Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure. Coronary Artery Calcium Volume?

Similar documents
Variability of Repeated Coronary Artery Calcium Scoring and Radiation. Dose on 64-slice and 16-slice CT by Prospective

Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study

Department of Cardiology, Grosshadern Clinic, University of Munich, Marchioninistrasse 15, Munich, Germany. Department of Cardiology,

European Journal of Radiology

A Rosetta Stone for Coronary Calcium Risk Stratification: Agatston, Volume, and Mass Scores in 11,490 Individuals

Electron Beam CT of the Heart

Repeatability Limits for Measurement of Coronary Artery Calcified Plaque with Cardiac CT in the Multi-Ethnic Study of Atherosclerosis

Cardiac CT - Coronary Calcium Basics Workshop II (Basic)

Coronary Artery Calcification Scoring in Low-Dose Ungated CT Screening for Lung Cancer: Interscan Agreement

M Marwan, D Ropers, T Pflederer, W G Daniel, S Achenbach

Aortic Valve Calcification as a Marker for Aortic Stenosis Severity: Assessment on 16-MDCT

Improvement of Image Quality with ß-Blocker Premedication on ECG-Gated 16-MDCT Coronary Angiography

Coronary CT Angiography

Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction

Chapter. Reproducibility of. Coronary calcium measuements

Medical Policy Electron Beam CT for Detection of Coronary Artery Disease

Sang Ho Lee, Byoung Wook Choi, Hee-Joung Kim*, Member, IEEE, Haijo Jung, Hye-Kyung Son, Won-Suk Kang, Sun Kook Yoo, Kyu Ok Choe, Hyung Sik Yoo

Coronary calcification detected by electron-beam computed tomography and myocardial infarction

Assessment of Agatston Coronary Artery Calcium Score Using Contrast-Enhanced CT Coronary Angiography

Coronary Artery Calcification

Non-invasive intravenous coronary angiography using electron beam tomography and multislice computed tomography

Soft and Intermediate Plaques in Coronary Arteries: How Accurately Can We Measure CT Attenuation Using 64-MDCT?

Chapter 4. Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands. Department of Radiology,

Low Dose Era in Cardiac CT

Adapted Transfer Function Design for Coronary Artery Evaluation

The recent introduction of multirow detector computed

Disclosures CORONARY CALCIUM SCORING REVISITED. Learning Objectives. Scoring Methods. Consultant for M2S, Inc. Coronary Calcium Scoring: Software

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 51, NO. 1, FEBRUARY

Page 2 of 16

Coronary Calcium Predicts Events Better With Absolute Calcium Scores Than Age-Sex-Race/Ethnicity Percentiles

Radiation dose of cardiac CT what is the evidence?

Study of estimation of coronary artery calcium by multi-slice spiral CT scan in post myocardial infarction cases

Fundamentals, Techniques, Pitfalls, and Limitations of MDCT Interpretation and Measurement

Cardiac Computed Tomography

Computed Tomography Imaging of the Coronary Arteries

Coronary artery disease (CAD) is the leading

Coronary Artery Imaging. Suvipaporn Siripornpitak, MD Inter-hospital Conference : Rajavithi Hospital

Diagnostic and Prognostic Value of Coronary Ca Score

Zurich Open Repository and Archive

Gender differences in CT calcium scoring: A phantom study

Multidetector Computed Tomography (MDCT) in Coronary Surgery: First Experiences With a New Tool for Diagnosis of Coronary Artery Disease

Kumar S, Sharma S. Department of Cardiac Radiology, AIIMS, New Delhi, India

Technology Assessment Institute: Summit on CT Dose Cardiac CT - Optimal Use of Evolving Scanner Technologies

NIH Public Access Author Manuscript Acad Radiol. Author manuscript; available in PMC 2009 September 16.

