Test Setup to Measure Gains of the Angra Photomultipliers Using the Single Photoelectron Technique

Similar documents
Effects of Magnetic field on Voltage response of PMT

Characteristics of PMT used in prototype array of ED

New 50 cm Photo-Detectors for Hyper-Kamiokande

CLAS12 FTOF Studies: Rate and Magnetic Shielding Effects on PMT Timing Resolutions

A study on the transmittivity of Tyvek

SoLID EM calorimeter study and general test system in SDU

NaI Detector and Photo-multiplier Tubes

Detailed performance evaluation of a new 20-inch photomultiplier tube with a Box and Line dynode

procedures Photomultiplier tests - Requirements for SNO and standardized testing ^o^-t/l-jo-^z, R.J.Boardman 1 Abstract April 5, 1990

Some recent results. Photomultiplier tests. S^o-sr^o-^n. 1 Abstract. R.J.Boardman. April 5, 1990

Luminosity Monitor and Test Run Results. Presented by: Luis Mercado UMASS Amherst 7/23/08

Performance evaluation of new photodetectors for Hyper-Kamiokande

A study of the new hemispherical 9-inch PMT

Plastic scintillation detectors for dose monitoring in digital breast tomosynthesis

Status of SHMS Electromagnetic Calorimeter

The 20 inch PMT system for the JUNO experiment

HiFFUT A New Class of Transducer

PMT Instrumentation of JUNO

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

MCP PMT with high time response and linear output current for neutron time-of-flight detectors

Hemodynamic Analysis System

Enriched Xenon Observatory for double beta decay. EXO-200 Progress. Liang Yang. Research Associate. July 8, 2008 SLAC Annual Program Review Page 1

N4i Installation Manual. (picture of N4i screen)

INSTALLATION INSTRUCTIONS

Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Preparation, testing and assembling of the inner LArGe infrastructure

Electromyography II Laboratory (Hand Dynamometer Transducer)

Detection. personal. Now, you can optimize your personal workflow. Promega instruments and reagents integrate easily.

Dynamic Hollow Cylinder System Proposal

General Single Ion Calibration. Pete 14-May-09

Lab 4: Introduction to Physiological Measurements - Cardiovascular

Supplementary Data. Supplementary Materials and Methods Measurement of NO formation by ozone-based chemiluminescence. Supplementary References

Recent Progress in Vacuum Photon Detectors at Hamamatsu

Operator s Manual Development of a Pulsatile Left Heart Simulator

Real Time Spectrometer for thermal neutrons from Radiotherapic Accelerators

Terahertz reflectometry imaging for low and high grade gliomas

LMJ - DP14 Neutron Pack Calibration

Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Lesson 2 EMG 2 Electromyography: Mechanical Work

Direct Reading (Pocket) Dosimeter

THYROID UPTAKE SYSTEMS

arxiv: v3 [physics.ins-det] 29 Jan 2016

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Neuron, Volume 63 Spatial attention decorrelates intrinsic activity fluctuations in Macaque area V4.

Theta sequences are essential for internally generated hippocampal firing fields.

1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

w. D. Jolly, F. A. Bruton, and C. Fedor

H. A. TANAKA (TORONTO/IPP/TRIUMF) MPMTS FOR HK A PERSONAL PERSPECTIVE

A bioimpedance-based cardiovascular measurement system

SpiderX. Portable Residual Stress X-Ray Diffractometer.

Gas gain: 2-5x10^4 Preamplifier peaking time: 10-15ns Gas pressure: 3bars. Discriminator 1. Charge. Discriminator 2. Control Logic (JTAG Interface)

Time Variation of Partial Discharge Activity Leading to Breakdown of Magnet Wire under Repetitive Surge Voltage Application

Task Test beam infrastructure in Frascati LNF, Ferrara & Perugia INFN structure and University of Bergen

Texas A&M University Electrical Engineering Department. ELEN 665 RF Communication Circuits Laboratory Fall 2010

Radiation Detection and Measurement

Descriptions of NDT Projects Fall 2004 October 31, 2004

RESULTS OF THE LHC DCCT CALIBRATION STUDIES

RiGO801 Laboratory Requirements

arxiv: v2 [astro-ph.im] 29 Oct 2014

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Thin Quartz Detector R&D and PMT Linearity Studies

Solar Illumination Study

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Development of Indigenous Ultrasonic Data Acquisition and Recording System for ISI of Pressure Tubes of PHWR

AMBCO 1000+P AUDIOMETER

Atomlab. Thyroid Uptake System BIODEX. A complete, mobile, Medical Spectrometer. The Clinical Advantage.

