Cardiac Rhythm Therapy // Advanced Features // ProMRI ProMRI Checklist and Quick Reference Guide
PAGE 2
Conditions for MR scans with a BIOTRONIK ProMRI system Usage of an MR scan on a patient having an implanted pacemaker (single- or dual-chamber), ICD, or CRT-D is only possible under highly specific prerequisites and conditions. BIOTRONIK offers a comprehensive, breakthrough portfolio of device systems approved for MR scans. These products are marked as ProMRI. Please use the matrix on the last page of this brochure or the BIOTRONIK ProMRI System Technical Manual to check whether the patient s ProMRI device-lead combination is approved for full body MR scans. The requirements listed in this guide apply to all ProMRI systems. PAGE 3
Conditions for MR scans with a BIOTRONIK ProMRI system (continued) Restrictions for the patient and the cardiac device system The device system consists of a pacemaker, ICD, or CRT-D with the respective leads that are separately labeled MR conditional and, when combined, constitutes an MR conditional device system (See charts on p. 10). There are no other active or abandoned cardiac implants (e.g., lead extensions, lead adapters or abandoned leads) in the patient s body. Other active or passive implants are permitted if they are identified as MR conditional by the manufacturer. Note: An MRI scan is permitted only if the product-specific conditions are met for all implants and if no metal implantable device longer than 5 cm is in the vicinity of a BIOTRONIK lead within a distance of less than 4 cm. The leads have been implanted for at least 6 weeks. The device system is implanted pectorally. The measured pacing threshold is not above 2.0 V at 0.4 ms pulse width. The device system should be functioning normally prior to an MRI. The battery status is neither ERI nor EOS. The device is programmed to an MRI mode immediately before the MR scan. PAGE 4
Conditions for MR scans with a BIOTRONIK ProMRI system (continued) Requirements of the MRI scanner Use of a clinical MRI machine with: A closed tube Cylindrical magnets A static magnetic field strength of 1.5 tesla. The slew rate of the MRI scanner s gradient fields should not exceed 200 T/m/s per axis. Restrictions during the MRI scan The MR scan should be performed with the patient in supine position. The mean specific absorption rate (SAR) for the whole body displayed by the MRI scanner must not exceed 2.0 W/kg. The head absorption rate displayed by the MRI scanner must not exceed 3.2 W/kg. Emergency equipment for resuscitation must be kept at hand and properly certified staff must be available. Continuously monitor the patient s condition during the entire MR scan using at least one of the following parameters: blood oxygen saturation, blood pressure or ECG. Note: The ECG function integrated in the MRI scanner is often not permitted for patient monitoring. Therefore, only use devices which are permitted for patient monitoring in an MRI environment. PAGE 5
ProMRI checklist for the radiology department Checklist before MRI scanning This checklist will help to ensure the safe application of an MR scan on patients with a BIOTRONIK device system that has been labeled MR Conditional. * We recommend checking the boxes off while using this checklist to ensure that an MR Conditional scan with BIOTRONIK ProMRI systems is permissible. Please check to see whether the patient s ProMRI device-lead combination is approved for a full body MR scan. (See charts on p. 10.) General requirements for patients with ProMRI systems Use of a clinical MRI system with a closed tube, cylindrical magnets and a static magnetic field strength of 1.5 tesla. The slew rate of the MRI scanner s gradient fields should not exceed 200 T/m/s per axis. The MR scan should be performed with the patient in supine position. The mean specific absorption rate for the whole body displayed by the MRI scanner must not exceed 2.0 W/kg. The head absorption rate displayed by the MRI scanner must not exceed 3.2 W/kg. Emergency equipment for resuscitation must be kept at hand and properly certified staff must be available. PAGE 6
