Absolute anchorage has been a long sought after,

Similar documents
Evaluation of cortical bone thickness of mandible with cone beam computed tomography for orthodontic mini implant installation

3D Cortical Bone Anatomy of the Mandibular Buccal Shelf: a CBCT study to define sites for extra-alveolar bone screws to treat Class III malocclusion

Bone density assessment for mini-implants position

Measurement of Crestal Cortical Bone Thickness at Implant Site: A Cone Beam Computed Tomography Study

In recent years, orthodontic mini-implants have been

Soft Tissue Thickness for Placement of an Orthodontic Miniscrew Using an Ultrasonic Device

The International Journal of Periodontics & Restorative Dentistry

CBCT evaluation of interdental cortical bone thickness at common orthodontic miniscrew implant placement sites

6. Timing for orthodontic force

Clinical indices for orthodontic mini-implants

There is little controversy regarding whether temporary

AAO 115th Annual Session San Francisco, CA May 17 (Sunday), 1:15-2:00 pm, 2015

Analysis of Buccal Cortical Plate Thickness for Placement of Mini-Implants A CBCT Study

THE USE OF TEMPORARY ANCHORAGE DEVICES FOR MOLAR INTRUSION & TREATMENT OF ANTERIOR OPEN BITE By Eduardo Nicolaievsky D.D.S.

Dental Implants: A Predictable Solution for Tooth Loss. Reena Talwar, DDS PhD FRCD(C) Oral & Maxillofacial Surgeon Associate Clinical Professor

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and.

Anchorage control is a critical consideration when

Influence of Third Molars on the Availability of Interradicular Spaces for Miniscrew Implant Placement

Assessment of palatal bone thickness in adults with cone beam computerised tomography

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

New treatment paradigms have reduced the importance

Osseointegrated dental implant treatment generally

KJLO. A Sequential Approach for an Asymmetric Extraction Case in. Lingual Orthodontics. Case Report INTRODUCTION DIAGNOSIS

Maxillary Molar Distalization with micro-implants:

Evaluation of alveolar cortical bone thickness and density for orthodontic mini-implant placement

Three-dimensional evaluation of maxillary anterior alveolar bone for optimal placement of miniscrew implants

Since their introduction, the use of orthodontic

Intramembranous autogenous bone graft is the gold

TADs Supported Tongue Crib: A New Minimalistic Design

Prosthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor

Intraoral molar-distalization appliances that

Cortical and cancellous bone thickness on the anterior region of alveolar bone in Korean: a study of dentate human cadavers

Tomographic mapping of the hard palate and overlying mucosa

Maxillary sinus perforation by orthodontic anchor screws

ORIGINAL ARTICLE , 1,2 onplant, 3,4, . 12, 13,14 , 7,8. 9, ,,. 16,17

With judicious treatment planning, the clinical

Open Access Relationship Between the Thickness of Cortical Bone at Maxillary Mid-palatal Area and Facial Height Using CBCT

Assessment of the relationship between the maxillary molars and adjacent structures using cone beam computed tomography

Monday Morning Pearls of Practice by Bobby Baig

Immediate Loading with Flapless Implant Surgery for Rehabilitation of Single Bound Edentulous Space

Easy placement is an advantage of orthodontic

Dental Morphology and Vocabulary

Utilizing Digital Treatment Planning and Guided Surgery in Conjunction with Narrow Body Implants. by Timothy F. Kosinski, DDS, MAGD

Orthodontic space opening during adolescence is

ORTHODONTIC CORRECTION Of OCCLUSAL CANT USING MINI IMPLANTS:A CASE REPORT. Gupta J*, Makhija P.G.**, Jain V***

Orthodontic Microimplants and Its

Molar distalisation with skeletal anchorage

Controlling Tissue Contours with a Prosthetically Driven Approach to Implant Dentistry

Anchorage conservation has always been a challenge

A Clinical Evaluation of Anatomic Features of Gingiva in Dental Students in Tabriz, Iran

PAIN PERCEPTION DURING MINIPLATE-ASSISTED ORTHODONTIC THERAPY

CASE REPORT. CBCT-Assisted Treatment of the Failing Long Span Bridge with Staged and Immediate Load Implant Restoration

