A General Unknown Screening for Drugs and Toxic Compounds in Human Serum. Thesis

Size: px
Start display at page:

Download "A General Unknown Screening for Drugs and Toxic Compounds in Human Serum. Thesis"

Transcription

1 A General Unknown Screening for Drugs and Toxic Compounds in Human Serum Thesis Submitted to the Faculty of Natural Sciences of the University of Basel, Switzerland for the Fulfilment of the Requirements for a Ph. D. Degree by Stefan Sturm from Biel (BE) Basel, 2005

2 Accepted by the faculty of natural sciences On proposal of Prof. Dr. Hauser Prof. Dr. Eberle Prof. Dr. Drewe Dr. Scholer Basel, 20. September 2005 Prof. Dr. Wirz Dean

3 Table of Content 1. List of Abbreviations Summary Introduction GENERAL UNKNOWN SCREENING CHOICE OF SPECIMEN EXTRACTION TECHNIQUES Liquid-Liquid Extraction Solid-Phase Extraction CHROMATOGRAPHY HPLC-DAD System LC-MS System Interfaces Electrospray Ionization Atmospheric Pressure Chemical Ionization Ion Transfer Capillary and Ion Optics Analysators and LC-MS Detection Modes Data-Dependent Acquisition (DDA) Ion Detection System Vacuum System Adduct Formation Retention Time Chromatographic Conditions Influences of Biological Matrix Influences of Biological Matrix on Chromatographic Behaviour Influences of Biological Matrix on Detection in HPLC Influences of Biological Matrix on Detection in MS Evaluation of Matrix Effects in MS MASS SPECTRA LIBRARIES Single Mass Spectrometry Tandem Mass Spectrometry Data Dependent Acquisition Sample Preparation in a General Unknown Screening Chromatic Conditions in a General Unknown Screening RESEARCH OBJECTIVE Materials and Methods GENERAL UNKNOWN SCREENING WITH HPLC Materials Sample and Buffer Preparation Procedures Extraction Procedure HPLC-DAD Method Extraction Recovery Yields Identification and Semi-Quantification Quantification GENERAL UNKNOWN SCREENING WITH MS Materials Sample and Buffer Preparation Procedures Extraction Procedure Chromatographic and Mass Spectral Conditions Mass Spectral Library Building Conditions Results and Discussion SAMPLE PREPARATION AND CHROMATIC CONDITIONS FOR A GUS

4 Table of Content Sample Preparation and Chromatic Conditions with HPLC Sample Preparation and Chromatographic Conditions with LC-MS LIBRARY SCREENING AND QUANTIFICATION Identification with HPLC A Semi-Quantitative Procedure with HPLC Method Development Automatic Library Searching with LC-MS A Semi-Quantitative Procedure with LC-MS Comparison of LC-MS and HPLC-DAD CASE REPORTS CONCLUSION Appendix PROPOFOL Introduction Materials and Methods Reagents and Chemicals Apparatus and Instrumental Conditions Standard Preparation Extraction Procedure Results and Discussion REMIFENTANIL Introduction Materials and Methods Reagents and Chemicals Apparatus and Instrumental Conditions Standard Preparation Extraction Procedure Results and Discussion APOMORPHINE Introduction Materials and Methods Reagents and Chemicals Apparatus and Instrumental Conditions Sample Preparation Extraction Procedure Results and Discussion References

5 List of Abbreviations 1. LIST OF ABBREVIATIONS Ac ACN amu APCI API AUC CID c max CNLS CV DAD Dc DDA DIS EMIT ESI FAB GC GUS HPLC IS LC LLE LOD ME MRM MS PB RF RI RP RT alternating current acetonitrile atomic mass unit atmospheric pressure chemical ionization atmospheric pressure ionization area under the curve collision induced dissociation peak concentration constant neutral-loss scanning coefficient of variation diode array detection direct current data-dependent acquisition daughter-ion scanning enzyme multiplied immunotechniques electrospray ionization fast atom bombardment gas chromatography general unknown screening high performance liquid chromatography internal standard liquid chromatography liquid-liquid extraction limit of detection matrix effect multiple-reaction monitoring mass spectrometry particle beam radio frequency retention index reversed-phase retention time 5

6 List of Abbreviations RRT SIM SPE SRM STA t 1 / 2 TIC t max TS TOF UV XIC relative retention time selected-ion monitoring mode solid phase extraction selected-reaction monitoring systematic toxicological analysis half-life total ion current time to peak concentration thermospray time-of-flight ultra violet extracted ion chromatogram 6

7 Summary 2. SUMMARY Screening for a wide range of drugs and toxic compounds in biological samples is an important task for forensic and clinical toxicological laboratories. Our object was to develop a method to detect and identify a wide range of compounds by using HPLC- DAD and LC-MS. A solid-phase extraction procedure using polymer-based columns was developed for the HPLC-DAD procedure. The extraction method appeared very universal. Ninteyfour of 100 drugs were extracted with a recovery of more than 50%. For the LC-MS procedure, the Prospekt solid-phase extraction was chosen for its ability to be linked to the atmospheric pressure chemical ionisation (APCI) ion source. With this procedure all of the tested drugs with LC-MS were successfully extracted. Hundred compounds from a variety of classes were investigated, if they can be identified and detected at low toxic serum concentration. Limits of detection were determined in spiked serum samples. It was found that HPLC-DAD was able to detect 61 out of 100 compounds with our procedure at low toxic serum concentration. The limit of detection (LOD) for the majority of the tested drugs (76%) was ng/ml. Drugs and toxic compounds were detected by comparison of the retention time and UV spectra with references compounds stored in a library. The LC-MS instrument was operated in the positive and in the negative mode using datadependent acquisition. Tandem mass spectrometry (MS-MS) was applied to identify toxicologically relevant substances in serum. An application program was created in order to detect automatically the unknown compounds. Drugs and metabolites were identified on the basis of their relative retention times, pseudo-molecular ions and fragment ions. A total of more than 400 spectra of more than 350 compounds were recorded. The corresponding relative retention times were added to the spectra in the constructed libraries (one for the positive and the other library for the negative mode). 7

8 Summary Eightyseven drugs of compounds were identified from serum using on-line solid-phase extraction with LC-MS-MS. The limit of detection (LOD) for the majority of compounds (67%) was 100 ng/ml, ranging from 10 to 4000 ng/ml. With the presented fully automated data-dependent LC-MS-MS procedure drugs can be analysed in serum with a high specificity and sensitivity. The LODs were sufficiently low to detect compounds at low toxic concentrations in serum. The integrated software drastically reduced the interpretation time. It was demonstrated that with the DDAmediated-LC-MS-MS screening approach almost all of the drugs detected by the conventional techniques as well as additional drugs were identified. This technique is useful for GUS and confirmation analysis in clinical and forensic toxicology. In general, LOD for compounds are lower with the LC-MS procedure than the HPLC procedure. For compounds not able to detect with the LC-MS procedure such as analgesics and barbiturates HPLC-DAD appeared to be the complement method. 8

9 Introduction 3. INTRODUCTION 3.1. GENERAL UNKNOWN SCREENING The identification of unknown compounds taken during an acute poisoning episode is a challenging task in clinical and forensic toxicology. The screening for drugs and toxic compounds is called general unknown screening (GUS) or systematic toxicological analysis (STA). This procedure is an analytical method aimed to detect and identify unknown xenobiotics in biological fluids. Acute poisoning with drugs is one of the most often occurring types of intoxication with exogenous compounds. In the year 2003, the number of intoxication incidents was about in Switzerland. Drugs are the reason for two-thirds of the poisonings in Switzerland with an assured medical diagnosis. The poisonings can be voluntary (suicidal intention) in 22% or accidental (mostly children) in 72% of the cases according to the annual report of the Swiss Toxicological Centre. Following the usual course of STA, samples will initially be analysed by immunochemical techniques (enzyme multiplied immunotechniques (EMIT), fluorescence polarization immunoassays, etc) (1, 2). These preliminary immunochemical screening procedures mainly concern rapid-response analytical tools providing a binary yes/no response, which indicates whether the target analytes are present above a preset concentration threshold or not. The samples providing a yes response to one or more compound classes or target substances are analysed with a conformation method. Apart from the first-line automated immunoassays available for the most common drugs, the GUS procedures currently used in clinical and forensic toxicology involve chromatographic techniques, ideally coupled to specific detectors (i.e., gas chromatography-mass spectrometry (GC-MS) (3) or high performance liquid chromatography (HPLC) coupled to diode array detection (DAD) (4-6)). None of these methods is sufficient to identify all possible toxic compounds. GC is limited to apolar, volatile and thermally stable compounds (7). For instance, some polar 9

10 Introduction compounds with no or little UV absorbency will neither be detected by GC-MS nor by HPLC-DAD. MS is more specific and reliable than DAD. The coupling of MS to HPLC seems to be a possibility to increase the range of compounds amenable to MS (8) CHOICE OF SPECIMEN The choice of specimen is often dictated by the clinical situation. The most common specimens used for the screening of drugs are blood, serum/plasma, and urine. Blood, plasma and serum are interchangeable in most methods. Urine is the most frequent specimen used in most hospital situations and may require hydrolysis prior to the isolation procedure to convert drug conjugates (glucuronides) to more easily measurable compounds. In forensic medicine the most investigated specimen is whole blood EXTRACTION TECHNIQUES Chromatographic techniques require a preanalytical isolation procedure to separate drugs from a biological matrix. There are two main suitable procedures for a GUS: liquid-liquid extraction and solid-phase extraction. The extract can be concentrated and dissolved in an appropriate volume of mobile phase. In some cases filtration or precipitation of proteins with an organic solvent (such as acetonitrile) prior to injection can provide a more direct means to introduce a sample into a HPLC. But these extraction procedures are not sufficient for a GUS Liquid-Liquid Extraction This has been the traditional method for isolating drugs from biological specimen. A distribution of the solute occurs between two not mixable liquids (specimen as the aqueous phase and a solvent as the organic phase). The isolation has to be performed at a ph at which the analyte is uncharged. For acidic drugs the aqueous phase has to be acidified, for basic drugs the aqueous phase has to be basified. In liquid-liquid extraction (LLE) ethyl acetate, acetone, chloroform, toluene, dichloromethane, butyl acetate and diethyl ether have been used as organic phases (9). The extraction power of these solvents differ not much from each other (10). 10 Although LLE is still used today, there are a number of drawbacks that limit its usefulness. Liquid-liquid extractions are difficult to automate. So this extraction procedure is labour intensive and time consuming. It s more difficult to reproduce the

11 Introduction resulting data than with SPE (11). Emulsion formation, increased solvent use and subsequent waste leading to environmental and safety issues are additional problems of LLE. These drawbacks have led to the development of other separation methods such as solid-phase extraction (SPE) Solid-Phase Extraction SPE on columns consists of four basic steps. Condition to wet and activate the column packing bed is the first step. Then the analytes of interest are loaded on the column to allow retention. In SPE, the analytes to be extracted are partioned between a solid and a liquid. They must have greater affinity for the solid phase than for the sample matrix. Ideally interfering compounds are rinsed off during the washing step and then the analytes are desorbed with a solvent appropriate for instrumental analysis. The principles of separation involve intermolecular forces between the analyte and the sorbent, the liquid phase or sample matrix. These intermolecular forces are dipoledipole forces, hydrogen bonding, ionic interactions and Van der Waal forces. The main sorbents in GUS used for solid phase extraction are reversed-phase and ionexchange. Reversed-phase (RP) chromatography partitions organic solutes from a polar phase (generally aqueous) to a non-polar phase, which may be in form of a hydrocarbon chain or polymeric sorbent. It involves non-polar interaction of the solute with the stationary phase through low-energy Van der Waals forces. The affinity of the solute for the sorbent depends upon its hydrophobicity. The solute is eluted with organic solvent, often acetonitrile, ethyl acetate or methanol. The most common RP sorbents for SPE are chemically bonded silica phases (12-14). Most widely used bonded phases are C 18 and C 8 sorbents (15). Pre-treatment is necessary with solvents such as acetonitrile or methanol. Other sorbents used for RP extractions include highly cross-linked co-polymers (styrene-divinylbenzene) (13, 16). Compared with silica sorbents, these sorbents are generally more hydrophobic, more retentive, stable over a larger ph range and do not have secondary (unbounded hydroxy silanol) groups. They can be used in automated on-line combined extraction and analysis systems. 11

