Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Test and Identification of Rare Polymorphisms Potentially Affecting Assay Performance

Size: px
Start display at page:

Download "Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Test and Identification of Rare Polymorphisms Potentially Affecting Assay Performance"

Transcription

1 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2010, p Vol. 48, /10/$12.00 doi: /jcm Copyright 2010, American Society for Microbiology. All Rights Reserved. Evaluation of the Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Test and Identification of Rare Polymorphisms Potentially Affecting Assay Performance Michael T. Pyne, 1 * Katherine L. Brown, 2 and David R. Hillyard 1,3 ARUP Institute for Clinical and Experimental Pathology, 500 Chipeta Way, Salt Lake City, Utah ; Roche Molecular Systems, Inc., 4300 Hacienda Drive, Pleasanton, California ; and University of Utah Department of Pathology, 15 N. Medical Drive East, Salt Lake City, Utah Received 16 April 2010/Returned for modification 10 June 2010/Accepted 17 June 2010 We evaluated the FDA-approved Roche Cobas AmpliPrep/Cobas TaqMan (CAP/CTM) HIV-1 viral load assay for sensitivity, reproducibility, linearity, HIV-1 subtype detection, and correlation to the Roche Amplicor HIV-1 monitor test, version 1.5 (Amplicor). The limit of detection calculated by probit analysis was 23.8 copies/ml using the 2nd International WHO Standard and 30.8 copies/ml using Viral Quality Assurance (VQA) material. Serial dilutions of six patient samples were used to determine inter- and intra-assay reproducibility and linearity, which were very good (<8% coefficient of variation [CV]; between 1.7 and 7.0 log 10 copies/ml). Subtype detection was evaluated in the CAP/CTM, Amplicor, and Bayer Versant HIV-1 bdna 3.0 (Versant) assays using a commercially available panel. Versant averaged log 10 copies/ml lower than CAP/CTM and Amplicor averaged log 10 copies/ml lower than CAP/CTM for the subtype panel. Correlation with samples previously tested by Amplicor was excellent (R ; average difference [Amplicor value minus CAP/CTM value], log 10 copies/ml). Of the 305 HIV samples tested, 7 samples generated CAP/CTM titers between 1.0 and 2.75 log 10 copies/ml lower than those for Amplicor. Three of these samples revealed primer and probe mismatches that could account for the discrepancies. Otherwise, the CAP/CTM assay exhibits excellent sensitivity, dynamic range, reproducibility, and correlation with Amplicor in an automated format. Measurement of HIV-1 RNA in the plasma of infected patients is critical for guiding treatment. Over the past decade, several commercial quantitative HIV-1 RNA assays have become available that utilize endpoint PCR, isothermal amplification, or signal amplification techniques. Most recently, Abbott Molecular (Des Plaines, IL) and Roche Molecular Systems (Branchburg, NJ) received FDA approval for their real-time PCR-based systems, the RealTime HIV-1 assay and Cobas AmpliPrep/Cobas TaqMan HIV-1 Test (CAP/CTM), respectively. Each assay has its own advantages and disadvantages in terms of sensitivity, equipment requirements, throughput, dynamic range, subtype detection, and cost (1a, 4, 6, 7, 8, 11, 13, 15, 16). The CAP/CTM test includes a docked version that permits automated sample in results out analysis of specimens without user intervention. We evaluated this configuration and compared its performance to those of the Roche Amplicor HIV-1 monitor test, version 1.5 (Amplicor), and the Bayer Versant HIV-1 bdna 3.0 assay (Versant). Seven samples which gave discrepant Amplicor versus CAP/CTM results were further evaluated by sequencing analysis. * Corresponding author. Mailing address: ARUP Institute for Clinical and Experimental Pathology, 500 Chipeta Way, Salt Lake City, UT Phone: (801) , ext Fax: (801) michael.pyne@aruplab.com. Published ahead of print on 23 June MATERIALS AND METHODS Limit of detection. To determine the limit of detection in the CAP/CTM assay, nine dilutions of the WHO reference material (HIV-1 RNA, 2nd International Standard, 97/650; NIBSC, Potters Bar, United Kingdom) or Viral Quality Assurance (VQA) (Rush University Medical Center, Chicago, IL) were prepared in Basematrix diluent (SeraCare Life Sciences, Milford, MA). Fourteen replicates at each concentration (25, 20, 15, 10, 7.5, 5, 2.5, 1, 0.5 copies/ml for WHO; 40, 35, 30, 25, 20, 10, 5, 2, 0.5 copies/ml for VQA) were tested. The conversion of IU to copies for the WHO used the conversion listed in the CAP/CTM product insert (1.7 IU/copy). Probit analysis was used to determine the limit of detection (95% detection rate) and 95% confidence interval. Reproducibility and linearity. Six patient samples with high viral titers, serially diluted 7-fold in Basematrix, were used to evaluate intra- and interassay reproducibility and linearity of the CAP/CTM assay. Sufficient volume for five aliquots of each dilution was prepared and stored at 4 C until use. For each sample, three of five replicates were tested on the same input rack to evaluate intra-assay reproducibility and linearity. Interassay reproducibility was evaluated by testing two additional aliquots, each on a separate day. The average values for the intra-assay replicates were combined with the values for the two additional aliquots to determine interassay reproducibility and linearity. Reproducibility was evaluated by determining the mean and of the mean for each dilution. Linearity was evaluated by linear regression analysis, excluding the aliquots with the highest measured concentration, which were set at identity to establish the expected values. Subtype panels. To assess the ability of the CAP/CTM assay to measure viral loads in samples containing different group M subtype viruses, the members of the HIV RNA genotype performance panel PRD202 (SeraCare Life Sciences, Milford, MA) were diluted 3-fold in Basematrix. Each subtype was tested in triplicate by the Amplicor, CAP/CTM, and Versant assays using procedures. Correlation samples. Archived frozen plasma samples previously tested by Amplicor were retrieved from storage, deidentified, and stored at 70 C before testing. Prior to being tested in the CAP/CTM assay, the samples were thawed at room temperature, vortexed briefly, and transferred to the S-input tubes. A total of 218 patient samples were selected to evaluate assay correlation. An additional 2852

