Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction

Size: px
Start display at page:

Download "Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction"

Transcription

1 370 Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction Seong-Ho Kang 1, Eun Hee Lee 2, Geon Park 1, Sook Jin Jang 1, and Dae Soo Moon 1 1 Department of Laboratory Medicine, Chosun University Medical School, Gwang-Ju, Korea; 2 Greencross Reference Laboratory, Yongin, Kyunggi-do, Korea. Abstract. This study was designed to compare two automated systems and one manual system for hepatitis B virus (HBV) nucleic acid extraction. The two automated systems were the MagNA Pure 96 system (Roche Applied Science, Manheim, Germany) and the Chemagic system (Chemagen, Baesweiler, Germany), and the manual system was the QIAamp system (Qiagen, Hilden, Germany). Sixty-eight samples that were within the detection range of the Cobas Ampliprep/Cobas TaqMan (CAP/CTM) platform (Roche Molecular Systems, Manheim, Germany) were selected. Extracted viral nucleic acids from the three systems were quantified using an AccuPower HBV Quantitative PCR kit (Bioneer, Daejon, Korea). The MagNA Pure 96 system and QIAamp system did not detect viral loads in one sample. The Chemagic system did not detect low viral loads in nine samples (range, IU/mL by the CAP/CTM platform). Comparisons of the viral loads of the samples from the MagNA Pure 96 system, the Chemagic system, and the QIAamp system with those from the CAP/CTM platform yielded correlation coefficients of 0.977, 0.914, and 0.967, respectively. Comparisons of the MagNA Pure 96 system and the Chemagic system with the QIAamp system yielded correlation coefficients of and 0.939, respectively. The MagNA Pure 96 system demonstrated better performance than the Chemagic system for HBV nucleic acid extraction. The MagNA Pure 96 system demonstrated comparable performance with the QIAamp system. Key words: Nucleic acid, Extraction, Hepatitis B virus, Quantification Introduction Because of their high sensitivity, a large number of polymerase chain reaction (PCR) tests are performed daily in clinical laboratories. Advances in these techniques, such as real-time PCR, have significantly improved the hands-on time and decreased contamination rates. Real-time PCR has enabled quantification of viral loads and fusion transcripts of various leukemia. Efficient nucleic acid extraction protocols are required to benefit from the advantages of nucleic acid amplification. Various techniques to extract nucleic acids have been introduced. Conventional manual extraction Address correspondence to Seong-Ho Kang, M.D., Ph.D., Department of Laboratory Medicine, Chosun University Medical School, 365 Pilmun-daero, Dong-gu, Gwangju, , South Korea; tel : ; fax : ; seonghomed@hanmail. net of nucleic acids is labor-intensive and susceptible to handling variations. Automated nucleic acid extraction is currently being widely performed and has become an essential part of molecular diagnostics. Hepatitis B virus (HBV) infection is a major cause of chronic liver disease, liver cirrhosis, and hepatocellular carcinoma. It is estimated that nearly 2 billion people worldwide are infected with HBV and that more than 350 million have a persistent and chronic infection [1]. Recent advances in antiviral therapy for chronic HBV infection, using new or more powerful nucleoside/nucleotide analogues, have led to better management of chronic HBV infection [2]. The introduction of highly sensitive tests for monitoring HBV DNA has supported the success of antiviral therapy. Quantification of HBV /12/ by the Association of Clinical Scientists, Inc.

2 Available online at Comparison of MagNA Pure 96, Chemagic MSM 1, and QIAmp MinElute 371 Table 1. Real-time quantitative PCR from extracted DNA by three systems (Number of samples) Real-time quantitative PCR MagNApure 96 Chemagic QIAamp Detected Not detected 1 a 9 b 1 c a 43 IU/mL by the CAP/CTM platform b Median value of 56 IU/mL (range, IU/mL) by the CAP/CTM platform c 26 IU/mL by the CAP/CTM platform viral load is one of the most frequently performed tests in molecular diagnostics, especially in Asia. This study was designed to compare HBV nucleic acid extraction in two automated systems and one manual system. One of the two automated systems was the MagNA Pure 96 DNA and Viral NA small volume kit (Roche Applied Science, Manheim, Germany) with MagNA pure 96 (Roche Applied Science). The second automated system was the Chemagic Viral DNA/RNA kit (Chemagen, Baesweiler, Germany) with Chemagic MSM1 (Chemagen). The manual system was the QIAamp MinElute Virus spin kit (Qiagen, Hilden, Germany). Both the MagNA Pure 96 system and the Chemagic system use magnetic bead technology for nucleic acid extraction. The QIAamp system uses a silica membrane-based method; extraction using this system is time-consuming and requires experienced personnel. The method relying on silica columns is considered the gold standard for nucleic acid extraction [3], so it served as the reference method in the present study. Materials and Methods Samples. Samples that were requested for HBV quantification were used. We first quantified HBV loads using the Cobas Ampliprep/Cobas TaqMan (CAP/CTM) platform (Roche Molecular Systems, Manheim, Germany). Sixty-eight samples that were within the detection range of CAP/CTM ( IU/mL) were selected. DNA extraction. Viral nucleic acids of these serum samples were extracted using the automated MagNA Pure 96 system, the Chemagic system, and the manual QIAamp system. For viral DNA extraction using the MagNA Pure 96 system, 200 µl samples were mixed with 300 µl lysis binding buffer. Magnetic glass particles were added to a mixture of samples. After the incubation, the nucleic acids bound to the beads were separated with a magnetic separator. After washing, the pure nucleic acids were eluted in a volume of 50 μl. For viral DNA extraction using the Chemagic system, samples were incubated with protease and lysis buffer with an input volume of 200 µl, and magnetic beads were added to the mixture. After the incubation, nucleic acids bound to the beads were separated with a magnetic separator. After washing, the nucleic acids were eluted in a volume of 50 μl. For viral DNA extraction using the QIAamp system, the manufacturer s protocol was used, input volume was 200 µl, and viral DNA was eluted in a volume of 50 μl. Real-time quantitative PCR. Extracted viral nucleic acids were quantified using an AccuPower HBV Quantitative PCR kit (Bioneer, Daejon, Korea) and an Exicycler 96 real-time Quantitative Thermal Block (Bioneer). The AccuPower HBV Quantitative PCR kit is a Conformité Européene certified product. Fifty microliters of each DNA

