3 Linearity and Detection Limit Linearity Detection limit... 4
|
|
- Marian Craig
- 5 years ago
- Views:
Transcription
1 Design Verification liquicolor 1 Introduction Imprecision, Reproducibility Data from internal study Date from external study (Cholesterol Reference Method Laboratory Network) Linearity and Detection Limit Linearity Detection limit Method comparison Friedewald vs. LDL-CHOLESTEROL Roche-BM vs. LDL-CHOLESTEROL liquicolor LDL CHOLESTEROL vs. Friedewald, performed at University hospital Mainz, Germany Roche Diagnostics vs. LDL CHOLESTEROL, performed at University hospital Mainz, Germany Traceability Interference study Endogeneous substances Anticoagulants Stability Lot real time stability at 8 C: Lot real time stability at 8 C: ot real time stability at 8 C QM-Element: ISO91:2 No Rev. 7 valid of /13
2 1 Introduction The LDL CHOLESTEROL liquicolor test is a direct homogeneous enzymatic assay for the quantitative determination of LDL-cholesterol (LDL). The test combines two steps, the first step eliminating chylomicrons, VLDL and HDL by specific enzymatic degradation. The next step determines the remaining LDL by the well-known enzymatic reaction cascade cholesterol esterase, cholesterol oxidase and peroxidase. In the final reaction hydrogen peroxide oxidises a chromogen (N-(2-hydroxy-3-sulfopropyl)-3,5-dimethoxyaniline) under the catalytic action of peroxidase. The resulting colour change is monitored spectrophoto-metrically (555 nm, [ nm] Hg 578 nm) and is directly proportional to the LDL concentration in the sample. For calculation of the LDL concentration a calibrator is employed. 2 Imprecision, Reproducibility The within-run imprecision of the LDL CHOLESTEROL liquicolor test was calculated from 1-fold determinations. The day-to-day imprecision was calculated from the results obtained on 4 different dates. Pooled human sera were employed as sample materials. 2.1 Data from internal study Sample Date 1 Date 2 Date 3 Date 4 Low 1 Medium 1 Low 1 Medium 1 Low 1 Medium 1 Low 1 Medium Mean, 84,7 141,9 84,7 142,4 83,5 141,8 84,7 142,1 SD,,82,99 1,16 1,7 1,8 1,3 1,16 1,2 CV, %,97,7 1,37,75 1,29,73 1,37,84 1 all results expressed in Overall results were as follows: Low Medium Overall mean, Overall SD, Overall CV, % Date from external study (Cholesterol Reference Method Laboratory Network) Three well characterised pool sera were employed as sample material and was tested over a period of 2 consecutive days with the LDL CHOLESTEROL test. The results are summarised in the following table. Day Sample 1 Sample 2 Sample 3 Result, Result, Result, Rev. 7 2/13
3 MEASURED ANALYTE CONCENTRATION (mg/dl ) Mean, SD, CV,% The above results demonstrate the excellent reproducibility of the test. Within-run and day-to-day imprecisions were found to be around or even below 1%. 3 Linearity and Detection Limit The linearity of the LDL CHOLESTEROL liquicolor method was controlled by employing a high concentration control serum and dilutions with phys. saline. The analysed concentrations were calculated vs. the regression line. Deviation from the regression line are expressed in absolute and relative values. As an example data measured on the Hitachi 717 are reported. 3.1 Linearity High Pool Analytical Data Regressed Data Deviation from Regression Line Content (%) (%) Linear Regression Error Bars = ± 2SD LINEARITY HIGH POOL CONTENT (%) Conclusion: From the above results a linearity range up to 5 LDL has been confirmed. It should be noted, however, that the linearity of the assay depends on the respective analyser application. Rev. 7 3/13