2 VOL. 48, 2010 CAP/CTM HIV-1 EVALUATION AND SEQUENCE POLYMORPHISMS 2853 TABLE 1. Limit of detection and probit analysis a 2nd International WHO Standard b Viral Quality Assurance c Copies/ml tested % Copies/ml tested % d a Nine dilutions of the 2nd International Standard or Viral Quality Assurance were prepared in Basematrix diluent. Fourteen replicates at each concentration were tested. Probit analysis was used to determine the limit of detection (95% detection rate) and 95% confidence interval. b Probit results: 23.8 copies/ml (95% CI: 20.1 to 27.5). c Probit results: 30.8 copies/ml (95% CI: 26.4 to 35.3). d Four samples were not tested due to an instrument. 63 samples with viral loads below the limit of detection of the UltraSensitive Amplicor assay ( 50 copies/ml) were selected for retesting in the CAP/CTM assay. Eighty-seven archived specimens that had previously been tested by Amplicor were also tested by CAP/CTM after the instrument was relocated to a different location in the laboratory. Linearity panel. Ten aliquots for each of the members of the AcroMetrix (Benicia, CA) ValiQuant HIV-1 RNA quantification panel were pooled to obtain a homogenous sample for testing. Each member was tested in triplicate on each of three days. The data were analyzed by Deming regression. Roche Amplicor assay. The Roche Amplicor HIV-1 monitor test, version 1.5, was performed as directed by the manufacturer using manual extraction (using either the UltraSensitive or Standard procedure, as specified by the ordering physician) and PCR setup. The amplifications were performed on Applied Biosystems GeneAmp PCR system 9600 or 9700 thermal cyclers. The microwell plates were washed using a Bio-Tek Instruments ELP-40 and read using a Bio-Tek Instruments ELx800 microtiter plate reader. Roche CAP/CTM assay. Nucleic acid extraction, PCR preparation, and amplification for the CAP/CTM assay were performed using the docked configuration of the Cobas AmpliPrep/Cobas TaqMan HIV-1 test system using protocols, reagents, and software provided by the manufacturer. The assay extracts nucleic acid from 0.85 ml of plasma using magnetic particles, previously described by Schumacher et al. (14). Versant HIV-1 RNA 3.0 assay (bdna). The Versant assay was performed as directed by the manufacturer, using the System 340 bdna analyzer (Siemens Healthcare Diagnostics, Deerfield, IL). Sequence analysis. Nucleic acids from 200 l of plasma were extracted from the samples using reagents from the Amplicor HIV-1 monitor kit. Briefly, 600 l lysis reagent was mixed with 200 l sample and the mixture was incubated for 10 min at room temperature. Then, 800 l isopropanol was added to each sample. Nucleic acids were pelleted by centrifugation at room temperature for 15 min in a microcentrifuge. The pellets were washed with 1 ml 70% ethanol and centrifuged for 5 min. The alcohol was removed, the tubes were pulse spun for 30 s, and residual ethanol was removed. The pellets were resuspended in 100 l HIV-1 sample diluent and stored at 80 C. An HIV-1 sequence flanking the CAP/CTM HIV-1 target region in gag was amplified by reverse transcriptase PCR (RT-PCR). MuLV reverse transcriptase was used for the RT reaction, and PCR amplification was carried out using Taq DNA polymerase. Hemi-nested amplifications were then performed to increase amplification sensitivity. The amplicons were treated with ExoSAP-IT (GE Healthcare, United Kingdom) and then were sequenced. The cycle sequencing reaction was set up using ABI BigDye Terminator v3.1 Chemistry (Applied Biosystems Inc., Foster City, CA) on the GeneAmp PCR System 9700 or 9800 fast thermal cycler (Applied Biosystems Inc., Foster City, CA). To determine the subtype of the virus in each sample, the gag sequence from each sample was aligned to sequences from HIV isolates of known genotypes from the GenBank sequence database using the ClustalW sequence alignment program, which also generates a phylogenetic tree, via the Lasergene software package (MegAlign version 5.07; DNASTAR Inc.). Sequence analysis was performed by Roche s Global Surveillance Program (RGSP) in Pleasanton, CA, using residual patient samples. The interpretation of sequencing results was done by Roche personnel in the context of Roche s proprietary HIV database and experience with the CAP/CTM assay. ARUP personnel were not granted access to the sequence data due to the proprietary nature of the data. RESULTS The limit of detection (LOD) of the CAP/CTM assay was determined using 14 replicates of nine concentrations of 2nd International WHO Standard material ranging between 0.5 and 25 copies/ml and VQA material ranging between 0.5 and 40 copies/ml (Table 1). Probit analysis predicts a 95% limit of detection of 23.8 copies/ml (95% confidence interval, 20.1 to 27.5) for WHO material and 30.8 copies/ml (95% confidence interval, 26.4 to 35.3) for VQA material. The reproducibility and linearity of the CAP/CTM assay were determined by replicate measurements of dilution series derived from six high-titer patient samples (Tables 2 and 3). Intra-assay slopes ranged from to , and interassay slopes ranged from to The intercept ranges were to and to for the intra- and interassay regressions, respectively. In addition to the patient sample dilutions to evaluate linearity, the six members of the ValiQuant panel, ranging from 100 to 5,000,000 copies/ml, were tested in triplicate on each of three days (Fig. 1). Some replicates of the 5,000,000-copies/ml panel member produced results above the upper limit of quantification for the assay (10,000,000 copies/ml). Deming regression analysis of the quantitative results produced the following results: measured log 10 copies/ml (assigned log 10 copies/ml) 0.254, S y/x D (R , n 47). All negative panel members were not by the CAP/CTM assay. A group M subtype panel was evaluated in the CAP/CTM, Amplicor, and Versant assays. This panel represents group M subtypes, including some common circulating recombinant forms (CRFs). As shown in Fig. 2, there were statistically significant differences (P 0.05) between assays for nearly every subtype. The following assay pairs were not significantly different (P 0.05): subtype A, Amplicor-Versant; subtype B, Amplicor-Versant and Amplicor-CAP/CTM; subtype D, Am-

3 2854 PYNE ET AL. J. CLIN. MICROBIOL. TABLE 2. Intra- and interassay precision a Dilution no. and comparison type Mean SEM log 10 HIV RNA copies/ml in the indicated sample (n) b A B C D E F Intra-assay ULQ LLQ LLQ (1) (2) (2) LLQ LLQ LLQ LLQ LLQ Interassay (1) (2) ULQ (1) LLQ LLQ (2) (1) LLQ LLQ LLQ LLQ LLQ a The intra- and interassay precision were evaluated by 7-fold serial dilutions of six high-titer patient samples (A to F). Five replicates were prepared and tested in the CAP/CTM assay. Three replicates were tested on a single rack to evaluate intra-assay reproducibility, and the remaining two replicates were tested on a second and third day (and compared to the mean of the intra-assay data) to evaluate interassay reproducibility. The expected values were calculated by setting the highest measured concentration at identity and using the dilution factor to calculate the remaining expected concentrations. The data assigned to identity in this manner were excluded from the regression analysis. b ULQ, greater than the upper limit of quantification ( 7 log 10 copies/ml). LLQ, less than the lower limit of quantification ( 48 copies/ml) or not. n 3 unless noted otherwise in parentheses (the number of replicates that produced quantitative results was less than the number of samples tested). plicor-versant; subtype G, Amplicor-CAP/CTM. Versant averaged log 10 copies/ml lower than CAP/CTM, and Amplicor averaged log 10 copies/ml lower than CAP/CTM. Sixty-three samples with UltraSensitive Amplicor results below the limit of detection ( 50 copies/ml) were retested with the CAP/CTM assay. One sample had an invalid QS (internal ). Forty-eight samples had no detectable HIV-1 RNA. Ten samples were but were below the quantification limit of the assay (48 copies/ml). Four samples were quantitated, measuring 60, 64, 94, and 116 copies/ml. These samples had insufficient volume for repeat testing. The results of the correlation of the CAP/CTM assay to the Amplicor assay using archived samples are shown in Fig. 3. A total of 218 samples were tested in the CAP/CTM assay; 1 sample had an invalid QS, 3 had clots, 17 were but not quantified, and 10 were not. A total of 187 samples produced valid quantitative results, with a Deming regression (measured as log 10 copies/ml) of CAP/CTM (Amplicor) (R , S y/x D 0.366). On average, the Amplicor assay results were log 10 copies/ml higher than those of the CAP/CTM assay. Ninety-six percent of the samples differed by fewer than 0.5 log 10 copies/ml; 98% of the samples differed by fewer than 1.0 log 10 copies/ml. One correlation sample (RIU025) had highly discrepant results in the correlation study. The Amplicor result was 26,000 copies/ml (4.4 log 10 copies/ml), while the CAP/CTM result was 59 copies/ml (1.8 log 10 copies/ml). The sample was retested with Amplicor, with a result of 46,000 copies/ml (4.7 log 10 copies/ml). There was insufficient sample for retesting in the CAP/CTM assay. Roche was notified of the discrepancy, and the remaining sample was sent to RGSP for sequence analysis (Table 4). Sequence information, including the gag target region of the CAP/CTM test, was obtained from the sample. Phylogenetic analysis determined the virus to be subtype B. Alignment of this sequence against the proprietary sequences of the CAP/ CTM assay primers and probe revealed mismatches to one of the primers and the probe. The three mismatches to the probe alone were determined to be unlikely to affect quantification of this specimen based on the location of the mismatches and TABLE 3. Intra- and interassay linear regression analysis a Intra-assay regression analysis Interassay regression analysis Sample A B C D E F a See Table 2 for an explanation of the method and calculations.

4 VOL. 48, 2010 CAP/CTM HIV-1 EVALUATION AND SEQUENCE POLYMORPHISMS 2855 FIG. 1. ValiQuant panel. The members of the AcroMetrix ValiQuant HIV-1 RNA quantification panel were tested in triplicate on each of 3 days. All replicates of the negative panel member were not. The data were analyzed by Deming regression, shown by the heavy line. The light line represents the line of identity. known mismatch tolerance of the assay probe (data not shown). In addition, quantification data from other HIV-1 samples with at least the same three probe mismatches showed a nominal titer difference (0.2 to 0.5 log 10 ) between the CAP/ CTM and Amplicor tests. The primer mismatch found in sample RIU025 is likely to affect quantification by the CAP/CTM test, since the mismatch is located near the 3 end of the primer. Other HIV-1 samples that contain the same mismatch have been shown to give titers that are 1.2 to 2.8 log 10 lower by the CAP/CTM test than by the Amplicor test. After completion of the evaluation described, the CAP/ CTM instrument was moved to a different location in the laboratory. To verify proper performance, additional experiments were performed. This included testing 87 additional archived specimens that had previously been tested using Amplicor (data not shown) (Deming regression [measured as log 10 copies/ml], CAP/CTM value [Amplicor value] 0.534; R ; S y/x D 0.496; n 86; average difference [Amplicor value minus CAP/CTM value], log 10 copies/ml). Although these were new samples, since they are deidentified, the possibility that one or more are from the same patient as the prior batch of samples cannot be excluded. Of the 87 samples, 6 had viral load measurements that were at least 1 log 10 copies/ml lower in the CAP/CTM assay than by Amplicor (Table 4, samples 1 to 6). To determine if these specimens also contained polymorphisms that may have interfered with quantification, aliquots were sent to RGSP for further testing. One sample FIG. 2. Subtype panel. The members of the HIV RNA genotype performance panel PRD202 (SeraCare Life Sciences, Milford, MA) were diluted 3-fold in Basematrix and tested in triplicate by the Amplicor, Versant, and CAP/CTM assays. The subtype designations and average log 10 copies/ml are shown. The bars represent 1 deviation. No HIV-1 RNA was by any assay for the negative-control panel member. For the pairs marked with an asterisk, P 0.05; all other pairs, P 0.05.

