Subjects and Methods 2017, 64 (8),

Similar documents
Comparison of glucose monitoring between Freestyle Libre Pro and ipro2 in patients with diabetes mellitus

Successful glycemic control using a flash glucose monitoring system for a pregnant woman with diabetes: a case report

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Flash Glucose Monitoring & Implantable Sensors

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

University of Groningen

Long-term effects of continuous glucose monitoring on HbA 1c levels: An audit

Objective: Established methods of self-monitoring of glucose levels include capillary self-monitoring of blood glucose (SMBG) and

Advances Towards the Bionic Pancreas.

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively

Diabetes Technology Update. Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018

Clinical Value and Evidence of Continuous Glucose Monitoring

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

ORIGINAL ARTICLE. Use of Freestyle Libre Pro TM Flash Glucose Monitoring System in Different Clinical Situations at a Diabetes Centre

Limitations of Conventional Methods of Self-Monitoring of Blood Glucose

Commissioning Policy

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

In response to an enquiry from the Scottish Diabetes Group (SDG)

Continuous Glucose Monitoring (CGM)

CADTH ISSUES IN EMERGING HEALTH TECHNOLOGIES

Analysis for the Improvement of Inadequate Glycemic Control in Patients with Type 2 Diabetes Mellitus in Nagano, Japan

Incorporating CGM Into Clinical Decision Making. Etie Moghissi, MD, FACE Clinical Associate Professor, David Geffen School of Medicine UCLA

Advances in Diabetes Care Technologies

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Review Article. Continuous Glucose Monitoring - A Step Towards Better Diabetes Management. Ajay V Kaduskar

The MiniMed Continuous Glucose Monitoring System. JOHN J. MASTROTOTARO, Ph.D.

Differences in Glycemic Variability Between Normoglycemic and Prediabetic Subjects

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

When Will CGM Replace SMBG? Roy W. Beck, MD, PhD. JAEB Center for Health Research Tampa, Florida

Usefulness of Ambulatory Glucose Profile (AGP) in Diabetes Care

Sponsor / Company: Sanofi Drug substance(s): insulin glargine (HOE901) According to template: QSD VERSION N 4.0 (07-JUN-2012) Page 1

Akio Ohta, Kaori Arai, Ami Nishine, Yoshiyuki Sada, Hiroyuki Kato, Hisashi Fukuda, Shiko Asai, Yoshio Nagai, Takuyuki Katabami and Yasushi Tanaka

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia

Quick Reference Guide

PREVALENCE AND EXTENT OF GLYCEMIC EXCURSIONS IN WELL- CONTROLLED PATIENTS WITH TYPE 2 DIABETES MELLITUS USING CONTINUOUS GLUCOSE-MONITORING SYSTEM

Systematic Review of the Effectiveness of Continuous Glucose Monitoring (CGM) on Glucose Control in Type 2 Diabetes

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Abstract CLINICAL APPLICATIONS. Objectives: Methods: Results: Conclusion: Journal of Diabetes Science and Technology

Supplementary Data. Formula S1. Calculation of IG fluctuation from continuous glucose monitoring profiles Z T 1 T

Figure 2.1: Glucose meter

Quick Reference Guide

Continuous Glucose Monitoring Devices Pharmacy Policy

1. Continuous Glucose Monitoring

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes

Quick Reference Guide

Description of the technology

Advances in Diabetes Care Technologies

Making Sense of Glucose Monitoring. My Journey with Glucose Monitoring Over the Last 37 Years 8/7/2017

Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

Flash Glucose Monitoring (Flash GM) Frequently asked questions (November 2018)

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M.

Dr Sander Zwart MD, FACP, ECNU Arizona-AACE 2016 Annual Meeting

Supplementary Appendix

Glycemic Variability:

Continuous Glucose Monitoring: The Future of Diabetes Management

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier?