Title. Author(s)Ono, Daisuke; Honma, Ken-Ichi; Honma, Sato. CitationScientific reports, 5: Issue Date Doc URL

By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun

LABORATORY INVESTIGATION

Supporting Online Material for

Annealing Influence on the Optical Properties of Nano ZnO

1 Exam Prep. SMACNA Architectural Sheet Metal Questions

SIMULATION RESPONSE OF RESISTIVE PLATE CHAMBER FOR FAST NEUTRONS USING GEANT4 MC CODE

Model 3500 Manual TLD Reader with WinREMS 3500-W-O-0602 Page 5-1 Operator's Manual

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

Exercise 6: Muscle Physiology II Twitch & Summation

Principles and Applications of Electrical Circuits and Signal Theories in EP

Cigarette Smoking Machine CSM-SCSM. User s Manual

PLL. The system blocks and their input and output signals are as follows:

Intravital Microscopic Interrogation of Peripheral Taste Sensation

Application(s) of Alanine

iworx Sample Lab Experiment HP-7: Hypothesis-driven Biofeedback Lab/Research Study

Xbox 360 S Hard Drive Bracket Replacement

s MASH Mobility (m/s)/n Frequency, Hz 2.5 x x x x x Average mobility

Technical Data Sheet SONOFLOW CO.55/xxx(H) V2.0

Supplementary Information

2The Concept of Periodic Functions

Modern Packaging and Weighing Technology for Industrial Food Producers

Multielement ultrasonic probes for projection imaging of biological media

Monitor Instructions for Models: CHB-R6 CHB-UV6

Experiment HH-3: Exercise, the Electrocardiogram, and Peripheral Circulation

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy

Insulin Effects on DPPE-succinyl Bilayer Resistance page 1 of 9

Neurobiology: The nerve cell. Principle and task To use a nerve function model to study the following aspects of a nerve cell:

Mechanical Simulation of Probing on SMART POWER POA devices

Dissection and Contraction of Frog Heart

Transcription:

Angra Neutrino Project AngraNote 003-2007 (Draft) Test Setup to Measure Gains of the Angra Photomultipliers Using the Single Photoelectron Technique W. Raposo, M. Vaz º Centro Brasileiro de Pesquisas Fisicas - CBPF, Rio de Janeiro, Brazil L. Villasenor Universidad Michoacana de San Nicolas de Hidalgo - UMSNH, Morelia, Mexico Abstract We report on a setup built to measure gains of the PMT proposed to be used in the Angra Neutrino detector. We illustrate the use of the standard single photoelectron technique supplemented with measurements of rise time in addition to the more conventional methods that measure only amplitude and / or charge of single photoelectron pulses. º Corresponding author, e-mail: mariovaz@cbpf.br

