Assessing Arrhythmia Performance ST/AR Algorithm

Similar documents
GE Healthcare. The GE EK-Pro Arrhythmia Detection Algorithm for Patient Monitoring

Helping you save more lives. Philips HeartStart Resuscitation Solutions Leading the way in quality resuscitation

MARS Ambulatory ECG Analysis The power to assess and predict

Pathfinder Holter Analyzer CARDIAC DIAGNOSTIC SOLUTIONS

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

The first MR with IQ. Philips Intera 1.5T With SmartExam

EP lab robotic support

Vital Responder: Real-time Health Monitoring of First- Responders

Field Strength. Regional Perfusion Imaging (RPI) matches cerebral arteries to flow territories

USING CORRELATION COEFFICIENT IN ECG WAVEFORM FOR ARRHYTHMIA DETECTION

Keeping pace with imaging innovation

Pocket Guide. System One REMstar Pro and REMstar Auto

Robust Detection of Atrial Fibrillation for a Long Term Telemonitoring System

High performance ultrasound

Speed - Accuracy - Exploration. Pathfinder SL

Assessment of ECG frequency and morphology parameters for automatic classification of life-threatening cardiac arrhythmias

Emergency Care Clinical Data Transmission Networks

HST-582J/6.555J/16.456J-Biomedical Signal and Image Processing-Spring Laboratory Project 1 The Electrocardiogram

Versatile, affordable ultrasound. Philips HD6 ultrasound system. The print quality of this copy is not an accurate representation of the original.

Natural performance. Introducing the BiPAP A30 - because ease of use and therapy efficacy are key to patient well-being

Assessment of the Performance of the Adaptive Thresholding Algorithm for QRS Detection with the Use of AHA Database

Welch Allyn HScribe Holter Analysis System

Panorama. Arrhythmia Analysis Frequently Asked Questions

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

CHAPTER IV PREPROCESSING & FEATURE EXTRACTION IN ECG SIGNALS

Feasibility evaluation of dual axis rotational angiography (XperSwing) in the diagnosis of coronary artery disease

CHAPTER 5 WAVELET BASED DETECTION OF VENTRICULAR ARRHYTHMIAS WITH NEURAL NETWORK CLASSIFIER

Genetic Algorithm based Feature Extraction for ECG Signal Classification using Neural Network

False arrhythmia alarm reduction in the intensive care unit

Arrhythmia Monitoring ST/AR Algorithm

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Rate: The atrial and ventricular rates are equal; heart rate is greater than 100 bpm (usually between bpm).

Wavelet Decomposition for Detection and Classification of Critical ECG Arrhythmias

Basic Dysrhythmia Interpretation

UNDERSTANDING YOUR ECG: A REVIEW

Ordinary person, extraordinary moment. Philips HeartStart OnSite Defibrillator

Mobile radiography realizes its full potential

diseases is highly important. Electrocardiogram (ECG) is an effective tool for diagnosis of the

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

The intelligent solution. System One Sleep Therapy System uses advanced intelligence to deliver excellent care while making patient management easy

Early defibrillation programs for the workplace. Give your workplace the power to save a life

ECG Beat Recognition using Principal Components Analysis and Artificial Neural Network

Changing your idea of clinical performance. Philips HD15 PureWave ultrasound system for echocardiography

ECG Signal Characterization and Correlation To Heart Abnormalities

Delineation of QRS-complex, P and T-wave in 12-lead ECG

ECG UPDATE COURSE DESCRIPTION

DIFFERENCE-BASED PARAMETER SET FOR LOCAL HEARTBEAT CLASSIFICATION: RANKING OF THE PARAMETERS

Meeting the big challenge of little faces. Philips Respironics pediatric patient interfaces for noninvasive ventilation

International Journal of Advanced Research in Computer Science and Software Engineering

ECG Signal Analysis for Abnormality Detection in the Heart beat

Developing Electrocardiogram Mathematical Model for Cardiovascular Pathological Conditions and Cardiac Arrhythmia