The treatment options for nongrowing skeletal Class

Diagnostics and treatment planning. Dr. Attila Szűcs DDS

Use of Palatal Miniscrew Anchorage and Lingual Multi-Bracket Appliances to Enhance Efficiency of Molar Scissors-Bite Correction

Bone Reduction Surgical Guide for the Novum Implant Procedure: Technical Note

The Brånemark osseointegration method, using titanium dental implants (fixtures)

Advanced Probing Techniques

Anchorage control is a fundamental prerequisite

Nonextraction Treatment of Upper Canine Premolar Transposition in an Adult Patient

The LOMAS System. Eric J. W. Liou, James C. Y. Lin

The majority of the early research concerning

Influence of Surface Characteristics on Survival Rates of Mini-Implants

International Journal of Current Medical and Pharmaceutical Research

Contemporary Implant Dentistry

Case Report. RapidSorb Rapid Resorbable Fixation System. Ridge augmentation in a one-step surgical protocol.

Quantitative assessment of interradicular bone density in the maxilla and mandible: implications in clinical orthodontics

Research Article Success Rate of Microimplants in a University Orthodontic Clinic

Dental Implant Planning using Cone Beam CT imaging: a pictorial guide.

Periodontal Disease. Radiology of Periodontal Disease. Periodontal Disease. The Role of Radiology in Assessment of Periodontal Disease

During the last 30 years, endosseous oral

Sang-Sun Han, 1 Kwang-Min Lee, 2 and Kee-Deog Kim Introduction

Simple Mechanics to Upright Horizontally Impacted Molars with Ramus Screws

Objective: The purpose of this study was to evaluate the reliability and accuracy of a

The Clinical Success of Self-tapping and Self-drilling Orthodontic Miniscrews

Root Proximity Characteristics and Type of Alveolar Bone Loss: A Case-control Study

PALATAL POSITIONING OF IMPLANTS IN SEVERELY RESORBED POSTERIOR MAXILLAE F. Atamni, M.Atamni, M.Atamna, Private Practice Tel-aviv Israel

CT Scanning Protocol For V2R Guided Surgery Solutions

A cone-beam computed tomography evaluation of buccal bone thickness following maxillary expansion

The Effect of X Ray Vertical Angulation on Radiographic Assessment of Alveolar Bone Loss

The International Journal of Periodontics & Restorative Dentistry

Lingual correction of a complex Class III malocclusion: Esthetic treatment without sacrificing quality results.

Case Report: Long-Term Outcome of Class II Division 1 Malocclusion Treated with Rapid Palatal Expansion and Cervical Traction

Two- and Three-dimensional Orthodontic Imaging Using Limited Cone Beam Computed Tomography

International Journal of Medical and Health Sciences

Overerupted upper molars due to missing lower

Bone density determination for the maxilla and the mandible in different age groups by using computerized tomography (Part I)

Crowded Class II Division 2 Malocclusion

SKELETAL ANCHORAGE IN ORTHODONTIC TREATMENT OF A CLASS II MALOCCLUSION

An Introduction to Dental Implants

The Application of Cone Beam CT Image Analysis for the Mandibular Ramus Bone Harvesting

Orthodontic Correction of Adult Gummy Smile with TADs and Periodontal Surgery

Molar intrusion with skeletal anchorage ; from single tooth intrusion to canting correction and skeletal open bite

Complications of postoperative swelling of the maxillary sinus membrane after sinus floor augmentation

Young-Jin Park, DDS,* and Sung-Am Cho, DDS, MS, PhD

Arch dimensional changes following orthodontic treatment with extraction of four first premolars

Orthodontic-prosthetic implant anchorage in a partially edentulous patient

Labial and lingual/palatal bone thickness of maxillary and mandibular anteriors in human cadavers in Koreans

Lever-arm and Mini-implant System for Anterior Torque Control during Retraction in Lingual Orthodontic Treatment

Mesial Step Class I or Class III Dependent upon extent of step seen clinically and patient s growth pattern Refer for early evaluation (by 8 years)

Transcription:

ORIGINAL ARTICLE Buccal cortical bone thickness for mini-implant placement Sebastian Baumgaertel a and Mark G. Hans b Cleveland, Ohio Introduction: The thickness of cortical bone is an important factor in mini-implant stability. In this study, we investigated the buccal cortical bone thickness of every interdental area as an aid in planning mini-implant placement. Methods: From the cone-beam computed tomography scans of 30 dry skulls, 2-dimensional slices through every interdental area were generated. On these, cortical bone thickness was measured at 2, 4, and 6 mm from the alveolar crest. Intraclass correlation was used to determine intrarater reliability, and analysis of variance (ANOVA) was used to test for differences in cortical bone thickness. Results: Buccal cortical bone thickness was greater in the mandible than in the maxilla. Whereas this thickness increased with increasing distance from the alveolar crest in the mandible and in the maxillary anterior sextant, it behaved differently in the maxillary buccal sextants; it was thinnest at the 4-mm level. Conclusions: Interdental buccal cortical bone thicknes varies in the jaws. There appears to be a distinct pattern. Knowledge of this pattern and the mean values for thickness can aid in mini-implant site selection and preparation. (Am J Orthod Dentofacial Orthop 2009;136:230-5) Absolute anchorage has been a long sought after, but rarely achieved, treatment ideal. Therefore, orthodontic mini-implants, which can provide this desired form of anchorage, are more popular than ever. This is also reflected in the increasing number of studies in the orthodontic literature. However, despite the recent surge in scientific articles, plenty is still unknown about this topic. 1 The literature is rich in case reports illustrating the successful use of mini-implants for various indications. Mini-implants are placed in many anatomic sites, depending on the indication and the biomechanics used. 2-5 Popular implant sites appear to be the palate, the lingual aspect of the maxillary alveolar process, the retromolar area in the mandible, and the buccal cortical plate in the maxilla and the mandible. 2-7 The latter has proven to be a versatile placement site and has thus been the subject of several investigations. 8-10 Many local anatomic factors must be considered; no single factor can be isolated to mark the ideal placement From the Department of Orthodontics, School of Dental Medicine, Case Western Reserve University, Cleveland, Ohio. a Assistant clinical professor. b Professor and chairman. The authors report no commercial, proprietary, or financial interest in the products or companies described in this article. Reprint requests to: Sebastian Baumgaertel, Case Western Reserve University, School of Dental Medicine, Department of Orthodontics, 10900 Euclid Ave, Cleveland, OH 44106; e-mail, sxb155@case.edu. Submitted, June 2007; revised and accepted, October 2007. 0889-5406/$36.00 Copyright Ó 2009 by the American Association of Orthodontists. doi:10.1016/j.ajodo.2007.10.045 site. Among the more important factors for placement in the buccal cortex are soft-tissue anatomy, interradicular distance, sinus morphology, nerve location, and buccolingual bone depth. 11,12 More recently, an interest in cortical bone thickness and quality has developed in conjunction with orthodontic skeletal anchorage systems. 9,13-15 The influence of bone quality on the long-term success of oral implants is undisputed and has been known for over a decade. 16 Bone quality can be expressed as the ratio of cortical to trabecular bone. In 1985, Lekholm and Zarb 17 suggested a classification system to categorize bone into 4 classes of bone quality. This ultimately became the standard classification in implant dentistry: (1) almost the entire jaw is homogenous compact bone, (2) a thick layer of compact bone surrounds a core of dense trabecular bone, (3) a thin layer of cortical bone surrounds a core of dense trabecular bone, and (4) a thin layer of cortical bone surrounds a core of low-density trabecular bone. However, further investigations showed that it is not primarily the ratio of cortical bone to trabecular bone that influences implant stability as much as the absolute amount of dense cortical bone. Cortical bone has a higher modulus of elasticity than trabecular bone, is stronger and more resistant to deformation, and will bear more load in clinical situations than trabecular bone. 18 What applies to traditional dental implants also applies to orthodontic mini-implants: thicker cortical bone provides greater primary stability. 19-21 Knowledge of the buccal cortical bone thickness in various areas can guide clinicians in selecting the 230

