A TWO DEGREE OF FREEDOM ROBOTIC EXERCISER FOR REHABILITATION

Similar documents
DESIGN OF MECHATRONIC ACTUATION SYSTEMS BASED ON FLUIDIC ARTIFICIAL MUSCLES

710. Movement control of active/passive exercisers

Navigator: 2 Degree of Freedom Robotics Hand Rehabilitation Device

NZQA Expiring unit standard 7026 version 4 Page 1 of 7. Apply knowledge of functional anatomy and biomechanics

KINEMATIC ANALYSIS OF ELBOW REHABILITATION EQUIPMENT

What is Kinesiology? Basic Biomechanics. Mechanics

ANNALS of the ORADEA UNIVERSITY. Fascicle of Management and Technological Engineering, Volume X (XX), 2011, NR2

Angular Measurements with BIOPAC Goniometers & Torsiometers

EMG-Driven Human Model for Orthosis Control

Lecture 2. In The Name of Allah. Instrumentation. Dr. Ali Karimpour Associate Professor Ferdowsi University of Mashhad

professional rehabilitation PRODUCT RANGE by

Control principles in upper-limb prostheses

Electromyogram-Assisted Upper Limb Rehabilitation Device

Design Considerations of a Human Arm Exoskeleton

Multi-joint Mechanics Dr. Ted Milner (KIN 416)

Five-Fingered Assistive Hand with Mechanical Compliance of Human Finger

DEVELOPMENT OF WEARABLE MASTER-SLAVE TRAINING DEVICE FOR UPPER LIMB CONSTRUCTED WITH PNEUMATIC ARTIFICIAL MUSCLES

Estimation of the Upper Limb Lifting Movement Under Varying Weight and Movement Speed

Hemiplegic Shoulder. Incidence & Rationale. Shoulder Pain Assessment & Treatment

Therapy Manual DO NOT PRINT

1. EXECUTIVE SUMMARY 2. PRODUCT SUMMARY

ECHORD call1 experiment MAAT

Simulator Based Experimental Motion Analysis of 3D Printed Artificial Shoulder Joint Geometries

DESIGN AND IMPLEMENTATION OF A MECHATRONIC SYSTEM FOR LOWER LIMB MEDICAL REHABILITATION

Development of a 6DOF Exoskeleton Robot for Human Upper-Limb Motion Assist

of the joint coordination in reach-to-grasp movements.

AN EXOSKELETAL ROBOT MANIPULATOR FOR LOWER LIMBS REHABILITATION

AT89C51 Microcontroller Based Control Model for Hybrid Assistive Limb (Knee )

New methods for the simulation with finite element of the human elbow

Description This lesson explains follow-through motion as it is applied to secondary actions as well as character animation.

The Human Machine: Biomechanics in Daily Life.

Among the non-operative methods, serial or wedge casting of the patient's arm in forced extension has been used with success. However, the weight and

Control of Finger Forces during Fast, Slow and Moderate Rotational Hand Movements

Etrouth Srikar 1, Habib Ali 2 1, 2 Dept. Biomedical Engineering, J.B Institute of Engineering and Technology, India

Verification of Immediate Effect on the Motor Function of a Plegic Upper Limb after Stroke by using UR-System 2.2

Design and Implementation study of Remote Home Rehabilitation Training Operating System based on Internet

Accuracy and validity of Kinetisense joint measures for cardinal movements, compared to current experimental and clinical gold standards.

Q: What is the relationship between muscle forces and EMG data that we have collected?

Overview of the Human Arm Anatomy

Learning Classifier Systems (LCS/XCSF)

FATIGUE DEVICE FOR TESTING ANKLE JOINT ENDOPROSTHESES

ASSESSMENT OF FLEXIBILITY

Thesis Rehabilitation robotics (RIA) Robotics for Bioengineering Forefront research at PRISMA Lab and ICAROS Center

Design and Modeling of an Upper Extremity Exoskeleton

PEMP-AML2506. Day 01A. Session Speaker Dr. M. D. Deshpande. 01A M.S. Ramaiah School of Advanced Studies - Bangalore 1

Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body weight, height, and sex in 178 healthy subjects

Higher National Unit specification: general information

Chapter 20: Muscular Fitness and Assessment

Wearable Hand Rehabilitation Robot Capable of Hand Function Assistance in Stroke Survivors*

CHAPTER 18 UNIVERSITY OF WISCONSIN AT MILWAUKEE

UNCORRECTED PROOFS. Figure 8.14 The first version of the NEUROExos 8.5 CASE STUDY: AN UPPER LIMB POWERED EXOSKELETON

