IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS. A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

Similar documents
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Intrafractional Junction Shifts Utilizing Multileaf Collimation: A Novel CSI Planning Technique. Rodney Hood RT(R)(T)CMD

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

3D-CRT Breast Cancer Planning

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

IMRT - Intensity Modulated Radiotherapy

The objective of this lecture is to integrate our knowledge of the differences between 2D and 3D planning and apply the same to various clinical

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

A simple technique for craniospinal radiotherapy in the supine position

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Hybrid VMAT/IMRT Approach to Traditional Cranio-Spinal Irradiation (CSI): A Case Study on Planning Techniques and Delivery

Many vendors are beginning to allow couch motion during radiation delivery.

Work partially supported by VisionRT

A Smart Setup for Craniospinal Irradiation

Radiotherapy techniques in brain tumors (2D, 3D CRT, IMRT) including craniospinal irradiation

Importance of daily portal imaging for Head and Neck IMRT treatments. Luciant Wolfsberger Brigham & Women s Hospital

Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer

Reducing excess radiation from portal imaging of pediatric brain tumors

Collimator rotation in volumetric modulated arc therapy for craniospinal irradiation and the dose distribution in the beam junction region

Feasibility of 4D IMRT Delivery for Hypofractionated High Dose Partial Prostate Treatments

Robust Optimization accounting for Uncertainties

Specification of Tumor Dose. Prescription dose. Purpose

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Treatment Planning & IGRT Credentialing for NRG SBRT Trials

Chapter 7 General conclusions and suggestions for future work

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Exploring the potential for clinical introduction of hippocampal-sparing IMPT of pediatric medulloblastoma - Past present and future

Future upcoming technologies and what audit needs to address

Overview of MLC-based Linac Radiosurgery

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Proton Stereotactic Radiotherapy: Clinical Overview. Brian Winey, Ph.D. Physicist, MGH Assistant Professor, HMS

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Role of Belly Board Device in the Age of Intensity Modulated Radiotherapy for Pelvic Irradiation

IMRT Planning Basics AAMD Student Webinar

Method to plan, administer, and verify supine craniospinal irradiation

Dose prescription, reporting and recording in intensity-modulated radiation therapy: a digest of the ICRU Report 83

Additional Questions for Review 2D & 3D

A review of RTTQA audit. Karen Venables

SRS Uncertainty: Linac and CyberKnife Uncertainties

Intensity modulated radiotherapy (IMRT) for treatment of post-operative high grade glioma in the right parietal region of brain

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

Original Date: April 2016 Page 1 of 7 FOR CMS (MEDICARE) MEMBERS ONLY

Clinical experience with TomoDirect System Tangential Mode

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida

ICRU Report 91 Was ist neu, was ändert sich?

Measurement of Dose to Critical Structures Surrounding the Prostate from. Intensity-Modulated Radiation Therapy (IMRT) and Three Dimensional

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS

Case Study. Institution Farrer Park Hospital

8/2/2017. Improving Dose Prescriptions for Safety, Reporting, and Clinical Guideline Consistency. Part III

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Treatment Planning for Lung. Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Page 1. Helical (Spiral) Tomotherapy. UW Helical Tomotherapy Unit. Helical (Spiral) Tomotherapy. MVCT of an Anesthetized Dog with a Sinus Tumor

Margins in SBRT. Mischa Hoogeman

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Address for Correspondence: Department of Medical Physics, Khwaja Yunus Ali University, Enayetpur, Sirajgonj ,

Defining Target Volumes and Organs at Risk: a common language

Mesothelioma XRT: From Old School to New School. A brief walk down memory lane at UCLA

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

NSABP PROTOCOL B-39B RTOG PROTOCOL 0413

IMPT with Carbon Ions

biij Initial experience in treating lung cancer with helical tomotherapy

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

TOMOTERAPIA in Italia: Esperienze a confronto

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Comparing conformal, arc radiotherapy and helical tomotherapy in craniospinal irradiation planning

Potential benefits of intensity-modulated proton therapy in head and neck cancer van de Water, Tara Arpana

Elekta 2017 Australasian User Meeting 12 th November 2017, Newcastle, NSW

Disclosures. Overview 8/3/2016. SRS: Cranial and Spine

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10

Optimization of RapidArc for Head-and-Neck Radiotherapy. Jessica Emily Salazar. Department of Medical Physics Duke University.

Quality Assurance of Helical Tomotherapy Machines

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

IVD TIMING test 1 test at 1 fraction 2 test at 2 fraction 3 test at 3 fraction 4 test during next week 5 test during next week 6 test during next week

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Department of Radiation Oncology SMART. Stereotactic MR-guided adaptive radiotherapy. Marloes Jeulink Omar Bohoudi

Savita Dandapani, MD, PhD City of Hope Duarte, CA

Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations

tomotherapy accurate treatment, fewer side effects bupacromwellhospital.com

EXACTRAC HIGHLY ACCURATE PATIENT MONITORING

Medical Errors in Radiation Therapy 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Kenny Guida, DMP, DABR March 21 st, 2015

CALCULATION OF OPTIMAL MARGINS BETWEEN CLINICAL TARGET VOLUME (CTV) AND PLANNING TARGET VOLUME (PTV)

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

Transcription:

IMRT FOR CRANIOSPINAL IRRADIATION: CHALLENGES AND RESULTS A. Miller, L. Kasulaitytė Institute of Oncolygy, Vilnius University

Content 1.Introduction 2.Methods and materials 3.Results and discussion 4.Conclusion

1.Introduction Craniospinal irradiation is very important part in the treatment of patients with primitive neuroectodermal tumors such as medulloblastoma. Craniospinal irradiation (CSI) is technically very challenging because of the difficult geometry of the treatment fields, high radiosensitivity of the spinal cord and other critical structures around.