American Journal of Orthodontics and Dentofacial Orthopedics Baumgaertel and Hans 231 Volume 136, Number 2 Fig 1. Two-dimensional sagittal view showing the orientation of the interdental slice on which measurements were made. Fig 3. Two-dimensional interdental slice, showing cortical (white) and trabecular (gray) bone with measurements of buccal cortical bone thickness. Fig 2. Two-dimensional horizontal view showing the orientation of the interdental slice on which measurements were made. placement site and the proper placement protocol. The purpose of this study was to investigate buccal cortical bone thickness in every interdental site in both jaws to provide a guideline for implant site selection and placement. MATERIAL AND METHODS The sample consisted of 30 adult dry skulls. Inclusion criteria were no more than 2 missing teeth per arch excluding third molars. Of these skulls, 9 had 1 missing tooth, and 3 had 2 missing teeth. All teeth, however, were lost postmortem, as shwon by the sockets. The skulls were imaged in maximum intercuspation with cone-beam computerized tomography (CBCT) with the Hitachi CB Mercuray CBCT Unit (Hitachi Medical, Tokyo, Japan) at a 9-in field of view, 100 kvp, and 10 ma. Two-dimensional slices, 0.28 mm thick, through each contact area were created, bisecting the interradicular distance (Fig 1) and oriented perpendicular to the bone surface (Fig 2) by using Accurex software (CyberMed, Seoul, Korea). This gave a clear image of cortical and cancellous bone (Fig 3). The demarcation between the 2 bone qualities was drawn manually by visual gray-white discrimination; gray was cancellous bone, and white was cortical bone. Each measurement area was coded with a number, beginning in the maxillary right quadrant, distal to the second molar with number 1 and ending in the

232 Baumgaertel and Hans American Journal of Orthodontics and Dentofacial Orthopedics August 2009 Fig 4. Schematic of interdental area coding and definition of sextants. Table I. Buccal cortical bone thickness at different measurement levels by sextants 2 mm SD 4 mm SD 6 mm SD Mean Sig MaxR 1.15 0.27 1.03 0.22 1.28 0.28 1.16 * MaxM 0.85 0.23 0.94 0.18 1.11 0.24 0.97 * MaxL 1.11 0.25 1.00 0.22 1.29 0.26 1.14 * MandR 1.48 0.47 1.89 0.51 2.22 0.6 1.87 NS MandM 0.85 0.15 0.99 0.21 1.15 0.26 1.00 * MandL 1.60 0.53 2.03 0.56 2.32 0.59 1.98 NS *P #0.001; NS, not significant; Sig, significance. mandibular right quadrant distal to the second molar with number 30. For analysis, these measurement areas were also grouped into sextants (Fig 4). Three measurement points were then defined at 2, 4, and 6 mm from the alveolar crest in each measurement area. From these, measurements were made perpendicular to the bone surface by using the function ruler in the Accurex software (Fig 3). Data were entered into an Excel 2003 spreadsheet (Microsoft, Redmond, Wash). On 10 randomly selected skulls, all measurements were made twice to assess intrarater reliability, 2 weeks apart. Statistical analysis SPSS software (version 15.0, SPSS, Chicago, Ill) was used for all statistical calculations. Preliminary data analysis indicated normal frequency distribution of the sample and equality of variances. Intraclass correlation was used to determine intrarater reliability, and analysis of variance (ANOVA) was used to test for differences in cortical bone thickness. The level of significance was set at P #0.05. RESULTS Intrarater reliability was high (r 5 0.91), and the measurements proved to be reproducible. Statistical analysis showed that, on average, buccal cortical bone was thicker in the mandible than in the maxilla, and that the buccal sextants of both jaws (maxilla left [MaxL], maxilla right [MaxR], mandible left [MandL], and mandible right [MandR]) had greater cortical bone thickness than the anterior sextants (maxilla middle [MaxM] and mandible middle [MandM]). When the means of the measurement levels (2, 4, and 6 mm) were compared in each sextant, the differences were significant in the entire maxilla (MaxL, MaxM, and MaxR) and the anterior mandibular sextant (MandM), but not in the mandibular buccal sextants (MandR and MandL). In the maxillary buccal sextants (MaxL and MaxR), cortical bone thickness was thickest at the 6-mm level and thinnest at the 4-mm level. In the maxillary anterior sextant (MaxM), thickness increased progressively with increasing distance to the alveolar crest and was thickest at the 6-mm level. The mandibular anterior sextant (MandM) was similar to its maxillary counterpart; cortical bone thickness increased away from the alveolar crest. In the mandibular buccal sextants (MandR and MandL), the same trend was observed (Table I). Further analysis showed that, overall, each contact differed significantly from the rest. Cortical bone thickness increased in both jaws with increasing distance from the midsagittal plane except distally to the maxillary second molars, where it decreased (Table II).