12 Introduction Oasis HLB (Waters) combines hydrophilic N-vinylpyrrolidone and lipophilic divinylbenzene, providing a hydrophilic-lipophilic balance. That means these cartridges have a large range to interact with different substances. This sorbent requires no conditioning step. Drying out does not destroy the structure of the sorbent compared to others (13). In ion-exchange sorbents, compounds are extracted by a high-energy ionic interaction with the sorbent. Strong anion-exchangers retain anions by quarternary amines as exchange groups, cation-exchangers groups retain cations by sulfonic acids. For extraction, the ph of the samples is adjusted so that the solute molecules are ionized. They are absorbed at oppositely charged sites of the sorbent. Changing the ph elutes them. Mixed-mode sorbents have both non-polar and ion-exchange functional groups. The combination is chosen so that an analyte is retained by both mechanisms (15). The SPE mechanism is more selective than liquid-liquid extraction because it is based on interaction between sample components and the sorbent, as well as on solubility. The more selective retention mechanism of SPE has a number of advantages over liquid-liquid extractions such as cleaner extracts, increased selectivity for the compounds of interest, no emulsion formation, larger selection of solvents available for use, and smaller volumes of solvent may be used for extraction. In general SPE can be used for smaller sample sizes. SPE also is less time consuming, and the procedure can be automated. Because of the short analysis time, volatile compounds also may be analyzed without significant loss (17). The majority of methods have used silica-based columns. Although successful, silicabased sorbents may have a number of drawbacks. First, for bonded phases, the reaction between the organosilane and the silica is incomplete, leaving unreacted silanol groups that can act as cation-exchange sites and creating a secondary retention mechanism. End-capping reduces the amount of free silanols by converting the hydroxyl group to a methoxy group but it is not 100% effective. Second, the sorbent is not stable at ph extremes. At ph below 2, there can be breakdown of the silyl ether linkage. 12

13 Introduction The polymer-based sorbents were developed to overcome the limitations of silicabased sorbents. The water-wettable sorbent is promoted for its ability to retain a wide spectrum of polar and nonpolar compounds and to remain stable from ph 1 to 14 (13). The automation of solid-phase extraction has different benefits. Precision, recovery and accuracy with fewer operator errors, reduced analysis time are some of these advantages (12, 14). In addition the automated SPE system can be connected directly to an analyser. Disadvantages can be the risk of carry-over and the loss of sensitivity (18) CHROMATOGRAPHY An autosampler is used to inject samples automatically into the liquid chromatography (LC) system. The high performance liquid chromatograph separates a sample into its chemical components by liquid chromatography. In liquid chromatography, the sample mixture partions between a solid stationary phase (column) of large surface area and a liquid mobile phase that passes through the column. The chemical properties of the components, the column and the mobile phase determine when each component elutes from the LC and enters the DAD system or the MS HPLC-DAD System Diode-array ultraviolet detectors record the absorbance of compounds over a range of wavelengths (e.g., nm) as they pass through the detector flow cell, thus allowing the on-line acquisition of UV spectra. The HPLC-DAD system has been extensively shown to be an efficient and practical method for both clinical and forensic toxicological investigations. This system is able to screen for a wide range of compounds including benzodiazepines, analgesics, diuretics, tricyclic antidepressants, drugs of abuse (amphetamines, cocaine) and barbiturates. Elliott and Hale had shown a screening of basic, acidic, and neutral drugs, providing reproducible results (19). Using the diode-array facility, it is also possible to detect any probable metabolites, which are not available as drug standards by comparing UV spectrum of the proposed metabolite and the parent compound. Some phase-one metabolic transformations 13

14 Introduction (especially N-dealkylation and sometimes hydroxylation) do not alter the UV chromophore of a drug. The system and the retention index (RI) database should be combined with additional techniques such as liquid chromatography-mass spectrometry, thus providing a powerful identification tool that combines RI values, UV spectra, molecular masses and ion fragmentation spectra LC-MS System The LC-MS system consists of an atmospheric pressure ionization (API) source, ion optics, mass analyzer, and ion detection system. Ionization of the sample takes place in the API source. The ions produced in the API source are transmitted by the ion optics into the mass analyser. Selected ions are ejected from the mass analyzer and reach the ion detection system where they produce a signal. Fig. 1 shows a schematic drawing of a liquid-chromatography mass spectrometer system. Atmospheric pressure ionization (APCI, ESI) Heated capillary Ion optics Ion trap / quadrupole / time-of-flight LC inlet Clustered ion Tube lens and skimmer Ion detection system Fig. 1. Schematic drawing of a liquid-chromatography mass spectrometer Interfaces LC-MS interfaces remove the mobile phase and ionize the analyte. Different LC-MS interface types such as moving belt, direct liquid introduction, continuous-flow or frit-terminated fast atom bombardment (FAB), particle beam (PB), thermospray (TS), electrospray (ESI) and atmospheric pressure chemical ionization (APCI) were developed. PB, FAB or TS have several limitations such as less sensitivity or less universality. Today, two relatively robust LC-MS interfaces types seem to have 14

15 Introduction become the golden standards of LC-MS, the atmospheric pressure ionization techniques, ESI and APCI. ESI and APCI have been used for a large majority of the applications of LC-MS in the last years. ESI and APCI involve a soft ionization process, which can be overcome by using collision-induced dissociation, which provides thorough fragmentation of the compounds Electrospray Ionization The ESI mode transfers ions in solution into the gas phase. Many samples, which were previously not suitable for mass analysis (for example, heat-labile compounds or high molecular weight compounds) can be analyzed by ESI. ESI typically produces mass spectra consisting of multiply charged ions depending on the structure of the analyte and the solvent. ESI can be used to analyze any polar compound that generates an ion in solution. Generated ions include adduct ions (for example, polyethylene glycols can be analyzed from a solution containing ammonium acetate, because of the adduct formation between the NH + 4 ions in solution and oxygen atoms in the polymer). With ESI the range of molecular weights that can be analyzed is greater than 100'000 atomic mass units (amu), due to multiple charging. So proteins as well as peptides can be analyzed. The flow rate from the LC into the MS detector can vary over a range of 1 µl/min to 1000 µl/min. Transported by the mobile phase the sample enters the ESI needle. To the needle a high voltage (+/- 5 kv) is applied. The needle with the aid of nitrogen gas sprays the sample solution into fine mist droplets. These droplets are electrically charged at their surface. Solvent evaporates from the droplets. The electrical charge density increases up to a critical point known as the Raylight stability limit. At this point, the droplets divide into smaller droplets (20, 21). This is because the electrostatic repulsion is greater than the surface tension. The process is repeated many times. Finally, sample ions are ejected from the cluster ions into the gas phase. This takes place in a heated capillary. 15

16 Introduction The ESI process is affected by droplet size, surface charge, liquid surface tension, solvent volatility, and ion solvation strength. Large droplets with high surface tension, low volatility, strong ion solvation, low surface charge, and high conductivity prevent a good electrospray process (20) Atmospheric Pressure Chemical Ionization Like ESI, APCI is a soft ionization technique. APCI provides molecular weight information for compounds of medium polarity that has some volatility. APCI is typically used to analyze molecules with molecular weight up to amu. APCI is a gas phase ionization technique. Therefore, the ph of the analyte in the gas phase and solvent vapour play an important role in the APCI process. APCI is a very robust ionization technique. It is less affected by minor changes in most variables such as changes in buffer and buffer strength. The rate of solvent flowing from LC into the MS detector in APCI mode is typically high (between 0.2 and 2 ml/min) (22). The APCI nozzle vaporizes the sample solution in a high temperature tube. The needle is located near the end of the tube. A high voltage is applied to the corona discharge needle. Reagent ions are formed through a series of chemical reactions. First primary ions are formed. These primary ions can be nitrogen, oxygen gas or solvent molecules. The primary ions react with the molecules in the sample to form ions (21). APCI as well as ESI can be used in positive or negative ion polarity mode. A positive applied voltage is used to generate positive ions and a negative applied voltage to generate negatives. For most molecules, the ion-positive mode produces a stronger ion current. This is especially true for molecules with one or more basic nitrogen (or other basic) atoms. Molecules, which generally produce strong negative ions, with acidic sites such as carboxylic acids and acidic alcohols, are an exception to this general rule. Although, in general, fewer negative ions are produced than positive ions, negative ion polarity can be more specific. This is because the negative ion polarity mode sometimes generates less chemical noise than does the positive mode. Thus, selectivity might be better in the negative ion mode than in the positive ion mode. 16

17 Introduction Ion Transfer Capillary and Ion Optics The ion transfer capillary assists in desolvating ions that are produced by ESI and APCI probes. Ions are focussed into the ion transfer capillary in the atmospheric pressure region and transported to the skimmer region by decreasing pressure gradient and electrostatic forces. The ion transfer capillary can be heated. Typical temperatures are 150 to 200 C. Ions from the ion transfer capillary enter the tube lens. The tube lens has a dependent potential to focus the ions towards the skimmer. If the tube lens offset voltage is high, collisions with the background gas can be energetic enough to cause fragmentation. This fragmentation is called the ion source collision induced dissociation. The skimmer acts to reduce the number of neutral molecules and large charged particles. These particles would create detector noise. Ion optics (in our apparatus octapoles) transmits ions from the API source to the mass analyzer. Magnetic fields are used to direct and focus the ion stream coming from the source Analysators and LC-MS Detection Modes Mass analysis of ionized substances is performed using one or two mass analyzers (MS-MS), which consist predominantly of ion traps and quadrupoles, sometimes of sector field and time-of-flight instruments (TOF). They can operate in the full scan mode or in the more sensitive selected-ion monitoring mode (SIM) detecting positive or negative ions. The MS-MS combination provides additional possibilities. Both analysers can be operated in a scan or in a selected-ion monitoring mode. The four combinations are: Parent-ion scanning (scan mode in the first, SIM in the second analyzer), daughter-ion scanning (DIS; SIM in the first, scan mode in the second analyzer), constant neutral-loss scanning (CNLS; scan mode in both analyzers) or selected-reaction monitoring (SRM; SIM in both analyzers). DIS is preferable for the identification of drugs and or their metabolites in complex matrices. Separation is performed on the LC and in the first mass analyzer, while structural information is obtained by fragmentation in the second analyzer. SRM 17

18 Introduction (also named multiple-reaction monitoring MRM) is the most powerful technique for quantification of small amounts of analyte in complex matrices (23, 24). Quadrupole is a quadrilateral array of square rods that acts as an ion transmission device. A radio-frequency voltage and direct current (dc) offset voltage, that are applied to the rods give rise to an electric field that guides the desired ions along the axis of the quadrupole. The other ions collides with the four square rods. During ion transmission, the offset voltage is negative for positive ions and positive for negative ions. Quadrupoles can also be placed in tandem to perform fragmentaion. Fragmentation takes place between the two quadrupoles in the collision cell. The most common set-up is the triple quadrupole mass spectrometer. The ion trap includes three electrodes: the entrance endcap electrode, the exit endcap electrode and the ring electrode. Both endcap electrodes have a small hole in their centre to permit the passage of ions into and out of the mass analyzer cavity. Various alternating current (ac) voltages are applied to the ring and endcap electrodes to trap, fragment, and eject ions according to their mass-to-charge ratios. The processes that occur in the mass analyzer can be divided in four steps: ion storage, ion isolation (SIM, SRM, and MS/MS full scan), collision induced dissociation (SRM and MS/MS full scan), and ion scan out. The application of a radio-frequency voltage to the ring electrode produces a threedimensional quadrupole field within the mass analyzer cavity. This time-varying field drives ionic motion in both the axial (towards the endcaps) and radial (from the ring electrode toward the centre) directions. Ionic motion must be stable in both the axial and radial directions for an ion to remain trapped. During ion scan out, the system produces a mass-dependent instability to eject ions from the mass analyzer in the axial direction. The voltage at which an ion is ejected from the mass analyzer is called its resonance voltage. Ac voltages that are applied to the endcap electrodes stimulate motion of the ions in the axial direction. The voltages applied to the endcap electrodes are equal in amplitude but are 180 out of phase to one another. When the radio-frequency 18

19 Introduction applied to the endcaps equals the resonance frequency of a trapped ion, the ion gains kinetic energy and the ion is ejected. During the collision induced dissociation a voltage is applied to the endcap electrodes to fragment ions into product ions. This voltage is not strong enough to eject an ion from the mass analyzer. However, ion motion in the axial direction is enhanced and the ion gains kinetic energy. The mass analyzer cavity contains helium as a collision activation partner. After many collisions with helium gas, the ion gains enough internal energy to cause it to dissociate into product ions Data-Dependent Acquisition (DDA) In the first step (so-called "survey scan") the MS-MS instrument is operated in the full-scan single-mass mode using the second mass filter in order to select ions above a predefined intensity threshold. Instantly, the first mass filter is set to selectively transmit these high-intensity ions to the collision cell where fragmentation energy is switched on and the resulting fragments are analysed by the second mass filter in the scan mode. Finally after a short time the instrument switches back to the initial conditions, with the possibility to set a refractory period for the last ions selected. The major advantage of this approach is its high specificity and selectivity, as the spectra recorded come from a single parent ion. It can be difficult to detect toxic compounds among background noise. The setting of a given threshold can be important due to the intense and highly variable background noise produced by extracts of real samples. If the threshold is given a high value it results in very poor sensitivity but high specificity, and if it is given a very low value it may result in too much "noisy" information Ion Detection System The ion detection system includes a 15 kv conversion dynode and a channel electron multiplier. A potential of + 15 kv for negative ion detection or 15 kv for positive ion detection is applied to the conversion dynode. When an ion strikes the surface of the conversion dynode, one or more secondary particles are produced. These secondary particles can include positive, negative ions, electrons, and neutrals. When positive ions strike a negatively charged conversion dynode, the secondary particles of interest are negative ions and electrons. When negative ions strike a positively 19

