Externally Applied Forces to the Palm Increase Carpal Tunnel Pressure

Similar documents
Main Menu. Wrist and Hand Joints click here. The Power is in Your Hands

Wrist & Hand Assessment and General View

Hand and Wrist Editing file. Color Code Important Doctors Notes Notes/Extra explanation

10/10/2014. Structure and Function of the Hand. The Hand. Osteology of the Hand

Wrist and Hand Anatomy

ARM Brachium Musculature

divided by the bones ( redius and ulna ) and interosseous membrane into :

Nerves of Upper limb. Dr. Brijendra Singh Professor & Head Department of Anatomy AIIMS Rishikesh

LECTURE 8 HANDS: BONES AND MUSCLES

WHAT CAN ULTRASOUND SEE IN THE CARPAL TUNNEL REGION?

Guyon's Canal Revisited: An Anatomic Study of the Carpal Ulnar Neurovascular Space

Muscles of the hand Prof. Abdulameer Al-Nuaimi

Endoscopic Carpal Tunnel Release System. SafeViewTM Endoscopic Carpal Tunnel Release System

Introduction. The wrist contains eight small carpal bones, which as a group act as a flexible spacer between the forearm and hand.

The Muscular System. Chapter 10 Part C. PowerPoint Lecture Slides prepared by Karen Dunbar Kareiva Ivy Tech Community College

Clinical examination of the wrist, thumb and hand

Endoscopic Carpal Tunnel Release System. SafeViewTM 360 Panoramic Visualization Sterile Packaged Fully Disposable

13 13/3/2012. Adel Muhanna

Hand / wrist Injections. MATS. June Condition Symptoms Conservative Treatments Location of injection CBA for surgery

A Patient s Guide to Ulnar Nerve Entrapment at the Wrist (Guyon s Canal Syndrome)

The hand is full with sweat glands, activated at times of stress. In Slide #2 there was a mistake where the doctor mentioned lateral septum twice.

Common Upper Extremity Neuropathies (Not Carpal Tunnel Syndrome)

Forearm and Wrist Regions Neumann Chapter 7

Physical therapy of the wrist and hand

Hand Anatomy A Patient's Guide to Hand Anatomy

MCQWeek2. All arise from the common flexor origin. The posterior aspect of the medial epicondyle is the common flexor origin.

Ergonomics Glossary. Force The amount of physical effort a person uses to do a task.

Centerline Carpal Tunnel Release

The hand. it's the most important subject of the upper limb because it has a clinical importance. the palm of the hand**

Initial experience with endoscopic carpal tunnel release surgery

Carpal Tunnel Syndrome (CTS)

In the name of Allah, Most gracious, Most merciful

Biceps Brachii. Muscles of the Arm and Hand 4/4/2017 MR. S. KELLY

Interesting Case Series. The Recurrent Motor Branch of the Median Nerve

Wrist and Hand Anatomy/Biomechanics

Anatomy - Hand. Wrist and Hand Anatomy/Biomechanics. Osteology. Carpal Arch. Property of VOMPTI, LLC

Anatomy Workshop Upper Extremity David Ebaugh, PT, PhD Workshop Leader. Lab Leaders: STATION I BRACHIAL PLEXUS

SmartRelease Endoscopic Carpal Tunnel Release (ECTR)

Viorel Nacu. The clinical anatomy of the Hand

8/25/2014. Radiocarpal Joint. Midcarpal Joint. Osteology of the Wrist

Kinesiology of The Wrist and Hand. Cuneyt Mirzanli Istanbul Gelisim University

Manual therapy approach to the Patient with Carpal Tunnel Syndrome.

10/15/2014. Wrist. Clarification of Terms. Clarification of Terms cont

Current Developments in the Prevention and Treatment of Repetitive Motion Injuries of the Upper Extremity

CHAPTER 6: THE UPPER EXTREMITY: THE ELBOW, FOREARM, WRIST, AND HAND

Clinical experience of open carpal tunnel release and Camitz operation in elderly Chinese patients

TECHNOLOGY AND HOW WE USE IT TO DAMAGE OURSELVES WILLIAM A. DELP, DO ASSISTANT PROFESSOR OF OMM GA PCOM

[[Sally Leaning Towards Peter To Take Cold Hand]]

Case 1. Your diagnosis

Variation of Superficial Palmar Arch: A Case Report

Names: Block: Date: Building an Articulated Model of the Hand

Safe Treatment of Trigger Thumb With Longitudinal Anatomic Landmarks

Nerves of the upper limb Prof. Abdulameer Al-Nuaimi. E. mail:

Introduction to Ultrasound Examination of the Hand and upper

SUSPECTS THE UNUSUAL. Often-Overlooked Muscles. Sternohyoid, longus colli, and longus capitis. Flexor pollicis longus.

