Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Similar documents
Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Determination of Polymerized Triglycerides by High Pressure Liquid Chromatography and Corona Veo Charged Aerosol Detector

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Determination of Olive Oil Adulteration by Principal Component Analysis with HPLC Charged Aerosol Detector Data

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Characterization of Used Cooking Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Impurity Profiling of Carbamazepine by HPLC/UV

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Novel, Universal Approach for the Measurement of Natural Products in a Variety of Botanicals and Supplements, Part 2

Direct Analysis of Multicomponent Adjuvants by HPLC with Charged Aerosol Detection

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Fast UHPLC Methods for Analysis of Amino Acids

Mercury Speciation Determinations in Asian Dietary Supplements

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

LC-MS/MS Method for the Determination of Tenofovir from Plasma

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Increased Identification Coverage and Throughput for Complex Lipidomes

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Determination of Phenolic Compounds in Apple Orchard Soil

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

LC-Based Lipidomics Analysis on QTRAP Instruments

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Glycerolipid Analysis. LC/MS/MS Analytical Services

Determination of Amantadine Residues in Chicken by LCMS-8040

The HPLC Preparative Scale-Up of Soybean Phospholipids Application

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Identification and Quantitation of Microcystins by Targeted Full-Scan LC-MS/MS

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Thermo Scientific Dionex CarboPac MA1 Column

Analysis of Commercially Available Products Containing Stevia

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Determination of Tigecycline in a Cell Lysate

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses

Lipid Analysis by Evaporative Light Scattering Detection

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Sensitive Analysis of Commonly Used Artificial and Natural Sweeteners Including Stevia and Their Impurities and Degradation Products

Lipid Analysis. Andréina Laffargue, IRD CRYMCEPT Montpellier workshop, October 17th Introduction to lipid structures

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Determination of Steviol Glycosides by HPLC with UV and ELS Detections

Results of Using Re-washed Vials and Closures

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Quantitative Method to measure Glycidol Fatty Acid Esters (GEs) in Edible Oils

Application Note. Authors. Abstract. Petrochemical

Application Note. Agilent Application Solution Analysis of fat-soluble vitamins from food matrix for nutrition labeling. Abstract.

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Transcription:

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: To develop HPLC methods for quantitation of phospholipids in natural products, including foods and ingredients, using an HPLC system with a charged aerosol detector. Methods: One normal-phase HPLC method, using the Thermo Scientific Hypersil GOLD Silica column, was created for the deteration of six phospholipids and was used to quantify phospholipids in foods and ingredients. Results: Samples of foods or ingredients were analyzed for content of phospholipids using the developed method. Sensitivity for the phospholipids varied from limits of detection between 10 22 ng o.c. Introduction Phospholipids are a broad class of lipids that can be divided into glycerophospholipids (GPLs) and sphingolipids. Both groups show great structural diversity. Phospholipids are amphiphilic molecules, having a hydrophilic head group, and a lipophilic fatty acid tail. Several families of GPLs exist biologically, differing in the type of polar head group present, and include: phosphatidylcholine (PC), phosphatidylethanolae (PE), phosphatidylserine (PS), phosphatidic acid (PA), cardiolipin, and phosphatidylinositol (PI). Each compound contains many species resulting from differences in their fatty acid composition. For example, PC may contain several different fatty ligands, which will result in mulitple peaks by reversed phase chromatography. Normal phase liquid chromatography (NP-HPLC) uses differences of polar moieties to separate analytes, and use of NP-HPLC will provide more quantitative data with less effort. Other diacyl phospholipids (e.g, dipalmitoylphosphatidylcholine (DPPC) and dipalmitoylphosphatidylethanolae (DPPE)) would exhibit similar properties. Lysophosphatides (e.g., lysophophatidylcholine (LPC)) are produced from the action of phospholipase enzymes, which removes the C2 fatty acyl side-chain. The Thermo Scientific Dionex Corona Veo Charged Aerosol Detector, a sensitive mass-based detector, is ideally suited for the direct measurement of phospholipids, as they are non-volatile and non-chromophoric compounds. It offers excellent sensitivity (down to low nanogram amounts on column), a dynamic range of over 4 orders of magnitude, and similar inter-analyte response independent of chemical structure. This developed method is based on an original publication by Rombaut, R., et al., (J. Dairy Sci., 2005, 88, 482), that enables the direct measurement of a number of GPL and SL species, each as near-single peaks. As shown in Figure 1, the detector uses nebulization to create aerosol droplets. The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogramlevel sensitivity. This technology has greater sensitivity, dynamic range and precision than ELSD and refractive index (RI), is gradient compatible and is simpler to operate than a mass spectrometer (MS). Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS). This sensitivity, combined with the linearity that is possible with use of the Corona Power Function, provides a unique and complete analytical solution for sensitive, reproducible, and routine analysis of non-chromophoric analytes. FIGURE 1. Schematic and functioning of charged aerosol detection. 1 3 10 2. 4 5 8 6 7 9 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charged gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge is measured by a highly sensitive electrometer 10. Signal transferred to chromatographic software 2 Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Methods Liquid Chromatography HPLC System: Thermo Scientific Dionex UltiMate LC System with LPG3400-SD pump (normal phase), WPS3000 RS autosampler, and TCC3000 RS column oven HPLC Column: Thermo Scientific Hypersil Silica 3 µm, 3.0 150 mm Column Temperature: 50 C Mobile Phase A: 0.5% diethylae-formate, ph 3.0 Mobile Phase B: 2-Propanol Mobile Phase C: iso-octane Flow Rate: 0.2 0.8 ml/ Injection Volume: 2 5 µl Detector: Corona Veo SD Charged Aerosol Detector Evaporator Temperature: 50 C Data rate: 10 Hz Filter: 3.6 s PowerFunction: 1.40 Flow Gradient: Time Flow (utes) (ml/) %A %B %C -4.0 0.8 1 64 35-0.5 0.8 1 64 35-0.2 0.2 1 64 35 0.0 0.2 1 64 35 0.1 0.8 1 64 35 2.0 0.8 4 61 35 7.0 0.8 10 60 30 13 0.8 10 60 30 14 0.8 1 64 35 Standard and Sample Preparations Standards were dissolved in methanol / chloroform (1:1), at a concentration of approximately 2.00 mg/ml, including the sodium salt for the PI and PS analytes. Standards were diluted sequentially by half to provide a calibration range of 78 10,000 ng o.c. of each analyte. Egg Yolk Sample: In a 2 ml glass vial, 90.3 mg yolk was placed. To this, 150 ul of methanol / chloroform (1:1) was added. The mixture was vortex-mixed. To precipitate proteins, 500 µl of acetonitrile was added and the mixture was vortex-mixed. Another 1,350 µl of methanol / chloroform (1:1) was added, vortex-mixed, and allowed to settle. The supernatant was centrifuged at 10,000 g for 3 utes, and the supernatant was analyzed directly. Lecithin: To an HPLC vial, 1.1 mg of Lecithin (Laboratory Grade, Fisher Scientific) was added. The sample was dissolved in 1100 µl of methanol / chloroform (1:1) and centrifugefiltered through a 0.2 micron filter, 10,000 g for 3 utes Supernatant was analyzed directly. Krill Oil: To an HPLC vial, 2.50 mg of krill oil was added. To this, 1000 µl of methanol/chloroform (1:1) was added. The solution was vortex-mixed and analyzed directly. Data Analysis All HPLC chromatograms were obtained and compiled using Thermo Scientific Dionex Chromeleon Chromatography Data System software, 6.8 SR 13. Thermo Scientific Poster Note PN71006_PITTCON_2014_E_03/14S 3