5 2856 PYNE ET AL. J. CLIN. MICROBIOL. FIG. 3. Correlation samples. A total of 218 samples previously tested in the Amplicor assay were tested in the CAP/CTM assay. The data were analyzed by the method of Bland and Altman (1). The bold line indicates the average difference, and the dashed lines indicate the average 2 deviations. (5A) was no longer available, but a plasma sample drawn from the same patient 2 weeks later was available. This sample (5B) was sent to RGSP for sequencing and testing in both Amplicor and CAP/CTM assays. Sequence surrounding the CAP/CTM target region was obtained from five of the six samples. No sequence information was obtained from sample 5B. This is likely due to the low titer of this sample. (Sample 5A was unable to be sequenced due to insufficient sample.) The sequences for the remaining five discrepant samples were determined to be subtype B based on phylogenetic analysis of their gag sequences. Each of the sequences from these samples contains at least one nucleotide mismatch to a primer or probe. Sample Amplicor log (copies/ml) CAP/CTM log (copies/ml) TABLE 4. Summary of samples with discrepancies a Difference log (copies/ml) For sample 1, there is one mismatch located near the 5 end of one of the primers. The location of this mismatch makes it unlikely to interfere with amplification. In addition, the identical mismatch has been found in five other samples previously analyzed by the RGSP that generated similar titers (titer differences, 0.6 to 0.3 log 10 ) between the CAP/CTM test and the Amplicor test. The sequence from sample 2 was found to have mismatches to one of the primers and the probe. The primer mismatches are located in the 5 half of the primer. A previous sample analyzed by the RGSP that contains the same primer mismatches plus one additional mismatch located closer to the 3 end of the primer generated titers that are nearly identical of mismatches present Mismatches likely to affect quantification Primers Probe Primers Probe Comments RIU /4.7 e e 1 3 Yes No Same sequence as sample No N/A No No Yes No Same sequence as RIU N/A No 5A b N/A N/A QNS; not sequenced 5B b 2.6/2.7 c,d,e 2.0 c 0.65 e Unknown Unknown Sequencing was not successful No Yes Mismatches not seen before by Roche a The seven samples that had large discrepancies between Amplicor and CAP/CTM results were further analyzed by sequencing analysis where possible. N/A, not applicable. b Sample 5A was of a quantity not sufficient (QNS) for retesting. However, a sample from the same patient drawn 2 weeks later (5B) was available. c These results were produced by Roche. d In addition to these results, four replicates had not results. e Sample was tested twice. The difference is based on the average of the results for the replicates.

6 VOL. 48, 2010 CAP/CTM HIV-1 EVALUATION AND SEQUENCE POLYMORPHISMS 2857 (difference, 0.1 log 10 ) between the CAP/CTM test and the Amplicor test, indicating that the primer mismatches seen in sample 2 are unlikely to be the cause of the titer discrepancy. The probe mismatch found in the sequence for sample 2 is common. There have been 40 samples sequenced by the RGSP that contain only this same probe mismatch, but all of these samples generated titers by the CAP/CTM test within one log 10 of the titers from the Amplicor test (titer differences, 0.44 to 0.63 log 10 ). Therefore, the probe mismatch found in the sequence of sample 2 is not likely to be the cause of the titer discrepancy. However, the possibility that the combination of the primer and probe mismatches may have contributed to the observed titer difference cannot be ruled out. The HIV-1 sequence from sample 4 contains two mismatches to the probe of the CAP/CTM test. Four additional samples (not connected with the current study) that have been sequenced by the RGSP have shown the same probe mismatches, with and without additional mismatches. These four samples generated similar titers (titer differences, 0.20 to 0.58 log 10 ) between the CAP/CTM test and the Amplicor test. Therefore, it is not likely that the probe mismatches found in sample 4 account for the titer discrepancy observed between the CAP/CTM and Amplicor tests. Based on the location of the mismatches and additional samples tested with the same mismatches, the mismatches found in samples 1, 2, and 4 are unlikely to be the cause of the underquantification of these samples. For samples 1, 2, and 4, the difference between Amplicor and CAP/CTM results for these samples was between 1.0 and 1.2 log 10 copies/ml, which could be accounted for by inherent variability in the two assays and might be resolved by replicate testing if sufficient material is available. The HIV-1 sequence obtained from sample 3 contains the same primer and probe mismatches as already described for sample RIU025. The gag sequences generated from these two samples were identical when aligned to each other. It is very unlikely for two different HIV-infected patients to contain identical gag sequences, suggesting that samples RIU025 and 3 are from the same source. Contamination is not likely the cause of the identical sequences, as the samples were analyzed at different locations and nearly 1 year apart. Since the samples were deidentified, it cannot be determined unequivocally if these are independent samples from the same patient. The HIV-1 sequence obtained from sample 6 contains four mismatches to the CAP/CTM probe and one mismatch to the downstream primer. The downstream primer mismatch is located near the 5 end of the primer. Based on its location, this mismatch is not likely to affect amplification efficiency. In addition, two samples analyzed previously by the RGSP contain at least the same downstream primer mismatch and were found to generate very similar titers (titer differences, 0.1 to 0.3 log 10 ) between the CAP/CTM test and the Amplicor test. However, based on their number and location, the four probe mismatches found in sample 6 may account for the titer discrepancy observed between the two assays. Although this exact mismatch pattern has not been observed in other samples collected by the RGSP, samples with sequences that contain similar numbers of mismatches in similar locations have been found to be underquantified by the CAP/CTM test by approximately one log 10 compared to those quantified by the Amplicor test (data not shown). DISCUSSION The of care for HIV-infected patients has long included viral load testing. Accurate measurements are critical for determining disease prognosis and guiding the course of treatment. Failure to accurately quantify HIV viral load can lead to inappropriate clinical management of patients and unnecessary virus transmission (3, 5). Viral load assays have evolved dramatically over the past decade. A trend for all quantitative viral testing during this period has been the adoption of real-time chemistries and instrumentation. Two assays (Abbott Molecular and Roche Molecular Systems) have received FDA approval for HIV-1 viral load testing in the real-time format. Compared to previous PCR testing which used endpoint detection, these realtime assays offer closed-tube automation with a wider analytical dynamic range. We evaluated several key performance characteristics of the FDA-approved CAP/CTM assay. This study documents the overall good performance of the Roche assay for clinical isolates and defined HIV subtype isolates. The LOD was determined by probit analysis to be slightly lower than the 40 copies/ml previously reported (14). Differences between the 1st and 2nd WHO s could account for this difference. The CAP/CTM assay exhibited excellent reproducibility and linearity over a 5-log 10 copies/ml range with variability of 10% or less. There were statistically significant differences between assays for most subtype panel members, with CAP/CTM measurements generally highest while Versant measurements were generally lowest. However, genetic variability between individual samples, as we have shown in this study, may have a greater consequence than variability of results between subtype members. Therefore, the results of subtype panel testing should be interpreted cautiously. Prior comparisons of the Amplicor and CAP/CTM assays have shown average differences (between Amplicor and CAP/ CTM) of (1a), 0.12 (8), 0.05 (11), (16), 0.01 (14), 0.48 (2), and 0.28 (6). Although the subtypes of the correlation samples in this study are unknown, they probably consist mainly of subtype B viruses, since they were derived from a pool of samples submitted from across the United States (reference 12 and unpublished observations at ARUP). This study also reveals the challenge of sequence diversity in the real-time format. We identified seven samples (from a total of 305 that were tested) with between 1.0 and 2.7 log 10 copies/ml underquantification by the CAP/CTM assay. Sequence analysis of these samples was successful in six of seven cases; unfortunately, no sequence was obtained from sample 5. The sequences from samples 1, 2, and 4 contain polymorphisms in the CAP/CTM primer and/or probe regions which are deemed by RGSP to be unlikely to contribute to the observed titer differences based on the location of the mismatches and data generated from other HIV-1 samples evaluated by RGSP. The sequences from samples RIU025, 3, and 6 contain polymorphisms that are likely to be contributing to the underquantification by the CAP/CTM test. The sequences generated from samples RIU025 and 3 are identical to each other,

