The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews

Similar documents
The International Journal of Periodontics & Restorative Dentistry

THE USE OF TEMPORARY ANCHORAGE DEVICES FOR MOLAR INTRUSION & TREATMENT OF ANTERIOR OPEN BITE By Eduardo Nicolaievsky D.D.S.

6. Timing for orthodontic force

ORIGINAL ARTICLE , 1,2 onplant, 3,4, . 12, 13,14 , 7,8. 9, ,,. 16,17

Bone density assessment for mini-implants position

Orthodontic Microimplants and Its

ORTHODONTIC CORRECTION Of OCCLUSAL CANT USING MINI IMPLANTS:A CASE REPORT. Gupta J*, Makhija P.G.**, Jain V***

In recent years, orthodontic mini-implants have been

Intraoral molar-distalization appliances that

Clinical indices for orthodontic mini-implants

Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic Device

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and.

The miniscrew has become one of the

Microscrew implant as an orthodontic anchoring device

Research Article Success Rate of Microimplants in a University Orthodontic Clinic

PAIN PERCEPTION DURING MINIPLATE-ASSISTED ORTHODONTIC THERAPY

Mini-implants in Orthodontics: A Review

Anchorage control is a critical consideration when

There is little controversy regarding whether temporary

International Journal of Medical and Health Sciences

3D Cortical Bone Anatomy of the Mandibular Buccal Shelf: a CBCT study to define sites for extra-alveolar bone screws to treat Class III malocclusion

MOLAR DISTALIZATION WITH MODIFIED GRAZ IMPLANT SUPPORTED PENDULUM SPRINGS. A CASE REPORT.

AAO 115th Annual Session San Francisco, CA May 17 (Sunday), 1:15-2:00 pm, 2015

Temporary Orthodontic Anchorage Devices

KJLO. A Sequential Approach for an Asymmetric Extraction Case in. Lingual Orthodontics. Case Report INTRODUCTION DIAGNOSIS

Molar distalisation with skeletal anchorage

Rotational Resistance of Surface-Treated Mini-Implants

INDIAN JOURNAL OF DENTAL ADVANCEMENTS

Anchorage control is a fundamental prerequisite

Correction of a maxillary canine-first premolar transposition using mini-implant anchorage

TADs Supported Tongue Crib: A New Minimalistic Design

Technique Guide. Orthodontic Bone Anchor (OBA) System. Skeletal implants for the orthodontic movement of the teeth.

Since their introduction, the use of orthodontic

Treatment of Class II, Division 2 Malocclusion with Miniscrew Supported En-Masse Retraction: Is Deepbite Really an Obstacle for Extraction Treatment?

Maxillary Molar Distalization with micro-implants:

Palatal Bone Density in Adult Subjects: Implications for Mini-Implant Placement

The use of mini-implants in en masse retraction for the treatment of bimaxillary dentoalveolar protrusion

Influence of Third Molars on the Availability of Interradicular Spaces for Miniscrew Implant Placement

Skeletal class III maloeclusion treated using a non-surgieal approach supplemented with mini-implants: a case report

Evaluation of cortical bone thickness of mandible with cone beam computed tomography for orthodontic mini implant installation

Many patients require prosthodontic rehabilitation

Sliding Mechanics with Microscrew Implant Anchorage

Osseointegrated dental implant treatment generally

Contemporary Implant Dentistry

Anchorage conservation has always been a challenge

Influence of Surface Characteristics on Survival Rates of Mini-Implants

Treatment of Long face / Open bite

Gentle-Jumper- Non-compliance Class II corrector

Ortho-surgical Management of Severe Vertical Dysplasia: A Case Report

The treatment options for nongrowing skeletal Class

Effectiveness of en-masse retraction using midpalatal miniscrews and a modified transpalatal arch: Treatment duration and dentoskeletal changes

Australian Dental Journal

Orthodontic miniscrews are becoming widely

The LOMAS System. Eric J. W. Liou, James C. Y. Lin

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws

Surgical technique. IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

Influence of the thickness of cortical bone on the stability of orthodontic miniscrews

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws

Lever-arm and Mini-implant System for Anterior Torque Control during Retraction in Lingual Orthodontic Treatment

