Swine H1N1 Influenza Human Pandemic Strain. genesig Easy Kit 2 Target Gene Kit for use on the genesig q reaction.

Similar documents
Human influenza A virus subtype (H3) genesig Easy Kit for use on the genesig q reaction. Primerdesign Ltd

Human Immunodeficiency Virus type 2. genesig Easy Kit for use on the genesig q reaction. Primerdesign Ltd

Swine H1N1 Influenza Human Pandemic Strain

Avian influenza A virus subtype (H5)

Avian influenza A virus subtype (H7)

Swine H1N1 Influenza Human Pandemic Strain

Avian influenza A virus subtype (H9)

Swine H1N1 Influenza Human Pandemic Strain

Swine H1N1 Influenza Human Pandemic Strain

Avian influenza A virus subtype (H7)

Human influenza A virus subtype (H3)

Avian Influenza A H5N8

Human influenza A virus subtype (H1)

Human Rotavirus A. genesig Standard Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only

Human Rotavirus C. genesig Advanced Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests

Human Rotavirus B. Non structural protein 5 (NSP5) 150 tests. Quantification of Human Rotavirus B genomes Advanced kit handbook HB10.01.

Human Influenza Types A(M1) & B and Respiratory Syncytial Virus. genesig PLEX kit. 100 tests. Primerdesign Ltd

Human Rotavirus A. genesig Advanced Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only

Cladosporium_spp genesig Standard Kit

Rotavirus A. genesig Standard Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests.

Cladosporium_spp genesig Advanced Kit

Human Immunodeficiency Virus Types 1 & 2 and Hepatitis Viruses B and C. genesig PLEX kit. 100 tests. Primerdesign Ltd

Rabies virus. 3 leader and nucleoprotein region. 150 tests. Techne qpcr test. Quantification of Rabies virus genomes Advanced kit handbook HB10.01.

Human Immunodeficiency Virus type 1

Influenza A virus subtype H5N1 (avian influenza)

Detection of Tamiflu resistant Swine H1N1 Influenza Human Pandemic Strain

Influenza A virus subtype H5N1 (avian influenza)

Introduction to Cladosporium_spp

Human Influenza A (Swine Flu) Rapid test

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.

HIV-1 Viral Load Real Time (RG)

Buy The Complete Version of This Book at Booklocker.com:

Rapid-VIDITEST Swine Flu

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay

PUBLIC HEALTH SIGNIFICANCE SEASONAL INFLUENZA AVIAN INFLUENZA SWINE INFLUENZA

Swine Flu, Fiction or Reality

MolecularMD. One-Step qrt-pcr BCR-ABL Kit. Product Description and User Manual. For Quantitative RT-PCR Analysis of BCR-ABL. Contact Us.

Sputum DNA Collection, Preservation and Isolation Kit 50 Individual Devices

Acute respiratory illness This is a disease that typically affects the airways in the nose and throat (the upper respiratory tract).

AVIAN FLU BACKGROUND ABOUT THE CAUSE. 2. Is this a form of SARS? No. SARS is caused by a Coronavirus, not an influenza virus.

Rotavirus Test Kit. Instructions For Use. Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA

Norgen s HIV proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad icycler

ABIOpure TM Viral (version 2.0)

1. Intended Use New Influenza A virus real time RT-PCR Panel is used for the detection of universal influenza A virus, universal swine Influenza A vir

ExiPrep Plus Viral DNA/RNA Kit User s Guide ExiPrep Plus Viral DNA/RNA Kit ExiPrep Plus Viral DNA Kit ExiPrep Plus Viral RNA Kit

Norgen s HIV Proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad T1000 Cycler

AmoyDx TM BRAF V600E Mutation Detection Kit

Bioneer Corporation 8-11,Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax:

ExiPrep TM Viral DNA/ RNA Kit

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

Running head: INFLUENZA VIRUS SEASON PREPAREDNESS AND RESPONSE 1

Kit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive

Product # Kit Components

Influenza: The Threat of a Pandemic

Storage: Logix Smart Zika Virus Master Mix and Logix Smart Zika Virus Positive Control must be stored at -20 ⁰C and can last up to 60 days.

