Validity of Data Extraction Techniques on the Kinetic Communicator (KinCom) Isokinetic Device

Similar documents
Effect of Preload and Range of Motion on Isokinetic Torque in Women

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75

***Note: Figures may be missing for this format of the document ***Note: Footnotes and endnotes indicated with brackets

Can Muscle Power Be Estimated From Thigh Bulk Measurements? A Preliminary Study

Effect of Gravity Correction on Shoulder Average Force and Reciprocal Muscle Group Ratios

Comparison of N-K Table Offset Angles with the Human Knee Flexor Torque Curve

Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers

Effect of cold treatment on the concentric and eccentric torque-velocity relationship of the quadriceps femoris

Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body weight, height, and sex in 178 healthy subjects

By: TEDDY W. WORRELL, MEd, PT, ATC, CRAID R. DENEGAR, PhD, ATC, SUSAN L. ARMSTRONG, MEd, ATC, and DAVID H. PERRIN, PhD, ATC

A Study on the Norm-Referenced Criteria for Isokinetic Functional Strength of the Wrist for Junior Baseball Players

performance in young jumpers

Resistive Eccentric Exercise: Effects of Visual

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

Peak Torque Comparison between isam 9000 and Biodex Isokinetic Devices

ISOKINETIC TESTING AND DATA INTERPRETATION

Eccentric and Concentric Force-Velocity Relationships of the Quadriceps Femoris Muscle

Isokinetic trunk flexion extension protocol to assess trunk muscle strength and endurance: Reliability, learning effect, and sex differences

Neither Stretching nor Postactivation Potentiation Affect Maximal Force and Rate of Force Production during Seven One-Minute Trials

Clinical Use of the Johnson Anti-Shear Device: How and Why to Use It

Comparative study of two isokinetics dynamometers: CYBEX NORM vs CON-TREX MJ

Effects of cutaneous and joint receptors on the in vivo quadriceps femoris torque-velocity relationship

differentiate between the various types of muscle contractions; describe the factors that influence strength development;

Comparison of isokinetic Strength and Flexibility Measures Between Hamstring Injured and Noninjured Athletes

Reliability of Stationary Dynamometer Muscle Strength Testing in Community-Dwelling Older Adults

strength ratios and power in tennis, squash and track athletes

Basics of kinetics. Kinesiology RHS 341 Lecture 7 Dr. Einas Al-Eisa

A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr.

DUAL POSITION BACK EX/FLEX ATTACHMENT

In the lower extremity, it is very important to maintain

Original Article. MUNENORI KATOH, PT, PhD 1), YUKINOBU HIIRAGI, PT, PhD 2), MANABU UCHIDA, PT, PhD 3)

Chapter 20: Muscular Fitness and Assessment

Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements

Tennis places high biomechanical demands upon

Test-Retest Reliability of Isokinetic Knee Extension and Flexion

The original publication is available at

Chapter 14 Training Muscles to Become Stronger

GK Jane Division of Physical Education, University of the Witwatersrand, Johannesburg, South Africa

DATA INTERPRETATION AND ANALYSIS

Isometric, Isotonic, and Isokinetic Torque Variations in Four Muscle Groups Through a Range of Joint Motion

Analyses of Isokinetic and Closed Chain Movements for Hamstring Reciprocal Coactivation

Reliability of isokinetic strength measurements of the knee in children with cerebral palsy

DAY 2 III. WORKOUT RULES

Biomechanics of Skeletal Muscle and the Musculoskeletal System

Lesson 7.2. Benefits of Flexibility

Specificity of Speed of Exercise. MARY T. MOFFROID, M.A. and ROBERT H. WHIPPLE, B.A PHYSICAL THERAPY

Exercise Therapy for Patients with Knee OA Knee Exercise Protocol Knee Home Exercise Programme

The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque

Muscle strength in patients with chronic pain

Section: Critically Appraised Topic. Article Title: The Effect of Contralateral Exercise on Patient Pain and Range of Motion

A Six Month Intervention Study on Strength Training in Youth Basketball Player in Albania

Chapter 20: Test Administration and Interpretation

The Star Excursion Balance Test (SEBT) is a unilateral, UNILATERAL BALANCE PERFORMANCE IN FEMALE COLLEGIATE SOCCER ATHLETES

TRUNK - 2 Flexion: Stretch Low Back Extensors (Supine)

EFFECTS OF TWO WARM-UP PROGRAMS ON BALANCE AND ISOKINETIC STRENGTH IN MALE HIGH SCHOOL SOCCER PLAYERS ACCEPTED

A new method for gravity correction of dynamometer data and determining passive elastic moments at the joint

Skeletal Muscles and Functions

The adaptations to resistance training are largely

Electrostimulation for Sport Training

Isokinetic strength testing does not predict hamstring injury in Australian Rules footballers

chapter Exercise Technique for Alternative Modes and Nontraditional Implement Training