05/06/14 Name Physics 130 Final Exam

12 th National Congress on Theoretical and Applied Mechanics September 2013, Saints Constantine and Helena, Varna, Bulgaria

Study on the control of variable resistance for isokinetic muscle training system

TRANSFORMER ALL IN ONE ISOKINETIC MACHINE ANKLE ELBOW HIP SHOULDER WRIST TRUNK

Design and Control of a Robotic Assistive Device for Hand Rehabilitation (RAD-HR)

Lecture 2. Statics & Dynamics of Rigid Bodies: Human body 30 August 2018

A Patient s Guide to Adult Forearm Fractures

Importance of Developmental Kinesiology for Manual Medicine

Development and application of a hand force measurement system

Work Environment Physics

Hand of Hope. For hand rehabilitation. Member of Vincent Medical Holdings Limited

Dimensional Requirements. Minimum space requirements for passageways and hallways.

Omo and Manu Neurexa plus

NEUROLOGICAL injury is a leading cause of permanent

Robot control using electromyography (EMG) signals of the wrist

FEASIBILITY OF EMG-BASED CONTROL OF SHOULDER MUSCLE FNS VIA ARTIFICIAL NEURAL NETWORK

Using the Kinect to Limit Abnormal Kinematics and Compensation Strategies During Therapy with End Effector Robots

Electromyography II Laboratory (Hand Dynamometer Transducer)

State of equilibrium is when the sum (or net effect) of forces acting on a body equals zero

Joints. Vi Michelle Austin

MOTOR EVALUATION SCALE FOR UPPER EXTREMITY IN STROKE PATIENTS (MESUPES-arm and MESUPES-hand)

SIMULATOR FOR TESTING THE FRICTION TORQUE OF THE FEMORAL HEAD OF A HIP PROSTHESIS

Design of a 7 Degree-of-Freedom Upper-Limb Powered Exoskeleton

Definition of Anatomy. Anatomy is the science of the structure of the body and the relation of its parts.

Project of a medical device to support the diagnosis of the ankle pathologies

DEVELOPMENT AND CONTROL OF A 3DOF UPPER-LIMB ROBOTIC DEVICE FOR PATIENTS WITH PARETIC LIMB IMPAIRMENT

Chapter 8 8/23/2016. Body Mechanics and Patient Mobility. Introduction to Body Mechanics and Patient Mobility

Optimization of Backward Giant Circle Technique on the Asymmetric Bars

Chapter 6. Results. 6.1 Introduction

NDTA BASIC ADULT COURSE KEY TO COURSE CONTENT

A ROBOT FOR WRIST REHABILITATION

Kinematic and Dynamic Adaptation of

Design and Control of an Exoskeletal Rehabilitation Device for Stroke and Spinal Cord Injury Patients

Sensory. Physical. Cognitive. Language 4/12/2016 LEARNING OBJECTIVES AT ASSESSMENT: SKILLS EVALUATION OVERVIEW

Orbital 360. Introducing: The Orbital 360. Joseph Noel Office: (515) NW 5 th Street Ankeny, IA 50023

Biomechanics. MCE 493/593 & ECE 492/592 Prosthesis Design and Control September

Fast Simulation of Arm Dynamics for Real-time, Userin-the-loop. Ed Chadwick Keele University Staffordshire, UK.

Anatomy. Anatomy deals with the structure of the human body, and includes a precise language on body positions and relationships between body parts.

WHAT is ERGONOMICS. What ergonomics does can be summed up in three questions: Who (Human) was it designed for? What (Task) was it designed for?

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

The Handmaster NMS1 surface FES neuroprosthesis in hemiplegic patients

Unilateral and Bilateral Rehabilitation of the Upper Limb Following Stroke via an Exoskeleton

REHABILITATION IN VIRTUAL REALITY VAST.REHAB IS DESIGNED FOR PHYSICAL, OCCUPATIONAL AND COGNITIVE THERAPY

Smart. Training. Developing advanced exercise machines

6.4 The Ankle. Body Divided into Planes. Health Services: Unit 6 Arms and Legs. Body Movement Vocabulary

Design considerations for an active soft orthotic system for shoulder rehabilitation

Re-establishing establishing Neuromuscular

rapael smart rehab solution

Transcription:

Journal Mechanisms and Manipulators, Vol. 5, Nr. 1, 2006, pp. 21-26 ARoTMM IFToMM A TWO DEGREE OF FREEDOM ROBOTIC EXERCISER FOR REHABILITATION Călin RUSU *, Dan MÂNDRU **, Vistrian MĂTIEŞ *** * ) Lecturer PhD. Eng. Technical University of Cluj-Napoca, e-mail: rusucalin74@yahoo.com ** ) Prof. PhD. Eng. Technical University of Cluj-Napoca, e-mail: mandrud@yahoo.com ** ) Prof. PhD. Eng. Technical University of Cluj-Napoca, e-mail: matiesvistrian@yahoo.com Abstract: Current therapy exercisers having one degree-of-freedom for rehabilitating single muscle groups are limited; an efficient functional rehabilitation process requires multiple muscles group s actions within complex movements of the limbs. In this paper an exerciser for active and passive movement of upper limb is proposed. Its structure is based on a five-bar mechanism actuated by two motors. This mechanism guides the patient s upper limb through a series of desired exercises. The dimensional synthesis of the five-bar mechanism is based on its workspace and anthropomorphic data of the patients. Some exercises aiming the recovery of motor and cognitive skills are described and discussed. The user interfaces are given too. 1. Introduction An important component of the actual medicine, beside the prophylactic and curative medicine, is rehabilitation, named also readaptation or reeducation, which is a complex medico-social assisting process that has as objective the reintegration of disabled in family and society, [11]. Its specific ways of action concern the achievement of optimal values for the morpho-functional capacity, psychical status, professional training, and social status. In order to obtain positive results, the rehabilitation treatment must be applied methodically and persistently, carefully controlled, aided by technical systems, which could make it more efficient, [2]. The advances that had occurred in the field of Rehabilitation Engineering are: prosthetics, orthotics, mobility aids, walk assist devices, sensorial augmentation systems, exercisers, rehabilitation robotics, assistive technology, a.s.o., [10]. Kinetotherapy is defined as an ensemble of procedures which promote motion as a basic element of rehabilitation treatment, [7]. Its task is the amelioration or liquidation of disabilities by performance s improvement (forces, precision, and mobility of locomotion apparatus). Kinetotherapy aspects are varied, including: walking, running, gym, games, training, rehabilitation of walk and prehension, hydrokinethoterapy, and so on. The kinetic treatment is recommended in the following cases: posttraumatic affections of the locomotion apparatus, diseases of the nervous system and respiratory system, rheumatic illnesses, cardiac, metabolic and nutrition-related diseases, [1], [12]. Baker [2] presented a therapeutic wrist rotator for the passive pronation - supination of the forearm. The device, fixed to a stationary table, includes an electric actuator driving a reduction drive to provide relatively slow rotational speed and relatively high torque to the output shaft and handgrip. On / off and rotational direction switches are provided to control the device. Bernardo et al. [4] studied the possibilities to use shape memory actuators in the structure of intelligent exercisers. They developed an actuation device based on active element U form, made of CuZnAl alloy with two-way shape memory effect. The device is suited to rehabilitation of the fingers joints. Gallasch et al. [9] proposed a therapeutic exerciser equipped with two angular voice coil actuators coupled by levers to the upper limb. The device designed by Gregorio et al. [13] is equipped with a controller for the sensing of the position of the handle at the end of an arm in the X and Y directions and in Z direction through a hinge which permits the arm to rotate upwardly. A therapy and training device for the shoulder joint with a trunk base having two shoulders extensions on a rotational joint base was described by Hassler [14]. It has a modular structure and can produce passive motions of the shoulder and elbow. Saringer et al. [8] presented a continuous passive motion device including an upper arm support suitably fixed to a drive actuator and an adjustable forearm support. Various cuffs are provided so a patient can secure the limb to the device; so pronation supination is passively created. More recently, Solomon et al. [17] described a therapeutic mobilisation and positioning device of joints having a control device that measures the force through the deformation of an elastic component. Elbow and wrist are the target joints and flexion-extension and pronationsupination are the assisted motions. Current therapy machines, [18], [19] having one degree of freedom for rehabilitating single muscle groups are limited in the rehabilitation process because the functional rehabilitation requires the rehabilitation of multiple muscle groups. The increasing of the flexibility and of the functional performances is assured by using robotic systems in the kinetotherapy processes. 21