7 2858 PYNE ET AL. J. CLIN. MICROBIOL. strongly suggesting that they were obtained from the same patient. This analysis suggests that 2 of the 304 (0.66%) unique HIV samples tested by the CAP/CTM test were confirmed to contain mismatches that are likely to interfere with quantification. The majority of the discrepant samples in this study had insufficient volume for repeat testing, with the exception of samples RIU025 and 5B (which were measured twice and six times in Amplicor, respectively). For samples containing similar sequence mismatches which have been previously analyzed by RGSP, the details of sample handling such as freeze/thaw cycles, age, and storage conditions of the samples are not available. These problems are compounded by the imperfect science of predicting effects of polymorphisms on PCR performance and the proprietary nature of the assay sequences, which require investigators to rely on manufacturer analysis (such as in this study) or to make assumptions about the primer and probe locations to interpret sequencing results (10). These factors pose a significant obstacle to achieving a rigorous understanding of effects of sequence variation on commercial HIV assays. The process of producing RNA transcripts from cloned PCR products as a surrogate for patient sample is one pathway for reducing the uncertainty of assay comparisons when sample is limiting. The approach is impractical for most laboratories and still leaves in question the basis for test failure for proprietary assays. However, the challenge posed by the continued diversification of the highly variable HIV genome in clinical isolates worldwide and its impact on testing is critically important and has increasingly become the focus of study. While most reports have emphasized underquantification of non-b subtypes (2, 3, 5, 6, 7, 9), the discrepant samples in this study and others (2, 10) showed that underquantification can also occur in subtype B and is not restricted to specific subtypes. Versions of the CAP/CTM product insert subsequent to the investigation of the discrepant samples described in this study have included a statement that though rare, mutations within the highly conserved region of the viral genome covered by the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test primers and/or probe may result in the underquantitation of or failure to detect the virus. Current practice dictates determining new baseline viral loads for patients when switching to a different assay when there is a significant average bias between the assays. Even when there is no average bias between the assays (such as in this study), it may be necessary to determine relative quantification between the methods for some patients, especially when test results do not agree with clinical symptoms. The frequency, severity, and subtype distribution of underquantified samples requires additional study. ACKNOWLEDGMENTS This study was performed in compliance with regulations concerning human subject research and was approved by the University of Utah Institutional Review Board. All components for Roche CAP/CTM testing, including the use of the instruments, disposables, reagents, genotype panels, reference s, and additional Amplicor and Versant testing, were provided or funded by Roche. The discrepant sample sequencing analysis and interpretation were performed by Roche. We extend our thanks to Denise Jones for preparing samples, the staff of the Molecular Hepatitis/Retrovirus Laboratory at ARUP for performing the Amplicor and Versant testing, Andrew Wilson for assistance with statistics, and Jody Harris and Monica Lin at Roche Molecular Systems for sequencing the discrepant samples. REFERENCES 1. Bland, J. M., and D. G. Altman Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i: a.Braun, P., R. Ehret, F. Wiesmann, F. Zabbai, M. Knickmann, R. Kühn, S. Thamm, G. Warnat, and H. Knechten Comparison of four commercial quantitative HIV-1 assays for viral load monitoring in clinical daily routine. Clin. Chem. Lab. Med. 45: Damond, F., B. Roquebert, A. Bénard, G. Collin, M. Miceli, P. Yéni, F. Brun-Vezinet, and D. Descamps Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays. J. Clin. Microbiol. 45: Delaugerre, C., B. Denis, G. Peytavin, P. Palmer, T. Mourez, J. Le Goff, J. Molina, and F. Simon Clinical and resistance consequences of misquantification of plasma and cerebrospinal fluid human immunodeficiency virus type 1 (HIV-1) RNA in an HIV-1 subtype G-infected patient. J. Clin. Microbiol. 47: Foulongne, V., B. Montes, M.-N. Didelot-Rousseau, and M. Segondy Comparison of the LCx human immunodeficiency virus (HIV) RNA Quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan assays for quantitation of HIV type 1 RNA in plasma. J. Clin. Microbiol. 44: Geelen, S., J. Lange, J. Borleffs, T. Wolfs, A. Weersink, and R. Schuurman Failure to detect a non-b HIV-1 subtype by the HIV-1 Amplicor Monitor test, version 1.5: a case of unexpected vertical transmission. AIDS 17: Gueudin, M., J. C. Plantier, V. Lemée, M. P. Schmitt, L. Chartier, T. Bourlet, A. Ruffault, F. Damond, M. Vray, and F. Simon Evaluation of the Roche Cobas TaqMan and Abbott RealTime extraction-quantification systems for HIV-1 subtypes. J. Acquir. Immune Defic. Syndr. 44: Holguín, A., M. López, M. Molinero, and V. Soriano Performance of three commercial viral load assays, Versant human immunodeficiency virus type 1 (HIV-1) RNA bdna v3.0, Cobas AmpliPrep/Cobas TaqMan HIV-1, and NucliSens HIV-1 EasyQ v1.2, testing HIV-1 non-b subtypes and recombinant variants. J. Clin. Microbiol. 46: Katsoulidou, A., M. Petrodaskalaki, V. Sypsa, E. Papachristou, C. G. Anastassopoulou, P. Gargalianos, A. Karafoulidou, M. Lazanas, T. Kordossis, A. Andoniadou, and A. Hatzakis Evaluation of the clinical sensitivity for the quantification of human immunodeficiency virus type 1 RNA in plasma: comparison of the new COBAS TaqMan HIV-1 with three current HIV- RNA assays LCx HIV RNA quantitative, VERSANT HIV-1 RNA 3.0 (bdna) and COBAS Amplicor HIV-1 Monitor v1.5. J. Virol. Methods 131: Kim, J. E., B. Beckthold, Z. Chen, J. Mihowich, L. Malloch, and M. J. Gill Identification of a novel HIV type 1 subtype H/J recombinant in Canada with discordant HIV viral load (RNA) values in three different commercial assays. AIDS Res. Hum. Retroviruses 23: Korn, K., B. Weissbrich, C. Henke-Gendo, A. Heim, C. M. Jauer, N. Taylor, and J. Eberle Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay. J. Clin. Microbiol. 47: Oliver, A. R., S. F. Pereira, and D. A. Clark Comparative evaluation of the automated Roche TaqMan real-time quantitative human immunodeficiency virus type 1 RNA PCR assay and the Roche AMPLICOR version 1.5 conventional PCR assay. J. Clin. Microbiol. 45: Osmanov, S., C. Pattou, N. Waler, B. Schwardländer, J. Esparza, and the WHO-UNAIDS Network for HIV Isolation and Characterization Estimated global distribution and regional spread of HIV-1 genetic subtypes in the year J. Acquir. Immune Defic. Syndr. 29: Pyne, M. T., E. Q. Konnick, A. Phansalkar, and D. R. Hillyard Evaluation of the Abbott investigational use only RealTime HIV-1 assay and comparison to the Roche Amplicor HIV-1 monitor test, version 1.5. J. Mol. Diagn. 11: Schumacher, W., E. Frick, M. Kauselmann, V. Maier-Hoyle, R. van der Vliet, and R. Babiel Fully automated quantification of human immunodeficiency virus (HIV) type 1 RNA in human plasma by the COBAS AmpliPrep/COBAS Taqman system. J. Clin. Virol. 38: Schutten, M Comparison of the Abbott Realtime HIV-1 and HCV viral load assays with commercial competitor assays. Expert Rev. Mol. Diagn. 8: Wolff, D., and A. Gerritzen Comparison of the Roche COBAS Amplicor Monitor, Roche COBAS Ampliprep/COBAS Taqman and Abbott RealTime test assays for quantification of hepatitis C virus and HIV RNA. Clin. Chem. Lab. Med. 45:

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA.

