Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Similar documents
BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Digital Breast Tomosynthesis from a first idea to clinical routine

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Dosimetry in digital mammography

Contrast-Enhanced Digital Mammography

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

Contrast Enhanced Spectral Mammography (CESM) Updates

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

Update on the ACR FFDM QC Manual

Mammograms Obtained with Rhodium vs Molybdenum Anodes: Contrast and Dose Differences

Mammography. What is Mammography? What are some common uses of the procedure?

Mammography Quality Control: A Refresher

Hacia la imagenología tomográfica de mama

Patient Dosimetry in Mammography and Tomosynthesis:

Update on MQSA and Mammography Accreditation

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

CURRENT METHODS IN IMAGE GUIDED BREAST BIOPSY

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company

Updates in Mammography. Dr. Yang Faridah A. Aziz Department of Biomedical Imaging University Malaya Medical Centre

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

Epworth Healthcare Benign Breast Disease Symposium. Sat Nov 12 th 2016

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Mammography. What is Mammography?

Mammography limitations. Clinical performance of digital breast tomosynthesis compared to digital mammography: blinded multi-reader study

Breast Imaging & You

Introduction 1. Executive Summary 5

Breast Tomosynthesis An additional screening tool in the fight against breast cancer

Direct half value layer measurements in mammography - is near enough good enough?

Stereotactic Breast Biopsy

The Radiology Aspects

Accreditation Case Review: Mammography and Stereotactic Biopsy

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

Innovations and Applications of Tomosynthesis. Andrew D. A. Maidment, Ph.D. University of Pennsylvania Department of Radiology

Breast positioning system for full field digital mammography and digital breast tomosynthesis system

ALARA in mammography screening. Hilde Bosmans et al.

Translating Protocols Across Patient Size: Babies to Bariatric

Excellence by experience Analog Mammography. MAMMOMAT 1000 and MAMMOMAT 3000 Nova. Answers for life.

Breast Imaging & You

The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 5000

Consensus Guideline on Image-Guided Percutaneous Biopsy of Palpable and Nonpalpable Breast Lesions

Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition

Hong Kong College of Radiologists Mammography Statement

Since its introduction in 2000, digital mammography has become

Breast Tomosynthesis. What is breast tomosynthesis?

Improving Methods for Breast Cancer Detection and Diagnosis. The National Cancer Institute (NCI) is funding numerous research projects to improve

Mammographic imaging of nonpalpable breast lesions. Malai Muttarak, MD Department of Radiology Chiang Mai University Chiang Mai, Thailand

Mammography: Analysis and Advanced Techniques $50. Special Price. 8 credit seminar October 28, 2017 Knoxville, TN. a division of Herzing University

Macrobiopsy under X-Ray Guidance

PURPOSE IMAGE-GUIDANCE MODALITIES IMAGE-GUIDED BREAST BIOPSY. US-Techniques. Ultrasound. US guided NLOBB. TH. Helbich

Min Jung Kim Department of Medicine The Graduate School, Yonsei University

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

Here are examples of bilateral analog mammograms from the same patient including CC and MLO projections.

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

Mammography: Analysis and Advanced Techniques

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis =

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

About me 10/9/2018. Digital Breast Tomosynthesis (DBT)

WOMEN S HEALTH SOLUTIONS. The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 8000

Prone table stereotactic breast biopsy

Acknowledgments. A Specific Diagnostic Task: Lung Nodule Detection. A Specific Diagnostic Task: Chest CT Protocols. Chest CT Protocols

Diagnostic Medical Physicist Via Christi Hospitals Wichita, Wichita, KS

Breast Imaging Essentials

Frank: Merrill's Atlas of Radiographic Positioning & Procedures, 12th Edition

Computer-Aided Diagnosis for Microcalcifications in Mammograms

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

-Ray Producing Equipment

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography

Survey of Mammography Practice: Initial Results from Serbia

Molecular Breast Imaging: History and Recent Developments

AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS

X-Ray & CT Physics / Clinical CT

Dedicated to Mammography

Breast Health and Imaging Glossary

Radiographic techniques in screen-film mammography

Andrew Karellas, PhD

Clinical evaluation of breast dose and the factors affecting breast dose in screen-film mammography

Tomosynthesis and breast imaging update. Dr Michael J Michell Consultant Radiologist King's College Hospital NHS Foundation Trust

Breast CT and Dosimetry

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care.