3 372 sample was mixed with the TaqMan probe and primers specific to the surface antigen S gene. PCR conditions were initial denaturation at 95 C for 10 min followed by 45 cycles of denaturation at 95 C for 20 sec, and extension at 55 C for 30 sec. The lower limit of detection was 22 IU/mL. The quantified viral loads from the three systems were compared with those from the CAP/CTM platform, and the loads from two automated systems (MagNA Pure 96 system and Chemagic system) were compared with those from the manual system (QIAamp system). In order to evaluate reproducibility, two positive control DNA samples with different concentrations ( copies/reaction and copies/reaction) were extracted twenty times using the three extraction methods described above and were subsequently quantified. HBV loads were log10 transformed for analysis. SPSS version 12.0 (SPSS, Chicago, IL, USA) was used for the statistical analysis. Results Real-time PCR from DNA extracted by the MagNA Pure 96 system detected HBV in 67 of the 68 samples but did not detect a low viral load in one sample (Table 1). The HBV load of the latter sample was 43 IU/mL using the CAP/CTM platform. Real-time PCR using DNA extracted by the Chemagic system detected HBV in 59 of 68 samples but did not detect the low viral load in nine samples. The median HBV load of samples that was not detected was 56 (range, IU/mL) in the CAP/CTM platform. Real-time PCR using DNA extracted by the QIAamp system detected HBV in 67 of 68 samples but did not detect HBV in one sample. The HBV value of the latter sample was 26 IU/mL in the CAP/CTM platform.

4 Comparison of MagNA Pure 96, Chemagic MSM 1, and QIAmp MinElute 373 Figure 2. Linear regression analysis of HBV load (log10 IU/mL). (A) MagNA Pure 96 sysem vs. QIAamp system (B) Chemagic system vs. QIAamp system. Comparisons of the viral loads of samples from the MagNA Pure 96 system, the Chemagic system, and the QIAamp system with those from the CAP/ CTM platform yielded correlation coefficients of 0.977, 0.914, and 0.967, respectively (Figure 1). The slopes for the MagNA Pure 96 system, the Chemagic system, and the QIAamp system compared with CAP/CTM platform were 0.953, 0.906, and 0.940, respectively. Comparisons of the MagNA Pure 96 system and the Chemagic system with the QIAamp system yielded correlation coefficients of and 0.939, respectively (Figure 2). The slopes of the MagNA Pure 96 system and the Chemagic system compared with the QIAamp system were and 0.964, respectively. In reproducibility testing using high positive control DNA, the means of the MagNA Pure 96 system, the Chemagic system, and the QIAamp system were 6.62 log 10 IU/mL, 6.65 log 10 IU/mL and 6.69 log 10 IU/mL, respectively. The standard deviations of the three systems were 5.51 log 10 IU/mL, 5.55 log 10 IU/mL, and 5.50 log 10 IU/mL, respectively. The coefficients of variations were 7.76%, 7.94%, and 8.08%, respectively. In reproducibility testing using low positive control DNA, the means of the MagNA Pure 96 system, the Chemagic system, and the QIAamp system were 2.80 log 10 IU/ ml, 2.74 log 10 IU/mL and 2.74 log 10 IU/mL, respectively. The standard deviations of the three systems were 1.90 log 10 IU/mL, 1.85 log 10 IU/mL, and 1.85 log 10 IU/mL, respectively. The coefficients of variations were 12.58%, 12.87%, and 12.89%, respectively. These results were not shown to be statistically different by analysis of variance (ANOVA). Discussion The present study compared the performance of three systems for HBV nucleic acid extraction. Some studies have evaluated Nuclisens easy MAG (biomerieux, Marcy l Etoile, France), QIAsymphony SP (Qiagen, Valencia, CA, USA), COBAS Ampliprep (Roche Molecular Systems), and MagNA Pure (Roche Applied Science) for nucleic acid extracts of cytomegalovirus and reported comparable performance with the manual system [3-6]. However, only one report has evaluated the performance of automated extraction systems for HBV quantification [7]. This could be due to the fact that large companies have developed HBV quantification systems for extraction and detection. HBV is a DNA virus, and quantification of HBV is the most frequent quantification test in molecular diagnostics in some countries. The extraction performance of HBV could be a barometer of viral DNA extraction in many extraction systems. In Korea, there are real-time quantification HBV kits that are commercially available and applicable to different real-time PCR machines. Thus,