AngraNote 003-2007 1 1 Introduction The measurement of photomultiplier (PMT) gains by sending a short flash of visible or UV light into the PMT photocathode is a simple and well known standard technique. The LED is fed by a fast (a few ns wide) pulse of variable amplitude adjusted to produce in average less than one photo-electron (PE) per pulse. Usually, the charge spectrum of single PE is analyzed either on-line or off-line to measure the mean value of the charge for single PE pulses. However, a practical difficulty arises when the PMT gain is not high enough to make the single photoelectron peak clearly visible in the charge spectrum. Another difficulty comes from the fact that the typical resolution of commercial ADCs (0.25 pc) is inadequate to measure the charge of PMT pulses with only one photoelectron which are, depending on the HV applied to the PMT base, of the order of one pc. These difficulties are more serious as one operates the PMT at relatively low voltages with gains lower than or of the order of 5x10 6. In this note we describe a test setup built to characterize PMT, its size was designed to accommodate small PMT like the Hamamatsu H7546 [1], to be used in the muon veto system of the Angra detector [2], up to large ones, like the Hamamatsu R5912 [3], to be used in the neutrino detector of Angra [2]. We also describe a simple method to obtain PMT gains that is free of the inconveniences mentioned in the previous paragraph. This technique makes use of a digital oscilloscope to capture the complete PMT traces into a PC. The voltage traces of thousands of LED-induced PMT pulses are recorded on the hard disk of a PC and subsequently analyzed off-line to extract their charge, amplitude and rise time. Finally, these data are analyzed to measure the charge spectrum of single PE. We illustrate the use of this technique by measuring the gain of a Hamamatsu R5912 PMT operated at 1500 V. 2 Experimental Setup The experimental setup schematic is shown in Figure 1, with its picture in Figure 2. The R5912 PMT is placed inside a sealed cylindrical chamber [7] consisting of a PVC tube with a thickness of 3mm, an internal diameter of 300mm and a height of 400 mm. The chamber has PVC covers. A schematic view of the dark chamber is shown in Figure 3 while Figure 4 shows the way in which the PMT and LED are placed inside the chamber. The L-shaped bottom is glued to the cylindrical tube while the L-shaped top cover is movable. The tube and covers are internally painted with black seal and externally covered with aluminum foil, totally involving the borders to improve light-tightness and electromagnetic shielding. This tube is in turn placed inside another cylinder made out of a 2mm thick aluminum sheet (400 mm by 900 mm with its borders tied together by an aluminum profile), with two aluminum covers also 2mm thick, Three rods with hexagonal nuts are used to press the aluminum covers against the cylinder, in addition to reinforcing light-tightness, this structure serves at the same time as a Faraday cage to provide electromagnetic shielding. A blue LED was placed inside the chamber on top of the photocathode. A cardboard fixed with insulating ribbon to the chamber tube was placed between the LED and the PMT, with a small hole to constraint the photons from the LED to reach only the central region of the photocathode. Cables for LED, PMT HV and PMT signal passed into the chamber through a hole on the chamber tube located near the top, this hole is subsequently sealed with black tape, cooper and insulating ribbons to ensure light tightness. The PMT we used possess a built-in voltage divider in its base.

AngraNote 003-2007 2 Figure 1 Setup schematic diagram Figure 2: Experimental setup photograph, from left to right: dark chamber with the PMT and LED inside; HV power supply Keithley 246; pulse generator Tektronix PG508; digital oscilloscope Tektronix TDS 1012B; IBM personal computer, with Microsoft Windows XP. Al Cover LED PVC cover Al tube PVC vase PMT under test Al Cover r rod w/ screw PVC vase Aluminum Profile Joint Al tube FOAM LED Figure 3a: Schematic view of the dark chamber built to house the PMT and the LED. Figure 4: Schematic view of the PMT and LED placed inside the dark chamber.

AngraNote 003-2007 3 We found the R5912 PMT signals pretty much noiseless once we operated the PMT inside the dark chamber just described, as can be seen in Figure 4, where in the oscilloscope screen the upper trace is the PMT signal output at 5 mv/div, and the lower trace is the signal send to the LED from a commercial pulse generator (Tektronix PG508) or, alternatively, with a home-made generator which will be described in separate technical note in the near future. Both pulse generators provide very short pulses (widths lower than 10 ns) at a controllable rate of 10 Hz. The stability of the LED pulser is a pre-requisite to obtain a Poisson distribution for the number of photoelectrons for long data-taking periods (usually 5-10 h are enough).the digital oscilloscope we used was a Tektronix TDS 1012B connected to a USB port of the PC. The maximum data acquisition rates obtained with the USB oscilloscope and the LabView-based DAQ system was around 5 Hz to transfer the 2500 point trace data into the PC for every event of driven the LED. Figure 5a: PMT with no light. Figure 5b: PMT output signal for one photon. Figure 5: Digital oscilloscope screen with LED pulse signal below ( 1 V/div) and above the PMT output signal (5 mv/div), horizontal scale 25 ns/div The charge, amplitude and rise time from 10 to 90% were subsequently measured off-line for every event by using another Lab View program. These parameters were analyzed and plotted using root [5]. 3 Results and Discussion Figure 6 shows a print-out of the front panel of the Lab View program used off-line to measure charge, amplitude and rise time of the oscilloscope traces previously recorded onto the PC s hard disk. This particular event shows an event with a single PMT pulse. The upper curve is the PMT trace (vertical scale in V) and the lower one the integrated charge (vertical scale in pc). Likewise, Figure 7 and Figure 8 show events with two and three pulses, respectively. Figure 8 shows raw data for charge vs rise time for a set of events taken on August 31th, 2007 for the Hamamatsu R5912 PMT operated at 1500 V. Throughout this note, this rise time is defined as the time it takes for the charge to go from 10% to 90% of its final value using an integration window of 150 ns; this window may contain zero, one o more single PE pulses. The accumulation of points on the lower left side region corresponds to events which contain only one