MR IMAGING OF THE WRIST

Ultrasound diagnostic of carpal tunnel syndrome

A Pressing Question - How Much Contact Pressure Is Too Much?

Key Points for Success:

Ultrasonography of Peripheral Nerve -upper extremity

Assessment of the Distal Extent of the A1 Pulley Release: A New Technique

Lecture 9: Forearm bones and muscles

eplasty: Vol. 8 Assessment of the Distal Extent of the A1 Pulley Release: A New Technique

Metacarpophalangeal Joint Implant Arthroplasty REHABILITATION PROTOCOL

MUSCULOSKELETAL DISORDERS: THE BIGGEST JOB SAFETY PROBLEM. What Are Musculoskeletal Disorders

Trapezium is by the thumb, Trapezoid is inside

medial half of clavicle; Sternum; upper six costal cartilages External surfaces of ribs 3-5

STRUCTURAL BASIS OF MEDICAL PRACTICE EXAMINATION 5. September 30, 2011

CADAVERIC STUDY OF VARIANT FLEXOR CARPI ULNARIS

Small muscles of the hand

Official Definition. Carpal tunnel syndrome, the most common focal peripheral neuropathy, results from compression of the median nerve at the wrist.

Evaluation of the carpal tunnel based on 3-D reconstruction from MRI

The Forearm 2. Extensor & lateral Compartments of the Forearm

Dr. Mahir Alhadidi Anatomy Lecture #9 Feb,28 th 2012

Muscular Nomenclature and Kinesiology - One

Integra. SafeGuard Mini Carpal Tunnel Release System SURGICAL TECHNIQUE

Interesting Case Series. Radial Tunnel Syndrome Complicated by Lateral Epicondylitis in a Middle-Aged Female

COMPOSITE WRIST FLEXOR STRETCH

The Elbow and the cubital fossa. Prof Oluwadiya Kehinde

The use of Ultrasound for the Diagnosis and Treatment of the Musculoskeletal System

RESEARCH PAPERS FOR NATIONAL ACADEMY OF OESTEOPATHY ON CARPAL TUNNEL SYNDROME BY JASBIR KAUR JAURA

Common Hand Problems and Case Examples. September 22, 2017 Carolyn Berg, MD Advanced Orthopedics of Oklahoma

Carpal Tunnel Syndrome/ Cubital Tunnel Syndrome. Nerve anatomy. Pathophysiology 6/14/2014. Most common compression neuropathies of the upper extremity

STRUCTURAL BASIS OF MEDICAL PRACTICE EXAMINATION 5 October 6, 2006

GENERAL EXERCISES THUMB, WRIST, HAND BMW MANUFACTURING CO. PZ-AM-G-US I July 2017

A case report submitted for the degree of. Doctor of Physical Therapy. Carroll University. Waukesha, WI. Kyle Vande Hei, SPT.

# Anatomy. Upper Extremities Muscles and anatomy of axilla. Tiba Al-Ani 9/10/2015 Nabil. Page 0 of 16

Modular Ulnar Head surgical technique. Transforming Extremities

Transverse Movement of the Median Nerve in the Carpal Tunnel during Wrist and Finger Motion in Patients with Carpal Tunnel Syndrome

Scapular Protraction in Sitting

Dr David Ruttenberg MBChB.,MSc(Med).,FAFOEM

Structure and Function of the Hand

Anatomic relationships of an endoscopic carpal tunnel device to surrounding structures

Carpal Tunnel Release

Compartment Syndrome

An Investigation of Hand Force Distribution, Hand Posture and Surface Orientation

Netter's Anatomy Flash Cards Section 6 List 4 th Edition

Wrist and Hand Complaints

Biomechanics of Two Reconstruction Techniques for Elbow Ulnar Collateral Ligament Insufficiency

High-resolution Ultrasound of the Thenar Motor Branch of the Median Nerve

Transcription:

Externally Applied Forces to the Palm Increase Carpal Tunnel Pressure Tyson K. Cobb, MD, Kai-Nan An, PhD, William P. Cooney, MD, Rochester, MN Change in carpal tunnel pressures that result from externally applied forces to the palm of the hand were assessed in five cadaveric specimens. MIKRO-TIP transducers were percutaneously placed into the carpal tunnel at the level of the hook of the hamate. A 1 kg force was applied to the palm of the hand in 16 separate locations. Significant elevations in carpal tunnel pressure were observed for external forces applied over the flexor retinaculum (103 mm Hg) and also for the hypothenar (37 mm Hg) and thenar (75 mm Hg) areas adjacent to the distal aspect of the carpal tunnel. These data demonstrate that the application of external forces on the palm of cadaver hands increases carpal tunnel pressure and the magnitude of the pressure change in the carpal tunnel depends on the location of the applied force. (J Hand Surg 1995;20A: 181-185.) Carpal tunnel syndrome is the most common entrapment syndrome in the upper extremity. Although this disease has been studied extensively, its cause remains unknown. A strong correlation has been shown between carpal tunnel syndrome and work related activities. Work activities involving repetitive motion of the hand and wrist are thought to predispose to the development of carpal tunnel syndrome.1 Furthermore, the incidence of carpal tunnel syndrome among employees correlates with the duration of employment. 2 The authors have observed increased carpal tunnel pressures as a result of externally applied forces to the palm of the hand. Previous studies have documented increased carpal tunnel pressure resulting from externally applied pressure over the flexor retinaculum. Gelberman et al.,3 Lundborg et al., 4 and Szabo et al? have each demonstrated that externally applied pressure over the flexor retinaculum results From the Department of Orthopedics, Mayo Clinic and Mayo Foundation, Rochester, MN. Received for publication Jan. 19, 1994; accepted in revised form June 4, 1994. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. Reprint requests: William P. Cooney, MD, Mayo Clinic, 200 First Street SW, Rochester, MN 55905. in an increase in carpal tunnel pressure. However, the authors are not aware of any investigations on the effect of external forces applied to the palm on carpal tunnel pressure. A better understanding of this phenomenon is required for proper interpretation of carpal tunnel pressure studies that incorporate a component of the external palmar forces, such as a tightly closed fist, and may be of benefit in preventive measures to minimize carpal tunnel pressures in the workplace through ergonomics. The magnitude of carpal tunnel pressure change relative to the location and amount of an externally applied force on the palm of the hand has not been described and is the basis for this report. Material and Methods Five fresh, frozen and thawed cadaveric upper extremities were removed by above elbow amputation. The specimens were kept frozen until they were used. After thawing, the palm of each specimen was marked (Fig. i). Pressures were measured with the MIKRO-TIP transducer system (SPR-524, Millar Instruments, Houston, TX). This system is a solid-state transducer in the tip of a 2.5F catheter. The transducer catheter was placed in the palm by insertion of an angiocatheter at the level of the proximal wrist flexion crease and adjacent to the palmaris longus tendon. The catheter was directed The Journal of Hand Surgery 181

182 Cobb et al. / Palm Increase Carpal Tunnel Pressure 9 @ @ / Figure 1. Sixteen locations where an external force (1 kg) was applied to the palm of the hand. Carpal tunnel pressures were recorded at the level of the hook of the hamate (near position 14). (By permission of Mayo Foundation.) dorsally and distally at an angle of 45 ~ until the proximal portion of the flexor retinaculum was penetrated. Penetration was identified by a marked increase in resistance followed by a pop and subsequent decrease in the resistance of catheter advancement. The catheter was advanced distally in the anterior aspect of the carpal canal next to the median nerve. The position of the hook of the hamate was marked by using topographic lines, 6 and the length of the transducer catheter needed to reach the hook of the hamate was measured. The needle was removed from the angiocatheter and the transducer catheter was placed within the angiocatheter and advanced the specific length to the hook of the hamate. The angiocatheter was withdrawn, and the transducer catheter was secured in place by suturing it to the skin. Force was applied to each of the numbered areas on the palm of the hand (Fig. 1) by means of a force dial (Wagner Instruments, Greenwich, CT) which applied! kg force to the palm in a palmar to dorsal direction at each of the specified locations. The MIKRO-TIP transducer and the control unit (model TC-510) were assembled, calibrated, and set \ at zero according to the manufacturer's specifications before placement. The system was attached through an analog to digital board to a computerized data-acquisition system. A conventional transducer system generated signals that served as event markers. A saline-filled syringe was attached to the transducer system. The conventional transducer was also hooked up through the analog to digital board of the computerized data-acquisition system. The tests were performed in a sequential manner with the head (5 mm diameter) of the force dial placed on each of the numbered locations of the palm of the hand (Fig. t). The plunger of the syringe connected to the conventional system was withdrawn so that a negative reading was recorded by the data-acquisition center (event marker) before placement of positive pressure by the force dial. After depressing the syringe of the event marker, the head of the force dial was depressed onto the specific location of the palm in a palmar to dorsal direction until the force dial read 1 kg. This pressure was maintained for approximately 1.5 second and then released. The event marker was depressed and the force dial was applied to the next specified location. The sequence was continued until all specified areas on the palm of the hand had been tested. Repeated measures analysis of variance was used to determine differences in carpal tunnel pressures resulting from forces applied to various regions on the palm. The paired t-test tested for differences between the proximal palm and the distal palm. Results The average carpal tunnel pressure resulting from pressure on the palm was calculated from the five specimens (Fig. 2). Some degree of increase in carpal tunnel pressure was observed as a result of an externally applied force for all locations tested. After withdrawal of the external force on the palm, the carpal tunnel pressure returned to baseline in each instance. Externally applied pressures applied to the distal aspect of the palm (positions 1, 4, 7, and 10) had little effect on pressures in the carpal tunnel. A 1 kg force applied to the palm of the hand in the region of the metacarpal heads resulted in a total mean pressure of 6 mm Hg in the carpal tunnel. Higher levels were noted in specimen 3. This specimen had a total mean pressure of 18 mm Hg carpal tunnel pressure for the metacarpal region. Table 1 contains the mean, median, standard deviation, and standard error of the mean for the carpal tunnel pressures resulting from a force applied to each of the 16 positions. Similar pressures were noted for the next more

