TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound

Similar documents
Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately.

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

Routine Quality Assurance Cookbook

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis & Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRANSDUCER TEST PHANTOMS

ULTRASOUND QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring DOPPLER FLOW PHANTOMS MULTI-PURPOSE PHANTOMS TRAINING PHANTOMS

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

Outline. QA/QC of Ultrasound Imagers: Basic Physics, Procedures and Experiences. Frame Rate Limitation. US Imaging Range Equation.

Ultrasound Physics and Knobology Alan Macfarlane. Consultant Anaesthetist Glasgow Royal Infirmary

Diploma of Medical Ultrasonography (DMU) Physical Principles of Ultrasound and Instrumentation Syllabus

Lesson 03: Sound Wave Propagation and Reflection. This lesson contains 15 slides plus 14 multiple-choice questions.

8/3/2016. Diagnostic Ultrasound Imaging Quality Assurance. Purpose. Information on US QA. James A. Zagzebski 1, Ph.D. Zheng Feng Lu 2, Ph.D.

Ultrasound Physics & Terminology

Current Ultrasound Quality Control Recommendations and Techniques

Hands-On Ultrasound Physics and Quality Control Workshop. Outline

Preamble (disclaimer)

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

1. Fig. 1 shows data for the intensity of a parallel beam of X-rays after penetration through varying thicknesses of a material

Principles of Ultrasound. Cara C. Prideaux, M.D. University of Utah PM&R Sports Medicine Fellow March 14, 2012

Underwater Acoustic Measurements in Megahertz Frequency Range.

ULTRASOUND. OB/Gyn (Core) Ultrasound PIEZOELECTRIC EFFECT. Principles of Ultrasound Physics and Instrumentation. Nathan Pinkney, BS, CDOS

Physical Principles of Ultrasound

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Ultrasound Applied Physics

Descriptions of NDT Projects Fall 2004 October 31, 2004

Basic Physics of Ultrasound and Knobology

Dr Emma Chung. Safety first - Physical principles for excellent imaging

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Ultrasound Principles cycle Frequency Wavelength Period Velocity

Quality Control Program Of Real Time Medical Ultrasound Machines In Sudan

Ultrasound. Principles of Medical Imaging. Contents. Prof. Dr. Philippe Cattin. MIAC, University of Basel. Oct 17th, 2016

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Introduction to Ultrasound Guided Region Anesthesia

Terminology Tissue Appearance

APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS

Performance Evaluation of Ultrasound Systems: Advanced Methods and Applications

Ultrasonic Testing Level I:

Debbie Childs RDMS, RVT Sonographer Murphy Medical Center Murphy, NC

Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion

DIGITAL IMAGE PROCESSING IN ULTRASOUND IMAGES

Basic of Ultrasound Physics E FAST & Renal Examination. Dr Muhammad Umer Ihsan MBBS,MD, DCH CCPU,DDU1,FACEM

Ultrasound Measurements and Non-destructive Testing Educational Laboratory

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

ULTRASOUND IMAGING EE 472 F2018. Prof. Yasser Mostafa Kadah

Tissue Strain Analytics Virtual Touch Tissue Imaging and Quantification

Point-of-Care Ultrasound: An Introduction

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

A new method of sonograph lateral resolution measurement using PSF analysis of received signal

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Ultrasound Accreditation and Regulatory Compliance from A to Z

Linear Ultrasonic Wave Propagation in Biological Tissues

4.17. RESEARCHING MODELS WITH AN ULTRASONIC ECHOSCOPE

Patient Safety Focused QA. LDR Brachytherapy Vrinda Narayana

The Physics of Ultrasound. The Physics of Ultrasound. Claus G. Roehrborn. Professor and Chairman. Ultrasound Physics

Ultrasound: Past and Present. Lecturer: Dr. John M Hudson, PhD

Augmentation of the In Vivo Elastic Properties Measurement System to include Bulk Properties

