Update on the ACR FFDM QC Manual

Similar documents
Update on MQSA and Mammography Accreditation

Overview. ACR Accreditation Update in Mammography. New ACR Activities. Requirements Today. What s New For Tomorrow. ACR: Recognized by FDA and CMS

Overview. ACR Update on FFDM Accreditation. New ACR Activities. QC Today. QC Tomorrow. ACR Breast Imaging Centers of Excellence BICOE

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

Kish chakrabarti, Ph.D. Senior Physicist CDRH/FDA

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

The American College of Radiology Digital Mammography QC Manual: Frequently Asked Questions (Revised 12/12/2018; new and updated items in red)

Mammography Quality Control: A Refresher

Implementation of a New Tomosynthesis Program: A Physicists Perspective

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Q. Where can I find out if my state currently requires stereotactic breast biopsy accreditation?

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

AAPM Annual Meeting. ACR Accreditation Update in CT

HISTORY OF MQSA AND ACR

CT Quality Control Manual FAQs

MQSA Physicist Qualifications. Jon J. Erickson, Ph.D., DABR

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Dosimetry in digital mammography

Maintaining a quality mammography facility: Overview of ACR and MQSA Requirements and Update on BI-RADS 5 TH Edition

Opportunities and Innovations in Digital Mammography John M. Sandrik, Ph.D. GE Healthcare Milwaukee, WI

Radiation Dosimetry in Digital Breast Tomosynthesis. March, 2015 William J. O Connel, Dr. Ph, Senior Medical Physicist

ACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program

IAEA TECDOC 1517 Quality Control in Mammography Software

The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 5000

Multicenter Trial, Phase I Assessment of 2-D FFDM Versus Combo of 2-D and 3-D Tomosynthesis in Breast Cancer Screening

Digital mammography imaging from Carestream Health solutions for great workflow, productivity, and patient care.

MAMMOGRAPHY QC: MORE THAN PHANTOM IMAGES. Stephanie Schofield RTR QC Technologist Nova Scotia Health Authority

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

Medical Diagnostic Imaging

Hong Kong College of Radiologists Mammography Statement

Radiographic techniques in screen-film mammography

2001 AAPM Summer School Seattle, Washington ACR R/F Phantom

Direct half value layer measurements in mammography - is near enough good enough?

Radiation Physics Consultants, Inc. Medical Physicist. Mammography Qualifications, Continuing Experience and Continuing Education

WOMEN S HEALTH SOLUTIONS. The power and promise of breast tomosynthesis is here. Selenia Dimensions system with Acquisition Workstation 8000

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions (Revised: July 31, 2017)

MRI FAQs. After reading these documents and checking your protocols, you can apply online here:

Nuts and Bolts of MRI Physics Tests or Selling Preparation H. Robert A. Bell, PhD. Learning Objectives

POSITIONING ACR REQUIREMENTS IMAGE REVIEW CATEGORIES DEFICIENCIES IN POSITIONING ACR REQUIREMENTS 3/28/2016 NUMBER 1 REASON FOR ACR FAILURE

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

Mammographic Accreditation Phantom

ACR Accreditation Update in MRI

ACR MRI Accreditation: Medical Physicist Role in the Application Process

Hands-On Ultrasound Physics and Quality Control Workshop. Outline

Effects of Quality Assurance Regulatory Enforcement on Performance of Mammography Systems: Evidence From Large- Scale Surveys in Taiwan

Routine Quality Assurance Cookbook

FDA Executive Summary

1. Introduction. 2. Technical procedures

CASE STUDIES. Population. Compression

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

The Mammography Examination

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Dental ConeBeam Computed Tomography (CBCT) X-ray Systems

Low Dose CT Lung Screening: What is Technically Required?