Coronary Calcification Improves Cardiovascular Risk Prediction in the Elderly

X-Ray & CT Physics / Clinical CT

Imaging. A Combined Positron Emission Tomography/Computed Tomography Study

Accuracy of Multislice Computed Tomography in the Preoperative Assessment of Coronary Disease in Patients With Aortic Valve Stenosis

Angio-CT: heart and coronary arteries

Electron-Beam Tomography Coronary Artery Calcium and Cardiac Events

Coronary Artery Calcium to Predict All-Cause Mortality in Elderly Men and Women

Coronary artery disease

Utility of Coronary Artery Calcium Measurement In Cardiovascular disease. California Technology Assessment Forum

Multidetector-row cardiac CT: diagnostic value of calcium scoring and CT coronary angiography in patients with symptomatic, but atypical, chest pain

Coronary artery calcium screening: implications for clinical practice

Dual Energy CT of the Heart: Perfusion and Beyond

International Journal of Cardiology

A new method for radiation dose reduction at cardiac CT with multi-phase data-averaging and non-rigid image registration: preliminary clinical trial

Studies with electron beam computed tomography (EBCT) Imaging

Coronary Heart Disease. Calcium Concentration of Individual Coronary Calcified Plaques as Measured by Multidetector Row Computed Tomography

Lesion-Specific Coronary Artery Calcium Quantification for Predicting Cardiac Event with Multiple Instance Support Vector Machines

Triple Rule-out using 320-row-detector volume MDCT: A comparison of the wide volume and helical modes

Role of Nonenhanced Multidetector CT Coronary Artery Calcium Testing in Asymptomatic and Symptomatic Individuals 1

CLINICAL. Aims. Keywords Coronary artery calcium score Agatston score Volume score Mass score Cross-vendor analysis Computed tomography Dual-source

Sixty four slice Computed Tomography Scan (64-slice

A head-to-head comparison of the coronary calcium score by computed tomography with myocardial perfusion imaging in predicting coronary artery disease

Cover Page. The handle holds various files of this Leiden University dissertation

Objective Calcium score carotid IMT hs-crp

RAMA-EGAT Risk Score for Predicting Coronary Artery Disease Evaluated by 64- Slice CT Angiography

ORIGINAL INVESTIGATION

Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT

2 Cardiovascular Computed Tomography: Current and Future Scanning System Design

Fellows on this rotation are expected to attend nuclear conferences and multimodality imaging conference.

Spiral Multislice Computed Tomography Coronary Angiography: A Current Status Report

A Snapshot on Nuclear Cardiac Imaging

General Cardiovascular Magnetic Resonance Imaging

Aortic valve stenosis has a prevalence of 2% to 7% in the

Εξελίξεις και νέες προοπτικές στην καρδιαγγειακή απεικόνιση CT. Σταμάτης Κυρζόπουλος Ωνάσειο Καρδιοχειρουργικό Κέντρο

Pushing the limits of cardiac CT. Steven Dymarkowski Radiology / Medical Imaging Research Centre

Ultralow Dose Chest CT with MBIR

High Coronary Artery Calcium Scores Pose an Extremely Elevated Risk for Hard Events

Combined Anatomical and Functional Imaging with Revolution * CT

An Introduction to Dual Energy Computed Tomography

Jacobo Kirsch Ivan Buitrago Tan-Lucien H. Mohammed Tianming Gao Craig R. Asher Gian M. Novaro

Modifi ed CT perfusion contrast injection protocols for improved CBF quantifi cation with lower temporal sampling

Coronary artery and abdominal aortic calcification are associated with cardiovascular disease in type 2 diabetes

ORIGINAL INVESTIGATION. Using the Coronary Artery Calcium Score to Predict Coronary Heart Disease Events

Cardiac Imaging Tests

Characteristics of Subclinical Coronary Artery Disease in Diabetic Patients without Known Coronary Artery Disease

The role of coronary artery calcium score on the detection of subclinical atherosclerosis in metabolic diseases

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT? 1

Non-invasive coronary angiography with high resolution multidetector-row computed tomography

Corporate Medical Policy

Key words: Coronary artery calcium; coronary artery disease; calcium score; cardiovascular risk

