The value of the craniocaudal mammographic view in breast cancer detection: A preliminary study

Similar documents
Title: Breast Screen Reader Assessment Strategy (BREAST): UK Collaboration

Improving Reading Time of Digital Breast Tomosynthesis with Concurrent Computer Aided Detection

Since its introduction in 2000, digital mammography has become

arxiv: v2 [cs.cv] 8 Mar 2018

Breast screening: understanding case difficulty and the nature of errors

FDA Executive Summary

Mammography is a most effective imaging modality in early breast cancer detection. The radiographs are searched for signs of abnormality by expert

Mammography limitations. Clinical performance of digital breast tomosynthesis compared to digital mammography: blinded multi-reader study

The Australian BreastScreen workforce: a snapshot

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography

The effect of digitising film prior mammograms on radiologists' performance in breast screening: a JAFROC study

The Radiology Aspects

CHAPTER 2 MAMMOGRAMS AND COMPUTER AIDED DETECTION

A comparison of the accuracy of film-screen mammography, full-field digital mammography, and digital breast tomosynthesis

Practitioner compression force variation in mammography : a 6 year study

Financial Disclosures

Mammographic Breast Density Classification by a Deep Learning Approach

Policy Library Clinical Advantages of Digital Breast Tomosynthesis in Symptomatic Patients

Detection and Classification of Calcifications on Digital Breast Tomosynthesis and 2D Digital Mammography: A Comparison

Women s Imaging Original Research

Session 4: Test instruments to assess interpretive performance challenges and opportunities Overview of Test Set Design and Use

Breast cancer detection: radiologists performance using mammography with and without automated whole-breast ultrasound

PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification.

Improvement in ROC curves of readers with next generation of mammography CAD

International Day of Radiology 2016 Interview on Breast Imaging Australia / Dr. Michelle Reintals. Breast imaging in Australia

Session 2: The Role of Specialist Radiology Technologists

Women s Imaging Original Research

Digital Breast Tomosynthesis

ACCREDITATION DOCUMENT THE RADIOGRAPHER

Can Digital Breast Tomosynthesis(DBT) Perform Better than Standard Digital Mammography Workup in a Breast Cancer Assessment Clinic?

Assessing Viewing Pattern Consistency in Mammogram Readers

Cairo/EG, Khartoum/SD, London/UK Biological effects, Diagnostic procedure, Ultrasound, Mammography, Breast /ecr2015/C-0107

Breast asymmetries in mammography: Management

The Effect of Image Processing on the Detection of Cancers in Digital Mammography

Proven clinical effectiveness at low radiation dose

Computer aided detection of clusters of microcalcifications on full field digital mammograms

Correlation between lesion type and the additional value of digital breast tomosynthesis

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

Mammographic interpretation training: how useful is handheld technology?

Characterization of the breast region for computer assisted Tabar masking of paired mammographic images

Radiologists detect 'gist' of breast cancer before overt signs appear April 2018

Corporate Medical Policy

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD

BreastScreen Victoria tomosynthesis screening trial, Maroondah pilot: Preliminary Outcomes

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

Deep-Learning Based Semantic Labeling for 2D Mammography & Comparison of Complexity for Machine Learning Tasks

Breast Imaging Update: Old Dog New Tricks

Jing Zhang, PhD, Lars J. Grimm, MD, MHS, Joseph Y. Lo, PhD, Karen S. Johnson, MD,

Min Jung Kim Department of Medicine The Graduate School, Yonsei University

Automatic classification of mammographic breast density

Purpose. [Purpose] [Introduction] Architectural distortion on mammography

Amammography report is a key component of the breast

Breast imaging in general practice

Fremtidens rolle for tomosyntese

Mammographic evaluation of palpable breast masses with pathological correlation: a tertiary care centre study in Nepal

Hacia la imagenología tomográfica de mama

Digital Breast Tomosynthesis from a first idea to clinical routine

Interval cancers in BreastScreen Aotearoa

Standard Breast Imaging Modalities. Lilian Wang, M.D. Breast Imaging Section Department of Radiology Northwestern Medicine

Contrast-Enhanced Digital Mammography

Diagnostic Dilemmas of Breast Imaging

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

S. Murgo, MD. Chr St-Joseph, Mons Erasme Hospital, Brussels

Diagnostic benefits of ultrasound-guided. CNB) versus mammograph-guided biopsy for suspicious microcalcifications. without definite breast mass

BARC/2013/E/019 BARC/2013/E/019. AUDIT OF MAMMOGRAPHY PERFORMED IN OUR HOSPITAL by Surita Kantharia Medical Division

Breast Tomosynthesis

Epworth Healthcare Benign Breast Disease Symposium. Sat Nov 12 th 2016

Women s Imaging Original Research

MEDICAL POLICY SUBJECT: MAMMOGRAPHY: COMPUTER- AIDED DETECTION (CAD) POLICY NUMBER: CATEGORY: Technology Assessment

EARLY DETECTION. Screening Mammography Programs. Knowledge Summary

Digital Breast Tomosynthesis in the Diagnostic Environment: A Subjective Side-by-Side Review

Breast Tomosynthesis

Outline. Digital Breast Tomosynthesis: Update and Pearls for Implementation. Tomosynthesis Dataset: 2D/3D (Hologic Combo Acquisition)

A superior chest x-ray computer-aided detection (CXR CAD) application: a reader study

Fundamentals of Breast Tomosynthesis

Breast Imaging! Ravi Adhikary, MD!