SenoBright Contrast Enhanced Spectral Mammography Technology. Ann-Katherine Carton Sylvie Saab-Puong Matt Suminski

Introduction 1. Executive Summary 5

Proven clinical effectiveness at low radiation dose

9/3/2015. New Mammographic Modality Training. Mammographer. Qualified Radiologic Technologist. Tomosynthesis Training

CT Low Dose Lung Cancer Screening. Part I. Journey to LDCT LCS Program

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

History of MQSA ACR creates testing procedures Oct Mammography Quality Standards Act Dec 1993 Interim Regulations Oct All mammog

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

The New Jersey Radiographic Quality Assurance Program at 5 Years

Ultrasound Accreditation and Regulatory Compliance from A to Z

Contrast-Enhanced Digital Mammography

Breast Tomosynthesis An additional screening tool in the fight against breast cancer

Dental Extraoral X-ray Systems

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company

Prepublication Requirements

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Dental Intraoral X-ray Systems

X-Ray & CT Physics / Clinical CT

About me 10/9/2018. Digital Breast Tomosynthesis (DBT)

Introduction and Background

Agenda: Dental Cone Beam Imaging

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Journal of Breast Cancer

Breast positioning system for full field digital mammography and digital breast tomosynthesis system

Dedicated to Mammography

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

Automated Practical System Qualification, Validation and Reporting acc. ASTM E

Steven Tate Nicholas

CHAPTER 2 MAMMOGRAMS AND COMPUTER AIDED DETECTION

Breast Imaging Essentials

Accreditation Case Review: Mammography and Stereotactic Biopsy

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

8/3/2016. DBT Physics Basic to Advanced: Primer On Tomosynthesis. Tomosynthesis Pedigree

Quality assurance in external radiotherapy

Computed tomography Acceptance testing and dose measurements

Mammography device use in Turkey, and quantity and quality analysis of mammography education

Ultrasound Accreditation Program Requirements

Example 1: How would you evaluate this transducer? What is the most common problem found in ultrasound QC testing? Uniformity is Subjective

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

ACR Accreditation Toolkit for Validation Site Surveys

Objectives. Explanation of Radiation Dose Terminology 10/9/2018. What are these lines?

-Ray Producing Equipment

Transcription:

Update on the ACR FFDM QC Manual Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO AAPM 2011 Vancouver, Canada 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000 3500 3000 2500 2000 1500 1000 500 0 FFDM in the US US Full-Field Digital Mammography (FFDM) Units and Facilities Facilities Units October 1 of each year 2004 2005 2006 2007 2008 2009 2010 As of 7/1/11 9,674 units at 6,714 facilities Over 79% of all units in US are FFDM FDA Approved ACR to Accredit GE 2000D, DS, Essential Hologic Digital Breast Fischer Tomosynthesis (DBT) SenoScan Lorad System approved for Selenia Siemens sale in US (Feb 2011) Novation, Inspiration Fuji FCRm (computed radiography) Carestream DirectView (computed radiography) *Sectra Microdose FDA Approved 4/28/2011 *In process of getting approval from FDA to accredit FDA s Current FFDM QC Requirements Follow latest version of mfr s QC manual procedures for unit tested Lorad (Hologic) allows facility to follow any of their manuals Meet mfr s performance standards Failures must be fixed before use on patients Most mfrs applied for alternative standards to allow 30 days for some QC tests 1

Workstations Monitors Printers VS. VS. All of the above may vary with QC manual revisions of same manufacturer/model Project ACR Subcommittee on Quality Assurance Clinical Representatives MITA Representatives ACR Representatives ACR Subcommittee on Quality Assurance Clinical Representatives Eric Berns, PhD University of Colorado - Chair Chris Adent-Delaney, RT - Northwestern Memorial Hospital Jay Baker, MD Duke University Medical Center Lawrence Bassett, MD UCLA Medical Center Chair, Joint Committee on Breast Imaging for Appropriateness Criteria and Guidelines Shelli Dixon, RT The Women s Imaging Center of Denver R. Edward Hendrick, PhD University of Colorado Hospital Debra Monticciolo, MD Texas A&M Health Sciences Center Chair of ACR Accreditation Program Chairs Chair of ACR Mammography Accreditation Douglas Pfeiffer, MS Boulder Community Hospital Margarita Zuley, MD University of Pittsburgh Medical Center 2