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

Learning Decision Tree for Selecting QRS Detectors for Cardiac Monitoring

A RECOGNITION OF ECG ARRHYTHMIAS USING ARTIFICIAL NEURAL NETWORKS

HST.582J / 6.555J / J Biomedical Signal and Image Processing Spring 2007

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Diagnostic capabilities of the implantable therapeutic systems

A MULTI-STAGE NEURAL NETWORK CLASSIFIER FOR ECG EVENTS

Waterproof quality IP X4

Rigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

High performance compact ultrasound

Equivalence of Activity Recordings and Derived Sleep Statistics

EKG Rhythm Interpretation Exam

One intelligent solution

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Electrocardiography for Healthcare Professionals

NOISE DETECTION ALGORITHM

3Replacement Transducer Cables

Non-Invasive Transcutaneous Pacing

Sequential algorithm for life threatening cardiac pathologies detection based on mean signal strength and EMD functions

Lab Activity 24 EKG. Portland Community College BI 232

Why to monitor AF ablation success?

Journal of Arrhythmia

EKG Abnormalities. Adapted from:

TELEMETRY BASICS FOR NURSING STUDENTS

One intelligent solution

BTL CardioPoint Relief & Waterfall. Relief & Waterfall. Abnormalities at first sight

High performance compact ultrasound

MRI to fit your planning. Philips Panorama HFO Oncology Configuration

ID: BFH_ Start time: :00:00. Examination summary

ProSim 3 and 2. Vital Signs Simulators

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

SEARCHING FOR NON-SENSE:

International Journal of Advance Engineering and Research Development

Somnolyzer 24x7. Addressing cost pressures in the sleep lab

Arrhythmia Care in the DGH What Still Needs to be Done? Dr. Sundeep Puri Consultant Cardiologist

ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

Keywords : Neural Pattern Recognition Tool (nprtool), Electrocardiogram (ECG), MIT-BIH database,. Atrial Fibrillation, Malignant Ventricular

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Identification of Arrhythmia Classes Using Machine-Learning Techniques

Biphasic Clinical Summaries

Premature Ventricular Contraction Arrhythmia Detection Using Wavelet Coefficients

The Function of an ECG in Diagnosing Heart Conditions. A useful guide to the function of the heart s electrical system for patients receiving an ECG

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms

Supraventricular Tachycardia (SVT)

The Cross-platform Application for Arrhythmia Detection

Xcelera version comparison of features and functionality

Transcription:

Assessing Arrhythmia Performance ST/AR Algorithm Application Note This report provides the arrhythmia performance of the ST/AR (ST and Arrhythmia) algorithm. For a description of the algorithm, see the Arrhythmia Monitoring ST/AR Algorithm Application Note 453564115631.. Introduction The performance of the algorithm used to detect and classify arrhythmias is fundamental to the effectiveness of computerized arrhythmia monitoring. It is important that the algorithm enable the system to alert the clinical staff to true arrhythmia events without unnecessarily distracting them from their duties. Such an algorithm will allow the monitoring system to generate accurate cumulative data that will be useful in supporting therapeutic decisions.. Procedures exist to facilitate the unbiased comparison of competing algorithms against identical data. A group of industry and user representatives under the auspices of the Association for the Advancement of Medical Instrumentation (AAMI) have published their recommendations for algorithm testing and reporting - Recommended Practice Document ANSI/AAMI EC57. N N V N V This application note summarizes the results of the PVC Performance Validation Test, Atrial Fibrillation detection test and the Ventricular Fibrillation Detection Test.

