Valencia Pedicle Screw Surgical Technique

Similar documents
Silverstone Interbody System Surgical Technique

EXCELLA ll. Spinal System

X-spine Surgical Technique

Imola Lateral IBF System Surgical Technique

Threshold Pedicular Fixation System Surgical Technique

HydraLok. Operative Technique. Polyaxial Pedicle Screw System

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine.

Surgical Technique & Product Catalogue. Guide for Open & MIS Procedures

The Most Reliable Spinal Solution

Y o u r Id e a s En g i n e e r e d t o Li f e

MATRIX Spine System Deformity

Technique Guide. MATRIX Spine System MIS Instrumentation. The total solution for simple and complex spine pathology.

EXCELLA MIS. Spinal System

operative technique Universal Application


VTI INTERLINK PEDICLE SCREW SYSTEM

Thunderbolt. surgical technique. MIS Pedicle Screw System. Where Nimble and Secure Intersect

MATRIX Spine System Deformity. A posterior pedicle screw, hook, and rod fixation system.

Technique Guide. Universal Spinal System (USS). Designed to achieve the goals of scoliosis surgery.

USS Variable Axis Screw

TiLock 2 Spinal System. Surgical Technique

Dymaxeon Spine System. Simple, Streamlined, Smart. Surgical Procedure

USS II ILIO-SACRAL Modular System for Stable Fixation in the Sacrum and Illium

Surgical Technique. Occipito-Cervico-Thoracic System

VERTEX SELECT. surgical technique. adjustability. Flexibility. adaptability. Reconstruction System

Surgical Technique FOR U.S. DOMESTIC USE ONLY

O PE RATIV E TE C HN IQ U E. Minimally Invasive Posterior Fixation

L8 Spine System SURGICAL TECHNIQUE. Add: No.1-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China

Ballista Percutaneous Screw Placement System. Surgical Technique

AccuVision Minimally Invasive Spinal Exposure System

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L.

Conventus CAGE PH Surgical Techniques

SURGICAL TECHNIQUE. SECURIS Pedicle Screw System for Minimally Invasive Surgery. 2 I SECURIS Pedicle Screw System

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

SURGICAL TECHNIQUE GUIDE SOLANAS. Posterior Cervico-Thoracic Fixation System Adjustable Bridge System

Ballista Percutaneous Screw Placement System

MATRIX Spine System Degenerative. A posterior pedicle screw and rod fixation system.

Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine.

The Versatile Polyaxial Solution for the Universal Spine Systems. USS II Polyaxial. Surgical Technique

RESPONSE SURGICAL TECHNIQUE

Handling instructions. USS Low Profile. Thoracolumbar posterior fixation system.

EXPEDIUM VERSE Spinal System

4.5 System. Surgical Technique. This publication is not intended for distribution in the USA.

System. Humeral Nail. Surgical Technique

SURGICAL TECHNIQUE. Standard Reduction HA Coated Modular. Discover the Difference

Cervical Solutions. Optio-C Anterior Cervical Plate. with Allograft/Autograft. Surgical Technique Guide

Biomet Omega21 TM Spinal Fixation System

M.I.S. MAKE IT SMART IN ONE SYSTEM. Surgical Technique. Hip Knee Spine Navigation

MATRIX Spine System MIS Instrumentation. A minimally invasive instrument system for use with the MATRIX Spine System.

TM TM Surgical Technique

PRODUCT SUMMARY Cobalt Chrome Tulip Multiple Options Triple Lead Thread

Technique Guide. Pangea Spine System. Top loading pedicle screw and hook system for posterior stabilization and correction of spinal deformities.

TABLE OF CONTENTS. 2 (8144 Rev 2)

POLYAXIAL SPINE SYSTEM D E G E N E R A T I V E

Dual-Opening USS. For deformity correction.

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique. Hip Knee Spine Navigation

A Fusion of Versatility, Performance and Ease.