Tomographic mapping of the hard palate and overlying mucosa

Management of three impacted teeth detected early

In biocreative therapy (C-therapy), torque control

Controlled Space Closure with a Statically Determinate Retraction System

Objective: The purpose of this study was to evaluate the reliability and accuracy of a

Finite Element Analysis of Stress and Strain Distribution in the Bone around the Implants used for Orthodontic Anchorage

Forty Consecutive Ramus Bone Screws Used to Correct Horizontally Impacted Mandibular Molars

SUCCESS RATES OF TEMPORARY ANCHORAGE DEVICES PLACED IN AN ORTHODONTIC CLINIC ALEX C. SCHREIBER

A finite element analysis of the effects of archwire size on orthodontic tooth movement in extraction space closure with miniscrew sliding mechanics

ORIGINAL ARTICLE. Key words: Osseointegration, Mini-implant, Finite element analysis, Von-Mises stress INTRODUCTION

Effective Tooth Movement Using Lingual Segmented Arch Mechanics Combined With Miniscrews

Invisalign technique in the treatment of adults with pre-restorative concerns

Scar formation and revision after the removal of orthodontic miniscrews

SKELETAL ANCHORAGE IN ORTHODONTIC TREATMENT OF A CLASS II MALOCCLUSION

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

Miniscrews were initially proposed as a means

Maxillary sinus perforation by orthodontic anchor screws

Introduction Subjects and methods

Fracture resistance of commonly used self-drilling orthodontic mini-implants

Forsus Class II Correctors as an Effective and Efficient Form of Anchorage in Extraction Cases

Absolute anchorage has been a long sought after,

Open Access Relationship Between the Thickness of Cortical Bone at Maxillary Mid-palatal Area and Facial Height Using CBCT

Maxillary Growth Control with High Pull Headgear- A Case Report

International Journal of Therapeutic Applications ISSN X

Overerupted upper molars due to missing lower

Case Report. Orthognathic Correction of Class II Open Bite. Using the Piezoelectric System and MatrixORTHOGNATHIC Plating System.

Case Report: Long-Term Outcome of Class II Division 1 Malocclusion Treated with Rapid Palatal Expansion and Cervical Traction

ËÕπøíπ â Ÿà â øíπ æ ËÕ øíπàπâ ªî : «ª»πå

Quantitative evaluation of palatal bone thickness for safe mini-implant placement using CBCT

ORTH easy. The Complete System for Cortical Anchorage. OrthoEasy Pins

Enhanced Control in the Transverse Dimension using the Unitek MIA Quad Helix System by Dr. Sven G. Wiezorek

Use of Palatal Miniscrew Anchorage and Lingual Multi-Bracket Appliances to Enhance Efficiency of Molar Scissors-Bite Correction

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews.

Miniscrew-supported coil spring for molar uprighting: Description

Molar intrusion with skeletal anchorage ; from single tooth intrusion to canting correction and skeletal open bite

Pre Prosthetic Adjunct Orthodontic Tooth Movement for Prosthetic Replacement Using an Osseointegrated Implant-A Case Report

Monday Morning Pearls of Practice by Bobby Baig

Mesial Step Class I or Class III Dependent upon extent of step seen clinically and patient s growth pattern Refer for early evaluation (by 8 years)

Comparison of the Failure Rate for Infra- Zygomatic Bone Screws Placed in Movable Mucosa or Attached Gingiva

A Novel Method of Altering the Buccal Segment Relationship

Orthodontic treatment of midline diastema related to abnormal frenum attachment - A case series.

Transcription:

Biological Forum An International Journal 7(2): 33-37(25) ISSN No. (Print): 975-3 ISSN No. (Online): 2249-3239 The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews Abbas Salehi Vaziri*, Hamed Naseri **, Mohsen Nourisari*, Hasanali Qaffari*, Mehdi Kashaniaraqbidi* and Reza Eslamiamirabadi* *Department of Orthodontics, School of Dentistry, University of Shahed, Tehran, IRAN. **Undergraduate student of Dentistry Medicine, School of Dentistry, University of Shahed, Tehran, IRAN. (Corresponding author: Hamed Naseri) (Received 2 April, 25, Accepted June, 25) (Published by Research Trend, Website: www.researchtrend.net) ABSTRACT: Researches stated that overheating during drilling, poor primary stability caused by overdrilling, inflammation, or local disturbances could have prevented normal healing in the early period. The theory was further supported by the higher cumulative survival of mini-implant systems in the maxilla. So, the aim of the current study was to investigate the clinical success of self-tapping and self-drilling orthodontic microimplants. Of the 57 patients, 49 were women and 8 were men. Self-tapping and self-drilling microscrew were randomly located in left and right side of maxillary of patients, respectively. All patients were under local anesthesia for insertion of the mini-implants. Then patients asked to impress their feeling for pain based on pain scale from to based on numeric rating NRS). In the second referred, the inflammation, hyperplasia and hemorrhage monitored in all patients. The mobility was checked from the second referred and continued for 6 months after placement. Force application was done using 2 g for 6 months. According to de results, of both methods was 93% and only 7% of patients faced the mobility of miniscrews. The inflammation was observed in 8.8% of miniscrews. The Chi-square test results revealed there was significant correlation between mobility and inflammation P =.22. 2.47 (based on NRS) but there was no significant difference among two groups. These results suggest there is no difference for self-tapping and self-drilling orthodontic miniscrews. Keywords: Self-tapping, Self-drilling, Orthodontic microimplants INTRODUCTION Orthodontic anchorage is defined as resistance to unwanted movement of the tooth. In the anteroposterior dimension, 3 anchorage situations are traditionally defined by the ratio of retraction to molar protraction. While moderate anchorage entails reciprocal space closure, maximum anchorage means that most of the space is closed by retraction of the incisors, and minimum anchorage means that most of the space is closed by protraction of the posterior segments. Absolute anchorage, when the anchorage parts keep on fully immobile, is occasionally desirable but is usually unreachable with traditional orthodontic mechanics. The exception is the incidence of ankylosed teeth in the anchorage part. Beneath these exceptional circumstances, forces applied to those teeth are entirely transferred to the nearby skeletal structures. This condition is occasionally named skeletal anchorage and, by the above definition could also be called absolute anchorage (Baumgaertel et al. 28). However, undesirable tooth movement, leads to anchorage loss, is a chief pitfall of these anchorage strengthening techniques. Within a decade, several kinds of noncompliance skeletal anchorage arrangements such as implants, onplants, mini plates and mini- or microscrews (Kim et al. 25) have gained in acceptance between clinicians, as a means to get absolute orthodontic Miniscrews, which were first introduced by Kanomi (997) to avoid undesirable tooth movement, can be easily placed into various locations in alveolar bone due to their small dimensions. The insertion procedure does not involve extensive trauma, and they can bear load immediately after placement. Moreover, they are easy to remove and involve relatively lower treatment costs compared with implants and onplants (Park et al Nevertheless, as a disadvantage, miniscrews could be remove simply with low elimination torques in comparison to implants because of their slight diameters and short lengths (Costa et al. 998).