RealLine Mycoplasma genitalium Str-Format

Influenza A H1N1 HA ELISA Pair Set

Rapid-VIDITEST. Influenza A

Cytomegalovirus (CMV) End-Point PCR Kit Product# EP36300

Influenza: The past, the present, the (future) pandemic

AVIAN INFLUENZA COMMODITIES TRAINING GUIDE MODULE 3 - SAMPLING, RAPID TESTING AND PACKAGING PARTICIPANT HANDOUTS

HDV Real-TM. Handbook

DOWNLOAD OR READ : WHEN BIRDS GET FLU AND COWS GO MAD HOW SAFE ARE WE 24 7 SCIENCE BEHIND THE SCENES PDF EBOOK EPUB MOBI

virotype Influenza A RT-PCR Kit Handbook

ab Nuclear Extract Kit

FREQUENTLY ASKED QUESTIONS 1 Avian Influenza in Birds

Single Cell Quantitative Polymer Chain Reaction (sc-qpcr)

SAFETY BULLETIN #3-05 November 11, 2005 Key Facts About Avian Influenza

ab HIV-1 Protease Inhibitor Screening Kit (Fluorometric)

ab HRV 3C Protease Inhibitor Screening Kit (Colorimetric)

Influenza A H1N1 (Swine Flu 2009) Hemagglutinin / HA ELISA Pair Set

RealLine HBV / HCV / HIV Str-Format

Influenza RN.ORG, S.A., RN.ORG, LLC

Real Time Kit. Pneumocystis jirovecii (carinii) Real-TM Handbook. for the detection of Pneumocystis jirovecii (carinii) in biological materials

RealLine HIV quantitative Str-Format

QUICK REFERENCE INSTRUCTIONS For use with Sofia only.

Devon Community Resilience. Influenza Pandemics. Richard Clarke Emergency Preparedness Manager Public Health England South West Centre

Microsart Calibration Reagent

RealLine HCV Qualitative Str-Format

1918 Influenza; Influenza A, H1N1. Basic agent information. Section I- Infectious Agent. Section II- Dissemination

A. No. There are no current reports of avian influenza (bird flu) in birds in the U.S.

DRAFT WGE WGE WGE WGE WGE WGE WGE WGE WGE WGE WGE WGE WGE WGE GETREADYNOWGE GETREADYNOWGE GETREADYNOWGE GETREADYNOWGE.

Avian influenza Avian influenza ("bird flu") and the significance of its transmission to humans

attomol HLA-B*27-Realtime LT 2 Assay for the detection of the human HLA-B*27-locus using LightCycler (Do not use for tissue typing!

Pandemic Influenza A Matter of Time

QUICK REFERENCE INSTRUCTIONS

Poliovirus Real-TM Handbook

Influenza A viruses Detection with real time RT-PCR reagents

Product Contents. 1 Specifications 1 Product Description. 2 Buffer Preparation... 3 Protocol. 3 Ordering Information 4

CMV FEP ALA For nucleic acid amplification of CMV DNA and Fluorescence detection with End Point analysis (FEP) on Aladin (Sacace)

2009 H1N1 Influenza ( Swine Flu ) Hemagglutinin ELISA kit

Avian Influenza (Bird Flu) Fact Sheet

Swine Flu. Background. Interim Recommendations. Infectious Period. Case Definitions for Infection with Swine-origin

INFLUENZA VIRUS. INFLUENZA VIRUS CDC WEBSITE

H1N1 Influenza. Influenza-A Basics. Influenza Basics. April 1, History of Influenza Pandemics. April 1 September 25, 2009

Influenza. What Is Influenza?