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

The transition from urinary

Nocturnal hypoglycemia is frequent in patients with

Update on Continuous Glucose Monitoring (CGM) Technology in Diabetes. Elena Toschi, MD November 12, 2016

DIABETES TECHNOLOGY: WHERE ARE WE NOW AND WHERE ARE WE GOING? Presented by: Tom Brobson

Continuous Glucose Monitoring System

Efficacy/pharmacodynamics: 85 Safety: 89

Continuous Glucose Monitoring (CGM)

The FreeStyle Libre Flash Glucose Monitoring System. Healthcare Professional Training - London

Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna. Ipoglicemie e Monitoraggio Glicemico

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

Performance-powered. The OneTouch. Ping insulin pump and meter-remote.

The Realities of Technology in Type 1 Diabetes

Early diagnosis, early treatment and the new diagnostic criteria of diabetes mellitus

Continuous Glucose Monitoring System

Commissioning Statement

Better Diagnostics for Life

Continuous Glucose Monitors for Diabetes Management

Continuous Glucose Monitoring System

Artificial Pancreas Device System (APDS)

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

REPORT INTERPRETATION

Sponsor / Company: Sanofi Drug substance(s): Insulin Glargine. Study Identifiers: NCT

Advances in Diabetes Care Technologies

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE

For Safe and Effective Diabetes Management

Efficacy of Combination Therapy With Sitagliptin and Low-Dose Glimepiride in Japanese Patients With Type 2 Diabetes

Diabetes Care Publish Ahead of Print, published online September 11, 2007

Test strip reading - Cats Negative Trace

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

Diabetes Care Publish Ahead of Print, published online April 7, 2017

FreeStyle Libre Report Overview

Continuous Glucose Monitoring

ROLE OF CONTINUOUS GLUCOSE MONITORING IN DIABETES TREATMENT

Continuous or Intermittent Glucose Monitoring (CGMS) in Interstitial Fluid Corporate Medical Policy

Since the discovery of insulin, the management of type 1 diabetes

Making the Most of Continuous Glucose Monitoring

Transcription:

2017, 64 (8), 827-832 Careful readings for a flash glucose monitoring system in nondiabetic Japanese subjects: individual differences and discrepancy in glucose concentrarion after glucose loading Keiko Sekido 1), Takashi Sekido 1), Atsuko Kaneko 1), Manami Hosokawa 1), Ai Sato 1), Yoshihiko Sato 1), Masanori Yamazaki 1), 2) and Mitsuhisa Komatsu 1) 1) Department of Diabetes, Endocrinology and Metabolism, Division of Internal Medicine, Shinshu University School of Medicine, Matsumoto, Japan 2) Department of Drug Discovery Science, Shinshu University School of Medicine, Matsumoto, Japan Abstract. The FreeStyle Libre Flash Glucose Monitoring System (FGM), which can continuously measure glucose concentration in the interstitial fluid glucose (FGM-ISFG), has been in clinical use worldwide. However, it is not clear how accurately FGM-ISFG reflects plasma glucose concentration (PG). In the present study, we examined the clinical utility of FGM by oral glucose tolerance test (OGTT). In eight healthy volunteers (3 males; mean age, 41.8 y) wearing