Subcommittee on Quality Assurance Project MITA Representatives Gail Rodriguez, PhD - MITA John Sandrik, PhD (Ret.) GE Medical Systems Robert Uzenoff - FUJIFILM Medical Systems Stephen Vastagh (Ret.) - MITA Moustafa Zerhouni Computerized Imaging References Systems ACR Representatives Marion Boston, RT Manager, ACR Breast Imaging Accreditation Priscilla Butler, MS Senior Director, ACR Breast Imaging Accreditation Programs Subcommittee Charge: Design ACR Accreditation Phantom for FFDM Write QC Manual for ACR FFDM Mammography Accreditation Program Project Subcommittee Goals: Standardize all QC tests for all digital manufacturers Standardize test frequencies Standardize performance criteria Project Subcommittee Goals: QC Tests: Tests come from a variety of sources (MQSA, ACR SFM, ACRIN DMIST, Manufacturer s QC programs, MITA, European Guidelines, subcommittee clinical experience, etc.) Clinically relevant User friendly This manual will become basis of new regulations 3

Project Project Subcommittee Goals: Update critical component of the ACR MAP Account for all past, present, and future FFDM systems Reasonable and appropriate for mass implementation Eliminate unnecessary complicated procedures & analysis Maximize user experience Especially for Techs & Rads & Facilities Theme Measurements be made with external equipment Dosimeters, photometers, etc. Minimal software requirements CNR & SNR Subcommittee Goals: Ensure that accredited systems Provide acceptable image quality While meeting or exceeding MQSA requirements Facilities (and Radiologists) Can feel confident that their system is performing in accordance with ACR requirements Are familiar with, and confident with their understanding of, their ACR FFDM QC Program ACR Digital QC Manual The QC Manual Structure of Manual: Radiologist s Section Clinical Image Quality Section Radiologic Technologist s Section Medical Physicist s Section Educational, Guidance, and Troubleshooting Section Glossary References Index 4

What Will Be New? What Will Be New? Radiologist Section Image ID regulations Hanging protocols (left vs. right) Monitor and viewing conditions guidance Section on diagnostic tools for analyzing poor images How to score the ACR FFDM Phantom Guides for understanding their role and responsibility for overseeing the QC program Tech Section Enhanced positioning and image quality section New Test: Monitor QC for the Radiologist New Test: Facility QC Review New Format: Action Log New Documentation: Facility Equipment Inventory Instructions for Mobile Units Eliminating calculations (Yet to be determined) What Will Be New? What Will Be New? Medical Physicist Section Include tests procedures and forms for all ACR and MQSA QC Tests Will include guidance on how to test Multiple units (FFDM s, AW s, RW s, Printers, etc) Multiple facilities Procedures for evaluating and documenting Tech QC Medical Physicist Section Theme: providing better documentation and communication Single MP Summary Form For Facility, ACR, State and MQSA Inspectors Include an Action Item Summary MP form for Tech for 0perating Levels and QC instructions MP letter to the Radiologist MP to use same Action Log form as Techs 5

What Will Be New? ACR Digital QC Manual Facility Guidance on how to handle multiple units at multiple locations. Guidance on who/what/when tests need to be performed when major and minor repairs are performed on unit. Facility QC Review (Tech Test) Quarterly Technologist QC Tests Test Number Name Frequency Required Action 1 ACR Phantom Image Quality Weekly Before Clinical Use 2 Acquisition Workstation (AW) Monitor QC Weekly Before Clinical Use 3 Radiologist Workstation (RW) Monitor QC Weekly Before Clinical Use 4 Laser Printer QC Weekly Before Clinical Use 5 Viewbox Cleanliness Weekly Before Clinical Use 6 Visual Checklist Monthly Before Clinical Use 7 Repeat Analysis Quarterly Within 30 Days 8 Monitor QC for the Radiologist Quarterly Before Clinical Use 9 Facility QC Review Quarterly Not Applicable 10 Compression Force Semiannual Before Clinical Use 11 Manufacturer Detector Calibration (If Applicable) Per Mfr Recommendation Before Clinical Use Supplemental Forms Action Log Facility Equipment Inventory Form ACR Digital QC Manual Medical Physicists QC Tests Test Number Name Frequency Required Action 1 ACR Phantom Image Quality Annual Before Clinical Use 2 Ghost Image Annual Before Clinical Use 3 Spatial Resolution Annual Before Clinical Use 4 Automatic Exposure Control System Performance Annual Before Clinical Use 5 Collimation Assessment Annual Within 30 Days 6 kvp Accuracy and Reproducibility MEE Only Before Clinical Use 7 Beam Quality (Half-Value Layer) Assessment Annual Within 30 Days 8 Average Glandular Dose Annual Before Clinical Use ACR Digital QC Manual Medical Physicists QC Tests Test Number Name Supplemental Forms 1 Medical Physicist Summary Report 2 Technologist Operating Level Information and QC Instruction Form 3 Medical Physicist Summary Letter for the Radiologist 4 Mammography Action Log 5 Technologist Pre-Inspection Interview Form 6 Technique Chart 9 Unit Checklist Annual Before Clinical Use 10 of Site s Technologist QC Program Annual Within 30 Days 11 MQSA Equipment Requirements MEE Only Before Clinical Use 12 Computed Radiography (If Applicable) Annual Before Clinical Use 13 Acquisition Workstation (AW) Monitor QC Annual Before Clinical Use 14 Radiologist Workstation (RW) Monitor QC Annual Before Clinical Use 15 Laser Printer QC Annual Before Clinical Use 16 Viewbox Luminance and Room Illuminance Annual Before Clinical Use 17 of Off-Site Technologist QC Program (If Applicable) Annual Before Clinical Use 6

