Panorama. Arrhythmia Analysis Frequently Asked Questions

Similar documents
Basic Dysrhythmia Interpretation

Arrhythmia Monitoring ST/AR Algorithm

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

Electrocardiography for Healthcare Professionals

Cardiac Telemetry Self Study: Part One Cardiovascular Review 2017 THINGS TO REMEMBER

The Nuts and Bolts of ICD Therapy

Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

ID: BFH_ Start time: :00:00. Examination summary

Telemetry Monitoring For Nurses and Monitor Technicians May, 2011

SARASOTA MEMORIAL HOSPITAL

Rate: The atrial and ventricular rates are equal; heart rate is greater than 100 bpm (usually between bpm).

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Cardiac Arrhythmias in Sleep

COURSE DESCRIPTION. Rev 2.0 7/2013. Page 1 of 26

ECG INTERPRETATION MANUAL

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

CRC 431 ECG Basics. Bill Pruitt, MBA, RRT, CPFT, AE-C

Bedside ECG Monitoring For Nurses BSM 6000 Series November, 2010

Defibrillator. BiomedGuy

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

GE Healthcare. The GE EK-Pro Arrhythmia Detection Algorithm for Patient Monitoring

Course Objectives. Proper Lead Placements. Review the ECG print paper. Review the mechanics of the Myocardium. Review basics of ECG Rhythms

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

Diagnostic capabilities of the implantable therapeutic systems

Electrocardiography for Healthcare Professionals

CORONARY ARTERIES HEART

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

HSCRIBE 5 HOLTER ANALYSIS SYSTEM

BEDSIDE ECG INTERPRETATION

INSTRUCTION MANUAL. phantom 320 ECG Simulator 1.4

Figure 2. Normal ECG tracing. Table 1.

ECG MACHINE KEC- Series

HTEC 91. Performing ECGs: Procedure. Normal Sinus Rhythm (NSR) Topic for Today: Sinus Rhythms. Characteristics of NSR. Conduction Pathway

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018

ABCs of ECGs. Shelby L. Durler

MULTI-PARAMETER PATIENT SIMULATOR

Dr.Binoy Skaria 13/07/15

Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms

Practice Name Practice Address Practice Address 2 Practice Phone Number

Appendix D Output Code and Interpretation of Analysis

Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG?

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Ambulatory ECG Monitoring. Gill Bouncir Blackpool Victoria Hospital

TELEMETRY BASICS FOR NURSING STUDENTS

Rigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective

Cardiac Arrhythmia How to approach นพ.พ น จ แกวส วรรณะ หน วยโรคห วใจและหลอดเล อด

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Supplemental Material

Advanced ICD Concepts

Delineation of QRS-complex, P and T-wave in 12-lead ECG

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

Rigel UNI-SiM. The world s smallest integrated NiBP, SpO2 and patient simulator. Key Features n Compact and cost-effective

FLB s What Are Those Funny-Looking Beats?

II. PROCEDURE DESCRIPTION A. Normal Waveform from an Electrocardiogram Figure 1 shows two cycles of a normal ECG waveform.

LABVIEW based expert system for Detection of heart abnormalities

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

2017/8/23 1. SE-2003&SE-2012 Holter Analysis System

Speed - Accuracy - Exploration. Pathfinder SL

COPYRIGHT Ramsey Medical, Inc.

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

NOISE DETECTION ALGORITHM

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Cardiology Flash Cards

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

I PART I. Timing Cycles and Troubleshooting Review COPYRIGHTED MATERIAL

RN-BC, MS, CCRN, FAHA

Miscellaneous Stuff Keep reading the Outline

Project Title Temporary Pacemaker Training Simulator

ECG ABNORMALITIES D R. T AM A R A AL Q U D AH

USING CORRELATION COEFFICIENT IN ECG WAVEFORM FOR ARRHYTHMIA DETECTION

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Supraventricular Arrhythmias. Reading Assignment. Chapter 5 (p17-30)

PEDIATRIC CARDIAC RHYTHM DISTURBANCES. -Jason Haag, CCEMT-P

The ABCs of EKGs/ECGs for HCPs. Al Heuer, PhD, MBA, RRT, RPFT Professor, Rutgers School of Health Related Professions

Is This Thing Working?

