- Chromium (Co-Cr L605) 243000 9.10E-06 0.3 476 Left hand side of the balloon is constrained in all DOF, whereas right hand side of the balloon is constrained, but axially free in Z-direction, and pressure applied on the inner surface of the balloon, frictionless contact is defined between balloon and stent it is shown in fig.5. B. Balloon Fig.3 Finite element meshed model of stent In this step, geometrical representation of balloon design is created based on reference [10]. The balloon as a medium to expand the stent was modeled to be 25mm length. The outer diameter and thickness of the balloon were 1.79mm and 0.05mm, respectively. A nylon material was used to represent the balloon. Similar to the stent model, the balloon was modeled by linear 8-noded membrane element which is defines as membrane M3D8R in abaqus. The finite element model of the stent consisted of 13805 nodes and 13750 elements as shown in fig.4. Fig.5 Finite element model of stent and balloon with boundary conditions A. Stent Analysis IV. RESULTS AND DISCUSSION Fig.4 Finite element meshed model of balloon 388

Von-mises in MPa Radial in mm 1 0.8 Fig.6 and stresses for PEEK 450G at different pressure load cases 0.6 0.4 0.2 0 0.5 1 1.5 Pressure in MPa PEEK450G Mg AZ31 Co-Cr L605 TABLE.2 DISPLACEMENT FOR PEEK 450G, MG AZ31 AND CO-CR L605 AT DIFFERENT PRESSURE LOAD CASES Materials Pressure MPa 0.5 0.25 0.24 0.15 1.0 0.51 0.5 0.32 1.5 0.8 0.77 0.5 Table.2 Shows radial displacement for PEEK 450G, Magnesium AZ31 and Cobalt-Chromium L605 are 0.8mm, 0.77mm and 0.5mm for different pressure. The radial displacement for PEEK 450G is maximum as compared to the other two materials, it is shown in fig.7. The stresses for PEEK 450G, Magnesium and Cobalt-Chromium are 67MPa, 221MPa and 542MPa as shown in Table.3. The stress generated for PEEK 450G is less as compared with other two materials, as shown in fig.8. TABLE.3 STRESS FOR PEEK 450G, MG AZ31 AND CO-CR L605 AT DIFFERENT PRESSURE LOAD CASES Materials Pressure MPa 0.5 23 162 477 1.0 42 195 499 Fig.7 for PEEK 450G, Mg AZ31 and Co-Cr L605 at different pressure 600 500 400 300 200 100 Fig.8 for PEEK 450G, Mg AZ31 and Co-Cr L605 At different pressure B. Stent Dogbone Ratio 0 When the stent expands, because of different distribution of circumferential stress between the free ends and the central part, it bends on edges that causes the diameter at the end sides becomes larger than that of the middle of the stent. This phenomenon is called dog boning. According to the findings of the clinical studies, a stent is expected to have the low dog boning. The dog boning is defined as: Dog boning = 0.5 1 1.5 Pressure in MPa PEEK450G Mg AZ31 Co-Cr L605 1.5 67 221 542 389

Fig.10 Foreshortening for PEEK 450G at 0.5MPa Fig.9 Dogbone ratio for PEEK 450G at 0.5MPa TABLE.4 DOG BONE RATIO FOR DIFFERENT MATERIALS Materials/Pressure Diameter expansion/dog bone ratio (%) 0.5 MPa 9.75 9.04 14.69 1.0 MPa 8.54 7.52 14.27 1.5 MPa 7.70 3.72 13.14 Table.4 shows the dogbone ratio for PEEK 450G, Magnesium AZ31 and Cobalt-Chromium L605 at different pressure loading cases. The dog bone ratio is 7.70, 3.72, and 13.14% at 1.5MPa pressure loading case. Dog bone ratio for Cobalt-Chromium L605 is maximum as compare to the PEEK 450G and Magnesium AZ31. C. Stent Foreshortening As mentioned above, because of the dog boning phenomenon, the diameter at the end sides of the stent becomes larger than that of the middle of the stent. This results a reduction in the stent length which is called foreshortening. In fact, foreshortening defines the effective length of the stent and according to the findings of the clinical studies; a stent is expected to have a low value of foreshortening. The foreshortening is defined as: TABLE.5 FORESHORTENING FOR DIFFERENT MATERIALS Materials/Pressure Foreshortening (%) 0.5 MPa 0.60 0.28 0.19 1.0 MPa 1.18 0.68 0.47 1.5 MPa 1.70 1.67 1.33 Table.5 shows the foreshortening values for PEEK 450G, Magnesium AZ31, and Cobalt-Chromium L605. The foreshortening values are 1.70, 1.67, and 1.33% at 1.5MPa load case. The foreshortening of PEEK 450G is maximum as compare to the other two materials. D. Stent Flexibility Stent will face swift disturbance due to artery movements for which it is displaced to understand the flexibility of stent, stent is displaced to 0.5mm at the more or less mid section when stent faces swift disturbance we will find stress, displacement and strains at different sections of stent. Foreshortening = Fig.11 Stent flexibility boundary conditions 390

