1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2
|
|
- Valentine Dalton
- 6 years ago
- Views:
Transcription
1 Design Verification liquicolor 1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results Linearity and Detection Limit Linearity Detection Limit Method comparison Comparison against -cyclodextrin and polyanion methods Comparison against classical phosphotungstate precipitation method Traceability Recovery of control sera Interferences Endogeneous substances Anticoagulants Stability Lot real time stability at 8 C: Lot real time stability at 8 C: Lot real time stability at 8 C:... 1 Form: Rev. 9 valid of /11
2 1 Introduction The HDL-CHOLESTEROL liquicolor test is a direct homogeneous enzymatic assay for the quantitative determination of HDL cholesterol (HDL). The test combines two steps, the first step eliminating chylomicrons, VLDL and LDL by specific enzymatic degradation. The next step determines the remaining HDL by the well-known enzymatic reaction cascade cholesterol esterase, cholesterol oxidase and peroxidase. In the final reaction hydrogen peroxide oxidises a chromogen (N-(2-hydroxy-3-sulfopropyl)-3,5-dimethoxyaniline) under the catalytic action of peroxidase. The resulting colour change is monitored spectrophotometrically (593 nm, [57 61 nm], Hg 578 nm) and directly proportional to the HDL concentration in the sample. For calculation of the HDL concentration a calibrator is employed. 2 Imprecision The within-run imprecision of the HDL-CHOLESTEROL liquicolor test was calculated from 3-fold determinations. The day-to-day imprecision was calculated from the results obtained on 2 consecutive days. Pooled human sera as well as commercially available controls were employed as sample materials. 2.1 Within-run imprecision, results Pool serum high Pool serum medium Pool serum low Mean, mg/dl Standard deviation, mg/dl Coefficient of variation, % Day-to-day imprecision, results Pool serum Control I Control II Mean, mg/dl Standard deviation, mg/dl Coefficient of variation, % Linearity and Detection Limit 3.1 Linearity The linearity of the HDL CHOLESTEROL liquicolor method was controlled by employing a high concentration control serum and dilutions with phys. saline. The analysed concentrations were calculated vs. the regression line. Deviation from the regression line are expressed in absolute and relative values. As an example data measured on the Hitachi 717 are reported. High Pool Analytical Data Regressed Data Deviation from Regression Line Content (%) mg/dl mg/dl mg/dl (%) Rev. 9 2/11
3 MEASURED ANALYTE CONCENTRATION (mg/dl ) 25 2 Linear Regression Error Bars = ± 2SD LINEARITY HIGH POOL CONTENT (%) From the above results a linearity range up to 15 mg/dl HDL-CHOLESTEROL has been confirmed. 3.2 Detection Limit A 2-fold determination of a sample (phys. saline) on a HumaStar 1 analyser revealed an absolute mean of.67mg/dl and a SD of.11mg/dl. From the three-fold standard deviation the detection limit can therefore be calculated on the base of mean +3 SD to 1.mg/dl. 4 Method comparison 4.1 Comparison against -cyclodextrin and polyanion methods 27 specimens have been determined by this method and by two other accepted methods (method A: - cyclodextrin/peg enzymatic method; method B: polyanion/surfactant method). Discrepant results have been retested and confirmed by ultracentrifugation method. The HDL-CHOLESTEROL liquicolor test showed an excellent correlation with each other methods. N = 27 r =.996 Y = a*x + b (method A) a =.97, b = 1.25 Y = a*x + b (method B) a =.97, b = 4.2 A total of 6 discrepant results have been observed with method A. When the respective samples have been further investigated by fractionation it could be demonstrated that the HDL-CHOLESTEROL liquicolor results best reflect the true HDL concentration while method A yielded low results. 4.2 Comparison against classical phosphotungstate precipitation method 5 samples have been determined by the "classical" phosphotungstate precipitation method (e.g. HDL Cholesterol HUMAN, cat.-no.: 118 & 117) and by the HDL Cholesterol liquicolor test. The data obtained have been evaluated by means of a non-parametric regression analysis according to Passing&Bablok. The regression analysis yielded the following equation: Rev. 9 3/11