Investigating the performance of a CAD x scheme for mammography in specific BIRADS categories

Updates in Mammography. Dr. Yang Faridah A. Aziz Department of Biomedical Imaging University Malaya Medical Centre

BreastScreen Queensland Gold Coast Service. General Practice Nurses Forum Thursday 4 May 2017

Imaging Guidelines for Breast Cancer Screening

Introduction 1. Executive Summary 5

Comparison Study: Impact of Thin Protective Coverlets on Positioning and Tissue Acquisition in Breast Imaging

The early detection programme for breast cancer

Features of Prospectively Overlooked Computer-Aided Detection Marks on Prior Screening Digital Mammograms in Women With Breast Cancer

European Cancer Network Aims and Activities

Mdiagnosing breast by using x-ray to both breasts. It

Ascertaining and reporting interval cancers in BreastScreen Aotearoa: A protocol NATIONAL SCREENING UNIT (NSU) MINISTRY OF HEALTH

Imaging in breast cancer. Mammography and Ultrasound Donya Farrokh.MD Radiologist Mashhad University of Medical Since

Ana Sofia Preto 19/06/2013

- The University of. - Student. Western Australia, Crawley, Western Australia

Mammogram Analysis: Tumor Classification

Women s Imaging Original Research

Mammography. What is Mammography? What are some common uses of the procedure?

Mammogram Analysis: Tumor Classification

American College of Radiology Annual Progress Report: 2009 Formula Grant

Breast Cancer Screening Factsheet

Contrast-Enhanced Spectral Mammography

Additional US or DBT after digital mammography: which one is the best combination?

Purpose: Materials and Methods: Results: Conclusion: Original Research n Breast Imaging

Sample Size Considerations. Todd Alonzo, PhD

Transcription:

The value of the craniocaudal mammographic view in breast cancer detection: A preliminary study P D Trieu 1, Prof. P C Brennan 1, Dr. W Lee 2, Dr. E Ryan 1, Dr. W Reed 1, Dr. M Pietrzyk 1 1. Medical Image Optimization and Perception (MIOPeG), Discipline of Medical Radiation Sciences, Faculty of Health Sciences, University of Sydney. 2. Breast Screen New South Wales, Cancer Institute. Abstract Mammography is considered a reliable modality in detecting early breast lesions. Many studies indicated that two-view (mediolateral oblique (MLO) and cranio-caudal (CC)) mammography provides radiologists with more information to detect breast cancer than a single medio-lateral oblique (MLO). Nevertheless, there have been growing concerns about the effects of radiation exposure on patients who absorb approximately double radiation dose with a routine two-view. Although one view MLO was proved to be less effective than two views in detecting cancer lesions, the value of cranio-caudal view has never been established. This paper aims to assess the value of the single cranio-caudal view mammogram in the detection of breast cancer. 129 radiologists were asked to report 60 two-view mammograms of the left and right breasts and 55 radiologists assessed a further set of 55 single craniocaudal views. Participants were asked to search for the presence of any breast lesions and provide confidence scores for their decisions. The sensitivity, specificity and localization sensitivity of each reader were analysed along with JAFROC (jackknife free response receiver operating characteristics) figure of merit and ROC (receiver operating characteristics) values. Results showed that two-view mammograms were more effective in detecting malignant nodules than single cranio-caudal view in terms of sensitivity, localized-sensitivity, ROC and JAFROC. The single cranio-caudal view had a higher specificity as compared to two-view mammography. Keywords Cranio-caudal mammogram, two-view mammogram, breast nodule detection, ROC, JAFROC 150

