axion Protocol Cell Culture on Microelectrode Arrays Cell Type: Cryopreserved Human ipsc-derived Neurons BioSystems v. 1.1

Similar documents
axion Protocol Cell Culture on Microelectrode Arrays Cell Type: GE Healthcare - Cytiva TM Plus Cardiomyocytes BioSystems v. 1.0

axion Protocol Cell Culture on Microelectrode Arrays Cell Type: QBM Cell Science - E14,15 Embryonic C57 Mouse Cortical Neurons BioSystems v. 1.

axion Protocol Cell Culture on Microelectrode Arrays Cell Type: QBM Cell Science - Neonatal (P1) Rat DRG Neurons BioSystems v. 1.1

Measuring Cardiac Activity:

Measuring Cardiac Activity:

Pluricyte Cardiomyocytes. using the Multiwell MEA System from Multi Channel Systems

Measuring Cardiac Activity:

Assessment of pro-arrhythmic effects using Pluricyte Cardiomyocytes. on the ACEA xcelligence RTCA CardioECR

Pluricyte Cardiomyocytes. Axion BioSystems Maestro TM MEA system

Pluricyte Cardiomyocytes

Modulating Glucose Uptake in Skeletal Myotubes: Insulin Induction with Bioluminescent Glucose Uptake Analysis

icell Cardiomyocytes 2 User s Guide

Assessment of pro-arrhythmic effects using Pluricyte Cardiomyocytes on the Axion Maestro system

Performing Bioenergetic Analysis: Seahorse XFe96 Analyzer

Assessment of pro-arrhythmic effects using Pluricyte Cardiomyocytes on the Axion BioSystems Maestro TM MEA system

Mouse Hepatic Progenitor Organoid Culture: Supplementary Protocols

icell Cardiomyocytes User s Guide Document ID: X1000

Human ipsc-derived Ventricular Cardiomyocytes. Protocol version 3.1

Human Induced Plutipotent Stem Cell (ipsc) Handling Protocols: Matrigel and mtesr/e8 Media

Striatal Neuron Medium Kit

Islet Assay Using the Agilent Seahorse XF24 Islet Capture Microplate

Assessment of pro-arrhythmic effects using Pluricyte Cardiomyocytes on the ACEA xcelligence RTCA CardioECR

NINDS Repository Human Induced Pluripotent Stem Cell (ipsc) Handling Protocols (Matrigel and mtesr Media)

MED64 Protocol Axol Human ipsc-derived Neural Stem Cells

Product Use HPSC-CC are for research use only. It is not approved for human or animal use, or for application in in vitro diagnostic procedures.

axion Application Note Modeling Pain with Rat Dorsal Root Ganglion Neurons on MEAs BioSystems

MED64 Protocol. Axol Human ipsc-derived Sensory Neuron Progenitor

Protocol for Thawing and Use of Plateable and Suspension Cryopreserved Hepatocytes

ab MitoTox Complex I OXPHOS Activity Microplate Assay

AXION PLATES INDUSTRY-LEADING 768 ELECTRODES AT THE THROUGHPUT YOU NEED AXION MEA PLATES

Cellartis Enhanced hips-hep v2 Kits User Manual

RODENT Hepatocytes Care Manual

Protocol for Thawing Cryopreserved Hepatocytes

TECHNICAL BULLETIN. Primary Human Hepatocytes LifeNet Health

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

Cryopreserved HepaRG Cells and Media Supplements

ab Cell Invasion Assay (Basement Membrane), 24-well, 8 µm

Follicle Dermal Papilla Cells

Islet Oxygen Consumption Assay using the XF24 Islet Capture Microplate Shirihai Lab and Seahorse Bioscience Aug

ab HIV-1 Protease Inhibitor Screening Kit (Fluorometric)

ab Lysosome/Cytotoxicity Dual Staining Kit

NF-κB p65 (Phospho-Thr254)

ab Lipid Peroxidation (MDA) Assay kit (Colorimetric/ Fluorometric)