5 374 it is important to select an efficient and sensitive extraction system when using these kits. The present study used the CAP/CTM platform as a reference method for HBV quantification. Previous studies have shown that the CAT/CTM HBV assay is sensitive, specific, and reproducible, and that it accurately quantifies HBV DNA levels [8, 9]. In the present study, the DNA viral loads detected by the MagNA Pure 96 system and the QIAamp system significantly correlated with those of the CAP/CTM platform, and the DNA viral loads of the Chemagic system also significantly correlated with those of the CAP/CTM platform. However, the correlation coefficient of the Chemagic system was lower than that of the MagNA Pure 96 system and that of the QIAamp system, and the slopes of the MagNA Pure 96 system and the QIAamp system were closer to 1 than was the slope of the Chemagic system. Compared with the QIAamp system, the reference method for viral nucleic acid extraction, the MagNA Pure 96 system correlated with a higher correlation coefficient than that of the Chemagic system. A previous report compared performance of the Chemagic system with the QIAamp system by testing qualitative rather than quantitative HBV PCR, and concluded that the Chemagic system is not distinguishable from the QIAamp [10]. Both the MagNA Pure 96 system and the Chemagic system use magnetic bead technology, but in the present study the Chemagic system did not detect low viral loads in nine samples in which viral DNA was detected by the MagNA Pure 96 system. These collective observations indicate that the nucleid acid extraction efficiency of the MagNA Pure 96 system is better than that of the Chemagic system. Both the MagNA Pure 96 system and the QIAamp system did not detect viral load in one sample with a low viral load. The viral load of the sample that the MagNA Pure 96 system did not detect (43 IU/ ml by the CAP/CTM platform) was slightly higher than that of the sample that the QIAamp system did not detect (26 IU/mL by the CAP/CTM platform). These samples might not have been detected due to random error or were perhaps close to the detection limit of the HBV quantification kit used in the present study. In the present study, specificity testing of the three systems with HBV negative samples or hepatitis C virus positive samples was not performed. The present study is a limited comparison and further study is needed to evaluate the specificity performance of the three systems. In conclusion, the MagNA Pure 96 system demonstrated better performance than the Chemagic system for HBV nucleic acid extraction. The MagNA Pure 96 system demonstrated comparable performance with the QIAamp system. Acknowledgements This study was supported by research funds from Chosun University Hospital in References 1. Yim HJ and Lok AS. Natural history of chronic hepatitis B virus infection: what we knew in 1981 and what we know in Hepatology 2006;43:S Chien RN and Liaw YF. Nucleos(t)ide analogues for hepatitis B virus: strategies for long-term success. Best Pract Res Clin Gastroenterol 2008;22: Pillet S, Bourlet T, Pozzetto B. Comparative evaluation of a commercially available automated system for extraction of viral DNA from whole blood: application to monitoring of epsteinbarr virus and cytomegalovirus load. J Clin Microbiol 2009;47: Mengelle C, Mansuy JM, Da Silva I, Davrinche C, Izopet J. Comparison of 2 highly automated nucleic acid extraction systems for quantitation of human cytomegalovirus in whole blood. Diagn Microbiol Infect Dis 2011;69: Miller S, Seet H, Khan Y, Wright C, Nadarajah R. Comparison of QIAGEN automated nucleic acid extraction methods for CMV quantitative PCR testing. Am J Clin Pathol 2010;133: Perandin F, Pollara PC, Gargiulo F, Bonfanti C, Manca N. Performance evaluation of the automated NucliSens easymag nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens. Diagn Microbiol Infect Dis 2009;64: Caliendo AM, Valsamakis A, Bremer JW, Ferreira-Gonzalez A, Granger S, Sabatini L, Tsongalis GJ, Wang YF, Yen-Lieberman B, Young S, Lurain NS. Multilaboratory evaluation of real-time PCR tests for hepatitis B virus DNA quantification. J Clin Microbiol 2011;49: Allice T, Cerutti F, Pittaluga F, Varetto S, Gabella S, Marzano A, Franchello A, Colucci G, Ghisetti V. COBAS AmpliPrep- COBAS TaqMan hepatitis B virus (HBV) test: a novel automated real-time PCR assay for quantification of HBV DNA in plasma. J Clin Microbiol 2007;45: Chevaliez S, Bouvier-Alias M, Laperche S, Pawlotsky JM. Performance of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for hepatitis B virus DNA quantification. J Clin Microbiol 2008;46: Kleines M, Schellenberg K, Ritter K. Efficient extraction of viral DNA and viral RNA by the Chemagic viral DNA/RNA kit allows sensitive detection of cytomegalovirus, hepatitis B virus, and hepatitis G virus by PCR. J Clin Microbiol 2003;41:

Multilaboratory Evaluation of Real-Time PCR Tests for Hepatitis B Virus DNA Quantification

Multilaboratory Evaluation of Real-Time PCR Tests for Hepatitis B Virus DNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2011, p. 2854 2858 Vol. 49, No. 8 0095-1137/11/$12.00 doi:10.1128/jcm.00471-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Multilaboratory

More information

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.2 08/2017 EN RealStar RealStar CMV PCR Kit 1.2 For research use only! (RUO) 021202 INS-021200-EN-S01 48 08 2017 altona Diagnostics GmbH Mörkenstr.

More information

Comparison of Three Roche HBV Viral Load Assay Formats

Comparison of Three Roche HBV Viral Load Assay Formats JCM Accepts, published online ahead of print on 25 April 2012 J. Clin. Microbiol. doi:10.1128/jcm.00746-12 Copyright 2012, American Society for Microbiology. All Rights Reserved. 1 Title 2 3 4 5 6 7 8

More information

Original Article Diagnostic Immunology INTRODUCTION https://doi.org/ /alm

Original Article Diagnostic Immunology INTRODUCTION https://doi.org/ /alm Original Article Diagnostic Immunology Ann Lab Med 2017;37:248-253 https://doi.org/10.3343/alm.2017.37.3.248 ISSN 2234-3806 eissn 2234-3814 Comparison of the QIAGEN artus HBV QS-RGQ Assay With the Roche

More information

Human diagnostics. Better be Sure: Quantify HDV & HBV viral load. RoboGene product family

Human diagnostics. Better be Sure: Quantify HDV & HBV viral load. RoboGene product family Human diagnostics Better be Sure: Quantify HDV & HBV viral load. RoboGene product family 2 RoboGene Product Family Improved patient management: Standardized monitoring of HBV DNA and HDV RNA viral load.

More information

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets (TaqMan Real-time ' FAM ' BHQ1

More information

Received 19 April 2011/Returned for modification 6 June 2011/Accepted 13 June 2011

Received 19 April 2011/Returned for modification 6 June 2011/Accepted 13 June 2011 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2011, p. 2899 2904 Vol. 49, No. 8 0095-1137/11/$12.00 doi:10.1128/jcm.00785-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Comparative

More information

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets Real- (TaqMan time ' FAM ' BHQ1

More information

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 2010, p. 3641 3647 Vol. 48, No. 10 0095-1137/10/$12.00 doi:10.1128/jcm.01306-10 Copyright 2010, American Society for Microbiology. All Rights Reserved. Performance

More information

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0 JCM Accepts, published online ahead of print on 25 September 2013 J. Clin. Microbiol. doi:10.1128/jcm.02102-13 Copyright 2013, American Society for Microbiology. All Rights Reserved. 1 2 Hepatitis C Virus

More information

Comparison of a novel real-time PCR method (RTA) and Artus RG for the quantification of HBV DNA and HCV RNA

