Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Similar documents
Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Non-Newtonian blood flow in human right coronary arteries: Transient simulations

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

FOR many decades, cardiovascular disease has been one of

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

Blood flow in vessels with artificial or pathological geometrical changes

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Blood Flow Simulation toward Actual Application at Hospital

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images

There has been extensive speculation in the literature about

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Finite Element Modeling of Three-Dimensional Pulsatile Flow in the Abdominal Aorta: Relevance to Atherosclerosis

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION

CPM Specifications Document Aortofemoral Normal:

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch?

Post-conditioning. P a g e 1. To my Thesis Committee,

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis

CARDIOVASCULAR disease is a leading cause of morbidity

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

4D model of hemodynamics in the abdominal aorta

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques

Effect of exercise on hemodynamic conditions in the abdominal aorta

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

Deposited on: 21 January 2009

3D Modeling of Plaque Progression in the Human Coronary Artery

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses

Analysis of the effects of plaque deposits on the blood flow through human artery

Numerical analysis of blood flow in human abdominal aorta

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

CPM Specifications Document Aortofemoral Normal:

Non-axisymmetric Flow Development in Pulsatile Blood Flow through an Aneurysm

CPM Specifications Document Aortic Coarctation: Exercise

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

Correlations of coronary plaque wall thickness with wall pressure and wall pressure gradient: A representative case study

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Computational design of Intracranial Stent using 3D visualization system

EFFECT OF BRANCHINGS ON BLOOD FLOW IN THE SYSTEM OF HUMAN CORONARY ARTERIES. Benchawan Wiwatanapataphee. Yong Hong Wu. Thanongchai Siriapisith

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

On the Existence of an Optimum End-to-side Junctional Geometry in Peripheral Bypass Surgery A Computer Generated Study

Particle Hemodynamics Analysis after Coronary Angioplasty

BLOOD FLOW IN ARTERIES

Research Article Numerical Study of Turbulent Pulsatile Blood Flow through Stenosed Artery Using Fluid-Solid Interaction

Geometry of the Carotid Bifurcation Predicts Its Exposure to Disturbed Flow

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

¼ density (kg/m 3 ) ij ¼ Reynolds stress tensor (kg/m-s 2 ) ! ¼ specific dissipation rate ("=k)

Comparison of Stent Designs using Computational Fluid Dynamics

Senior Honors Thesis Prospectus

Influence of Oscillating Flow on LDL Transport and Wall Shear Stress in the Normal Aortic Arch

Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms

Blood Flow Patterns in the Proximal Human Coronary Arteries: Relationship to Atherosclerotic Plaque Occurrence

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

FSI Analysis of Diseased Coronary using Patient Specific Data

Flow in re-stenosed artery after angioplasty

Quantifying turbulent wall shear stress in a subject specific human aorta using large eddy simulation

Determining the effect of congenital bicuspid aortic valves on aortic dissection using computational fluid dynamics

Vascular Hemodynamics CFD Modeling

FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall

Research Article Effect of Pulsatile Flow Waveform and Womersley Number on theflowinstenosedarterialgeometry

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery

The study on hemodynamic effect of series type LVAD on aortic blood flow pattern: a primary numerical study

Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

A Two-layered Model for the Analysis of Arterial Rheology

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Rheology update in Cardiovascular disease

CFD study on effect of branch sizes in human coronary artery

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD

BLOOD SIMULATION AND VIRTUAL BYPASS SURGERY

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling

Transcription:

Blood flow in S-shaped in-plane and out-of-plane coronary arteries Author Johnston, Peter, Johnston, Barbara Published 2008 Journal Title The A N Z I A M Journal DOI https://doi.org/10.21914/anziamj.v49i0.330 Copyright Statement 2008 Cambridge University Press. The attached file is reproduced here in accordance with the copyright policy of the publisher. Please refer to the journal's website for access to the definitive, published version. Downloaded from http://hdl.handle.net/10072/22491 Griffith Research Online https://research-repository.griffith.edu.au