OSS Orthopedic Salvage System Compress Device. Surgical Technique

Dynesys Top-Loading System

XRL A modular expandable radiolucent vertebral body replacement system

Virage OCT Spinal Fixation System Surgical Technique

Surgical Technique 1

I. II. SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE

Technique Guide MAXIMUM ACCESS SURGICAL PLATFORM

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique

TiLock XT Minimally Invasive Surgery (MIS) Pedicle Screw System

Surgical Technique. Available In Titanium And Stainless Steel

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

LUMBAR POSTERIOR MINIMALLY INVASIVE SYSTEM. Surgical Technique

A U X I L I A R Y C O N N E C T O R S Surgical Technique

Xia 3 Spinal System. AIS surgical technique

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

fassier DUVal TelescoPIc IM system s RESCUE SYSTEM 2.0

VIPER Cortical Fix Fenestrated Screw System. Surgical Technique for OPEN & MIS Procedures

Surgical Technique Guide

Thoracolumbar Solutions. Vital Spinal Fixation System. Degenerative. Surgical Technique Guide

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

DIPLOMAT FIXATOR. open" and MIS" without having to change systems Spontaneous augmentation. Minimally invasive posterior instrumentation

Technique Guide. USS Small Stature/Pediatric. A multifaceted deformity system for use in patients of small stature.

Surgical Technique. Unisyn Hip System

OPTIONAL/ADDITIONAL INSTRUMENTS

The Top-loading Pedicle Screw and Rod System Designed for the Posterior Stabilization of the Lower Back. Click X System. Surgical Technique

EXPEDIUM 5.5 Spine System Family Product Catalogue

Occipital Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

TiLock XT Minimally Invasive Surgery (MIS) Pedicle Screw System

SerratoTM. Surgical technique

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

BAK/C Cervical Anterior Interbody Fusion System

FASSIER DUVAL TELESCOPIC IM SYSTEM S RESCUE SYSTEM 2.0

TECHNICAL BROCHURE. Capture Facet Fixation System

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

O PE RATIV E TE C HN IQ U E. ProView. Expandable Retractor System U.S. EDITION

Surgical Technique Guide

LCP Distal Tibia Plate

STRENGTH FROM WITHIN. Surgical Technique for Representatives

Trilogy Acetabular System

100 Interpace Parkway Parsippany, NJ

SFX Cross Connector System. Surgical Technique Guide

As edited by Dr. Oheneba Boachie-Adjei, Dr. Matthew Cunningham, Dr. John Kostuik, Dr. Raymund Woo and the Complex Spine Study Group et al

Technique Guide. NFlex. Semi-rigid rods for posterior lumbar stabilization.

Transcription:

Valencia Pedicle Screw Surgical Technique

VALENCIA CIRCUIT

TABLE OF CONTENTS Design Rationale Indications for Use Surgical Technique 1. Pedicle Preparation 2. Screw Insertion 3. Rod Placement 4. Locking Cap Insertion 5. Rod Reduction 6. Final Locking 7. Compression & Distraction 8. Cross Connector 9. Implant Removal

DESIGN RATIONALE 2 3 1 4 Solid, cannulated and reduction head polyaxial screws Low-torque hybrid square thread T25 dual lead screw body Up to 60 polyaxial angulation 4.5mm - 8.5mm screw diameters 25mm - 100mm screw lengths

INDICATIONS FOR USE The Altus Spine Valencia Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. Reference product insert (PI-001) for complete system indications for use, description, warnings and precautions

1. PEDICLE PREPARATION PREPARE PEDICLES Choose from the Awl, Straight Probes, Curved Probes, Pedicle Sound, and / or Taps for pedicle preparation Confirm depth using the indicators on the Bone Probes and Taps