20 Introduction charged conversion dynode, the secondary particles of interest are positive ions. These secondary particles are focused by the curved surface of the conversion dynode and are accelerated by a voltage gradient into the electron multiplier. Secondary particles from the conversion dynode strike the inner walls of the electron multiplier with sufficient energy to eject electrons. The ejected electrons are accelerated farther into the cathode, drawn by the increasingly positive gradient. Due to the funnel shape electrons do not travel far. They strike again the surface, thereby causing the emisson of more electrons. Thus, a cascade of electrons is created that finally results in a measurable current. The current is proportional to the number of secondary particles striking the cathode Vacuum System Vacuum is necessary to perform and maintain a mass spectra analysis. At normal pressure the produced ions would collide with molecules (N 2, O 2, etc.) passing from the API stack to the ion detection system. High vacuum levels cause reduced sensitivity, and reduced lifetime of the electron multiplier. The vacuum system evacuates the region around the API stack, ion optics, mass analyzer, and ion detection system. A forepump establishes the vacuum necessary for the proper operation of the turbomolecular pump. The pump maintains a pressure of approximatively 1.33 mbar. A turbomolecular pump provides the vacuum for the ion optics and analyzer regions. Under normal operating conditions the vacuum in the API region is 1020 mbar and 2.67x10-5 mbar in the analyzer region Adduct Formation Besides matrix effects, adduct formation also adds to the complexity of LC-MS. Generally, ESI or APCI result in deprotonated (M-H) - molecules in the negative mode and in protonated (M+H) + molecules in the positive ionization mode. Several adduct ions such as (M+Na) +, (M+K) +, or (M+NH 4 ) + were also reported (25, 26). The exact mechanism in adduct formation is not clearly understood, carboxyl or carbonyl ether or ester groups are believed to be responsible for binding the alkali metal ions. Adduct formation process is not reproducible and consequently it is not clear, what adduct ion can be used for MRM. Addition of ammonium results in (M+NH 4 ) +. The first step in fragmentation is the loss of neutral NH 3. The latter can then fragment further. Sodium adduct ions are much more stable and yield less 20

21 Introduction fragments. It is clear that the evaluation of both matrix effects as well as adduct formation should be included in the validation procedure of a LC-MS method Retention Time It is unusual for a substance to have an absolute retention time that remains constant either over a period of time (because of a gradual change of stationary phase) or between different batches of columns with nominally identical packing material. In order to account for this variability, relative retention time or retention index (RI) can be used as a method of correction. They are based on the relationship between a substance s retention time and those of a reference compound. The use of retention index was shown to be advantageous (27) Chromatographic Conditions Drug screening is a qualitative technique for the identification of the presence of drugs. Gradient chromatography is needed for the analysis of the large diversity of substances. For quantitative LC analyses of compounds identified by screening, it can be more efficient to use isocratic chromatography Influences of Biological Matrix Influences of Biological Matrix on Chromatographic Behaviour Bogusz et al. showed that the co-extracted biological matrix did not exert any specific influence on the chromatographic behaviour (28). The biological matrix did not affect the chromatography of acidic, neutral and basic drugs analyzed by means of gradient HPLC. To identify drugs extracted from biological samples, it is possible to use the retention time of pure drugs Influences of Biological Matrix on Detection in HPLC The presence of extracted matrix substances can affect the identification of drugs by HPLC with a diode array detector spectrum library. These compounds can co-elute at the same time as some toxicologically relevant substances and overlap the corresponding UV spectra. The purity of the biological extract is therefore of critical importance for successful HPLC-DAD identification. 21

22 Introduction Influences of Biological Matrix on Detection in MS Matrix effect (ME) in mass spectrometry can be defined as any change in the ionization process of an analyte due to a co-eluting compound (29). This can result either in an enhancement or in a suppression of the ionization. Ion suppression or enhancement affects the precision, sensitivity and accuracy of an analytical procedure (30). Sample matrix, co-eluting compounds, and cross-talk can contribute to ion suppression. ESI is more impacted to ion suppression than APCI (31). In electrospray ionization (ESI) the ionization process is taking place in the liquid phase. Matrix effects in ESI are due to a competition of matrix compounds and molecules of interest for access to the droplet surface and subsequent gas-phase emission. Besides that, matrix constituents can change properties of the surface tension and viscosity. These factors are known to affect the ionization process (29). In APCI, the ionization is taking place in the gas phase. The non-volatile matrix compounds are thought to co-precipate with the analyte molecule. This influences the ionization process (32). In summary, ion suppression results form the presence of less volatile compounds that can change the efficiency of droplet formation or droplet evaporation, which in turn affects the amount of charged ion in the gas phase that reaches the detector Evaluation of Matrix Effects in MS Matuszewski et al. described a procedure for the evaluation of matrix effects (33). Three sets of samples are necessary. Set A consists of neat standard solutions. For Set B blank matrices are supplemented (after extraction) with the same amount of standards as used for set A. Set C consists of extracts of different blank matrices, supplemented with the same amount of standards but added before extraction. Matrix effects (ME%), recovery (RE%) and process efficiency (PE%) can be calculated with different formula. For the calculation the resulting peak areas are needed. 22

23 Introduction ME% = B/A x 100 RE% = C/B x 100 PE% = C/A x 100 Bonfiglio et al. described another procedure to evaluate matrix effects (34). This procedure is based on the postcolumn infusion of an analyte in a chromatographic run of an extract or a blank matrix. This signal is compared to the signal obtained with the post-column infusion of this same model analyte in a chromatographic run with eluent only. This procedure indicates also critical areas in the chromatogram. Typical examples of substances known to influence the ionization process include salts and other endogenous compounds (fatty acids, triglycerides), dosing vehicles (polyethylenglycol, propylenglycol and cremophore), anticoagulants, and constituents of sampling material (e.g. polymers) (35, 36). It was also demonstrated, that the ionization efficiency of a co-eluting internal standard (IS) is influenced by high levels of the compound of interest. Sample preparation also influences the amount of ion suppression. Protein precipitation alone results in pronounced matrix effects over the whole chromatographic run. SPE was able to remove efficiently hydrophilic interfering compounds. But hydrophobic interferences were increased. There are a few strategies to eliminate matrix effects, but often it is very difficult. A reduction of matrix constituents injected can help. This can be done by injecting a smaller sample volume, a diluted sample or by applying more selective extraction techniques. Then, co-elution of the analyte and matrix constituents should be avoided. But the separation on the liquid chromatographic system is limited. The use of co-eluting internal standard(s) such as labelled ISs seems ideal because it is expected that the matrix effect on the analyte and on the IS is identical. But labelled internal standards are not always available and costs are often very expensive. In addition, these standards cannot be used in a GUS. 23

24 Introduction The mobile phase influences the ionization efficiency in LC-MS. Additives can also have effects on matrix induced ion suppression or enhancement of an analyte. The addition of small amounts of ammonium formate resulted in better ME% values. Higher levels or to high levels of other acids suppressed the signal (37) MASS SPECTRA LIBRARIES Gas chromatography-mass spectrometry was applied for GUS over the last decades. It was shown that GC-MS is very specific, sensitive and is able to build very large libraries of standardized spectra (3). A successful GC-MS procedure requires a volatile, thermally stable analyte. But GC-MS fails often to detect polar and thermally labile and non-volatile compounds. These requirements generally necessitate extraction from the biological matrix, followed in many cases by derivatization. These limitations led to investigate other possibilities for analyzing biological specimens (HPLC-DAD, LC-MS (-MS)) Single Mass Spectrometry A proposed alternative for a GUS is single MS spectra. The mass spectrometer works in the scan mode and applies in-source CID. The sample is screened at variable orifice voltages. The peaks exceeding a preset intensity are identified by comparison of their spectra with in-source CID spectrum libraries. The spectra obtained using in source- CID are mostly identically as those produced by conventional CID in the collision cell of a triple quadrupole mass spectrometry or of an ion trap. Marquet et al. described a procedure using full-scan from 100 to amu, with a step of 0.2 amu (38). In source CID was set at four continuously alternated voltages. In the positive mode CID was with low energy 20 ev, in high CID with high energy 80 ev, in the negative mode with low 20 ev, with high CID energy 80 ev. Several teams showed that fragmentation ensured reproducible in source CID spectra, at least using the same type of instruments (39). To obtain informative spectra (fragment ions and protonated molecule respectively molecular ion) pairs of full mass spectra were reconstructed by adding spectra at 20 ev and 80 ev on one hand and 20 ev and 80 ev on the other. These spectra were compared with spectra in the library. One library was built in the positive mode 24

25 Introduction containing reconstructed mass spectra; the other library was built in the negative mode containing 500 reconstructed spectra. Software was developed in co-operation with the manufacturer to automatically reconstruct such spectra and compare each positive and negative spectrum, together with their relative retention time to those in the library. This LC-ES-MS technique was compared to GC-MS and HPLC-DAD GUS procedures for the identification of unknown compounds. 75% of the spiked compounds in these samples were identified by this method versus 66% for GC-MS and 71% for HPLC-DAD. The conclusion was that it was complementary to GC-MS and HPLC-DAD and helped enlarge the range of drugs detected (40). Gergov et al. also built such a library of reconstructed CID-MS spectra using a triplequadrupole mass spectrometry. They obtained the reconstructed spectra by the addition of spectra recorded at 25 ev and 90 ev (39). The authors concluded that the later mentioned MS/MS library showed better results. MS/MS spectra showed no interference caused by either co-elution of β-blocking drugs or matrix material as a result of its superior selectivity compared with single MS-CID. Better match indices were obtained. Weinmann et al. built a spectra library of 400 drugs using a triple quadrupole mass spectrometry (41). In source CID was applied at 20, 50, and 80 V only in the positive mode. No reconstructed spectra were obtained. Each spectrum obtained was compared to the library. This procedure cannot be called GUS because selective extraction procedures were applied and when interfering compounds or background noise occurred, they used MS/MS fragmentation. In summary, in-source CID mass spectra showed that they were reproducible with the same type of instruments (42). The main drawback of these techniques is the low specificity and selectivity, as the spectra recorded do not result from a single parent ion. To obtain good mass spectra of signals against a high background is almost impossible. This technique requires a completely separation of drugs and matrix compounds, which is not achievable. 25

26 Introduction Reports give hints of poor interinstrument reproducibility, which would require the development of mass spectral libraries for each brand. A standardization of the crucial parts of the instruments with respect to in-source CID would be necessary. This is one explanation of the absence of any large, commercial in-source CID mass spectral library so far (43) Tandem Mass Spectrometry Weinmann et al. built an MS-MS library of more than 500 therapeutic or illicit drugs using LC-ES-triple-quadrupole spectrometry (41). One or a few parent ions were selected in the first quadrupole (generally the protonated molecules in the positive ionization mode or the molecular ions in the negative mode). These ions were fragmented in the collision cell. The fragments were analyzed in the third quadrupole in the scan mode. Four different collision energies (20, 30, 40, 50 ev) were applied resulting in four different positive product-ion spectra. These spectra were recorded in the library for each compound. Gergov et al. developed a library of MS-MS spectra for almost 400 therapeutic or illicit drugs (39). Generally collision energy of 35 ev was applied to obtain MS-MS spectra. Additional spectra were acquired at 20 or 50 ev for those compounds giving no informative spectrum at 35 ev. In the first step of the operation, samples were prescreened using LC-MS-SIM. Any matches to a previously defined list of retention times and [M+H] + were recorded. From this match a multi-period LC-MS-MS-CID product ion method was automatically created, that means these ions were selected as parent ions for product-ion scanning and library searching. This is a rather complicated process for a β-blocking screening, when direct product ion-scanning could have been applied. This method described a preliminary version of a datadependent acquisition. The same authors found a high reproducability of MS-MS CID-spectra between the instruments (44). They were using three different triple quadrupole instruments from the same manufacturer. Their results suggested good interinstrument fragmentation reproducability. 26

