Methods DEVICES AND TECHNOLOGY

Similar documents
Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

DEVICES AND TECHNOLOGY

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

Home blood pressure measurement with oscillometric upper-arm devices

When the BHS dissolved its Working Party on Blood Pressure Measurement, the Working Group on Blood

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

a ARSMED (Association for Research and Development of Biomedical Received 11 October 2007 Revised 27 March 2008 Accepted 27 March 2008

A GUIDE TO OUR NIBP TECHNOLOGY

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

greater than 90% of patients having blood pressure measured consistently according to the protocol.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

doi: /fampra/cmw053

There is convincing evidence in clinical studies

Special articles from the ESH working group on blood pressure monitoring 23

The introduction of oscillometric automated blood pressure

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

In clinical practice, blood pressure (BP) treatment

Arm position and blood pressure: a risk factor for hypertension?

Validity study of oscillometric blood pressure measurement devices using an oscillometric waveform simulator

Home monitoring of blood glucose

Validation of Missouri Aneroid Sphygmomanometer to Measure Blood Pressure in Patients with Cancer

Validation Study Result Form

Device Judgement Recommended

Automated versus manual blood pressure measurement: A randomized crossover trial

Automatic oscillometric blood pressure devices are frequently

Validation Study Result Form

The increasing awareness of hypertension as a serious

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

The Value of a BP Determination Method Using a Novel Non-Invasive BP Device against the Invasive Catheter Measurement

a University of São Paulo School of Nursing, b Obesity Unit, Endocrinology Division

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

THE ACCURATE ASSESSMENT

STATE OF THE ART BP ASSESSMENT

Céline Guggiari a, Christophe Büla a, Katia Iglesias b and Bernard Waeber b,c

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

Biomedical Instrumentation E. Blood Pressure

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Data Collection Worksheet

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

Comparison Between Automated and Manual Sphygmomanometer for Measuring Blood Pressure in Children

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Comparison of blood pressure measured at the arm, ankle and calf

Assessing Blood Pressure Accuracy of an Aneroid Sphygmomanometer in a National Survey Environment

Maker Spacelabs Healthcare Manufacturer Spacelabs Healthcare. Brand Spacelabs Model OnTrak

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

Blood Pressure Measurement in SPRINT

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

An Automated System for Blood Pressure Determination During Exercise

Non-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Validation Study Registration Form

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

CONCORD INTERNAL MEDICINE HYPERTENSION PROTOCOL

B lood pressure is the pressure exerted by the circulating blood against vessel walls, and is a vital sign in clinical

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

Evaluation of high-definition and conventional oscillometric blood pressure measurement in anaesthetised dogs using ACVIM guidelines

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

External Oscillatory Blood Pressure - EOBPTM

A chronic increase in blood pressure is a major risk factor for cardiovascular disease, whereas reducing

Device Equivalence Evaluation Form

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Device Equivalence Evaluation Form

Accuracy and reliability of wrist-cuff devices for selfmeasurement

THE EFFECT OF CLOTHES ON SPHYGMOMANOMETRIC BLOOD PRESSURE: COMPARISON OVER SLEEVED ARM VERSUS UNSLEEVED (BARE) ARM

MX3 Plus English. Introduction. Table of contents. 1 How to obtain meaningful blood pressure readings

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Comparison of the Microlife A2 Basic (BP3GQ1-3P) with the Microlife BP3BT0-A and Microlife BP A100 Plus

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT

Device Equivalence Evaluation Form

Transcription:

DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. Design The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed. Methods For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. Results The difference between the two observers was 1 7 2 mmhg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were 1 7 7 and 5 7 6 mmhg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmhg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmhg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmhg for 94 of the systolic readings and 96 of the diastolic readings. Conclusions The Omron HEM-907 device passes the two phases of the international validation protocol. Blood Press Monit 7: 237 241 & 2002 Lippincott Williams & Wilkins. Blood Pressure Monitoring 2002, 7:237^241 Keywords: Omron HEM-907, blood pressure measuring device, validation, international protocol L Institut CardioVasculaire, Paris, France. Sponsorship: Supported in part by Omron Japan. Potential conflicts of interest: None Correspondence and requests for reprints to Professor Roland Asmar, L Institut CardioVasculaire, 21 Boulevard Delessert, 75016 Paris, France. Fax: þ33 1 5574 6665; e-mail: icv@icv.org Received 12 December 2001 Revised 22 April 2002 Accepted 26 April 2002 Introduction Banning the use of mercury in clinical practice because of its toxicity may lead to the total replacement of mercury sphygmomanometers by automatic blood pressure (BP) measuring devices in the near future. Prior to marketing, these devices should be assessed for safety, accuracy, and reliability [1]. In 1987, the Association for the Advancement of Medical Instrumentation (AAMI) [2] published a standard for electronic or automated sphygmomanometers that included a protocol for the evaluation of the accuracy of devices and in 1990 the protocol of the British Hypertension Society (BHS) [3] was published. These protocols, which differed in details, had a common objective, the standardization of validation of BP measuring devices [4]. Both protocols [5,6] have since been revised. Recently, a suggestion was made that it would seem timely, therefore, for the AAMI and the BHS to join together in producing a revised common protocol for the validation of BP measuring devices that might be acceptable as an international protocol. A proposal for the international validation protocol has been established during the session Devices and Validation of the First Consensus Conference on Self Blood Pressure Measurement [1,7]. The aim of our study was to validate the Omron HEM-907 device according to the proposed published international validation protocol [1,7]. Methods The tested device The Omron HEM-907 device records BP oscillometrically with an electrostatic capacity semi-conductor pressure sensor in the range 0 299 mmhg and heart rates in the range 30 199 beats/min. Systolic blood pressure (SBP), diastolic blood pressure (DBP) and heart rate are displayed on a digital display. The inflation is by an automatic pumping system and the deflation is by means of an automatic pressure-releasing electromagnetic control valve. Error codes (from 1 to 9) indicate when the device is malfunctioning or being used inappropriately. The unit is powered by an AC adapter (230 VAC, 50 Hz, output DC 1359-5237 & 2002 Lippincott Williams & Wilkins

238 Blood Pressure Monitoring 2002, Vol 7 No 4 8 V 500 ma). The unit s weight is approximately 910 g (32 oz) and it measures 139 mm (5 1/2 00 ) 203 mm (8 00 ) 131 mm (5 1/6 00 ) (width, height and depth, respectively). A cuff/bladder set M (applicable arm circumference: 22 32 cm or 8 2/3 00 12 3/5 00 ) is provided. Two other cuff/bladder sets, L (arm circumference 32 42 cm or 12 3/ 5 00 16 1/2 00 ) and set S (arm circumference 17 22 cm or 6 2/ 3 00 8 2/3 00 ) are optional. Device validation Observer training was achieved prior to the validation process. Two physicians (observer 1 and observer 2) were trained and certified on the basis of a CD-ROM [8] specifically developed by the French Society of Hypertension for the certification of observers involved in clinical studies. The validation of the Omron HEM-907 was performed according to the international protocol [1,7]. In a first step, validation was performed by the two observers on a group of 15 subjects, five subjects in each of the BP categories defined by the international guidelines [1,7]; then inclusion was carried out until 33 subjects in total had been included, with 11 subjects being in each BP category (for SBP: o 130, in the 130 160 range, and 4 160 mmhg; for DBP: o 80, in the 80 100 range, and 4 100 mmhg). According to the guidelines [1,7], a device failing the first phase of validation is eliminated from further testing. The subjects were seated in a quiet room and BP measurements started after a 5 10 min rest. All measurements were made on the left arm with a cuff adapted to the arm circumference. Blood pressure was measured simultaneously (Y tube) with two calibrated mercury sphygmomanometers by the two observers alternately with the automatic device. The observers were blinded to each other s readings. Measurements were carried out in the following sequence: BP1, mercury; BP2, device; BP3, mercury; BP4, device; BP5, mercury; BP6, device; and BP7, mercury. All pressures were recorded with the patient seated, with the arm being at heart level. For each subject, the device measurements BP2, BP4, and BP6 were first compared to the mean of the two observer measurements BP1, BP3, and BP5 respectively to form a set and then to observer measurements BP3, BP5, and BP7 respectively to form a second set. In total, 45 pairs of measurements were available for analysis for the primary phase and 99 pairs for the secondary phase of the international protocol. Accuracy criteria For each set and each observer, the number of testinstrument measurements differing from the mercury sphygmomanometer by 5, 10, and 15 mmhg or less were calculated separately for SBP and DBP. The three pairs from the set whose comparisons were more favourable to the device were retained. The number of differences in each class was compared to the number required by the international protocol. Table 1 shows the international protocol evaluation criteria. The difference (device observer), for SBP and DBP separately (using the data on which the final grade is based), was plotted against the mean of the device pressure and the mean observer pressure, using all 99 points [9]. In order to obtain 33 subjects to fulfil the inclusion criteria of the international protocol, 96 subjects in total were preincluded in the study. Also, 52 additional subjects (the first 52) out of the remaining 63, were selected on the basis of the BHS criteria in order to perform the analysis according to the AAMI [6] and BHS [5] guidelines; however analysis according to BP categories as recommended in the BHS protocol [5] was not performed since the number of patients required in each BP category was not achieved (Table 2). Table 1 The international protocol evaluation criteria for devices using sequential same-arm measurement r 5 mmhg r 10 mmhg r 15 mmhg Phase 1 At least one of 20 30 35 Phase 2 At least two of 50 75 90 All of 45 70 85 Table 2 BP ranges of the subjects according to the BHS categorization criteria SBP (mmhg) o 90 90 129 130 160 161 180 4 180 BHS (n) 8 20 20 20 8 Omron study (n) 0 32 37 12 4 DBP (mmhg) o 60 60 79 80 100 101 110 4 110 BHS (n) 8 20 20 20 8 Omron study (n) 1 27 46 7 4 SBP, systolic blood pressure; DBP, diastolic blood pressure; n, number of subjects.