Human Pluripotent Stem Cell Cardiomyocyte Differentiation Kit (PSCCDK) Introduction Kit Components Cat. # # of vials Reagent Quantity Storage

Human ipsc-derived Microglial Precursors

RayBio Human Granzyme B ELISA Kit

Human Apolipoprotein A1 EIA Kit

Lipid Droplets Fluorescence Assay Kit

Human TSH ELISA Kit. User Manual

ab Human Citrate Synthase (CS) Activity Assay Kit

Morinaga Mouse C-peptide ELISA Kit

91154 PRIME-XV T Cell CDM Liquid 1 L Additional package sizes are available at request

Human HIV (1+2) antigen&antibody ELISA Kit

Receiving Cryopreserved Human Hepatocytes

ROS Activity Assay Kit

Midi Plant Genomic DNA Purification Kit

Enzyme Immunoassay for

Mouse C-Peptide ELISA Kit

Multi-Parameter Apoptosis Assay Kit

Cellartis Hepatocyte Diff Kit User Manual

CHO α 1 β 2 γ 2 GABAA Cell Line

TECHNICAL BULLETIN. Catalog Number RAB0447 Storage Temperature 20 C

QS S Assist KINASE_ADP-Glo TM Kit

CytoPainter Lysosomal Staining Kit - Blue Fluorescence

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

Intracellular (Total) ROS Activity Assay Kit (Red)

Fatty Acid Oxidation Assay on the XF24 Analyzer

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma

Product # R8132 (Explorer Kit) R8133 (Bulk Kit)

Procine sphingomyelin ELISA Kit

nachr α 4 β 2 CHO Cell Line

HbA1c (Human) ELISA Kit

Global Histone H3 Acetylation Assay Kit

For research or further manufacturing use only. Not for injection or diagnostic procedures.

ASSAY OF using AZO-FRUCTAN S-AZFR5 11/17

TECHNICAL BULLETIN. Phospho-Akt (pser 473 ) ELISA Kit for detection of human, mouse, or rat phospho-akt (pser 473 ) in cell and tissue lysates

INSTRUCTIONS Pierce Primary Cardiomyocyte Isolation Kit

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / CD81)

RayBio Human ENA-78 ELISA Kit

Lumino Firefly Luciferase Assay

ab65336 Triglyceride Quantification Assay Kit (Colorimetric/ Fluorometric)

IncuCyte NeuroPrime Cell Kit Catalog Number: 4585

Follicle Dermal Papilla Cell

ab Ascorbic Acid Assay Kit (Fluorometric)

Product Size Catalog Number. 500,000 proliferating cells. cryopreserved cells that have been thawed and cultured for three days at PromoCell.

Human Umbilical Cord Blood CD34 + /133+ Progenitor Cell Care Manual

EPIGENTEK. EpiQuik Global Histone H3 Acetylation Assay Kit. Base Catalog # P-4008 PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE

To place an order, please visit lifelinecelltech.com or call customer service at

LDL Uptake Flow Cytometry Assay Kit

Total Histone H3 Acetylation Detection Fast Kit (Colorimetric)

ab Lipase Activity Assay Kit (Colorimetric)

For Research Use Only Ver

ab HRV 3C Protease Inhibitor Screening Kit (Colorimetric)

FOCUS SubCell. For the Enrichment of Subcellular Fractions. (Cat. # ) think proteins! think G-Biosciences

WiCell Neural Stem Cell Protocols

Transcription:

axion BioSystems Cell Culture on Microelectrode Arrays Cell Type: Cryopreserved Human ipsc-derived Neurons Protocol v. 1.1

