Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

Similar documents
A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

IMRT Planning Basics AAMD Student Webinar

Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Potential benefits of intensity-modulated proton therapy in head and neck cancer van de Water, Tara Arpana

IMAT: intensity-modulated arc therapy

3D-CRT Breast Cancer Planning

Chapter 2. Level II lymph nodes and radiation-induced xerostomia

Chapters from Clinical Oncology

IMRT - Intensity Modulated Radiotherapy

Intensity Modulated Radiation Therapy (IMRT)

A guide to using multi-criteria optimization (MCO) for IMRT planning in RayStation

Radiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Optimization of RapidArc for Head-and-Neck Radiotherapy. Jessica Emily Salazar. Department of Medical Physics Duke University.

Direct machine parameter optimization for intensity modulated radiation therapy (IMRT) of oropharyngeal cancer a planning study

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Master s science Thesis. Evaluation of gantry angle optimization for IMRT treatment planning systems using the Pareto front approach.

Relationship of segment area and monitor unit efficiency in aperture-based IMRT optimization

TomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016

Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida

Dosimetric Effects of Using Generalized Equivalent Uniform Dose (geud) in Plan Optimization

Elekta - a partner and world-leading supplier

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

IMRT IN HEAD NECK CANCER

I. Equipments for external beam radiotherapy

Implementation of advanced RT Techniques

Intensity Modulated RadioTherapy

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10

Int J Clin Exp Med 2015;8(9): /ISSN: /IJCEM Gang Zhou, Yanze Sun, Jianjun Qian, Ye Tian, Xueguan Lu

VOLUMETRIC-MODULATED ARC THERAPY VS. 3D-CONFORMAL RADIOTHERAPY FOR BREAST CANCER

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Defining Target Volumes and Organs at Risk: a common language

Page 1. Helical (Spiral) Tomotherapy. UW Helical Tomotherapy Unit. Helical (Spiral) Tomotherapy. MVCT of an Anesthetized Dog with a Sinus Tumor

Corporate Medical Policy

Knowledge-Based IMRT Treatment Planning for Prostate Cancer: Experience with 101. Cases from Duke Clinic. Deon Martina Dick

Helical Tomotherapy Experience. TomoTherapy Whole Brain Head & Neck Prostate Lung Summary. HI-ART TomoTherapy System. HI-ART TomoTherapy System

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

SBRT TREATMENT PLANNING: TIPS + TRICKS. Rachel A. Hackett CMD, RT(T)

Helical tomotherapy for head and neck squamous cell carcinoma: Dosimetric comparison with linear accelerator-based step-and-shoot IMRT

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

A feedback constraint optimization method for intensity-modulated radiation therapy of nasopharyngeal carcinoma

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Dosimetric Comparison of Intensity-Modulated Radiotherapy versus 3D Conformal Radiotherapy in Patients with Head and Neck Cancer

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

From position verification and correction to adaptive RT Adaptive RT and dose accumulation

Intensity Modulated Radiation Therapy (IMRT)

Applications of Modern Radiotherapy Systems

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

biij Initial experience in treating lung cancer with helical tomotherapy

BLADDER RADIOTHERAPY PLANNING DOCUMENT

Completion of Treatment Planning. Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast

Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT RapidArc TomoTherapy HD. Barbara Agrimson, BS RT(T)(R), CMD

Impact of MLC properties and IMRT technique in meningioma and head-and-neck treatments

BCCR ORIGINAL ARTICLE

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

A STUDY OF PLANNING DOSE CONSTRAINTS FOR TREATMENT OF NASOPHARYNGEAL CARCINOMA USING A COMMERCIAL INVERSE TREATMENT PLANNING SYSTEM

Advances in external beam radiotherapy

Clinical Impact of Couch Top and Rails on IMRT and Arc Therapy

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Evaluation of scatter distribution through segmented intensity modulated radiation therapy (IMRT) fields in treatment of head and neck cancer

Electron Beam ET - Reloaded Therapy - Reloaded

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

Radiation therapy treatment plan optimization accounting for random and systematic patient setup uncertainties

A new homogeneity index based on statistical analysis of the dose volume histogram

REVISITING ICRU VOLUME DEFINITIONS. Eduardo Rosenblatt Vienna, Austria

Utilizzo delle tecniche VMAT nei trattamenti del testa collo Marta Scorsetti M.D.