ProMRI checklist for the radiology department (continued) The patient should be continiously monitored in an appropriate manner during the entire MR scan. The following parameters can be observed: blood oxygen saturation, blood pressure, ECG. Note: The ECG function integrated in the MRI scanner is often not permitted for patient monitoring. Therefore, only use devices which are permitted for patient monitoring in an MRI environment. Ensure with the cardiology department that the device is programmed to MRI mode prior to the MR scan. Name and signature Date * Details on these conditions and requirements can be found in the BIOTRONIK ProMRI System Technical Manual. Or, contact your local BIOTRONIK representative and visit www.biotronikusa.com/promri. PAGE 7
ProMRI checklist for the cardiology department Use the following checklist to ensure that patients who have been implanted with a BIOTRONIK device system labeled MR Conditional can receive an MR scan safely. * We recommend that you check off the boxes in order to be certain that a full body MR Conditional scan with BIOTRONIK ProMRI systems is permissible. Patient Name Address City Cardiac device system (please complete) Device * Detailed information about ProMRI can be found in the BIOTRONIK ProMRI System Technical Manual. You can download this document as a PDF file from the website: www.biotronikusa.com/manuals. Alternatively, contact your local BIOTRONIK representative and visit the website www.biotronikusa.com/promri. PAGE 8
ProMRI checklist for the cardiology department (continued) The device system consists of a pacemaker, ICD, or CRT-D with the respective leads that are separately labeled MR conditional and, when combined, constitutes an MR conditional device system (See charts on p. 10). There are no other active or abandoned cardiac implants (e.g., lead extensions, lead adapters or abandoned leads) in the patient s body. Other active or passive implants are permitted if they are identified as MR conditional by the manufacturer. Note: An MRI scan is permitted only if the product-specific conditions are met for all implants and if no metal implantable device longer than 5 cm is in the vicinity of a BIOTRONIK lead within a distance of less than 4 cm. The leads have been implanted for at least 6 weeks. The device system is implanted pectorally. The measured pacing threshold is not above 2.0 V at 0.4 ms pulse width. The device system should be functioning normally prior to an MRI. The battery status is neither ERI nor EOS. Ensure that the device is programmed to an MRI mode immediately before the MR scan and reprogrammed to the initial settings after the scan. Post MR scan requirements met (See p. 10). Name and signature Date * Details on these conditions and requirements can be found in the BIOTRONIK ProMRI System Technical Manual. Or, contact your local BIOTRONIK representative and visit www.biotronikusa.com/promri. PAGE 9
Post MR scan requirements The patient must undergo follow-up device interrogation, which is necessary for the patient s safety for two reasons: To reprogram device to original pacing parameters. To access the device system for any adverse effects caused by the MR scan. ProMRI system check Pacemakers Bradycardia Devices Eluna DR-T Entovis DR-T Eluna SR-T Entovis SR-T Setrox 60 Setrox 53 Siello S 45 DX ICDs Siello S 53 Siello S 60 Solia S 45 Solia S 53 Solia S 60 Linox smart S DX 65/15 Linox smart S DX 65/17 Tachycardia Devices Iperia VR-T DX Inventra VR-T DX Iforia VR-T DX PAGE 10
DR-T DF-1 ICDs Tachycardia Devices Iperia DR-T (DF-1) Iforia DR-T (DF-1) Setrox 53 Siello S 45 Siello S 53 Atrial Siello S 60 Solia S 45 Solia S 53 Solia S 60 RV Linox smart S 65 Linox smart SD 65/18 DR-T DF-4 ICDs Tachycardia Devices Iperia DR-T (DF4) Iforia DR-T (DF4) Setrox 53 Siello S 45 Siello S 53 Atrial Siello S 60 Solia S 45 Solia S 53 Solia S 60 PAGE 11 Continued >
DR-T DF-4 ICDs (continued) Tachycardia Devices Iperia DR-T (DF4) Iforia DR-T (DF4) Protego S 65 Protego S 75 RV Protego SD 65/16 Protego SD 65/18 Protego SD 75/18 HF-T DF-1 CRT-Ds Tachycardia Devices Iperia 7 HF-T (DF-1) Inventra 7 HF-T (DF-1) Atrial Setrox S 53 Linox smart S 65 RV Linox smart SD 65/18 Corox OTW 75 BP Corox OTW 85 BP Corox OTW-L 75 BP LV Corox OTW-L 85 BP Corox OTW-S 75 BP Corox OTW-S 85 BP PAGE 12
HF-T DF4 CRT-Ds Tachycardia Devices Iperia 7 HF-T (DF4) Inventra 7 HF-T (DF4) Atrial Setrox S 53 Protego S 65 Protego S 75 RV Protego SD 65/16 Protego SD 65/18 Protego SD 75/18 Corox OTW 75 BP Corox OTW 85 BP Corox OTW-L 75 BP LV Corox OTW-L 85 BP Corox OTW-S 75 BP Corox OTW-S 85 BP PAGE 13
ProMRI Checklist and Quick Reference Guide www.biotronikusa.com/promri BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035-5369 (800) 547-0394 (24-hour) (800) 291-0470 (fax) www.biotronik.com M4186-C 06/16 2016 BIOTRONIK, Inc. All rights reserved. BR316r2 6/10/16