The value of craniocaudal mammographic view in breast cancer detection P D Trieu 1. Introduction Breast cancer is one of the most common types of cancer diagnosed in women. According to Australian Institute of Health and Welfare (2010), the rate of female developing breast cancer before 85 years old is one in nine and the risk of mortality due to this disease is one in thirty seven. Mammography with reasonably high sensitivity (60% 80%) and specificity (73% 95%) has been considered to be a reliable modality in detecting early breast lesions through abnormal signs such as masses, calcifications, bilateral asymmetry and distortion (Maggio, 2004; Kavanagh et al., 2000). When breast screening was firstly introduced, a single mediolateral oblique (MLO) view of each breast was considered as the routine examination and only later was the caudo-cranial (CC) added as a subsequent view. Wald et al. (1995) through a randomized controlled trial found that two view mammography (CC plus MLO) was able to detect breast cancer signs, especially small lesions, better than one view image, with a 24% increase in sensitivity (4). Other studies by Blanks, Moss & Wallis (1997), Law & Faulkner (2002) and Seigneurin, Exbrayat, Labarere & Colonna (2009) also indicated that standard two view breast radiography at subsequent screens provided radiologists with more information in detection of cancer than only the MLO. Nevertheless, there has been growing concerns about the effects of radiation exposure on patients who absorb approximately double the radiation dose with a routine two-view rather than a single projection (Law & Faulkner, 2002). Although one view MLO was proved to be less effective than two views, the role of one view CC in diagnosis of breast cancer has not yet been explored. The aim of this study is to assess the value of cranio-caudal mammogram in screendetected breast cancers. 151

2. Methods and Materials Two retrospective experiments were conducted at the 2011 Royal Australian and New Zealand College of Radiologists Breast Imaging Group meeting in Hobart. In the first experiment, 60 two-view mammograms (CC&MLO) of both breasts comprising 20 malignant lesions and 40 cancer-free cases were collected. The second study consists of 55 single CC views (either left or right breast) of which 23 contained malignant lesions. All lesions were biopsy proven. Figure 1: The two view mammogram: Medio-lateral oblique (MLO) (left) and Cranio - caudal (CC) (right) The portrait Eizo Radiforce GS510 Specs (21.3 inch) monochrome high- class LCD diagnostic monitors were utilized to display images in the both experiments. Average viewing distance was approximately 40 cm from the display and the ambient light in the reading room ranged from 20 to 30 lux. 129 radiologists who currently report breast images in Australia and New Zealand were involved in the first experiment and 25 readers were recruited for the second experiment. 14 radiologists involved in both sessions. The average experience of reporting mammography of participants was approximately 10 years (Figure 2). 152

The value of craniocaudal mammographic view in breast cancer detection P D Trieu Figure 2: Participant demographic details Two view Experiment One view Experiment Common readers between two experiments Number of readers Average mammo reading experience (years) Average number of mammo readings per year Profession General radiologist Breast specialist Radiographer 129 9 4790.6 111 13 4 1 25 10.9 5008.6 23 2 0 0 14 11.5 6478.5 12 2 0 0 Radiology resident The same general procedure was used for both experiments. Participants were asked to undergo one session of image interpretation and decide on the presence or absence of breast lesions on each image. Radiologists were allowed to digitally manipulate the images (panning, zooming). When the readers were ready to provide a decision, they were asked to indicate the confidence score from 1 to 5; whereby 1 representing complete confidence that the case was normal, 2 was dedicated for a benign finding and 3 to 5 represented a malignant lesion. For the images with confidence scores from 2 to 5, lesions locations were indicated by using a mousecontrolled cursor to mark the coordinates (x, y). The readers moved on to the next case with a mouse-click when they were certain that they had reported all lesions. There was no restriction on search time and number of mouse-clicks. Clinical history and information on the distribution of normal and abnormal cases were not provided to participants. In term of data analysis, the locations and the confidence levels reported by observers were classified into true positive, false positive, true negative and false negative basing on truth of the cases (Figure 3). The localizations were assessed based on visual angle of 1 degree between the mouse-click coordinates and nodule center. 153

Figure 3: Classification of confidence levels according to the truth of the cases Cases Confidence levels Transferred data Normal/Benign 1, 2 True negative 3, 4, 5 False positive Malignant 1, 2 False negative 3, 4, 5 True positive As participants were allowed to indicate more than one finding per case, the confidence scores were analyzed by using the receiver operating characteristics (ROC) (Metz, 2006) and jackknife free response receiver operating characteristics (JAFROC) methods (Chakraborty, 2004). ROC provides the comprehensive description of participant performance in term of sensitivity combined with specificity while JAFROC method allows quantitative analysis of observer data when radiologists interpret images, which could contain more than one lesion and a location can be reported for each perceived lesion. T-test was utilized to compare all reader/all case metrics and DBM MRMC ROC (Dorfman Berbaum Metz Multi Reader Multi Case) was used for same-reader same-case analysis. In addition sensitivity, specificity and localization sensitivity of each reader were calculated with the data collected from both sessions of the study. 3. Results 3.1. All cases/all readers: The results showed that two view (CC & MLO) mammography demonstrated higher efficacy in detecting malignant cases in terms of sensitivity (0.84 > 0.62; p<0.001), localized-sensitivity (0.6 > 0.5; p=0.006) as compared to one view. Whilst two-view verified a higher sensitivity than 154