4 3.2 Detection limit A 2-fold determination of a sample (phys. saline) on a HumaStar 1 analyser revealed an absolute mean of -.15 and a SD of.5. From the three-fold standard deviation the detection limit can therefore be calculated on the base of mean +3 SD to Method comparison Independent studies have been performed to verify the comparability of the LDL CHOLESTEROL liquicolor test with the CDC reference method, other LDL tests and the LDL results calculated according to the Friedewald formula. 4.1 Friedewald vs. LDL-CHOLESTEROL The LDL CHOLESTEROL test has been compared against results obtained with by the classical Friedewald formula. 15 serum samples have been tested and evaluated by a non-parametric regression analysis. Y = A * X + B A to 1.35 B, to 1.35 r.938 Low, for Y: 27.1 for X: High, for Y: for X: Median, for Y: 12.2 for X: Method comparison LD L FR IEDEWALD Rev. 7 4/13
5 4.2 Roche-BM vs. LDL-CHOLESTEROL liquicolor The LDL CHOLESTEROL liquicolor test has been compared against results obtained with by a competitor s test (Roche-BM). 15 serum samples have been tested with both methods and evaluated by a non-parametric regression analysis. Y = A * X + B A to 1.51 B, to r.938 Low, for Y: 27.1 for X: 37.9 High, for Y: for X: Median, for Y: 12.2 for X: Method comparison LDL 2 LD L Ch 15 ole stemg rol /dl HU M AN Roche-BM 4.3 LDL CHOLESTEROL vs. Friedewald, performed at University hospital Mainz, Germany The LDL CHOLESTEROL test has been compared against results obtained with by the classical Friedewald formula. 12 serum samples have been tested on an analyzer (CHOLESTEROL, HDL CHOLESTEROL, triglycerides and LDL CHOLESTEROL liquicolor). Results were evaluated by a non-parametric regression analysis. Y = A * X + B A B, r.974 Low, for Y: 32 for X: 29 High, for Y: 263 for X: 239 Median, for Y: 133 for X: 12 Rev. 7 5/13
6 Method comparison University hospital Mainz, Germany FR IEDEWALD 4.4 Roche Diagnostics vs. LDL CHOLESTEROL, performed at University hospital Mainz, Germany The LDL CHOLESTEROL test has been compared against results obtained with by a competitor s test (Roche-BM). 6 serum samples have been tested with both methods and evaluated by a non-parametric regression analysis. Y = A * X + B A B, r.968 Low, for Y: 45 for X: 56 High, for Y: 263 for X: 25 Median, for Y: 135 for X: 134 Rev. 7 6/13
7 mg /d l ROCHE Diagn ostics mg /dl Conclusion: The method comparison studies demonstrated a very well correlation with long-standing classical methods as well as with competitor tests. 5 Traceability The LDL-CHOLESTEROL liquicolor test for the determination of LDL-Cholesterol is calibrated with a calibrator, which is traceable to the CDC reference method. The recovery of the calibrators have been tested as well as the comparison of 5 patient samples. The results are summarised in the tables below. Hitachi 717 Calibration: Reference Human assign. abs assign. abs cal cal Recovery: assign Reference Human Reference Human Reference Human rep rep rep Mean Recovery % 1.3% 12.9% 11.4% Rev. 7 7/13
8 Human LDL [mg/dl] Sample: X: Reference Y: Human Sample: X: Reference Y: Human y =,998x -,593 R 2 =, Reference method LDL [mg/dl] Conclusion: The LDL-CHOLESTEROL liquicolor calibrator is traceable to the international reference method. Rev. 7 8/13
9 6 Interference study 6.1 Endogeneous substances Any potential interfering influence by ascorbic acid, hemoglobin and bilirubin has been studied by adding pure substances to a normal pool serum. The obtained results demonstrate that ascorbic acid up to 5, hemoglobin up to 5 and bilirubin up to 3 do not interfere with the test (Glick value = 1). VLDL triglycerides up to 13, VLDL cholesterol (total) up to 5, chylomicron triglycerides up to 7 and chylomicron cholesterol (total) up to 17 showed only minor or negligible interference (Glick value = 1-2). 6.2 Anticoagulants To pooled plasma different additives in different concentrations were added to show, if they affect the test results. The results are summarised in the following table: Concentr Anticoag. Na-Citr. mg/l LDL Na-Hep. mg/l LDL Na-Fl. mg/l LDL Na-oxal. mg/l LDL EDTA mg/l LDL Conclusion: The results gave clear evidence that none of the anticoagulants has an effect on the test results. 