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

The impact of dose prescription on treatment volume

Clinical experience with TomoDirect System Tangential Mode

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Original Article ABSTRACT

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

The role of Radiation Oncologist: Hi-tech treatments for liver metastases

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Treatment of exceptionally large prostate cancer patients with low-energy intensity-modulated photons

Dose prescription, reporting and recording in intensity-modulated radiation therapy: a digest of the ICRU Report 83

Additional Questions for Review 2D & 3D

First, how does radiation work?

Ranking radiotherapy treatment plans: physical or biological objectives?

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance

A Patient s Guide to SRS

Address for Correspondence: Department of Medical Physics, Khwaja Yunus Ali University, Enayetpur, Sirajgonj ,

Database Plan Quality Impact on Knowledgebased Radiation Therapy Treatment Planning of Prostate Cancer

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Report of ICRU Committee on Volume and Dose Specification for Prescribing, Reporting and Recording in Conformal and IMRT A Progress Report

Transcription:

Investigations and research Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO M. Kunze-Busch P. van Kollenburg Department of Radiation Oncology, Radboud University Nijmegen Medical Centre, Nijmegen, the Netherlands. In the Netherlands, about 2,300 people per year are diagnosed with a malignant tumor in the head and neck region. About 450 of them are referred to the Radboud University Nijmegen Medical Centre. Most of them undergo radiotherapy, either as a sole treatment, or in combination with surgery or chemotherapy. As part of the tailored radiation therapy most patients will be treated with Intensity Modulated Radiotherapy (IMRT). IMRT treatment planning of Head and Neck (H&N) patients can be a complex and timeconsuming task due to concave, irregular target volumes which are close to or even overlap with critical structures such as the parotid glands. The radiation oncologist defines the clinical aims, e.g. the prescribed dose to the target volume(s) and the maximum dose to the critical structures. In the IMRT treatment planning process, these clinical aims are specified mathematically in the form of objective functions. During treatment planning several decisions have to be made: e.g. how many beams should be employed, under which angle should they enter the patient and what type of objectives functions have to be used to find a good balance between tumor coverage and sparing of critical structures. However, in a clinical environment where it is essential for the patient to start treatment shortly after diagnosis, it is important that the planning procedure is straightforward and time-efficient. The goal is therefore to find a standard set of initial conditions for the optimization process, i.e. number of beams and gantry angles, IMRT parameters (such as the number of segments), and planning objectives with which a satisfactory clinical plan can be achieved for most of the patients. IMRT In external beam radiotherapy the tumor is irradiated with shaped beams of ionizing radiation (X-rays or electrons) that are generated by a linear accelerator. The individual shaping of the beams to match the 3D shape of the tumor has become known as 3D Conformal Radiation Therapy (3D-CRT). Although it is called conformal radiotherapy, it does not have the ability to conform the isodose distribution well to concave shaped target volumes. In Intensity Modulated Radiotherapy (IMRT) a more conformal dose distribution can be achieved, to include concave targets, by introducing an additional degree of freedom: the beam intensity across each of the radiation beams. Figure 1 is a schematic representation of the beam intensity profile for one of these beams. IMRT (which is basically an advanced form of 3D-CRT) stepped into the limelight of radiation oncology more than a decade ago, and since then it has become a hot topic among medical physicists and radiation oncologists. Following technical developments by manufacturers of treatment delivery systems and treatment planning systems, the technique has been implemented in many centers around the world, where it is used to treat a variety of cancers such as H&N, brain and prostate cancer. The aim of 1 E Most patients with a malignant head and neck tumor are treated with Intensity Modulated Radiotherapy (IMRT). F Figure 1. An intensity-modulated field that spares the organ at risk (yellow: low dose, red: high dose). In our institute a planning study was performed for patients suffering from oropharyngeal cancer, using the Pinnacle 3 treatment planning system (Philips Medical Systems, Best, the Netherlands). The results are reported below. MEDICAMUNDI 51/2+3 2007/11 95