27 Introduction Baumann et al built a MS-MS library of more than 500 spectra using an ion-trap mass spectrometer in the positive mode (45). To obtain rich product ion spectra, resonance excitation at 20 amu below the parent ion selected was used, to further dissociate the [M+H-H 2 O] + ions. A mass dependent correction was automatically applied to the collision energy, because generally fragmentation energy decreases linearly when mass increases. Specific MS-MS spectra of different substances could be recorded. The described libraries and methods are theoretically interesting but of little use in clinical toxicology. These procedures cannot be used as GUS methods, because samples are monitored only for a previously selected limited number of compounds. A solution to this limitation could be data-dependent acquisition Data Dependent Acquisition The data dependent acquisition method was first used in a preliminary GUS procedure by Decaestecker et al (46). They used a time-of-flight detector in the positive electrospray mode. The quadrupole initially transmitted all masses (50 to 450). If an ion reached a predefined threshold the quadrupole selectively transmitted these highintensity ions to the collision cell operating with single fragmentation energy. The resulting fragments were analyzed in the TOF detector. The instrument switched back to the initial conditions. A refractory period was applied to the last selected ions for 2 minutes. Fitzgerald et al used also the technique of data-dependent acquisition (47). They modified the commercially available column-switching instrument, the REMEDi HS from BioRad Diagnostics, to make it compatible with atmospheric pressure ionization. Urine samples were injected directly on the column switching system. The compounds were analyzed in the full scan mode between 50 and 500 amu. When any ion exceeded a preset threshold, they were selected and fragmented by collision-induced dissociation in the ion trap. The major advantage of these techniques is their high specificity and selectivity, as the spectra recorded come from a single parent ion. The main limitations of these works are that they do not explain the procedure to detect unknown substances. Also they did not mention the LOD of any substances. 27

28 Introduction In a further paper Decaestecker specified the different LOD of some drugs but still did not explain the mechanism of drug finding. They described that setting a low switching intensity threshold will increase the number of interferences detected. As a consequence, the interpretation of the data set was much more complicated. And no automated data interpretation was described. Marquet et al described a preliminary procedure for the screening of drugs using a quadrupole-linear ion-trap mass spectrometer (48). For the first detection step (so called survey scan ) an enhanced MS mode was used. Ions were accumulated and then filtered in the Q3-linear ion-trap. MS-MS mode for on-the-fly fragmentation of the ions above the user-defined data-dependent acquisition was applied. Four insource collision-induced fragmentation conditions are altered, low and high fragmentation in the positive and negative modes (20, 50, -15 and 40 V, respectively). Positive and negative reconstructed spectra were obtained by adding two spectra on the other. Better signal-to noise ratio was obtained with this method compared to the in source CID-MS technique of the same group; because the reconstructed mass spectra decreased the influence of contaminant ions Sample Preparation in a General Unknown Screening Sample preparation is a key step in a screening procedure because the analytical system can not detect a compound if it is not extracted and so not injected in the system. The ideal extraction procedure would isolate the compounds of interest and eliminate the biological matrix compounds. The critical point is the signal-tobackground noise ratio. The signal-to-background noise is determined by the height of the chromatographic peaks and the baseline (purity of the extracts). Decaestecker et al tested an entire series of SPE sorbents for a LC-MS-MS general unknown screening (49). The sorbents were divided in three categories: apolar, mixedmode and polymeric. The extraction procedure was formulated individual for each column. Only neutral and basic compounds were chosen. From the apolar sorbents (Isolute C 2, C 4, C 8, C 18, C 18 MF, PH and CN) the C 8 packing material demonstrated the best overall extraction yield. MCX gave the best extraction 28

29 Introduction yields from the type from the mixed-mode category (MCX, HCX, HCX3, HCX5) and was slightly better than HLB from the polymeric category. Maurer et al proposed two extraction procedures for a GUS one with LLE and the other with SPE (50-52). In the LLE procedure method the sample was extracted at ph 7 and at ph 12 with a mixture of diethyl ether/ethyl acetate (1:1, v/v). The organic phases were evaporated to dryness. The combined residues were dissolved in organic solution. In the SPE method samples were worked up with HCX columns. The compounds of interest were eluted with methanol/aqueous ammonia (98:2, v/v). A drawback of this extraction procedure is that the methods are not suitable for all drugs. For example benzodiazepines could not be extracted in a sufficient way with the SPE procedure. Venisse et al applied different extraction techniques using Extrelut, Oasis HLB and MCX cartridges. Both types of Oasis columns resulted in higher background noise, but the extraction and detection of tested compounds was much better than with Extrelut columns (53). In summary, only liquid-liquid extraction and solid phase extraction method were used. Liquid-liquid extraction requires two procedures, one for acidic and one for basic drugs. SPE was performed with mixed-mode cartridges, classic hydrophobic bonded phase or a polymer Chromatic Conditions in a General Unknown Screening Most commonly used stationary phases were reversed-phase C 18 columns. A short guard column with the same type of stationary phase is recommended. In LC-MS the mobile phases consisted of mixtures of volatile buffers (e.g. ammonium acetate or formate) with acidic ph and organic modifiers such as methanol and acetonitrile. In HPLC-DAD the mobile phase consisted of phosphate buffer with acidic ph and acetonitrile or methanol. Gradient elution was performed. 29

30 Introduction 3.6. RESEARCH OBJECTIVE Acute poisonings account for 10 to 30% of the admissions to non-specialized care units, and 7 to 15% of autopsies performed in forensic institutions (4). Often, an intoxicated patient entering the hospital is consciousness or the patient or his relatives give misleading information. Consequently, the development of a reliable technique for general-unknown screening of pharmaceutical, toxicants and drugs of abuse in biological fluids is necessary. Such screening methods should be simple, rapid, reproducible, and available to cover a broad spectrum of toxicologically relevant substances, in one analytical run, with sufficient specifity. The combinations of mass spectrometry with suitable chromatographic procedures are the methods of choice, because they are very sensitive, precise, specific, universal and if coupled to automated extraction system very fast. It is necessary to speed up the process of a STA. The results should be given automatically. The aim of the project was to develop a fully-automated LC-MS method in combination with a HPLC-DAD method to detect a wide range of toxicologically relevant compounds. 30

31 Materials and Methods 4. MATERIALS AND METHODS 4.1. GENERAL UNKNOWN SCREENING WITH HPLC Materials Test substances obtained from various pharmaceutical companies were of pharmaceutical purity. Organic solvents and reagents were of analytical grade. Acetonitrile and methanol were purchased from Merck (Darmstadt, Germany). Solidphase extractions (SPE) were carried out on Oasis HLB cartridges (Waters, Rupperswil, Switzerland). Deionised water was generated with a Milli-Q water purification system from Millipore (Kloten, Switzerland) Sample and Buffer Preparation Procedures Separate stock solutions were prepared in methanol-water (1:1, v/v) at a concentration of 100 µg/ml. Standards in drug free serum (Biorad, Reinach, Switzerland) were prepared by spiking stock solutions of drug mixtures to make concentrations ranging from to 5 µg/ml, resulting in a set of standards with the following concentrations: 0.250, 0.500, 1.000, and µg/ml. The buffer (ph 2.2) was prepared with a 20 mmol/l KH 2 PO 4 solution, adjusted to the desired ph by appropriate addition of orthophosphoric acid Extraction Procedure One ml of serum was acidified by addition of 20 µl phosphoric acid 85%. The solidphase extraction (SPE) procedure was carried out on Oasis HLB cartridges. The Oasis HLB cartridges were conditioned initially with methanol 1 ml, followed by 1 ml Millipore water. 1 ml per serum sample was loaded onto the cartridges. Cartridges were subsequently washed with 1 ml Millipore water and dripped dry under charging pressure. The compounds of interest were eluted with 1 ml methanol. The extracts were evaporated to dryness under a stream of nitrogen and then reconstituted in 1 ml of a 10% acetonitrile solution for the general unknown screening. For the quantification methods, the extract was solved in the corresponding mobile phase. 31

32 Materials and Methods HPLC-DAD Method A Varian Pro Star HPLC system was used for this analysis. It consisted of a Pro Star 230 solvent delivery module (SDM), a Pro Star 330 PDA Detector, and a Star chromatography workstation system with software version The analytical column used was a LiChrospher 60 RP-select B (5 µm, 125 x 4 mm) protected by a LiChrospher 60 RP-select B (5 µm, 4 x 4 mm) both obtained from Merck (Darmstadt, Germany). A HPLC method for the GUS and methods for the quantification of specific substance groups were developed. The gradient and the running time was method dependent. The HPLC was performed with a gradient of acetonitrile in ph 2.2, 20 mmol/l potassium phosphate buffer (for the GUS: 5% ACN for 1 min, increased linearly to 50% in 14 min, increased linearly to 90% in 3 min, maintained for 4 min) delivered at a flow rate of 1.5 ml/min. The method for the GUS was 22 minutes. The injection volume was 150 µl in a loop of 20 µl. Detection was performed scanning the nm wavelength range. In-house libraries of spectra of drugs, toxic compounds and metabolites were used for compound identification. Table I shows the different quantification methods including the running time and the gradient Extraction Recovery Yields Drug standard prepared in serum were extracted as previously described. For calculation of the recovery yields, the peak area of the compound obtained after the extraction was compared with those of standards prepared in methanol and diluted with the mobile phase to the appropriate concentrations Identification and Semi-Quantification At the completion of the chromatographic procedure, peaks were detected for drugs. Peaks were identified based on their relative retention times. The purity and identity of each peak was assessed by examination of the UV spectra of the peak in comparison to library entries of drugs with similar RTs. Concentrations of drugs were calculated by comparing the peak area of the analyte and the corresponding standard containing a known amount of the analyte. 32

33 Materials and Methods Quantification Quantification methods were developed to determine the extraction recoveries. Table I shows the different quantification methods using all the same mobile phase and the same column described above. Table I. Quantification Methods by HPLC for more than 100 Substances. substance group method time (min) phosphate buffer (%) ACN (%) Anaesthetics Anaesthetics Analgesics Analgesics Analgesics Antidepressants Antidepressants Antidiabetics Antidiabetics Antiretroviral Antiretroviral Barbiturates Barbiturats Benzodiazepines Benzodiazepines Betablocking Betablocking Coumarines Coumarines Coumarines Diuretics Diuretics Diuretics Drugs of Abuse Drugsofabuse Laxatives Laxatives Neuroleptics Neuroleptics Neuroleptics Stimulants Stimulans THC THC

34 Materials and Methods 4.2. GENERAL UNKNOWN SCREENING WITH MS Materials Test substances obtained from various pharmaceutical companies were of pharmaceutical purity. Organic solvents and reagents were of analytical grade. Acetonitrile and methanol were purchased from Merck (Darmstadt, Germany), ammonium formate from Aldrich (Steinheim, Germany) and formic acid from Fluka (Buchs, Switzerland). Deionised water was generated with a Milli-Q water purification system from Millipore (Kloten, Switzerland). HySphere Resin GP cartridges were purchased from Spark Holland (Emmen, Netherlands). The Prospekt 2 TM unit from Spark Holland (Emmen, Netherlands) consists of an automatic cartridge-exchange module, dual cartridge clamps, solvent delivery unit (SDU) (including a high-pressure dispenser (HPD) and solvents). The solvent delivery unit is used to condition and wash the cartridges with solvents and to transfer the contents of the sample loop of the austosampler into the cartridge. The high-pressure dispenser provides the SPE solvents Sample and Buffer Preparation Procedures Separate stock solutions were prepared in methanol-water (1:1, v/v) at a concentration of 100 µg/ml. Serum standards were prepared by spiking with stock solutions of drug mixtures to make concentrations ranging from to 4 µg/ml, resulting in a set of standards with the following concentrations: 0.005, 0.010, 0.025, 0.050, 0.100, 0.250, 0.500, 1.000, and µg/ml. d 3 -Benzoylecgonine was prepared as internal standard (IS) at a concentration of 5 µg/ml. The ph 3.0 buffer was prepared with a 10 mmol/l ammonium formate solution, adjusted to the desired ph by appropriate addition of concentrated formic acid Extraction Procedure One ml of serum was acidified by addition of 20 µl concentrated formic acid and 100 µl of internal standard solution were pipetted into each sample. 34