Validation of the Omron HEM-907 device El Assaad et al. 239 Statistical analysis was performed using the NCSS s software (Number Cruncher Statistical Systems, Kaysville, Utah, USA). According to BHS 93 the grade is A for SBP and B for DBP (Table 4). The device also passes the validation criteria of the AAMI (Table 4). Results The mean age of the 33 subjects was 51.1 7 13.9 years (19 men and 14 women). The difference between the two observers was 1 7 2 mmhg for the SBP and for the DBP. The 99 measurements were 139 7 27 / 84 7 16 mmhg with the Omron HEM-907 device and 140 7 27 / 89 7 15 mmhg with the standard mercury sphygmomanometer. In total, 45 (3 15) sets of measurements were available for analysis in the first phase of the validation process, and 99 (3 33) in the second phase. The number of measurements differing from the mercury standard by 5, 10, and 15 mmhg or less, are shown in Table 3. These results are in concordance with the requested criteria of the international protocol for the primary and secondary phases. Thus the Omron HEM-907 device fulfils the validation criteria of the international protocol. The difference between the device readings and observer readings and the mean BP of the device and the two observers for all 99 points for SBP and DBP are displayed in Figure 1. The standard deviation tends to increase with the level of BP. This tendency seems more marked for SBP than for DBP. Discussion The Omron HEM-907 device fulfils the validation criteria of the international protocol for SBP and for DBP. The International recommendations [1,7] have been published as a proposed draft aiming to simplify the two main available guidelines, the BHS protocol, and the AAMI protocol. These two validation protocols have many similarities but there are some important differences [1,10]. It has been demonstrated in practice that validation studies can be performed in such a way as to satisfy the criteria of both protocols [1,7]. The main advantages of the international protocol are that it requires a lower number of subjects, simplifies the procedure, without affecting the accuracy of the validation [1,7]. To our knowledge, this is the first study to be published following the published proposal of the international protocol. The Omron HEM-907 has been previously validated according to the AAMI validation protocol [11]. The present study confirms that the international protocol on 33 subjects does not substantially alter the results but greatly simplifies the process of validation. This analysis shows that with the Omron HEM-907, the device observer limits of agreement widened at higher SBP values. The same conclusion was reached with the AAMI validation protocol but after analysis on 100 subjects [11]. Table 3 Number of measurements differing by 5, 10, and 15mmHg from the mercury standard, and mean7 SD of the differences between the device and the mercury standard (Phase I: 15 subjects, 45 measurements and Phase II: 33 subjects, 99 measurements Mean 7 SD (mmhg) Differences between Omron device and mercury sphygmomanometer Mean 7 SD of differences (mmhg) r 5 mmhg r 10 mmhg r 15 mmhg Phase I BP1 BP3 BP5 SBP 143 7 26 24 35 38 3 7 9 DBP 89 7 15 26 39 43 4 7 6 BP3 BP5 BP7 SBP 140 7 25 27 39 44 1 7 7 DBP 89 7 14 26 40 43 4 7 7 Final Grading SBP 140 7 25 27 39 44 1 7 7 DBP 89 7 14 26 40 43 4 7 7 Phase II BP1 BP3 BP5 SBP 142 7 28 53 76 89 3 7 8 DBP 89 7 16 52 83 96 5 7 5 BP3 BP5 BP7 SBP 140 7 27 61 85 94 1 7 7 DBP 89 7 15 52 85 96 5 7 6 Final Grading SBP 140 7 27 61 85 94 1 7 7 DBP 89 7 15 52 85 96 5 7 6 SBP, systolic blood pressure; DBP, diastolic blood pressure; SD, standard deviation.