Comparison of a novel real-time PCR method (RTA) and Artus RG for the quantification of HBV DNA and HCV RNA Original Article Comparison of a novel real-time PCR method (RTA) and Artus RG for the quantification of HBV DNA and HCV RNA Sibel Aydogan 1, Ziya Cibali Acikgoz 2, Aysegül Gozalan 1, Fisun Kirca 1, Tuba

More information

Instructions for Use. RealStar Influenza Screen & Type RT-PCR Kit /2017 EN

Instructions for Use. RealStar Influenza Screen & Type RT-PCR Kit /2017 EN Instructions for Use RealStar Influenza Screen & Type RT-PCR Kit 4.0 05/2017 EN RealStar Influenza Screen & Type RT-PCR Kit 4.0 For research use only! (RUO) 164003 INS-164000-EN-S01 96 05 2017 altona

More information

INTRODUCTION SUMMARY. KEY WORDS: Quantification, HBV DNA, TaqMan, Automation, Antiviral therapy

INTRODUCTION SUMMARY. KEY WORDS: Quantification, HBV DNA, TaqMan, Automation, Antiviral therapy NEW MICROBIOLOGICA, 31, 27-35, 2008 Comparison of the Cobas Ampliprep /Cobas TaqMan HBV Test versus the Cobas Amplicor HBV monitor for HBV-DNA detection and quantification during antiviral therapy Tiziano

More information

Instructions for Use. RealStar Influenza S&T RT-PCR Kit /2017 EN

Instructions for Use. RealStar Influenza S&T RT-PCR Kit /2017 EN Instructions for Use RealStar Influenza S&T RT-PCR Kit 3.0 01/2017 EN RealStar Influenza S&T RT-PCR Kit 3.0 For research use only! (RUO) 163003 INS-163000-EN-S02 96 01 2017 altona Diagnostics GmbH Mörkenstr.

More information

A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA

A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA Original Article A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA Xiaoyu Fu 1,Deming Tan 1, Xiaoguang Dou 2,Jinjun Chen 3,Juan Wu 1 1

More information

Technical Bulletin No. 162

Technical Bulletin No. 162 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 162 cobas 6800 HCV Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA.

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA. 000000 00000 0000 000 00 0 bdna () 00000 0000 000 00 0 Nuclisens () 000 00 0 000000 00000 0000 000 00 0 Amplicor () Comparison of Amplicor HIV- monitor Test, NucliSens HIV- QT and bdna Versant HIV RNA

More information

PERFORMANCE OF THE COBAS AMPLIPREP/COBAS TAQMAN (CAP/CTM) REAL-TIME POLYMERASE CHAIN REACTION ASSAY ACCEPTED FOR HEPATITIS B VIRUS DNA QUANTIFICATION

PERFORMANCE OF THE COBAS AMPLIPREP/COBAS TAQMAN (CAP/CTM) REAL-TIME POLYMERASE CHAIN REACTION ASSAY ACCEPTED FOR HEPATITIS B VIRUS DNA QUANTIFICATION JCM Accepts, published online ahead of print on 20 February 2008 J. Clin. Microbiol. doi:10.1128/jcm.01248-07 Copyright 2008, American Society for Microbiology and/or the Listed Authors/Institutions. All

More information

Viral RNA / DNA purification products from MACHEREY-NAGEL. MN guide for viral RNA / DNA purification Multiple solutions for many needs

Viral RNA / DNA purification products from MACHEREY-NAGEL. MN guide for viral RNA / DNA purification Multiple solutions for many needs Viral RNA / DNA Purification Guide Viral RNA / DNA purification products from MACHEREY-NAGEL MN guide for viral RNA / DNA purification Multiple solutions for many needs Single or high throughput Human

More information

Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for. hepatitis B virus DNA quantification.

Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for. hepatitis B virus DNA quantification. Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for hepatitis B virus DNA quantification. Stéphane Chevaliez, Magali Bouvier-Alias, Syria Laperche, Christophe Hézode,

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Test, version 2.0 (TaqMan 48) Number: PQDx 0126-046-00 Abstract The COBAS AmpliPrep/COBAS TaqMan

More information

Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets

Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets Liu et al. Virology Journal (2017) 14:94 DOI 10.1186/s12985-017-0759-8 RESEARCH Open Access Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets Chao Liu 1,2,3,

More information

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2007, p. 2529 2536 Vol. 45, No. 8 0095-1137/07/$08.00 0 doi:10.1128/jcm.00058-07 Copyright 2007, American Society for Microbiology. All Rights Reserved. Comparison

More information

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up JCM Accepts, published online ahead of print on 11 May 2011 J. Clin. Microbiol. doi:10.1128/jcm.00721-11 Copyright 2011, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights

More information

Antiviral Therapy 2014; 19: (doi: /IMP2723)

Antiviral Therapy 2014; 19: (doi: /IMP2723) Antiviral Therapy 214; 19:449 454 (doi: 1.3851/IMP2723) Original article Performance characteristics of the COBAS Ampliprep/COBAS TaqMan v2. and the Abbott RealTime hepatitis C assays implications for

More information

Results of the College of American Pathologists (CAP) Proficiency Program for CMV, EBV, and BKV Viral Load Testing

Results of the College of American Pathologists (CAP) Proficiency Program for CMV, EBV, and BKV Viral Load Testing 2009 College of American Pathologists. The College does not permit reproduction of any substantial portion of the material in this Report without its written authorization. The College hereby authorizes

More information

HIV-1 Viral Load Real Time (RG)

HIV-1 Viral Load Real Time (RG) -1 Viral Load Real Time (RG) Real Time RT-PCR type 1 RNA quantification assay MSP Reg. pending Valdense 3616. 11700. Montevideo. Uruguay. phone (598) 2 336 83 01. Fax (598) 2 336 71 60. Info@atgen.com.uy

More information

For purification of viral DNA and RNA from a wide range of sample materials

For purification of viral DNA and RNA from a wide range of sample materials QIAamp virus kits For purification of viral DNA and RNA from a wide range of sample materials Automatable on QIAGEN s proven QIAamp Kits set the standard for purification of viral DNA and RNA. QIAamp virus