Electron Beam Computed Tomographic Coronary Calcium Score Cutpoints and Severity of Associated Angiographic Lumen Stenosis

Computed Tomography of the Coronary Arteries

Given the burden of coronary artery disease (CAD), Prognostic Value of Coronary Artery Calcification CORONARY ARTERY CALCIFICATION

Computed Tomography to Detect Coronary Artery Calcification. Original Policy Date

X-ray coronary angiography is considered the diagnostic

Transcription:

Electron Beam CT versus 16-slice Spiral CT: How Accurately Can We Measure Coronary Artery Calcium Volume? 1

Objective: The purpose of this study is to investigate how accurately we can measure CAC volume using electron beam computed tomography (CT) and 16-slice spiral CT. Materials and Methods: CAC models with known volume attached to a cardiac phantom were scanned. The error of measurement, variability between measured and real volumes and inter-scan measurement variability were obtained. For spiral CT, 7 different parameters; i.e. (1) slice-thickness (0.625mm, 1.25mm and 2.5mm), (2) retrospective spiral electrocardiograph (ECG)-gated or prospective axial ECG-triggering, (3) overlapping or non-overlapping, were included. Results: The error of measurement was 15% on electron beam CT and 8-20% on spiral CT. CAC volumes were underestimated in 92% and overestimated in 8% of the electron beam CT scans. Volumes were underestimated in 79%, correct in 5% and overestimated in 16% of the spiral CT scans. The best measurement and the least variability was observed on 0.625mm retrospective spiral ECG-gated CT (error of 8%), a significant result (t-test: p<0.01) when compared with electron beam CT. Conclusion: CAC volume measurement on CT scanners may be significantly different and often underestimates the real volume of CAC. For precise evaluation of CAC volume, thin-slice retrospective spiral ECG-gated scan using a spiral CT scanner is desirable. 2

The risk of myocardial infarction and sudden cardiac death is related to the coronary artery calcium (CAC) score [1,2] as defined by electron beam CT. The score is the product of volume and density of calcification as defined by Agatston [3]. Monitoring CAC is suggested to assess the progression or regression of coronary atherosclerosis, thereby documenting a risk factor which might suggest future medical intervention [4,5]. Originally, Agatston score and then volume [6] and mass [7] scoring were introduced for the quantification of CAC. However, whether CT quantification of moving CAC reflects actual CAC volume remains a matter for debate. The purpose of this study is to assess measurement accuracy by quantification of CAC models with known volume, attached to a cardiac phantom. Materials and Methods: An originally designed cardiac phantom, consisting of 5 components: driver, control, support, rubber balloon phantom and an ECG, was used. The phantom, mimicking the left ventricle, was filled with a mixture of water and contrast medium (58HU) to simulate non-contrast blood, and was submerged in corn oil (-118HU) simulating epicardial and pericardial fat. For CAC models, putty (501HU) with 2 sizes/volumes: 3mm/30mm 3 and 5mm/50mm 3 was packed inside rubber tubes (mimicking coronary arteries) attached to the balloon surface (Fig. 1). For heart rate sequences, 1) low heart rate (60bpm), 2) high heart rate (85bpm), 3) high heart rate (85bpm) with shifting and 4) arrhythmia sequences were set (Fig. 2). The ends of the balloon were stabilized to a fixed support at a distance of 10cm. However, deformity of the balloon resulted in some through-plane motion of the 3