2 G Figure 2. Fields shaped with an MLC overlap to form an intensity modulated field (yellow: low dose, red: high dose). E IMRT planning with Pinnacle 3 has become less time-intensive and more straightforward. IMRT is to conform the radiation more tightly to the tumor than would be possible with 3D-CRT, so that the total dose to the tumor can be increased while at the same time the dose to the neighboring tissues is minimized. The intensity-modulated beams can be delivered by means of a Multi Leaf Collimator (MLC), which is also used in conventional radiotherapy. The MLC allows the beam to be shaped in a fast and controllable way. It consists of tungsten leaves that can be moved independently of each other to achieve the desired aperture shape. In our clinic, IMRT beams are created by delivering several fields (or segments) with different shapes from the same beam direction (see Figure 2). A mathematical optimization process - so-called inverse planning - is used to determine the shape and intensity of these MLC radiation apertures. An optimal intensity distribution, i.e. one that meets the objectives as closely as possible, is found by dividing the beam into very small beams ( beamlets ) and by varying the intensity of each individual beamlet. However, the optimal intensity distribution then has to be transformed into a deliverable series of MLC segments. This process is called leaf-sequencing or segmentation. Until recently, the treatment planning process comprised several steps for the treatment planner. First an ideal fluence (intensity) map had to be created and the corresponding treatment plan reviewed. If the treatment plan was clinically acceptable, the segmentation process was started. This conversion usually led to a deterioration of the quality of the treatment plan, so that it had to be carefull reviewed once more. Finally, to improve the quality of the treatment plan, a segment weight optimization procedure had to be started, in which the weights of the segments were further optimized. optimization simply with a touch on a button : the conversion of the ideal fluence map into segments takes place during the optimization process. The treatment planner has to decide beforehand after how many iterations the ideal fluence map is to be converted. After the conversion, DMPO continues to optimize the weight and the aperture form of the individual segments by moving the leaf positions (the number of segments stays the same). Now only the final, clinically deliverable treatment plan needs to be reviewed and evaluated. Pinnacle 3 also adds the possibility of creating command files (known as scripts ) with which the beam set-up, additional contours and a set of objectives are set up automatically. This speeds up the planning process. H&N IMRT in Nijmegen In the Radboud University Nijmegen Medical Centre, patients with advanced oropharyngeal tumors are treated with a Simultaneous Integrated Boost (SIB) technique [3]. 50.3 Gy is given to the elective volume and 68 Gy to the boost target volume in 34 treatment sessions (fractions) over 5.5 weeks, delivering one fraction per day during the first 4 weeks and twice daily during the last 1.5 weeks (according to an accelerated scheme). The IMRT treatment ( step-and shoot ) is applied using Elekta SLi-Precise machines (Elekta, desktop v. 5) with the Oncentra -Visir Record & Verify Software (Nucletron, v. 2.1). Step by step to the class solution To keep the planning procedure time-efficient and straightforward, a standard set of beams and objectives has to be found. 96 MEDICAMUNDI 51/2+3 2007/11 Inverse planning with DMPO With the introduction of the DMPO (Direct Machine Parameter Optimization) algorithm [1], IMRT planning with Pinnacle 3 has become less time-intensive [2] and more straightforward. The treatment planner can start the whole The number of beams In an earlier planning study (before working with DMPO), 5-, 7-, and 9-beam set-ups were compared. The 7-beam set-up was found to be a good compromise between plan quality and treatment time on the linear accelerator.