Trina Lynd, M.S. Medical Physicist Lifefirst Imaging & Oncology Cullman, AL Tri-State Alabama, Louisiana and Mississippi Spring 2016 Meeting April

Basic Ultrasound Physics Board Review Questions

INTRODUCTION. Getting the best scan. Choosing a probe. Choosing the frequency

Pulse-Echo Ultrasound Imaging. Resolution in Ultrasound Imaging. Doppler Ultrasound. Resolution vs Penetration. Medical Imaging (EL582/BE620/GA4426)

CHARACTERIZATION OF ANNULAR ARRAY TRANSDUCER

Annular Array Transducer and Matched Amplifier for Therapeutic Ultrasound

ACR Accreditation Update in MRI

1 Fundamentals. Basic Definitions and Physics Principles. Fundamentals

Automatic quality control processing for detection of element/channel dropout in ultrasound transducers

Hands-on. Physics Workshop. 2-day Workshop June 1-2, 2018 Ann Arbor, MI. Program Director: Brian Fowlkes, PhD

Ultrasound in Medicine

Features and Benefits

Ultrasonic Thickness Procedure Pulse-Echo Thickness Gage TABLE OF CONTENTS. 1.0 Introduction Reference Documents Summary of Practice 2

Available online at ScienceDirect. Physics Procedia 70 (2015 )

Evaluation of the Quality of Thick Fibre Composites Using Immersion and Air- Coupled Ultrasonic Techniques

Review of TG-186 recommendations

High resolution ultrasound scanner for skin imaging

10 Years Experience in Industrial Phased Array Testing of Rolled Bars

ULTRASONIC ARRAY APPROACH FOR THE EVALUATION OF ELECTROFUSION JOINTS OF POLYETHYLENE GAS PIPING

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Example 1: How would you evaluate this transducer? What is the most common problem found in ultrasound QC testing? Uniformity is Subjective

Compact Gamma Camera for Detection of Prostate Cancer

Ultrasound Physics & Doppler

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Interim Report. Prostate Cancer Fight Foundation Motorcycle Ride for Dad. February 7, Supporting Kingston s university hospitals:

Feng Xiujuan National Institute of Metrology (NIM),China

Characterization of Ultrasound Elevation Beamwidth Artefacts for Brachytherapy Needle Insertion

AAPM Annual Meeting. ACR Accreditation Update in CT

Performance of phased array and conventional ultrasonic probes on the new ISO reference block

Musculoskeletal Ultrasound: Basics, Utility, and Clinical Applications

Quality Assurance of Ultrasound Imagers: Procedures, Expectations, and Philosophies

Confident and Conclusive Diagnosis with Flexible Solutions

API. Defined Procedure. for. Ultrasonic Thickness Measurement API-UT-21

Joon-Il Choi, MD, Pyo Nyun Kim, MD, Woo Kyoung Jeong, MD, Hyun Cheol Kim, MD, Dal Mo Yang, MD, Sang Hoon Cha, MD, Jae-Joon Chung, MD

Skin Characterization with High-Frequency Ultrasound

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy

On the feasibility of speckle reduction in echocardiography using strain compounding

Pipeline Technology Conference 2007

Imaging Rotation. University of Michigan Department of Radiation Oncology Division of Radiation Physics. Resident:

What is Ultrasound? What is Ultrasound? B A. Basic Principles of Ultrasound. Basic Principles of Ultrasound. Basic Principles of Ultrasound

Quality management for Breast Brachytherapy.

OPTION I TEST REVIEW

Transcription:

TG-128: Quality Assurance for Prostate Brachytherapy Ultrasound STEVEN SUTLIEF DOUG PFEIFFER (HEATHER PIERCE, WENGZHENG FENG, JIM KOFLER) AAPM ANNUAL MEETING 2010

Educational Objectives To describe the methods of TG-128: Quality assurance tests for prostate brachytherapy ultrasound systems. To highlight special techniques which may be used to accomplish the tests described in TG-128. To review the materials and time needed to complete the TG-128 tests.