35 Materials and Methods On-line SPE and elution was performed using the Prospekt 2 system. The HySphere Resin GP cartridge was conditioned with 1 ml methanol (5 ml/min) and with 1 ml water (5 ml/min). The injection volume was 290 µl in a loop of 100 µl. This 100 µl of the serum mixture was loaded on the cartridge. The sorbent was washed with 1 ml water (2 ml/min), and eluted with the mobile phase over 15 minutes. Cartridges were used only one time in order to avoid contamination Chromatographic and Mass Spectral Conditions The chromatographic system consisted of a Rheos 2000 Micro HPLC pump (ThermoFinnigan, Allschwil, Switzerland) and a Midas Symbiosis Autosampler from Spark (Emmen, Netherlands) equipped with a 100 µl loop. A four-channel degasser has been integrated into the Rheos CPS LC System. The LC-MS-MS apparatus was a LCQ Advantage MAX from Thermo Finnigan (Allschwil, Switzerland) equipped with an APCI device operating in the positive and in the negative detection mode. The chromatographic separation was performed on a CC Nucleodur C18 Gravity 3 µm (4 x 125 mm) with an integrated guard column 3 µm (4 x 8 mm) from Macherey- Nagel (Oensingen, Switzerland). The mobile phase was delivered at a flow rate of 400 µl/min. Each chromatographic run was performed with a binary, linear A/B gradient (Solvent A was 10 mmol/l ammonium formate, ph 3.0. Solvent B was 90% acetonitrile, 10% 10 mmol/l ammonium formate, ph 3.0.). The program was as follows: 0-1 min, 6% B; 1-8 min, 6 to 100% B; 8-20 min 100% B; min column equilibration with 6% B. The solvents were degassed. A data-dependent acquisition was used, generating a full-scan between 80 and 750 amu in the first mode. In the second mode a MS-MS spectrum of the most intense ion of the previous full-scan was performed. Only this high-intensity ion was kept in the ion trap where normalized collision energy is switched on and the resulting fragments were analysed. In the positive mode, normalized collision energy was 40.0%, in the negative mode 35.0%. The mass spectrometer reverted back to the full-scan mode. Dynamic exclusion was enabled. Thirty seconds refractory period was applied to the last selected ion. 35

36 Materials and Methods The following APCI inlet conditions were applied. The heated vaporizer was kept at 465 C. Both the sheath gas and the auxiliary gas were nitrogen set at 60 and 15 relative units, respectively. The capillary entrance to the ion trap was at an offset of 28 V in the positive mode, -4 V in the negative mode and was maintained at 220 C. The corona current was 5 µa. Table II shows the data dependent and global data dependent settings. Table II. Data Dependent and Global Data Dependent Settings. Data Dependent Settings Default Charge State 1 Default Isolation Width 4.0 Normalized Collision Energy (%) 40.0 resp Min. Signal Required Global Data Dependent Settings Exclusion Mass Width 0.5 Reject Mass Width 1.0 Dynamic Exclusion enabled Repeat Count 1 Repeat Duration 0.5 Exclusion List Size 25 Exclusion Duration 0.5 Exclusion Mass Width Mass Spectral Library Building Conditions Standard solutions were prepared in methanol-water (1:1, v/v) at a concentration of 1-2 µg/ml. Two mass spectral libraries were created, one for each ionization mode, by injecting 20 µl of these solutions directly without HPLC separation into the MS system. The MS-MS spectra were obtained at 40.0% normalized collision energy in the positive mode and 35.0% normalized collision energy in the negative mode. The obtained MS-MS spectra were added to the library. Relative retention data were acquired by actual LC-MS (-MS) analysis running a mixture of each compound with the internal standard. 36

37 Materials and Methods A library was constructed to include per spectra, the name of the compound, the molecular formula, and its protonated molecular ion, together with the relative retention time. The mass spectral library comprises a total of more than 400 spectra of more than 350 compounds. For a semi-quantitative procedure peak areas of the compounds (at a concentration of 1 µg/ml) were recorded and attached to the library. Obtained MS-MS data from a chromatographic run were compared to the MS-MS library using the NIST Mass Spectral Program 2.0 from ThermoFinnigan. A computer program (XcLibraryScreening) was created to automate the searching process and to include the RRT and the molecular ion in the identification of unknown compounds. 37

38 Results and Discussion 5. RESULTS AND DISCUSSION 5.1. SAMPLE PREPARATION AND CHROMATIC CONDITIONS FOR A GUS The sample preparation must be suitable for various compounds in a large polarity range (acidic, neutral and basic). Thus, it requires a nonclass-specific extraction procedure able to isolate the widest range of relevant molecules from biological matrices. Such procedures may involve a single liquid-liquid extraction, two liquidliquid extraction in parallel (one for acidic and neutral and one for basic compounds) (51, 54-56), solid-phase extraction with classical hydrophobic phase (49, 57), or mixedmode phase SPE (49, 57-60). SPE was chosen as extraction technique. The SPE mechanism is more selective than liquid-liquid extraction. SPE has a number of advantages over liquid-liquid extractions such as cleaner extracts, increased selectivity for the compounds of interest, and smaller volumes of solvent may be used for extraction. SPE takes less time, and the procedure can be automated (11). A polymer-based sorbent was selected because it is stable over a larger ph range and does not have secondary (unbounded hydroxy silanol) groups compared to classical hydrophobic phase. A mixed-mode phase involves both reversed-phase retention and cation-exchange properties. The column elution requires two steps. First the acidic and neutral substances were eluted with methanol. In the second step the basic substances were eluted with a methanol/nh 4 OH mixture. The acid-neutral and the basic fractions were mixed before injection in the HPLC-DAD system in order to save time. Absolute recoveries were measured for the polymer (Oasis HLB) and the mixed-modebased sorbent (Oasis MCX). The Oasis HLB column showed the better extraction recoveries for most of the tested substances than the Oasis MCX column. 38

39 Results and Discussion Sample Preparation and Chromatic Conditions with HPLC Before loading a sample on the cartridge, 20 µl of orthophoshoric acid (85%) (or 20 µl of formic acid in the LC-MS method) was added to the sample. The addition of acids influenced the protein-binding. The ph was lowered resulting in a disruption of the drug-protein interaction yielding better recoveries. Above all, highly-protein bound drugs showed a much better recovery. For example the extraction recovery for torasemide was improved from 49% to 86%. For each compound of the library, absolute recovery was measured at least two times by extracting and assaying drug-free serum spiked with the drugs tested, then comparing the peaks areas of these extracted standards with those of methanolic standards at the same concentration. Of the tested drugs, 94 of 100 were successfully extracted. A successful extraction is defined as a recovery of more than 50%. The extraction of more than 50% of the tested drugs results in a recovery of more than 90%. Fig. 2 shows the absolute recovery of the 100 tested compounds. This method proved to extract a large range of drugs. Acid, neutrals as well as basic drugs were extracted in a satisfying way. Only 6 of the hundred tested drugs showed an extraction recovery of less than 50%. These drugs are bromadiolone (extraction recovery 31%), chlorophacinone (8%), delta-8-thc (34%), delta-9-thc (47%), emodine (40%) and rhein (19%). They are all hydrophobic and mostly of them acidic substances. 1 ml of methanol was not sufficient to elute these compounds in a satisfying way ( 50%) from the column. Improvement of the extraction recovery for these substances could be achieved by using a higher volume of methanol, a stronger solvent than methanol or another column. Acidic substances were not charged at the present ph because of the addition of the ortho-phosphoric acid. Also a change of the ph could improve the extraction recovery. The extraction also proved to be reproducible. Table III shows the extraction recovery and coefficient of variation of the 100 tested compounds. The coefficient of variation was always less than 15% for all of the compounds with an extraction recovery of 39

40 Results and Discussion more than 50%. The only exception was secobarbital showing a coefficient of variation of 16.2% number of compounds > <20 absolute % recovery Fig. 2. Absolute % recovery of the 100 tested compounds with HLB Oasis columns. Gradient studies were performed to achieve a fast elution ramp without losing too much sensitivity and resolution. It was decided that a 5 to 50% acetonitrile gradient over 14 minutes, and a 50 to 90% acetonitrile gradient over 3 minutes would be suitable for this method. A slow gradient was chosen for the first gradient because most of the compounds were eluted in a short time from the column with 50% of acetonitrile. These compounds showed a retention time of less than 15 minutes. The other substances were eluted with the fast 50-90% acetonitrile gradient. Analytical results can be interpreted after 22 minutes. This is an acceptable turnaround time for the analysis of emergency drug screening results. In Table III the retention time of the 100 substances are listed. 40

41 Results and Discussion Table III. LOD, Recoveries of Drugs and their Coefficient of Variation Using the HLB Column and Recorded Retention Time of 100 Compounds. Compounds LOD (ng/ml) Recovery (%) CV (%) Retention time (min) 6-acetylmorphine acebutolol acenocoumarol acetaminophen acetazolamide acetylsalicylic acid aloeemodine alprazolam alprenolol amiloride amitryptiline amobarbital amprenavir atenolol benzoylecgonine bisacodyl brallobarbital bromadiolone bupivacaine butalbital cannabidiol cannabinol canrenone carbromal chlordiazepoxide chlorophacinone chlorthalidone cinchocaine citalopram cocaethylene cocaine codeine coumachlor coumaphos coumatetralyl crimidine delta-8-thc delta-9-thc diazepam diclofenac dihydrocodeine efavirenz emodine ephedrine flupenthixol <ER <ER >

42 Results and Discussion Table III (continued). LOD, Recoveries of Drugs and their Coefficient of Variation Using the HLB Column and Recorded Retention Time of 100 Compounds. Compounds LOD (ng/ml) % Recovery CV (%) Retention time (min) furosemide glibenclamide glibornuride gliclazide hydrochlorothiazide hydrocodone hydromorphone ibuprofen imipramine indinavir levomepromazine lidocaine lopinavir lorazepam mefenamic acid mepivacaine methylphenidate metoprolol morphine nadolol nalbuphine naproxen nelfinavir nevirapine norcodeine olanzapine oxprenolol pentobarbital phenobarbital phenolphthalein phenprocoumon pindolol propranolol propyphenazone pseudoephedrine quetiapine rhein ritalinic acid ritonavir salicylic acid saquinavir secobarbital sertraline sotalol spironolactone <ER >

43 Results and Discussion Table III (continued). LOD, Recoveries of Drugs and their Coefficient of Variation Using the HLB Column and Recorded Retention Time of 100 Compounds. Compounds LOD (ng/ml) % Recovery CV (%) Retention time (min) thiopental thioridazine timolol tolbutamide torasemide trimipramine tubocurarine venlafaxine warfarin zuclopenthixol Sample Preparation and Chromatographic Conditions with LC-MS On-line SPE was chosen as an extraction technique because this procedure is universal, rapid and can be automated. This method is becoming popular in bioanalytical analysis (61). The Prospekt solid phase extraction can be linked to the APCI ion source of the LC-MS-MS instrument (62). The system couples and automates sample extraction and instrumental analysis. Benefits of this technology include improved precision of all extraction steps. This method has a time saving advantage compared to other techniques because vaporizing the sample extract is not necessary. In addition, the procedure presented in this study extracts acidic, neutral as well as basic drugs. The chromatographic conditions (column and mobile phase) must be chosen in an appropriate way. The most polar must be retained and the most lipophilic compounds must be eluted. The mobile phases must be compatible with API sources. Most of the HPLC-DAD applications use non-volatile buffers such as phosphate and borate buffers. These non-volatile buffers cause problems in combination with MS and should be avoided. They are blocking the capillary in the probe, and are causing salt buildup on the spray head and thus compromise the integrity of the spray. Volatile buffers like ammonium acetate, ammonium formate, acetic acid and ammonium carbonate buffers can be used instead. Ammonium formate was chosen because of its 43

44 Results and Discussion suitability as a buffer at a ph of 3. Ammonium is necessary for a better elution of the compounds in combination with the Prospekt and often avoids the tailing of peaks. An appropriate column has to be chosen in combination with the Prospekt. The column has to retain all the compounds stronger than the extraction column of the Prospekt to obtain a satisfying chromatography. A C 18 Nucleodur Gravity column was chosen because of its capability to be linked with the Prospekt, its stability within a large ph range and its robustness. The separation of the drugs was carried out under acidic conditions (ph = 3) in order to limit secondary interaction on the free silanol groups of the C 18 column. The first peak eluted at 5.9 minutes (morphine) and the last at 18.4 minutes (delta-8-thc). The absolute retention times are shown in Table IV. Separation of substances is necessary to detect low-level analytes. Ionization of mobile phase components (acetonitrile, ammonium formate) and endogenous compounds is the main source of background noise. Contamination of the mass spectra by these compounds potentially hampers the identification of low concentrations of analytes. The screening of a sample can be performed in less than one hour with a chromatographic run taking 23 minutes for each mode (positive and negative), which is an acceptable analytical time for a GUS (including library search and interpretation). APCI and ESI were compared with each other in order to examine with which ion source the higher signals were detected. More than hundred compounds were investigated in aqueous solutions. In summary, higher signals were detected with APCI for hydrophobic substances like cannabinoids or coumarines and acidic drugs. Better results were obtained with ESI for basic drugs such as neuroleptics and antidepressants. Basic drugs are positively charged at a ph of 3.0 (ph of the mobile phase). These results are not astonishing. The relation of the molecular mass range and the polarity of analytes that can be analyzed by GC-MS, APCI LC-MS and ESI LC-MS techniques are sketched in Fig. 3. APCI allows sensitive determination of analytes with moderate polarity and molecular mass. ESI enables the better determination of 44