American Journal of Orthodontics and Dentofacial Orthopedics Baumgaertel and Hans 233 Volume 136, Number 2 Table II. Buccal cortical bone thickness at the measurement levels in each interdental area (1-30) Area 2 mm SD 4 mm SD 6 mm SD Mean SD Sig 1 1.00 0.31 0.98 0.21 1.17 0.27 1.05 0.28 2 1.23 0.23 1.10 0.19 1.39 0.27 1.24 0.26 3 1.25 0.27 1.09 0.24 1.34 0.30 1.23 0.29 4 1.16 0.24 1.01 0.24 1.29 0.27 1.15 0.27 5 1.11 0.28 0.98 0.23 1.22 0.26 1.10 0.27 6 0.91 0.18 0.99 0.22 1.13 0.27 1.01 0.25 7 0.89 0.28 1.00 0.20 1.12 0.26 1.00 0.25 8 0.75 0.18 0.85 0.21 0.99 0.24 0.86 0.23 9 0.88 0.39 0.88 0.13 1.16 0.21 0.97 0.21 10 0.85 0.15 1.01 0.15 1.17 0.22 1.01 0.22 11 1.16 0.18 1.03 0.17 1.28 0.19 1.16 0.21 12 1.14 0.24 1.03 0.20 1.32 0.25 1.16 0.26 13 1.18 0.26 1.04 0.21 1.34 0.28 1.19 0.28 14 1.20 0.28 1.06 0.27 1.35 0.26 1.20 0.29 15 0.90 0.30 0.86 0.26 1.15 0.31 0.97 0.32 16 2.28 0.79 2.86 0.72 3.05 0.76 2.73 0.82 17 2.00 0.94 2.49 0.82 2.89 0.74 2.46 0.91 18 1.44 0.37 1.85 0.47 2.20 0.55 1.83 0.56 19 1.26 0.30 1.66 0.42 1.96 0.57 1.63 0.53 20 1.01 0.23 1.28 0.29 1.49 0.32 1.26 0.34 21 0.90 0.19 1.05 0.22 1.27 0.23 1.07 0.26 22 0.82 0.15 0.98 0.22 1.11 0.26 0.97 0.24 23 0.82 0.23 0.93 0.14 1.03 0.20 0.93 0.17 24 0.85 0.17 0.96 0.24 1.10 0.32 0.97 0.27 25 0.86 0.14 1.04 0.25 1.25 0.30 1.05 0.29 26 0.91 0.19 1.14 0.27 1.39 0.28 1.15 0.32 27 1.16 0.31 1.50 0.42 1.88 0.54 1.51 0.52 28 1.32 0.41 1.74 0.51 2.13 0.65 1.73 0.62 29 1.45 1.35 2.01 0.67 2.60 0.72 2.02 0.76 30 2.58 1.08 3.08 0.74 3.12 0.80 2.92 0.91 * *P #0.05; P #0.01; P #0.001; Sig, significance. DISCUSSION Turkyilmaz et al 22 demonstrated that it is possible to estimate primary implant stability from presurgical computerized tomography (CT) diagnosis. Therefore, it would be desirable to image every patient with CT, but not every clinician has access to a CTor CBCT scanner and might lack some potentially important information. This investigation was intended to supply a guideline when CT imaging is not possible. In this study, we found an interesting pattern of variation in buccal cortical bone thickness. In general, the mandible provides more buccal cortical bone than the maxilla. Friberg et al 23 found similar results in their morphometric study; they reported more compact cortical bone in the mandible than in the maxilla. In the mandible and the maxillary anterior sextant, buccal cortical bone increases in thickness as the distance of the measurement points from the alveolar crest increases. This is no surprise, since one would expect cortical bone thickness to increase from the alveolar bone to the basal bone. In the maxillary buccal sextants, however, buccal cortical bone thickness decreases at the 4-mm mark before it increases again at the 6-mm mark. This finding agrees with the study of Kim et al, 15 who found that, in the buccal interproximal areas of the maxilla that they investigated, the cortical bone was thickest closest to and farthest from the CEJ and thinnest in the middle. This was an interesting result because it demonstrates that, to maximize cortical bone anchorage in the maxillary buccal sextants, the mini-implant should be placed more than 4 mm apically from the alveolar crest. This means that most mini-implants in the maxillary buccal sextants must be placed close to the mucogingival junction or perhaps even in mucosa. Another interesting finding was that buccal cortical bone is thinnest in the anterior sextants of both jaws and increases progressively toward the posterior, except distally to the maxillary second molars, where the buccal cortex is on average thin. The implications for clinical purposes are that, to place the implant in a more favorable spot, biomechanics should be ideally structured so that placement is possible in the buccal sextants, except