The displacement of PEEK 450G, Cobalt-Chromium L605 and Magnesium AZ31 are 0.61mm, 0.56mm and 0.55mm. The displacement of PEEK450G is maximum as compare with other two materials. The stress generated in Cobalt-Chromium L605 is maximum as compare to the PEEK 450G and Magnesium AZ31. Plastic strain is very less in Magnesium AZ31 as shown in table.6. V. CONCLUSION Realistic features of materials on the basis of results obtained from Finite Element Analysis, considering the different pressure loads exerted by blood pressure of human body on Stent implants are concluded as follows with respect to displacement, stresses, dogbone ratio, foreshortening and flexibility. Radial displacement of PEEK 450G is better as compare to the Magnesium AZ31 and Cobalt-Chromium L605, analysis shows PEEK 450G is better material than Mg and Co-Cr with applied pressure load, PEEK 450G does not have plastic strains and stresses are within the yield limits. Dog bone ratio should reduce with more pressure which is observed in calculations and results are reasonable for all materials, with no plastic strains PEEK 450G is better with dog bone ratio. Foreshortening should increase with more pressure which is observed in calculations and results show more foreshortening for PEEK than the other materials. A Stent flexibility result shows PEEK450G is more flexible with more displacements for swift changes inside the artery. Finally PEEK 450G is recommended for future coronary Stents. Fig.12, and Plastic Strain for PEEK 450G at 0.5mm displacement TABLE.6 DISPLACEMENT, STRESS AND PLASTIC STRAIN FOR PEEK 450G AT 0.5MM DISPLACEMENT Materials Stent flexibility 0.61 0.55 0.56 184 238 850 PEEQ (%) 13.52 0.8 4.33 REFERENCES [1] David Chua SN, Mac Donald BJ, and Hashmi MSJ, Finite element simulation of stent and balloon interaction, journal of materials processing technology 2003, pp 143-144. [2] Cardiovascular disease statistics, American Heart Association. 1988: ICD/9, pp 390-459. [3] Pericevic I, Lally C, Tonner D, and Kelly DJ. The influence of plaque composition on underlying arterial wall stress during stent expansion, 2009, pp 428-433. [4] S.Garg and P.W.Serruys, Coronary stents: Current status, vol 56, 2010, pp S1-S42. [5] S.Yusuf, D.Zucker, E.Passamani, T.Takaro, and L.D Fisher, Effect of coronary artery bypass graft surgery on survival, vol 344, 1994, pp 563-570. [6] Lim D,Cho SK,Park WP, Kristensson A, and Al-Hassani STS, Suggestion of potential stent design parameters to reduce restenosis risk driven by foreshortening or dog bonning due to non uniform balloon stent expansion, 2008, pp 1118-1129. [7] Li N, Zhang H, and Ouyang H, Shape optimization of coronary artery stent based on a parametric model. J finite element in analysis and design, 2009, pp 468-475. [8] Shravan S. Pargaonkar, Mangesh S. Prabhuue, A polyayletherketone biomaterial for use in medical implant applications, vol 5, 2015, pp 1-3. [9] Craig Bonsignore, Open stent design, NDC, 47533 Westinghouse Drive Fremont, CA, 94577, 2011, pp 20-47. [10] Schiavone, Tensile behavior for SS316L, Co-Cr L605 and Mg alloy, phd thesis, 2015, pp 65-67. [11] VICTREX, The mechanical properties of PEEK 450G medical document data is considered as per room temperature, 2014, pp 1-4. [12] Schiavone. A study of balloon type, system constrained and artery constitutive model used in finite element simulation of stent deployment, 2015, pp1-15. 391