Laboratory for Clinical and Biological Studies, University of Miami Miller School of Medicine, Miami, FL, USA. 000000 00000 0000 000 00 0 bdna () 00000 0000 000 00 0 Nuclisens () 000 00 0 000000 00000 0000 000 00 0 Amplicor () Comparison of Amplicor HIV- monitor Test, NucliSens HIV- QT and bdna Versant HIV RNA

More information

Comparison of Three Roche HBV Viral Load Assay Formats

Comparison of Three Roche HBV Viral Load Assay Formats JCM Accepts, published online ahead of print on 25 April 2012 J. Clin. Microbiol. doi:10.1128/jcm.00746-12 Copyright 2012, American Society for Microbiology. All Rights Reserved. 1 Title 2 3 4 5 6 7 8

More information

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26 1 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Viral Load Quality Assessment Program Summary

More information

Technical Bulletin No. 161

Technical Bulletin No. 161 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 161 cobas 6800 HIV-1 Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

Technical Bulletin No. 162

Technical Bulletin No. 162 CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 162 cobas 6800 HCV Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations

More information

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up JCM Accepts, published online ahead of print on 11 May 2011 J. Clin. Microbiol. doi:10.1128/jcm.00721-11 Copyright 2011, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights

More information

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012

VQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012 1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral

More information

Received 12 September 2008/Returned for modification 4 November 2008/Accepted 12 April 2009

Received 12 September 2008/Returned for modification 4 November 2008/Accepted 12 April 2009 JOURNAL OF CLINICAL MICROBIOLOGY, July 2009, p. 2209 2217 Vol. 47, No. 7 0095-1137/09/$08.00 0 doi:10.1128/jcm.01761-08 Copyright 2009, American Society for Microbiology. All Rights Reserved. Evaluation

More information

HIV-1 Load Comparison Using Four Commercial Real-Time Assays

HIV-1 Load Comparison Using Four Commercial Real-Time Assays JOURNAL OF CLINICAL MICROBIOLOGY, Jan. 2011, p. 292 297 Vol. 49, No. 1 0095-1137/11/$12.00 doi:10.1128/jcm.01688-10 Copyright 2011, American Society for Microbiology. All Rights Reserved. HIV-1 Load Comparison

More information

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0

Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Clinical Chemistry / EVALUATION OF QUANTITATIVE HEPATITIS C VIRUS ASSAY Performance Characteristics of the COBAS AMPLICOR Hepatitis C Virus MONITOR Test, Version 2.0 Maria Erali, MS, 1 Edward R. Ashwood,

More information

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2017Oct27

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2017Oct27 1 The National Laboratory for HIV Reference Services is Accredited to ISO 15189 and ISO 17043 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology

More information

It takes more than just a single target

It takes more than just a single target It takes more than just a single target As the challenges you face evolve... HIV mutates No HIV-1 mutation can be considered to be neutral 1 Growing evidence indicates all HIV subtypes may be prone to

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Test, version 2.0 (TaqMan 48) Number: PQDx 0126-046-00 Abstract The COBAS AmpliPrep/COBAS TaqMan

More information

Introduction: Table/Figure Descriptions:

Introduction: Table/Figure Descriptions: Introduction: We have completed the analysis of your HIV RNA Validation Study. The validation plan was designed to verify the installation of an unmodified FDA-approved HIV RNA assay into your laboratory.

More information

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification

Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Oct. 2010, p. 3641 3647 Vol. 48, No. 10 0095-1137/10/$12.00 doi:10.1128/jcm.01306-10 Copyright 2010, American Society for Microbiology. All Rights Reserved. Performance

More information

Evaluation of the new Abbott mplus feature: Impact on clinical laboratory efficiencies with Abbott RealTime HIV-1, HCV, HBV and CT/NG assays

Evaluation of the new Abbott mplus feature: Impact on clinical laboratory efficiencies with Abbott RealTime HIV-1, HCV, HBV and CT/NG assays JCM Accepts, published online ahead of print on 2 October 2013 J. Clin. Microbiol. doi:10.1128/jcm.01672-13 Copyright 2013, American Society for Microbiology. All Rights Reserved. 1 2 3 4 5 Evaluation

More information

Performance evaluation of the new Roche COBAS AmpliPrep. COBAS TaqMan HIV-1 test version 2.0 for the quantification of HIV-1 RNA

Performance evaluation of the new Roche COBAS AmpliPrep. COBAS TaqMan HIV-1 test version 2.0 for the quantification of HIV-1 RNA JCM Accepts, published online ahead of print on 17 February 2010 J. Clin. Microbiol. doi:10.1128/jcm.01832-09 Copyright 2010, American Society for Microbiology and/or the Listed Authors/Institutions. All

More information

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS

altona RealStar Instructions for Use RealStar CMV PCR Kit /2017 EN DIAGNOSTICS altona DIAGNOSTICS Instructions for Use RealStar CMV PCR Kit 1.2 08/2017 EN RealStar RealStar CMV PCR Kit 1.2 For research use only! (RUO) 021202 INS-021200-EN-S01 48 08 2017 altona Diagnostics GmbH Mörkenstr.

More information

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO)

Diagnostic Methods of HBV infection. Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Diagnostic Methods of HBV infection Zohreh Sharifi,ph.D of Virology Research center, Iranian Blood Transfusion Organization (IBTO) Hepatitis B-laboratory diagnosis Detection of HBV infection involves

More information

Received 18 September 2009/Returned for modification 18 November 2009/Accepted 8 February 2010

Received 18 September 2009/Returned for modification 18 November 2009/Accepted 8 February 2010 JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 2010, p. 1195 1200 Vol. 48, No. 4 0095-1137/10/$12.00 doi:10.1128/jcm.01832-09 Copyright 2010, American Society for Microbiology. All Rights Reserved. Performance

More information

VQA Control SOP Version 4.0 Roche Amplicor HIV-1 DNA Test, v August 2007

VQA Control SOP Version 4.0 Roche Amplicor HIV-1 DNA Test, v August 2007 1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral

More information

CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD

CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD CALIBRATION OF ANALYTICAL STANDARDS FOR HBV-DNA, HCV-RNA AND HIV-1 RNA IN GENOME COPIES BY A REFERENCE METHOD AAJ van Drimmelen, E.R. Bax and W.G.V. Quint, BioQControl (BQC), Delft Diagnostic Laboratories

More information

on May 2, 2018 by guest

on May 2, 2018 by guest JCM Accepts, published online ahead of print on 5 September 2012 J. Clin. Microbiol. doi:10.1128/jcm.01221-12 Copyright 2012, American Society for Microbiology. All Rights Reserved. 1 2 3 4 Title: Ten

More information

Molecular Diagnosis Future Directions

Molecular Diagnosis Future Directions Molecular Diagnosis Future Directions Philip Cunningham NSW State Reference Laboratory for HIV/AIDS & Molecular Diagnostic Medicine Laboratory, SydPath St Vincent s Hospital Sydney Update on Molecular

More information

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection.

For in vitro Veterinary Diagnostics only. Kylt Rotavirus A. Real-Time RT-PCR Detection. For in vitro Veterinary Diagnostics only. Kylt Rotavirus A Real-Time RT-PCR Detection www.kylt.eu DIRECTION FOR USE Kylt Rotavirus A Real-Time RT-PCR Detection A. General Kylt Rotavirus A products are

More information

Multilaboratory Evaluation of Real-Time PCR Tests for Hepatitis B Virus DNA Quantification

Multilaboratory Evaluation of Real-Time PCR Tests for Hepatitis B Virus DNA Quantification JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2011, p. 2854 2858 Vol. 49, No. 8 0095-1137/11/$12.00 doi:10.1128/jcm.00471-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Multilaboratory

More information

on October 4, 2018 by guest

on October 4, 2018 by guest JCM Accepts, published online ahead of print on 3 July 2012 J. Clin. Microbiol. doi:10.1128/jcm.01249-12 Copyright 2012, American Society for Microbiology. All Rights Reserved. 1 Title: Performance of

More information

Instructions for Use. RealStar Influenza Screen & Type RT-PCR Kit /2017 EN

Instructions for Use. RealStar Influenza Screen & Type RT-PCR Kit /2017 EN Instructions for Use RealStar Influenza Screen & Type RT-PCR Kit 4.0 05/2017 EN RealStar Influenza Screen & Type RT-PCR Kit 4.0 For research use only! (RUO) 164003 INS-164000-EN-S01 96 05 2017 altona

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: Abbott RealTime HIV-1 (m2000sp) Number: PQDx

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: Abbott RealTime HIV-1 (m2000sp) Number: PQDx WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: Abbott RealTime HIV-1 (m2000sp) Number: PQDx 0145-027-00 Abstract Abbott RealTime HIV-1 (m2000sp) assay with product code 2G31, which

