Is This Thing Working?

Similar documents
Friedman, Rott, Wokhlu, Asirvatham, Hayes 201. Figure 65.7 Shortening of the AV interval during pacing.

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

Timing cycles. Objectives Upon completion of this program the participant will be able to:

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Pacing and Device Jargon Made Simple. Dr Jonathan Timperley MB ChB MD FRCP Consultant Cardiologist Northampton General Hospital

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

Programming of Bradycardic Parameters. C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany

How to prevent unecessary right ventricular pacing

Cardiac rhythm detailed monitoring by an implanted pacemaker: The iecg solution

Device Interrogation- Pacemakers, ICD and Loop Recorders. Dulce Obias-Manno, RN, MHSA, CCDS,CEPS, FHRS Device Clinic Coordinator, MHVI

Implant Procedure Concepts

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

Electrocardiography for Healthcare Professionals

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

How to Approach the Patient with CRT and Recurrent Heart Failure

Pediatric pacemakers & ICDs:

Figure 2. Normal ECG tracing. Table 1.

Diagnostic capabilities of the implantable therapeutic systems

Teaching Rounds in Cardiac Electrophysiology

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Intermittent episodes of paced tachycardia: what is the cause?

Electrocardiography for Healthcare Professionals

Supplementary Online Content

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Device Diagnostics and Stroke Prevention: State of the Art

Sinus rhythm with premature atrial beats 2 and 6 (see Lead II).

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

CONSULTA CRT-P Model C4TR01

ICD: Basics, Programming and Trouble-shooting

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER

Case Report Intermittent Right Ventricular Outflow Tract Capture due to Chronic Right Atrial Lead Dislodgement

SARASOTA MEMORIAL HOSPITAL

Pacing Codes and Modes Concepts

GDT1000 SENSING ACUTE STUDY

DON T FORGET TO OPTIMISE DEVICE PROGRAMMING

PARAD/PARAD+ : P and R Based Arrhythmia Detection

His Bundle Pacing: Where is it going? Kenneth A. Ellenbogen, M.D. Kontos Professor, VCU School of Medicine November 17, 2017

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Temporary pacemaker : Programming &Trouble shooting

BHRS Prep course Pub style Quiz NOT A

Chapter 80 Implantable Cardiac Devices

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

6to. Congreso Virtual de Cardiología - 6th. Virtual Congress of Cardiology

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

Electrocardiography for Healthcare Professionals

8/20/2012. Learning Outcomes (Cont d)

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

BEDSIDE ECG INTERPRETATION

I PART I. Timing Cycles and Troubleshooting Review COPYRIGHTED MATERIAL

HRC Carole Joyce. Bradford Royal Infirmary. Senior Chief Cardiac Physiologist. Pacing & Invasive Services.

Contents. Example: Pacemaker in DDD Mode. I. Introduction II. Website and Relevant Documents III. Background Information IV.

Pacemaker Concepts and Terminology*

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

His Bundle Pacing in Bundle Branch Block May 11, 2017

TIP CARD. Model Temporary External Pacemaker

a lecture series by SWESEMJR

Cardiac resynchronisation therapy (biventricular pacing) for the treatment of heart failure

Everything you need to know about His bundle pacing October 20, 2017

Project Title Temporary Pacemaker Training Simulator

Treatment of AF by Pacing Therapy Is there Anything Else Beyond AVN Ablation?

Supplemental Material

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Arrhythmia Study Guide 3 Junctional and Ventricular Rhythms

12 th International Winter Arrhythmia School Collingwood, Ontario February 7, Dr. Derek Yung. MD, FRCPC

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART II

BIPN100 F15 Human Physiology I (Kristan) Problem set #5 p. 1

Automatic assessment of atrial pacing threshold in current medical practice

December D. EL ALLAF Centre Hospitalier Hutois, Huy, Belgium. P. ATTUEL Centre Chirurgical Val d'or, Saint Cloud, France

Troubleshooting ICD. NASPE Training Lancashire & South Cumbria Cardiac Network

Model 5392 EPG Temporary Pacer

Advanced ICD Troubleshooting: Part II REVIEW

TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER)

Pacing troubleshooting. NASPE training Lancashire & South Cumbria Cardiac Network

Ventricular Parasystole

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Update on Cardiac Implanted Electronic Devices (CIEDs)

Antony French Consultant Cardiologist & Electrophysiologist

The Nuts and Bolts of ICD Therapy

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

Dipartimento di Scienze Cardiovascolari Università Campus Bio-Medico di Roma Dott. Vito Calabrese