234 Baumgaertel and Hans American Journal of Orthodontics and Dentofacial Orthopedics August 2009 Fig 5. Map of buccal cortical bone thickness in each measurement area: measurement sites are ranked by overall buccal cortical bone thickness in each sextant (1 is the thickest). Red is the thinnest measurement in a site; yellow is the medium measurement; green is the thickest measurement. for the maxillary retromolar area that provides insufficient buccal cortical bone. This finding reinforces a similar one by Deguchi et al, 24 who found significantly less buccal cortical bone distal to the maxillary second molars. Placement in the anterior sextants of both jaws should be avoided for several reasons: this area provides little cortical bone for anchorage of the implant and little attached gingiva, and it frequently lacks sufficient interradicular distances. 8,25 It can be debated whether these findings are also clinically significant considering all other factors that can affect mini-implant success rates. 10 It is also well understood that shortcomings in cortical bone thickness can be compensated for by variations of mini-implant angulation or perhaps implant design (cylindrical vs conical shank). 24 However, as Miyamoto et al 20 demonstrated, overall cortical bone thickness is important in implant stability and therefore should be considered when selecting the preferred implant site. Therefore, in addition to the statistical significance, our findings appear to have clinical significance also. These results were charted in a visual format that might help clinicians in selecting the most ideal placement sites for cortical bone availability, and it can serve as a guideline for preparing the implant site for maximum primary stability and minimal osseous damage in the implant site (Fig 5). When this aid is used, it must be kept in mind that measurements from CBCT slice data, although sufficiently accurate, tend to slightly underestimate the anatomic truth. 26 There might be slightly more bone available than the mean values indicate. A possible shortcoming of this study was that the distributions of sex and age in this sample are unknown; these might have an impact on the thickness of cortical bone. However, this could not be determined in a previous study. 24 Even if this influences the absolute measured values slightly, it would probably not change the observed pattern. Although over one third of the sample had at least 1 missing tooth, this cannot be regarded as a limitation, since these teeth were lost postmortem. This is indicated by the alveolar sockets of the missing teeth. If teeth are lost after death as a result of the fixation process or handling, cortical bone thickness is not affected. CONCLUSIONS Interdental buccal cortical bone thickness appears to vary according to a distinctive pattern. Knowledge of this pattern can aid clinicians in implant site selection and proper site preparation. A visual aid was created to facilitate the clinical application of these results. Future studies are needed to determine the exact relationship between cortical bone thickness, the method of implant site preparation, and success rates. REFERENCES 1. Cornelis MA, Scheffler NR, De Clerck HJ, Tulloch JF, Nyssen- Behets C. Systematic review of the experimental use of temporary skeletal anchorage devices in orthodontics. Am J Orthod Dentofacial Orthop 2007;131(4 Suppl):S52-8. 2. Park HS, Kwon OW, Sung JH. Nonextraction treatment of an open bite with microscrew implant anchorage. Am J Orthod Dentofacial Orthop 2006;130:391-402. 3. Roth A, Yildirim M, Diedrich P. Forced eruption with microscrew anchorage for preprosthetic leveling of the gingival margin. J Orofac Orthop 2004;65:513-9. 4. Park YC, Choi YJ, Choi NC, Lee JS. Esthetic segmental retraction of maxillary anterior teeth with a palatal appliance and