OMICS - 3rd Int. Conference & 2

Comparing Dominant and Non-Dominant Torque and Work using Biodex 3 Isokinetic Protocol for Knee Flexors and Extensors INTRODUCTION PURPOSE METHODS

ACL Injury Prevention: Considerations for Children and Adolescents

Hakan Giir, MD, PhD, Bedrettin Akova, MD, Selfuk Kiifiiko~lu, MD

Summary Chart 1 2 months

Biomechanics of Resistance Training. Tutor: Phil Watkins

Technical Report. Differences in Quadriceps Femoris Muscle Torque When Using a Clinical Electrical Stimulator Versus a Portable Electrical Stimulator

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System

Role of Body and Joint Position on lsokinetic Exercise and Testing

Most Dangerous Man! FACTS ON STREGNTH. Session #627 Functionally Fit Seniors

Do Persons with PFP. PFJ Loading? Biomechanical Factors Contributing to Patellomoral Pain: The Dynamic Q Angle. Patellofemoral Pain: A Critical Review

by Hand-Held, isometric, and Isokinetic Dynamometry in Patients With Knee Dysfunction

Knee Flexor:Extensor Isokinetic Ratios in Young Male Gymnasts and Swimmers

MUSCLE BEHAVIOR IN ARTIFICIALLY ACTIVATED MUSCLE - MEASUREMENTS ON NEUROLOGICALLY INTACT AND PARAPLEGIC SUBJECTS METHODS

Eccentric torque-velocity relationship of the elbow flexors

Obesity is associated with reduced joint range of motion (Park, 2010), which has been partially

ISOKINETIC ASSESSMENT OF THE FOREARM AND WRIST MUSCLES

A descriptive profile of age-specific knee extension flexion strength in elite junior tennis players

RECOMMENDED STRETCHES

PROCEDURAL OPTIONS FOR MEASURING MUSCLE STRENGTH

Biomechanics of Skeletal Muscle and the Musculoskeletal System

Equipment Operation. Checklist Exercises

POST-ACTIVATION POTENTIATION AND VERTICAL JUMP PERFORMANCE. Cody Hardwick

TRANSFORMER ALL IN ONE ISOKINETIC MACHINE ANKLE ELBOW HIP SHOULDER WRIST TRUNK

MODULE 8: Breaking Down the Exercises Balance Exercises

The following pages constitute the final, accepted and revised manuscript of the article:

Relationship between Selected Physical Characteristics and Functional Capacity in the Anterior Cruciate Ligament-Insufficient Athlete

By Dr. Ellyn Robinson

Relative Isometric Force of the Hip Abductor and Adductor Muscles

What is Kinesiology? Basic Biomechanics. Mechanics

Lecture 2. Statics & Dynamics of Rigid Bodies: Human body 30 August 2018

APPENDIX: The Houston Astros Stretching Program

Name /scre/18_127 01/26/ :53AM Plate # 0-Composite pg 144 # 1

' Assistant professor, director, Graduate Athletic Training, Curry School of Education, University of Virginia,

Older Adult Advanced

Εffects of specific resistance training program on

Isokinetic Testing of the Shoulder Abductors and Adductors: Windowed vs Nonwindowed Data Collection

CHAPTER 15: KINESIOLOGY OF FITNESS AND EXERCISE

Andrew J Murphy

Transcription:

Validity of Data Extraction Techniques on the Kinetic Communicator (KinCom) Isokinetic Device By: Laurie L. Tis, PhD, AT,C * and David H. Perrin, PhD, AT,C Tis, L.L., & Perrin, D.H. (1993). Validity of the data extraction technique on the Kinetic Communicator isokinetic device. Isokinetics and Exercise Science, 3, 96-100. ***Note: Figures may be missing from this format of the document Abstract: The purpose of this study was to examine the validity of the average and peak torque values of the knee extensor and flexor musculature obtained through the data extraction technique. Twenty women (age = 20.2 ± 1.01 yr; height = 169.0 ± 6.8 cm; weight = 60.8 ± 5.5 kg) were assessed for isokinetic peak and average torque of the knee musculature at a velocity of 90 deg/sec. Subjects were randomly assessed through a range of motion of 5-90 deg and 25-70 deg. Peak and average torque values within a range of motion of 25-70 deg were extracted from the tested range of motion of 5-90 deg. Correlational analyses between the tested and extracted variables revealed relationships ranging from r = 0.79 to 0.95. Paired t tests found no significant differences (p < 0.006). These findings suggest that the data extraction techniques are valid. Article: INTRODUCTION Hislop and Perrine 1 first introduced isokinetic dynamometry to the scientific community in 1967. Several mechanical properties of muscle, including torque, total work, and average power, have been used as the traditional isokinetic measurements. 4,6 Isokinetic assessment has also historically occurred through a full range of motion. However, the newer generations of isokinetic dynamometers have developed several additional evaluation options and data analysis parameters. The KinCom software package (Chattecx Corp., Chattanooga, TN) allows the clinician to extract data from a smaller range of motion within the total tested range of motion for data analysis purposes. Data extraction can be accomplished by moving the range of motion cursors after data collection to delineate a smaller range of motion. Two common uses of the data extraction technique have recently been illustrated in the literature. The first use was to test a healthy population through a full range of motion and extract data through a shorter range of motion to infer to an injured population. Ng and Kramer5 reported peak and average torque values of the shoulder internal and external rotator muscle groups. The researchers assessed the shoulder rotators through a range of 135 deg at a test velocity of 60 deg/sec. However, they reported torque values through a 60-deg range of motion. The stated purpose of this data extraction technique was to infer data of injured patients with restricted ranges of motion. * Assistant Professor, Georgia State University, Department of HPRD, College of Education, Atlanta, Georgia. Associate Professor and Director, Sports Medicine Research Laboratory and Graduate Athletic Training Education, University of Virginia, Charlottesville, Virginia.