calcium models themselves. The volumes of the balloon phantom were approximately 100 and 200ml at the systolic and diastolic phases, respectively. The diastolic phase of the phantom corresponded to around 80% of the R-R interval, however the duration of the diastolic phase was inconstant, depending on the heart rate. Three sequential CT scans were repeated using both an electron beam CT scanner (C-150 XL; Imatron, South San Francisco, CA) and a 16-slice spiral CT scanner (LightSpeed Ultrafast 16; GE Medical Systems, Waukesha, WI). The scan was performed at three different starting points for the table (-1, 0 and +1mm) in order to determine the effect of the position of the patient on the scoring result (to mimic interscan variability). Electron Beam CT Protocol The standard electron beam CT protocol used was as follows; 100msec acquisition time, 35-40 continuous gapless slices of 3mm thickness, 130kV, 625mA. The single-section mode images were obtained by ECG-triggered to 80% of the R-R interval. Image reconstruction was performed with a 512 x 512 pixel matrix using a sharp reconstruction kernel. The display field was 18cm. Retrospective ECG-gated spiral CT protocol Volumetric data of the phantom were obtained by helical mode with scan parameters of 0.625mm collimation width x 16 detectors, gantry rotation speeds of 0.5 (sequence 1, 4) and 0.6 (sequence 2, 3) sec/rotation, 120kV and 100mA. The pitch was set to 0.275, enabling for multisector reconstruction (where pitch is defined as table feed per gantry rotation divided by the total x-ray beam width (NxT) where N is the number of active DAS channels and T is the single DAS channel width). Multisector reconstruction 4

means that images, to improve temporal resolution, are retrospectively reconstructed by combining some (n=2 to 4) adjacent cardiac cycles [8]. Four kinds of reconstructions were created; (1) 0.625mm thickness images with 0.625mm increment, (2) 1.25mm/1.25mm, (3) 2.5mm/1.25mm; i.e. overlapping reconstruction and (4) 2.5mm/2.5mm. The center of the temporal window was set to 80% of the R-R interval. The matrix size and field of view were the same as for the electron beam CT protocol, and the reconstruction kernel was standard. Almost all spiral CT images had a temporal resolution from 100 to 250msec, although the temporal resolution achieved by multisector reconstruction differed according to the heart rate and the number of cardiac cycles used for image reconstruction. Prospective ECG-triggering axial CT protocol Axial images with thickness of 0.625mm, 1.25mm and 2.5mm were scanned at prospective triggering so that the center of the temporal window corresponded to 80% of the R-R interval. Scanning parameters of a gantry rotation speed 0.5 sec/rotation, 120kV and 100mA were used. The matrix size and field of view were the same as for the electron beam CT protocol, and the reconstruction kernel was standard. Temporal resolution was 250msec. Phantom scan and reconstruction on electron beam CT and 16-slice CT is summarized in Table 1. Error of measurement and variability between measured and real volumes We measured volume (not the so called calcium volume score) of the calcium models. In the measurement, volume score [6] was used, with the calcium detection 5

threshold set to half the CT value of the calcium model (250HU). With this setting, if more than half of a pixel included calcium, the volume score became positive, and if less than half, this led to a negative measurement. This concept, strictly speaking, is justified when calcium is surrounded by water (CT value zero). In this study however, apart from calcium materials, pixels might include a part or all parts of the surrounding structure, i.e. a mixture of water and contrast medium, corn oil and rubber tube. The error of measurement and variability of the measured volume were calculated using the following equation: Error = abs(measurement real volume) / real volume Variability = 2 x abs(measurement real volume) / (measurement + real volume). This was performed on each of the three sequential scans for the calcium models and the results were compared between CT protocols as well as between calcium sizes. It was tested to see whether calcium had been under or overestimated. The calcium models were also scanned in a static state. Inter-scan variability The inter-scan variability was calculated using the following equation: Inter-scan variability = 2 x abs(s1 S2) / (S1 + S2) where S1 is CAC score on the first scan. This was calculated between scan1 and scan2, scan2 and scan3 and between scan3 and scan1. Statistical Analysis: For statistical analysis, t-tests were used to determine differences in variability. P-values < 0.05 were considered to identify significant differences. 6