Trademarks Axion BioSystems, Inc. and the logo are trademarks of Axion BioSystems, and may not be used without the express written permission of Axion BioSystems, Inc. All other brands, product names, company names, trademarks and service marks are the properties of their respective owners. Restrictions and Liabilities This document is provided as is and Axion BioSystems will assume no responsibility for any typographical, technical or other inaccuracies in this document. Axion BioSystems does not make any commitment to provide any changes, updates, enhancements, or other additions to this document to you within any time frame or at all. This document might contain references to third-party sources of information, hardware or software, products or services and/or third-party websites (collectively the Third-Party Information ). Axion BioSystems has no control over and is not responsible for any Third-Party Information, including, without limitation the content, accuracy, copyright compliance, compatibility, performance, trustworthiness, legality, decency, links or any other aspect of Third-Party Information. The inclusion of Third-Party Information in this document does not imply endorsement with your use of Axion BioSystems technology. Conditions of Use You are responsible for understanding and performing the protocols that are described within. Axion BioSystems makes no guarantee for any results you may achieve. These protocols are provided as a recommendation by Axion BioSystems based on its use and experience. Origin Axion BioSystems Microelectrode Arrays are manufactured in the United States of America. Copyright Notice 2014 Axion BioSystems, Inc. All rights reserved. This document may not be reproduced, distributed, modified or publicly displayed without the express written permission of Axion BioSystems. Acknowledgement axion BioSystems Axion BioSystems would like to thank Cellular Dynamics International, Inc. for providing their experience and resources toward the creation of this application note. Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays i

Table of Contents 1. Before you Begin p.1 2. Introduction p.2 3. Technical Support p.3 4. Required Materials p.4 4.1 Consumables p.4 4.2 Equipment p.4 5. Workflow p.5 6. Methods p.6 6.1 Preparing the MEA Plates p.6 6.2 Thawing Cryopreserved icell Neurons p.6 6.3 Plating icell Neurons onto the MEA p.7 7. Data Acquisition & Analysis p.9 8. Summary p.9 Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays ii

Before You Begin 1. Read this entire manual before using cells or the microelectrode arrays. 2. Check the Axion Maestro system for correct performance. Contact Axion at support@axion-biosystems.com with any issues. 3. Consult with Axion about untested experimental variables if there is concern with the safety of the equipment. 4. The User Guide created by Cellular Dynamics International, Inc. is referenced throughout this manual, please refer to this document for further information: CDI icell Neurons User Guide Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 1

axion BioSystems Introduction icell Neurons are human induced pluripotent stem cell (ipsc)-derived neurons that exhibit biochemical, electrophysiological, and pathophysiological properties characteristic of native human neurons. Due to their high purity, functional relevance, and ease of use, icell Neurons represent an optimal in vitro test system for neurobiology interrogations in basic research and in many areas of drug development. The Maestro multielectrode array (MEA) system from Axion BioSystems is a non-invasive, label-free platform that measures local field potentials of electrically active cells reflecting underlying activity of ion channels and synapses. With proper handling, icell Neurons can be thawed and cultured directly on MEAs to form neuronal networks amenable to electrophysiological interrogation. Together, icell Neurons and Axion Biosystems MEA technology provide an excellent, non-invasive platform for assessing effects of compounds on neuronal excitability and neurotransmission. This Application Protocol describes how to handle icell Neurons for use on the Maestro MEA system and provides basic instructions for compound treatments, data acquisition, and analysis. Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 2

Technical Support For any questions about cell plating or Maestro system operation, please contact Axion BioSystems Support using the information below. Telephone: (404) 477-2557 Fax: (404) 385-4638 E-mail: support@axion-biosystems.com Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 3