Reference Databases for Ventricular Arrhythmia Testing Databases are available to the public for the testing of arrhythmia detection algorithms. These databases consist of records of patient ECG waveforms, together with a set of annotation files in which each beat has been labeled by expert cardiologists. Two of the databases that have been used to assess the performance of the ST/AR arrhythmia algorithm are as follows: The American Heart Association (AHA) database, distributed by the Emergency Care Research Institute (ECRI), consists of data from 80 patients, two of whom are paced. The Massachusetts Institute of Technology and Beth Israel Hospital (MIT-BIH) database consists of data from 48 patients, four of whom are paced. The AHA and MIT-BIH databases were developed specifically for the purpose of evaluating ventricular arrhythmia detection algorithms. Both of these databases have two simultaneous channels of ECG. Even though both the AHA and MIT-BIH databases contain some paced patients, data from paced patients (in digital form) do not retain sufficient pace pulse signal quality to allow for reliable detection of pace pulses. Therefore, the current recommended practice document recommends that the paced patients not be included in the evaluation. QRS and PVC Performance Validation Test The QRS and PVC Performance Validation Test quantifies the performance of the arrhythmia algorithm in QRS detection and PVC classification. Performance Measures The performance measures quantify the extent to which the algorithm results agree with the expert annotation of a given patient record. The annotation or labeling of each beat in the reference database by clinical experts is called the true classification of that beat. QRS Detection Sensitivity is the percentage of true QRSs that the algorithm classifies as QRSs. It is calculated as the number of QRSs the algorithm correctly classifies, divided by the number of true QRSs, and A consistently high value for this performance measure indicates that the system is able to detect a high percentage of the QRS complexes. QRS Detection Positive Predictivity is the percentage of beats that the algorithm classifies as QRSs that are true QRSs. It is calculated as the number of QRSs the algorithm correctly classifies, divided by the total number of beats that the algorithm classifies as QRSs, and A consistently high value for this performance measure indicates that the beats designated as QRSs by the algorithm are truly QRSs a high percentage of the time. PVC Detection Sensitivity is the percentage of true PVCs that the algorithm classifies as PVCs. It is calculated as the number of PVCs the algorithm correctly classifies, divided by the number of true PVCs, and If the algorithm detects the PVCs of a majority of patients with a high level of sensitivity, one can expect the system to detect PVCs with a high degree of reliability. A sensitivity of 100%, that is, correctly identifying all PVCs, is the ultimate goal. Actual performance will tend to fall short of that goal. PVC Detection Positive Predictivity is the percentage of beats that the algorithm classifies as PVCs that are true PVCs. It is calculated as the number of PVCs the algorithm correctly classifies, divided by the total number of beats that the algorithm classifies as PVCs, and A consistently high value for this performance measure indicates that the beats designated as PVCs by the algorithm are truly PVCs a high percentage of the time. PVC Detection False Positive Rate is the percentage of beats that the algorithm incorrectly labels as PVCs relative to all non-pvc beats. It is calculated as the number of non-pvcs the algorithm incorrectly classifies as PVCs, divided by the number of non-pvc beats, and then converted to a percentage. This measure is an indication of the most visible annoyance to the clinical staff. If the false positive rate is high, the arrhythmia monitoring system tends to generate a large number of false alarms. If a large percentage of patients have a low false positive rate, the system maintains a relatively high level of credibility. It is important that an arrhythmia algorithm have: High QRS Detection Sensitivity High QRS Detection Positive Predictivity High PVC Detection Sensitivity High PVC Detection Positive Predictivity A low PVC Detection False Positive Rate These five parameters together show how well an algorithm correctly detects QRS complexes and PVCs. AAMI recommends that when comparing performance results and evaluating arrhythmia algorithm performance, it is important that all patient records in the database (except those of paced patients) be included. 2

