Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Similar documents
Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

LIPASE liquicolor. Design Verification. Multipurpose Reagent

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

liquicolor (AMP Buffer, IFCC) Design Verification

Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #:

For In Vitro Diagnostic Use

liquicolor Design Verification

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

BIOO RESEARCH PRODUCTS. Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #:

TABLE OF CONTENTS GENERAL INFORMATION... 1

METHOD VALIDATION: WHY, HOW AND WHEN?

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

BIOO LIFE SCIENCE PRODUCTS

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

Standard Operating Procedure

Patient: 55 y female (ambulatory)

3 Linearity and Detection Limit Linearity Detection limit... 4

Reference Material Institute for Clinical Chemistry Standards (ReCCS)

METHOD VALIDATION CASE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

Hexagon PSA. Design Verification. Contents

Development and Validation of Stability Indicating HPTLC Method for Estimation of Seratrodast

ab Sphingomyelin Assay Kit

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

SCIEX Vitamin D 200M Assay for the Topaz System

Evaluation of new MiniCollect Z Serum (Separator) Tubes

Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer

Blood Urea Nitrogen Enzymatic Kit Manual Catalog #:

Delta Check Calculation Guide

C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

See external label 2 C-8 C = C-REACTIVE PROTEIN (CRP) LATEX SLIDE TEST

The NGAL Turbidimetric Immunoassay Reagent Kit

Total Thyroxine ELISA (T4)

Total Bilirubin Assay Kit Manual Catalog #:

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

ab Factor Xa Activity Assay Kit (Fluorometric)

Determination of hemoglobin is one of the most commonly

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

ab HMG-CoA Reductase Activity Assay Kit (Colorimetric)

The Virtues and Pitfalls of Implementing a New Test

Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent

See external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

DetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine

METHOD VALIDATION CASE

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA

Dr Bill Bartlett Blood Sciences, Ninewells Hospital & Medical School, NHS Tayside, Scotland, UK.

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Kit for assay of thioredoxin

Internal Validation Guide of Y-STR Systems for Forensic Laboratories Printed in USA. 11/12 Part# GE713

EXPERIMENT 3 ENZYMATIC QUANTITATION OF GLUCOSE

Investigating the recheck rules for urine analysis in children

Guide to Fulfillment of Measurement Uncertainty

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Creatinine (serum, plasma)

Blood Urea Nitrogen Enzymatic Kit Manual Catalog #:

HumaTex CRP. Design Verification. Contents

ab65344 Uric Acid Assay Kit (Colorimetric/Fluorometric)

Aina Blood Monitoring System

25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL

Bias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University

Technical Bulletin No. 161

Design Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...

QUALITROL 2H Insert. 01 Inglês - Ref.: 72. Ref.:72. Precautions and warnings

Research Article Derivative Spectrophotometric Method for Estimation of Metformin Hydrochloride in Bulk Drug and Dosage Form

Validated UV Spectrophotometric Method Development And Stability Studies Of Acamprosate Calcium In Bulk And Tablet Dosage Form

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Reference Material Institute for Clinical Chemistry Standards (ReCCS)

100 Test Cuvette Assay for 2-Thiobarbituric Acid Reactive Substances (TBARS) For Research Use Only INTRODUCTION H +

ab Creatinine Assay Kit (Colorimetric)

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

CLINICAL CHEMISTRY REAGENTS. Product Profile

ab Lipoxygenase Inhibitor Screening Assay Kit

Enzyme Immunoassay for

The Adiponectin Turbidimetric Immunoassay Reagent Kit

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

Bovine Insulin ELISA

Validation Case studies: the good, the bad and the molecular

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.

Instructions for Use. Tg Antibody ELISA

Standardization of. S.M.Boutorabi DCLS,PhD

Exendin-4 (Exenatide) ELISA Kit

Single Cell Quantitative Polymer Chain Reaction (sc-qpcr)

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Human Creatinine Urinary Detection Kit

Galactose Colorimetric Detection Kit

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase

Transcription:

Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo, Ying Wang, Fanjun Kong, Rui Wang Beijing You an Hospital Clinical Laboratory Center, Capital Medical University, Beijing 100069 Received August 25, 2014; Accepted October 30, 2014; Epub November 15, 2014; Published November 30, 2014 Abstract: Objective: This study was designed to evaluate whether the high concentration of bilirubin is able to interfere the determination of alkaline phosphatase (ALP). Methods: Clinical tests evaluating the interfering substance of bilirubin of various concentrations on the determination of ALP were conducted based upon the Clinical and Laboratory Standards Institute (CLSI) document EP7-A2, the most recent guideline on interference testing approved in 2005. Results: Paired t-test comparing different doses of bilirubin revealed that the concentration of 1 000 µmmol/l bilirubin negatively interfered the determination of ALP levels. The experiment designed with five different concentrations of bilirubin showed that bilirubin can exert negative interference on the measurement of ALP in a linear pattern. Conclusion: High concentration of bilirubin can cause false measurement of ALP levels, probably interfering with the clinical prognosis of liver diseases. Keywords: Alkaline phosphatase, bilirubin, interference Introduction Alkaline phosphatase (ALP) is a routine laboratory test used to screen for liver disease and elevated levels are often the first indication of chronic hepatic lesions, which is also an important index for evaluating the severity and prognosis of liver illnesses [1]. Recent data also indicated that elevated serum alkaline phosphatase levels are associated with increased mortality in patients with metabolic syndrome and they also suggested that higher levels of serum alkaline phosphatase is a predictor of mortality independent of the baseline prevalence of metabolic syndrome [2]. Hyperbilirubinemia is one of the common symptoms encountered by patients with liver diseases. A variety of previous investigations have proven that high concentrations of bilirubin exert significant interference on the determination of multiple biochemical substances [3-5]. This paper is designed to investigate whether high levels of bilirubin are able to pose interference upon the detection of ALP in patients with hyperbilirubinemia in laboratory settings, thereby affecting the clinical diagnosis and prognosis of hepatic diseases in clinical practice. The laboratory procedures of the interference test were conducted strictly according to the Clinical and Laboratory Standards Institute (CLSI) document EP7-A2, the most recent guideline on interference testing approved in 2005 by the U.S. Committee for Clinical Laboratory Standards [6]. Materials and methods Instruments and reagents Instruments Olympus AU5400 analyzer (Olympus America, Melville, NY). It has been verified that the linear range of total bilirubin is: 0-1009 μmol/l. Reagents The purity of standard bilirubin was 99%. The ALP activity was measured using an ALP kit based upon AMP buffermethod. The levels of total bilirubin kits were measured by total bilirubin kits by employing the Diazonium method.

Table 1. Number of replicates corresponding to /S /S No. of replicates /S No. of replicates 0.8 41 1.5 12 1 26 1.6 10 1.1 22 1.8 8 1.2 18 2 7 1.3 16 2.5 5 1.4 14 3 3 Note: where (within-run) standard deviations were calculated from 20 times repeated measurements of the control samples; is the maximum allowable interference when measuring concentration. Methods The end-point method was performed for the measurement of ALP and total brilirubin on the biochemical analyzer. Preparation of bilirubin stock solution A total of 60 mg bilirubin was accurately weighed and added into a test tube. Then 5 ml of 0.4% NaOH solution was supplemented and constantly shaken for even mixture. When approaching the presence of complete dissolution of bilirubin, 0.4% NaOH solution was supplemented by droplets until the presence of full dissolution. The saturated bilirubin solution was prepared and stored at -20 C freezer for subsequent preparation. After calculation, the concentration of bilirubin in the experiment was expected to be approximately 13 650 μmol/l. Considering the degradation property of bilirubin when placed under light, the final concentration of bilirubin in this experiment was determined as the mean value of 10 times measurements. Preparation of basic samples The concentration of the test specimens should be designed to include high, low and medical decision levels. Fresh serum was obtained from healthy individuals who had not taken medication in the past 3 days. The serum samples should have no signs of hemolysis, jaundice or lipemia. Since bilirubin is an endogenous metabolite, therefore, the final concentration of bilirubin may be excessively high, which probably influences the credibility of the experimental outcomes. To properly prevent the incidence of such events, the concentration of bilirubin in the basic samples was lower than 10 μmol/l. The final concentrations of ALP and bilirubin in the mixed serum were determined by laboratory instruments. The respective serum samples with high and low levels of ALP were chosen and mixed with a rational proportion to adjust the final concentration of ALP up to 50, 150 and 400 U/L strictly in accordance with the medical decision level proposed by Stander et al. Preparation of the control and test samples Based upon the 3-year clinical experience in our hospital, the limit of high concentration of bilirubin was approximately 1 000 μmol/l. Therefore, the bilirubin level of the test samples was adjusted to 1 000 μmol/l. As required by document EP7-A2, the proportion of diluted sample should be controlled under 5%, suggesting that the concentration of interfering substances should be 20 times (or above) higher compared with the test samples. However, the maximal level of bulirubin of the test samples (saturation samples) did not exceed 13 650 μmol/l, significantly lower than the required level of 20 000 μmol/l. Thus, the volume of the undiluted solution was approximately 1/13 of the total volume and the sample concentration was diluted by approximately -7.7%. Interfering sample preparation: 1200 μl basic sample + 100 μl stock solution of bilirubin. Control sample preparation: 1200 μl basic sample + 100 μl 4% NaOH. Number of replicates The number of replicates was determined by initially calculating the ratio of the maximal allowable error ( ) and the within-run standard deviation (s) and then identified the number of replicates according to corresponding /s, as indicated in Table 1. In this study, the maximum allowable error was determined based upon the biological variation, and the total error (TE) of ALP achieved 11.7% [1]. The of each item was equal to the means of the basic sample * the maximum allowable error %. 4245 Int J Clin Exp Med 2014;7(11):4244-4248