SCOTTISH AMBULANCE SERVICE Strategic Co-ordination Centre (SCC) Bulletin 01/ April Swine Flu-Information Sheet

Transcription:

TM Primerdesign Ltd Swine H1N1 Influenza Human Pandemic Strain genesig Easy Kit 2 Target Gene Kit for use on the genesig q16 50 reaction For general laboratory and research use only 1

genesig Easy: at a glance guide For each RNA test Component Volume M1 primer/probe mix 5 µl Your RNA sample 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette Component Volume N1 primer/probe mix 5 µl Your RNA sample 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette For each positive control Component Volume M1 primer/probe mix 5 µl Positive control template 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette Component Volume N1 primer/probe mix 5 µl Positive control template 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette For each negative control Component Volume M1 primer/probe mix 5 µl Water 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette Component Volume N1 primer/probe mix 5 µl Water 5 µl oasig OneStep Master Mix 10 µl Lab-in-a-box pipette 2

Kit Contents M1 specific primer/probe mix (BROWN) Once resuspended the kits should remain at -20ºC until ready to use. N1 specific primer/probe mix (BROWN) Once resuspended the kits should remain at -20ºC until ready to use. Lyophilised oasig TM OneStep Master Mix Lyophilised oasig TM OneStep Master Mix resuspension buffer (BLUE lid) M1 positive control template (RED lid) N1 positive control template (RED lid) Internal extraction control RNA (BLUE lid) RNase/DNase free water (WHITE lid) Template preparation buffer (YELLOW lid) 108 x genesig q16 reaction tubes Reagents and equipment to be supplied by the user genesig q16 instrument genesig Easy DNA/RNA Extraction Kit This kit is designed to work well with all processes that yield high quality RNA but the genesig Easy extraction method is recommended for ease of use. genesig Lab-In-A-Box The genesig Lab-In-A-Box contains all of the pipettes, tips and racks that you will need to use a genesig Easy kit. Alternatively if you already have these components and equipment these can be used instead. 3

Step-by-step guide 1. Resuspend the test components This step must be performed for both primer/probe mixes. Blue pipette (500µl) Use the blue pipette to transfer 500µl* of the oasig OneStep Master Mix resuspension buffer into the tube of lyophilised oasig OneStep Master Mix and mix well by gently swirling. *Transfering 525µl of the oasig OneStep Master Mix resuspension buffer to your oasig OneStep Master Mix (instead of the 500µl recommended above) will enable you to take full advantage of the 50 reactions by accounting for volume losses during pipetting. In order to do so with the genesig Easy fixed volume pipettes use 1x blue, 2x red and 1x grey pipettes to make the total volume. Please be assured that this will not adversely affect the efficiency of the test. Blue pipette (500µl) Then use the blue pipette to transfer 500µl of water into the brown tube labelled M1 or N1 primers/probe. Cap and shake tube to mix. A thorough shake is essential to ensure that all components are resuspended. Failure to mix well can produce poor kit performance. These components are now ready to use. Store them in the freezer from hereon. Top tip Ensure that the primer/probe mix is mixed thoroughly before each use by shaking. Once resuspended do not expose genesig Easy kit to temperatures above -20 C for longer than 30 minutes at a time. 4

2. Internal extraction control Blue pipette (2 x 500µl) Use the blue pipette to transfer 1000µl (2 x 500µl) of template preparation buffer into the Internal Extraction Control RNA tube. Cap and shake tube to mix. Your kit contains Internal Extraction Control RNA. This is added to your biological sample at the beginning of the RNA extraction process. It is extracted along with the RNA from your target of interest. The q16 will detect the presence of this Internal Extraction Control RNA at the same time as your target. This is the ideal way to show that your RNA extraction process has been successful. If using an alternative extraction kit: Use the red pipette to transfer 10µl of Internal Extraction Control RNA to your sample after the lysis buffer has been added then follow the rest of the extraction protocol. If using samples that have already been extracted: Use the grey pipette to transfer 5µl of Internal Extraction Control RNA to your extracted sample. 3. Add primer/probe mix to all reaction tubes Grey pipette (5µl) For each reaction to be run, use the grey pipette to add 5µl of your M1 or N1 primers/probe mix to every tube. Top tip Always pipette the primer/probe mix directly into the bottom of the tube. You can label the tube lids to aid your reaction setup but avoid labelling tube sides. 5