Vaziri, Naseri, Nourisari,Qaffari, Kashaniaraqbidi and Eslamiamirabadi 34 Because of the challenge of controlling reciprocal tooth movement in noncompliant patients, periodontically compromised patients, and those with a reduced number of teeth undergoing orthodontic treatment (Block and Hoffman, 995; Baumgaertel, 24) anchorage systems have been described in the literature over the past decades (Melsen et al. 998; da Cunha et al. 25). The wide acceptance orthodontic miniimplant as a temporary anchorage device can especially be attributed to minimal anatomic limitations for placement, thus broadening the clinical applications. Obtaining an efficient interface between mini-implant and bone tissue & continues to be the key point to achieving higher success rates (Papadopoulos and Tarawneh, 27; Coletti et al. 27). The recommended position is between the first molar and second premolar level of the attached and unattached gingiva. In this position avoidance of injury to the roots cannot be guaranteed (Roccia et al. 25; Gibbons al. 27; Dao et al. 29). Researches stated that overheating during drilling, poor primary stability caused by overdrilling, inflammation, or disturbances could have prevented normal healing in the early period. The theory was further supported by the higher cumulative survival of mini implant systems in the maxilla than in the mandible (Park, 23). Because bone density is high in the mandible, implants can have high torque and good initial stability. Overheating of the pilot drill causing bone damage, might contribute to the high failure rate, so copious irrigation with saline solution was needed. The self-drilling method, a new technique, was used in recent studies (Xun et al. 27). Its placement procedure is simplified, without pilot drilling and incision. Even though success rates were diverse, it was believed that failure rates might be further reduced with increasing clinical experience and perfecting of the placement technique. To date, few clinical studies have assessed miniscrew success rates, the predictability of placement techniques, or the management of risk factors for failure. So, the aim of the current study was to investigate the clinical success of self-tapping and self-drilling orthodontic miniscrew on subsequent inflammation and pain in patients. MATERIAL AND METHODS The study sample consisted as in vivo study. The miniimplants were designed originally for the purpose of achieving anchorage in orthodontic tooth movement. Of the 57 patients, 49 were women and 8 were men. The minimum age for the patients was 5 years. Then the patients were treated for Cl II malocclusion and retraction of maxillary teeth. Special attention was required during mini implant placement to decrease the incidence of injury to delicate anatomic structures like vessels, nerves and dental roots. All patients were under local anesthesia for insertion of the mini-implants Selftapping and self-drilling miniscrew were randomly with respect of priority located in left and right side of maxillary, self-drilling for the right side and selftapping for the left side. Screws purchased from Jeil Medical Corporation, Seoul, South Korea (the image is presented below). The coordination of the screws was:.4 mm in diameter and 8 mm height, titanium alloy. Screws were placed between first molar and second placed at 9ø to buccal surface of alveolar of the maxillary bone (Park et al. 26). All miniscrews placed by an expert orthodontist, to subsequent experimental error. A pilot hole was drilled with physiologic serum on that part which self-tapping screws intentioned Park et al. 26). Then patients asked to impress their feeling for pain based on pain scale from to based on numeric (Breivik et al. 28). In the second referred, the inflammation, hyperplasia and hemorrhage monitored in all patients. The mobility was checked in the second referred and continued for 6 months after placement. Force application was done after 2weeks to month after application, using 2 g for 6 months. Statistical analysis: Data obtained from pain score or inflammation proceed in excel then analyzed by oneway analysis of variance (ANOVA) using SPSS 6. for Windows. For treatment showing a main effect by ANOVA, Chi-square test and t-test were used. P<.5 was considered as significant differences between treatments. RESULTS The effects of self-tapping and self-drilling orthodontic microimplants on inflammation and pain on patients are presented in figs. -5. Effect of self-tapping and self-drilling orthodontic miniscrew s on pain frequency is shown in Fig.. According to the data, the observed pain score was between -7 based on NRS ( stands for less and 7 for high) but this score is from - and in this study there was no report for scores 8, 9 and. As seen, the frequency for score was much more than the others. Also, the score 3 was reported by 3 miniscrews. According to the results, the in the inflammation was only in 8.8% of patients (green section ). So, it seems the self-tapping and self-drilling orthodontic miniscrew had scarce effect on inflammation accuracy.

Vaziri, Naseri, Nourisari,Qaffari, Kashaniaraqbidi and Eslamiamirabadi 35 4 3 2 PAIN As seen in Fig. 3, the mobility was detected in only. percent of self-tapping and self-drilling orthodontic miniscrews -treated patients. Fig. 4 reveals effect of age on success of self-tapping and self-drilling orthodontic miniscrews. The mean of age was 25.7. AGE Frequency 2 3 4 5 6 7 4 3 PAIN Fig.. Effect of self-tapping and self-drilling orthodontic microimplants on pain frequency. 2 INFLAMMA Frequency 7.5 2. 22.5 25. 27.5 3. 32.5 35. Std. Dev = 4.45 Mean = 25.7 N = 4. AGE Fig. 4. Effect of age on success of self-tapping and self-drilling orthodontic microimplants. Fig. 2. Effect of self-tapping and self-drilling orthodontic miniscrews on inflammation. : inflammation, : without inflammation. MOBILITY As seen in Fig. 5, there was no significant difference on right or left self-tapping and self-drilling on inflammation incidence. According to the Chi-square test the difference was not significant p =.7. Also, Chi-square test revealed there was a significant difference on pain between both sexes P =.3. Also, a significant correlation observed between mobility and inflammation P =.22. 6 5 4 3 2 Count INFLAMMA Fig. 3. Effect of self-tapping and self-drilling orthodontic miniscrews on mobility. : Mobility, : without mobility. L_OR_R Fig. 5. Effect of right or left self-tapping and selfdrilling on inflammation. : self-drilling, : selftapping.