The ECG Course. Boone County Fire Protection District EMS Education

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Percepta CRT-P MRI SureScan W1TR01

Case Description A 25 years old male served in a combat unit for 3 years (GOLANY) Implantation of Dual chamber ICD in 7/2010 due to Severe Non ischemi

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

Anesthesia Assistants Review Course

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

Pacing Lead Implant Testing. Document Identifier

WHAT S THAT RHYTHM I AM HEARING? GUIDE TO AUSCULTATION OF ARRHYTHMIAS IN HORSES

Introduction to a new "X out of Y" Mode-Switching Algorithm in Inos² CLS and Logos Dual-Chamber Pacemakers

Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG?

Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead e A necessary evil

PACING SYSTEMS.

UNDERSTANDING YOUR ECG: A REVIEW

Dr.Binoy Skaria 13/07/15

Transcription:

Is This *#@!* Thing Working? Pacemaker (and ICD) ECG and Telemetry Pitfalls Wayne O. Adkisson, MD adki0004@umn.edu

Disclosures I currently receive research support from Medtronic, Inc. I have been compensated by Medtronic, Inc for participating in Adverse Events Adjudication Committees (AEAC). There is a conflict of interest plan in place, administered by the University of Minnesota.

Objectives At the end of the presentation participants should be able to discuss 1. How rhythm devices may fail 2. How normal device function may mimic device failure 3. When to worry

How Is the Device Supposed to Work? The earliest pacemakers of the era of modern medicine had only one function: THEY PACED THE HEART.

How Is the Device Supposed to Work? If your patient was attached to a pacemaker that paced at 60 bpm that s all it did pace at 60 bpm.

What s the Only Thing That Can Go Wrong? FAIL TO PACE What could cause this type of early pacemaker to fail to pace the heart? 1. No output 2. Output insufficient to capture the heart

The next innovation was the ability to sense intrinsic cardiac events. Why? To avoid unnecessary pacing.

The next innovation was the ability to sense intrinsic cardiac events.

Pacemaker is programmed in a VVI mode Do the starred pacemaker spikes represent a failure of sensing? Absolutely.

The next step in pacing was AV synchrony, dual-chamber pacing. A DDD pacemaker should: 1. Pace in both chambers 2. Sense in both chambers 3. Sensed events will both inhibit and trigger pacemaker output That gives us four possible combinations of events: 1. AS-VS 2. AP-VS 3. AS-VP 4. AP-VP

Four possible combinations of dual-chamber pacing:

One more thing: What the heck is this? Biventricular pacing

AV synchrony is wonderful but what happened here?

We already saw an example of under sensing in a VVI pacemaker:

This is an example of over sensing in a DDD pacemaker. This is an example of cross-talk. The ventricular lead senses the output from the atrial lead (starred) and assumes what it sensed was an intrinsic QRS and withheld pacing.

Ways a pacemaker can fail: 1. FAILURE TO PACE Tried to pace but couldn t failure of output or of capture Didn t attempt to pace when it should have 2. FAILURE TO SENSE Oversensing results in Underpacing Undersensing results in Overpacing

Back to cross talk What if every atrial output was sensed as a ventricular sensed beat in a pacemaker-dependent patient? Asystole And that s bad

Back to cross talk What if ever atrial output was sensed as a ventricular sensed beat in a pacemaker-dependent patient? Asystole. And that s bad Post-atrial ventricular blanking period Ventricular Safe Pace Window

Too many pacemaker spikes? Failure to sense?

Failure to sense? No, this is an example of Ventricular Safety Pace, our first mimic. How can I tell? The shortened AV delay.

Are there too many pacing spikes? i.e. Is there under sensing? Is this a malfunction? No. This is an example of fusion and pseudo-fusion.

Pacemaker fusion and pseudo-fusion When the pacemaker sees the ventricular depolarization may be after you see it on the ECG.

Pacemaker fusion and pseudo-fusion

Pacemaker fusion and pseudo-fusion

Too many spikes?

Too many spikes?

Too many spikes? Cardiac resynchronization therapy only works if both ventricles are paced 100% of the time. Therefore, if the patient has a cardiac resynchronization therapy (CRT) you may see pacemaker spikes that appear to suggest a failure of sensing during: Atrial fibrillation Premature ventricular beats This is not a malfunction. It is done in an attempt to maintain a high degree of CRT

Too many spikes? CRT devices are designed to maximize RV-LV pacing. That s CRT devices. What about standard, non-biv, pacing? In standard pacing the opposite is true. In standard pacing the goal is to minimize RV pacing. This can lead to situations where it appears the device is failing to pace.