The second recently observed use of the data extraction technique is to eliminate the first and last 10 deg from the tested range of motion. The stated purpose of this technique is to eliminate from analysis any torque production during acceleration to the preset velocity and to eliminate any potential impact artifact from the torque curve analysis. 2,3,5 Although use of the data extraction technique in isokinetic research is becoming more frequent, the validity of extracted data has not been determined. Therefore, the purpose of this study was to determine the validity of the data extraction capability of the KinCom software package. METHODS Subjects Twenty women (age = 20.2 ± 1.01 yr; height = 169.0 ± 6.8 cm; weight = 60.8 ± 5.5 kg) participated in the study on giving their informed consent in accordance with institutional human investigation committee guidelines. Subjects were excluded from the study if they reported any history of injury sustained to the right knee. Test Protocol Subjects were assessed for isokinetic concentric and eccentric average and peak torque (Nm) of the right side knee extensor and flexor musculature on a Kinetic Communicator (KinCom) isokinetic dynamometer at a velocity of 90 deg/sec. Subjects were randomly assessed through 5-90 deg ROM and 25-70 deg ROM. A preload of 75 N was used, all data were gravity corrected, and the KinCom was calibrated before each test session. Knee Extensor Assessment Torque/force output of the knee extensor muscle group was assessed in an upright, seated position using the KinCom back rest attachment. The hip was positioned in approximately 80 deg of flexion. Subjects were secured with velcro straps at the distal thigh and across the waist. The distal pad was placed slightly proximal to the malleoli. The axis of the dynamometer was aligned with the axis of rotation of the knee. Knee Flexor Assessment Torque/force output of the knee flexor muscle groups was assessed in a prone position. Subjects were secured at the pelvis and distal thigh with Velcro straps. The distal pad was placed slightly proximal to the malleoli, and the axis of the dynamometer was aligned with the axis of rotation of the knee. Warm-up Procedures All subjects participated in stretching exercises for the quadriceps and hamstring muscle groups both before and after the testing session. All subjects participated in a warm-up session before both isokinetic test sessions. The warm-up consisted of three to five submaximal and one maximal concentric and eccentric contractions for both the knee extensor and flexor muscle groups. Assessment procedures followed a 1-min rest period. Data Collection

After a 1-min rest period, subjects performed several maximal concentric and eccentric contractions of the knee extensor muscle group. The data management capabilities of the KinCom's computer software allowed the researcher to examine the torque curves produced by each subject as the contractions were executed. Therefore, as torque curves were produced and examined, three reproducible torque curves were selected and used for subsequent data analysis. Subjects performed the number of contractions needed to obtain the three reproducible torque curves. After assessment of knee extensor strength, subjects were positioned for assessment of knee flexor strength. Data collection procedures were identical for the knee flexor muscle group. Data Extraction After data collection, the KinCom software package has many options for data reduction. The system will calculate gravity-corrected average and peak torque values from the gravity correction value and from the torque produced by the subjects. Average torque values are produced immediately when extracting data. By moving a cursor across the torque curve, the peak torque values were located and used for further analysis. Similarly, the software package allows the range of motion cursors to be moved, such that a shorter range of motion may be analyzed from within a tested, larger range of motion. Using the same techniques as previously mentioned, peak and average torque values from a range of 25-70 deg range of motion were extracted from within the tested 5-90 deg range of motion, and used for subsequent data analysis. Statistical Analysis To determine the validity of the KinCom computer software's data extraction capability, correlational analyses (Pearson product moment correlation) were used to determine the relationship between the tested versus extracted concentric and eccentric average and peak torque values of the knee extensors and flexors. Additionally, eight paired t tests were computed to determine if significant differences existed between the tested and extracted average and peak torque values produced by the knee extensor and flexor muscle groups. A Bonferroni correction (0.05 / 8) was used to adjust for the possibility of an inflated alpha level. The adjusted level of significance was set at p < 0.006. RESULTS Average and peak tested and extracted torque values are presented in Table 1. Pearson correlation coeffi-