2. The Developed Exerciser The proposed exerciser is designed to be used in passive and active upper limb s mobilization, aiming the reattainment of the shoulder articulations, elbow s and wrists movement capacity and patient s motor skills. It permits a large variety of exercises, which can be accomplished automatically, and modified by programming them and allows the disabled persons tele-assistance. The several patients control by a single therapist is possible and with the aid of adequate equipment, the system can monitorize the most important of the patient s biomechanic parameters. 3 M 2 1 a b c Fig. 1 The structure of the exerciser (a, b) and normal zones of convenient reach of upper limbs (c) According to figure 1a, user 1, in the sitting position, has its upper limb linked with mechanical structure 2 (the hand is in direct contact with a handle). The upper limb is mobilized, actively or passively, in the marked plan. Mechanical structure 2 is connected to computer 3. The upper limb s movement is occurring into a plane, it results that the mechanical structure must be based on two degrees of freedom mechanism. The mechanical structure presented in figure 1b is consisting in using a five-bar mechanism, the driving joints being placed at the base (elements 4 and 5 are input elements). A similar robotic exerciser is described in [3]. The dimensional synthesis of the pentalater mechanism is based on its workspace, taking into account the so-called zones of convenient reach (ZCR) of upper limbs, for the sitting position, [15], [16]. In figure 1c, 8 represent the normal ZCR, for the left upper limb, 9 for the right upper limb, and 10 for the both. The exerciser responds to the requirements of passive or active motions of the arm (backwardforward projection, inner-outer rotation and abduction-adduction), of the forearm (flexion-extension and pronation supination) and of the hand (abduction adduction and flexion extension), [11]. As is displayed in figure 2, the actuated joint variables are φ 1 and φ 3, while the cartesian variables are the (x, y) coordinates of the revolute center P. Lengths d, l and l 1 define the geometry of this mechanism entirely. The relationship between the linear velocity of the characteristic point of the mechanism P(x,y) and the angular velocities of the driving elements is : V x 1 Jp J V (2.1) y 3 22

where, J p and J φ are, respectively, the direct-kinematics and the inverse-kinematics matrices of the mechanism, defined as, l1 cos 2 l1 sin 2 ll1 sin( 1 2) 0 J p ; J (2.2) l1 cos 4 l1 sin 4 0 ll1 sin( 3 4 ) The singular configurations associated to J p respectively J φ are called direct-kinematic singularities and inverse-kinematic singularities. The direct-kinematic singularities are located inside the Cartesian workspace and the inverse-kinematic singularities are often located on its boundaries. When the mechanism is close to a direct-kinematic singularity, a small joint rate can generate a large velocity of the characteristic point. This means that the positioning accuracy is lower in some directions for some configurations close to direct-kinematic singularities because the encoder resolution is amplified. In addition, velocity amplification in one direction is equivalent to a loss of stiffness in this direction. For an efficient therapy exerciser, the detection of the singular configurations is an important problem because in this case, exist some directions, for which the mechanism cannot move or can become difficult to control. The condition number k of matrix J p defined as: 1 cos( 4 2 ) k (2.3) 1 cos( 4 2 ) quantifies the proximity to a configuration where the velocity amplification is equal in any direction. When matrix J p attains a condition number of unity, it is termed isotropic, its inversion being performed without any round-off-error amplification. When the condition number has a value of zero, the mechanism is in a singular position. Mechanism postures for which condition 2 holds are thus the most 4 2 / accurate for purposes of the direct kinematics of the mechanism. Correspondingly, the locus of points whereby matrix J p is isotropic is called the isotropy locus in the Cartesian workspace (figure 3). Fig. 2 The five-bar mechanism Fig. 3 Isotropy locus in the workspace 3. The Rehabilitation Exercises For the active exercises, the therapist learns the patient about the motions to be executed, and the exerciser displaces trajectories and targets to be accomplished. In the first exercise the patient has to displace the mechanism s characteristic point on the target circle, with displaying the precision of this task. This application is intended for active exercises. The user interface for this application is presented in figure 4. In order recover the control abilities, it is necessary that the cursor 6 to overlap on the target 4 by moving the mechanism of the exerciser. The target is placed on the circle 1 and its position can be modified. The 23

proximity is marked by indicators 8. The positioning error can be controlled with button 7. The position and radius of the circles 1 and target 4 can be modified using cursors 2 respectively buttons 3 and 5. This facility allows various motions in the shoulder, elbow and wrist joints. The results are depicted in figure 4. Fig. 3 The user interface for the first application a b c d Fig. 4 Different target positions For passive exercises, the exerciser interacts with the upper limb, moving it through the marked points and along the desired trajectories. The user s interface for the second application (for passive motions) is presented in figure 5. The characteristic point of the mechanism is programmed to move along an arbitrary trajectory 5. Its form and position in the workspace could be modified, using table 1. Hereby, the range of motions performed by the patient s limb is controlled, and the desired anatomical movements could be selected. The speed and the direction of motion are possible to be controlled by moving the cursor along the programmed trajectory. By interpolation of the x, y values from table 1, it s possible to obtain wide variety of trajectories. Two examples are presented in figure 6. In figure 7 in presented the prototype of the symmetric five-bar mechanism used for the applications. The lengths of the elements are 272 mm respectively 400 mm and the distance between the two fixed joints is 600 mm. The driving elements are actuated by two DC servomotors with reduction gears.the servomotors 24