The Journal of Hand Surgery / Vol. 20A No. 2 March 1995 183 623 @ o / f Figure 2. The mean carpal tunnel pressure in mm Hg at each location from an externally applied force to the palm for the five specimens. (By permission of Mayo Foundation.) Table 1. Mean, Median, Standard Deviation, and Standard Error of the Mean for the Carpal Tunnel Pressures Resulting From a Force Applied to Each of the 16 Positions* Position Mean Median SD SE 1 7.00 5 5.24 2.35 2 4.20 3 3.03 1.36 3 25.40 15 20.50 9.17 4 8.60 5 8.73 3.91 5 8.00 9 4.85 2.17 6 74.60 66 25.60 11.50 7 4.80 3 5.40 2.42 8 4.60 4 3.58 1.60 9 37.00 39 18.29 8.18 10 2.80 2 2.78 1.24 11 3.20 3 2.05 0.92 12 18.60 18 16.59 7.42 13 98.00 105 63.10 28.20 14 136.40 160 70.20 31.40 15 77.00 29 86.70 38.80 16 9.40 1 18.26 8.16 * N = 5 for all positions. SD, standard deviation; SE, standard error of the mean. \ proximal row (positions 2, 5, 8, and 11), that included the region of the palm over the metacarpal necks and distal to the thumb web space. The total mean carpal tunnel pressure for a 1 kg force applied to this region was 5 mm Hg. The range for this region was 0 to 14 mm Hg in the five specimens. Intermediate pressures were noted for the proximal hypothenar area (positions 9 and 12). The total mean carpal tunnel pressure for a 1 kg force applied to this region was 28 mm Hg (range, 4 to 54 mm Hg). Pressure applied at position 9, just distal to the hook of the hamate, resulted in the highest mean carpal tunnel pressure for this region (mean, 37 mm Hg; range, 7 to 54 mm Hg). The region of the thenar eminence (positions 3 and 6) resulted in higher carpal tunnel pressures than those of the hypothenar eminence, with a total mean carpal tunnel pressure of 50 mm Hg. The medial aspect of the base of the thenar eminence (position 6) yielded the highest pressure in the region, with a mean of 75 mm Hg (47 to 113 mm Hg). The region over the flexor retinaculum in the central aspect of the proximal region of the hand had the greatest effect on the carpal tunnel pressures (positions 13, 14, and 15). The resulting total mean carpal tunnel pressure for a 1 kg force applied to this region was 103 mm Hg (mean, 1 to 226 mm Hg). The highest pressures were generated by pressure applied in the midline of the palm adjacent to the hook of the hamate at position 14 (mean, 136 mm Hg; range, 60 to 226 mm Hg). Repeated pressure analysis of variance demonstrated that the carpal tunnel pressures were higher for forces applied to positions 13 and 14 than for other regions (p <.05). Position 13 had higher carpal tunnel pressures compared to other positions, except for positions 15 and 16, which were not significantly different from position 13 (t9 <.05). The 8O 70 Proximal half Area i Distal half Figure 3. Distribution of the mean differences in carpal tunnel pressure resulting from a palmar force for each of the five specimens. Note that the proximal half of the palm has consistently higher differences in pressure.