FGM sensors for 14 days, OGTT was performed during days 1 7 and days 8 14, and then both FGM-ISFG and PG were compared. Parkes error grid analysis indicated that all of 65 FGM-ISFG values were within Zone A (no effect on clinical action) and Zone B (little or no effect on clinical outcome). However, in OGTT, the mean FGM-ISFG was higher than the mean actual PG at 30, 60, and 90 minutes after loading (155.5 vs. 139.2 mg/dl, 166.2 vs. 139.2 mg/dl, 149.5 vs. 138.2 mg/dl, respectively; p<0.05). Moreover, the area under the curve of FGM-ISFG was also significantly larger than that of PG (17,626.2 vs. 15,195.0 min mg/dl; p<0.05). In four of eight subjects, FGM-ISFG tended to be higher than PG in both OGTTs, and the greatest difference between the two values was 58 mg/dl. FGM is useful for glycemic control, whereas it is not appropriate to change therapeutic regimens based on the judgment of nocturnal hypoglycemia and postprandial hyperglycemia by FGM-ISFG. Careful attention is required for proper application of FGM. Key words: Flash glucose monitoring, Interstitial fluid glucose GLUCOSE MEASUREMENT using capillary samples is useful for blood glucose control in patients with diabetes mellitus. Recently, continuous glucose monitoring system (CGM), which can provide information on blood glucose levels by continuously measuring interstitial fluid glucose concentration (ISFG), is one of the most commonly used tools to detect nocturnal hypoglycemia and postprandial hyperglycemia. Although there is a time lag between fluctuation in ISFG and that in blood glucose level, ISFG is generally thought to be an acceptable indicator for blood glucose control [1]. In conventional CGM, it is necessary to measure self-monitoring capillary glucose concentrations 3 4 times a day for calibration of ISFG. However, the FreeStyle Libre Flash Glucose Monitoring System Submitted May 10, 2017; Accepted Jul. 13, 2017 as EJ17-0193 Released online in J-STAGE as advance publication Jul. 25, 2017 Correspondence to: Masanori Yamazaki, M.D., Ph.D., Department of Diabetes, Endocrinology and Metabolism, Division of Internal Medicine, Shinshu University School of Medicine, 3-1-1 Asahi, Matsumoto 390-8621, Japan. E-mail: macha@shinshu-u.ac.jp The Japan Endocrine Society (FGM), which can monitor ISFG for 14 days with only attachment of a small sensor patch to the upper arm without calibration, was recently approved for use in Japan. The system is so easy to handle that it can be widely used in clinical settings requiring glycemic control in prediabetic and diabetic subjects [2]. However, little information is available regarding the how accurately ISFG measured by FGM (FGM- ISFG) reflects plasma glucose concentration (PG). Therefore, we examined the clinical utility of FGM in non-diabetic subjects by comparing FGM-ISFG with PG, followed by oral glucose tolerance test (OGTT). Subjects and Methods Eight healthy volunteers without diabetes participated in this study. This study was approved by the Shinshu University Medical Ethics Committee, and written informed consent was obtained from all subjects. FGM was precisely attached according to the package insert and ISFG was monitored for 14 days.