7 Stability The stability of the LDL-CHOLESTEROL liquicolor test has been tested on real-time storage conditions over a period of 21 months for three independent production lots with different commercial controls. The measurement was made on Hitachi 717. The individual results are summarised in the following tables. 7.1 Lot real time stability at 8 C: Duration (days) (months) 6M 9M 13M 18M 21M Sensitivity BLK STD mabs./1mg/dl (deviation from T=) (1%) (97%) (1%) (98%) (97%) (93%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cont. I mean Lot recovery% 1% 98% 11% 98% 96% 98% Rev. 7 9/13
10 LDL-C(mg/dL) Deviation from T= Lipid Cont. II mean Lot recovery% 1% 98% 98% 98% 99% 99% Linearity % % % % % % % % % % % Lin. material lot # A-12 A-12 A-13 A-14 A-15 A-15 Sensitivity [per 1mg/dL] 14% 12% 1% 8% 6% 4% 2% % Duration (days) Linearity % 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% 6M 9M 13M 18M 21M Rev. 7 1/13
11 L D L - C ( m g / d L ) 7.2 Lot real time stability at 8 C: Duration (days) (months) 6M 9M 13M 18M 21M Sensitivity BLK STD mabs./1mg/dl (deviation from T=) (1%) (98%) (1%) (98%) (97%) (93%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cont. I mean Lot recovery% 1% 97% 11% 97% 96% 95% Lipid Cont. II mean Lot recovery% 1% 97% 99% 97% 11% 98% Linearity % % % % % % % % % % % Lin. material lot # A-12 A-12 A-13 A-14 A-15 A-15 De viat ion fro m T= 14% 12% 1% 8% Sensitivity [per 1mg/dL] 6% 4% 2% % Duration (days) L i n e a r i t y % 1 % 2 % 3 % 4 % 5 % 6 % 7 % 8 % 9 % 1 % 6 M 9 M 1 3 M 1 8 M 2 1 M Rev. 7 11/13
12 Deviation from T= 7.3 ot real time stability at 8 C Duration (days) (months) 6M 9M 13M 18M 21M Sensitivity BLK STD mabs./1mg/dl (deviation from T=) (1%) (97%) (99%) (97%) (96%) (92%) Calibrator lot # Assigned value (mg/dl) Specificity Lipid Cont. I mean Lot recovery% 1% 97% 12% 98% 97% 98% Lipid Cont. II mean Lot recovery% 1% 11% 11% 98% 1% 1% Linearity % % % % % % % % % % % Lin. material lot # A-12 A-12 A-13 A-14 A-15 A-15 Sensitivity [per 1mg/dL] 14% 12% 1% 8% 6% 4% 2% % Duration (days) Rev. 7 12/13
13 LDL-C(mg/dL) Linearity % 1% 2% 3% 4% 5% 6% 7% 8% 9% 1% 6M 9M 13M 18M 21M Conclusion: The above results support the claimed stability of the test of 18 months from the date of production. Rev. 7 13/13
1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...
More informationliquicolor (AMP Buffer, IFCC) Design Verification
Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...
More informationliquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationLIPASE liquicolor. Design Verification. Multipurpose Reagent
Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of
More informationHexagon PSA. Design Verification. Contents
Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation
More informationHumaTex CRP. Design Verification. Contents
Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...
More informationTotal Thyroxine ELISA (T4)
Design Verification Total Thyroxine ELISA (T4) Contents 1 Assay Principle... 2 2 Imprecision... 2 Within-run Imprecision... 2 Between run Imprecision... 2 3 Comparison of Methods, Accuracy... 2 4 Linearity...
More informationDesign Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...
Design Verification IMTEC-TSH RECEPTOR-ANTIBODIES CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 2.1 Determination of Diagnostic Sensitivity and Diagnostic Specificity...
More informationBCH 447. Triglyceride Determination in Serum
BCH 447 Triglyceride Determination in Serum Introduction: Triglycerides are esters of fatty acids and are hydrolyzed by lipase to glycerol and free fatty acids. Triglyceride determinations when performed
More information1. Reagent Store at 2-8 ºC.
LDL Insert Ref.:146 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. Professional use. [For in vitro diagnostic use.] Test principle.
More informationLDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings
LDL LD Insert Ref.:129 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. [For in vitro diagnostic use.] Test principle. The method
More informationDesign Verification. Form: /5 Rev. 007 valid of
Design Verification g 1 Determination of Diagnostic Sensitivity and Diagnostic Specificity... 2 Crossreactivity... 3 Potentially interfering substances... 3 2 Imprecision... 4 3 Standardisation... 4 4
More informationTriglyceride determination
Triglyceride determination Introduction: - Triglycerides are esters of fatty acids and are hydrolyzed to glycerol and free fatty acids (by lipase) - Triglyceride determinations when performed in conjunction
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by
More informationLipid Profile Analysis of Aircrew
Original Article Lipid Profile Analysis of Aircrew Sampath S *, Richa + ABSTRACT Most laboratories including all our Armed Forces Centers, estimate Low Density Lipoproteins Cholestorol (LDL-C) levels with
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More informationEstimation of glucose in blood serum
Estimation of glucose in blood serum Enzymatic estimation of glucose uses a reagent containing two enzymes and a chromogen. Glucose oxidase catalyses the oxidation of glucose to gluconolactone with the
More informationSuppl. Table 1: CV of pooled lipoprotein fractions analysed by ESI-MS/MS
Supplement VLDL LDL HDL PC 3.3 1.77 1.3 LPC 4.82 2.5.35 SM 3.1 4.6 1.92 CER 2.17 6.3 4.15 PE 3.18 1.93 2.79 PE-pl 13.18 1.9 2.32 CE 2.9.65.4 FC.36 3.5 2.54 Suppl. Table 1: CV of pooled lipoprotein fractions
More informationHumaTex Febrile Antigens
Design Verification HumaTex Febrile Antigens Contents 1 Function... 2 2 Product Line... 2 3 Sensitivity... 2 Description of Control Materials... 2 Kit controls Positive and Negative... 3 4 Interferences...
More informationApplicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.
Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.
en CHOLESTEROL 7D62 304796/R02 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. Read Highlighted Changes Revised October 2012 Package insert
More informationIMTEC-ANA-LIA MAXX. Design Verification
Design Verification IMTEC-ANA-LIA MAXX CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Specificity... 2 3 Interferences... 4 4 Imprecision... 6 4.1 Within-Run Imprecision... 6 4.2 Between-Run
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationhc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine
Design Verification hc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine 1 Function... 2 2 Sensitivity and Dynamic Range... 2 2.1 Analytical Sensitivity... 2 3 Validation of Efficacy... 3 3.1
More informationHigh-density Lipoprotein Cholesterol (HDL-C) Assay Kit
(FOR RESEARCH USE ONLY. DO NOT USE IT IN CLINICAL DIAGNOSIS!) High-density Lipoprotein Cholesterol (HDL-C) Assay Kit (Double reagents) Catalog No: E-BC-K221 Method: Colorimetric method Specification: 96T
More informationINSTRUCTIONS FOR USE
VITROS Chemistry Products Reagent 680 1728 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry
More informationEvaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser
Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser
More informationEvaluation of Calculated Low-Density Lipoprotein Against a Direct Assay
Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay Navapun Charuruks MD*, Anna Milintagas BSc** * Department of Laboratory Medicine, Faculty of Medicine, Chulalongkorn University **
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.
CHOLESTEROL 7D62-21 304342/R1 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System. NOTE: This package insert must be read
More informationWhy do we need SD LipidoCare?
SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationHDL CHOLESTEROL. 01 English - Ref.: 13. Ref.:13. Insert
HDL CHOLESTEROL Insert Ref.:13 Intended use. System for selective precipitation of Low and Very Low Density Lipoproteins (LDL and VLDL) and HDL cholesterol measurement in the supernatant, by an end point
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationDesign Verification. Form:
Design Verification SYPHILIS TPHA liquid CONTENTS 1 Function... 2 2 Sensitivity... 2 Preparation of Serum Control Panel... 2 Kit Controls Positive and Negative... 3 3 Traceability... 3 4 Specificity and
More informationLDL Liquiform. 01 English - Ref.: 111. Ref.:111. Insert. Intended use. Test principle. Methodology. Reagents Precautions and warnings
LDL Liquiform Insert Ref.:111 Intended use. System for direct homogenous determination of LDL cholesterol in serum and plasma Test principle. On the first stage of the reaction, a specific surfactant present
More informationUREA CE Insert. 01 English - Ref.: 27. Ref.:27
UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]
More informationKinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase
Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,
More informationG. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626
Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi
More informationLIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE
LIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE Revised 3/2014 LIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE For questions concerning protocol and assay development, validation of results, etc., please
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,
More informationGeneral Chemistry Scheme Guide
General Chemistry Scheme Guide Copyright WEQAS. All rights reserved. No part of this document may be reproduced or utilised in any form without permission from WEQAS Contents. Scheme details and repertoire.....