More information

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx

WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: Alere q HIV-1/2 Detect WHO reference number: PQDx WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: Alere q HIV-1/2 Detect WHO reference number: PQDx 0226-032-00 Alere q HIV-1/2 Detect with product codes 270110050, 270110010 and 270300001,

More information

Trends in molecular diagnostics

Trends in molecular diagnostics Trends in molecular diagnostics Detection of target genes of interest Quantification Infectious diseases HIV Hepatitis C & B TB / MAC Cytomegalovirus Herpes simplex Varicella zoster CT/GC HPV Profiling

More information

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0

Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas. TaqMan HCV Test, version 1.0 JCM Accepts, published online ahead of print on 25 September 2013 J. Clin. Microbiol. doi:10.1128/jcm.02102-13 Copyright 2013, American Society for Microbiology. All Rights Reserved. 1 2 Hepatitis C Virus

More information

Panther has new prey

Panther has new prey Raising the Bar for Performance Testing Panther has new prey The Aptima HIV-1 Quant Dx assay leads the hunt for HIV-1 diagnosis and viral load monitoring. Freedom to work the way you choose Run what assays

More information

A Sensitive Branched DNA HIV-1 Signal Amplification Viral Load Assay with Single Day Turnaround

A Sensitive Branched DNA HIV-1 Signal Amplification Viral Load Assay with Single Day Turnaround A Sensitive Branched DNA HIV-1 Signal Amplification Viral Load Assay with Single Day Turnaround Mark A. Baumeister, Nan Zhang, Hilda Beas, Jesse R. Brooks, Jesse A. Canchola, Carlo Cosenza, Felix Kleshik,

More information

CMV Diagnostic Strategies: Current and Future

CMV Diagnostic Strategies: Current and Future CMV Diagnostic Strategies: Current and Future Tony Mazzulli, MD, FRCPC, FACP Microbiologist-in-Chief Mount Sinai Hospital & University Health Network, Toronto Faculty/Presenter Disclosure Relationships

More information

Roche Molecular Biochemicals Application Note No. HP 1/1999

Roche Molecular Biochemicals Application Note No. HP 1/1999 Roche Molecular Biochemicals Application Note No. HP 1/1999 Nucleic Acid Purification High Pure Viral Nucleic Acid Kit High Pure 16 System Viral Nucleic Acid Kit Efficiency of Hepatitis C Virus sample

More information

Quantification of HBV, HCV genotype and HIV subtype panels

Quantification of HBV, HCV genotype and HIV subtype panels Quantification of HBV, HCV genotype and HIV subtype panels Harry van Drimmelen 1,2, Wim Quint 2, Nico Lelie 3 and the international NAT study group 1. Biologicals Quality Control, 2. DDL Diagnostic Laboratory,

More information

Instructions for Use. RealStar Influenza S&T RT-PCR Kit /2017 EN

Instructions for Use. RealStar Influenza S&T RT-PCR Kit /2017 EN Instructions for Use RealStar Influenza S&T RT-PCR Kit 3.0 01/2017 EN RealStar Influenza S&T RT-PCR Kit 3.0 For research use only! (RUO) 163003 INS-163000-EN-S02 96 01 2017 altona Diagnostics GmbH Mörkenstr.

More information

Product # Kit Components

Product # Kit Components 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Pneumocystis jirovecii PCR Kit Product # 42820 Product Insert Background Information

More information

Switch from Signal Amplification to COBAS AmpliPrep/COBAS TaqMan 48: Is There a Need to Re-Baseline?

Switch from Signal Amplification to COBAS AmpliPrep/COBAS TaqMan 48: Is There a Need to Re-Baseline? Switch from Signal Amplification to COBAS AmpliPrep/COBAS TaqMan 48: Is There a Need to Re-Baseline? MADHUCHHANDA CHOUDHARY, GLORIA CALDITO, JANICE M MATTHEWS-GREER ABSTRACT Background: Accurate quantitation

More information

RealLine HIV quantitative Str-Format

RealLine HIV quantitative Str-Format Instructions for use DETECTION AND QUANTIFICATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Research Use Only (RUO) Attention! Please read the information about quantification process carefully!

More information

Norgen s HIV Proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad T1000 Cycler

Norgen s HIV Proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad T1000 Cycler 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com HIV Proviral DNA PCR Kit Product# 33840 Product Insert Intended

More information

Antiviral Therapy 2014; 19: (doi: /IMP2723)

Antiviral Therapy 2014; 19: (doi: /IMP2723) Antiviral Therapy 214; 19:449 454 (doi: 1.3851/IMP2723) Original article Performance characteristics of the COBAS Ampliprep/COBAS TaqMan v2. and the Abbott RealTime hepatitis C assays implications for

More information

Pelagia Research Library. European Journal of Experimental Biology, 2015, 5(10):1-5

Pelagia Research Library. European Journal of Experimental Biology, 2015, 5(10):1-5 Available online at www.pelagiaresearchlibrary.com European Journal of Experimental Biology, 2015, 5(10):1-5 ISSN: 2248 9215 CODEN (USA): EJEBAU Molecular diagnosis of human immuno deficiency virus (HIV)

More information

VARIATION IN MEASUREMENT OF HIV RNA VIRAL LOAD

VARIATION IN MEASUREMENT OF HIV RNA VIRAL LOAD VARIATION IN MEASUREMENT OF HIV RNA VIRAL LOAD SOURCES OF VARIATION (RANDOM VS SYSTEMATIC) MAGNITUDE OF EACH SOURCE CONSEQUENCES FOR CONFIDENCE LIMITS AROUND MEASUREMENTS AND CHANGES DATA FROM ROCHE HIV

More information

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012

altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 always a drop ahead. 11/2012 altona DIAGNOSTICS altona DIAGNOSTICS RealStar CMV PCR Kit 1.0 11/2012 altona Diagnostics GmbH Moerkenstr. 12 22767 Hamburg Germany phone +49 40 548 06 76-0 fax +49 40 548 06 76-10 e-mail info@altona-diagnostics.com

More information

Norgen s HIV proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad icycler

Norgen s HIV proviral DNA PCR Kit was developed and validated to be used with the following PCR instruments: Qiagen Rotor-Gene Q BioRad icycler 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com HIV Proviral DNA PCR Kit Product # 33840 Product Insert Background Information

More information

HIV viral load testing in the era of ART. Christian Noah Labor Lademannbogen, Hamburg

HIV viral load testing in the era of ART. Christian Noah Labor Lademannbogen, Hamburg HIV viral load testing in the era of ART Christian Noah Labor Lademannbogen, Hamburg 1 Life expectancy of patients on ART Data from the UK Collaborative HIV Cohort (UK CHIC) Requirements: Early diagnosis

More information

Diagnostic Methods of HBV and HDV infections

Diagnostic Methods of HBV and HDV infections Diagnostic Methods of HBV and HDV infections Zohreh Sharifi,ph.D Blood Transfusion Research Center, High Institute for Research and Education in Transfusion Medicine Hepatitis B-laboratory diagnosis Detection

More information

The quantification of hepatitis C virus (HCV)

The quantification of hepatitis C virus (HCV) Overestimation and Underestimation of Hepatitis C Virus RNA Levels in a Widely Used Real-Time Polymerase Chain Reaction Based Method Stéphane Chevaliez, 1,2 Magali Bouvier-Alias, 1,2 Rozenn Brillet, 1,2

More information

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay

Phosphate buffered saline (PBS) for washing the cells TE buffer (nuclease-free) ph 7.5 for use with the PrimePCR Reverse Transcription Control Assay Catalog # Description 172-5080 SingleShot Cell Lysis Kit, 100 x 50 µl reactions 172-5081 SingleShot Cell Lysis Kit, 500 x 50 µl reactions For research purposes only. Introduction The SingleShot Cell Lysis

More information

Author's response to reviews

Author's response to reviews Author's response to reviews Title: Stable Hepatitis C Virus RNA Detection by RT-PCR During Four Days Storage Authors: Anne-Isabelle de Moreau de Gerbehaye (ng.deradigues@yucom.be) Monique Bodeus (bodeus@mblg.ucl.ac.be)

More information

A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA

A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA Original Article A multi-center clinical study comparing Sansure Magb and CAP/CTM HBV tests in the quantitative detection of HBV DNA Xiaoyu Fu 1,Deming Tan 1, Xiaoguang Dou 2,Jinjun Chen 3,Juan Wu 1 1

More information

Human Rotavirus A. genesig Standard Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only

Human Rotavirus A. genesig Standard Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only TM Primerdesign Ltd Human Rotavirus A Non structural protein 5 (NSP5) genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus A Rotavirus is a genus