Minimizing RV pacing The most common method of reducing RV pacing is by using a technique called hysteresis deficiency or lagging behind. To a physicist hysteresis is - a retardation of an effect when the forces acting upon a body are changed. For pacing systems hysteresis refers to gradually extending the programmed AV interval.

Too many spikes?

Too many spikes?

What would have happened if the patient had complete heart block?

Too many spikes?

Too few spikes? i.e. Over sensing or failure to pace?

Too few spikes? i.e. Over sensing or failure to pace?

Too few spikes? i.e. Over sensing or failure to pace?

Too few spikes? i.e. Over sensing or failure to pace?

Too few spikes? i.e. Over sensing or failure to pace?

Failure to pace? Often devices will have an algorithm intended to conserve battery life by adjusting the energy output based on automated determinations of the capture threshold.

Failure to pace? Often devices will have an algorithm intended to conserve battery life by adjusting the energy output based on automated determinations of the capture threshold.

How else can these devices confuse me? The lower rate is supposed to be 60 bpm and the heart rate is slower than that. Confirm that the rate that the ECG reports is correct. Is it dark outside? Is the patient asleep?

How else can these devices confuse me? The lower rate is supposed to be 60 bpm and the heart rate is faster than that. Confirm that the rate that the ECG reports is correct. Is the heart rate fast due to pacing? Is rate response programmed on? Is the patient having atrial fibrillation or frequent atrial ectopic beats?

How else can these devices confuse me? The lower rate is supposed to be 60 bpm and the heart rate is faster than that. Is the patient having atrial fibrillation or frequent atrial ectopic beats? Many devices have features that attempt to avoid pauses in the atrial rhythm in an attempt to minimize atrial fibrillation or smooth out the ventricular rate either during AF or when AF terminates.

When NOT to worry? If the pacing rate varies from what the programmed rate is supposed to be. Seeing occasional pacemaker spikes on top of, or inside, PVCs. Seeing occasional pacemaker spikes in the QRS during atrial fibrillation if the device is a CRT device. Even if there is true failure of sensing resulting in unneeded pacing that is rarely an emergency.

When NOT to worry? Even if there is true failure of sensing resulting in unneeded pacing that is rarely an emergency. EXCEPT WHEN IT RESULTS IN THIS:

When NOT to worry? Failure of capture when it is: Followed by pacing with a short AV interval Occurs at fixed intervals 1 min, 2 min, 4 min; every 12 hours; etc

When TO worry? The pause in ventricular pacing is more than twice the programmed lower rate. If the lower rate is 60 bpm, you shouldn t see a failure of pacing that causes the true (not computer reported) heart rate to be less than 30 bpm.

The pause in ventricular pacing is more than twice the programmed lower rate. If the lower rate is 60 bpm, you shouldn t see a failure of pacing that causes the true (not computer reported) heart rate to be less than 30 bpm. The blue bars indicate the current lower interval. The red bars are twice the lower interval or half the lower rate. There is a V paced beat at less than half the lower rate and the AV interval on the paced beat is short. This is a case of DON T worry.

When TO worry? = P wave = ventricular escape beats = paced QRS complexes = conducted beat = pacemaker spikes Is this Managed Ventricular Pacing? NO!

When TO worry? Is this a type of automated threshold testing? NO, again. Threshold testing includes a back-up pulse to ensure capture.

When TO worry? Threshold testing includes a back-up pulse to ensure capture.

Conclusions - I Cardiac Implantable Electronic Devices (CIEDs) are complicated. If the patient is being paced the level of concern is immediately lower. Pacemaker output on time with or shortly after the QRS (or PVC) is usually normal. Pacemaker output during atrial fibrillation is usually normal if the device is one that provides cardiac resynchronization therapy. In general, abnormal sensing is less of a concern. A rare missing pacemaker output, esp if it occurs in a repeating pattern, is usually not a problem.

Conclusions II Ventricular output ( spike ) without a QRS or evidence of back-up output is worrisome. If you aren t sure, call someone. St. Jude Medical Toll-free number: +1-800-722-3774 BIOTRONIK, Inc. Tel +1 800-547-0394 Medtronic, Inc Clinician Services & Technical Support (800) 328-0810 Boston Scientific Cardiac Rhythm Support (800) CARDIAC (1-800-227-3422)