Required Materials Consumables Item Vendor Catalog Number icell Neurons Kit Cellular Dynamics International NRC-100-010-001 HBSS Life Technologies 14170-112 50% Polyethylenimine Solution (PEI) Sigma-Aldrich P3143 Boric Acid Fisher Scientific A73-500 Sodium Tetraborate Sigma-Aldrich 221732 Hydrochloric Acid Sigma-Aldrich H1758 Laminin Sigma-Aldrich L2020 Kimwipes Pipettes and Pipettors 15 ml and 50 ml Centrifuge Tubes Pipette Aid and Sterile Pipettes Sterile 70% Ethanol Equipment Item Vendor Catalog Number Maestro MEA System Axion BioSystems 12-Well MEA Axion BioSystems M768-GLx 48-Well MEA Axion BioSystems M768-KAP 96-Well MEA Axion BioSystems M768-KAP-96 Axion Integrated Studio (AxIS) Axion BioSystems 37 o C Water Bath Cell Culture Incubator Hemocytometer or Automated Cell Counter Biological Safety Cabinet Tabletop Centrifuge Phase Contrast Microscope Liquid Nitrogen Storage Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 4

Workflow icell Neurons are thawed and plated onto the MEA plates coated with polyethylenimine (PEI). On day 3 post-plating, electrical activity can first be checked using the Maestro. On day 4, cells can be used for experimentation such as compound screening. Thaw and Plate onto the MEA Begin Compound Screening Days in Culture -1 0 3 4 Coat the MEA Check MEA Activity Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 5

Methods Preparing the MEA Plates 1. Prepare 1 L of borate buffer by dissolving 3.10 g of boric acid and 4.75 g of sodium tetraborate in distilled water. Adjust the ph to 8.4. 2. Prepare a 0.1% PEI solution by diluting 50% PEI solution in borate buffer. Filter the 0.1% PEI solution through a 0.22 µm filter. Tip 0.05-0.1% PEI solution can be stored at 4 C for up to 1 month. 3. Add 200 µl of 0.1% PEI solution to each well of the MEA plate. Incubate at room temperature for 1 hour. 4. Aspirate the PEI solution from each well of the MEA plate. Do not allow the wells to dry. 5. Rinse each well 4 times with 300 µl of sterile water. 6. Air-dry the MEA plate with the lid off in a sterile biological safety cabinet overnight.! It is essential to allow the MEA plate to air-dry overnight to achieve optimal cell adhesion. Thawing Cryopreserved icell Neurons 7. Prepare Complete icell Neurons Maintenance Medium (Complete Maintenance Medium) according to the icell Neurons User s Guide*. 8. Dilute the laminin solution (1 mg/ml) in 15 ml Complete Maintenance Medium to a final concentration of 20 µg/ml. Gently mix by inverting the tube.! Thaw stock laminin solution at room temperature or overnight at 4 C. Do not thaw stock laminin solution in a 37 C water bath. Do not vortex laminin solution. 9. Thaw cryopreserved icell Neurons according to the icell Neurons User s Guide* to a final volume of 5 ml in Complete Maintenance Medium. 10. Remove a sample of the cell suspension and count the viable neurons using a hemocytometer. * Refer to: http://www.cellulardynamics.com/products/lit/cdi_icellneuronsusersguide121106.pdf Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 6