More information

Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction

Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction 370 Comparison of MagNA Pure 96, Chemagic MSM1, and QIAamp MinElute for Hepatitis B Virus Nucleic Acid Extraction Seong-Ho Kang 1, Eun Hee Lee 2, Geon Park 1, Sook Jin Jang 1, and Dae Soo Moon 1 1 Department

More information

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ

Micropathology Ltd. University of Warwick Science Park, Venture Centre, Sir William Lyons Road, Coventry CV4 7EZ www.micropathology.com info@micropathology.com Micropathology Ltd Tel 24hrs: +44 (0) 24-76 323222 Fax / Ans: +44 (0) 24-76 - 323333 University of Warwick Science Park, Venture Centre, Sir William Lyons

More information

HIV-1 Viral Load Real Time (RG)

HIV-1 Viral Load Real Time (RG) -1 Viral Load Real Time (RG) Real Time RT-PCR type 1 RNA quantification assay MSP Reg. pending Valdense 3616. 11700. Montevideo. Uruguay. phone (598) 2 336 83 01. Fax (598) 2 336 71 60. Info@atgen.com.uy

More information

Kit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive

Kit Components Product # EP42720 (24 preps) MDx 2X PCR Master Mix 350 µl Cryptococcus neoformans Primer Mix 70 µl Cryptococcus neoformans Positive 3430 Schmon Parkway Thorold, ON, Canada L2V 4Y6 Phone: 866-667-4362 (905) 227-8848 Fax: (905) 227-1061 Email: techsupport@norgenbiotek.com Cryptococcus neoformans End-Point PCR Kit Product# EP42720 Product

More information

Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for. hepatitis B virus DNA quantification.

Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for. hepatitis B virus DNA quantification. Performance of version 2.0 of the Cobas AmpliPrep/Cobas TaqMan real-time PCR assay for hepatitis B virus DNA quantification. Stéphane Chevaliez, Magali Bouvier-Alias, Syria Laperche, Christophe Hézode,

More information

For purification of viral DNA and RNA from a wide range of sample materials

For purification of viral DNA and RNA from a wide range of sample materials QIAamp virus kits For purification of viral DNA and RNA from a wide range of sample materials Automatable on QIAGEN s proven QIAamp Kits set the standard for purification of viral DNA and RNA. QIAamp virus

More information

Stability of native, lyophilized and inactivated standards

Stability of native, lyophilized and inactivated standards Stability of native, lyophilized and inactivated standards Harry van Drimmelen Bio Quality Control, Heiloo, Netherlands Satellite Meeting before IPFA-PEI 25 th Workshop Twenty-five Years Standardization

More information

Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets

Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets Liu et al. Virology Journal (2017) 14:94 DOI 10.1186/s12985-017-0759-8 RESEARCH Open Access Real-time PCR assays for hepatitis B virus DNA quantification may require two different targets Chao Liu 1,2,3,

More information

Human Rotavirus B. Non structural protein 5 (NSP5) 150 tests. Quantification of Human Rotavirus B genomes Advanced kit handbook HB10.01.

Human Rotavirus B. Non structural protein 5 (NSP5) 150 tests. Quantification of Human Rotavirus B genomes Advanced kit handbook HB10.01. PCR Max Ltd TM qpcr test Human Rotavirus B Non structural protein 5 (NSP5) 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus B Rotavirus is a genus of double-stranded

More information

Hepatitis C Virus (RNA)A

Hepatitis C Virus (RNA)A Hepatitis C Virus (RNA)A 2012 EQA Programme Final Report QAV994112 (HCVRNA12A) Professor Jacques Izopet Scientific Expert on behalf of QCMD Report authorised by the QCMD Executive in July 2012 A UKAS accredited

More information

Molecular diagnosis of infectious disease - Method validation and environmental setting 傳染病的分子診斷 - 方法確認與環境背景. WC Yam 任永昌

Molecular diagnosis of infectious disease - Method validation and environmental setting 傳染病的分子診斷 - 方法確認與環境背景. WC Yam 任永昌 Molecular diagnosis of infectious disease - Method validation and environmental setting 傳染病的分子診斷 - 方法確認與環境背景 WC Yam 任永昌 Dept of Microbiology Queen Mary Hospital The University of Hong Kong Clinical and

More information

Official Journal of the European Communities COMMISSION

Official Journal of the European Communities COMMISSION 16.5.2002 EN Official Journal of the European Communities L 131/17 COMMISSION COMMISSION DECISION of 7 May 2002 on common technical specifications for in vitro-diagnostic medical devices (notified under

More information

Abbott RealTime HIV-1

Abbott RealTime HIV-1 E RealTime HIV-1 6L18 51-602146/R2 Abbott RealTime HIV-1 Customer Service: 1-800-553-7042 This package insert must be read carefully prior to use. Package insert instructions must be followed accordingly.

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx

WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: VERSANT HIV-1 RNA 1.0 Assay (kpcr) Number: PQDx 0115-041-00 Abstract The VERSANT HIV-1 RNA 1.0 Assay (kpcr) with product codes 10375763,

More information

Supplemental Materials and Methods Plasmids and viruses Quantitative Reverse Transcription PCR Generation of molecular standard for quantitative PCR

Supplemental Materials and Methods Plasmids and viruses Quantitative Reverse Transcription PCR Generation of molecular standard for quantitative PCR Supplemental Materials and Methods Plasmids and viruses To generate pseudotyped viruses, the previously described recombinant plasmids pnl4-3-δnef-gfp or pnl4-3-δ6-drgfp and a vector expressing HIV-1 X4

More information

Experience with Standardisation of Blood Virology NAT. Clare Morris Division of Retrovirology National Institute for Biological Standards and Control

Experience with Standardisation of Blood Virology NAT. Clare Morris Division of Retrovirology National Institute for Biological Standards and Control Experience with Standardisation of Blood Virology NAT Clare Morris Division of Retrovirology National Institute for Biological Standards and Control Background of Blood Virology Standardisation In the

More information

Interpreting quantitative HBV, HCV and HIV-1 nucleic acid testing

Interpreting quantitative HBV, HCV and HIV-1 nucleic acid testing ISSN : 0974-7435 Volume 10 Issue 24 Interpreting quantitative HBV, HCV and HIV-1 nucleic acid testing Manoj K.Rajput National Institute of Biologicals, Noida 201309, (INDIA) E-mail : mkrajput@hotmail.com

More information

Received 19 April 2011/Returned for modification 6 June 2011/Accepted 13 June 2011

Received 19 April 2011/Returned for modification 6 June 2011/Accepted 13 June 2011 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2011, p. 2899 2904 Vol. 49, No. 8 0095-1137/11/$12.00 doi:10.1128/jcm.00785-11 Copyright 2011, American Society for Microbiology. All Rights Reserved. Comparative

More information

HBV PUBLIC HEALTH IMPLICATIONS

HBV PUBLIC HEALTH IMPLICATIONS جزايری دکتر سيد محمد آزمايشگاه ھپاتيت B -دانشکده بھداشت ويروس شناسی- گروه دانشگاه علوم پزشکی تھران کنگره ارتقا کيفيت- ١٣٩٢ HBV PUBLIC HEALTH IMPLICATIONS 2 billion people have been infected by HBV worldwide.

More information

HIV-RNA reference panels

HIV-RNA reference panels The kit insert contains a detailed protocol and should be read carefully before testing the run control to ensure optimal performance Table of contents Overview HIV-RNA panels for sensitivity analysis...