More information

AmpliX HBV Quantitative

AmpliX HBV Quantitative Instructions for use REAL TIME PCR DETECTION AND QUANTITATION KIT OF HEPATITIS B VIRUS DNA Research Use Only (RUO) (Lyo-format) VBD0595 96 rcs valid from May 2013 Explanation of symbols used in labeling

More information

Simple Solutions for Patient Monitoring. Maximum Flexibility 2 Configuration: 8 to 1000 viral load tests/day 2 Sample: Plasma/DBS

Simple Solutions for Patient Monitoring. Maximum Flexibility 2 Configuration: 8 to 1000 viral load tests/day 2 Sample: Plasma/DBS Simple Solutions for Patient Monitoring What is the interest of having a high level of sensitivity? Higher sensitivity means you are able to offer better patient monitoring, particularly in the low viral

More information

Virological Tools and Monitoring in the DAA Era

Virological Tools and Monitoring in the DAA Era Virological Tools and Monitoring in the DAA Era Prof. Jean-Michel Pawlotsky, MD, PhD National Reference Center for Viral Hepatitis B, C and delta Department of Virology & INSERM U955 Henri Mondor Hospital

More information

RealLine HIV quantitative Str-Format

RealLine HIV quantitative Str-Format Instructions for use DETECTION AND QUANTIFICATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Research Use Only (RUO) Attention! Please read the information about quantification process carefully!

More information

Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood

Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood Brief Communication Clinical Microbiology Ann Lab Med 2017;37:147-150 https://doi.org/10.3343/alm.2017.37.2.147 ISSN 2234-3806 eissn 2234-3814 Performance of the Real-Q EBV Quantification Kit for Epstein-Barr

More information

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring Monitoring of HCV RNA Hepatitis C Requirements of Antiviral Therapy Monitoring Hepatitis C Requirements of Antiviral Therapy Monitoring Author: Ute Hofmann,, Konrad-Zuse-Str. 1, 07745 Jena, Germany Beatrix

More information

4th International HIV/Viral Hepatitis Co-Infection Meeting

4th International HIV/Viral Hepatitis Co-Infection Meeting 4th International HIV/Viral Hepatitis Co-Infection Meeting The Rocky Road to Viral Hepatitis Elimination: Assuring access to antiviral therapy for ALL co-infected patients from low to high income settings

More information

Molecular Diagnosis Future Directions

Molecular Diagnosis Future Directions Molecular Diagnosis Future Directions Philip Cunningham NSW State Reference Laboratory for HIV/AIDS & Molecular Diagnostic Medicine Laboratory, SydPath St Vincent s Hospital Sydney Update on Molecular

More information

ExiPrep TM Dx Fully Automated Nucleic Acid Extraction

ExiPrep TM Dx Fully Automated Nucleic Acid Extraction 20150629 Ver.1(EN) ExiPrep TM Dx Fully Automated Nucleic Acid Extraction 8-11 Munpyeongseo-ro, Daedeok-gu, Daejeon, 34302, Republic of Korea Tel: +82-42-930-8777 Fax: +82-42-930-8688 E-mail: sales@bioneer.com

More information

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO)

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Diagnostic Methods of HBV infection Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Hepatitis B-laboratory diagnosis Detection of HBV infection involves

More information

HBV PUBLIC HEALTH IMPLICATIONS

HBV PUBLIC HEALTH IMPLICATIONS جزايری دکتر سيد محمد آزمايشگاه ھپاتيت B -دانشکده بھداشت ويروس شناسی- گروه دانشگاه علوم پزشکی تھران کنگره ارتقا کيفيت- ١٣٩٢ HBV PUBLIC HEALTH IMPLICATIONS 2 billion people have been infected by HBV worldwide.

More information

Technical Bulletin No. 161

Technical Bulletin No. 161 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 161 cobas 6800 HIV-1 Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

artus EBV QS-RGQ Kit Performance Characteristics May 2012 Sample & Assay Technologies Analytical sensitivity plasma

artus EBV QS-RGQ Kit Performance Characteristics May 2012 Sample & Assay Technologies Analytical sensitivity plasma artus EBV QS-RGQ Kit Performance Characteristics artus EBV QS-RGQ Kit, Version 1, 4501363 Check availability of new electronic labeling revisions at www.qiagen.com/products/artusebvpcrkitce.aspx before

More information

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012 altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 11/2012 altona Diagnostics GmbH Moerkenstr. 12 22767 Hamburg Germany phone +49 40 548 06 76-0 fax +49 40 548 06 76-10 e-mail info@altona-diagnostics.com

More information

on October 4, 2018 by guest

on October 4, 2018 by guest JCM Accepts, published online ahead of print on 3 July 2012 J. Clin. Microbiol. doi:10.1128/jcm.01249-12 Copyright 2012, American Society for Microbiology. All Rights Reserved. 1 Title: Performance of

More information

QUANTITATION OF HEPATITIS B VIRUS DNA IN PLASMA USING A SENSITIVE COST-EFFECTIVE IN-HOUSE REAL-TIME PCR ASSAY

QUANTITATION OF HEPATITIS B VIRUS DNA IN PLASMA USING A SENSITIVE COST-EFFECTIVE IN-HOUSE REAL-TIME PCR ASSAY Indian Journal of Medical Microbiology, (2009) 27(2): 111-5 Original Article QUANTITATION OF HEPATITIS B VIRUS DNA IN PLASMA USING A SENSITIVE COST-EFFECTIVE IN-HOUSE REAL-TIME PCR ASSAY Hubert Darius

More information

Purification of viral nucleic acid from serum, plasma, cell-free biological fluids MACHEREY- NAGEL

Purification of viral nucleic acid from serum, plasma, cell-free biological fluids MACHEREY- NAGEL Purification of viral nucleic acid from serum, plasma, cell-free biological fluids Purification of viral nucleic acid from serum, plasma, cell-free biological fluids viral RNA: viral DNA: NucleoSpin RNA