240 Blood Pressure Monitoring 2002, Vol 7 No 4 Fig. 1 30 20 10 Device pressure - mean observer pressure (mmhg) 0-10 -20-30 90 100 110 120 130 140 150 160 170 180 190 200 30 20 10 0-10 would not significantly differ at, for example, a SBP of 190 mmhg versus one of 200 mmhg [11]. One limit of the present study is that these results are based on only one device and the validation was done in only one centre, however the international protocol does not specify the number of devices to be tested or the number of study sites recommended to enhance the heterogeneity of the study population. The AAMI protocol [6] recommends more than one study site without specifying the number of study sites and without noting the number of devices to validate, whereas the BHS protocol [5] does not make any recommendations regarding the number of study sites but does recommend that the capability of a number of devices of the model being tested are assessed in order to give consistent measurements, and if any substantial differences between instruments of the same device occur, then further device validation is not appropriate. Conclusion In conclusion, the Omron HEM-907 BP measuring device has passed the validation criteria of the published proposal of the international protocol for the validation of BP measuring devices and can be used for BP measurement. -20-30 30 40 50 60 70 80 90 100 110 120 130 140 Mean of device pressure and two observers pressure (mmhg) Plot of pressure difference (mmhg) between the mean of the two observers and the Omron HEM-907 device and the mean pressure (mmhg) for the mean of the two observers and the Omron HEM-907 device in 33 subjects (n ¼ 99). The upper panel corresponds to the systolic blood pressure readings whereas the lower panel shows the values for diastolic blood pressure. Reference lines, 0, 7 5, 7 10 and 7 15 mmhg difference. The increased error at extremes of BP occurs in virtually all non-invasive devices, but the degree of error varies [11,12]. It is, however, also important to recognize that this usually bears little clinical relevance since therapeutic decisions References 1 O Brien E, De Gaudemaris R, Bobrie G, Agabiti Rossei E, Vaisse B. Proceeding from a conference on self BP measurement. Devices and validation. Blood Press Monit 2000; 5:93 100. 2 Association for the Advancement of Medical Instrumentation. The national standard for electronic or automated sphygmomanometers. Arlington, VA USA: AAMI; 1987. 3 O Brien E, Petrie J, Littler WA, Padfield PL, O Malley K, Jamieson M, et al. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8:607 619. 4 O Brien E, O Malley K. Twenty-four-hour ambulatory blood pressure monitoring: a review of validation data. J Hypertens 1990; 8(Suppl 6): S11 S16. 5 O Brien E, Petrie J, Littler W, de Swiet M, Padfield P, Altman D, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(Suppl 2):S43 S62. 6 White W, Berson A, Robbins C, Jamieson M, Prisant L, Roccella E, et al. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993; 21: 504 509. Table 4 British Hypertension Society (BHS) grading and Association for the Advancement of Medical Instrumentation (AAMI) criteria for the Omron HEM-907 device (85 subjects, 255 measurements) BHS AAMI r 5 mmhg r 10 mmhg r 15 mmhg Grade Mean 7 SD (mmhg) Grade Final grading SBP 62% 87% 95% A 0 7 7 Pass DBP 58% 90% 98% B 4 7 5 Pass Observer comparison SBP 100% 100% 100% 1 7 2 DBP 100% 100% 100% 1 7 2 SBP, systolic blood pressure; DBP, diastolic blood pressure; SD, standard deviation.

Validation of the Omron HEM-907 device El Assaad et al. 241 7 O Brien E. Proposals for simplifying the validation protocols of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation. Blood Press Monit 2000; 5: 43 45. 8 Groupe d Evaluation et de Mesure de la Société Française d Hypertension Artérielle. La prise de la tension artérielle au cabinet médical. CD-Rom. Meudon (Haute de Seine): France: Editions RanD, 1998. 9 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; i: 307 310. 10 O Brien E, Atkins N. A comparison of the British Hypertension Society and Association for the Advancement of Medical Instrumentation protocols for validating blood pressure measuring devices: can the two be reconciled? J Hypertens 1994; 12: 1089 1094. 11 White W, Anwar Y. Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults. Blood Press Monit 2001; 6: 107 110. 12 Geddes LA, Voellz M, Combs C, Reiner D, Babbs CF. Characterization of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng 1982; 10: 271 280.