Outline Ultrasound physics Equipment TG128 tests Tolerances Time estimates Materials

Report AAPM Task Group 128: Quality assurance tests for prostate brachytherapy ultrasound systems. Medical Physics, Vol. 35, No. 12, December 2008, pages 5471-5489. The report is available on the Publications page of the AAPM web site.

Beam geometry Ultrasound physics

Ultrasound Physics Wavelength (axial resolution) λ = c f where f = ultrasound frequency c = speed of sound in the tissue Material Velocity ( m/s) Air 330 Water 1497 Bone 3500 Fat 1440 Blood 1570 Soft tissue 1540

Ultrasound Physics Sound Power Reflection Coefficient [ ] 2 [ Z 2 + Z 1 ] 2 α r = Z 2 Z 1 Z is the acoustic impedance Z 1 = 3500 (bone) Z 2 = 1540 (soft tissue) α r = 0.15 Stronger boundary signal Z 1 = 1450 (fat) Z 2 = 1540 (soft tissue) α r = 0.0009 Weaker boundary signal

Ultrasound physics The attenuation coefficient for a typical clinical tissue density and ultrasound frequency is 0.5 db/cm/mhz. P P o P P o = 0.5 db/cm/mhz 3 MHz 5 cm 2 = = = 0.5 db/cm/mhz 6 MHz 5 cm 2 = 15/10 15 db = 10 0.03 (deeper resolution) 30/10 30 db = 10 = 0.001 (less deep resolution)

Ultrasound Physics So why would you ever use 6 MHz? λ = c f You can t measure smaller than your wavelength. 1540 m/s λ = = 0.5 mm 3 MHz (finer resolution) 1540 m/s λ = = 0.3 mm 6 MHz (coarser resolution)

US equipment Probe

System US equipment

Prostate Brachytherapy US QA Complete QA will include: Ultrasound unit Needle template Treatment planning system Fluoroscope CT for post implant Template Needle Prostate Ultrasound Probe Ultrasound System

Setup for Phantom Measurements Clinical perspective, but coupling gel can leak out

Setup for Phantom Measurements Coupling gel stays in place, but image can be confusing

QC Testing Limit inter-observer variability Ideally performed by a single individual If multiple people, train for consistency

The Phantom CIRS Model 45 phantom. Used here for illustrative purposes only; no endorsement is implied. Wires spaced at known intervals Volumetric objects. TG128 report recommends a phantom design, but no manufacturer has implemented it yet

Test 1: Grayscale visibility Monitor must be adjusted appropriately for optimal visibility Calibration and brightness can drift over time Phosphors can fade

Test 1: Grayscale visibility Locate the gray scale strip on the side of the ultrasound screen. Depending on the type of strip, count the number of gray levels or measure the length of the gradation.

Test 2: Depth of penetration Reduced depth of penetration can reduce visibility of anterior capsule border Any increase in system noise will reduce depth of penetration Dead transducer elements (reducing signal) will also reduce depth of penetration

Test 2: Depth of penetration Find a relatively homogeneous region in the phantom. Using the digital calipers, determine the maximum depth that the static ultrasound speckle pattern of the phantom can be clearly distinguished from the dynamic electronic noise.

Test 3: Spatial resolution Negatively impacted by Poor probe condition Problems with pulse formation circuit boards Problems with pulse send/receive circuit boards 1 mm 2 mm 3 mm 4 mm 5 mm

Test 3: Spatial resolution Find a region of the phantom having single filament targets at various depths. Measure the dimensions of the filament image in both the axial and lateral directions. These dimensions are effectively the axial and lateral resolution limits. Switch the probe to the orthogonal direction and repeat

Test 4: Distance measurement accuracy Distance determined by pixel size and pixel calibration Pixel depth determined by range equation Differences from 1540 m/s sound speed will lead to errors in distance measurement Lateral distance calibration determined by FOV and pixel calibration. Errors can stem from image processing board and other circuitry

Test 4: Distance measurement accuracy Axial measurement: Align a column of fiber targets near to the center of the image, if possible. Freeze the image. Using the electronic calipers, measure the distance between the most proximal and the most distal targets. Lateral measurement: Repeat using a row of targets, measuring most lateral targets.