More informationFor in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date
AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2
More informationMaterials and Methods
Materials and Methods The study was conducted at School of Health Sciences, Kannur University Thalassery Campus, Palayad, and Pushpagiri Institute of Medical Sciences and Research Center, Thiruvalla, Kerala,
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationCholesterol (blood, plasma, serum)
1 Cholesterol (blood, plasma, serum) 1 Name and description of analyte 1.1 Name of analyte Cholesterol (plasma; also blood, serum) 1.2 Alternative names 2,15-dimethyl-14-(1,5-dimethylhexyl)tetracyclo[8.7.0.0
More informationTABLE OF CONTENTS GENERAL INFORMATION... 1
BIOO RESEARCH PRODUCTS Glucose Assay Kit Manual Catalog #: 5611-01 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure Overview... 1 Required Materials
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More informationLDLC3. English System information For cobas c 311/501 analyzers:
Order information Analyzer(s) on which cobas c pack(s) can be used 07005717 190 LDL Cholesterol Gen.3 (200 tests) System ID 07 7565 7 Roche/Hitachi cobas c 311, cobas c 501/502 12172623 122 Calibrator
More informationLPO-CC (Lipid Peroxides) Kit
K-ASSAY LPO-CC (Lipid Peroxides) Kit For the Quantitative Colorimetric Determination of High Levels of Lipid Peroxides in Serum, Plasma and Other Biological Samples. Cat. No. CC-004 For Research Use Only.
More informationSmall dense low-density lipoprotein is a risk for coronary artery disease in an urban Japanese cohort: The Suita study
Small dense low-density lipoprotein is a risk for coronary artery disease in an urban Japanese cohort: The Suita study Hidenori Arai 1, Yoshihiro Kokubo 2, Makoto Watanabe 2, Tatsuya Sawamura 3, Tomonori
More informationCertificate of Analysis
Siemens Healthcare Diagnostics Products GmbH Emil-von-Behring-Str. 76 35041 Marburg Germany Customer To whom it may concern Certificate of Analysis Material: 10446446 Kit-: 47066 Total lot size: 445 PCK
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationPLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL
PLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL Lipids are characterized by low polarity and limited solubility in water. Their plasma concentration is about 500-600
More informationAlere Cholestech LDX Laboratory Procedure Lipid Profile GLU TC HDL GLU TC GLU Lipid Profile TC HDL TC
Laboratory Name: Laboratory Address: Date of this packet: Insert Revision: 26196en Rev. A 2010/12 Alere Cholestech LDX Laboratory Procedure Lipid Profile GLU TC HDL GLU TC GLU Lipid Profile TC HDL TC This
More informationPlasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana
Plasma Hemoglobin Determination: Two Manual Methods Compared An Honors Thesis (HONRS 499) by Anne M. Emerick Sharlene Strahl St. Francis Hospital and Health Centers Beech Grove, Indiana Dr. Nancy Behforouz
More informationEvaluation Report. Eurolyser PT (INR) test kit (ST0180) on CUBE and smart analysers
Evaluation Report Eurolyser PT test kit (ST0180) on CUBE and smart analysers Locations Location 1: Eurolyser Diagnostica GmbH Operator: Simone Wieser Date: 30.06.2015 Specimens The specimens used for analysis
More informationLipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives
ISPUB.COM The Internet Journal of Health Volume 8 Number 1 Lipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives P Igbinaduwa, B Igbinaduwa, I Oforofuo Citation P Igbinaduwa,
More informationTraceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.
Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff UK www.weqas.com Programme
More informationHbA1c enzymatic assay evaluation at AN academic
Peer reviewed ORIGINAL ARTICLE HbA1c enzymatic assay evaluation at AN academic laboratory B Phiri-Ramongane Mmed Chem Path, FCP Path(SA)-Chem Department of Chemical Pathology, Sefako Makgatho Health Sciences
More informationDYSLIPIDAEMIC PATTERN OF PATIENTS WITH TYPE 2 DIABETES MELLITUS. Eid Mohamed, Mafauzy Mohamed*, Faridah Abdul Rashid
Malaysian Journal of Medical Sciences, Vol. 11, No. 1, January 2004 (44-51) ORIGINAL ARTICLE DYSLIPIDAEMIC PATTERN OF PATIENTS WITH TYPE 2 DIABETES MELLITUS Eid Mohamed, Mafauzy Mohamed*, Faridah Abdul
More informationFree Glycerol Assay Kit (Colorimetric)
Product Manual Free Glycerol Assay Kit (Colorimetric) Catalog Number STA-398 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Glycerol is the backbone of Triglycerides
More informationSITA 100 mg (n = 378)
Supplementary Table 1. Summary of Sulfonylurea Background Therapy at Baseline and During the Treatment Period. Sulfonylurea at baseline, n (%) SITA 100 mg (n = 378) CANA 300 mg (n = 377) Total (N = 755)
More informationENZYMATIC CREATININE Insert
ENZYMATIC CREATININE Insert Ref.:17 Intended use. System for determination of creatinine in serum, plasma, and urine samples by end-point reaction. Professional use. [ Only for in vitro diagnostic use.]
More informationOxalate (urine, plasma)
Oxalate (urine, plasma) 1 Name and description of analyte 1.1 Name of analyte Oxalate 1.2 Alternative names 1.3 NLMC code To follow 1.4. Function of analyte Oxalate is a metabolic end product primarily
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationEvaluation of a Rapid Homogeneous Method for Direct Measurement of High-Density Lipoprotein Cholesterol
Clinical Chemistry / ORIGINAL ARTICLE Evaluation of a Rapid Homogeneous Method for Direct Measurement of High-Density Lipoprotein Cholesterol R. Scott Hubbard, MD,1 Shaina V. Hirany, MS,2 Sridevi Devaraj,
More informationJLR Papers In Press. Published on October 16, 2003 as Manuscript D JLR200
JLR Papers In Press. Published on October 16, 2003 as Manuscript D300024-JLR200 A method of direct measurement for the enzymatic determination of cholesterol esters Toshimi Mizoguchi 1, Toshiyuki Edano,
More informationBIOO LIFE SCIENCE PRODUCTS
BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit
More informationOriginal Article Interfering effect of bilirubin on the determination of alkaline phosphatase
Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,
More informationGentian Canine CRP Immunoassay Application Note for scil VitroVet*
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of
More informationAspartate Aminotransferase Test Code: AST
Aspartate Aminotransferase Test Code: AST Intended Use For in vitro diagnostic use only CARESIDE Aspartate Aminotransferase (AST) cartridges are used with the CARESIDE Analyzer to quantitatively measure
More informationPhosphorus Test Code: PHOS
Intended Use Phosphorus Test Code: PHOS For in vitro diagnostic use only CARESIDE cartridges are used with the CARESIDE Analyzer to quantitatively measure inorganic phosphorus in anticoagulated whole blood,
More informationFree Fatty Acid Assay Kit (Fluorometric)
Product Manual Free Fatty Acid Assay Kit (Fluorometric) Catalog Number STA-619 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Triglycerides (TAG) are a type of lipid
More informationDesign Verification Function... 2 Analytical sensitivity... 2 Specificity, cross-reactivity and interferences... 3
Design Verification 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity and interferences...
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationGlutathione Assay Kit
Glutathione Assay Kit Catalog Number KA1649 250 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...
More informationTurbidos. Design Verification. Contents
Design Verification Turbidos Contents 1 adjustment... 2 2 Target value determination... 2 3 Stability... 2 3.1 Accelerated stress data... 2 3.2 Open vial stability... 3 3.3 Real Time Stability... 4 4 Result
More informationHDL LE. 01 Inglês - Ref.: 98. Ref.:98. Insert. Methodology. Intended use. Reagents. Test principle Precautions and warnings. Summary.