1. PEDICLE PREPARATION (CONT.) PEDICLE MARKING Pedicle Markers can be used to determine the orientation and depth of pedicle cavities Right and left specific Pedicle Markers are supplied Select the appropriate Pedicle Marker and load into the Utility Driver Pedicle Marker assembly: pull back on the sleeve to the unlock position; insert the marker and release the sleeve to lock the assembly Place the Pedicle Marker into the prepared hole, unlock the sleeve and remove the Utility Driver, use fluoroscopy to confirm position Pull Back Sleeve Insert Marker

2. SCREW INSERTION ASSEMBLE DRIVER, STANDARD LOCKING Attach the Locking Driver to a Ratcheting Handle Load the appropriate Pedicle Screw onto the Locking Driver by engaging the T25 tip of the driver into the drive feature of the screw body Insertion Guides Note: Assembly requires alignment of distal tip of driver with slots of screw head Once the tip is properly seated in the screw, lower the sleeve of the driver into the threads of the screw Rotate the thumb tab clockwise until the screw is firmly attached (as shown) INSERT SCREW Insert the screw until the tip of the screw reaches the pedicle entry point, and confirm desired trajectory Advance the screw by rotating the Ratcheting Handle clockwise Correct Incorrect Note: Do not hold thumb tab while inserting screw as there is potential for locking sleeve to loosen during insertion REMOVE THE DRIVER Turn the thumb tab counterclockwise until it completely disengages the screw head Repeat screw insertion steps until all screws are in place Use the Utility Driver with the T25 Tip to make minor height adjustments after all screws are inserted

2. SCREW INSERTION (CONT.) ASSEMBLE DRIVER, SUPER LOCKER Attach the Locking Driver to a Ratcheting Handle Unlock the driver by pulling the tightening knob in the direction of the arrow marked Load Load the appropriate Pedicle Screw onto the Locking Driver by engaging the T25 tip of the driver into the drive feature of the screw body Note: Assembly requires alignment of distal tip of driver with slots of screw head Once the tip is properly seated in the screw, lower the sleeve of the driver into the threads of the screw Rotate the knob clockwise until the screw is firmly attached Lock the driver by pushing the knob in the direction of the arrow marked Lock INSERT SCREW Identical to standard driver, see prior section REMOVE THE DRIVER Unlock the driver by pulling the knob in the direction of the arrow marked Load Turn the knob counterclockwise until it completely disengages the screw head Repeat screw insertion steps until all screws are in place Use the Utility Driver with the T25 Tip to make minor height adjustments after all screws are inserted

3. ROD PLACEMENT SCREW HEAD ALIGNMENT Use the Head Positioner to achieve desired alignment Insert the Head Positioner into the head of the screw and rotate to desired position To aid in Rod placement, utilize the Rod Holder SELECT APPROPRIATE ROD Determine Rod length by trial or using the Rod Template ROD CONTOURING Ensure the appropriate contour (small, medium, large) is set on the main cam prior to bending Note: Do not reverse bend rods. Reverse bending may produce internal stresses which may become the focal point for eventual breakage of the implant

4. LOCKING CAP INSERTION LOAD LOCKING CAPS Load a T25 Locking Caps onto each and of the Double Sided Cap Inserter Introduce the tip of the inserter into the cap; a firm push will engage the cap properly Place the cap into the screw head and turn clockwise Use the Thread Start markings to aid in starting cap Align Markers

5. ROD REDUCTION MINOR REDUCTIONS Utilize the Rod Pusher to gain the additional clearance needed to start threading the cap MAJOR REDUCTIONS Engage the Rod Rocker to either side of the screw head Rock the instrument against the rod to reduce With the rod reduced, introduce the T25 Locking Cap

5. ROD REDUCTION (CONT.) SIGNIFICANT REDUCTIONS Place the distal end of the Rod Persuader down over the tulip head of the screw The inside edge of the Rod Persuader must be lower than the ledge on the side of the screw head Firmly squeeze the handles together to reduce the rod Note: If the Rod Persuader is not positioned correctly it may result in a jam. To release jam, hold the release knob down while using both hands to pull the handle levers apart Once the rod is appropriately reduced, insert the T25 Inserter loaded with a cap Turn the T25 Inserter clockwise, either manually or with a Ratcheting Handle, to insert the cap After provisionally tightening the cap, remove the Inserter and the persuader To remove the persuader, pull the release knob and allow the handles to retract