Introduction (cont) The mechanical properties of available equipment make possible treatment only with treatment field junctions and several isocenters. Thus causing potential problems under/over dosage of the spinal cord in the areas of field junction and over dosage of adjacent structures. IMRT technique can be used in craniospinal and extra-cranial tumor treatment due to better dose coverage, more homogenous dose distribution and fewer doses to adjacent normal tissues.

1.1 Aim of the study The aim of our study was to develop IMRT technique for CS irradiation, to measure patient set-up errors during treatment and to compare possible affection of these uncertainties to the dose statistics of 3D CRT and IMRT treatment plans.

2. Methods and materials Two young patients (age 31-35 years) with medulloblastomas were treated between 2009-2010. One patient had craniospinal treatment administered, and the second only extracranial (spinal) treatment. A dose of 32,4 Gy was prescribed to the PTV with 18 fractions. The PTV was created by adding 1,0 cm margin to CTV in all directions. Outlined organs at risk (OAR) included eyes, lenses and kidneys.

2.1 Treatment plans There were three plans generated for each patient - two 3D CRT and one IMRT treatment plans. Two 3DCRT plans had differences in field junction. First plan had field junction with 3 mm gaps in cranial-spinal area and spinal-spinal area. It was set on even days. Second treatment plan had no gaps (exact match) in cranial-spinal and spinal-spinal junction areas, and it was set on odd days.

3D CRT treatment plan consisted of 2 opposing cranial fields for craniospinal case and three spinal fields one upper and two lower fields. All spinal treatment fields had couch rotation to 90 degrees. Extracranial treatment plan consisted of tree spinal treatment fields.

IMRT treatment plan consisted of 2 cranial fields and 6 spinal fields. No spinal field had couch rotation.

2.2 Treatment plan verification Patient related IMRT QA was performed before treatment. Verification plan created within the treatment planning system with portal dose prediction algorithm was verified on linac portal imaging system. During the treatment on-line verification was performed before every session, and subsequently required adjustment of the patient position was made before plan delivery.

3.Results and discussion The IMRT and 3D CRT treatment plans comparison was performed and it revealed the apparent advantage of IMRT plan in terms of dose distribution homogeneity in the target volume.

3.1 Dose distribution homogeneity Extracranial 3D CRT treatment plan.

3.2 Dose distribution homogeneity Extracranial (spinal) and CSI IMRT treatment cases.

3.3 Set up verification It has been proven that IMRT method, rightfully applied, provide noticeably better dose distribution for targets and critical organs for many different sites. It is very important to demonstrate the IMRT plan stability and robustness to every day variation in patient position.

Therefore, the data of on-line patient setup errors was analyzed. Set up verification (cont)

3.4 Evaluation of set up verification Maximum setup error was found to be in our case 3mm in LNG and VRT directions. Influence of these maximum Original plan Cranial isocenter 3 mm lateral shift 3mm vertical shift Upper spinal isocenter 3 mm lateral shift 3 mm vertical shift Lower spinal isocenter 3 mm lateral shift 3 mm vertical shift Max dose Mean dose 108% 96 % 111,3% 99% 120% 94,6% 109,7% 94% 108,4% 94,6% 109,5% 93% 107,5% 94,3%

Evaluation of set up verification (cont) Simulated change in isocenter position for 3D CRT plans in order of 3mm magnitude gives rise of up to 147% maximum dose in the junction area. Additionally, 3D CRT plans might have more than 3 mm isocenter move as soon as field setup involves table rotation to 90-degree angle.

3.5 Whole body dose This issue is of paramount importance in case of medulloblastoma treatments, because of young patient age and relatively good prognosis. The dose to the normal surrounding tissues strongly depends on the number of fields involved and because of the interleaf leakage, partially on the number of monitor units for the particular plan. It is possible to produce sufficiently good IMRT plan with number of

Dose to the surrounding tissue is in all cases much better than for the 3D CRT plans. Whole body dose (cont)

Whole body dose (cont) Thought, the body dose from 1 to approximately 5 Gy is slightly lower for 3D CRT plan, the difference is small and the risk of secondary cancer incidence in this dose range is still unclear. The difference in body dose up from this point, however, is around two fold in favor of IMRT plan. The risk of secondary cancer development in this dose range (up to 4 cm from the field edges) is well known and depends strongly on the dose-volume parameter

4. Conclusions The target volume of medulloblastoma is usually very complex. Due to target long shape and different distance from skin surface it is possible to achieve homogenous dose distribution in the target only using IMRT treatment technique. IMRT technique also enables to ensure good treatment accuracy as this technique seems to be less sensitive to patient set-up uncertainties and there is no need for couch

Thank you.