4. Add Master Mix to all reaction tubes Red pipette (10µl) For reaction to be run, use the red pipette to add 10µl of the oasig OneStep Master Mix to the tubes containing primer/probe mix. Move swiftly to begin your q16 run, as any delay after the oasig OneStep Master Mix has been added can effect the sensitivity of your test. Top tip Always add the oasig OneStep Master Mix to the side of the tube to reduce the introduction of bubbles. 5. Negative control Grey pipette (5µl) For each test you will require a negative control. Instead of RNA, water is used. This sample should typically prove negative thus proving that all of your positive samples really are positive. To create a negative control reaction simply use the grey pipette to add 5µl of the water to the required reaction tubes. Close these tubes after adding the water. Because some genesig kit targets are common in the environment you may occasionally see a late signal in the negative control. The q16 software will take this into account accordingly. Top tip Always add water to the side of the tube to reduce the introduction of bubbles. 6

6. Set up a test Grey pipette (5µl) For each sample you wish to analyse, use the grey pipette to add 5µl of your RNA sample to the required reaction tubes. Close these tubes after adding the sample. Always change pipette tips between samples. Top tip Always add the RNA sample to the side of the tube to reduce the introduction of bubbles. 7. Positive control Blue pipette (500µl) Use the blue pipette to transfer 500µl of template preparation buffer into each of the positive control template tubes. Cap and shake tube to mix. Each time you run a test you will require a positive control. This is a small portion of RNA from your target of interest. It serves two purposes: 1. It will always test positive so it shows that everything is working as it should be. 2. The q16 software knows how much RNA is present in the positive control. So it can automatically compare your sample of interest with the positive control to calculate the amount of target RNA in your sample. To create a positive control reaction simply use 5µl of the positive control instead of your RNA sample. 7

Grey pipette (5µl) Take great care when setting up your positive control. The positive control template has the potential to give you a false positive signal in your other samples. Set positive controls up last after all other sample tubes are closed. Always change your pipette tip between samples. You may even choose to set up positive controls in a separate room. Top tip Always add the positive control template to the side of the tube to reduce the introduction of bubbles. 8. Running the test Place the tubes into the correct positions in your q16 as defined by the software, this may include positioning of empty tubes to ensure that the q16 lid is balanced. The run can then be started. Top tip Before loading tubes into the q16, check for bubbles! Flick the bottom of the tubes to remove any bubbles that may have formed during the test setup. Apply centrifugal force with a sharp wrist action to ensure all solution is at the bottom of the reaction tube. When repeating a test you can use a previous file as a template by clicking open then selecting File name > Files of Type > Experiment file as template 8

International Units No international units What do my results mean? Analysis of your data is carried out automatically by the genesig q16. The following information is designed to help you fully understand a result or to troubleshoot: The kit contains two Influenza A primer and probe sets. The M1 primer and probe set are designed to detect all influenza A sequences regardless of the species of origin. Samples that test positive with the M1 primer set contain an influenza A infection of unknown serotype. The N1 primer and probe set is specific to the swine H1N1 pandemic strain. Samples that test positive with the N1 primer contain the pandemic swine H1N1 Influenza. Samples that test positive for M1 and negative for N1 are indicative of a naturally circulating influenza A infection which is not swine flu. Positive Explanation Your sample has produced a positive result. Your target of interest is present and you can use the reported quantity. As this is a two target gene kit, both target genes must be positive to confirm the test as a genuine positive. Negative Explanation Your sample has produced a negative result. The target is not present in your sample. Test contaminated Explanation The Negative Control should be completely free of any DNA/RNA. If you see this error message it means that at some point during the setup, the Negative Control has been contaminated with DNA/RNA and has given a positive signal. This contamination has invalidated the test. The Positive Control and your test samples are both possible sources of contaminating DNA/RNA. The genesig q16 reaction tubes from previous runs will also contain very high amounts of DNA so it is important that these are carefully disposed of after the run is completed and NEVER OPENED. It may be the case that your kits have become contaminated which will lead to the same problem occurring repeatedly. Solutions 1. Clean your working area using a commercial DNA remover solution to ensure the area is DNA free at the start of your run and re-run the test. 2. If the problem persists then the kit has become contaminated and it will have to be discarded and replaced with a new kit. When you open the new kit, run a simple test to show that changing the kit has solved the problem. Prepare a test which includes only the Positive Control, the Negative Control and one mock sample. For the mock sample add water instead of any sample RNA. The result for the Negative Control and the mock sample should be negative indicating that contamination is no longer present. 9