828 Sekido et al. ISFG is measured automatically every 10 minutes and can be displayed on a remote device when necessary. For estimation of FGM performance, 75-g OGTT was conducted during the first period from days 1 7 after installing FGM and a second from days 7 14 day. Venous blood was sampled at 0, 30, 60, 90, and 120 minutes for PG measurements, and FGM-ISFG was recorded simultaneously. PG was measured by the glucose oxidation method. Unfortunately, some problems with FGM occurred in three subjects: in case 2, a scan failure on day 10 led to cancellation of the second OGTT; in case 4, a new sensor was reattached and then remeasured because FGM-ISFG was continuously low, and greatly deviated away from the capillary blood glucose level; in case 8, FGM-ISFG was always about 80 100 mg/dl higher than the blood glucose from SMBG, exceeding 200 mg/dl with no relation to meals, which resulted in cancellation of OGTT. Therefore, a total of 13 OGTTs were performed in seven subjects. Using the Japan Diabetes Society criteria [3], participants were categorized as having the following glucose tolerance based on fasting plasma glucose levels (FPG) and 2-h post-load plasma glucose levels (2-h PG) in 75-g OGTT: normal glucose tolerance, NGT (FPG <110 mg/dl and 2-h PG <140 mg/ dl); impaired glucose tolerance, IGT (FPG <126 mg/ dl and 2-h PG 140 199 mg/dl; diabetes mellitus, DM (FPG 126 mg/dl and/or 2-h PG 200 mg/dl). Parkes error grid analysis was performed to evaluate the consistency of FGM-ISFG with actual PG. In this analysis, actual PG and FGM-ISFG were plotted on the abscissa and ordinate, respectively. The zones in the graph indicate clinical usefulness as follows: Zone A: no effect on clinical action; Zone B: altered clinical action (little or no effect on clinical outcome); Zone C: altered clinical action (likely to affect clinical outcome); Zone D: altered clinical action. Next, we analyzed the individual data. The differences between FGM-ISFG and PG were examined at 0, 30, 60, 90, and 120 minutes. Pearson s product-moment correlation coefficient was calculated to determine the correlation between FGM-ISFG and PG. The Mann Whitney U test was used to detect statistical significance of differences, and p<0.05 was taken to indicate significance. Results The seven participants (3 men and 4 women) had a mean age of 41.4 years (range: 25 56 years). Four were judged as having normal glucose tolerance, and the others had impaired glucose tolerance (Table 1). There was a significant and strong correlation between FGM-ISFG and PG obtained from OGTT in all participants (r=0.850, p<0.001). The following correlation formula was determined: FGM-ISFG (mg/dl) = 24.942 + 0.94707 Actual PG (mg/dl) As shown in Fig. 1, plotting actual PG and FGM- ISFG by the Parkes error grid method indicated that 64 (98.5%) of 65 data sets fell within Zone A, and the remaining one data set (1.5%) fell within with Zone B. There were no FGM-ISFG values in Zones C and D. The overall result of OGTT is shown in Fig. 2. FGM-ISFG was significantly higher than PG at 30, 60, and 90 minutes after glucose loading (155.5 vs. 139.2 mg/dl, 166.2 vs. 139.2 mg/dl, 149.5 vs. 138.2 mg/ dl, respectively; p<0.05). The area under the curve of FGM-ISFG was also significantly larger than that of PG (17,626.2 vs. 15,195.0 min mg/dl; p<0.05). Overall mean absolute difference (MAD) and mean absolute relative difference (MARD) between FGM-ISFG and actual PG were 19.6±10.0 mg/dl and 17.1±10.0%, respectively. MAD and MARD in each OGTT were indicated in Table 1. MAD ranged from 5.4 to 37.0 mg/dl and MARD also varied from 3.5 to 33.5%. The differences between FGM-ISFG and actual PG in each subject are shown in Fig. 3. The mean values of Table 1 Mean absolute difference and mean absolute relative difference between interstitial fluid glucose and actual plasma glucose Case Age/Sex Glucose tolerance status MAD (mg/dl) MARD (%) 1 51/M IGT 16.4 9.9 5.4 3.5 2 25/F NGT 5.4 5.2 N.A. N.A. 3 27/F NGT 22.2 20.9 27.0 31.7 4 47/F NGT 27.2 27.1 37.0 33.5 5 56/M IGT 24.0 19.3 29.6 22.0 6 51/F IGT 25.8 20.9 16.0 13.0 7 33/M NGT 8.6 6.0 10.8 9.0 MAD, mean absolute difference; MARD, mean absolute relative difference; IGT, impaired glucose tolerance; NGT, normal glucose tolerance; N.A., not available.