ACR Digital QC Manual Tech & MP Test Number Name Educational and Example Forms 1 Complete set of forms with example data, scores, and calculations 2 ACR Phantom Scoring Guide 3 SNR & CNR Calculation Guide 4 Monitor Test Pattern Guide 5 Printed Film Guide 6 FFDM Artifact Guide The ACR FFDM Phantom Design Summary Phantom Design Principles Base on existing ACR Accreditation Phantom Similar imaging and scoring to current SFM phantom Can be used on both SFM & FFDM Total attenuation matched to current SFM phantom Similar thickness Similar total dose Permits testing of 3.0 mgy dose limit Design Summary Phantom Design Principles Cover all (or most) of detector on all digital systems Single exposure results in all relevant information All evaluation (including artifacts) can be done at one WW, WL Provide detailed specifications to manufacturer s Manufacturing will be open to all qualifying vendors Must receive approval from ACR (ACR will test sample phantom) 7

Differences from screen-film phantom Eliminate subtraction for artifacts Add Fail for artifacts Improve specific rules for scoring Change pass/fail criteria from 4,3,3 To: 2,3,2 Design Summary **But, objects are the same (effective) size as SFM Phantom The ACR FFDM Phantom The ACR FFDM Phantom The ACR FFDM Phantom vs. SFM ACR Phantom 8

The ACR FFDM Phantom Wax Insert Comparison Compensator FFDM SFM ID Tag Wax Insert ID Tag Acrylic Body Image of Entire Phantom Expanded view of Wax Insert Serial Number Serial Number *Note: Gray dot in lower left corner of wax insert is an artifact due to a bubble in wax insert. 9

Expanded view of Wax Insert Expanded view of Wax Insert With Pass/Fail criteria Pass Criteria: 2 Fibers, 3 Specks, 2 Masses Pass Serial Number CNR Image of Entire Phantom The ACR FFDM Phantom Effects of Thickness Equalization Mean = 542.3 Mean = 498.5 St. Dev = 7.8 Serial Number New FFDM phantom equalizes attenuation inside and outside wax insert. This permits evaluation of artifacts over entire phantom area with same WW and WL used to score test objects. 10

The ACR FFDM Phantom CR 24x30 The ACR Prototype Screen-Film 18x24 The ACR Prototype Screen-Film 18x24 The ACR Prototype Screen-Film 18x24 Heel effect Screen defects Non-uniform OD from dark center to light edge Roller marks Dust 11

Design Summary AEC Technique Comparison Measurements using the FFDM Phantom Lorad Mo Lorad - W Fuji CR 18 x 24 cm Fischer Phantom used in 3 of 11 Tech tests 12 of 24 Tech sub-tests 7 of 17 Physicists tests 50% Tech Tests 41% Physics Tests Mode Auto-Filter Auto-Filter AA Auto-Technique Phantom FFDM SFM FFDM SFM FFDM SFM FFDM SFM Compression 5.2 5.2 5.2 5.2 4.0 4.0 5.74 4.05 Thickness (cm) Target/Filter Mo/Mo Mo/Mo W/Rh W/Rh Mo/Mo Mo/Mo W/Al W/Al kvp 29 29 28 28 27 27 31 27 19 of 48 Physicist sub-tests mas 66.4 65.4 92.5 97.6 90 89 177 ma 158 ma Machine Reported Dose 1.64 1.61 1.03 1.08 ** ** 0.954 1.211 (mgy) Manual Technique Signal Comparison Mode Lorad Mo Manual Phantom FFDM SFM Target/Filter Mo/Mo Mo/Mo kvp 29 29 mas 65 65 Signal Wax 542.0 546.5 St. Dev. Wax 9.7 9.7 ACR Digital QC Manual Benefits of Phantom Design Provides view of entire detector artifact evaluation W/L optimized for test objects optimizes for artifact eval Finer gradations of test objects Test objects go to smaller sizes AGD measurement & limit same as SFM Meets MQSA Provides single image/exposure for evaluation(s) Minimal training (~ 25,000 Techs currently trained) Provides basis for monitor and laser printer QC ACR Physics Reviewers Can see scores and artifacts on single submitted film (or image) Do not need different WW/WL settings 12