Preventive action An ideal lab set-up has a Clean area where the test reagents are prepared and a sample area where DNA/RNA samples and the Positive Control templates are handled. The best workflow involves setting up all the test components (excluding the positive control template) in the clean area and then moving the tests to the sample area for sample and Positive Control addition. If this method is followed then the kit components are always kept away from possible sources of contamination. For extra security the Negative Control can be completely prepared and sealed in the clean area. All work areas should be decontaminated regularly with DNA remover. Sample preparation failed Explanation The test has failed because the quality of the sample was not high enough. The Internal Extraction Control component identifies whether the sample has been prepared correctly and is of suitable quality. This error message means that this quality control test has failed and the sample quality is not high enough for analysis. Solutions 1. Check the sample preparation protocol for any user errors then repeat. 2. Poor quality samples can result from overloading the sample preparation protocol with too much starting material. Try reducing the amount of starting material then repeat. 3. Failing to add the Internal Extraction Control RNA to your sample during the sample preparation protocol can also lead to a reported result of sample preparation failed. Ensure that this step has not been overlooked or forgotten. If your samples are derived from an archive store or from a process separate from your genesig Easy extraction kit; you must add 5µl of Internal Extraction Control RNA into each 0.5ml of your sample to make it suitable for use on the q16. Positive result, poor quality sample Explanation The test is positive so if you are only interested in obtaining a present or absent answer for your sample then your result is reliable. However, the test contains an Internal Extraction Control component that identifies if the sample is of high quality. This quality control test has failed and the sample is not therefore of high enough quality to accurately calculate the exact copy number of RNA present. If you require quantitative information for your sample then proceed with the solutions below. Solutions 1. For appropriate solutions, read the Sample preparation failed section of this handbook. 10

Test failed Explanation The test has failed because the Positive Control has not worked. The Positive Control is present to show that all aspects of the test are working correctly together. When this control test fails, the test as a whole is invalidated. This finding indicates that a problem has occurred in the reaction set-up part of the experiment and has nothing to do with sample preparation. Solutions 1. Check the entire workflow and test set-up to look for any user errors, then repeat the test e.g. have the right colour pipettes and solutions been used with the correct tubes? 2. Ensure the positive and negative controls are inserted into the correct wells of your q16. 3. A component of the test may have gone off due to handing errors, incorrect storage or exceeding the shelf life. When you open a new kit, run a simple test to show that changing the kit has solved the problem. Prepare a test which includes only the Positive Control, the Negative Control and one mock sample. For the mock sample add internal control template instead of any sample RNA. If the Positive Control works, the mock sample will now be called as a negative result. Test failed and is contaminated Explanation The Positive Control is indicating test failure, and the Negative Control is indicating test contamination. Please read the Test Failed and Test contamination sections of this technical support handbook for a further explanation. Solution 1. For appropriate solutions, read both the Test failed and Test contaminated sections of this handbook. 11