Table 2. Preparation of dose-response sample formulation Sequence No. Formulation Estimated concentrations of total bilirubin (μmol/l) 1 (L) Low Pool 1300 μl < 10 2 (3L + 1H) Low Pool 300 ul+ High Pool 100 μl 250 3 (2L + 2H) Low Pool 200 μl+ High Pool 200 μl 500 4 (1L + 3H) Low Pool 100 μl + High Pool 300 μl 750 5 (H) High Pool 1300 μl 1000 Table 3. Evaluation of the interfering effects of total bilirubin on the determination of ALP Level Base pool (mmol/l) Total allowable error (TE%) Withim SD (S) /S Number of replicates Control pool (U/L) Test pool (U/L) dobs dc Interference judgment LOW 53 11.70% 6.20 0.52 11.93 3 49.5 13.1-36.4 0.49 Yes MED 165.6 11.70% 19.38 1.06 18.28 3 152.2 118.4-33.8 1.01 Yes HIGH 433.3 11.70% 50.70 3.41 14.87 3 396.5 369.3-27.2 3.24 Yes Table 4. Interference effects of high concentrations of bilirubin on the measurement of ALP levels Control sample (U/L) Test sample (U/L) Interference effects error Imprecision Bias Total allowable error LOW 49.5 13.1-73.54% 3.20% 6.40% 11.70% MED 152.2 118.4-22.21% HIGH 396.5 369.3-6.86% Sample analysis The test (T) and control (C) samples were analyzed in an alternating order, e.g. C1T1C2- T2C3T3...CnTn. To prevent the system affected by the test samples, deionized water was additionally supplemented to protect the control samples influenced by the test samples. As shown in the formula C1T1CxCxC2T2CxCxC3T3...CxCxCnTn, the results of the additional control sample (Cx) were eliminated. The concentrations of bulirubin and ALP were detected simultaneously. Data analysis The obtained data were subject to point estimation in terms of the interfering effect by comparing the d obs (defined as the difference between the average values of the test and control samples) and the cut-off value (d c ) using the following equation, d obs =x test -x control. The interfering effect existed when d obs >d c. Note: n is the actual sample size. The cut-off value (d c ) can be computed for a two-sided test using the following equation: d c =(d null + sz 1-α/2 )/ n Note: d null was the value stated in the null hypothesis, constantly equal to 0. The 95% confidence interval (CL) of the interfering effect can be calculated according to the following equation 95% Confidence Interval (CL)=( _ x test - _ x control ) ± t 2 s 2 0.975, n-1 n Note: s is within-run standard deviation, n is the number of replicates for each sample, t 0.975, n-1 is taken from a Student t-table as the 97.5 th percentile of t-distribution with n-1 degrees of freedom. (For n > 30, substituting 2.0 for t 0.975, n-1 is a reasonable approximation). Preparation of dose-response samples The dose-response experiment determines the relationship between the interfering concentration and the magnitude of interference, which permits the estimation of the effect at any interfering concentration within the range tested, as illustrated in Table 2. The low concentration samples can be prepared by referring to the preparation method for the control samples; The high concentration samples can be prepared according to the procedures for the test samples. 4246 Int J Clin Exp Med 2014;7(11):4244-4248