ANZIAM J. 49 (EMAC2007) pp.c341 C358, 2008 C341 Blood flow in S-shaped in-plane and out-of-plane coronary arteries P. R. Johnston 1 B. M. Johnston 2 (Received 27 July 2007; revised 10 December 2007) Abstract The distribution of wall shear stress (wss) in coronary arteries is an initiating factor for coronary artery disease. The complicated three dimensional structure of coronary arteries makes it difficult to isolate the factors affecting wss distributions. Here we present a computational fluid dynamics study of transient flow in simplified S-shaped arteries with in-plane and out-of-plane bends. For in-plane arteries there are two regions of consistently low wss during the cardiac cycle. The wss distribution is related to secondary velocities which appear as symmetric counter-rotating eddies and are approximately 8% of the axial velocity. For out-of-plane arteries the wss distributions are similar, but the eddies are no longer symmetric with secondary velocities only 7% of the axial velocities. The symmetric velocity patterns produced by the in-plane model could only exist in a single bend, which limits the usefulness of this model. The results obtained give useful insights into the nature of blood flow in realistically shaped coronary arteries. See http://anziamj.austms.org.au/ojs/index.php/anziamj/article/view/330 for this article, c Austral. Mathematical Soc. 2008. Published January 15, 2008. ISSN 1446-8735

Contents C342 Contents 1 Introduction C342 2 The model C344 2.1 Geometry........................... C344 2.2 Governing equations..................... C345 2.3 Computational mesh..................... C347 2.4 Parameters for comparison.................. C348 3 Results and discussion C348 3.1 Wall shear stress....................... C348 3.2 Secondary velocities...................... C351 3.3 Oscillatory shear index.................... C354 4 Conclusions C355 References C355 1 Introduction Atherosclerotic lesions develop preferentially at certain sites in arteries such as the aortic arch, carotid bifurcation, infrarenal, femoral and coronary arteries [1, 4, 5, 10, 16]. Clinical observations found that plaques form predominantly on the inside of curves and the outer walls of vessel bifurcations [1, 7, 10, 16] in the arteries listed above and studies have shown that these sites co-localise with regions of low shear stress [8]. Several anatomical studies have shown that there is something different about plaque formation in coronary arteries compared to other arteries [10, 15]. The main differences are that plaques form in both straight and curved sections of the arteries, which is not the case for other arteries. Rea-

1 Introduction C343 sons for such differences have been postulated as the complex pulsatile nature of the blood flow [14], the rheological properties of blood [3] and the geometry of the arteries, including branching, bifurcation and curvature [7, 8]. Possible factors that may cause differences in the coronary arteries compared with other arteries are the particular character of the coronary inlet velocity waveform [11] and the fact that coronary velocities are lower than most of the other arteries mentioned [9, 11], which, with their smaller diameters, leads to lower Reynolds numbers [11]. Other factors may be their tortuous nature [1], which includes out-of-plane bends [2]. Although there are numerous studies of blood flow in various models of different human arteries, the starting point for this paper will be a recent study by Qiao et al. [13], who considered pulsatile flow in two in-plane S- shaped curves of 6 and 3 mm radii, respectively. The idea of their study was to numerically simulate blood flow in the aorta as well as other minor arteries. The study used an inlet velocity waveform corresponding to flow in the ascending aorta with a maximum velocity of 1.4 m/s over a 0.4 second cycle, corresponding to a rather rapid pulse rate of 150 beats per minute. This article presents simulations of blood flow in the coronary (small diameter) arteries. The aim is to use realistic coronary velocities up to 0.22 m/s at a frequency of 60 beats per minute, a resting pulse rate. Two S-shaped coronary arteries are considered: firstly an in-plane artery (similar to that of Qiao et al. [13]) and secondly an artery with an out-of-plane bend. Simulations are presented for a left coronary artery using a velocity waveform measured at the entrance of the artery. In Section 3, comparisons are made between the in- and out-of-plane arteries, which are compared in terms of wall shear stress (wss), secondary velocities and oscillatory shear index (osi).

2 The model C344 Inlet A A Inlet B B C A A C D D B B C Outlet E E C Outlet Figure 1: In-plane (left) and out-of-plane (right) artery models 2 The model 2.1 Geometry Consider the in-plane S-shaped model shown in Figure 1(left). The straight inlet and outlet sections are of length 3 cm and 2 cm, respectively, and the radius of curvature, R c, of each curved section is 3.5 cm, giving an overall artery length of 16 cm. The cross-section of the artery model is circular with a radius of 2 mm. Several particular cross-sections are indicated in the figure by the lines AA, BB, CC, DD and EE and these will be used to discuss results in Section 3. To create the out-of-plane artery model, Figure 1(right), the second half