are driven by linear voltage amplifiers. These amplifiers receive the command signals from the 12 bits digital to analogue converters, which are integrated in the data acquisition and control board DAQPad 1200. The transducers are, in this case, two precision rotary potentiometers Fig. 5 The user interface for the second application Fig. 6 Different trajectories for passive motions Fig.View 7 Thesoftware prototypefollows of the exerciser For the first application, the Lab these stages: 25

- based on direct kinematic model, the coordinates X and Y of the characteristic point are calculated; - these values are compared with the ones imposed by the position of the target, resulting the positioning error in Cartesian space - if the positioning error is into the admissible range, the light indicators is settled For the second application: - the values for angles φ 1,φ 3 are calculated, taking into consideration the coordinates X and Y of the points on the trajectory. - these values are compared with the ones received from the transducers, resulting the position and velocity errors ; - at every 5 ms an PI controller with feed-forward compensation, processes the information received from the transducers, and establishes the strategy to command the motors, aiming the error s cancelling. 4. Conclusions A 2 DOF exerciser was developed allowing active or passive complex motions of upper limbs taking into account individual needs. The mechanical structure is composed by a five-bar mechanism actuated by two DC servomotors. The exerciser offers the control opportunities of exercises parameters. It can adapt to anthropometric dimension and to the morpho-functional residual capability of each patient offering a more flexibility, concerning the exercise variety. Two applications were developed and described: one for active motion and the other one, for passive motion of upper limb. References 1. Baciu, C., Aparatul locomotor - Anatomie funcţională, semiologie clinică, diagnostic diferenţial, Editura Medicală, Bucureşti, 1981. 2. Bronzino, J.D. (editor), Biomedical Engineering Handbook, CRC Press & IEEE Press, Hartford, 1995. 3. Hogan, N., Interactive Robotic Therapist, U.S. Patent no. 5.466.213, 1995. 4. Bernardo, D., Martinet, N., Hean, C., Shape memory intelligent actuator for application to rehabilitation, Proc. of the 9 th International Conference on Mechanics in Medicine and Biology, pp. 127 131, 1996. 5. Baker, E., Therapeutic wrist rotator, U.S. Patent no. 5.788.607, 1998. 6. Kroemer, K., Ergonomics How to design for ease and efficiency, Prentice Hall Englewood Cliffs, 1998. 7. Kiss, I. Fizio-kinetoterapie şi recuperarea medicală în afecţiunile aparatului locomotor, Editura Medicală, Bucureşti, 1999 8. Saringer, J., Culhane, J., Continuous passive motion device for upper extremity forearm therapy, U.S Patent no. 5.951.499, 1999. 9. Gallasch, E., Taferner, H.P., Fend, M., Stiffness controlled voice coil actuator for muscoskeletal studies, Proceedins of the International Conference ACTUATOR 2000, pp. 639 642, 2000. 10. Mândru, D., Mătieş, V., Tătar, O., Research concerning the rehabilitation of the prehension function, Acta Electrotehnica Napocensis, vol. 41, nr. 1, pp. 46 51, 2000. 11. Mândru, D., Ingineria protezării şi reabilitării, Editura Casa Cărţii de Ştiinţă, Cluj-Napoca, 2001. 12. Plas, H.F., Kinetoterapie activă, Editura Polirom, Iaşi, 2001. 13. Gregorio, P. et al., Hand controller, U.S. Patent no. 6.781.569, 2004. 14. Hassler, A., Therapeutic and training device for the shoulder joint, U.S. Patent no. 6.824.498, 2004. 15. Mândru, D., Crişan, R., Tătar, O., Noveanu, S., Acţionări în mecanica fina si mecatronică, Alma Mater, Cluj- Napoca, 2004. 16. Mândru, D., Rusu, C., Noveanu, S., Research concerning the development of a robotic system for rehabilitation exercises, Mecatronica, nr. 2, pp. 55 60, 2004. 17. Solomon, A., Culhane, J., Cotterel, D., Control device for the therapeutic mobilization of joints, U.S. Patent no., 6.743.187, 2004. 18. www. alimed.com 19. www. powertrainer.com 26