Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Similar documents
Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Determination of Polymerized Triglycerides by High Pressure Liquid Chromatography and Corona Veo Charged Aerosol Detector

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Determination of Olive Oil Adulteration by Principal Component Analysis with HPLC Charged Aerosol Detector Data

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Characterization of Used Cooking Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

Direct Analysis of Multicomponent Adjuvants by HPLC with Charged Aerosol Detection

Impurity Profiling of Carbamazepine by HPLC/UV

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Novel, Universal Approach for the Measurement of Natural Products in a Variety of Botanicals and Supplements, Part 2

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Fast UHPLC Methods for Analysis of Amino Acids

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Determination of Carbohydrates in Kombucha Using HPAE-PAD

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Mercury Speciation Determinations in Asian Dietary Supplements

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Determination of Phenolic Compounds in Apple Orchard Soil

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Analysis of Commercially Available Products Containing Stevia

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Thermo Scientific Dionex CarboPac MA1 Column

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Increased Identification Coverage and Throughput for Complex Lipidomes

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Determination of Amantadine Residues in Chicken by LCMS-8040

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Determination of Tigecycline in a Cell Lysate

Flupyradifurone. HPLC Method

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Sensitive Analysis of Commonly Used Artificial and Natural Sweeteners Including Stevia and Their Impurities and Degradation Products

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Determination of Steviol Glycosides by HPLC with UV and ELS Detections

ASSAY AND IMPURITY METHOD FOR DURACOR TABLETS BY HPLC

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

A New Era in SFC Column Technology

Improving Food Process Control Using the

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Application Note. Authors. Abstract. Petrochemical

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA200

Thermo Scientific ART Pipette Tips

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

Quantitative Method to measure Glycidol Fatty Acid Esters (GEs) in Edible Oils

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography

High protein content determination in food and animal feed by combustion using the Thermo Scientific FlashSmart Elemental Analyzer

Transcription:

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA

Overview Purpose: To develop HPLC methods for quantitation of emulsifiers in foods, using an HPLC system with a charged aerosol detector. Methods: Two methods, using the Thermo Scientific Accucore C18 column and the Thermo Scientific Hypersil Gold Silica column, were used to quantify hydroxypropylmethylcellulose (HPMC, modified cellulose) and Lecithin (via phosphatidylcholine amount), respectively. Results: Samples of foods or ingredients were analyzed for content of lecithin or HPMC using the developed methods. Introduction Emulsifiers are used to maintain a uniform suspension of immiscible materials. These compounds are typically surfactants, and can be designed for use in specific applications and products. Acylglycerols are used in food products containing oil and water (e.g. margarine, mayonnaise); lecithin is commonly found in chocolate and spray oils 1 ; acid esters of monoglycerides are used for dough conditioners; and hydroxypropylmethyl cellulose (HPMC) is use to thicken dairy products and help improve flavor characteristics. 2 HPMC is also an important emulsifier used in the pharmaceutical industry. 3 The analysis of emulsifiers is becog increasingly important, for product quality, consistency and stability properties. High performance liquid chromatography (HPLC) is one of the more prevalent methods for analyzing these compounds. However, the majority of these analytes do not contain suitable chromophore characteristics for UV detection, which then requires the use of a universal detector, such as evaporative light scattering, refractive index, or charged aerosol detection. The Thermo Scientific Dionex Corona Veo charged aerosol detector was used in the analyses of different emulsifiers that were extracted from food products. The Corona Veo, a sensitive mass-based detector, is ideally suited for the direct measurement of emulsifiers, as they are non-volatile and non-chromophoric compounds. It offers excellent sensitivity (down to low nanogram amounts on column), a dynamic range of over 4 orders of magnitude, and similar inter-analyte response independent of chemical structure. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity, dynamic range and precision than ELSD and refractive index (RI), is gradient compatible and is simpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS). This sensitivity, combined with the linearity that is possible with use of the Corona Power Function, provides a unique and complete analytical solution for sensitive, reproducible, and routine analysis of non-chromophoric analytes. Several examples of emulsifier HPLC separations are detailed. FIGURE 1. Schematic and functioning of charged aerosol detection. 1 3 10. 2 4 5 8 6 7 9 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charged gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge is measured by a highly sensitive electrometer 10. Signal transferred to chromatographic software 2 Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Methods Liquid Chromatography Lecithin / DPPC HPLC System: Thermo Scientific Dionex UltiMate 3000 System with LPG3400-SD pump (normal phase), WPS3000 RS autosampler, and TCC3000 RS column oven HPLC Column: Hypersil Silica 5 µm, 3.0 150 mm Column Temperature: 50 C Mobile Phase A: Water, 18.2 MΩ cm Mobile Phase B: 2-Propanol Mobile Phase C: iso-octane Flow Rate: 0.2 1.5 ml/ Injection Volume: 2 10 µl Detector: Flow Gradient: Corona Veo SD Evaporator Temperature: 50 C Filter: 5 s Data rate: 10 Hz PowerFunction: 1.40 Time Flow (utes) (ml/) %A %B %C -2.0 1.5 2 63 35 0.5 1.5 2 63 35 0.2 0.2 2 63 35 0.0 0.2 2 63 35 0.2 1.5 7 58 35 7.0 1.5 11 54 35 11 1.5 11 54 35 12 1.5 2 63 35 Standard and Sample Preparations Standards and samples were dissolved in methanol / chloroform (1:9). If the solutions were clear, samples were used as is. If the sample was not clear, samples were centrifuged at 10,000 g for 5 utes and the supernatant was used. If the sample was still cloudy, the samples were then centrifuge-filtered under the same conditions using a 0.2 micron membrane. Liquid Chromatography HPMC HPLC System: UltiMate 3000 System with LPG3600-DGP pump, WPS3000 RS autosampler, and TCC3000 RS column oven HPLC Columns: Imtakt Presto FF-C18, 2.0 µm, 150 x 4.6 mm Column Temperature: 40 C Mobile Phase A: Water Mobile Phase B: Acetonitrile Mobile Phase C: 2-Propanol Flow Rate: 0.4 ml/ Injection Volume: 2 10 µl Detector: Corona Veo SD Evaporation Temperature: 50 C Filter: 3.6 s Data rate: 5 Hz PowerFunction: 1.90 Flow Gradients: non-dairy: 5% B for 5 utes before injection, hold 1 ute after injection, gradient to 100% B to 3 utes, hold to 8 utes, return to 5% to 9 utes. Dairy (extended gradient): 5% B for 5 utes before injection, hold 1 ute after injection, gradient to 100% B to 3 utes, hold to 8 utes, gradient to 100% C to 9 utes, hold 100% to 12.5 utes, gradient to 100% B to 13 utes, return to 5% B to 14 utes. Standard and Sample Preparations Standards and samples were dissolved in water. If the solutions were clear, samples were used as is. If the sample was not clear, samples were centrifuged at 10,000 g for 5 utes and the supernatant was used. Samples requiring more thorough cleaning were centrifuged-filtered using a 0.2 micron membrane. Data Analysis All HPLC chromatograms were obtained and compiled using Thermo Scientific Dionex Chromeleon Chromatography Data System software, 6.8 SR 13. Thermo Scientific Poster Note PN70995_PITTCON_2014_E_03/14S 3