More information

Use of Viral Load Testing in Managing CMV Infections in SOTR

Use of Viral Load Testing in Managing CMV Infections in SOTR Use of Viral Load Testing in Managing CMV Infections in SOTR Angela M. Caliendo, MD, PhD, FIDSA Professor and Vice Chair, Medicine Alpert Medical School of Brown University Providence, RI Disclosures Scientific

More information

Hepatitis C Virus (RNA)A

Hepatitis C Virus (RNA)A Hepatitis C Virus (RNA)A 2012 EQA Programme Final Report QAV994112 (HCVRNA12A) Professor Jacques Izopet Scientific Expert on behalf of QCMD Report authorised by the QCMD Executive in July 2012 A UKAS accredited

More information

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26 1 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Viral Load Quality Assessment Program Summary

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Qualitative Test, version 2.0 (TaqMan 48) Number: PQDx 0221-046-00 Abstract COBAS AmpliPrep/COBAS

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Qualitative Test, version 2.0 (TaqMan 96) Number: PQDx 0200-046-00 Abstract COBAS AmpliPrep/COBAS

More information

Diagnostic Methods of HBV and HDV infections

Diagnostic Methods of HBV and HDV infections Diagnostic Methods of HBV and HDV infections Zohreh Sharifi,ph.D Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine Hepatitis B-laboratory diagnosis Detection

More information

Received 12 July 2007/Returned for modification 19 November 2007/Accepted 4 February 2008

Received 12 July 2007/Returned for modification 19 November 2007/Accepted 4 February 2008 JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 2008, p. 1241 1245 Vol. 46, No. 4 0095-1137/08/$08.00 0 doi:10.1128/jcm.01403-07 Copyright 2008, American Society for Microbiology. All Rights Reserved. Detection

More information

Hepatitis B Virus Genemer

Hepatitis B Virus Genemer Product Manual Hepatitis B Virus Genemer Primer Pair for amplification of HBV Viral Specific Fragment Catalog No.: 60-2007-10 Store at 20 o C For research use only. Not for use in diagnostic procedures

More information

RealLine HBV / HCV / HIV Str-Format

RealLine HBV / HCV / HIV Str-Format Instructions for use ASSAY KIT FOR THE QUALITATIVE AND DIFFERENTIAL DETECTION OF HEPATITIS B VIRUS DNA, HEPATITIS C VIRUS RNA, AND HUMAN IMMUNODEFICIENCY VIRUS TYPES 1 AND 2 RNA USING THE PCR/RT-PCR METHOD

More information

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection. For in vitro Veterinary Diagnostics only. Kylt Rotavirus A Real-Time RT-PCR Detection www.kylt.eu DIRECTION FOR USE Kylt Rotavirus A Real-Time RT-PCR Detection A. General Kylt Rotavirus A products are

More information

Screening and Diagnosis of Hepatitis Virus Infections

Screening and Diagnosis of Hepatitis Virus Infections Screening and Diagnosis of Hepatitis Virus Infections Prof. Jean-Michel Pawlotsky, MD, PhD National Reference Center for Viral Hepatitis B, C and delta Department of Virology & INSERM U955 Henri Mondor

More information

Trends in molecular diagnostics

Trends in molecular diagnostics Trends in molecular diagnostics Detection of target genes of interest Quantification Infectious diseases HIV Hepatitis C & B TB / MAC Cytomegalovirus Herpes simplex Varicella zoster CT/GC HPV Profiling

More information

RealLine HIV qualitative Str-Format

RealLine HIV qualitative Str-Format Instructions for Use REAL TIME PCR DETECTION KIT FOR HUMAN IMMUNODEFICIENCY VIRUS RNA Research Use Only (RUO) RealLine HIV Qualitative (Str-format) VBD0196 96 Tests valid from July 2016 Rev01072016_EN

More information

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012 1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral

More information

RealLine HCV Qualitative Str-Format

RealLine HCV Qualitative Str-Format Instructions for use REAL TIME PCR DETECTION KIT FOR THE HEPATITIS C VIRUS RNA (HCV) Research Use Only (RUO) (Str-format) VBD0795 96 Tests valid from: October 2018 Rev05_1018_EN Page 1 of 8 Explanation

More information

CMV Diagnostic Strategies: Current and Future

CMV Diagnostic Strategies: Current and Future CMV Diagnostic Strategies: Current and Future Tony Mazzulli, MD, FRCPC, FACP Microbiologist-in-Chief Mount Sinai Hospital & University Health Network, Toronto Faculty/Presenter Disclosure Relationships

More information

COMPARISON OF HBV RIBONUCLEASE H DOMAIN IN NAÏVE AND DRUG RESISTANT PATIENTS

COMPARISON OF HBV RIBONUCLEASE H DOMAIN IN NAÏVE AND DRUG RESISTANT PATIENTS HBV RIBONUCLEASE H DOMAIN IN PATIENTS WITH DRUG RESISTANT COMPARISON OF HBV RIBONUCLEASE H DOMAIN IN NAÏVE AND DRUG RESISTANT PATIENTS Surachai Amornsawadwattana, Pattaratida Sa-Nguanmoo, Preeyaporn Vichaiwattana,

More information

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2017Oct27

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2017Oct27 1 The National Laboratory for HIV Reference Services is Accredited to ISO 15189 and ISO 17043 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology

More information

The quantification of hepatitis C virus (HCV)

The quantification of hepatitis C virus (HCV) Overestimation and Underestimation of Hepatitis C Virus RNA Levels in a Widely Used Real-Time Polymerase Chain Reaction Based Method Stéphane Chevaliez, 1,2 Magali Bouvier-Alias, 1,2 Rozenn Brillet, 1,2

More information

Bristol-Myers Squibb

Bristol-Myers Squibb A Study of the Safety and Efficacy of plus Tenofovir in Adults with Chronic Hepatitis B Virus Infection with Previous Nucleoside/Nucleotide Treatment Failure () FINAL CLINICAL STUDY REPORT EUDRACT Number:

More information

Evaluation of New Rapid Antigen Test for the Detection of Pandemic. Influenza A/H1N Virus