More information

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring

Monitoring of HCV RNA. Hepatitis C Requirements of Antiviral Therapy Monitoring Monitoring of HCV RNA Hepatitis C Requirements of Antiviral Therapy Monitoring Hepatitis C Requirements of Antiviral Therapy Monitoring Author: Ute Hofmann,, Konrad-Zuse-Str. 1, 07745 Jena, Germany Beatrix

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Qualitative Test, version 2.0 (TaqMan 48) Number: PQDx 0221-046-00 Abstract COBAS AmpliPrep/COBAS

More information

AccuSet HBV Worldwide Performance Panel

AccuSet HBV Worldwide Performance Panel PACKAGE INSERT 0805-033 INTENDED USE The is intended for use by diagnostic manufacturers, researchers, and clinical laboratories to develop, evaluate, or troubleshoot HBV test methods. Characterized samples

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: COBAS AmpliPrep/COBAS TaqMan HIV-1 Qualitative Test, version 2.0 (TaqMan 96) Number: PQDx 0200-046-00 Abstract COBAS AmpliPrep/COBAS

More information

A Novel Duplex Real-Time Reverse-Transcription PCR Assay for the Detection of Influenza A and the Novel Influenza A(H1N1) Strain

A Novel Duplex Real-Time Reverse-Transcription PCR Assay for the Detection of Influenza A and the Novel Influenza A(H1N1) Strain Viruses 2009, 1, 1204-1208; doi:10.3390/v1031204 OPEN ACCESS viruses ISSN 1999-4915 www.mdpi.com/journal/viruses Article A Novel Duplex Real-Time Reverse-Transcription PCR Assay for the Detection of Influenza

More information

Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood

Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood Brief Communication Clinical Microbiology Ann Lab Med 2017;37:147-150 https://doi.org/10.3343/alm.2017.37.2.147 ISSN 2234-3806 eissn 2234-3814 Performance of the Real-Q EBV Quantification Kit for Epstein-Barr

More information

High Failure Rate of the ViroSeq HIV-1 Genotyping System for Drug Resistance Testing in Cameroon, a Country with Broad HIV-1 Genetic Diversity

High Failure Rate of the ViroSeq HIV-1 Genotyping System for Drug Resistance Testing in Cameroon, a Country with Broad HIV-1 Genetic Diversity JOURNAL OF CLINICAL MICROBIOLOGY, Apr. 2011, p. 1635 1641 Vol. 49, No. 4 0095-1137/11/$12.00 doi:10.1128/jcm.01478-10 Copyright 2011, American Society for Microbiology. All Rights Reserved. High Failure

More information

Human Rotavirus C. genesig Advanced Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests

Human Rotavirus C. genesig Advanced Kit. DNA testing. Everything... Everyone... Everywhere... Non structural protein 5 (NSP5) 150 tests TM Primerdesign Ltd TM Primerdesign Ltd Human Rotavirus C Non structural protein 5 (NSP5) genesig Advanced Kit 150 tests DNA testing Everything... Everyone... Everywhere... For general laboratory and research

More information

AccuVert HIV-1 Seroconversion Panel PRB974 ( )

AccuVert HIV-1 Seroconversion Panel PRB974 ( ) PACKAGE INSERT PRB974 (0600-0258) INTENDED USE PRB974 (0600-0258) is a group of serial bleeds from an individual plasma donor during seroconversion. This panel is intended for use by diagnostics manufacturers

More information

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR

(DNA) Real-time PCR. Exicycler 96 Rotor-Gene Q/6000 PCR Real-Time (DNA) Real-time Exicycler 96 Rotor-Gene Q/6000 IU Mix1 Mix2 IU/μl IU/μl IU/μl IU/μl IU/μl IPC NTC C 1 Lot# 2 Freeze & thawing 1 MSDS: Material Safety Data Sheets (TaqMan Real-time ' FAM ' BHQ1

More information

HIV-1 AccuVert TM Seroconversion Panel

HIV-1 AccuVert TM Seroconversion Panel PACKAGE INSERT PRB954 (0600-0238) INTENDED USE The is intended for use by diagnostic manufacturers, researchers, and clinical laboratories to develop, evaluate, or troubleshoot test methods. Characterized

More information

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007

Received 8 January 2007/Returned for modification 20 February 2007/Accepted 4 June 2007 JOURNAL OF CLINICAL MICROBIOLOGY, Aug. 2007, p. 2529 2536 Vol. 45, No. 8 0095-1137/07/$08.00 0 doi:10.1128/jcm.00058-07 Copyright 2007, American Society for Microbiology. All Rights Reserved. Comparison

More information

Clinical Performance of Roche COBAS 4800 HPV Test

Clinical Performance of Roche COBAS 4800 HPV Test JCM Accepts, published online ahead of print on 9 April 2014 J. Clin. Microbiol. doi:10.1128/jcm.00883-14 Copyright 2014, American Society for Microbiology. All Rights Reserved. 1 1 2 3 4 5 6 Clinical

More information

DATA SHEET. Provided: 500 µl of 5.6 mm Tris HCl, 4.4 mm Tris base, 0.05% sodium azide 0.1 mm EDTA, 5 mg/liter calf thymus DNA.

DATA SHEET. Provided: 500 µl of 5.6 mm Tris HCl, 4.4 mm Tris base, 0.05% sodium azide 0.1 mm EDTA, 5 mg/liter calf thymus DNA. Viral Load DNA >> Standard PCR standard 0 Copies Catalog Number: 1122 Lot Number: 150298 Release Category: A Provided: 500 µl of 5.6 mm Tris HCl, 4.4 mm Tris base, 0.05% sodium azide 0.1 mm EDTA, 5 mg/liter

More information

AccuVert HBV Seroconversion Panel PHM941(M) ( )

AccuVert HBV Seroconversion Panel PHM941(M) ( ) PACKAGE INSERT PHM941(M) (0605-0061) INTENDED USE PHM941(M) (0605-0061) is a group of serial bleeds from an individual plasma donor during HBV seroconversion. This panel is intended for use by diagnostics

More information

P0141 HBV 1000 copies/ml genotype reference panel

P0141 HBV 1000 copies/ml genotype reference panel P0141 HBV 1000 copies/ml genotype reference panel P0141 The kit insert contains a detailed protocol and should be read carefully before testing the run control to ensure optimal performance Table of contents

More information

Human Rotavirus A. genesig Advanced Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only

Human Rotavirus A. genesig Advanced Kit. Non structural protein 5 (NSP5) 150 tests. Primerdesign Ltd. For general laboratory and research use only TM Primerdesign Ltd Human Rotavirus A Non structural protein 5 (NSP5) genesig Advanced Kit 150 tests For general laboratory and research use only 1 Introduction to Human Rotavirus A Rotavirus is a genus

More information

Khalid Alquthami (Correspondence) Regional Lab, Makkah. Saudi Arabia

Khalid Alquthami (Correspondence) Regional Lab, Makkah. Saudi Arabia 1 Regional Lab, Makkah. Saudi Arabia stract: Clinical specificity and genotype/subtype detection of viruses using the Cobas TaqScreen MPX system V 2.0, which is a nucleic acid test (NAT) that uses multiples

More information

Case Studies, or Verification Vignettes

Case Studies, or Verification Vignettes Case Studies, or Verification Vignettes 1 Vignette #1 Change from One Automated AST to Another Your lab is changing from one FDA-cleared automated AST to another Is a verification study required? 2 Yes

More information

Human diagnostics. Better be Sure: Quantify HDV & HBV viral load. RoboGene product family

Human diagnostics. Better be Sure: Quantify HDV & HBV viral load. RoboGene product family Human diagnostics Better be Sure: Quantify HDV & HBV viral load. RoboGene product family 2 RoboGene Product Family Improved patient management: Standardized monitoring of HBV DNA and HDV RNA viral load.

More information

Evaluation of Dried Blood Spots (DBS) for Human Immunodeficiency Virus (HIV-1) Drug Resistance Testing

Evaluation of Dried Blood Spots (DBS) for Human Immunodeficiency Virus (HIV-1) Drug Resistance Testing Evaluation of Dried Blood Spots (DBS) for Human Immunodeficiency Virus (HIV-1) Drug Resistance Testing Dawit Assefa 2, Woldaregay E.Abegaz 3, Teferi Gedif 2, Belete Tegbaru 1, Dereje Teshome 1, Tesfaye

More information

HIV-1 Seroconversion Panel PRB973

HIV-1 Seroconversion Panel PRB973 A SERACARE PANEL PRODUCT INTENDED USE The is a group of serial bleeds from an individual plasma donor during seroconversion. This panel is intended for use by diagnostics manufacturers and clinical laboratorians

More information

Human HBcAb IgM ELISA kit

Human HBcAb IgM ELISA kit Human HBcAb IgM ELISA kit Catalog number: NR-R10163 (96 wells) The kit is designed to qualitatively detect HBcAb IgM in human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES

More information

Worlock et al. Virology Journal (2017) 14:66 DOI /s

Worlock et al. Virology Journal (2017) 14:66 DOI /s Worlock et al. Virology Journal (2017) 14:66 DOI 10.1186/s12985-017-0727-3 METHODOLOGY Open Access Analytical characteristics and comparative evaluation of Aptima HCV quant Dx assay with the Abbott RealTime

More information

Avian Influenza A H5N8

Avian Influenza A H5N8 TM Primerdesign Ltd Avian Influenza A H5N8 Hemagglutinin (HA) gene & Neuraminidase (NA) gene genesig Standard Kit 150 tests For general laboratory and research use only 1 Introduction to Avian Influenza

More information