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

UNDERSTANDING YOUR ECG: A REVIEW

EKG Competency for Agency

Basic EKG Interpretation. Nirja Parikh, PT, DPT

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital

Blocks & Dissociations. Reading Assignment (p47-52 in Outline )

DEVELOPMENT OF A SIMPLE SOFTWARE TOOL TO DETECT THE QRS COMPLEX FROM THE ECG SIGNAL

ProSim 3 and 2. Vital Signs Simulators

Arrhythmic Complications of MI. Teferi Mitiku, MD Assistant Clinical Professor of Medicine University of California Irvine

WHAT S THAT RHYTHM I AM HEARING? GUIDE TO AUSCULTATION OF ARRHYTHMIAS IN HORSES

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

ECG UPDATE COURSE DESCRIPTION

La strategia diagnostica: il monitoraggio ecg prolungato. Michele Brignole

ProSim 3 and 2 Vital Signs Simulators

Conduction Problems / Arrhythmias. Conduction

Chad Morsch B.S., ACSM CEP

EKG. Danil Hammoudi.MD

This presentation will deal with the basics of ECG description as well as the physiological basics of

4/14/15 HTEC 91. Topics for Today. Guess That Rhythm. Premature Ventricular Contractions (PVCs) Ventricular Rhythms

Basic ECG Interpretation Module Notebook

Transcription:

Panorama Arrhythmia Analysis Frequently Asked Questions

What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead set 5-wire lead set and 12 lead Viewed ECG vector II and V Viewed ECG vector I, II, and V What ECG vectors are used for V-Fib Detection? 3-wire lead set 5-wire lead set and 12 lead Viewed ECG vector II and V 2

How is the Heart Rate calculated? Leads use the mean R-R interval of the last 16 beats if the HR > 48 bpm. If the Heart Rate derived from the last 4 beats is 48 bpm, then that rate is used for HR. PVCs and Paced beats are included. What is the algorithm s definition of Asystole? No QRS detection for a configurable time period (3-10 Seconds) 3

What is the algorithm s definition of V-Fib? Presence of a Fib waveform with no recognizable P, QRS, or T waves in a 5 second window What is the algorithm s definition of V-Tach? When the PVC rate and number of consecutive PVCs for the configurable V-Tach setting have been met. 4

What are the user configurable options for the V-Tach rate? The V-Tach rate can be set between 100-180 bpm What are the user configurable options for the V-Tach threshold? The V-Tach threshold can be set between 3-15 consecutive PVCs What is the algorithm s definition of Run? The PVC Rate meets criteria for V-Tach rate but did not meet the number of consecutive PVCs to be called V-Tach 5

What is the algorithm s definition of Ventricular Rhythm? Number of consecutive PVCs is >2 and the V-Tach rate is not reached What is the algorithm s definition of Irregular Heart Rate? When the measured variations in the R-R interval over a period of time exceed a preset limit established by the arrhythmia analysis algorithm 6

What is the algorithm s definition of Bigeminy? Three or more N-V or V-N consecutive sequences What is the algorithm s definition of Trigeminy? Three or more N-N-V or V-N-N consecutive sequences What is the algorithm s definition of Bradycardia? Heart rate 10% lower than selected Low HR alarm limit 7

What does Relearn do? Selecting Relearn updates the ST and Arrhythmia templates Note: Relearn sets new ST baselines and the ability to compare to initial baseline will be lost. Note: Ensure waveform being learned is the patient s baseline rhythm and not a lethal arrhythmia or artifact. 8

When does an Automatic Relearn occur? 3-wire lead set Change in viewed ECG vector 3-wire lead set, 5-wire lead set and 12 lead Initial admission Resume monitoring after a Standby mode Arrhythmia and/or ST is enabled After correction of lead fault of >60 sec 9

When should a Manual Relearn be done? Suspected false alarm calls Change in patient s dominant beat and/or rhythm occurs 10

What arrhythmia calls can be made during a Learn period? Only calls during Learn: Asystole (once 3 beats have been detected) V-Fib (once 3 beats have been detected) What does the message ECG Noise mean? The algorithm has detected noise in the ECG waveform and cannot perform arrhythmia analysis during this time. Correct the noise to return to arrhythmia processing. Does adjusting the wave gain (size) of the ECG waveform impact algorithm performance or heart rate calculation? No. 11

The monitor is calling Asystole or Low Heart Rate despite the visual presence of an ECG waveform. What could be the issue? Alarm may be latched (Asystole) Acknowledge a latched alarm by selecting over the patient s digital or waveform area to clear the latched alarm if condition is resolved. The QRS complexes are too small for detection by the algorithm. Move ECG patches closer together to increase the voltage of the QRS complexes. Select Relearn to establish new baseline templates Change displayed lead (if using 3-wire lead set) 12

Why is the V-Tach alarm called when V-Tach is not present in the waveform? Alarm may be latched (V-Tach) Acknowledge a latched alarm by selecting over the patient s digital or waveform area to clear the latched alarm if condition is resolved. Baseline rhythm has changed and arrhythmia templates need to be updated Select Relearn to establish new baseline templates Artifact may be present Poor connectivity or skin contact Secure ECG lead wires and cables Electrocautery interference The algorithm may not clearly distinguish the QRS complexes from T waves that are similar in height Reposition ECG patches to establish taller QRS complexes 13

Are there any special considerations for patients with permanent Pacemakers? HR is calculated from the patient s QRS complex, not the Pacemaker enhancement. Reposition electrodes away from the pacemaker device to reduce pacemaker noise 14