CT Quality Control Manual FAQs

ACR MRI Accreditation: Medical Physicist Role in the Application Process

ACCREDITATION DOCUMENT THE RADIOGRAPHER

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology

Physics of MBI (~10 slides)

Contrast Enhanced Spectral Mammography (CESM) Initial UK Experience. Dr Sarah L Tennant BMedSci, BMBS, MRCP, FRCR

Diagnostic Dilemmas of Breast Imaging

Policy Library Clinical Advantages of Digital Breast Tomosynthesis in Symptomatic Patients

Contrast-Enhanced Spectral Mammography

CT SCAN PROTOCOL. Shoulder

Cone-Beam CT for MSK Extremities

Breast Care Unit. 1. The triple assessment means that your breast will be examined by a doctor trained in breast disease.

Computed tomography Acceptance testing and dose measurements

Transcription:

Disclosures Stereotactic Biopsy vs Mammography: and Dose None Vikas Patel, PhD, DABR Upstate Medical Physics 2014 Annual Meeting The American Association of Physicists in Medicine Austin, TX Learning Objectives Outline quality and dose characteristics of Stereotactic Biopsy (SBB) systems Screen-Film and Full-Field Digital Mammography systems Introduction quality characteristics of: Stereotactic Biopsy (SBB) systems Screen-Film mammography (SFM) systems Full-field digital mammography (FFDM) systems Patient dose: Average Glandular Dose (AGD) Summary Introduction Introduction Mammograms show abnormalities such as a suspicious solid mass, microcalcification clusters etc. Biopsy done to determine the abnormalities are benign or malignant Reduces number of benign open surgeries Types of breast biopsy: Fine Needle Aspirations (FNA) Vacuum Assisted (VA) Core Needle (CN) Open Surgical http://www.cancer.gov/cancertopics/screening/understanding breast changes/page6 1

Introduction -guided Biopsy Accuracy of localization of the lesion is critical in biopsy procedures It depends on: Accuracy of the equipment used Sample volume size Skill of the clinician Grid coordinate system Uses dedicated mammography equipment Only 2D localization lack of depth accuracy Ultrasound-guided biopsy systems Lack of visualization and accuracy of localization Stereotactic breast biopsy systems (SBB) Stereotactic Biopsy (SBB) Systems Stereotactic Biopsy (SBB) Systems Localization in 3D and accurate to within 1mm Less dependent on operator skills compared to other methods Limitations include localization of lesions: Widely scattered throughout the breast Near the chest wall or high in axilla In very thin breasts courtesy Siemens, Hologic and IMS (Giotto) Stereotactic Biopsy (SBB) Systems Stereotactic Biopsy (SBB) Systems Based on a zero-degree scout and two radiographs 30 degrees apart (+15 deg) Carr et. al., Radiographics, 2001:21:463 473 www.radiologictechnology.org 2

Requirements System Design Mammography SBB (prone systems) Mammography Detection and characterization of abnormalities SBB Localization, targeting and sampling the abnormalities Mo/Mo, Mo/Rh, Rh/Rh, W/Rh, W/Ag Large target angle 0.1 and 0.3 mm focal spot Grid for scatter removal (contact mode) Strong compression Not much geometric blur in contact mode SID ~65cm Mo/Mo and Mo/Rh Steep target angle 0.25 mm focal spot Air gap for scatter removal Light Compression Possible high geometric blur SID ~88 cm System Design Digital Receptors Mammography SBB (prone systems) 18 cm x 24 cm and 24 cm x 30 cm FOV Film screen, direct and indirect flat-panel receptors 100 micron pixel size (digital systems) 5 cm x 5 cm FOV -based lens-coupled or fiber-optic coupled receptors 50 micron effective pixel size Indirect Flat Panel Yaffe M. J. (2010), Digital Mammography, Bick U. and Diekmann, F. (Eds.), Springer, p. 13 31 Digital Receptors Digital Receptors Direct Flat Panel Detectors Yaffe M. J. (2010), Digital Mammography, Bick U. and Diekmann, F. (Eds.), Springer, p. 13 31 image courtesy R. J. Pizzutiello 3