Performance Results The following results pertain to the ST/AR arrhythmia algorithm. ST/ AR can perform multi-lead or single-lead analysis. This setting depends on your unit s default setting or can be changed on an individual basis. In certain cases, such as the M3/M4 Monitor only single-lead analysis is available. In the following document both multi-lead and single-lead performance will be described. Multi-Lead Results Summary The following table summarizes the performance results of the multilead analysis.. Table 1: Multi-Lead Performance Summary Average Performance Measures Single-Lead Results Summary AHA MIT-BIH QRS detection sensitivity (%) 99.80 99.66 QRS Detection Positive 99.87 99.86 PVC Detection Sensitivity (%) 95.96 94.25 PVC Detection Positive PVC Detection False Positive Rate (%) 98.34 96.38 0.16 0.27 The following table summarizes the performance results of the QRS and PVC performance validation test when using the best single lead for analysis. Table 2: Single-Lead Performance Summary Average Performance Measure AHA MIT-BIH QRS detection sensitivity (%) 99.65 99.12 QRS Detection Positive 99.88 99.90 PVC Detection Sensitivity (%) 94.20 94.75 PVC Detection Positive PVC Detection False Positive Rate (%) 98.56 96.39 0.14 0.27 These results show that the algorithm performs well on both of the reference databases whether using single-lead or multi-lead analysis. The MIT-BIH database presents more difficult situations in which to detect PVCs because the second channel of this database is very noisy; there is a low signal-to-noise ratio and, in many cases, low signal amplitude. The high values for PVC Detection Sensitivity indicate that the algorithm detects PVCs with a high degree of reliability. The consistently high values for PVC Detection Positive Predictivity indicate that the beats designated as PVCs by the algorithm are truly PVCs a high percentage of the time. The PVC Detection False Positive Rates of less than 0.3% indicate that for every 1000 beats, the algorithm incorrectly classifies less than 3 PVCs. Ventricular Fibrillation Detection Test The Ventricular Fibrillation Detection Test quantifies the performance of the arrhythmia algorithm in detection of ventricular fibrillation (VF). Performance Measures The following three performance measures together describe algorithm performance in ventricular fibrillation detection: Relative VF Detection Time is measured as the length of time from the beginning of a VF episode to the time that the arrhythmia system generates a Ventricular Fibrillation/Ventricular Tachycardia red alarm. In some instances, a negative alarm time results when the arrhythmia algorithm detects alarms before the annotated episode beginning time. In these cases, the alarms are triggered by the ventricular tachycardia which immediately preceded the onset of fibrillation or flutter. These are considered to be early detections of life-threatening episodes, not false-positive errors. VF Detection Sensitivity is the percentage of true ventricular fibrillation episodes that the algorithm detects as ventricular fibrillation. VF False Positive Detection is the number of times that the algorithm incorrectly detects ventricular fibrillation. Results Summary The following two histograms summarizes the VF detection time results of the VF detection test for multi-lead and for single-lead analysis. There are a total of 15 VF episodes in the AHA and MIT-BIH databases. With single-lead analysis each channel is treated seperately resulting in 30 VF episodes. The horizontal axis in the histogram represents a range of relative VF detection times from 45 seconds before the onset of VF to 25 seconds after the onset of the episode. The vertical axis represents the number of VF episodes. The solid bars 3

indicate the number of VF episodes for which the relative VF detection time was in the range spanned by that bar. arrhythmia algorithm will achieve 100% detection accuracy in all clinical episodes. However, the ST/AR arrhythmia algorithm has demonstrated 100% sensitivity in generating red alarms during lifethreatening episodes of ventricular fibrillation using the AHA and MIT- BIH databases. Atrial Fibrillation Detection Test The atrial fibrillation detection test quantifies the performance of the arrhythmia algorithm in detecting atrial fibrillation (AF). Figure 1 Multi-Lead Algorithm Relative VF Detection Time AHA and MIT-BIH Databases Reference Database For performance evaluation, several databases, which were not used during the algorithm development, were used for the testing. The dataset consists of a total of 234 cardiologists annotated adult patientrecords with 75 records contain AF episodes. The dataset includes three publicly available databases, including the MIT AF database with 23 patient-records, the AHA testing database with 75 patient-records, and the long-term ST database with 86 patient-records. The dataset also includes four proprietary databases collected from 4 different hospitals. These proprietary databases are labeled as Pafdb A, B, C, and D. The number of AF, AFL, and AF/FL episodes and their total duration are summarized in Table 3 below for all the individual databases. Figure 2 Single-lead Algorithm Relative VF Detection Time AHA and MIT-BIH Database In terms of VF detection sensitivity and the AHA and MIT-BIH databases, a Ventricular Fibrillation or Ventricular Tachycardia red alarm is always generated by the arrhythmia system immediately preceding or within a few seconds of the VF start time. In terms of VF false positive detection, the algorithm detected only two false VF episodes across both databases. Both of these alarms occurred on ventricular tachycardia episodes with high heart rate. These are considered an appropriate announcement of lifethreatening events, not false-positive errors. Ventricular fibrillation and flutter is a chaotic and unpredictable event. There is no existing database that can be used to ensure that an 4