Results Calibrations Five phospholipids were used to calibrate the method for sample analysis: PE, PI, PS, PC, and LPC. A chromatogram of the standard solution at approximately 5000 ng o.c. is shown in Figure 2. Each analyte solution was injected threefold, plotted, and fit to a linear regression, as shown in Figure 3. As can be seen from the plot, all plots were linear, and four of the five analytes had nearly the same response factor, which is a typical characteristic of charged aerosol detectors with non-volatile analytes. Other phospholipids, PA, PG, and SPH were not included in these analyses. PA and PG eluted at the retention time of 4.8 utes, or the small peak seen just after PE, in Figure 2. The correlation coefficients, R 2, for all analytes was greater than 0.995 across the entire dynamic range, from 78 to over 10,000 ng. o.c. Along with these values, the sensitivity results, based on signal to noise calculations for LOD and LOQ, are provided in Table 1. Figure 2. HPLC-CAD chromatogram overlays of phospholipid standards at 5000 ng o.c. in triplicate. 200 1 - PE - 4.393 150 3 - PI - 5.533 100 4 - PS - 6.883 6 - PC - 9.407 7 - LPC - 12.163 50 2-4.803 5-7.533-20 Figure 3. Linear regression calibrations curves for PE, PI, PS, PC, and LPC, from 78 10,000 ng o.c., each in triplicate. 70 60 50 40 30 20 10 0 0 2000 4000 6000 8000 10000 12000 14000 Amount on column (ng) PE PI PS PC LPC Table 1. Calibration and Sensitivity for PE, PI, PS, PC, and LPC for 78 10,000 ng o.c. Phospholipid R 2 LOD (ng o.c.)* LOQ (ng o.c.)** PE 0.9991 14 42 PI 0.9973 14 42 PS 0.9988 22 67 PC 0.9955 10 31 LPC 0.9974 14 41 * LOD calculated at signal to noise ratio of 3.3, at lowest on column amount for each analyte. ** LOQ calculated at signal to noise ratio of 10, at lowest on column amount for each analyte. 4 Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Sample Analysis Egg Yolk The sample of egg yolk was analyzed directly, as prepared, with a 2 µl injection volume. The resulting chromatogram was clean of interfering peaks, as shown in Figure 4. The compositional results 1 match literature values (obtained by extraction and 31 P-NMR) very well, as provided in Table 2. The actual amount of phospholipids differed from a quantitative study, 2 possibly due to differences between eggs and / or the extraction efficiency of the method used to clean the sample. Figure 4. HPLC-CAD chromatogram showing phospholipids found in egg yolk. Figure 6. HPLC-CAD ch 200 Egg yolk 150 3 - PC - 9.404 100 1 - PE - 4.384 50 6 - LPC - 12.214 2 - PI - 5.557 4-9.990 5-11.917-5 Table 2. Results of egg yolk analysis compared to literature values, including total, relative composition of total phospholipids (PL). Phospholipid Literature (PL-%) 1 Found (PL-%) PE 19.86 18.9 PI 1.73 2.22 PS 0.0 0.0 PC 75.40 75.6 LPC 0.04 0.92 Total 2 97.03 mg/g 86.6 mg/g Sample Analysis Lecithin (Laboratory Grade) The lecithin sample was analyzed directly, with chromatogram shown in Figure 5, finding PE, possibly PA (peak at 4.59 utes), PI, PS, PC. The LPC peak retention time was obscured by matrix peak. Composition and total phospholipids results are provided in Table 3. Table 4. Results of krill and relative compositio Phospholipid PE PI PS PC LPC Total * AOCS Official Method J Figure 5. HPLC-CAD chromatogram showing phospholipids found in lecithin. 90 Lecithin 1.0 mg/ml 60 1 - PE - 4.363 6 - PC - 9.410 40 3 - PI - 5.517 20 2-4.593 5 - PS - 7.373 4-6.777-10 Table 3. Results of laboratory-grade lecithin analysis, including total phospholipids and mass-% of each phospholipid relative to amount of lecithin. Phospholipid Found (mass-%) PE 25.2 PI 17.0 PS 1.2 PC 33.6 Total 792 mg/g Thermo Scientific Poster Note PN71006_PITTCON_2014_E_03/14S 5

Sample Analysis Krill Oil The krill oil sample was analyzed directly, with chromatogram shown in Figure 6, finding the major phospholipids present, mainly PC, PE, and LPC and the or PI and PS phospholipids. Relative phospholipid composition and total phospholipids results are provided in Table 4. The values obtained matched the AOCS method results within 15%. Figure 6. HPLC-CAD chromatogram showing phospholipids found in krill oil. 100 Krill Oil 2.5 mg/ml 4 - PC - 9.424 80 60 40 20 1 - PE - 4.407 3 - PS - 7.404 2 - PI - 5.644 5 - LPC - 12.244-5 Table 4. Results of krill oil analysis compared to literature values, including total and relative composition of total phospholipids. Phospholipid AOCS (PL-%) * Found (PL-%) PE 4.6 3.5 PI -- 0.14 PS -- 1.05 PC 81.2 86.7 LPC 10.6 8.64 Total 43.0 mg / g 44.1 mg/g * AOCS Official Method Ja 7c-07 3 Conclusions spholipids found in lecithin. A method for the deteration of phospholipids in various samples was developed using normal phase HPLC and the Corona Veo Charged Aerosol Detector. Sensitivities for the phospholipids were < 25 ng o.c. LOD. Linear calibration curves were found for all five phospholipid analytes, with correlation coefficients > 0.995 over four orders of magnitude. - PC - 9.410 Phospholipids compositional results matched orthogonal results very well, and quantity amounts were similar, whereby differences were more likely due to sample preparation rather than chromatographic method conditions. References 10.0 11.0 12.0 13.0 14.0 15.0 16.0 lysis, including total d relative to amount of lecithin. ound (mass-%) 25.2 17.0 1.2 33.6 792 mg/g 1. El Bagir, N.M.; et al. Lipid Composition of Egg Yolk and Serum in Laying Hens Fed Diets Containing Black Cu (Nigella sativa). International Journal of Poultry Science 2006, 5(6), 574 578. 2. Walker, L.A.; Wang, T.; Xin, H.; Dokle, D. Supplementation of Laying-Hen Feed with Palm Tocos and Algae Astaxanthin for Egg Yolk Nutrient Enrichment. J. Agric. 6 Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752 PN71006_E 03/14S