American Journal of Orthodontics and Dentofacial Orthopedics Baumgaertel and Hans 235 Volume 136, Number 2 orthodontic mini-implants. Am J Orthod Dentofacial Orthop 2007;131:537-44. 5. Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997;31:763-7. 6. Xun C, Zeng X, Wang X. Microscrew anchorage in skeletal anterior open-bite treatment. Angle Orthod 2007;77:47-56. 7. Park HS, Lee SK, Kwon OW. Group distal movement of teeth using microscrew implant anchorage. Angle Orthod 2005;75:602-9. 8. Schnelle MA, Beck FM, Jaynes RM, Huja SS. A radiographic evaluation of the availability of bone for placement of miniscrews. Angle Orthod 2004;74:832-7. 9. Poggio PM, Incorvati C, Velo S, Carano A. Safe zones : a guide for miniscrew positioning in the maxillary and mandibular arch. Angle Orthod 2006;76:191-7. 10. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop 2006;130:18-25. 11. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop 2007;131(4 Suppl): S43-51. 12. Baumgaertel S, Razavi MR, Hans MG. Mini-implant anchorage for the orthodontic practitioner. Am J Orthod Dentofacial Orthop 2008;133:621-7. 13. Henriksen B, Bavitz B, Kelly B, Harn SD. Evaluation of bone thickness in the anterior hard palate relative to midsagittal orthodontic implants. Int J Oral Maxillofac Implants 2003;18:578-81. 14. Kang S, Lee SJ, Ahn SJ, Heo MS, Kim TW. Bone thickness of the palate for orthodontic mini-implant anchorage in adults. Am J Orthod Dentofacial Orthop 2007;131(4 Suppl):S74-81. 15. Kim HJ, Yun HS, Park HD, Kim DH, Park YC. Soft-tissue and cortical-bone thickness at orthodontic implant sites. Am J Orthod Dentofacial Orthop 2006;130:177-82. 16. Zarb GA, Schmitt A. Implant prosthodontic treatment options for the edentulous patient. J Oral Rehabil 1995;22:661-71. 17. Lekholm U, Zarb GA. Patient selection and preparation. In: Branemark PI, Zarb GA, Albrektsson T, editors. Tissue-integrated prosthesis: osseointegration in clinical dentistry. Chicago: Quintessence; 1985. p. 199-209. 18. Holmes DC, Loftus JT. Influence of bone quality on stress distribution for endosseous implants. J Oral Implantol 1997;23:104-11. 19. Homolka P, Beer A, Birkfellner W, Nowotny R, Gahleitner A, Tschabitscher M, et al. Bone mineral density measurement with dental quantitative CT prior to dental implant placement in cadaver mandibles: pilot study. Radiology 2002;224:247-52. 20. Miyamoto I, Tsuboi Y, Wada E, Suwa H, Iizuka T. Influence of cortical bone thickness and implant length on implant stability at the time of surgery clinical, prospective, biomechanical, and imaging study. Bone 2005;37:776-80. 21. Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop 2006;67:162-74. 22. Turkyilmaz I, Tozum TF, Tumer C, Ozbek EN. Assessment of correlation between computerized tomography values of the bone, and maximum torque and resonance frequency values at dental implant placement. J Oral Rehabil 2006;33:881-8. 23. Friberg B, Sennerby L, Roos J, Lekholm U. Identification of bone quality in conjunction with insertion of titanium implants. A pilot study in jaw autopsy specimensclin Oral Implants Res199;6:213-9. 24. Deguchi T, Nasu M, Murakami K, Yabuuchi T, Kamioka H, Takano-Yamamoto T. Quantitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. Am J Orthod Dentofacial Orthop 2006;129:721.e7 12. 25. Tenenbaum H, Tenenbaum M. A clinical study of the width of the attached gingiva in the deciduous, transitional and permanent dentitions. J Clin Periodontol 1986;13:270-5. 26. Lascala CA, Panella J, Marques MM. Analysis of the accuracy of linear measurements obtained by cone beam computerized tomography (CBCT-NewTom). Dentomaxillofac Radiol 2004;33:291-4.