184 Cobb et al. / Palm Increase Carpal Tunnel Pressure paired t-test demonstrated significantly higher carpal tunnel pressures for the proximal aspect of the palm (including positions 3, 6, 9, 12, 13, 14, 15, and 16) compared to the distal aspect of the palm (including positions 1, 2, 4, 5, 7, 8, and 10) (p =.002). Since position 11 is on the border of the proximal and distal aspect of the palm, the test was performed with position 11 on both sides without a change in significance (Fig. 3). Discussion This study showed that an externally applied force to the proximal aspect of the hand results in significantly increased carpal tunnel pressures. Gelberman et al. 7 found the average intracarpal canal tissue pressure for patients with carpal tunnel syndrome was 32 mm Hg. With this value as a reference, significant mean pressures were observed at position 9 (mean, 28 mm Hg), just distal to the hook of the hamate; position 6 (mean, 50 mm Hg), the medial aspect of the base of the thenar eminence; position 13 (mean, 98 mm Hg), distal aspect of the aponeurotic portion 8 of the flexor retinaculum; position 14 (mean, 136 mm Hg), adjacent to the hook of the hamate; and position 15 (mean, 75 mm Hg), roughly at the junction of the proximal and central portions of the flexor retinaculum. The findings are consistent with those of earlier studies 4'5 that examined carpal tunnel pressures. These authors found that increased pressures (ranging from 30 to 90 mm Hg) caused dysfunction of the median nerve, including numbness and tingling within 15 minutes and complete sensory block 7 after 30 minutes of sustained pressure. The response of the median nerve to sustained pressure is a function of both time and pressure. Sensory response decreased rapidly and disappeared totally at 60 mm Hg, whereas nerves exposed to 30 mm Hg responded more slowly and with a less complete response. 7 The exact position of pressure measurement relative to the hamate hook was not described in these studies. The authors are not aware of any studies that have evaluated the effect of intermittent pressure on the median nerve. Although pressures measured in the carpal tunnel in our study are within the range of those found to induce symptoms of carpal tunnel syndrome, such a mechanism of pressure induction would be of short duration and the clinical significance has not been established. The additive effect of numerous short-lived elevations in carpal tunnel pressure, as expected in many work environments, may contribute to the development of carpal tunnel syndrome and should be evaluated in vivo. Significant increases in pressure at positions 6 and 9 are not explained by direct effects on the carpal tunnel because these positions are located radial and ulnar to the carpal tunnel, respectively. It is likely that pressure on the thenar and hypothenar muscles, that contribute to the distal portion of the flexor retinaculum, 8 results in stretch on this structure and increases in pressure within the canal. A second likely cause for increased pressures in the carpal tunnel as a result of externally applied pressure over positions 6 and 9 is the distribution of pressure through compartmental anatomy, including the flexor tendon sheaths. Forces applied at these positions may exert their effect through the bursal connections, as described by Parona. 9 A third likely cause is simple soft tissue compression effects, which are transmitted to adjacent tissues. This phenomenon was described by Guyton ~~ and Guyton et al. u in their early studies of tissue pressure. They used a blood pressure cuff to exert a mechanical force, resulting in increases in both solid and fluid pressures. This phenomenon has also been observed by the present authors in the laboratory. Externally applied pressure over a catheter in a muscular compartment results in pressure increase. The amount of pressure increase recorded by the catheter is a function of the distance of the catheter from the externally applied pressure and the tissue through which the pressure is transmitted. The concentric distribution of pressures observed in this study is consistent with this type of phenomenon. The highest pressures recorded were those directly over the catheter tip (position 14 in Fig. 1), and they decreased as a function of the distance away from the catheter. The findings in this study have helped the present authors explain some of the findings of other pressure studies in which unexplainably high pressures were measured within the canal when a component of external pressure at the base of the hand was included as a portion of the variable assessed. An example of this is pressures measured within the carpal canal as the result of flexion of the fingers. As the fingers are brought into tight flexion, the force of the fingertips on the base of the hand has a profound effect on pressures measured in the canal with active grip. These findings are also of clinical importance for the purpose of tool design. Alterations in median nerve conduction in humans, which imitate the first stages of carpal tunnel syndrome, have been shown to occur at a pressure of 30 mm Hg. 4 Nerve dysfunction and decrease of intraneura112,13 blood flow as a result of increased pressure have been shown to occur in animal models at 30 mm Hg. 14-16 Therefore, median nerve dysfunction occurs with pressures at or less than those recorded in our study for a 1 kg force applied in positions 6, 9, 13, 14, and 15. Fur-