DISCUSSION Vaziri, Naseri, Nourisari,Qaffari, Kashaniaraqbidi and Eslamiamirabadi 36 Anchorage control is the most important factor in successful orthodontic treatment. Various techniques have been devised and used in orthodontic practice to strengthen anchorage. Traditional techniques, like use of a multi-bracket appliance, extra oral anchorage by headgear and functional therapies, cannot effectively control anchorage, especially in adult patients. The reinforcement of anchorage requires complicated biomechanics and good patient compliance. Application of the mini-implant as alternative anchorage for various types of tooth movement has been demonstrated (Tseng et al. 26). Maino et al. (23) u used implants to provide anchorage when retracting the maxillary. Kyung et al. (23) applied mini-implant anchorage to stretch inferior second molars to first molar extraction sites. Carano et al. (25) used the mini-implants to provide anchorage during incisor intrusion. Lee et al. (24) that midpalatal mini-implants could be effectively used for intrusion of maxillary molars. The principle gateway of stability for mini-implants is a mechanical lock within the bone. Poor quality or an insufficient quantity of available bone may cause lack of retention of the mini-implant. Oral hygiene could also influence the success rate of mini-implants. According to the results, it seems the self-tapping and self-drilling orthodontic miniscrews had scarce effect on inflammation accuracy when hygiene quality is optimum. Also, mobility was detected in only. percent of tapping and self-drilling orthodontic miniscrews -treated patients. Additionally, there was no significant difference on self-tapping and self-drilling groups on inflammation incidence. According to the data, the observed pain score was between -7 based on NRS. A commonly used clinical measure of pain is the NRS. Patients are asked to indicate the intensity of pain by reporting a number that best represents it. The NRS is easy to administer verbally in a clinical setting and is a familiar clinical tool. Also, the visual analog scale (VAS) has been used extensively in clinical research. An advantage of the VAS is that pain is measured continuously. The reliability and validity of the VAS in the ED setting have been demonstrated. The lowest clinically significant difference in pain can be identified by patients has been recognized and validated in several diverse data sets (Bijur et al. 23). Screw implants fails for numerous causes. The reasons of dental implant failure are host aspects (osteoporosis and diabetes and smoking), surgical factors of improper surgical method. These factors have adverse effect on failure of screw implants, though, should be elucidated in a forthcoming study (Ashley et al. 23). Surgical factors include improper surgical techniques like lack of initial stability, overheating during placement, and the fitness of the pilot hole to the diameter of the screw implant. In current study, all screw implants were placed by same dentist using similar technique the surgical factors had no effect on the clinical success. Nevertheless, using this procedure, dentists may have acceptable success in practice. Management causes can count as poor care, inflammation, oral hygiene and even excessive load. In a previous study 6 of 2 failed screw implants failed within 2 months post placement. The other 6 screw implants failed until months, and the cause might be might indicate surgical and management actions are crucial for screw implant success (Park, 23). Current study is a novel field our knowledge is less about the reasons that affect the success of screw implants. Screw implant features, host related issues such as recipient sites, procedure and environmental management were evaluated. Amongst them, local host and management factors are the most important ones (Park et al. 27). ACKNOWLEDGMENTS This paper was carried out from the thesis in doctoral in dentistry medicine of Dr. Hamed Naseri under supervision of Dr. Abbas Salehi vaziri in University of Shahed, Tehran, Iran. REFERENCES Ashley ET, Covington LL, Bishop BG, Breault LG. (23). Ailing and failing endosseous dental implants: a literature review. J. Contemp. Dent. Pract. 4: 35-5. Baumgaertel S, Razavi MR, Hans MG, (28). Mini-implant anchorage for the orthodontic practitioner. Am. J. Orthod. Dentofacial. Orthop. 33: 62-7. Baumgaertel S, (24). Te mporary skeletal anchorage devices: the case for miniscrews. Am. J. Orthod. Dentofacial. Orthop. 45: 558-64. Bijur PE, Latimer CT, Gallagher EJ, (23). Validation of a verbally administered numerical rating scale of acute pain for use in the emergency department. Academic Emer. Med. : 39-392. Block MS, Hoffman DR, (995). A new device for absolute anchorage for orthodontics. Am. J. Orthod. Dentofacial. Orthop. : 25-8. Breivik H, Borchgrevink PC, Allen SM, Rosseland LA, Romundstad L, Breivik EK, Kvarstein G, Stubhaug A, (28). Assessment of pain. Br. J. Anaesthesia, (): 7-24.