Results Sample Analysis- Lecithin Lecithin was detered by the analysis of phosphatidyl choline in samples, as measured against a dipalmitoylphosphatidylcholine (DPPC) standard calibration curve. Because the material is comprised of various acyl groups, the peak shape between the samples and the DPPC standard were different, thus requiring a linear calibration curve. A solution of DPPC, 1 mg/ ml was prepared in methanol/chloroform (1:9), sequentially diluted and analyzed for calibration. A chromatogram of the DPPC standard is shown in Figure 2. To provide a linear calibration fit for DPPC for amounts 39 10,000 ng on column (ng o.c.) a power function value (PFV) of 1.4 was identified. The linear calibration curve shown in Figure 3 had a linear regression coefficient, R 2 = 0.9996 and %RSD of 2.84. Figure 2. HPLC-CAD chromatogram of DPPC, 2500 ng on column. 25.0 pa 1 - DPPC/PC - 8.042 12.5-5.0 0.0 2.5 5.0 7.5 12.0 Figure 3. Linear regression fit of HPLC-CAD calibration data, 39 to 10,000 ng o.c., each amount in triplicate, inset 39-1250 ng o.c. 25.0 22.5 DPPC/PC External CAD_1 Area [pa*] 20.0 17.5 15.0 12.5 10.0 7.5 3.00 Area [pa*] 2.50 2.00 1.50 5.0 1.00 0.50 2.5 ng o.c. 0.00 0 125 250 375 500 625 750 875 1,000 1,125 1,300 0.0 0 1,250 2,500 3,750 5,000 6,250 7,500 8,750 10,000 12,000 The linear fit spanned four order of magnitude. For sensitivity, the limit of quantitation, based on signal-to-noise ratio of 10, was detered to be 20 ng o.c. Three samples were dissolved, clarified, and analyzed, including lecithin, a granola bar, shown in Figure 4, and krill oil. The results are shown in Table 1, below. Table 1. Phosphatidylcholine found in food samples. Phosphatidylcholine Sample Claim amount found (mass-%) Lecithin, Laboratory Grade Percent target 47.4 N/A N/A Granola Bar 0.05 < 2% N/A Krill Oil 34.1 34.9* 97.7 *AOCS Official Method Ja 7c-07 4 4 Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Figure 4. HPLC chromatogram of lecithin (DPPC/PC) in granola bar, extracted in methanol / chloroform (1:9), centrifuge-filtered. 50.0 45.0 pa 40.0 35.0 Current [pa] 30.0 25.0 20.0 15.0 3 - DPPC/PC - 7.988 10.0 5.0 1-2.753 2-5.070 21-2.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 Sample Analysis- HPMC Retention Time [] Hydroxypropylmethyl cellulose (HPMC) was dissolved in water and analyzed for calibration, from 156 to 5000 ng o.c., and the peak areas were fit to a linear regression line, as shown in Figure 5. The resulting correlation coefficient was 0.998 with a %RSD of 5.53. A PFV of 1.9 was used, and the estimated limit of quantitation is approximately 10 ng o.c. The peak area %RSD values ranged from 0.53 (625 ng o.c.) to 5.21 (156 ng o.c.). Figure 5. Linear HPMC calibration curve, from 156 to 5000 ng o.c., each amount analyzed in triplicate. 0.400 HPMC External CAD_1 Area [pa*] 0.350 0.300 0.250 0.200 0.150 0.100 0.050 ng o.c. 0.000 0 500 1,000 1,500 2,000 2,500 3,000 3,500 4,000 4,500 5,000 5,500 6,000 Two samples were prepared and analyzed: a popsicle and a more complex frozen milk product, each contained less than 1% of HPMC ( modified cellulose ). The popsicle was melted and diluted, 99.5 mg (100 µl) into 900 µl water, and analyzed directly. The frozen milk product, however was diluted and vortex-mixed (44.7 mg in 900 µl of water) and centrifuge-filtered. Due to the added lipids of the milk product, an extended gradient was also used to clean the column of the lipids. A spike-recovery sample of the dairy product was also prepared, using 47.9 mg of product and 10 µl of 10,000 ng o.c. standard solution (additional 1000 ng o.c.). The results of the two products and the spike recovery sample are provided in Table 2. Table 2. HPMC found in food samples Sample HPMC found (mass-%) Claim amount Recovery (%) Popsicle 0.05 < 1% N/A Dairy Product 0.21 < 1% N/A Spiked Dairy Product 835 ng o.c. (spiked) 1000 ng o.c. spiked 83.5 Thermo Scientific Poster Note PN70995_PITTCON_2014_E_03/14S 5