Results: In a static state, error of measurement was 5% on electron beam CT and 2-6% on spiral CT. The best measurement was observed on 0.625mm spiral CT. Error of measurement on electron beam CT and spiral CT is shown (Fig. 3). The error on electron beam CT was 15%. On 16-slice spiral CT scanner, the error was different between scan protocols (8% to 20%). The error of 8% on 0.625mm spiral CT and 9% on 1.25mm spiral CT was significantly lower than that on electron beam CT (p<0.01, 0.02). Variability between measured and real volumes for calcium, and inter-scan variability is shown (Fig. 4). The variability between measured and real volumes was 16% on electron beam CT and 8-24% on spiral CT. The inter-scan variability was 12% on electron beam CT and 8-26% on spiral CT. The best results were obtained from 0.625mm spiral CT, both on the variability between real and measured volumes (8%) and inter-scan variability (8%). Figure 5 shows the type of scan taken and whether the resulting measurement was an underestimation, correct or an overestimation. In 92% on electron beam CT and 63 to 100% (79% in overall) on spiral CT, CT measurement resulted in underestimation. When compared between 3mm and 5mm calcium models, variability between measured and real volumes in the 3mm model was significantly higher than that in the 5mm model (Fig. 6). Discussion: There are many studies on inter-scan variability in CAC measurement, however 7

variability between real CAC volume and that measured on CT has not received much focus due to its not being assessable in vivo. We have found only one phantom study [9] calculating volume score [6] of calcified plaques with known volumes. Factors influencing inter-scan variability on CAC measurement reported are as follows; partial volume effect [10], the use of the step function in the Agatston method to quantitate calcium [11], coronary artery motion [12], image noise [13], field inhomogeneity [7], lack of calibration [14], total volume of CAC [15], scoring parameters [16], etc. In the current study, factors contributing to improving measurement accuracy on spiral CT are thinner slice thickness and use of retrospective spiral ECG-gated scan. These also act in reducing variability between measured and real volumes and inter-scan variability [17, 18]. This is explainable as related to factors, partial volume effect and coronary artery motion; i.e. temporal resolution. Despite relatively low inter-scan variability (10%), the level of measurement error (14%) on 2.5mm overlapping spiral CT seems to be important as it may lead to an inappropriate assignment in CAC risk stratification [19, 20]. Our result is different from that attained by Kopp et al. which, using a 4-slice spiral CT scanner (2.5mm thickness with 1mm increment), showed very good accordance. The reason for this is not clear, however. One possibility considered is that our heart rate protocols included shift and arrhythmia sequences which produce more through-plane motion of calcium than stable heart rates do. In addition, on multisector reconstruction, which we used, small variations in heart rate are an important factor for optimizing temporal resolution [21]. We were also surprised by strong the similarity between volume score and real volume in the study by Kopp et al.. This is because it is emphasized by Ulzheimer et al. [18] 8

that volume score does not correspond to a real physical measurement as it strongly depends on the thresholds used. Schlosser et al. suggest that motion may result in smearing artifacts and an underestimation of the score [22]. In most scans, our results support this theory. However, correct or overestimation of calcium volume existed in a minor number of cases. This indicates it is not adequate to hypothesize that real CAC volume is higher than what is actually measured on CT. Variability between measured and real volumes was higher in the 3mm model scan. This seems to be explainable mainly by partial volume effect. Inter-scan variability of the Agatston score on electron beam CT is also known to be high on a low-score group [5]. The CAC models used in the current study do not have the exact properties as a real CAC in vivo. The putty material we used has a regular shape, homogenous density and greater x-ray absorption. The reason we selected such models was that our main purpose was to evaluate measurement accuracy rather than test inter-scan variability. Furthermore, putty was place inside the tube rather than into or on the wall of the tube, as is the case with CAC in vivo. The setting of threshold may affect volume measurement. In the current study, we set the CAC threshold to half the CT value of the calcium model which is measured on spiral CT. A recent study [23] shows that image attenuation values vary by CT scanners, therefore highlighting the importance of the use of calibration phantoms. However, the fact that error of measurement was only 5% in a static state on electron beam CT and that this was comparable to those on spiral CT, seems to suggest that the threshold 9