The Journal of Hand Surgery / Vol. 20A No. 2 March 1995 185 thermore, the amount of force exerted in the palm of the hand with stressful labor is considerably greater than the 1 kg force used in this study. Therefore, tool grip design should include characteristics that would allow externally applied forces to the palm to be minimized in these regions. Specifically designed gloves may also allow distribution of forces away from these sensitive areas. The duration of pressure increase has been shown to be an important factor in determining the pathologic effect ~5 and the recovery of nerve function. 17 The chronic effects of intermittent pressure increases as might be seen during prolonged repetitive activities have not been investigated but could theoretically contribute to carpal tunnel syndrome. Therefore, preventive measures such as specially designed gloves may prove beneficial in preventing work-related carpal tunnel syndrome. The authors thank Dean R. Fisher, Department of Anatomy, for help with cadaveric specimens and Duane M. Ilstrup, MS and Steven L. Wallrichs, BS, Section of Biostatistics, for help with statistics. References 1. Occupational disease surveillance: carpal tunnel syndrome. JAMA 1989;262:886-9. 2. Muffiy-Elsey D, Flinn-Wagner S. Proposed screening tool for the detection of cumulative trauma disorders of the upper extremity. J Hand Surg 1987;12A:931-5. 3. Gelberman RH, Szabo RM, Williamson RV, Hargens AR, Yaru NC, Minteer-Convery MA. Tissue pressure threshold for peripheral nerve viability. Clin Orthop 1983;178:285-91. 4. Lundborg G, Gelberman RH, Minteer-Convery M, Lee YF, Hargens AR. Median nerve compression in the carpal tunnel-functional response to experimentally induced controlled pressure. J Hand Surg 1982; 7A:252-9. 5. Szabo RM, Gelberman RH, Williamson RV, Hargens AR. Effects of increased systemic blood pressure on the tissue fluid pressure threshold of peripheral nerve. J Orthop Res 1983;1:172-8. 6. Cobb TK, Cooney WP, An K-N. Relationship of deep structures of the hand and wrist to topographical landmarks. Clin Anat 1993;6:300-7. 7. GelbermanRH, HergenroederPT, Hargens AR, Lundborg GN, Akeson WH. The carpal tunnel syndrome: a study of carpal canal pressures. J Bone Joint Surg 198l;63A:380-3. 8. Cobb TK, Dalley BK, Posteraro RH, Lewis RC. Anatomy of the flexor retinaculum. J Hand Surg 1993; 18A:91-9. 9. Parona F. Dell'oncotomia negli accessi profundi diffusi dell'avambrachio. Annali Universali di Medicine e Chirurgia, Milano, 1876. 10. Guyton AC. A concept of negative interstitial pressure based on pressures in implanted perforated capsules. Circ Res 1963;12:399-414. 11. Guyton AC, Granger HJ, Taylor AE. Interstitial fluid pressure. Physiol Rev 1971;51:527-63. 12. Ogata K, Naito M. Blood flow of peripheral nerve effects of dissection, stretching and compression. J Hand Surg 1986;11B: 10-4. 13. Rydevik B, Lundborg G, Bagge U. Effects of graded compression on intraneural blood flow: an in vivo study on rabbit tibial nerve. J Hand Surg 1981;6A: 3-12. 14. Dellon AL, Schneider RJ, Burke R. Effect of acute compartmental pressure change on response to vibratory stimuli in primates. Plast Reconstr Surg 1983;72: 208-16. 15. Hargens AR, Romine JS, Sipe JC, Evans KL, Mubarak S J, Akeson WH. Peripheral nerve-conduction block by high muscle-compartment pressure. J Bone Joint Surg 1979;61A:192-200. 16. Lundborg G, Meyers R, Powell H. Nerve compression injury and increased endoneurial fluid pressure: a "miniature compartment syndrome." J Neurol Neurosurg Psychiatry 1983 ;46:1119-24. 17. RorabeckCH, ClarkKM. Thepathophysiologyofthe anterior tibial compartment syndrome: an experimental investigation. J Trauma 1978;18:299-303.