Technologist Tests Test # Test Name Test Element Frequency [1] Performance Criteria Action Time The QC Tests 1 ACR Phantom Image Quality (Acquisition Workstation) Phantom Average Glandular Dose (AGD) Check Automatic Exposure Control Consistency Artifact Machine reported AGD must be 3.0 mgy Machine reported AGD must be within ± 0.20 mgy of Operating Level (OL) No clinically significant artifacts NA if AGD is NA OL value obtained from MP, use mas if AGD is NA Phantom Scoring Acquisition Workstation Fibers 2.0 Masses 3.0 Specks 2.0 [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. Technologist Tests Technologist Tests Test # Test Name Test Element Frequency [1] Performance Criteria Action Time Test # Test Name Test Element Frequency [1] Performance Criteria Action Time Monitor Cleanliness Monitor(s) must be clean Monitor Cleanliness Monitor(s) must be clean 2 Acquisition Workstation QC Test Pattern Manufacturer s Recommended QC Tests Visual evaluation must pass (SMPTE or TG18) Per Manufacturer Per Manufacturer Per Manufacturer On monitors where test pattern is available Mfr s test(s) established by MP (If Applicable) 3 Radiologist Workstation QC Test Pattern Artifact ACR Phantom Scoring Visual evaluation must pass (SMPTE or TG18) No clinically significant artifacts Fibers 2.0 Masses 3.0 Specks 2.0 Manufacturer s Recommended QC Tests Per Manufacturer Per Manufacturer Per Manufacturer Mfr s test(s) established by MP (If Applicable) [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. 13

Technologist Tests Technologist Tests Test # Test Name Test Element Frequency [1] Performance Criteria Action Time Test # Test Name Test Element Frequency [1] Performance Criteria Action Time 4 Laser Printer QC Artifact Check ACR Phantom Scoring Background Optical Density Check No clinically significant artifacts Fibers 2.0 Masses 3.0 Specks 2.0 Background OD must be within ± 0.15 OD of OL OL value give by MP 5 6 Viewbox Cleanliness Check Visual Checklist Viewboxes Clean Visual Checklist Monthly Document all cleaning procedures and frequencies Must pass all tests Contrast Optical Density Check Dmax Optical Density Check Density difference (DD) must be within 0.05 of OL 3.5 OD OL value give by MP 7 Repeat Analysis Repeat Analysis Quarterly If the total repeat changes from previously determined rate by more than 2.0% of the total images included in the analysis, the reason(s) for the change must be determined test date [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. Test # Test Name Test Element 8 9 10 11 Monitor QC For The Radiologist Facility QC Review Compression Force Manufacturer Detector Calibration (If App). Imaging Chain Spot Check QC Review with Lead Interpreting Physician & Facility Manager Compression Force Manufacturer Detector Calibration Technologist Tests Frequency [1] Performance Criteria Action Time Quarterly Must pass all tests Quarterly Semi-annually Per Manufacturer Recommendation QC review signed by lead interpreting physician, facility manager, and QC technologist Initial power drive max force between 25 and 45 lbs (11.1 and 20.0 decanewtons) Per Manufacturer Recommendation NA [1] Daily and weekly tests must be performed on all days/weeks that the component of the mammography system is used. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. No. Test Name Test Element 1 ACR Phantom Image Quality (Acquisition Workstation) Medical Physicist s Tests Artifact Phantom Scoring Acquisition Workstation Exposure Duration Signal-to-Noise Ratio Measurement Contrast-to-Noise Ratio Verification Distance Measurement Check Frequency Action Limits Action Time Fibers 2.0 Masses 3.0 Specks 2.0 No clinically significant artifacts Total exposure time for the ACR Phantom must be 2.0 seconds SNR must be > 40 CNR TBD 70 mm + 7.0 mm All clinical target/filter combinations Contact and Mag modes Action Limit TBD during pilot testing Action Limit TBD during pilot testing Applicable if software provides measurement tool [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. 14