Table 3: Summary of AF Testing Database (not used in development) Database Name Total Number of Records Records with AF Epi AFL Epi AF/FL Epi Non AF AF AFL AF/FL # D(sec) # D(sec) # D(sec) D(sec) afdb 23 23 8 0 291 336232 14 5877 0 0 501290 ahatest 75 4 0 1 4 41665 0 0 1 10800 756000 ltstdb 86 2 0 0 2 89730 0 0 0 0 2147484 Pafdb A 10 9 0 1 24 37490 0 0 1 7200 6599 Pafdb B 5 5 0 0 5 34638 0 0 0 0 0 Pafdb C 1 1 0 0 1 60008 0 0 0 0 0 Pafdb D 34 31 0 2 31 190120 0 0 10 14294 4096 Total 234 75 8 4 358 789883 14 5877 12 32294 3415469 Total Hours 219.41 1.63 8.97 948.74 Abbreviations: AF = Atrial Fibrillation, AFL = Atrial Flutter, AF/FL = Combination of AF and AFL, NonAF = Rhythm is not AF or AFL, Epi = Episode, D(sec) = Duration of episode in seconds Performance Measures The performance measures quantify the extent to which the algorithm results agree with the cardiologists' annotated AF episodes. The performance measures used to measure AF detection accuracy include the following four measures as described in the Recommended Practice Document ANSI/AAMI EC57: for reporting AF detection performance. AF Duration Detection Sensitivity is calculated as the duration of overlapping between the algorithm-marked episode and the cardiologist-annotated episode divided by the duration of the cardiologist-annotated episode, and This performance measure indicates how well the algorithm is able to marked the true AF episode. AF Duration Detection Positive Predictivity is calculated as the duration of overlapping between the algorithm-marked episode and the cardiologist-annotated episode divided by the duration of the algorithm-marked episode, and This performance measure indicates what percentage of the algorithmmarked episode duration is actually correct. the number of cardiologist-annotated episodes, and converted to a percentage. AF Episode Detection Positive Predictivity is the percentage of the AF episodes that the algorithm detected as AF that are true AF. It is calculated as the number of AF episodes the algorithm correctly detected divided by the total number of AF episodes that the algorithm detected as AF, and Results Summary For the final performance calculation, episodes less than 1 minute in duration and all atrial flutter episodes are excluded. The following table summarizes the performance results of the AF detection algorithm using the performance measures described above. These performance evaluation results show that the AF detection algorithm has achieved a high level of accuracy in detecting AF episode over a large patient population. AF Episode Detection Sensitivity is the percentage of true AF episodes that the algorithm detected as AF. It is calculated as the number of AF episodes the algorithm correctly detected divided by 5

Table 4: AF Detection Performance Summary Database Record 234 Records (All AF testing database) 232 Records (Excluding 2 sinus arrhythmia records) 23 records (MIT AF Database) AF Duration Detection AF Episode Detection Sensitivity Positive Predictivity Sensitivity Positive Predictivity 94% 97% 95% 86% 94% 98% 95% 93% 94% 98% 92% 97% Conclusion The testing procedures recommended by AAMI using the reference databases provide a way to compare the performance of the arrhythmia monitoring algorithms. From the results presented in this application note, the excellent performance of the ST/AR arrhythmia algorithm is evident. This ability is basic to the effective monitoring of arrhythmia events in the clinical setting. 6

Philips Healthcare is part of Royal Philips Electronics Interested? Would you like to know more about our imaginative products? Please do not hesitate to contact us. We would be glad to hear from you. On the web www.philips.com/healthcare Via e-mail healthcare@philips.com By fax +31 40 27 64 887 By mail Philips Healthcare Global Information Center P.O Box 1286 5602 BG Eindhoven The Netherlands By phone Asia Tel: +852 2821 5888 Europe, Middle East, Africa Tel: +49 7031 463 2254 Latin America Tel: +55 11 2125 0764 North America Tel: +1 425 487 7000 800 285 5585 (toll free, US only) 2008 Koninklijke Philips Electronics N.V. All rights are reserved. Philips Medical Systems Nederland B.V. reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Printed in The Netherlands. 453564115641 Oct 2008