The value of craniocaudal mammographic view in breast cancer detection P D Trieu single cranio-caudal mammogram, CC showed higher (non-significant) specificity (0.79) than CC & MLO (0.72) (p=0.065) (Figure 4). Figure 4: Averaged sensitivity, specificity, localized sensitivity, ROC and JAFROC based on data of one view (CC) and two view (CC&MLO) experiments. * * * * * demonstrates the significant difference in the metric comparision between one view and two views. Sensitivity LOC- Sensitivity Specificity ROC JAFROC CC 0.622 0.499 0.787 0.740 0.620 CC&MLO 0.840 0.600 0.720 0.820 0.670 P values <0.001 0.006 0.065 <0.001 0.020 There were also significant differences found in ROC (0.82 > 0.74; p<0.001) and JAFROC (0.67 >0.62; p=0.02) across one view and two views. Unpaired comparisons between two experiments indicated that radiologists localized malignant lesions more accurately in two views than in one view. 3.2. Same cases same readers: Of 15 cases and 14 readers common between the one view and two view experiments, there 155

was no significant difference in ROC and JAFROC. However, noticeable changes have been demonstrated in the specificity and the localization-sensitivity. CC view continued showing higher capability in identifying normal cases (0.81 > 0.66; p=0.001) while CC-MLO was twice more sensitive than CC in lesion detection (0.75>0.39; p=0.16), although the level of localization of both experiments were relatively low ( less than 0.5, p=0.016) (Figure 5). Figure 5: Averaged sensitivity, specificity, localized sensitivity, ROC and JAFROC based on data of same readers and same cases in one view (CC) and two view (CC&MLO) experiments. * * * * demonstrates the significant difference in the metric comparision between one view and two views Sensitivity LOC- Sensitivity Specificity ROC JAFROC CC 0.390 0.189 0.812 0.680 0.540 CC&MLO 0.750 0.375 0.659 0.840 0.490 P values 0.016 0.016 0.001 0.110 0.675 4. Breakthrough work The results demonstrate that malignant lesions tended to be missed with a single CC view compared with two-view mammography. This was shown to be true with all the analytical 156

The value of craniocaudal mammographic view in breast cancer detection P D Trieu metrics used. However the enhanced specificity of CC may suggest some benefit for low risk women having a follow-up breast screening. There were some limitations with the work, one being the laboratory effect. In observer experiments, radiologists were not affected by the pressure of being responsible for patients hence the reporting of some readers might be different from clinical reporting. Besides, the absence of CC views of the other breast in the second experiment might be a reason why the sensitivity of CC view was relatively low. Therefore, more studies with adequate views and larger number of cases are necessary for more accurate analysis. 5. Conclusion Radiologists have a better performance in detection of breast nodules when they have access to both the MLO and CC views. Whilst the CC view alone may miss important diagnostic information, some benefit for screening programs may be evident with single view examinations. This requires further investigation. 157

References Australian Institute of Health and Welfare & Australasian Association of Cancer Registries (2010). Cancer in Australia: An overview (Cancer series no. 60. Cat. no. CAN 56). Canberra: AIHW. Blanks R.G., Moss S.M., Wallis M.G. (1997). Use of two view mammography compared with one view in the detection of small invasive cancers: Further results from the National Health Service breast screening programme. Journal of Medical Screening, 4, 98-101. Chakraborty, D.P. (2004). Jackknife free-response ROC methodology. In D.P. Chakraborty & M.P. Eckstein (Eds.), Proceedings of the SPIE, Medical Imaging 2004: Image perception, observer performance and technology assessment, (Vol. 5372, pp. 144-153). Kavanagh A.M., Giles G.G., Mitchell H., Cawson J.N. (2000).The sensitivity, specificity, and positive predictive value of screening mammography and symptomatic status. BMJ, 7(2), 105-110. Law J., Faulkner K. (2002). Two view screening and extending the age range: the balance of benefit and risk. British Journal of Radiology, 75, 889-94. Maggio, C.D (2004). State of the art of current modalities for the diagnosis of breast lesions, European Journal of Nuclear Medicine and Molecular Imaging 31 (Suppl. 1), 56-69. Metz, C.E. (2006). Receiver Operating Characteristic Analysis: A tool for the Quantitative Evaluation of Observer Performance and Imaging System. American Journal of Radiology, 3, 413-422. Seigneurin A., Exbrayat C., Labarere J., Colonna M. (2009). Comparison of interval breast cancer rates for two-versus single-view screening mammography: A population-based study. The Breast, 18, 284-288. Wald N.J., Murphy P., Major P., Parkes C., Townsend J., Frost C. (1995). UKCCCR multicentre randomised controlled trial of one and two view mammography in breast cancer screening. BMJ, 311, 1189-93. 158