Accuracy of flash glucose monitoring 829 Fig. 1 Parkes error grid analysis The validity of FGM was determined by examining the Parkes error grid on which actual PG and FGM-ISFG data in OGTTs were plotted. All data fell within Zones A and B. FGM, flash glucose monitoring system; PG, plasma glucose concentration; FGM-ISFG, interstitial fluid glucose concentration by FGM. Fig. 2 Overall results of OGTTs FGM-ISFG was significantly higher than PG at 30, 60, and 90 minutes after glucose loading. All data are expressed as means ± standard deviation (SD). * p<0.05. OGTT, oral glucose tolerance test; FGM- ISFG, interstitial fluid glucose concentration by flash glucose monitoring; PG, plasma glucose concentration. Fig. 3 Differences between FGM-ISFG and actual PG in OGTTs Over half of the subjects showed higher FGM-ISFG than actual PG. OGTT, oral glucose tolerance test; FGM-ISFG, interstitial fluid glucose concentration obtained by flash glucose monitoring; PG, plasma glucose concentration; NGT, normal glucose tolerance; IGT, impaired glucose tolerance; FGM-ISFG, interstitial fluid glucose concentration obtained by flash glucose monitoring; PG, plasma glucose concentration; N/A, not available.

830 Sekido et al. FGM-ISFG and PG obtained from two OGTTs were used in this analysis. In Cases 2 5, FGM-ISFG tended to be higher than PG in both OGTTs. The greatest difference between FGM-ISFG and actual PG was 58 mg/dl at 60 minutes after glucose loading in Case 4. Furthermore, a line graph of all measured FGM-ISFG and actual PG revealed that FGM-ISFG maintained higher levels than actual PG regardless of a delayed change of ISFG (Fig. 4). In Case 7, FGM-ISFG was higher than PG in the initial OGTT, but the two values were similar in the second OGTT. In Case 1, both values in the OGTTs were almost identical. Discussion FGM provides accurate measurements of ISFG corresponding to capillary PG reference values, remaining stable over 14 days of wear and unaffected by patient characteristics [4]. Furthermore, in patients with type 1 diabetes, this system had a mean absolute relative difference of 13.2% and mean absolute difference of 19.8 mg/dl for the entire study period, which was similar to CGMs reported in previous studies under at-home conditions [5]. In two randomized controlled studies performed in insulin-treated adult patients with type 1 or 2 diabetes, FGM intervention failed to decrease HbA1c levels, but achieved a 38 43% reduction of time in hypoglycemia (<70 mg/dl) with less undesirable sensor-wear reactions [6, 7]. Thus, it is clinically helpful for meticulous glucose control and prevention of hypoglycemia for patients with diabetes. We found that FGM-ISFG could be used to evaluate PG in terms of clinical decision making because all data were plotted in Zones A and B on Parkes error grid analysis. The 2013 version of ISO15197 requires 95% of measured glucose values to be within Zone A or B on the Parkes error grid [8]. The FGM system was found to meet the criteria and to be fully available, indicating that the system is beneficial for non-diabetic Japanese to get information about variation patterns of blood glucose. However, it is a serious clinical problem that FGM-ISFG was higher than actual PG overall. Physiologically, it is unlikely that the glucose concentration in the interstitial fluid is higher than that in the blood. Glucose monitoring using subcutaneous interstitial fluid is less invasive, but this method can delay the detection of glycemic variability and decrease the sensitivity to local glycemic fluctuations related to the insertion sites of sensor, leading to unexplained Fig. 4 Variation of FGM-ISFG and PG during OGTT in Case 4 FGM-ISFG levels were always higher than PG levels. OGTT, oral glucose tolerance test; FGM-ISFG, interstitial fluid glucose concentration by flash glucose monitoring; PG, plasma glucose concentration. fluctuations of measured values [9, 10]. As these features are considered equally true in non-diabetic as well as diabetic individuals, FGM-ISFG in non-diabetic persons may tend to be displayed at a higher level due to measurement problems. Interestingly, some subjects showed little difference between FGM- ISFG and PG, whereas others showed marked differences with no tendencies related to gender, age, or glucose tolerance. Bailey et al. [4] reported that the accuracy of the sensor was unaffected by factors such as BMI, age, type of diabetes, insertion site, presence of insulin injection, or HbA1c. This strongly suggests that absolute values of FGM-ISFG should not be used for clinical decision making, at least in some subjects, without calibration of the values. The reasons for individual differences are unknown at present, but each showed a similar tendency in the first and second tests. Thus, appropriate calibration of absolute values may overcome these discrepancies. FGM-ISFG was higher than PG in all OGTTs. This tendency was particularly prominent with time points after glucose loading. Theoretically, the rise in FGM-ISFG should be delayed compared to that in PG at 30 minutes after glucose loading, suggesting that accuracy is particularly problematic for values after loading. Another study showed similar findings that called into question the reliability of FGM. In a study performed in 20 subjects (8 with type 1 diabetes mellitus, 12 with type 2 diabetes mellitus), the overall correlation of FGM-ISFG showed lower values from