Vaziri, Naseri, Nourisari,Qaffari, Kashaniaraqbidi and Eslamiamirabadi 37 Carano A, Velo S, Leone P, Siciliani G, (25). Clinical applications of the Miniscrew Anchorage System. J. Clin. Orthod. 39: 9-24. Coletti D, Salama A, Caccamese J, (27). Application of intermaxillary fixation screws in maxillofacial trauma. J. Oral Maxillofac. Surg. 65: 746. Costa A, Raffainl M, Melsen B, (998). Miniscrews as orthodontic anchorage: a preliminary report. Int. J. Adult Orthodon. Orthognath. Surg. 3: 2-9. Da Cunha AC, Marquezan M, Lima I, Lopes RT, Nojima LI, Sant'Anna EF, (25). Influence of bone architecture on the primary stability of different mini-implant designs. Am. J. Orthod. Dentofacial. Orthop. 47: 45-5. Dao V, Renjen R, Prasad HS, Rohrer MD, Maganzini AL, Kraut RA, (29). Cementum, Pulp, Periodontal Ligament, and Bone Response After Direct Injury With Orthodontic Anchorage Screws: A Histomorphologic Study in an Animal Model. J. Oral Maxillofac. Surg. 67: 244-2445. Gibbons AJ, Hodder SC, (27). A self -drilling intermaxillary screw. Br. J. Oral Maxillofac. Surg. 45: 46. Kanomi R, (997). Mini -implant for orthodontic anchorage. J. Clin. Orthodonti. 3():763-767. Kim JW, Ahn SJ, Chang YI, (25). Histomorphometric and mechanical analyses of the drill-free screw as orthodontic anchorage. Am. J. Orthod. Dentofacial. Orthop. 28(2): 9-94. Kyung SH, Choi JH, Park YC, (23). Miniscrew anchorage used to protract lower second molars into first molar extraction sites. J. Clin. Orthod. 37: 575-579. Lee JS, Kim DH, Park YC, Kyung SH, Kim TK. (24). The efficient use of midpalatal miniscrew implants. Angle Orthod. 74: 7-74. Maino BG, Bednar J, Pagin P, Mura P, (23). The spider screw for skeletal anchorage. J. Clin. Orthod. 37: 9-97. Melsen B, Petersen JK, Costa A, (998). Zygoma ligatures: an alternative form of maxillary anchorage. J. Clin. Orthod. 32: 54-8. Papadopoulos MA, Tarawneh F, (27). The use of miniscrew implants for temporary skeletal anchorage in orthodontics: A comprehensive review. Oral Surg. Oral Med. Oral Path. Oral Radiol. Endod. 3: 6-5. Park HS, Bae SM, Kyung HM, Sung JH, (2). Microimplant anchorage for treatment of skeletal Class I bialveolar protrusion. J. Clin. Orthodonti. 35(7): 47-422. Park HS, (23). Clinical study on success rate of microscrew implantsb for orthodontic anchorage. Korean J. Orthod. 33: 5-6. Park HS, Jeong SH, Kwon OW, (26). Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am. J. Orthod. Dentofacial. Orthop. 3: 8-25. Roccia F, Tavolaccini A, Dell'Acqua A, et al, (25). An audit of mandibular fractures treated by intermaxillary fixation using intraoral cortical bone screws. J. Craniomaxillofacial. Surg. 33: 25. Tseng YC, Hsieh CH, Chen CH, Shen YS, Huang IY, Chen CM, (26). The application of mini -implants for orthodontic anchorage. Int. J. Oral Maxillofac. Surg. 35: 74-77. Xun C, Zeng X, Wang X. (27). Microscrew anchorage in skeletal anterior open-bite treatment. Angle Orthod. 77: 47-56.