Figure 6. Overlaid HPLC-CAD chromatograms showing HPMC in a popsicle, melted and diluted, in duplicate. 0.500 0.400 pa 1-2.447 0.300 0.200 2 - HPMC - 6.857 0.100-0.020 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 Figure 7. Overlaid HPLC-CAD chromatograms of HPMC in a frozen dairy product, 44.7 mg dissolved in 900 µl water, centrifuged, analyzed in duplicate. 4.00 pa 1-2.417 3.00 2 - HPMC - 6.740 2.00 1.00 Lipids -0.10 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 Conclusions Two HPLC methods were developed to analyze two, different emulsifiers commonly found in food products, lecithin and HPMC (modified cellulose). The lecithin method was able to detere the amount of phosphatidylcholine found in a food sample, in the ingredient itself, as well as in a natural nutraceutical product, krill oil, with results matching the official AOCS method for phospholipids. Sensitivity was 20 ng o.c. LOQ. HPMC was calibrated over a wide range of concentrations, and the method was able to detere HPMC in two food products, including a spike-recovery of 83.5% and sensitivity to 10 ng o.c. LOQ. Both of these analytes varied in their composition, mainly in their acyl groups, so linear calibration was required for accurate deterations. The Corona Veo charged aerosol detector was able to provide linear calibration curves for these analytes. The sensitivity and linear dynamic range provided useful, analytical data for these analytes that also lack a chromophore. References 1. Van Nieuwenhuyzen, W. Emulsions in food products: preparation, stabilization and applications in Emulsions and Nanosuspensions for the Formulation of Poorly Soluble Drugs; Müller, R.H.; Benita, S.; Böhm, B. H. L., Eds.; medpharm GmbH Scientific Publishers, Stuttgart, 1998; pp 82-99. 2. de Mariscal, P.; Bell, D.A. Fiber-Based Mimetics: Methylcellulose Gums in Handbook of Fat Replacers; Roller, S.; Jones, S.A., Eds.; CRC Press, LLC: Boca Rotan, FL, 1996. 3. Jones, D. In Pharmaceutical Applications of Polymers for Drug Delivery; Humphreys, S., Ed.; Rapra Review Reports, 15, Smithers Rapra Press: Shropshire, UK, 2004; pp 1-124. 4. http://search.aocs.org/methods/search_methods_view_method.cfm?method= JA7C_07.pdf (last accessed 07 Feb 2014). 6 Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752 PN70995_E 03/14S