5. ROD REDUCTION (CONT.) EXTENSIVE REDUCTIONS Place the distal end of the Threaded Rod Persuader down over the tulip head of the screw The inside edge of the Rod Persuader must be lower than the ledge on the side of the screw head Twist the shaft handle clockwise to reduce the rod Once the rod is appropriately reduced, insert the T25 Inserter loaded with a cap Turn the T25 Inserter clockwise, either manually or with a Ratcheting Handle, to insert the cap Note: You may insert the T25 with loaded cap through the hollow shaft to reach screw if maintaining reduction is necessary for insertion After provisionally tightening the cap, remove the Inserter and the reducer To remove the reducer, turn shaft handle counterclockwise until tip comes free of the screw

6. FINAL LOCKING ASSEMBLE T25 DRIVE SHAFT TO TORQUE LIMITING HANDLE Pull back the plunger and insert the shaft until the load line is flush with the plunger, then release Lower the Counter Torque over the head of the screw until seated against the rod Insert the driver into the Counter Torque Note: It is important to ensure that the T25 Drive Shaft is properly engaged in the Locking Cap prior to final locking. Failing to do so may result in a damaged construct or instrument Turn the Torque Limiting Handle clockwise until it clicks The audible and tactile feedback confirms the construct is locked down to its proper specification Remove Driver and Counter Torque from the screw Repeat Section 6 steps for remainder of the construct Note: Final locking must be secured using a Torque Limiting Handle of 65 in-lb (7.3 N-m)

7. COMPRESSION & DISTRACTION (OPTIONAL) Ensure that one locking cap has been through Final Locking (section 6). Keep the second cap is seated on the rod, but not locked Position the Compressor or the Distractor such that the instrument straddles the rod, while still making contact with the screw heads Squeeze the handles together until the desired amount of compression / distraction has been achieved The instrument has a ratcheting lock at its proximal end if needed, or it can be folded inside the handle if it is not needed 8. CROSS CONNECTORS (OPTIONAL) The Cross Connectors come in 4 different sizes, (small, medium, large, and extra-large) and automatically adjust length, rotation, and angulation to seat properly on each rod Place each hook end of the Cross Connector over the rod, in the desired location Connect the Cross Connector Cam Driver to the X/C Torque Limiter Handle and place the instrument over the hook slots. Turn clockwise until audible click Connect the Cross Connector Nut Driver to the X/C Torque Limiter Handle and place the instrument over the center nut. Turn clockwise until audible click Repeat this for each Cross Connector

9. IMPLANT REMOVAL (OPTIONAL) CROSS CONNECTOR REMOVAL Connect the Cross Connector Nut Driver to the X/C Torque Limiter Place the instrument over the center nut and turn counterclockwise until loose Connect the Cross Connector Cam Driver to the X/C Torque Limiter Place the instrument into the cam socket and turn counterclockwise until loose and in the unlock position PEDICLE SCREW REMOVAL Engage the Counter Torque over the screw head Insert the Screw Driver into the Counter Torque, turn counterclockwise Repeat for each locking cap Remove rods from construct If necessary, use the Head Positioner to mobilize the polyaxial screw head of implant Engage the Screw Driver T25 Tip into the screw body socket, turn counterclockwise and remove from the pedicle; repeat for each pedicle screw

Contact Us to Order Product 2017 Altus Spine. U.S. and International Patents Pending Info@altus-spine.com www.altus-spine.com BR-2301 Rev A CORPORATE HEADQUARTERS 1340 Enterprise Drive West Chester, PA 19380 Office: (610) 355-4156 // Fax: (610) 300-3049