Figure 1. Dose-response experiment of bilirubin interference (low ALP). Figure 3. Dose-response experiment of bilirubin interference (high ALP). Evaluation of interfering effects Different concentrations of interfering errors can be estimated by linear regression analysis (Figures 1-3). Discussion Figure 2. Dose-response experiment of bilirubin interference (medium ALP). All samples were analyzed in a random order to avoid run-to-run variables A total of five concentrations were analyzed within the same analytical run. The first set of replicates was analyzed in an ascending order, the second set in a descending order and the third set in an ascending order, etc., with the purpose to average out any systematic drift effects [2]. SPSS16.0 statistical software SPSS16.0 statistical software was utilized for data analysis. Results Evaluation of the interference effects of total bulirubin was illustrated in Table 3. The interference effects of high concentrations of bilirubin on the detection of ALP levels were shown in Table 4 below. Dose-response experiment of bilirubin interference ALP. New protocols for measuring interfering substances are needed. These could be developed by a Clinical and Laboratory Standards Institute (CLSI) committee on interfering substances. The most recent CLSI guideline on interference testing in clinical (document EP7-A2) was approved in 2005 [6]. The testing of both interfering substances as well as analytical accuracy might be best performed by a notified body instead of the monitor s manufacturer. The evaluation scheme can be widely used in clinical laboratory testing methods, instruments and various types of specimens, especially for analyzing the potential interfering substances. In this study, paired-difference experiments at higher concentrations of potential interfering substances (bilirubin) for initial screening. After verifying that bilirubin is the interfering substance, dose-response experiments were conducted to evaluate the association between the concentration of interfering substances and the degree of interference. Whether the interfering effects of bilirubin upon ALP level was equally evaluated objectively. As shown in Table 4, high concentrations of bilirubin played a role in the negative interference of ALP levels from low- and moderate-concentration samples were -73.54% and -22.21%, respectively, which exceeded the total allowable error of 11.7%. In this experiment, the outcomes and regression formulas proved that the interfering effects caused by different concentrations of bilirubin can be estimated. 4247 Int J Clin Exp Med 2014;7(11):4244-4248

These results suggest that when patients were clinically diagnosed with hyperbilirubinemia, ALP alone should not be directly utilized as a diagnostic parameter if the interfering effects of bilirubin can not be confirmed. Bilirubin probably inhibits the enzyme activity, especially by exerting negative interfering effects since bilirubin is a type of reducing agent, which is able to weaken electron and proton transmission [7]. This study indeed has certain limitations as below: First, the substance of bilirubin supplemented to the serum samples may differ from the endogenous bilirubin yielded within human body. Second, the test samples used in this experiment are unable to represent the samples in clinical settings. Third, the concentration of the test samples may be excessively low, high. Acknowledgements The ethics committee of Beijing You an Hospital approved this study. Disclosure of conflict of interest None. Address correspondence to: Dr. Zhong Wang, Department of Clinical Laboratory Center, Beijing You an Hospital, Capital Medical University, Beijing 100069, China. Tel: 86010-87556698; Fax: 86010-87556698; E-mail: docwangzhong@126.com References [1] Yingwu Y, Yusan W, Ziyu S. National Clinical Laboratory Procedures [M] 3rd edition Nanjing: Southeast University Press; 2006. 80, 476,485-490. [2] Krishnamurthy VR, Baird BC, Wei G, Greene T, Raphael K, Beddhu S. Associations of Serum Alkaline Phosphatase with Metabolic Syndrome and Mortality, Am J Med 2011; 124: 566.e1-566.e7 [3] Hullin DA, McGrane MT. Effect of bilirubin on uricase-peroxidase coupled reactions. Implications for urate measurement in clinical samples and external quality assessment schemes Ann Clin Biochem 1991; 28: 98-100. [4] James DR, Price CP. Interference in colorimetric reactions for measuring hydrogen peroxide. Ann Clin Biochem 1984; 21: 398-404. [5] Cobbaert CM, Baadenhuijsen H, Weykamp CW. Prime Time for Enzymatic Creatinine Methods in Pediatrics. Clin Chem 2009; 55: 549-558. [6] McEnroe RJ, Burritt MF, Powers DM, Rheinheimer DW, Wallace BH. Interference testing in clinical chemistry; approved guideline second; 2005. [7] Pyles LA, Stejskal EJ, Einzig S. Spectrophotometric measurement of plasma 2-thiobarbituric acid-reactive substances in the presence of hemoglobin and bilirubin interference. Proc Soc Exp Biol Med 1993; 202: 407-419. 4248 Int J Clin Exp Med 2014;7(11):4244-4248