System Design System Design Side View Front View Side View Front View Focal Spot X-ray Tube Focal Spot X-ray Tube Small area Collimator Small area Collimator Compressed Compressed Lens Optical coupling/mirror system Light reflection from phosphor Mirror Phosphor A/D DMA courtesy R. J. Pizzutiello Phosphor Fiber Optic 2:1 fiberoptic taper demagnification Light transmission through phosphor A/D DMA courtesy R. J. Pizzutiello Type of SBB systems Outline Cost and space Less Patient fatigue and motion Less Patient fear No patient weight issue Better for handicapped patients Access to the breast Prone Systems Add on Upright Systems Introduction quality characteristics of: Stereotactic Biopsy (SBB) systems Screen-Film mammography (SFM) systems Full-field digital mammography (FFDM) systems Patient dose: Average Glandular Dose (AGD) Summary Stroke margin safety Overall imaging performance Characteristics Factors Affecting Target/ Filter Digital Display Receptor Processing Scatter 4

Characteristics Screen-film systems Acceptable density range Acceptable exposure range Yaffe, M. J., RadioGraphics 1990, 10: 341 363 4096 Pixel Value Digital receptors Dynamic range depends upon bit depth processing Digital displays Window width and level functions 0 Exposure Characteristics Geometric blur Higher in SBB system compared to mammography systems Increased mag due to air gap No compression in the open part of the compression paddle Patient motion blur SBB procedures take significantly more time compared to typical mammography procedures 5

receptor blur SBB: 50 micron pixel (7-10 lp/mm in 1024x1024 mode) Screen-film: grain size dependent (15-20 lp/mm) Digital flat panel: approx. 100 micron pixel (8-10 lp/mm) Matrix size 512 x 512 has less resolution than 1024 x 1024 Yaffe M. J. (2010), Digital Mammography, Bick U. and Diekmann, F. (Eds.), Springer, p. 13 31 Characteristics Yaffe M. J. (2010), Digital Mammography, Bick U. and Diekmann, F. (Eds.), Springer, p. 13 31 Electronic dark noise for digital systems Speed is an issue for screen-film systems eg. Kodak s dual emulsion screen-film system Technique factors 58 mas Effects of mas 28 kvp 113 mas 227 mas 339 mas courtesy R. J. Pizzutiello 6

Characteristics Dead pixel Read out failure Ghosting Dust Gridlines Patient accessories 7

Outline Patient Dose Introduction quality characteristics of: Stereotactic Biopsy (SBB) systems Screen-Film mammography (SFM) systems Full-field digital mammography (FFDM) systems Patient dose: Average Glandular Dose (AGD) Summary Lamborghini Huracan LP 610 4 http://abcnews.go.com/blotter/dear abc news fixer york owes money/story?id=19980537 Average Glandular Dose (AGD) Factors Affecting AGD Dose index to estimate average dose to the glandular tissue just like CTDI is in CT Displayed on the units and calculated periodically Standard breast (4.2 cm compressed thickness and 50% Fatty and 50% Glandular composition) equivalent phantom Repeat/ Retake Spectrum Dose Technique Factors www.gammex.com/n-portfolio/productpage.asp?id=299&category= Mammography&name=Mammographic+Accreditation+Phantom%2C+Gammex+156 Density Thickness Average Glandular Dose (AGD) Outline Typical doses for one cc view SF systems 180 250 mrad Digital systems 100 190 mrad Use of AEC and proper technique chart along with less repeats will minimize patient dose Introduction quality characteristics of: Stereotactic Biopsy (SBB) systems Screen-Film mammography (SFM) systems Full-field digital mammography (FFDM) systems Patient dose: Average Glandular Dose (AGD) Summary 8

Summary SBB systems have different requirements compared to mammography systems which leads to unique system design quality characteristics were reviewed Factors affecting dose were reviewed Future? Screen Film SBB (Prone systems) FFDM 15 20 lp/mm 7 10 lp/mm (1024 mode) 8 10 lp/mm (contact mode) Display Viewbox CRT: 480x640, 0.4 mm x 0.4 mm pixel pitch Flatpanel/LCD: 1280x1024 (1.3 MP), 0.285 mm x 0.285 mm pixel pitch, 550:1 contrast ratio Flatpanel/LCD: 2560 x 2048 (5MP), 0.165 mm x 0.165 mm pixel pitch, 850:1 contrast ratio Future? FFDM SBB (Prone systems) SBB from Add ons Side View Front View Focal Spot X-ray Tube Small area Collimator Compressed Mirror Lens Phosphor Optical coupling/mirror system Light reflection from phosphor A/D DMA Side View Front View Focal Spot X-ray Tube Small area Collimator Compressed Phosphor Fiber Optic 2:1 fiberoptic taper demagnification Light transmission through phosphor A/D DMA 9