Accuracy of flash glucose monitoring 831 FGM than expected based on point of care (POC) glucose measurement in the lower glucose range, and higher than expected values in the higher ranges. Furthermore, 75-g OGTT resulted in a slower rise in glucose level as determined by FGM than that obtained by the perchloric acid hexokinase method, POC testing, test strip, and conventional CGM during the first 45 60 minutes after glucose loading [11]. Moreover, a study using CGM including FGM for type 1 diabetes patients indicated that all sensors performed less accurately during hypoglycemia and best during hyperglycemia [12]. These observations strongly suggest that values obtained from FGM should be examined in a clinical context. FGM allows blood glucose fluctuation to be detected easily, and this technique may become the central method of evaluating glycemic control of diabetes in future. However, at present, it is not appropriate to change therapeutic regimens based on the judgment of nocturnal hypoglycemia and postprandial hyperglycemia by FGM-ISFG per se. Combined use of more reliable devices, including instruments for self-monitoring of blood glucose and subsequent appropriate calibration, is required if FGM is adopted. In some cases, it may be necessary to measure actual PG proactively if there are doubts regarding the results obtained with FGM. According to the current diagnostic criteria in Japan, diagnosis of diabetes is made partly based on fasting plasma glucose level, 75 g OGTT 2-hour value, and casual plasma glucose level [3]. In a potential individual with diabetes, a missed diagnosis may occur if blood glucose levels at the points described above do not happen to meet the criteria. Meanwhile, continuous blood glucose measurement with the FGM systems may be surely allowed to catch glucose variation pattern including hyperglycemia required for diagnosis. Moreover, adoption of dynamic index such as AUC obtained from FGM-based glucose values may lead to more early and correct diagnosis of diabetes. The FGM system will have potential to become an alternative diagnostic tool for diabetes by improvement of overestimation of IFSG-based glucose levels ISFGs after glucose loading. This study had a small number of subjects and did not include diabetic patients. To confirm our findings and obtain more detailed information on which subjects exhibit large discrepancies in FGM-ISFG and PG, it is desirable to perform clinical experiments in large numbers of subjects, including those with diabetes mellitus. In conclusion, the individual differences in FGM- ISFG in terms of consistency with actual PG were large, and FGM-ISFG tended to be higher. This tendency became more pronounced after glucose loading. Careful attention is required for proper application of FGM. Disclosure None of the authors have any potential conflicts of interest associated with this research. References 1. Sato J, Hirose T, Watada H (2012) Continuous glucose monitoring system: Is it accurate, safe and clinically useful? J Diabetes Investig 3: 225-230. 2. Heinemann L, Freckmann G (2015) CGM versus FGM; or, continuous glucose monitoring is not flash glucose monitoring. J Diabetes Sci Technol 9: 947-950. 3. Committee of the Japan Diabetes Society on the Diagnostic Criteria of Diabetes Mellitus, Seino Y, Nanjo K, Tajima N, Kadowaki T, et al. (2010) Report of the committee on the classification and diagnostic criteria of diabetes mellitus. J Diabetes Investig 1: 212-228. 4. Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S (2015) The Performance and Usability of a Factory- Calibrated Flash Glucose Monitoring System. Diabetes Technol Ther 17: 787-794. 5. Ólafsdóttir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, et al. (2017) A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes. Diabetes Technol Ther 19: 164-172. 6. Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R (2016) Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 388: 2254-2263. 7. Haak T, Hanaire H, Ajjan R, Hermanns N, Riveline JP, et al. (2017) Flash Glucose-Sensing Technology as a Replacement for Blood Glucose Monitoring for the Management of Insulin-Treated Type 2 Diabetes: a Multicenter, Open-Label Randomized Controlled Trial. Diabetes Ther 8: 55-73. 8. Garg SK, Reed K, Grey J, Westerman A (2013) Possible Impacts of New Accuracy Standards of Self-Monitoring of Blood Glucose. US Endocrinol 9: 28-31.

832 Sekido et al. 9. Burnett DR, Huyett LM, Zisser HC, Doyle III FJ, Mensh BD (2014) Glucose sensing in the peritoneal space offers faster kinetics than sensing in the Subcutaneous Space. Diabetes 63: 2498-2505. 10. Keenan DB, Mastrototaro JJ, Voskanyan G, Steil GM (2009) Delays in minimally invasive continuous glucose monitoring devices: a review of current technology. J Diabetes Sci Technol 3: 1207-1214. 11. Fokkert MJ, van Dijk PR, Edens MA, Abbes S, de Jong D, et al. (2017) Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus. BMJ Open Diabetes Res Care 5: e000320. 12. Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, et al. (2017) Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes. Diabetes Obes Metab 19: 1051-1055.