4 Y = a*x + b a =.9641 Confidence interval 95%.923 to 1.1 b = Confidence interval 95% to 3.6 Method comparison HDL CHOLESTEROL HDL Cholesterol liquicolor, mg/dl The above results demonstrate that HDL CHOLESTEROL liquicolor test recovers the independent controls within the assigned target range and that there is no deviation against a comparison method. 5 Traceability The HDL-CHOLESTEROL liquicolor test for the determination of HDL-Cholesterol is calibrated with a calibrator, which is traceable to the CDC reference method. The recovery of the calibrators as well as the comparison of 5 patient samples have been tested. The results are summarised in the tables below. Hitachi 717 Calibration: Reference Human assign. abs assign. abs cal-1. mg/dl 1. mg/dl cal mg/dl mg/dl Recovery: assign.. mg/dl 57.8 mg/dl 71.4 mg/dl Reference Human Reference Human Reference Human rep rep rep Mean Recovery % 1.7% 13.9% 11.% Rev. 9 4/11
5 Human HDL [mg/dl] Sample: X: Reference Y: Human Sample: X: Reference Y: Human , y =,987x -,46 R 2 =, Reference method HDL mg/dl The HDL-CHOLESTEROL liquicolor calibrator is traceable to the international reference method. 6 Recovery of control sera Commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. Fivefold determinations of each control serum have been performed with the HDL CHOLESTEROL liquicolor test method. The means of the fivefold determinations have been calculated and compared with the target values. Rev. 9 5/11
6 CONTROL SERUM RECOVERY HUMAN fresh Reference HUMAN 8 h 25 C Cal. factor Control LOT Target Range Result Deviation Result Deviation Result Deviation serum g/dl g/dl g/dl (%) g/dl (%) g/dl (%) Serodos # SeroDOS plus # Precinorm L # Lip control 1 # Lip control 2 # Sum No significant differences could be observed between fresh and stressed reagents, confirming the stability of the reagents. Controls were all recoverd well within the specified ranges. 7 Interferences 7.1 Endogeneous substances Any potential interfering influence by ascorbic acid, hemoglobin, bilirubin and turbidity (VLDL-cholesterol and VLDL-triglycerides) has been studied by adding pure substances to a normal pool serum. The data are summarised in the following table. Ascorbic acid, % %.4 %.8 %.12 %.16 %.2 % Found HDL, mg/dl Hemoglobin, mg/dl Found HDL, mg/dl Bilirubin, mg/dl Found HDL, mg/dl VLDL triglycerides, mg/dl Found HDL, mg/dl VLDL cholesterol, mg/dl Found HDL, mg/dl The above results demonstrate that ascorbic acid up to.2 %, hemoglobin up to 5 mg/dl, bilirubin up to 3 mg/dl do not interfere with the test (Glick value = 1) while VLDL-triglycerides up to 13 mg/dl showed only minor interference (Glick value = 2). 7.2 Anticoagulants To pooled plasma different additives in different concentrations were added to show, if they affect the test results. The results are summarised in the following table: Concentr Anticoag. Na-Citr. mg/l HDL mg/dl Na-Hep. mg/l HDL mg/dl Na-Fl. mg/l HDL mg/dl Na-oxal. mg/l HDL mg/dl EDTA mg/l HDL mg/dl The results gave clear evidence that none of the anticoagulants has an effect on the test results. 8 Stability The stability of the HDL-CHOLESTEROL liquicolor test has been tested on real-time storage conditions over a period of 2 months for three independent production lots with different commercial controls. The measurement was made on Hitachi 717. The individual results are summarised in the following tables. Rev. 9 6/11