Now seven uniformly spaced coplanar beams are routinely used. 3 The number of segments The allowed maximum number of segments was varied (40, 60, 80, 100) to find a good compromise between quality and feasibility. More than 60 segments did not lead to significant improvements in the treatment plans [4], but would result in a longer time for the patient on the treatment couch. Currently a maximum number of 60 segments is employed. Before starting the optimization process, due consideration has to be given to the use of additional contours which help to sculpt the dose distribution and subsequently simplify the planning process. Additional contours Organ At Risk (OAR) Figure 4a shows a transverse slice with the clinical Regions of Interest (ROI). Next to it, Figure 4b shows the same slice with additional contours added to guide the optimization process. Any overlapping ROI s are separated to avoid conflicting objectives. For example, the Planning Target Volume (PTV) is subtracted from the parotid gland and the resulting parotid gland minus PTV ROI (Figure 4b) is designated as a separate objective. To achieve an additional sparing of the parotid gland further away from the PTV, a third contour for the parotid gland is created. This extra parotid contour is the same as parotid gland minus PTV but with an extra margin towards the PTV to allow for the application of a more strict objective. Additional contours - Planning target volume To further guide the optimization and conform the dose to the target volume, additional ROI s known as guiding volumes are created. These may be adjusted after the first optimization run. An inner ring is created for the elective volume (red) as well as the boost target volume (blue) (Figure 4b). These inner rings receive a minimum dose objective to prevent an underdosage at the periphery of the PTV. An outer ring (light blue) is also created for the boost target volume in order to avoid a spreading of high dose into the elective target volume. The elective volume is split into a cranial and a caudal part (Figure 5) to help keep the dose in the caudal part between 95% and 107% of the prescribed dose, as required by the ICRU. Objectives Different sets of objective functions for target volumes and critical structures were tested, including dose, dose-volume (DV) and Equivalent Uniform Dose (EUD) based objectives [5]. For example, a minimum (or maximum) dose objective tries to keep the dose in the tumor (critical structure) above (or below) a specified value. The set of objectives below resulted in clinically acceptable treatment plans for all 25 patients G Figure 3. The 7-beam set-up offers a good compromise between plan quality and treatment time. Figure 4. Transverse slice showing Regions of Interest (ROI). 4a 4b guiding volume F F Figure 4a. Transverse slice showing the elective volume (outlined in red), the boost volume (outlined in blue) and the parotid glands (outlined in green). F Figure 4b. Transverse slice showing additional contours in colorwash: parotid gland minus PTV (green), inner ring elective volume (red), inner ring (blue) and outer ring (light blue), boost volume and guiding volumes (yellow). MEDICAMUNDI 51/2+3 2007/11 97

E Figure 5. The elective volume is split into cranial and caudal parts. The red colorwash represents, the orange colorwash represents PTV>cran minus ( + ring outside). 5 ROI ring inside Objective E E Table 1. Set of objectives currently used for oropharynx patients. PTV < boost volume, PTV > elective volume, Ex-ROI external minus the other ROI s ring outside Max DVH studied and it is now used for all our clinical patients. The exact dose levels and weights still require some fine-tuning during the planning process and are therefore not shown. Of course, this set of objectives is not the only one to produce satisfactory treatment plans. Others may produce similar results as well. However, it is a set that works well in our clinic with the given combination of planning system, treatment technique, treatment equipment, delineation and evaluation. PTV> ring inside Larynx - PTV Parotid gland- PTV Brain stem Spinal cord Guiding volumes Ex -ROI Max DVH E Figure 6. Dose distributions for four different patients planned with the set of objectives shown in Table 1. 6a 6b 6c 6d 98 MEDICAMUNDI 51/2+3 2007/11

Results For 25 consecutively treated clinical patients this set of objectives has turned out to be adequate to obtain a satisfactory plan. Due to the standardized set of beams, objectives, IMRT parameters and additional ROI s, the whole planning process can be done in one day, and hence does not require more planning time than a conventional plan. With the additional use of scripts the planning time can even be shortened. Conclusions With the DMPO module in the Philips Pinnacle 3 treatment planning system, inverse planning for IMRT has become more straightforward and less time intensive. After initially investing time in finding a good set of objectives, oropharynx treatment planning with DMPO can be done in the same time period as conventional planning. This allows the clinical introduction of IMRT for all oropharynx patients and raises the standard of care for the patients K References [1] Hårdemark B, Liander A, Rehbinder H, Loef J, Robinson D: P 3 IMRT. Biological Optimisation and EUD. Pinnacle White Paper No. 4535 983 02482, Philips 2004. [2] S. Tung et al. Clinical Evaluation of Direct Machine Parameter Optimization Algorithm for Head and Neck IMRT Treatment, Medical Physics 2005; 32: 1971. [3] R. Mohan et al. Radiobiological Considerations in the Design of Fractionation Strategies for Intensity-Modulated Radiation Therapy of Head and Neck Cancers. Int J Radiat Oncol Biol Phys 2000; 46: 619-30. [4] P.v. Kollenburg et al. Feasibility and Quality of Head & Neck SIB- IMRT Techniques - A Planning Study with the New Optimizer in Pinnacle (DMPO) for an Elekta Linac. Radiotherapy and Oncology 2005; 76, Suppl.2: S189-S190. [5] Q. Wu et al. Optimization of Intensity-Modulated Radiotherapy Plans based on the Equivalent Uniform Dose. Int J Radiat Oncol Biol Phys 2002; 52: 224-235. MEDICAMUNDI 51/2+3 2007/11 99