Test 5: Area measurement accuracy Area measurement is central to the implant procedure Since area measurement and distance measurement are so closely related, they have similar fault causes

Test 5: Area measurement accuracy Scan an object of known dimension such that the ultrasound beam intercepts it normally Using the appropriate tool on the ultrasound system, carefully trace the boundary of the object and record the calculated area of the object Measured area = 3.1 Nominal area = 3.05

Test 6: Volume measurement accuracy Complementary to distance and area measurements Errors are compounded multiplicatively

Test 6: Volume measurement accuracy Measured volume: 20.8 cc Certified volume: 20.6 cc Locate the base and apex of the phantom target; zero the stepper at the base Using the typical clinical procedure, perform a volume study After contouring the entire target, record the calculated volume

Test 7: Needle template alignment Depends on Accuracy of electronic template Distance measurement accuracy Physical needle template location

Test 7: Needle template alignment Place the probe with the needle template attached vertically in the water bath. Place needles at each corner of the needle template and one at the center. On the US system, verify that needle flashes in the image correspond to locations of needles on electronic grid overlay.

Test 8: TPS volume accuracy Perform a volume study of 3D target in the US phantom Import ultrasound images into treatment planning computer Retrace contours in treatment planning software Compare TPS volume to volume calculated by US system Variseed volume: 21.4 cc (3.9%) US Measured volume: 20.8 cc Certified volume: 20.6 cc

Tolerances Test Test name Tolerances 1 Grayscale visibility Δ > 2 steps or 10% from baseline 2 Depth of penetration Δ > 1 cm from baseline 3 Axial and lateral resolution Δ > 1 cm from baseline 4 Axial and lateral distance accuracy Error > 2 mm or 2% 5 Area measurement accuracy Error > 3 mm or 3% 6 Volume measurement accuracy Error > 5% 7 Needle template alignment Error > 3 mm 8 Treatment planning computer volume accuracy Error > 5%

Frequency # Test name Frequency 1 Grayscale visibility Annual 2 Depth of penetration Annual* 3 Axial and lateral resolution Annual* 4 Axial and lateral distance measurement accuracy Annual 5 Area measurement accuracy Annual 6 Volume measurement accuracy Annual 7 Needle template alignment Annual** 8 Treatment planning computer volume accuracy At Acceptance *Or following transport to another facility. **Or when a new template is used.

Time Estimates Test # Test name Typical duration 0 Gather and filling in preliminary information 10 minutes 1 Grayscale visibility 2 minutes 2 Depth of penetration 2 minutes 3 Axial and lateral resolution 1-5 minutes 4 Axial and lateral distance measurement accuracy 5 minutes 5 Area measurement accuracy 5 minutes 6 Volume measurement accuracy 10 minutes 7 Needle template alignment 15 minutes 8 Treatment planning computer volume accuracy 15 minutes Total: 70 minutes

Materials # Test name Equipment 1 Grayscale visibility Phantom 2 Depth of penetration Phantom 3 Axial and lateral resolution Phantom 4 Axial and lateral distance measurement accuracy Phantom 5 Area measurement accuracy Phantom 6 Volume measurement accuracy Phantom 7 Needle template alignment Water bath 8 Treatment planning computer volume accuracy TPS, Phantom

Ring-down artifact The signal is reflected multiple times within the needle.

Moisture in probe connection Electronic artifact

Poor contact between probe and object Contact artifact

Concluding Remarks Each of the tests rely on establishing a set of baseline measurements against which future measurements can be compared. The overall time commitment is manageable: the full set of measurements should be performed annually and will take about 90 minutes.