A. 11. Further dilute the cell suspension with 5 ml of 20 µg/ml laminin solution to achieve a final laminin concentration of 10 µg/ml. 12. Transfer the cell suspension to a 15 ml centrifuge tube. 13. Concentrate the neurons by centrifuging at 380 x g for 5 minutes. B. 14. Dilute 500 µl of 20 µg/ml laminin solution in additional 500 µl of Complete Maintenance Medium to achieve a 10 µg/ml laminin solution. 15. Aspirate the supernatant to just above the cell pellet, being careful not to disturb the pellet. 16. Resuspend the cell pellet in 10 µg/ml laminin solution to a final concentration of 28,000 viable neurons/µl. 17. Transfer the cell suspension to a 1.5 ml centrifuge tube. C. Plating icell Neurons onto the MEA 18. Dispense a 5 µl droplet of the cell suspension (140,000 neurons) directly over the recording electrode area of each well of the MEA precoated with PEI solution. See Figure 1 for appropriate drop placement and size. Figure 1: Drop Placement Diagram The layouts above represent the bottom surfaces of wells in a 12-well MEA (A), a 48-well MEA (B), and a 96-well MEA (C). Diagram A represents a 12-well MEA and the inner 64 dots of the electrode array with the 4 ground electrodes located in the corners. Diagram B represents a 48-well MEA and the inner 16 dots of the electrode array with the 4 ground electrodes located in the corners. Diagram C represents a 96-well MEA and the inner 8 dots of the electrode array with the 4 ground electrodes located in the corners. The red circles indicate the approximate size and location for the drop placement.! Do not allow the cell suspension droplets to incubate in the MEA plate for longer than 1 hour before beginning addition of Complete Maintenance Medium. 19. Add 6 ml of sterile water to the area surrounding the wells (MEA reservoirs) of the MEA plate to prevent substrate evaporation. Do not allow water into the wells of the MEA plate. Tip MEA reservoir water is no longer required following the media addition in Steps 21 and 22. 20. Incubate the MEA plate with the seeded neurons in a cell culture incubator at 37 C, 5% CO 2 for 1 hour. 21. Gently add 100 µl of Complete Maintenance Medium down the side of each well of the MEA one row at a time using a multi-channel pipettor. Adding the medium too quickly will dislodge the adhered neurons. 22. Repeat step 21 two times to reach a final volume of 300 µl/well. Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 7

! Timing is critical in this step. The performance is compromised if the droplets are allowed to dry. CDI recommends adding a small volume of medium to all wells first rather than adding the total volume in each well at once. 23. Incubate in a cell culture incubator at 37 C, 5% CO 2. 24. Exchange 50% of medium every 7 days by aspirating 150 µl of spent medium from each well and adding 150 µl of Complete Maintenance Medium. 25. Test electrical activity of icell Neurons using the Maestro MEA system on day 3 post-plating.! Peak activity will occur on day 4 to day 5 postplating. For optimal experimental reproducibility, wait until these days to test effects of compounds Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 8

Data Acquisition & Analysis Axion AxIS Software is used to record raw voltage data and detect spikes for rate analysis. Product manuals can be downloaded at www.axion-biosystems.com for more comprehensive instructions. To test the sensitivity of icell neurons to well-characterized neuroactive compounds, well-wide spontaneous spike activity was quantified pre and post-dose. Results are presented in Fig. 3. Substantial changes were observed for all compounds except negative controls. For example, the inhibitory neurotransmitter GABA reduces spontaneous activity, whereas disinhibition with the GABA antagonists PTX and gabazine increases activity. % Change in MFR 150 100 50 0 Compound Conc. DMSO 0.10% Picrotoxin 25µM Gabazine 10µM Lidocaine 50µM Domoic Acid 0.1µM GABA 3µM TEA 10mM -50-100 -150 Figure 3: Effect of Neuroactive Compounds on Spontaneous Spike Activity in icell Neurons Cultured for 4 Days in Axion 48-well MEA plates Well-wide Mean Firing Rate (MFR) was calculated pre- and post-dose as total spikes/time. Electrodes with < 5 spikes/min were considered inactive and were not included in the calculations. Values represent means ± SEM for n=3 wells. The vehicle, DMSO, reduced MFR by only 12%. Summary DMSO Picrotoxin Gabazine Lidocaine Domoic Acid GABA TEA icell Neurons can be thawed and cultured directly on Axion multiwell MEAs, where neuronal networks are established and spontaneous electrical signals can be monitored. Neuronal spike activity can be modulated with compounds and functional responses can be robustly quantified. The methods and data presented here highlight the ease of using icell Neurons on the Maestro MEA system from Axion BioSystems. Together, these products offer a high throughput in vitro platform for gathering relevant data on the electrophysiological activity of human neuronal cells. Axion BioSystems Cryopreserved Human ipsc-derived Neurons on Microelectrode Arrays 9