7 8.1 Lot real time stability at 8 C: Duration (month) Sensitivity Blank STD mabs./1mg/dl (deviation from T=) (1%) (13%) (99%) (12%) (11%) (99%) (98%) (93%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cal.-D 6.2 6, Lot , Target: 6.5 mean , recov.% 1% 99% 99% 1% 1% 99% 99% 1% Lipid Cont. I , Lot , Target: 82. Mean , recov.% 97% 98% 11% 1% Lipid Cont. I Lot Target: 85. mean recov.% 13% 14% 14% 15% 18% Lipid Cont. II , Lot , Target: 5.5 mean , recov.% 97% 96% 1% 98% Lipid Cont. II Lot Target: 52.5 mean recov.% 99% 98% 99% 1% 99% Precision n=2 mean Liquid Cont. N SD CV n=2 mean Liquid Cont. SD AN CV Linearity % % % % % % % % % % % Lin. material lot # X-3 X-3 X-4 X-4 X-4 X-4 X-4 X-4 Rev. 9 7/11
8 HDL-C(mg/dL) Deviation from T= 14% Sensitivity [per 1mg/dL] 12% 1% 8% 6% 4% 2% % Duration (month) 2 Linearity % 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% M 3M 6M 9M 12M 15M 18M 2M 8.2 Lot real time stability at 8 C: Duration (month) Sensitivity Blank STD mabs./1mg/dl (deviation from T=) (1%) (12%) (11%) (12%) (11%) (97%) (97%) (92%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cal.-D Lot Target: 6.5 mean recov.% 1% 11% 1% 1% 12% 12% 11% 1% Lipid Cont. I Lot Target: 82. mean recov.% 97% 11% 99% 12% Lipid Cont. I Lot Target: 85. mean recov.% 14% 13% 17% 15% 16% Rev. 9 8/11
9 Lipid Cont. II Lot Target: 5.5 mean recov.% 98% 97% 98% 1% Lipid Cont. II Lot Target: 52.5 mean recov.% 99% 98% 13% 11% 99% Precision n=2 mean Liquid Cont. N SD CV n=2 mean Liquid Cont. AN SD CV Linearity % % % % % % % % % % % Lin. material lot # X-3 X-3 X-4 X-4 X-4 X-4 X-4 X-4 14% Sensitivity [per 1mg/dL] De via tio n fro m T= 12% 1% 8% 6% 4% 2% % Duration (month) Rev. 9 9/11
10 HDL-C(mg/dL) 2 Linearity % 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% M 3M 6M 9M 12M 15M 18M 2M 8.3 Lot real time stability at 8 C: Duration (month) Sensitivity Blank STD mabs./1mg/dl (deviation from T=) (1%) (12%) (11%) (11%) (11%) (98%) (93%) (93%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cal.-D Lot Target: 6.5 mean recov.% 13% 11% 99% 12% 11% 1% 97% 99% Lipid Cont. I Lot Target: 82. mean recov.% 11% 1% 11% 13% Lipid Cont. I Lot Target: 85. mean recov.% 15% 14% 14% 14% 14% Lipid Cont. II Lot Target: 5.5 mean recov.% 1% 97% 99% 1% Lipid Cont. II Lot Target: 52.5 mean recov.% 11% 99% 11% 99% 97% Precision n=2 mean Liquid Cont. N SD CV n=2 mean Liquid Cont. AN SD CV Rev. 9 1/11
11 HDL-C(mg/dL) Deviation from T= Linearity % % % % % % % % % % % Lin. material lot # X-3 X-3 X-4 X-4 X-4 X-4 X-4 X-4 14% Sensitivity [per 1mg/dL] 12% 1% 8% 6% 4% 2% % Duration (month) 2 Linearity % 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% M 3M 6M 9M 12M 15M 18M 2M The above results clearly demonstrate the stability of the test for more than 18 months after production. In addition to the real-time stability studies, temperature stress tests are routinely performed on each production batch as part of the end control of the product. Rev. 9 11/11
3 Linearity and Detection Limit Linearity Detection limit... 4
Design Verification liquicolor 1 Introduction... 2 2 Imprecision, Reproducibility... 2 2.1 Data from internal study... 2 2.2 Date from external study (Cholesterol Reference Method Laboratory Network)...
More informationliquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationLIPASE liquicolor. Design Verification. Multipurpose Reagent
Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of
More informationliquicolor (AMP Buffer, IFCC) Design Verification
Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...
More information1. Reagent Store at 2-8 ºC.
LDL Insert Ref.:146 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. Professional use. [For in vitro diagnostic use.] Test principle.