AngraNote 003-2007 4 PE. Figure 9 shows the distribution of amplitude versus rise time for the same data. Figure 10 shows the distribution of rise time for events with amplitude greater than 5 mv. The single peak between 0 and 15 ns corresponds to single photoelectron pulses, the flat distribution above 15 ns up to around 120 ns is due to our using a slow LED, i. e., the probability for a second PE pulse to occur within 100 ns after a first PE pulse is approximately constant. Figure 6: Event with a single isolated PMT pulse. The upper curve is the PMT trace in V and the lower one the PMT output charge in pc. The horizontal scale is 25 ns/div. Figure 7: Event with two isoated PMT pulses. The upper curve is the PMT trace in V and the lower one the charge in pc. The horizontal scale is 25 ns/div. are measured. Therefore, the technique illustrated in this note can be applied to PMTs with gains well below 107, where it would be impossible to measure gains with a commercial ADC because the typical 0.25 pc resolution is inadequately coarse.

AngraNote 003-2007 5 Figure 8: Event with three isolated PMT pulses. The upper curve is the PMT trace in V andthe lower one the charge in pc. The horizontal scale is 25 ns/div. Figure 9: Charge vs risetime from 10 to 90% for a set of events taken on August 31th., 2007. Figure 10: Amplitude vs risetime (from 10% to 90%) for the same data shown in Figure 9. Figure 11 shows the raw distribution of charge (integrated off-line as illustrated in Figure 6, the termination resistor used to match impedances is 50 -, as usual). Likewise, Figure 11 shows the distribution of pulse amplitudes for events with Q >.1 pc and risetime > 15 ns. In order to select 1- PE pulses we cut on risetime lower than 15 ns, as shown in Figure 7 and Figure 9. The result is shown in Figure 13 where the peak for a single PE is clearly visible at a charge of 2.2 pc as indicated by the gaussian fit superimposed (this not so good fit is only meant to show the position of the most probable value for the charge duistribution). We can also exclude 1-PE events by requiring the risetime to be greater than 15 ns, this is shown in Figure 14. The mean value of this peak is twice the charge value for single PE events; this means that two photoelectron events dominate in this plot.

AngraNote 003-2007 6 Figure 11: Risetime for events with V > 5 mv. The single peak between 0 and 15 ns corresponds to single photoelectron pulses, the flat distribution above 15 ns is due to our using a slow LED. Figure 12: Pulse amplitude for events with Q >.1 pc and risetime > 15 ns. Figure 13: Pulse amplitude for events with Q >.1 pc and risetime > 15 ns. Figure 14: PMT charge for events with Q >.1 pc and risetime < 15 ns. Figure 15: PMT charge for events with Q >.1 pc and risetime > 15 ns. The most probable value is twice that of a single PE.

AngraNote 003-2007 7 4 References [1] See http : //www.sales.hamamatsu.com/en/products/electron tube division/detectors/photomultiplier modules/part h7546b 200.php [2] J. C. Anjos et al., Braz. J. Phys. vol.36 no.4a So Paulo Dec. 2006. [3] See http : //www.sales.hamamatsu.com/en/products/electron tube division/detectors/photomultiplier tubes/part r5912.php [4] See http : //www.ni.com/labview/ [5] http : //root.cern.ch/ [6] http : //www.tek.com/products/oscilloscopes/tds1000tds2000/index.html [7] Vaz, Mario - Class notes on Basic Instrumentation in High Energy Physics, CBPF, May 2007