Influenza, commonly known as the flu, is an infectious disease of birds and mammals caused by an RNA virus of the family Orthomyxoviridae (the influenza viruses). In people, common symptoms of influenza are fever, sore throat, muscle pains, severe headache, coughing, and weakness and fatigue. In more serious cases, influenza causes pneumonia, which can be fatal, particularly in young children and the elderly. Although the common cold is sometimes confused with influenza, it is a much less severe disease and caused by a different virus. Similarly, gastroenteritis is sometimes called "stomach flu" or "24-hour flu", but is unrelated to influenza. Typically, influenza is transmitted from infected mammals through the air by coughs or sneezes creating aerosols containing the virus, and from infected birds through their droppings. Influenza can also be transmitted by saliva, nasal secretions, feces and blood. Infections either occur through direct contact with these bodily fluids, or by contact with contaminated surfaces. Flu viruses can remain infectious for over 30 days at 0 C (32 F) and about one week at human body temperature, although they are rapidly inactivated by disinfectants and detergents. Flu spreads around the world in seasonal epidemics, killing millions of people in pandemic years and hundreds of thousands in non-pandemic years. Three influenza pandemics occurred in the 20th century each following a major genetic change in the virus and killed tens of millions of people. Often, these pandemics result from the spread of a flu virus between animal species. Since it first killed humans in Asia in the 1990s a deadly avian strain of H5N1 has posed the greatest influenza pandemic threat. However, this virus has not yet mutated to spread easily between people. Vaccinations against influenza are most common in high-risk humans in industrialised countries and farmed poultry. The most common human vaccine is the trivalent flu vaccine that contains purified and inactivated material from three viral strains. Typically this vaccine includes material from two influenza A virus subtypes and one influenza B virus strain. A vaccine formulated for one year may be ineffective in the following year, since the Influenza A Virus changes every year and different strains become dominant. Antiviral drugs can be used to treat influenza, with neuraminidase inhibitors being particularly effective. 12

Specificity The Primerdesign genesig Kit for (H1N1-swine) genomes is designed for the in vitro quantification of H1N1-swine genomes. The kit is designed to have the broadest detection profile possible whilst remaining specific to the H1N1-swine genome. The primers and probe sequences in this kit have 100% homology with a broad range of H1N1- swine sequences based on a comprehensive bioinformatics analysis. The kit has 2 primer and probe sets. The M1 primers and probe detect ALL influenza A isolates including those originally isolated from both human and swine. The primers have 100% homology to all pandemic H1N1 sequences. The N1 primer and probe set detects ONLY the pandemic swine flu H1N1. The primers and probe have 100% homology to all pandemic strains. The primers will not detect N1 from any seasonal endemic human or swine N1 strains. If you require further information, or have a specific question about the detection profile of this kit then please send an e.mail to enquiry@primerdesign.co.uk and our bioinformatics team will answer your question. 13

Kit storage and stability This lyophilised kit is stable at room temperature but should be stored at -20ºC on arrival. Once the lyophilised components have been resuspended they should not be exposed to temperatures above -20 C for longer than 30 minutes at a time and unnecessary repeated freeze/thawing should be avoided. The kit is stable for six months from the date of resuspension under these circumstances. Primerdesign does not recommend using the kit after the expiry date stated on the pack. Suitable sample material All kinds of sample material suited for PCR amplification can be used. samples are suitable in terms of purity, concentration, and RNA/DNA integrity. Please ensure the Dynamic range of test Under optimal PCR conditions genesig H1N1-swine detection kits have very high priming efficiencies of >95% and can detect less than 100 copies of target template. Notices and disclaimers This product is developed, designed and sold for research purposes only. It is not intended for human diagnostic or drug purposes or to be administered to humans unless clearly expressed for that purpose by the Food and Drug Administration in the USA or the appropriate regulatory authorities in the country of use. During the warranty period Primerdesign genesig detection kits allow precise and reproducible data recovery combined with excellent sensitivity. For data obtained by violation to the general GLP guidelines and the manufacturer s recommendations the right to claim under guarantee is expired. PCR is a proprietary technology covered by several US and foreign patents. These patents are owned by Roche Molecular Systems Inc. and have been sub-licensed by PE Corporation in certain fields. Depending on your specific application you may need a license from Roche or PE to practice PCR. Additional information on purchasing licenses to practice the PCR process may be obtained by contacting the Director of Licensing at Roche Molecular Systems, 1145 Atlantic Avenue, Alameda, CA 94501 or Applied Biosystems business group of the Applera Corporation, 850 Lincoln Centre Drive, Foster City, CA 94404. In addition, the 5' nuclease assay and other homogeneous amplification methods used in connection with the PCR process may be covered by U. S. Patents 5,210,015 and 5,487,972, owned by Roche Molecular Systems, Inc, and by U.S. Patent 5,538,848, owned by The Perkin-Elmer Corporation. Trademarks Primerdesign TM is a trademark of Primerdesign Ltd. genesig is a registered trademark of Primerdesign Ltd. 14