HDL LE Insert Ref.:98 Intended use. For the quantitative determination of high-density lipoprotein cholesterol (HDL-C) in serum and plasma. Test principle. This method is based on accelerating the reaction
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationKatsuyuki Nakajima, PhD. Member of JCCLS International Committee
Katsuyuki Nakajima, PhD Member of JCCLS International Committee Visiting Professor and Scientist Tufts University, Boston, MA & Framingham Offspring Study, Framingham, MA August 20 th, 2011, Tokyo Framingham
More informationEFFECT OF DYKURE A POLYHERBAL FORMULATION ON TYPE-I AND TYPE-II DIABETES
EFFECT OF DYKURE A POLYHERBAL FORMULATION ON TYPE-I AND TYPE-II DIABETES Address for Correspondence: Harshul Parikh a, R. Balaraman b Prof. R. Balaraman Pharmacy Department Faculty of Tech. & Engg., The
More informationSupplementary Material 1. Statistical methods used to conduct power calculations.
Supplementary Material 1. Statistical methods used to conduct power calculations. Post-hoc power calculations and patient numbers needed to detect changes were conducted considering (i) the observed partial
More informationEvaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser
Evaluation Report Eurolyser Phenobarbital test kit (VT0211) on solo Analyser Location Location: Eurolyser Diagnostica GmbH Operator: Bettina Schönauer Date: February 2016 - March 2016 Specimens The specimens
More informationHDL LE Insert. 01 English - Ref.: 98. Ref.:98
HDL LE Insert Ref.:98 Intended use. For the quantitative determination of high-density lipoprotein cholesterol (HDL-C) in serum and plasma. Professional use. [For in vitro diagnostic use.] Test principle.
More informationEvaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples
Evaluation of the EUROLYSER-SOLO Total T4 assay for veterinary samples Dr. rer.-nat. Volker Schlüter Zacherlweg 18 D-82061 Neuried Date: 2011 April 27 th Scope of the study Goal of this study was the validation
More information* * : : : Final. (Automated Strip Test, Microscopy) Colour Specific Gravity Nil
LL - LL-ROHINI (NATIONAL REFERENCE 136235212 Age Unknown Gender Unknown 5/6/2017 103400AM 5/6/2017 105702AM 5/6/2017 35028M Ref By Final Swasth lus Health Basic anel URINE EXAMINATION, ROUTINE; URINE,
More informationCreatinine (serum, plasma)
Creatinine (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Creatinine 1.2 Alternative names None 1.3 Description of analyte Creatinine is a heterocyclic nitrogenous compound (IUPAC
More informationReference Material Institute for Clinical Chemistry Standards (ReCCS)
Reference Material Institute for Clinical Chemistry Standards (ReCCS) Certified Reference Material for Measurement of Total Cholesterol and Glycerides in Human Serum Certificate of Analysis Intended use
More informationSerum Triglyceride Quantification Kit (Colorimetric)
Product Manual Serum Triglyceride Quantification Kit (Colorimetric) Catalog Number STA-396 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Triglycerides (TAG) are a type
More informationab Sphingomyelin Assay Kit
ab133118 - Sphingomyelin Assay Kit Instructions for Use For the specific, sensitive and convenient method of quantifying Sphingomyelin in plasma or serum. This product is for research use only and is not
More informationTechnical Bulletin. Cholesterol Reference Method Laboratory Network (CRMLN) Overview. TB Rev. 0
Technical Bulletin Reference Method Laboratory Network (CRMLN) Overview TB000020 Rev. 0 Reference Method Laboratory Network (CRMLN) - General Information With guidance from the US Centers for Disease Control
More informationCLINICAL CHEMISTRY REAGENTS. Product Profile
Product Profile Why Clinical Chemistry Reagents? Quantitative determination of specific analytes associated with various types of disease. Diagnosis Identifying a disease already present Prognosis Forecasting
More information15/9/2017 4:23:00PM 15/9/2017 4:26:23PM 20/9/2017 5:00:25PM A/c Status. Test Name Results Units Bio. Ref. Interval < >40.00 mg/dl <150.
Lab No 135091257 Age 30 Years Gender Male 15/9/2017 42300M 15/9/2017 42623M 20/9/2017 50025M Ref By UNKNWON Final Test Results Units Bio Ref Interval SWASTH LUS DIABETES ANEL LIID ROFILE, BASIC, SERUM
More information