2 The model C345 of the artery, after the CC plane, is rotated through π/2 in the plane of CC so the outlet now comes out of the page. The lines DD and EE are still present but not shown in Figure 1(right). 2.2 Governing equations In these simulations blood is assumed to be incompressible fluid, which is governed by the Navier Stokes equations and the continuity equation ρ v t + ρv v = µ 2 v P, v = 0, where the velocity v is a three dimensional vector field, P denotes pressure, ρ denotes density (1056 kg/m 3 ) and µ denotes viscosity (0.00345 Pa s). This model assumes that blood is a Newtonian fluid, which has been demonstrated to be a reasonable assumption for the transient simulation of blood flow in arteries [6]. These equations are solved subject to the following boundary conditions. 1. The artery walls are assumed rigid (which is a common assumption [3, 12, 13, 14] used to ease computational load), upon which a no-slip condition is imposed on v. 2. At the outlet, the velocities are free but the normal and tangential stresses are constrained to be zero and the gauge pressure is set to zero. 3. At the inlet, a uniform velocity profile is used with a time varying forcing function representative of flow in the left coronary artery, Figure 2 [12]. The forcing function is scaled to have a maximum inlet velocity v max = 0.214 m/s with a period T = 1 second (60 heart beats per minute).

2 The model C346 0.25 0.2 LCA Velocity (m/s) 0.15 0.1 0.05 0-0.05-0.1 0 0.2 0.4 0.6 0.8 1 Time (seconds) Figure 2: Waveform of the velocity forcing function for a left coronary artery.

2 The model C347 To compare this model with other flow studies, the following dimensionless parameters are used: the Reynolds number, Re; the Womersley number, α, and the Dean number, Dn. For the simulations presented here, the parameters are defined as and evaluate to: 1. Peak Reynolds number Re = 2ρRV max µ = 226, 2. Womersley parameter α = 2πρ T µ R = 0.057, 3. Dean Number Dn = Re R R c = 54. 2.3 Computational mesh The nonlinear governing equations were solved using the cfd-ace suite of packages (cfdrc, Huntsville, Alabama, usa). The solution procedure utilises the finite volume method and is based on the simple family of algorithms. Meshes were created using the geom package of the cfd-ace suite. The inlet mesh is a typical butterfly mesh with 18 nodes along each side of the centre square and 25 nodes along the wings of the mesh. Nodes are clustered towards the walls of the artery to improve the calculation of the wall shear stress. The volume mesh is created by extruding the inlet mesh along the centreline of the artery employing a total of 96 nodes over the length of the artery, giving a total of 192,882 nodes for the mesh. This figure was determined by increasing the number of nodes until changes in the oscillatory shear index were less than 0.1% over the entire artery wall.

3 Results and discussion C348 2.4 Parameters for comparison To study the differences between the in-plane and out-of-plane arteries the following comparisons will be used: 1. The wall shear stress (wss) τ w = µ v n, wall which is related to the outward normal derivative of the velocity at the wall. 2. Secondary velocities in the planes AA, BB, CC, DD and EE along the artery. 3. The oscillatory shear index osi = 1 T 1 τ 0 w dt 2 T τ, 0 w dt which is a form of time averaged wall shear stress over one heart beat. 3 Results and discussion 3.1 Wall shear stress During the majority of the cardiac cycle there is forward flow (Figure 2) and the wss is qualitatively similar to that shown in Figure 3, with high wss on the outside of curves and low wss on the inside of curves. Figure 3(left) shows

3 Results and discussion C349 Figure 3: Wall shear stress distributions at peak flow in the in-plane (left) and out-of-plane (right) artery models.

3 Results and discussion C350 WSS (Pa) 3.5 3 2.5 2 1.5 1 0.5 0-0.5-1 -1.5 In-plane C Out-of-plane C In-plane C Out-of-plane C 0 0.2 0.4 0.6 0.8 1 Normalised Time Figure 4: Trace of wss at the points C and C throughout the heart beat cycle. wss for in-plane artery model at (normalised) time t = 0.52, which corresponds to peak forward flow for that artery. Interestingly, the pattern of wss is the same for the out-of-plane model (Figure 3(right)) but the wss distribution wraps smoothly around the artery between the planes CC and DD. At other times of forward flow, the pattern of wss is very similar to those shown in Figure 3, but the magnitudes vary with the velocity waveform (Figure 2). Figure 4 shows that wss at C is always greater than or equal to wss at C. This is consistent with the wss plots in Figure 3, which show spots of high wss at C and low wss at C, for both artery models, at the particular