Medical Physicist s Tests Medical Physicist s Tests No. Test Name Test Element Frequency Action Limits Action Time No. Test Name Test Element Frequency Action Limits Action Time 2 Ghost Image Ghost Image Ghost Image Value must be within + 0.3 X-ray Field & Detector Alignment X-ray field does not extend beyond IR by more than 2% of SID X-ray field extends all the way to chest wall Measured for largest paddle only 3 Spatial Resolution Bar Pattern Contact & Mag Modes See table in QC Forms 5 Collimation Assessment Light Field & X-ray Field Alignment Compression Paddle Alignment Light field & X-ray field alignment (Length or Width) must not exceed 2% of SID Chest edge of compression paddle must not extend beyond chest-wall edge of IR by more than 1% of SID CR must measure at both IR sizes 4 Automatic Exposure Control System Performance SNR Thickness Tracking (2, 4, 6 cm, Large, Mag) Density Control Function (if applicable) Must maintain signal value to within + 10% of average Each step should result in 12% to 15% change in mas & mean signal value 6 kvp Accuracy and Reproducibility kvp Accuracy and Reproducibility Only upon installation Must be accurate within ± 5% of the indicated kvp at -Lowest clinical kvp that can be measured by a kvp test device -Most commonly used clinical kvp -Highest available clinical kvp Coefficient of variation must be 0.02 [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. Medical Physicist s Tests Medical Physicist s Tests No. Test Name Test Element Beam Quality (Half-Value 7 HVL Layer) Assessment Frequency Action Limits Action Time Must meet upper and lower criteria in 1999 ACR Mammography QC Manual. (note: both upper and lower criteria are important to maintain a check on beam quality in the recommended absence of routine kvp tests; if necessary, the MP may conduct kvp testing to investigate outliers) No. Test Name Test Element Frequency Action Limits Action Time 9 Unit Checklist Unit Annually Must pass all tests 10 of Site s Technologist QC Program of Site s QC Annually Tech QC tests pass MP review and corrective action documented correctly 8 Average Glandular Dose Average Glandular Dose Measurement AGD delivered during a single cranio-caudal view of an attenuator simulating the attenuation of a standard breast must be 3.0 mgy (0.3 rad) per exposure 11 MQSA Equipment Requirements MQSA Equipment Requirements Upon installation (MEE) Must meet MQSA requirements MQSA Equipment Requirements Machine Indicated Average Glandular Dose (AGD) Check Machine reported AGD must be within + 25% of measured AGD SNR Inter-plate Consistency Variation in mas must be within ± 10 of average Variation in SNR must be within ± 15% of average 12 Computed Radiography Tests (if applicable) Artifact on All Cassettes All plates - no clinically significant artifacts CR Reader Scanner Performance Edges on must appear smooth with no jagged edges [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted using the component of the mammography system that failed the test. 15