More informationBCH 447. Triglyceride Determination in Serum
BCH 447 Triglyceride Determination in Serum Introduction: Triglycerides are esters of fatty acids and are hydrolyzed by lipase to glycerol and free fatty acids. Triglyceride determinations when performed
More informationLDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings
LDL LD Insert Ref.:129 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. [For in vitro diagnostic use.] Test principle. The method
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More informationTriglyceride determination
Triglyceride determination Introduction: - Triglycerides are esters of fatty acids and are hydrolyzed to glycerol and free fatty acids (by lipase) - Triglyceride determinations when performed in conjunction
More informationHexagon PSA. Design Verification. Contents
Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation
More informationHumaTex CRP. Design Verification. Contents
Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...
More informationTotal Thyroxine ELISA (T4)
Design Verification Total Thyroxine ELISA (T4) Contents 1 Assay Principle... 2 2 Imprecision... 2 Within-run Imprecision... 2 Between run Imprecision... 2 3 Comparison of Methods, Accuracy... 2 4 Linearity...
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by
More informationIMTEC-ANA-LIA MAXX. Design Verification
Design Verification IMTEC-ANA-LIA MAXX CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Specificity... 2 3 Interferences... 4 4 Imprecision... 6 4.1 Within-Run Imprecision... 6 4.2 Between-Run
More informationDesign Verification. Form: /5 Rev. 007 valid of
Design Verification g 1 Determination of Diagnostic Sensitivity and Diagnostic Specificity... 2 Crossreactivity... 3 Potentially interfering substances... 3 2 Imprecision... 4 3 Standardisation... 4 4
More informationDesign Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...
Design Verification IMTEC-TSH RECEPTOR-ANTIBODIES CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 2.1 Determination of Diagnostic Sensitivity and Diagnostic Specificity...
More informationEvaluation of a Rapid Homogeneous Method for Direct Measurement of High-Density Lipoprotein Cholesterol
Clinical Chemistry / ORIGINAL ARTICLE Evaluation of a Rapid Homogeneous Method for Direct Measurement of High-Density Lipoprotein Cholesterol R. Scott Hubbard, MD,1 Shaina V. Hirany, MS,2 Sridevi Devaraj,
More informationSuppl. Table 1: CV of pooled lipoprotein fractions analysed by ESI-MS/MS
Supplement VLDL LDL HDL PC 3.3 1.77 1.3 LPC 4.82 2.5.35 SM 3.1 4.6 1.92 CER 2.17 6.3 4.15 PE 3.18 1.93 2.79 PE-pl 13.18 1.9 2.32 CE 2.9.65.4 FC.36 3.5 2.54 Suppl. Table 1: CV of pooled lipoprotein fractions
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.
en CHOLESTEROL 7D62 304796/R02 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. Read Highlighted Changes Revised October 2012 Package insert
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationLipid Profile Analysis of Aircrew
Original Article Lipid Profile Analysis of Aircrew Sampath S *, Richa + ABSTRACT Most laboratories including all our Armed Forces Centers, estimate Low Density Lipoproteins Cholestorol (LDL-C) levels with
More informationHigh-density Lipoprotein Cholesterol (HDL-C) Assay Kit
(FOR RESEARCH USE ONLY. DO NOT USE IT IN CLINICAL DIAGNOSIS!) High-density Lipoprotein Cholesterol (HDL-C) Assay Kit (Double reagents) Catalog No: E-BC-K221 Method: Colorimetric method Specification: 96T
More informationPLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL
PLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL Lipids are characterized by low polarity and limited solubility in water. Their plasma concentration is about 500-600
More informationHDL CHOLESTEROL. 01 English - Ref.: 13. Ref.:13. Insert
HDL CHOLESTEROL Insert Ref.:13 Intended use. System for selective precipitation of Low and Very Low Density Lipoproteins (LDL and VLDL) and HDL cholesterol measurement in the supernatant, by an end point
More informationEvaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser
Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser
More informationINSTRUCTIONS FOR USE
VITROS Chemistry Products Reagent 680 1728 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More informationSeparation of HDL Particles by Immunoprecipitation
Sun Diagnostics, LLC Separation of HDL Particles by Immunoprecipitation Rae-Anne Nguyen and John H. Contois 13 Introduction We all know that HDL cholesterol concentration is inversely associated with coronary
More informationKinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase
Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,
More informationTurbidos. Design Verification. Contents
Design Verification Turbidos Contents 1 adjustment... 2 2 Target value determination... 2 3 Stability... 2 3.1 Accelerated stress data... 2 3.2 Open vial stability... 3 3.3 Real Time Stability... 4 4 Result
More informationEstimation of glucose in blood serum
Estimation of glucose in blood serum Enzymatic estimation of glucose uses a reagent containing two enzymes and a chromogen. Glucose oxidase catalyses the oxidation of glucose to gluconolactone with the
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationFor in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date
AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationApplicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.
Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required
More informationEvaluation Report. Eurolyser PT (INR) test kit (ST0180) on CUBE and smart analysers
Evaluation Report Eurolyser PT test kit (ST0180) on CUBE and smart analysers Locations Location 1: Eurolyser Diagnostica GmbH Operator: Simone Wieser Date: 30.06.2015 Specimens The specimens used for analysis
More informationG. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626
Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi
More informationGlutathione Assay Kit
Glutathione Assay Kit Catalog Number KA1649 250 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationLDLC3. English System information For cobas c 311/501 analyzers:
Order information Analyzer(s) on which cobas c pack(s) can be used 07005717 190 LDL Cholesterol Gen.3 (200 tests) System ID 07 7565 7 Roche/Hitachi cobas c 311, cobas c 501/502 12172623 122 Calibrator
More informationLIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE
LIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE Revised 3/2014 LIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE For questions concerning protocol and assay development, validation of results, etc., please
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationLPO-CC (Lipid Peroxides) Kit
K-ASSAY LPO-CC (Lipid Peroxides) Kit For the Quantitative Colorimetric Determination of High Levels of Lipid Peroxides in Serum, Plasma and Other Biological Samples. Cat. No. CC-004 For Research Use Only.
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationMaterials and Methods
Materials and Methods The study was conducted at School of Health Sciences, Kannur University Thalassery Campus, Palayad, and Pushpagiri Institute of Medical Sciences and Research Center, Thiruvalla, Kerala,
More informationUREA CE Insert. 01 English - Ref.: 27. Ref.:27
UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]
More informationDesign Verification. Form:
Design Verification SYPHILIS TPHA liquid CONTENTS 1 Function... 2 2 Sensitivity... 2 Preparation of Serum Control Panel... 2 Kit Controls Positive and Negative... 3 3 Traceability... 3 4 Specificity and
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationhc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine
Design Verification hc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine 1 Function... 2 2 Sensitivity and Dynamic Range... 2 2.1 Analytical Sensitivity... 2 3 Validation of Efficacy... 3 3.1
More informationLDL Liquiform. 01 English - Ref.: 111. Ref.:111. Insert. Intended use. Test principle. Methodology. Reagents Precautions and warnings
LDL Liquiform Insert Ref.:111 Intended use. System for direct homogenous determination of LDL cholesterol in serum and plasma Test principle. On the first stage of the reaction, a specific surfactant present
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationLipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives
ISPUB.COM The Internet Journal of Health Volume 8 Number 1 Lipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives P Igbinaduwa, B Igbinaduwa, I Oforofuo Citation P Igbinaduwa,
More informationSerodos and Serodos plus
Design Verification Serodos and Serodos plus Contents 1 Value Adjustment... 2 2 Target Determination... 2 3 Stability... 2 Real-Time Stability... 3 Stability after Reconstitution... 4 Stability after Reconstitution
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,
More informationGeneral Chemistry Scheme Guide
General Chemistry Scheme Guide Copyright WEQAS. All rights reserved. No part of this document may be reproduced or utilised in any form without permission from WEQAS Contents. Scheme details and repertoire.....
More informationTABLE OF CONTENTS GENERAL INFORMATION... 1
BIOO RESEARCH PRODUCTS Glucose Assay Kit Manual Catalog #: 5611-01 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure Overview... 1 Required Materials
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.