Evaluation of New Rapid Antigen Test for the Detection of Pandemic. Influenza A/H1N Virus JCM Accepts, published online ahead of print on 31 March 2010 J. Clin. Microbiol. doi:10.1128/jcm.02392-09 Copyright 2010, American Society for Microbiology and/or the Listed Authors/Institutions. All

More information

Roche Molecular Biochemicals Application Note No. HP 1/1999

Roche Molecular Biochemicals Application Note No. HP 1/1999 Roche Molecular Biochemicals Application Note No. HP 1/1999 Nucleic Acid Purification High Pure Viral Nucleic Acid Kit High Pure 16 System Viral Nucleic Acid Kit Efficiency of Hepatitis C Virus sample

More information

Introduction: Table/Figure Descriptions:

Introduction: Table/Figure Descriptions: Introduction: We have completed the analysis of your HIV RNA Validation Study. The validation plan was designed to verify the installation of an unmodified FDA-approved HIV RNA assay into your laboratory.

More information

Reassessment necessity of patients with serogroup 2 of hepatitis C virus and undetectable serum

Reassessment necessity of patients with serogroup 2 of hepatitis C virus and undetectable serum 1 Reassessment necessity of patients with serogroup 2 of hepatitis C virus and undetectable serum HCV RNA Yuki Moritou 1, Fusao Ikeda 1,2,#, 1 Yasuto Takeuchi 1, 1 Hiroyuki Seki 1, 1 Shintaro Nanba 1,

More information

Evaluation of the new Abbott mplus feature: Impact on clinical laboratory efficiencies with Abbott RealTime HIV-1, HCV, HBV and CT/NG assays

Evaluation of the new Abbott mplus feature: Impact on clinical laboratory efficiencies with Abbott RealTime HIV-1, HCV, HBV and CT/NG assays JCM Accepts, published online ahead of print on 2 October 2013 J. Clin. Microbiol. doi:10.1128/jcm.01672-13 Copyright 2013, American Society for Microbiology. All Rights Reserved. 1 2 3 4 5 Evaluation

More information

EJOHG ORIGINAL ARTICLE ABSTRACT INTRODUCTION

EJOHG ORIGINAL ARTICLE ABSTRACT INTRODUCTION Evaluation of a Rapid One-step Real-time PCR Method as a High-throughput 10.5005/jp-journals-10018-1121 Screening for Quantification ORIGINAL ARTICLE Evaluation of a Rapid One-step Real-time PCR Method

More information

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: Alere q HIV-1/2 Detect WHO reference number: PQDx 0226-032-00 Alere q HIV-1/2 Detect with product codes 270110050, 270110010 and 270300001,

More information

Khalid Alquthami (Correspondence) Regional Lab, Makkah. Saudi Arabia

Khalid Alquthami (Correspondence) Regional Lab, Makkah. Saudi Arabia 1 Regional Lab, Makkah. Saudi Arabia stract: Clinical specificity and genotype/subtype detection of viruses using the Cobas TaqScreen MPX system V 2.0, which is a nucleic acid test (NAT) that uses multiples

More information

Evaluation of NucliSens easymag for Automated Nucleic Acid Extraction from Various Clinical Specimens

Evaluation of NucliSens easymag for Automated Nucleic Acid Extraction from Various Clinical Specimens JOURNAL OF CLINICAL MICROBIOLOGY, Feb. 2007, p. 421 425 Vol. 45, No. 2 0095-1137/07/$08.00 0 doi:10.1128/jcm.00894-06 Copyright 2007, American Society for Microbiology. All Rights Reserved. Evaluation

More information

B19 Virus EQA Programme Final Report QAV (B19DNA14)

B19 Virus EQA Programme Final Report QAV (B19DNA14) B19 Virus 2014 EQA Programme Final Report QAV034116 (B19DNA14) Prof. Hubert GM Niesters Scientific Expert on behalf of QCMD Report authorised by the QCMD Executive in July 2014 A UKAS accredited proficiency

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK West of Scotland Specialist Virology Centre New Lister Building Level 5 Glasgow Royal Infirmary 10-16 Alexandra Parade Glasgow G31 2ER Contact:

More information

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Clinical Chemistry / EVALUATION OF QUANTITATIVE HEPATITIS C VIRUS ASSAY Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Maria Erali, MS, 1 Edward R. Ashwood,

More information

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.0 08/2017 EN RealStar RealStar CMV PCR Kit 1.0 For use with Mx 3005P QPCR System (Stratagene) VERSANT kpcr Molecular System AD (Siemens Healthcare)

More information

Final Clinical Study Report. to the Dossier SYNOPSIS. Final Clinical Study Report for Study AI463110

Final Clinical Study Report. to the Dossier SYNOPSIS. Final Clinical Study Report for Study AI463110 BMS-475 AI463 Name of Sponsor/Company: Bristol-Myers Squibb Individual Study Table Referring to the Dossier For National Authority Use Only) Name of Finished Product: Baraclude Name of Active Ingredient:

More information

Supplementary Figure 1. mrna targets were found in exosomes and absent in free-floating supernatant. Serum exosomes and exosome-free supernatant were

Supplementary Figure 1. mrna targets were found in exosomes and absent in free-floating supernatant. Serum exosomes and exosome-free supernatant were Supplementary Figure 1. mrna targets were found in exosomes and absent in free-floating supernatant. Serum exosomes and exosome-free supernatant were separated via ultracentrifugation and lysed to analyze

More information

RealLine HCV quantitative Str-Format

RealLine HCV quantitative Str-Format Instructions for use QUANTITATIVE KIT FOR REAL TIME PCR DETECTION FOR RNA OF HEPATITIS C VIRUS Attention! Please read the information about quantification process carefully! Research Use Only (RUO) (Str

More information

DETECTION AND QUANTITATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR

DETECTION AND QUANTITATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Lyo-Format Instructions for use Lyo-Format DETECTION AND QUANTITATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR (Lyo-format) 5645 96 rcs valid from May 2013 Distribué par / Distributed by:

More information

Received 30 July 2011/Returned for modification 1 September 2011/Accepted 14 September 2011