45 Results and Discussion analytes of high molecular mass (up to several hundred units such as peptides or proteins) and high polarity. Table IV. Recorded LC Retention Times of 87 Compounds. Compounds morphine amiloride atenolol hydromorphone sotalol codeine dihydrocodeine norcodeine 6-acetylmorphine acetaminophen olanzapine hydrocodone pseudoephedrine ephedrine nadolol tubocurarine nalbuphine benzoylecgonine acetazolamide ritalinic acid pindolol mepivacaine acebutolol timolol lidocaine metoprolol cocaine oxprenolol hydrochlorothiazide venlafaxine cocaethylene bupivacaine propranolol alprenolol chlorthalidone quetiapine indinavir torasemide citalopram levomepromazine nevirapine chlordiazepoxide flupenthixol cinchocaine Retention time (min) Compounds imipramine phenobarbital amitryptiline canrenone trimipramine brallobarbital nelfinavir sertraline zuclopenthixol saquinavir thioridazine crimidine furosemide phenolphthalein alprazolam lorazepam propyphenazone amprenavir tolbutamide rhein aloeemodine acenocoumarol gliclazide warfarin glibornuride bisacodyl glibenclamide ritonavir diazepam lopinavir phenprocoumon diclofenac coumachlor coumatetralyl efavirenz emodine mefenamic acid bromadiolone chlorophacinone cannabidiol cannabinol delta-9-thc delta-8-thc Retention time (min)

46 Results and Discussion Electrospray Molecular mass APCI GC-MS Analyte Polarity Fig. 3. Relation of the molecular mass range and the polarity of analytes analyzable by GC-MS and LC-MS interface techniques (APCI and ESI). This was confirmed by Bogusz et al. They showed that between these two ionization sources, atmospheric pressure chemical ionization appeared more universal and assured generally higher sensitivity. Only in the case of very polar drugs (e.g. psilocin or psilocybin) electrospray ionization was more sensitive (63). For a GUS it is not necessary to detect high molecular mass. Most of drugs have masses under 500 amu and are moderately polar. But at an acidic ph (required in this method) a lot of compounds are positively charged and become more polar. None of the two ion sources showed a large benefit in this examination. APCI was preferred to electrospray ionization in order to reduce ion suppression. This phenomenon affects the amount of charged ion in the gas phase that reaches the detector. Although ion suppression can have effects on both electrospray ionization and atmospheric pressure chemical ionization, evidence indicates that the electrospray interface is more impacted. Experiments have shown that with the same extraction procedure and method ESI was more affected by ion suppression than APCI. Serums or aqueous solutions were spiked with the same amount of a compound. Ionization with ESI resulted in a large variability between the matrix specimen and the aqueous solution. The signal of the analyte in serum was mostly decreased. The extent of the 46

47 Results and Discussion ion suppression is not being predictable. In a GUS ion suppression is responsible for higher LODs. In the worst case ion suppression can result in false negative results LIBRARY SCREENING AND QUANTIFICATION Identification with HPLC The present method allows easy and rapid identification of 100 substances belonging to different pharmacological classes. The LOD of all the substances are listed in Table III. UV spectra and retention time databases were established by spiking aqueous with pure drugs for the identification of drugs extracted from serum. Bogusz et al showed that the co-extracted biological matrix did not exert any influence on the chromatographic behaviour of drugs analyzed by means of gradient HPLC (28). Serum samples spiked with decreasing concentrations of the tested drugs were analyzed in order to determine LODs. Each concentration was extracted and analysed two times. The LOD was set at the lowest concentration where the signal of the compound was three times higher than the background noise and the spectral similarity was above the cut-off (> 0.9). Identification of drugs with diode array detection and a spectrum library are affected by the presence of co-extracted matrix substance. These compounds can co-elute at the same time as toxicologically relevant substances and overlap the corresponding UV spectra. The purity of the extract is therefore important for a successful HPLC-DAD identification of a drug even at small concentrations. In order to increase the specificity of the method, each drug of the library is also characterized by its retention time. The retention time has to be in a time window of ± 0.5 min. Identification errors require coincidence of both RTs and UV spectra. 47

48 Results and Discussion The LOD was set at the lowest concentrations where both samples fulfilled the mentioned requirements (spectra similarity and retention time). The limit of detection (LOD) for the majority of the tested drugs (76%) was ng/ml. Shifts of the retention time of compounds are characteristics of changes in the molarity or ph of the mobile phase or ageing of the column. Also two columns of the same brand can cause a shift of the retention time. The use of retention times relative to an internal standard reduces the influence of the altering of the column or changing the column. Relative retention time is more reproducible than the retention time. Bogusz et al showed that relative retention time data can be used interlaboratory (64, 65). With this procedure the retention time can be corrected and is less affected by the mentioned influences A Semi-Quantitative Procedure with HPLC Not only the identification of compounds is important, also the determination of the concentration is crucial in a GUS to estimate the severity of the intoxication. A semiquantitative procedure was performed. Drug-free serum spiked with the drugs tested were extracted and analyzed. Peak areas of extracted serum sample were compared to peak areas of methanolic standards of the database considering the extraction recovery. Results are shown in Table IV. The measured concentration of 72 of totally 94 substances was within 20%. These results indicated the possibility to estimate the drug concentration with this procedure. Co-extracted matrix substances affect the quantitation of drugs eluting at the same time. This is one of the major causes of errors in chromatographic analysis. An erroneous result can be caused by a coeluting matrix substance that mimics the spectrum of a compound listed in Table V or interfere with the detection of a part of a peak. The basis of the assay was to provide a means to identify a drug or poison and an estimate of its concentration. More accurate quantitation was then conducted using the quantification method. 48

49 Results and Discussion 49 Table V. Extraction rate of spiked serum samples compared to methanolic standard concentrations. Drug Measured (µg/ml) Difference (%) Drug Measured (µg/ml) Difference (%) 6-acetylmorphine acebutolol acenocoumarol acetaminophen acetazolamide acetylsalicylic acid aloeemodine alprazolam alprenolol amiloride amitryptiline amobarbital amprenavir atenolol benzoylecgonine bisacodyl brallobarbital bupivacaine butalbital cannabidiol cannabinol canrenone carbromal chlordiazepoxide chlorthalidone cinchocaine citalopram cocaethylene cocaine codeine coumachlor coumaphos coumatetralyl crimidine diazepam diclofenac dihydrocodeine efavirenz ephedrine flupenthixol furosemide glibenclamide glibornuride gliclazide hydrochlorothiazide 0.99± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± % 10.9% 14.8% -16.3% -4.8% -41.9% 0.8% -46.5% -7.4% -29.0% 45.8% -54.3% 21.5% 2.5% 17.5% 44.3% -6.7% -19.8% -3.0% -46.4% -40.9% -1.2% 25.7% -7.8% 1.2% 5.4% -0.1% 2.2% -2.6% 14.3% -0.8% 8.9% -2.9% 0.8% -13.1% -14.3% -22.5% 3.9% -8.2% -16.0% 16.4% -0.5% -25.2% -10.4% 13.0% hydrocodone hydromorphone ibuprofen imipramine indinavir levomepromazine lidocaine lopinavir lorazepam mefenamic acid mepivacaine methylphenidate metoprolol morphine nadolol nalbuphine naproxen nelfinavir nevirapine norcodeine olanzapine oxprenolol pentobarbital phenobarbital phenolphthalein phenprocoumon pindolol propranolol propyphenazone pseudoephedrine quetiapine ritalinic acid ritonavir salicylic acid saquinavir secobarbital sertraline sotalol spironolactone thiopental thioridazine timolol tolbutamide torasemide trimipramine 0.70± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± % -2.4% 30.8% -9.8% -5.7% -4.1% -11.7% -21.3% 9.2% -28.9% 12.9% -3.5% -7.9% -10.3% 5.7% 12.2% 14.3% 12.7% -14.5% -10.9% -30.7% -0.3% 16.6% -13.7% -13.0% -5.1% -41.3% 4.3% -4.3% -19.2% 6.9% -20.7% -48.4% -14.6% -17.3% -14.4% 38.3% -7.2% -31.4% 16.7% 12.2% 3.2% -18.7% 8.8% -13.6%

50 Results and Discussion Table V (continued). Testing of the Semi-Quantitative General Unknown Screening procedure. Drug Measured (µg/ml) Difference (%) Drug Measured (µg/ml) Difference (%) tubocurarine venlafaxine 1.19± ± % 13.9% warfarin zuclopenthixol 0.80± ± % -12.3% Method Development MS-MS data results higher specificity and selectivity and more structural information for the identification of an unknown substance. First the unknown compound has to be detected. In the second step a product-ion scan has to be applied. This procedure is not compatibile for a GUS with the classic MRM mode. Mass spectra recorded with insource CID don t result from a single parent ion resulting in low specificity and selectivity. The method of choice was a data-dependent acquisition procedure. In the first step (so-called "survey scan") the MS-MS instrument is operated in the full-scan singlemass mode. In the second mode the most intense ion above a predefined threshold was selected. This ion was fragmented in the collision cell. The resulting fragments were analyzed in the scan mode. A refractory period was set to the last ion selected for 30 seconds. Less intense ions of compounds eluting at the same time would not be detected without a refractory period. A refractory period longer than 30 seconds can result in a loss of identification of a compound with the same molecular mass ion. With a shorter refractory period the method can fail to detect substances eluting at the same time. After a short time the instrument switched back to the initial conditions. Mass spectra data were established by spiking methanol-water (1:1; v/v) with pure drugs for the identification of drugs extracted from serum. Mass spectra were recorded in the negative as well as in the positive mode; because some substances can only be detected in one mode or they have different LOD for both modes. For example morphine was better detected with the positive mode, bromadiolone better detected in the negative mode. For both modes a library was created. In best cases, substances 50

51 Results and Discussion with positive and negative spectra can be identified in both modes with the respective libraries. So results from both libraries confirm each other in most cases. The MS-MS spectra were obtained at 40.0% normalized collision energy in the positive mode and 35.0% normalized collision energy in the negative mode. Examples of product ion mass spectra are shown in Fig. 4 and Fig. 5. The MS-MS spectrum of cocaine in the positive and the spectrum of saquinavir in the negative mode are presented. The normalized collision energies of 40.0% in the positive and 35.0% in the negative mode were chosen in order to obtain fragmentation of the compounds. A decrease of the normalized collision energy would yield less fragmentation. Applying higher normalized collision energy would result in much lower peak intensities of the fragments. Fig. 4. MS-MS spectrum of cocaine obtained in the positive mode with normalized collision energy of 40.0%. Fig. 5. MS-MS spectrum of saquinavir obtained in the negative mode with normalized collision energy of 35.0%. The relative retention time data were gathered by running a mixture of each compound with the internal standard by an actual LC-MS (-MS) analysis. 51

52 Results and Discussion A library was constructed to include per spectra, the name of the compound, the molecular formula, and its protonated molecular ion, together with its relative retention time. This mass spectral library comprises spectra of drugs and metabolites from a large diversity of substance classes. With this procedure acidic, neutral as well as basic drugs can be detected and identified Automatic Library Searching with LC-MS A small application program was developed for the automated identification of unknown compounds with LC-MS. In order to identify unknown compounds in a serum sample a chromatographic run was performed in each mode. In the next step the developed program compared each recorded MS-MS spectrum to those in the library from the Xcalibur software. With this program the number of best hits that the unknown spectrum should be compared to can be specified. In the procedure described in this study the ten best hits were chosen. In the presented example a run was loaded of a serum sample spiked with 5 substances (Fig. 6). The run was performed in the positive mode. Phenolphthalein, gliclazide, bisacodyl, glibornuride and glibenclamide were added to a serum sample at a concentration of 1 µg/ml. The similarity between the library spectra and the unknown spectra is characterized by the match factor and the reverse match factor. The match factor indicates the correlation between the unknown spectrum and the library spectrum (presence and relative intensities of mass-to-charge ratios). The reverse match factor indicates an inverse search. The presence and the relative intensity of the ions of the library spectrum are compared to those of the unknown spectrum. This parameter ignores the ions present in the unknown spectrum if absent in the reference spectrum. 52

53 Results and Discussion Fig. 6. All MS-MS spectra were loaded from a run. They were compared to the spectra in the library. The match factor and the inverse match factor range between 0 and with 0 indicating no similarity and indicating perfect similarity. In our procedure for both factors the threshold was set at 400. With this threshold parameter the best results were obtained. It is also possible to set another threshold. A higher threshold can result in a higher LOD and in a more specific result. A lower threshold can result in false negatives but also in a lower LOD. 53