No. Test Name Test Element Medical Physicist s Tests Frequency Action Limits Action Time No. Test Name Test Element Medical Physicist s Tests Monitor Screen Frequency Action Limits Action Time Screen must be free of scratches, defects, fingerprints, dust, or marks that may impede image interpretation. Monitor Screen Monitor screen must be free of scratches, defects, fingerprints, dust, or marks that may impede image interpretatoin. Defective Pixel Check Monitors must be free of any defective pixels that may impede image interpretation 13 Acquisition Workstation QC Defective Pixel Check Test Pattern Luminance Check (Black and White Level, Luminance Ratio) Luminance Uniformity DICOM Gray-Scale Display Function (GSDF) Check Manufacturer s Recommended QC Tests Monitors must be free of any defective pixels that may impede image interpretation Test pattern visual evaluation must pass (TG18- QC or SMPTE) Mfr Recommendations Typically: Lmin < 1.0 cd/m 2 Typically: Lmax > 450 cd/m 2 Lum Ratio > 250 <30% difference from average for CRT Recommendations <10% difference from average for LCD Error rate < + 10% of GSDF Per Manufacturer Per Manufacturer Applies only if monitor is capable of test Applies only if monitor is capable of test Applies only if monitor is capable of test Applies only if monitor is capable of test MP to decide if Mfr QC tests apply to FFDM 14 Radiologist Workstation QC Test Pattern ACR Phantom Scoring & Artifact Check Clinical Image Check (Monitor Comparison) Ambient Light Conditions Luminance Check (Black and White Level, Luminance Ratio) Luminance Uniformity Gray-Scale Display Function (GSDF) Check Test pattern visual evaluation must pass (TG18-QC) Fibers 2.0 Masses 3.0 Specks 2.0 No clinically significant artifacts Distance measured must be 70 mm + 7.0 mm When same clinical image is on each monitor: Background light levels must match Color tone must match Contrast must appear the same Ambient light should be < 10 lux Total darkness is not recommended Mfr Recommendations Typically: Lmin < 1.0 cd/m 2 Typically: Lmax > 450 cd/m 2 Lum Ratio > 250 Right-Left Ratio for Lmin & Lmax < 10% difference < 30% difference from average for CRT < 10% difference from average for LCD Error rate < + 10% of GSDF [1] All tests must be done during the Mammography Equipment (MEE) following installation of new systems; MEEs performed after equipment relocation or service must include relevant tests. Any MEE test failure must be corrected immediately. [2] Problems must be corrected within the specified time limit; after that, further examinations may not be performed nor any images printed or interpreted Manufacturer s Recommended QC Tests Per Manufacturer Per Manufacturer MP to decide if Mfr QC tests apply to FFDM using the component of the mammography system that failed the test. No. Test Name Test Element Medical Physicist s Tests Frequency Action Limits Action Time Tech Tests ACR Phantom Artifact No clinically significant artifacts ACR Phantom Scoring Fibers 2.0 Masses 3.0 Specks 2.0 ACR Phantom - Background Optical Density Check Background OD must be > 1.4 OD Background OD should be within ± 0.30 of 1.9 OD Background OD must be within + 0.15 OD of OL (if applicable) Background OD 15 Laser Printer QC ACR Phantom - Contrast Optical Density Contrast OD must be > 0.10 OD Contrast OD must be within 0.05 of OL (if applicable) ACR Phantom - Dmax Check 3.5 OD ACR Phantom - Printed Size Check Distance measured must be 70 mm + 7.0 mm Size of printed image appears to be correctly sized Phantom appears to be correctly positioned on the film Test Pattern Check Test pattern must pass visual evaluation (TG-18) 16

Tech Tests Tech Tests Contrast OD Dmax No. Test Name Test Element 16 Viewbox Luminance and Room Illuminance Medical Physicist s Tests Viewbox Luminance Room Illuminance Frequency Action Limits Action Time ACR Recommends > 3,000 cd/m2 ACR Recommends: Illuminance at surface < 50 lux Illuminance seen by observer < 50 lux QC Tests Summary We did take into consideration the following: 17 of Off-Site Technologist QC Program of Off- Site s QC Annually Tech QC tests pass MP review and corrective action documented correctly MQSA, ACR SFM Manual, ACRIN DMIST results, Manufacturer s QC programs (FFDM, CR, Monitor, Printer), MITA, European Guidelines, AAPM TG18, and others Subcommittee and others clinical experience 17

What s Next The Approval Process Preemptive Questions When ready, draft will be sent to manufacturers for their input before it is sent to FDA We hope manufacturers will adopt this manual Draft should be completed in 2011 for review by FDA When final, ACR will apply for FDA alternative standard under current regulations Alternative standard will allow facilities to use this instead of the manufacturer s manuals Potential for ACR QC Manual to be basis for new MQSA Regulations Cost of phantom? Don t know. Reason to believe it will be affordable. Implementation and roll-out? ACR to develop a plan to include some sort of training. When? Can t be too soon! We re working double-time to get this completed. Preemptive Questions Tomo? Perhaps. Preliminary testing looks very promising. CEDM? Phantom does show objects in single plane End of Presentation Questions? Yet to be tested. 18