CHOLESTEROL 7D62-21 304342/R1 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. NOTE: This package insert must be read
More informationComparison of a Homogeneous Assay With a Precipitation Method for the Measurement of HDL Cholesterol in Diabetic Patients
Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Comparison of a Homogeneous Assay With a Precipitation Method for the Measurement of HDL Cholesterol in Diabetic Patients TONNY JENSEN, MD
More informationab Sphingomyelin Assay Kit
ab133118 - Sphingomyelin Assay Kit Instructions for Use For the specific, sensitive and convenient method of quantifying Sphingomyelin in plasma or serum. This product is for research use only and is not
More informationClinical Chemistry Specific Proteins Presentation
Clinical Chemistry Specific Proteins Presentation November 2004 AEROSET and c8000 are registered trademarks of Abbott Laboratories. All other trademarks, brands, trade names and product names are the property
More informationEvaluation of Calculated Low-Density Lipoprotein Against a Direct Assay
Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay Navapun Charuruks MD*, Anna Milintagas BSc** * Department of Laboratory Medicine, Faculty of Medicine, Chulalongkorn University **
More informationEvaluation of the Abbott CELL-DYN 4000 Hematology Analyzer
Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationS4. Summary of the GALNS assay validation. Intra-assay variation (within-run precision)
S4. Summary of the GALNS assay validation (i.) Intra-assay variation (within-run precision) Intra-assay variation was determined by measuring standard blood samples (low activity standard; medium activity
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationTechnical Bulletin. Cholesterol Reference Method Laboratory Network (CRMLN) Overview. TB Rev. 0
Technical Bulletin Reference Method Laboratory Network (CRMLN) Overview TB000020 Rev. 0 Reference Method Laboratory Network (CRMLN) - General Information With guidance from the US Centers for Disease Control
More informationNew immunoseparation-based homogeneous assay for HDL-cholesterol compared with three homogeneous and two heterogeneous methods for HDL-cholesterol
Clinical Chemistry 44:7 1443 1451 (1998) Lipids and Lipoproteins New immunoseparation-based homogeneous assay for HDL-cholesterol compared with three homogeneous and two heterogeneous methods for HDL-cholesterol
More informationCreatinine (serum, plasma)
Creatinine (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Creatinine 1.2 Alternative names None 1.3 Description of analyte Creatinine is a heterocyclic nitrogenous compound (IUPAC
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationPhosphorus Test Code: PHOS
Intended Use Phosphorus Test Code: PHOS For in vitro diagnostic use only CARESIDE cartridges are used with the CARESIDE Analyzer to quantitatively measure inorganic phosphorus in anticoagulated whole blood,
More informationBÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:
More informationPatient: 55 y female (ambulatory)
Disclosures Speaking Honoraria Radiometer (Canada) Nova Biomedical, Draeger Roche Diagnostics (Canada) Research Support (Reagents, Instrumentation, Travel) Nova Biomedical Abbott Laboratories (Canada)
More informationElectrolyte Analyzer with Ion-Selective Electrode and Blood Gas analyzer
Biochemistry & Fundamental Reference Laboratory, Yokohama Our Head Office and Biochemistry Reference Laboratory moved to Yokohama with aims to have further development on the 21 st September, 2018. [Reference
More informationCholesterol (blood, plasma, serum)
1 Cholesterol (blood, plasma, serum) 1 Name and description of analyte 1.1 Name of analyte Cholesterol (plasma; also blood, serum) 1.2 Alternative names 2,15-dimethyl-14-(1,5-dimethylhexyl)tetracyclo[8.7.0.0
More informationELEGANCE Chlamydia pneumoniae IgG ELISA KIT
INTENDED USE The ELEGANCE Chlamydia pneumoniae IgG ELISA has been designed for the in vitro diagnostic measurement of anti- Chlamydia pneumoniae IgG in the screening of human serum. PRINCIPLES OF THE ELEGANCE
More informationab65390 HDL and LDL/VLDL Cholesterol Assay kit
ab65390 HDL and LDL/VLDL Cholesterol Assay kit (Colorimetric/Fluorometric) Instructions for Use For the rapid, sensitive and accurate measurement of HDL and LDL/VLDL Cholesterol in various samples. This