54 Results and Discussion MS-MS spectrum using DDA Library MS-MS spectrum Fig. 7. Comparison of the positive MS-MS spectrum of glibornurid in a serum sample using DDA to the MS-MS spectrum of the library. Each product ion mass spectrum was subjected to an automated library searching routine compared to the library spectra. Fig. 7 shows the MS-MS spectrum of gliclazide from spiked serum (above in red colour) obtained with this procedure compared to the MS-MS spectrum of the library (below in blue colour). The match factor of the presented mass spectra was 889, the reverse match factor was 989. Compound identification took in account the mass-to-charge ratio of the unknown compound selected before fragmentation. This mass-to-charge ratio has to be within ± 2 m/z of the reference mass-to charge ratio recorded in the library. The width of the mass-to-charge ratio window was chosen to include the isotopes of the compound. The pseudomolecular ion (usually, protonated in the positive mode, deprotonated in the negative mode) and its fragments were obtained and compared to references in the library. Each MS-MS spectrum recorded is derived from one single mass-to-charge ratio (representing the most intense ion of the previous full scan). Other authors used collision-induced dissociation at different voltages to obtain the same information. Mass spectra were acquired by continuously switching between a low and a high orifice voltage throughout the run to obtain both protonated molecular ion (low-voltage scan) and mass spectral fragments (high voltage scan) from the CID in the ion source. The low-voltage spectrum was dominated by the pseudomolecular ion, whereas the high-voltage spectrum contained mass fragment ions. Mass spectra from the two different voltage scans were then summed to produce a mass spectrum 54

55 Results and Discussion for each compound in the library in order to maximize the information. With the procedure presented in this study it is not necessary to switch between different orifice voltages. RRT was also included in the identification procedure. The RRT of the unknown compound has to be within ± 5%. This large time-window was chosen because the MS-MS spectra can be obtained during the whole peak width and the refractory period was set at 30 s. The RT of the IS was registered in the positive mode with a value of approximately 6.9 min. Variations in RT occur when using different lots of columns with the same absorption material. Only if all the parameters were within the fixed areas a hit was reported. In summary the match factor and the reverse match factor had to be above 400, the mass-to-charge had to be ± 2 m/z and the RRT had to be within 5% (Fig. 8). Each MS-MS spectrum, which fulfilled these conditions, was reported. The report consists of the different hits with their names, together with the match factor and reverse match factor of the comparison with the library spectrum, relative retention time and mass-to-charge in this run and the one in the library. The new program automatically releases a report, which consists of the different hits with the substance names, together with the match factor, the reverse match factor, the RRTs and mass-to-charge relations compared to the ones in the library. 55

56 Results and Discussion Fig. 8. The library search parameters and the library can be specified. In this method the threshold of the match factors were set at 400, the RRT has to be within ± 5% and the maximum deviation of the mass-to-charge was two. It was allowed to return the ten best hits. All 5 substances in a concentration of 1 µg/ml were identified with our procedure. Fig. 9 shows the output generated by the LC-MS-MS and the software. Gliclazide, bisacodyl, glibornuride and glibenclamide all have similar retention times. However, this was a minor problem because a refractory period was applied enabling to identify the compounds even though they were not chromatographically separated. Importantly, the co-eluting substances did not affect the MS-MS spectra in the presented procedure in contrast to single MS CID methods. Therefore, the analysis of unknown compound is more rapid using MS-MS. Spectra obtained using MS with in source-cid are mostly identically as those produced by conventional MS-MS CID. But these MS spectra would be affected by co-eluting substances. 56

57 Results and Discussion Fig. 9. In this result file the different hits are presented with their substance name, match factor, reverse match factor, RRT and mass-to-charge relation compared to the corresponding parameters in the library. The same substance can be found several times with different MS-MS spectra. Serum samples spiked with decreasing concentrations of the tested drugs were analyzed in order to determine LODs. Each concentration from to 4 µg/ml was extracted and analyzed two times. The LOD was set at the lowest concentrations where both samples fulfilled the mentioned requirements to identify a compound and hits were reported. The compounds were listed as hits by the small application program. Out of 100 tested compounds, only 13 (amobarbital, acetylsalicylic acid, butalbital, carbromal, coumaphos, ibuprofen, methylphenidate, naproxen, pentobarbital, salicylic acid, secobarbital, spironolactone and thiopental) were not detectable with this LC-MS method. These compounds were identified neither at high therapeutic concentrations 57

Introduction to LC/MS/MS

Introduction to LC/MS/MS Trends in 2006 Introduction to LC/MS/MS By Crystal Holt, LC/MS Product Specialist, Varian Inc. Toxicology laboratories Increased use of LC/MS Excellent LD Cheaper (still expensive) Much more robust Solves

More information

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution A simple definition of a Mass Spectrometer A Mass Spectrometer is an analytical instrument that can separate charged molecules according

More information

Measuring Lipid Composition LC-MS/MS

Measuring Lipid Composition LC-MS/MS Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven

More information

Performance of an ultra low elution volume 96-well plate

Performance of an ultra low elution volume 96-well plate Performance of an ultra low elution volume 96-well plate Claude R. Mallet, Ziling Lu, Jeff R. Mazzeo, Uwe D. Neue Waters Corporation PittCon 2003 March 10-14 2003 Orlando, Florida Today s Challenges Faced

More information

Rapid and Robust Detection of THC and Its Metabolites in Blood

Rapid and Robust Detection of THC and Its Metabolites in Blood Rapid and Robust Detection of THC and Its Metabolites in Blood Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 95051 USA Abstract A robust method

More information

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample

More information

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions Simplifying Qual/Quan Analysis in Discovery DMPK using UPLC and Xevo TQ MS Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION The determination of the drug metabolism

More information

Sample Concentration and Analysis of Human Hormones in Drinking Water

Sample Concentration and Analysis of Human Hormones in Drinking Water Sample Concentration and Analysis of Human Hormones in Drinking Water Carl Fisher Applications Chemist Ion Chromatography/Sample Preparation Thermo Fisher Scientific March 1, 215 1 The world leader in

More information

Components of a Mass Spectrometer

Components of a Mass Spectrometer Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,

More information

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value Agilent 64 Triple Quadrupole LC/MS Sensitivity, Reliability, Value Sensitivity, Reliability, Value Whether you quantitate drug metabolites, measure herbicide levels in food, or determine contaminant levels

More information

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s) Ionization Methods Neutral species Charged species Removal/addition of electron(s) M + e - (M +. )* + 2e - electron ionization Removal/addition of proton(s) M + (Matrix)-H MH + + (Matrix) - chemical ionization

More information

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate

More information

Metabolomics: quantifying the phenotype

Metabolomics: quantifying the phenotype Metabolomics: quantifying the phenotype Metabolomics Promises Quantitative Phenotyping What can happen GENOME What appears to be happening Bioinformatics TRANSCRIPTOME What makes it happen PROTEOME Systems

More information

LECTURE 3. Ionization Techniques for Mass Spectrometry

LECTURE 3. Ionization Techniques for Mass Spectrometry LECTURE 3 Ionization Techniques for Mass Spectrometry Jack Henion, Ph.D. Emeritus Professor, Analytical Toxicology Cornell University Ithaca, NY 14850 Lecture 3, Page 1 Contents Electron ionization (EI)

More information

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF PUSHER TOF DETECTOR ZSPRAY TM Ion Source SAMPLING CONE SKIMMER RF HEXAPOLE RF HEXAPOLE QUADRUPOLE IN NARROW BANDPASS MODE

More information

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS FORENSICS AND TOXICOLOGY ANALYSIS DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS Solutions for Your Analytical Business Markets and Applications

More information

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS) Michael Rummel, Matthew Trass, Michael Campognone, and Sky Countryman Phenomenex, Inc., 411 Madrid Avenue, Torrance,

More information

LC/MS Method for Comprehensive Analysis of Plasma Lipids

LC/MS Method for Comprehensive Analysis of Plasma Lipids Application Note omics LC/MS Method for Comprehensive Analysis of Plasma s Authors Tomas Cajka and Oliver Fiehn West Coast Metabolomics Center, University of California Davis, 451 Health Sciences Drive,

More information

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No.

Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection. EPL-BAS Method No. Page 1 of 10 Determination of 6-Chloropicolinic Acid (6-CPA) in Crops by Liquid Chromatography with Tandem Mass Spectrometry Detection EPL-BAS Method No. 205G881B Method Summary: Residues of 6-CPA are

More information

More structural information with MS n

More structural information with MS n PRODUCT SPECIFICATIONS The LTQ XL linear ion trap mass spectrometer More structural information with MS n The LTQ XL linear ion trap mass spectrometer delivers more structural information faster and with

More information

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis A Fast and Sensitive LC/MS/MS Method for the Quantification and Confirmation of 3 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine Samples Cliquid Software for Routine LC/MS/MS Analysis

More information

Determination of Amantadine Residues in Chicken by LCMS-8040

Determination of Amantadine Residues in Chicken by LCMS-8040 Liquid Chromatography Mass Spectrometry Determination of Amantadine Residues in Chicken by LCMS-8040 A method for the determination of amantadine in chicken was established using Shimadzu Triple Quadrupole

More information

Supporting information

Supporting information Supporting information Figure legends Supplementary Table 1. Specific product ions obtained from fragmentation of lithium adducts in the positive ion mode comparing the different positional isomers of

More information

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS SCIEX QTRAP 4500 LC-MS/MS System and Spark Holland PICO Adrian M. Taylor 1, Peter Ringeling 2, Martin Sibum 2, Stefan Sturm

More information

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS TECHNICAL NOTE 21882 Robust extraction, separation, and quantitation of structural isomer steroids human plasma by SPE-UHPLC-MS/MS Authors Jon Bardsley 1, Kean Woodmansey 1, and Stacy Tremintin 2 1 Thermo

More information

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010 MASS SPECTROMETRY BASED METABOLOMICS Pavel Aronov ABRF2010 Metabolomics Research Group March 21, 2010 Types of Experiments in Metabolomics targeted non targeted Number of analyzed metabolites is limited

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc. SAMHSA-Compliant LC/MS/MS Analysis of 11-nor-9-carboxy-D 9 - Tetrahydrocannabinol in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120 Application Note Forensic Toxicology Authors Irina

More information

Dienes Derivatization MaxSpec Kit

Dienes Derivatization MaxSpec Kit Dienes Derivatization MaxSpec Kit Item No. 601510 www.caymanchem.com Customer Service 800.364.9897 Technical Support 888.526.5351 1180 E. Ellsworth Rd Ann Arbor, MI USA TABLE OF CONTENTS GENERAL INFORMATION

More information

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes Catalin E. Doneanu, Weibin Chen, and Jeffrey R. Mazzeo Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B E N E F

More information

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 9 Infinity II/63B LC/TOF system Application Brief Authors Mike Adams, Karen Kaikaris, and A. Roth CWC Labs Joan Stevens, Sue D

More information

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition Evelyn Goh Waters Pacific, Singapore APPLICATION BENEFITS This method allows for the simultaneous analysis of 12 water-soluble

More information

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS Method for the Determination of Tenofovir from Plasma LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract

More information

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS

Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Detection of Low Level of Chloramphenicol in Milk and Honey with MIP SPE and LC-MS-MS Olga Shimelis, An Trinh, and Michael Ye Supelco, Div. of Sigma-Aldrich, Bellefonte, PA T407125 Introduction Molecularly

More information

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid Joanne ( 乔安妮 ) Mather Senior Scientist Waters Corporation Data courtesy of Erin Chambers and Mary

More information

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA

More information

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer Robert Plumb, Michael D. Jones, and Marian Twohig Waters Corporation, Milford, MA, USA INTRODUCTION The detection and characterization of impurities and degradation products of an active pharmaceutical

More information

Agilent s LC/MS Portfolio and Applications Examples

Agilent s LC/MS Portfolio and Applications Examples Agilent s LC/MS Portfolio and Applications Examples Sean Orlowicz LC/MS Product Specialist Agilent Technologies 31August Agilent s > 40 year heritage in mass spectrometry 1971, 5930A GC/MS 1971-2011 2005,

More information

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes Eric S. Grumbach,, Diane M. Diehl and Jeffrey R. Mazzeo 2003 Waters Corp. Outline Introduction HILIC Definition

More information

Analyzing Phenyl Ureas and Carbamates with HPLC/API-MS. Presented by Chen-Kai Meng, Ph.D. Applications Chemist

Analyzing Phenyl Ureas and Carbamates with HPLC/API-MS. Presented by Chen-Kai Meng, Ph.D. Applications Chemist Analyzing Phenyl Ureas and Carbamates with HPLC/API-MS Presented by Chen-Kai Meng, Ph.D. Applications Chemist Outline Overview of ESI, APPI, and APCI sources Comparing ESI, APPI, and APCI results Identifying

More information

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS

Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS Fast quantitative Forensic Analysis of THC and its Metabolites in Biological Samples using Captiva EMR- Lipid and LC/MSMS Christophe Deckers, M.Sc. Sample prep Application Scientist Types of ``Interferences``