factor is not considered to have a significant effect. We do not imply that we propose to change the CAC quantification approach; i.e. measuring actual CAC volume instead of Agatston, volume score and mass score. Absolute volume measurement is never possible in vivo. What is most important is validating a reproducible technique which can predict cardiac events. Although the best measurement and the least variability was observed on 0.625mm retrospective spiral ECG-gated CT, it brings it with the most radiation exposure to maintain the image quality needed, which is an important factor in CAC scoring [13]. Determination of slice thickness and reduction of the tube current or voltage should be validated in further clinical studies. One limitation of our study is that we did not test 1.5mm-thickness on electron beam CT. Different from the electron beam CT scanner we used, a new generation electron beam CT scanner is able to cover a whole heart using 1.5mm-thickness. This, of course, is expected to improve measurement accuracy. In conclusion, we need to be aware of the fact that CAC volume measurement on either electron beam or spiral CT may be significantly different and often underestimates real CAC volume. For precise evaluation of CAC volume, thin-slice retrospective spiral ECG-gated scan using a spiral CT scanner is desirable. 10

References: 1. Arad Y, Spadaro LA, Goodman K, Newstein D, Guerci AD. Prediction of coronary events with electron beam computed tomography. J Am Coll Cardiol 2000;36:1253-1260. 2. Wong ND, Hsu JC, Detrano RC, Diamond G, Eisenberg H, Gardin JM. Coronary artery calcium evaluation by electron beam computed tomography and its relation to new cardiovascular events. Am J Cardiol 2000;86:495-498 3. Agatston AS, Janowitz WR, Hildner FJ, Zusmer NR, Viamonte M, Detrano R. Quantification of coronary calcium using ultrafast computed tomography. J Am Coll Cardiol 1990;15:827-832 4. Callister TQ, Raggi P, Cooil B, Lippolis NJ, Russo DJ. Effect of HMG-CoA reductase inhibitors on coronary artery disease as assessed by electron-beam computed tomography. N Engl J Med 1998;339:1972-1978 5. Budoff MJ, Lane KL, Bakhsheshi H, et al. Rates of progression of coronary calcium by electron beam tomography. Am J Cardiol 2000;86:8-11 6. Callister TQ, Cooil B, Raya SP, et al. Coronary artery disease: improved reproducibility of calcium scoring with an electron-beam CT volumetric method. Radiology 1998;208:807-814 7. Detrano R, Kang X, Mahaisavariya P, et al. Accuracy of quantifying coronary hydroxyapatite with electron beam tomography. Invest Radiol 1994;29:733-738 8. Horiguchi J, Nakanishi T, Ito K. Quantification of coronary artery calcium using multidetector CT and a retrospective ECG-gating reconstruction algorithm. AJR 2001;177:1429-1435 11

9. Kopp AF, Ohnesorge B, Becker C, et al. Reproducibility and Accuracy of Coronary Calcium Measurements with Multi Detector Row versus Electron-Beam CT. Radiology 2002;225:113-119 10. Kajinami K, Seki H, Takekoshi N, Mabuchi H. Quantification of coronary artery calcification using ultrafast computed tomography: reproducibility of measurements. Coron Artery Dis 1993;4:1103-1108 11. Yoon HC, Greaser III LE, Mather R, et al. Coronary artery calcium: alternate methods for accurate and reproducible quantification. Acta Radiol 1997; 4:666-673 12. Mao S, Bakhsheshi H, Lu B, Liu SCK, Oudiz RJ, Budoff MJ. Effect of electrocardiogram triggering on reproducibility of coronary artery calcium scoring. Radiology 2001;220:707-711 13. Bielak LF, Kaufmann RB, Moll PP, McCollough CH, Schwartz RS, Sheedy PF. Small lesions in the heart identified at electron beam CT: calcification or noise? Radiology 1994;192:631-636 14. McCollough CH, Kaufmann RB, Cameron BM, Katz DJ, Sheedy PF, Peyser PA. Electron-beam CT: use of a calibration phantom to reduce variability in calcium quantification. Radiology 1995;196:159-165 15. Achenbach S, Ropers D, Mohlenkamp S, et al. Variability of repeated coronary artery calcium measurements by electron beam tomography. Am J Cardiology 2001;87:210-213 16. van Ooijen PM, Vliegenthart R, Witteman JCM, Oudkerk M. Influence of scoring parameter settings on Agatston and volume scores for coronary calcification. Eur Radiol 2005;15:102-110 12