3 Results and discussion C351 instant in time (t = 0.52) chosen. However, the magnitude of the negative wss is greater at C than at C and so, for the portion of the cycle where there is back-flow (Figure 2, t = 0.08 to t = 0.37), wss is larger on the inside rather than the outside of the curve, since negative wss indicates reversed direction. Another feature of Figure 4 is that wss is very similar in both models and this is in agreement with the similarities between in the distributions in Figure 3. In addition, similar graphs (not presented) at E and E show wss which is almost the same as at C and C, respectively, where the dash and non-dash reversal occurs because E is now on the inside of the bend and therefore corresponds to C, and so on. 3.2 Secondary velocities Examination of secondary velocities in the planes BB, CC, DD and EE indicates differences in flow patterns between the two artery models. For the majority of the cycle the flow profiles show two eddies, one in each of the top and bottom halves of the artery, with flow in these in opposite directions. A series of these plots was constructed for various time intervals and the characteristics of these plots are summarised in Table 1. The entries in this table are of the form #$* where #=number of eddies, $=rotation direction of the topmost eddy (c=clockwise, a=anti-clockwise) and *=symmetry of eddies about the plane of curvature (s=symmetric, n=not symmetric). Four key features observed from Table 1 are the following 1. During forward flow, two eddies are always seen, while during back flow two eddies are seen at BB and DD, but never at EE and only once at CC. 2. At BB and CC, the top eddy rotates in a clockwise direction (except at CC during back flow), whereas at DD and EE, the rotation direction of the top eddy is anti-clockwise.

3 Results and discussion C352 Table 1: Summary of secondary velocity flow patterns at various points of the in-plane (ip) and out-of-plane (oop) artery models (c=clockwise, a=anticlockwise, s=symmetric, see text for full explanation of the notation). Time ip oop ip oop ip oop ip oop Period BB BB CC CC DD DD EE EE 0.01 0.10 2cs 2cs 2cs 2cs 2as 2an 2as 2as 0.11 0.19 2cs 2cs 0 0 2as 2an 0 0 0.20 0.37 2cs 2cs 2as 2an 2as 2as 0 0 0.38 0.41 2cs 2cs 0 0 2as 2as 0 0 0.42 1.00 2cs 2cs 2cs 2cs 2as 2as 2as 2as 3. In the in-plane artery model, the eddies are symmetric with respect to the plane of curvature. 4. In the out-of-plane artery model, the eddies are symmetric at BB, CC and EE ; at DD, they are often not symmetric. The symmetric flow patterns produced by the in-plane artery model reflects a limitation of that model. For this case it would be almost possible to predict the flow patterns by considering a series of single bends in the artery. However, since coronary arteries sit on the curved surface of the heart, no symmetry plane can exist and the in-plane model is of limited use. The ratio of mean secondary velocity to mean axial velocity in the CC plane for the in-plane model was 8.2% and for the out-of-plane model it was 7.3%. In the EE plane the ratio was 6.3% for both models.

3 Results and discussion C353 Figure 5: Oscillatory shear index in the in-plane (left) and out-of-plane (right) artery models.

3 Results and discussion C354 Table 2: Values of mmwss and osi at the points of mmwss and next mmwss for the in-plane (ip) and out-of-plane (oop) artery models Artery Time of Value of osi at Time of Value of osi at Model mmwss mmwss mmwss next next next mmwss mmwss mmwss ip 0.44 1.61 0.14 0.44 1.70 0.14 oop 0.44 1.71 0.13 0.44 1.73 0.14 3.3 Oscillatory shear index A plot of osi for both artery models is shown in Figure 5 and when this is compared with Figure 3, which shows wss for the same arteries, it is apparent that positions of low wss are positions of high osi and vice versa. For the inplane model (Figure 5(left)) there is an abrupt change in osi just downstream of the CC plane. However, for the out-of-plane model (Figure 5(right)) there is a gradual change in osi wrapping around the artery in a helical fashion from C to E. To examine the relationship between extremes in wall shear stress and extremes in osi, positions of the lowest maximum wss at any point over the heart beat cycle were calculated and denoted by mmwss. These points, denoted C and E (see Figure 3), are the same for both arteries and are very close to the points C and E. At these points the osi was also determined and the results are given in Table 2. In general, these values of osi are very near to maximum values. Table 2 shows that low values of wss correspond to high values of osi, as discussed above. The osi in the in-plane model is a little higher than in out-of-plane model at the same point.