cients ranged from 0.79 to 0.95 between the tested and extracted variables and paired t tests showed no significant differences between the tested and extracted torque measures (Table 1). DISCUSSION The KinCom software package allows the researcher to extract data from a smaller range of motion within the total tested range of motion for the purpose of data analysis. The results of the present study indicate that this is a valid procedure. Pearson product moment correlation coefficients showed moderate to high relationships between the tested and extracted measures. In further support of the validity of the procedure, the differences between the tested and extracted values were nominal and not statistically significant, ranging from 1 to 8%.

Closer examination of the data showed some interesting patterns in the relationships between the tested and extracted torque values. Most of the comparisons demonstrate that the extracted average and peak torque values were greater than the tested values, with the exception of

eccentric extension (Figures 1-4). This finding may be related to the type of torque curves produced by the knee extensor and flexor musculature. The eccentric extensor torque curve was an ascending curve at 90 deg/sec and extraction removed the highest points in the curve, which lowered the average torque value. However, the other curves were either parabolic (concentric extension) or relatively flat (flexion), with the lowest points in the curve existing at the ends. As such, extraction eliminated the low points and thus increased the average torque value. Although these differences were not significant in the present study, the findings are limited because the relationships will vary depending on the muscle group assessed, the velocity of assessment, the subject population used, and the portion of a total curve that will be extracted. For example, assessment of the knee extensor musculature at a lower velocity, or in a population of athletic males, may generate torque curves that are more parabolic in nature. Extracting from the center of the curve may hypothetically significantly increase the torque values in comparison with actual tested values. Another interesting observation in the present study was that the extracted peak torque values were not exactly the same as the tested values, indicating that, depending on the amount of curve that is extracted, the point at which peak torque occurs may be eliminated. Again, the degree that peak torque was affected in the present study was minimal. However, this finding is specific to the muscle groups assessed, the subject population, test velocity, and the amount of curve extracted. Although at the present time limited use of the data extraction techniques are found in the literature, the potential for further use exists. One demonstrated application is to infer data of an injured population with restricted range of motion. Ng and Kramer 5 reported peak and average torque values of the shoulder internal and external rotator muscle groups. The researchers assessed the shoulder rotators through a range of 135 deg at a test velocity of 60 deg/sec. However, they reported average and peak torque values through 60 deg range of motion. Their observation that peak torque did not occur within the extracted range of motion adds credence to the findings of the present study in that peak torques may not be exact if the point at which peak torque occurs is not included in the extracted portion of the curve. A second recently observed use of the data extraction technique was to eliminate the first and last 10 deg from the tested range of motion. The stated purpose of this technique is to eliminate from analysis any torque production during acceleration to the preset velocity and to eliminate any potential impact artifact from the torque curve analysis. 2,3,5 With respect to the KinCom isokinetic device, this may be an unnecessary procedure because the software controls the time to accelerate to the preset velocity through what it terms acceleration turn points. Through selection of either a low, medium, or high acceleration turn point, the preset velocity is reached at varying time frames. Similarly, preload should control for impact artifact by standardizing the amount of force a subject must produce before lever arm movement can be initiated. Although this may not be a necessary use of the data extraction procedure, the torque values are valid based on the findings of this study. In conclusion, the findings of this investigation suggest that the data extraction technique is a valid procedure. Further research is needed to examine the relationship between tested and

extracted torque values for other muscle groups, additional subject populations, and additional test velocities. Additionally, further research should also examine the relationship between tested and extracted values when extracting through different portions of a torque curve as well as differing amounts of a torque curve. REFERENCES 1. Hislop HJ, Perrine JJ: The isokinetic concept of exercise. Phys Ther 47:114-117,1967. 2. Kramer JF: Effect of hand position on knee extension and knee flexion torques of intercollegiate rowers. J Orthop Sports Phys Ther 11:367-371,1990. 3. Kramer JF, Vaz MD, Hakansson D: Effect of activation force on knee extensor torques. Med Sci Sports Exerc 23:231-237,1991. 4. Moffroid M, Whipple R, Hofkosh J, et al.: A study of isokinetic exercise. Phys Ther 49:735-746,1969. 5. Ng LR, Kramer JS: Shoulder rotator torques in female tennis and nontennis players. J Orthop Sports Phys Ther 13:40-45,1991. 6. Thistle HG, Hislop Hi, Moffroid M, Lowman EW: Isokinetic contraction: a new concept of resistive exercise. Arch Phys Med Rehabil 48:279-282,1967.