17. Ohnesorge B, Flohr T, Fischbach R, et al. Reproducibility of coronary calcium quantification in repeat examinations with retrospectively ECG-gated multisection spiral CT. Eur Radiol 2002;12:1532-1540 18. Ulzheimer S and Kalender WA. Assessment of coronary artery scoring performance in cardiac computed tomography. Eur Radiol 2003;13:484-497 19. Rumberger JA, Brundage BH, Rader DJ, Kondos G. Electron beam computed tomographic coronary calcium scanning: a review and guidelines for use in asymptomatic patients. Mayo Cli Proc 1999;74:243-252 20. Becker CR, Majeed A, Crispin A et al, CT measurement of coronary calcium mass: impact on global cardiac risk assessment. Eur Radiol 2005;15:96-101 21. Wicky S, Rosol M, Hamberg LM, et al. Evaluation of retrospective multisector and half scan ECG-gated multidetector cardiac CT protocols with moving phantoms. JCAT 2002;26:768-776 22. Schlosser T, Hunold P, Schmermund A, et al. Coronary artery calcium score: Influence of reconstruction interval at 16 detector row CT with retrospective electrocardiographic gating. Radiology 2004;233:586-589 23. Nelson JC, Kronmal RA, Carr JJ, et al. Measuring coronary calcium on CT images adjusted for attenuation differences. Radiology 2005;235:403-414 13

Figure Legends Fig. 1 Cardiac phantom Picture shows two sizes of CAC models packed inside rubber tubes (mimicking coronary arteries) and attached to the balloon surface. 14

Fig. 2 Heart rate sequences Graph shows the four types of heart rate sequences used in the study; 1) low heart rate (60bpm), 2) high heart rate (85bpm), 3) high heart rate (85bpm) with shifting and 4) arrhythmia. 15

Fig. 3 Error of measurement Graph shows the error of measurement (%) on electron beam CT and 16-slice spiral CT scanners. The error was 8% on 0.625mm spiral CT and 9% on 1.25mm spiral CT, significantly lower than 15% on electron beam CT (p<0.01, 0.02). 16

Fig. 4 Variability between measured and real volumes and inter-scan variability Graph shows variability (%) between measured and real volumes (white bars) and inter-scan variability (gray bars). Thinner slice thickness and use of retrospective spiral ECG-gated scan are factors in reducing both variabilities. Variability between measured and real volumes tended to be higher than inter-scan variability on electron beam CT (p=0.07) and 2.5mm overlapping spiral CT (p=0.07). 17

Fig. 5 Estimation of calcium volume on CT Graph shows underestimation (white bars) and overestimation (gray bars) in CAC volume measurement. CT often underestimates the real volume of CAC. The sum of underestimation and overestimation ratios does not come to 100% in some of the CT protocols that have the correct measurement. 18

Fig. 6 3mm vs 5mm calcium models in variability between real and measured volume Graph shows variability (%) between real and measured volumes in 3mm calcium models (white bars) is higher than that in 5mm models (gray bars), and reached a significant level (p<0.05) on spiral CT. 19

Table 1: Summary of phantom scan and reconstruction on electron beam CT and 16-slice CT No. of phantom 1 No. of CAC models 2 (3mm/30mm3 and 5mm/50mm3, putty) No. of heart rate curves per scan 4 (60, 85, 85 shifting and arrythmia; in bpm) No. of scans per protocol 3 (home position = -1, 0 and +1 mm) No. of protocols 8 1 on EBCT: 3/3mm 4 on MSCT Retrospective: 0.625/0.625mm, 1.25/1.25mm, 2.5/1.25mm, 2.5/2.5mm 3 on MSCT Prospective: 0.625mm, 1.25mm, 2.5mm No. Number, CAC. coronary artery calcium EBCT. electron beam CT, MSCT. multi slice (16-slice) CT 20