More informationEliKine Free Thyroxine (ft4) ELISA Kit
EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE
More informationForum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007
Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease
More informationOriginal Article Interfering effect of bilirubin on the determination of alkaline phosphatase
Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,
More informationEvaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser
Evaluation Report Eurolyser Phenobarbital test kit (VT0211) on solo Analyser Location Location: Eurolyser Diagnostica GmbH Operator: Bettina Schönauer Date: February 2016 - March 2016 Specimens The specimens
More informationActive-B12 (Holo-TC) ELISA KIT
Active-B12 (Holo-TC) ELISA KIT Cat. No.:DEIA6312 Pkg.Size:96T Intended use The Active-B12 (Holotranscobalamin) assay is an enzyme-immunoassay (ELISA) for the quantitative determination of holotranscobalamin
More informationPlasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana
Plasma Hemoglobin Determination: Two Manual Methods Compared An Honors Thesis (HONRS 499) by Anne M. Emerick Sharlene Strahl St. Francis Hospital and Health Centers Beech Grove, Indiana Dr. Nancy Behforouz
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationPiccolo Lipid Panel Reagent Disc
Piccolo Lipid Panel Reagent Disc Customer and Technical Service: 800-822-2947 Customers outside the US should contact their local Abaxis representative for customer service July 2011 PN: 400-7144-1 Rev.:
More informationINTERNAL QUALITY CONTROL ( Q I C) QC)
EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)
More informationCreatinine Test Code: CRE
Creatinine Test Code: CRE Intended Use For in vitro diagnostic use only CARESIDE cartridges are used with the CARESIDE Analyzer to quantitatively measure creatinine concentration in anti-coagulated whole
More informationAspartate Aminotransferase Test Code: AST
Aspartate Aminotransferase Test Code: AST Intended Use For in vitro diagnostic use only CARESIDE Aspartate Aminotransferase (AST) cartridges are used with the CARESIDE Analyzer to quantitatively measure
More informationESTIMATION OF LECITHIN CHOLESTROL ACYL TRANSFERASE ACTIVITY AND HIGH DENSITY LIPOPROTEIN CHOLESTROL IN TYPE 2 DIABETES MELLITUS
ESTIMATION OF LECITHIN CHOLESTROL ACYL TRANSFERASE ACTIVITY AND HIGH DENSITY LIPOPROTEIN CHOLESTROL IN TYPE 2 DIABETES MELLITUS *Tahmeen Jameel 1 and Syed Mahmood Ahmed 2 1 Department of Biochemistry,
More informationStandard Operating Procedure
Subject Alkaline Phosphatase C111 Index Number Lab-1504 Section Laboratory Subsection Regional Clinic/Affiliate Hospital Laboratories Category Departmental Contact Kamprud, Elizabeth Last Revised 4/12/2017
More informationENZYMATIC CREATININE Insert
ENZYMATIC CREATININE Insert Ref.:17 Intended use. System for determination of creatinine in serum, plasma, and urine samples by end-point reaction. Professional use. [ Only for in vitro diagnostic use.]
More informationAccuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips
Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Sponsored by Arthur Roberts, MD of The Living Heart Foundation The Living Heart
More informationRat TSH ELISA KIT. Please, read this instruction carefully before use.
Rat TSH ELISA KIT Research Reagent Please, read this instruction carefully before use. This is an ELISA (Enzyme Linked ImmunoSorbent Assay) kit for measurement of rat TSH (thyroid-stimulating hormone)
More informationDetermination of hemoglobin is one of the most commonly
ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning
More informationMethod Comparison Report Semi-Annual 1/5/2018
Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community
More informationSmall dense low-density lipoprotein is a risk for coronary artery disease in an urban Japanese cohort: The Suita study
Small dense low-density lipoprotein is a risk for coronary artery disease in an urban Japanese cohort: The Suita study Hidenori Arai 1, Yoshihiro Kokubo 2, Makoto Watanabe 2, Tatsuya Sawamura 3, Tomonori
More informationFree Glycerol Assay Kit (Colorimetric)
Product Manual Free Glycerol Assay Kit (Colorimetric) Catalog Number STA-398 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Glycerol is the backbone of Triglycerides
More information