4 Conclusions C355 4 Conclusions Simulations have been presented for blood flow in coronary arteries of simplified geometry. In-plane and out-of-plane models were used in order to avoid the complex nature of realistic coronary arteries. Simulations show that differences exist between flow parameters (wss, secondary velocities, osi) in these two simplified models. The in-plane model is too much of a simplification for considering multiple bends in an artery. It produces a symmetric flow pattern which could only exist in a single bend. The out-of-plane model allows for a smooth transition region between two bends in the artery. In order to achieve useful insights into blood flow in realistic coronary arteries using simplified models it is necessary to at least consider flow with some form of out-of-plane bend. The degree of out-of-plane nature of the model does not need to be the relatively extreme π/2 rotation used here. Useful insights would be obtained by using less extreme degrees of rotation and flow through bends of less than 90. References [1] Toshihisa Asakura and Takeshi Karino. Flow patterns and spatial distribution of atherosclerotic lesions in human coronary arteries. Circ. Res., 66:1054 1066, 1990. C342, C343 [2] C. G. Caro, D. J. Doorley, M. Tarnawski, K. T. Scott, Q. Long, and C. L. Dumoulin. Non-planar curvature and branching of arteries and non-planar-type flow. Proc. R. Soc. Lond. A, 452:185 197, 1996. C343 [3] Jie Chen and Xi-Yun Lu. Numerical investigation of the non-newtonian pulsatile blood flow in a bifurcation model with a non-planar branch. Journal of Biomechanics, 39(5):818 832, 2006. doi:10.1016/j.jbiomech.2005.02.003 C343, C345

References C356 [4] M. H. Friedman, O. J. Deters, F. F. Mark, C. B. Bargeron, and G. M. Hutchins. Aterial geometry affects hemodynamics: A potential risk factor for athersclerosis. Atherosclerosis, 46:225 231, 1983. doi:10.1016/0021-9150(83)90113-2 C342 [5] H. W. Hoogstraten, J. G. Kootstra, B. Hillen, J. K. B. Krijger, and P. J. W. Wensing. Numerical simulation of blood flow in an artery with two successive bends. Journal of Biomechanics, 29:1075 1083, 1996. doi:10.1016/0021-9290(95)00174-3 C342 [6] B. M. Johnston, P. R. Johnston, S. Corney, and D. Kilpatrick. Non-Newtonian blood flow in human right coronary arteries: Transient simulations. Journal of Biomechanics, 39(6):1116 1128, 2006. doi:10.1016/j.jbiomech.2005.01.034 C345 [7] D. Liepsch. An introduction to biofluid mechanics basic models and applications. Journal of Biomechanics, 35:415 435, 2002. doi:10.1016/s0021-9290(01)00185-3 C342, C343 [8] A. M. Malek, L. A. Seth, and S. Izumo. Hemodynamic shear stress and its role in atherosclerosis. JAMA, 282(21):2035 2042, 1999. C342, C343 [9] R. M. Nerem, J. A. Rumberger, D. R. Gross, W. W. Muir, and G. L. Gelger. Hot-film coronary artery velocity measurements in horses. Cardiovasc. Res., 10:301 313, 1976. C343 [10] M. Ojha, R. L. Leask, J. Butany, and K. W. Johnston. Distribution of intimal and medial thickening in the human right coronary artery: a study of 17 RCAs. Atherosclerosis, 158:147 153, 2001. doi:10.1016/s0021-9150(00)00759-0 C342 [11] T. J. Pedley. The fluid mechanics of large blood vessels. Cambridge University Press, 1980. C343

References C357 [12] K. Perktold, R. M. Nerem, and R. O. Peter. A numerical calculation of flow in a curved tube model of the left main coronary artery. J. Biomechanics, 24(3 4):175 189, 1991. doi:10.1016/0021-9290(91)90176-n C345 [13] A. K. Qiao, X. L. Guo, S. G. Wu, Y. J. Zeng, and X. H. Xu. Numerical study of nonlinear pulsatile flow in S-shaped curved arteries. Medical Engineering and Physics, 26:545 552, 2004. doi:10.1016/j.medengphy.2004.04.008 C343, C345 [14] D. A. Steinman, D. A. Vorp, and C. R. Ethier. Computational modeling of arterial biomechanics: Insights into pathogenesis and treatment of vascular disease. Journal of Vascular Surgery, 37(5):1119 1127, 2003. doi:10.1067/mva.2003.122 C343, C345 [15] J. C. Wang, S.-L. T. Normand, L Mauri, and R. E. Kuntz. Coronary artery spatial distribution of acute myocardial infarction occlusions. Circulation, 110:278 284, 2004. doi:10.1161/01.cir.0000135468.67850.f4 C342 [16] C. K. Zarins, D. P. Giddens, B. K. Bharadvaj, V. S. Sottiurai, R. F. Mabon, and S. Glagov. Carotid bifurcation atherosclerosis: quantitative correlation of plaque localization with flow velocity profiles and wall shear stress. Circ. Res., 53:502 514, 1983. C342

References C358 Author addresses 1. P. R. Johnston, School of Biomolecular and Physical Sciences Griffith University, Nathan, Australia. mailto:p.johnston@griffith.edu.au 2. B. M. Johnston, School of Biomolecular and Physical Sciences Griffith University, Nathan, Australia.