More information

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis

Comprehensive Study of SLE as a Sample. Preparation Tool for Bioanalysis Comprehensive Study of SLE as a Sample Preparation Tool for Bioanalysis Wan Wang, Warren Chen, Jerry Wang Bonna-Agela Technologies 179 Southern Street, West TEDA, Tianjin, China Abstract A simple, fast,

More information

The Raptor HILIC-Si Column

The Raptor HILIC-Si Column The Raptor HILIC-Si Column With Raptor LC columns, Restek chemists became the first to combine the speed of superficially porous particles (also known as SPP or core-shell particles) with the resolution

More information

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D Introduction Components of pharmaceutical products and food products often include high

More information

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management CONTENT Chapter No. Title Page No. Abbreviations List of Figures List of Tables 1 Introduction 1 1.1 Practical aspects 4 1.2 Stability-indicating assay method (SIAM) 5 1.3 Regulatory aspects 6 1.4 Techniques

More information

Analysis of HMF by HPLC

Analysis of HMF by HPLC COST Action 927 Training School Building Skills on the Analysis of Thermal Process Contaminants in Foods 22-26 October 2007, Ankara Analysis of HMF by HPLC Vural Gökmen O O OH Background O COOH O R 2 Carbonyl

More information

Polymer Additive Analysis by EI and APCI

Polymer Additive Analysis by EI and APCI Polymer Additive Analysis by EI and APCI Kate Yu, Eric Block LC/MS Waters Corporation Introduction to polymer additive analysis by LC/MS. Electron Ionization (EI) Advantages Classical spectra Library searchable

More information

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry Application Note Food Safety Authors Chenhao Zhai Agilent

More information

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results

Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results Sample Preparation Techniques for Biological Matrices: Finding the right balance to achieve optimal results Presenter: Diana Wang Applications Engineer SPP-Support@agilent.com 1 Today s Agenda - Why sample

More information

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column Clinical, Forensic & Toxicology Applications 4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column By Amanda Rigdon Abstract A rapid, sensitive method

More information

Application Note. Author. Abstract. Introduction. Food Safety

Application Note. Author. Abstract. Introduction. Food Safety Determination of β2-agonists in Pork with SPE eanup and LC-MS/MS Detection Using Agilent BondElut PCX Solid-Phase Extraction Cartridges, Agilent Poroshell 120 column and Liquid Chromatography-Tandem Mass

More information

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W

Shuguang Li, Jason Anspach, Sky Countryman, and Erica Pike Phenomenex, Inc., 411 Madrid Ave., Torrance, CA USA PO _W Simple, Fast and Accurate Quantitation of Human Plasma Vitamins and Their Metabolites by Protein Precipitation Combined with Columns Using HPLC-UV, HPLC-FLD or LC/MS/MS Shuguang Li, Jason Anspach, Sky

More information

MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager

MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager Quick Start Guide Step 1. Calibrate the Q-TOF LC/MS for low m/z ratios 2 Step 2. Set up a Flow Injection Analysis (FIA) method for

More information

SeQuant ZIC -HILIC For all who expect more...

SeQuant ZIC -HILIC For all who expect more... SeQuant ZIC -ILIC For all who expect more... The better choice for PLC and LC/MS of all types of polar and hydrophilic compounds EMD Millipore Corp. is a subsidiary of Merck KGaA, Darmstadt, Germany Poor

More information

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS A RAPID AD SESITIVE AALYSIS METD OF SUDA RED I, II, III & IV I TOMATO SAUCE USIG ULTRA PERFORMACE LC MS/MS Choon Keow G, aomi TAAKA, Michelle KIM, Swee Lee YAP Waters Asia, Regional Technology Center,

More information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer APPLICATION NOTE 72819 Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer Author Maria Grübner; Thermo Fisher Scientific, Germering,

More information

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Application Note: 346 MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine Gargi Choudhary and Diane Cho, Thermo Fisher Scientific, San Jose, CA Wayne Skinner and

More information

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID James Little, Eastman Chemical Company, Kingsport, TN Overview Phospholipids and

More information

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods

More information

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples WA20714 The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples Michael S. Young and Kevin M. Jenkins Waters Corporation, 34 Maple Street, Milford, MA

More information

Edgar Naegele. Abstract

Edgar Naegele. Abstract Simultaneous determination of metabolic stability and identification of buspirone metabolites using multiple column fast LC/TOF mass spectrometry Application ote Edgar aegele Abstract A recent trend in

More information

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System Keywords Introduction Application Note PHA0413 High Pressure Liquid Chromatography

More information

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Application ote #LCMS-2 esquire series Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices Introduction The simple monitoring

More information

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples WA20259 The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples Michael S. Young and Kevin M. Jenkins Waters Corporation, 34 Maple Street, Milford, MA 01757 Introduction

More information

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff Analysis of several common organic acids in tobacco leaf, snus, and moist snuff S. C. Moldoveanu, W. A. Scott R.J. Reynolds Tobacco Co. Background Among the organic acids commonly present in tobacco, are

More information

Application Note. Abstract. Authors. Pharmaceutical

Application Note. Abstract. Authors. Pharmaceutical Analysis of xycodone and Its Metabolites-oroxycodone, xymorphone, and oroxymorphone in Plasma by LC/MS with an Agilent ZRBAX StableBond SB-C18 LC Column Application ote Pharmaceutical Authors Linda L.

More information

What You Can t See Can Hurt You. How MS/MS Specificity Can Bite Your Backside

What You Can t See Can Hurt You. How MS/MS Specificity Can Bite Your Backside What You Can t See Can Hurt You How MS/MS Specificity Can Bite Your Backside Johan van den Heever, Tom Thompson, and Don Noot Agri-Food Laboratories Branch Advances in Trace rganic Residue Analysis early

More information

SeQuant ZIC -HILIC For all who expect more...

SeQuant ZIC -HILIC For all who expect more... SeQuant ZIC -ILIC For all who expect more... The better choice for PLC and LC/MS of all types of polar and hydrophilic compounds EMD Millipore is a division of Merck KGaA, Darmstadt, Germany Poor retention

More information

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract.

Application Note. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice. Author. Abstract. Agilent Application Solution Analysis of ascorbic acid, citric acid and benzoic acid in orange juice Application Note Author Food Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India 8 6 4 2

More information

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer A Robustness Study for the Agilent 7 LC-MS/MS Mass Spectrometer Application Note Clinical Research Authors Linda Côté, Siji Joseph, Sreelakshmy Menon, and Kevin McCann Agilent Technologies, Inc. Abstract

More information

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations For Research Use Only. Not for use in diagnostic procedures. 1 2015 AB Sciex RUO-MKT-11-1018-A For research

More information

Summary Chapter 8 CHAPTER 8. Summary. Page 173

Summary Chapter 8 CHAPTER 8. Summary. Page 173 CHAPTER 8 Summary Page 173 Chapter 2: Liquid Chromatography/Tandem Mass Spectrometry Method for Quantitative Estimation of PEG 400 and its Applications A rapid sensitive and selective MRM based method

More information

Fundamentals of Soft Ionization and MS Instrumentation

Fundamentals of Soft Ionization and MS Instrumentation Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:

More information

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for

More information

Authors. Abstract. Introduction. Environmental

Authors. Abstract. Introduction. Environmental Determination of Ultratrace Amitrol in Water Samples by in situ Derivatization-Solid Phase Extraction-Liquid Chromatography-Mass Selective Detector Application Environmental Authors Gerd Vanhoenacker,

More information

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Identification of Steroids in Water by Ion Trap LC/MS/MS Application Identification of Steroids in Water by Ion Trap LC/MS/MS Application Environmental Author Paul Zavitsanos and Christine Miller Agilent Technologies, Inc. Centerville Road Wilmington, DE 9- USA Abstract

More information

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Application Note Clinical Research Authors Derick Lucas and Limian Zhao Agilent Technologies, Inc. Abstract Lipids from

More information

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS Application Note Forensic Analysis Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS Authors Joan Stevens and Limian Zhao Agilent Technologies,

More information

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is 2, 4, 5-T. 1. Instrument Liquid Chromatograph-tandem mass spectrometer (LC-MS/MS)

More information

IC-MS Environmental Applications - Water Testing. Application Notebook

IC-MS Environmental Applications - Water Testing. Application Notebook IC-MS Environmental Applications - Water Testing Application Notebook Table of Contents Environmental Sample Using IC-MS 3 Environmental Applications: Water Testing Acrylamide in Water 4 Trace Urea in

More information

Author. Introduction. Abstract

Author. Introduction. Abstract Simultaneous Measurement of Sirolimus and Tacrolimus Concentrations in Blood by Semi-Automated Extraction and Liquid Chromatography-Electrospray Ionization Mass Spectrometry Application Author Gary L.

More information

SPE Cartridges. All Finneran Products Are Certified. Conditioning Sample application Wash Elution

SPE Cartridges. All Finneran Products Are Certified. Conditioning Sample application Wash Elution All Finneran Products Are Certified Finneran Solid Phase Extraction (SPE) cartridges provide a fast and efficient sample clean-up and concentration prior to analysis through GC, HPLC, and/or other instrument

More information

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1 Development of an SPE-LC/MS/MS Assay for the Simultaneous Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid in Support of Alzheimer s Research Dr. Erin E. Chambers Waters Corporation Presented

More information

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides Mohammed Abrar Unilabs York Bioanalytical Solutions Outline Introduction Why LC-MS/MS? Limitations of

More information

Neosolaniol. [Methods listed in the Feed Analysis Standards]

Neosolaniol. [Methods listed in the Feed Analysis Standards] Neosolaniol [Methods listed in the Feed Analysis Standards] 1 Simultaneous analysis of mycotoxins by liquid chromatography/ tandem mass spectrometry [Feed Analysis Standards, Chapter 5, Section 1 9.1 ]

More information

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Dominic Foley, Michelle Wills, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION

More information

MALDI-TOF. Introduction. Schematic and Theory of MALDI

MALDI-TOF. Introduction. Schematic and Theory of MALDI MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein

More information

ApplicationNOTE. Abstract

ApplicationNOTE. Abstract Abstract CAPILLARY ELECTROPHORESIS-MASS SPECTROMETRIC AND TANDEM MASS SPECTROMETRIC ANALYSIS AND IDENTIFICATION OF NON-STEROIDAL ANTI-INFLAMMATORY DRUGS AND THEIR METABOLITES IN HUMAN URINE * Alison E.

More information

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS Mark Ritchie and Evelyn Goh Waters Pacific Pte Ltd., Singapore A P P L I C AT ION B E N E F I T S Delivers a rapid 10-min MRM method

More information

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these

More information

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System Application Note Food Testing and Agriculture Mass-Based Purification of Natural Product Impurities Using an Agilent 126 Infinity II Preparative LC/MSD System Authors Florian Rieck and Jörg Hippler Agilent

More information

130327SCH4U_biochem April 09, 2013

130327SCH4U_biochem April 09, 2013 Option B: B1.1 ENERGY Human Biochemistry If more energy is taken in from food than is used up, weight gain will follow. Similarly if more energy is used than we supply our body with, weight loss will occur.

More information

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection Michael Jones Waters Corporation, Wilmslow, UK APPLICATION BENEFITS The ACQUITY Arc System is

More information

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System Mary E. Lame, Erin E. Chambers, Sukhdev S. Bangar, and Kenneth J. Fountain

More information

California TOF/QTOF Users Meetings: MS Settings. Jim Lau, Agilent Technologies.

California TOF/QTOF Users Meetings: MS Settings. Jim Lau, Agilent Technologies. California TOF/QTOF Users Meetings: MS Settings Jim Lau, Agilent Technologies. Data Acquisition Overview How to Improve Performance Optimize Source Parameters: Drying Gas Temperature, Gas Flow, and Neb

More information

Identification & Confirmation of Structurally Related Degradation Products of Simvastatin

Identification & Confirmation of Structurally Related Degradation Products of Simvastatin Identification & Confirmation of Structurally Related Degradation Products of Simvastatin Power of QTRAP Systems for Identification and Confirmation of Degradation Products Dilip Reddy 1, Chandra Sekar

More information

Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS

Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS Jean-Daniel Berset 1 and André Schreiber 2 1 Water and Soil Protection Laboratory,

More information

PosterREPRINT EVALUATION OF A RF-ONLY STACKED RING BASED COLLISION CELL WITH AXIAL FIELD FOR THE LC-MS-MS ANALYSIS OF THE METABOLITES OF RABEPRAZOLE

PosterREPRINT EVALUATION OF A RF-ONLY STACKED RING BASED COLLISION CELL WITH AXIAL FIELD FOR THE LC-MS-MS ANALYSIS OF THE METABOLITES OF RABEPRAZOLE OVERVIEW Comparison of the performance of a stacked ring based collision cell with axial field and a hexapole collision cell. Full scan LC-MS/MS-MS using precursor, product and neutral loss modes of operation

More information