Received 30 July 2011/Returned for modification 1 September 2011/Accepted 14 September 2011 JOURNAL OF CLINICAL MICROBIOLOGY, Dec. 2011, p. 4089 4093 Vol. 49, No. 12 0095-1137/11/$12.00 doi:10.1128/jcm.05292-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Comparison

More information

ORIGINAL ARTICLE. Asian Pacific Journal of Allergy and Immunology. Abstract

ORIGINAL ARTICLE. Asian Pacific Journal of Allergy and Immunology. Abstract Asian Pacific Journal of Allergy and Immunology ORIGINAL ARTICLE Performance Evaluation of the Bioneer AccuPower HIV-1 Quantitative RT-PCR kit: Comparison with the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx 0115-041-00 Abstract The VERSANT HIV-1 RNA 1.0 Assay (kpcr) with product codes 10375763,

More information

Leukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit

Leukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit Revision No.: ZJ0003 Issue Date: Aug 7 th, 2008 Leukemia BCR-ABL Fusion Gene Real Time RT-PCR Kit Cat. No.: TR-0126-02 For use with ABI Prism 7000/7300/7500/7900(96 well); Smart Cycler II; icycler iq 4/iQ

More information

Application of μmacs Streptavidin MicroBeads for the analysis of HIV-1 directly from patient plasma

Application of μmacs Streptavidin MicroBeads for the analysis of HIV-1 directly from patient plasma Excerpt from MACS&more Vol 8 1/2004 Application of μmacs Streptavidin MicroBeads for the analysis of HIV-1 directly from patient plasma L. Davis Lupo and Salvatore T. Butera HIV and Retrovirology Branch,

More information

Product # Kit Components

Product # Kit Components 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Pneumocystis jirovecii PCR Kit Product # 42820 Product Insert Background Information

More information

Blocking the expression of the hepatitis B virus S gene in hepatocellular carcinoma cell lines with an anti-gene locked nucleic acid in vitro

Blocking the expression of the hepatitis B virus S gene in hepatocellular carcinoma cell lines with an anti-gene locked nucleic acid in vitro Blocking the expression of the hepatitis B virus S gene in hepatocellular carcinoma cell lines with an anti-gene locked nucleic acid in vitro Y.-B. Deng, H.-J. Qin, Y.-H. Luo, Z.-R. Liang and J.-J. Zou

More information

About HPV. Human papillomavirus (HPV) is a group of viruses that are extremely common worldwide. Theree are more than 100 types of HPV, of which at

About HPV. Human papillomavirus (HPV) is a group of viruses that are extremely common worldwide. Theree are more than 100 types of HPV, of which at Cervical Cancer & HPV What is HPV Human papillomavirus (HPV) is a group of viruses that are extremely common worldwide. About HPV Theree are more than 100 types of HPV, of which at least 13 are cancer-causing

More information

QIAsymphony DSP Circulating DNA Kit

QIAsymphony DSP Circulating DNA Kit QIAsymphony DSP Circulating DNA Kit February 2017 Performance Characteristics 937556 Sample to Insight Contents Performance Characteristics... 4 Basic performance... 4 Run precision... 6 Equivalent performance

More information

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ www.micropathology.com info@micropathology.com Micropathology Ltd Tel 24hrs: +44 (0) 24-76 323222 Fax / Ans: +44 (0) 24-76 - 323333 University of Warwick Science Park, Venture Centre, Sir William Lyons

More information

Laboratory and Clinical Diagnosis of HCV Infection

Laboratory and Clinical Diagnosis of HCV Infection Laboratory and Clinical Diagnosis of HCV Infection Jean-Michel Pawlotsky,, MD, PhD Department of Virology (EA 3489) Henri Mondor Hospital University of Paris XII Créteil,, France I Nonspecific Liver Tests

More information

Circulating Cell-Free DNA Pre-analytics: Importance of ccfdna Stabilization and Extraction for Liquid Biopsy Applications

Circulating Cell-Free DNA Pre-analytics: Importance of ccfdna Stabilization and Extraction for Liquid Biopsy Applications Circulating Cell-Free DNA Pre-analytics: Importance of ccfdna Stabilization and Extraction for Liquid Biopsy Applications Martin Schlumpberger, Ass.Director R&D, QIAGEN GmbH IQNpath 2017 - Circulating

More information

Kit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive

Kit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cryptococcus neoformans End-Point PCR Kit Product# EP42720 Product

More information

Manual innuprep DNA/RNA Virus PLUS Kit - KFFLX

Manual innuprep DNA/RNA Virus PLUS Kit - KFFLX Manual Order No.: 845-.)- reactions 845-.)- reactions Publication No.: HB_KF-5196_e_150414 This documentation describes the state at the time of publishing. It needs not necessarily agree with future versions.

More information

Practical Aspects of Standardisation for a Global Controls Manufacturer

Practical Aspects of Standardisation for a Global Controls Manufacturer Practical Aspects of Standardisation for a Global Controls Manufacturer MRSA SoGAT Clinical Diagnostic Meeting NIBSC, UK June 25, 2008 Frank Opdam PhD, AcroMetrix Standardisation A fundamental goal of

More information

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.2 08/2017 EN RealStar RealStar CMV PCR Kit 1.2 For use with SmartCycler II (Cepheid) LightCycler 1.2/1.5/2.0 Instruments (Roche) 021212 MAN-021210-EN-S02

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Royal Liverpool & Broadgreen University Hospitals NHS Trust Department of Virology Liverpool Clinical Laboratories Royal Liverpool and Broadgreen

More information

NEW EXTRACTION PROTOCOL NucliSENS easymag : «Whole Blood Viral Extraction Protocol» COMPOSITION:

NEW EXTRACTION PROTOCOL NucliSENS easymag : «Whole Blood Viral Extraction Protocol» COMPOSITION: CMV HHV6,7,8 R-gene REF 69-100 REF 69-100B NEW EXTRACTION PROTOCOL NucliSENS easymag : «Whole Blood Viral Extraction Protocol» HHV6 Premix r-gene REF 69-100R6 HHV7 Premix r-gene REF 69-100R7 HHV8 Premix

More information