LIPASE liquicolor. Design Verification. Multipurpose Reagent
|
|
- Roberta Horton
- 5 years ago
- Views:
Transcription
1 Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction Imprecision Linearity and Detection Limit Linearity Detection Limit Recovery of Control Sera Comparison of Methods Stability, Accelerated Stress Linearity Recovery of Control Sera Real Time Stability Recovery of Control Sera Open Vial Stability Linearity Recovery of Control Sera Interferences Traceability... 9 Form: /9 valid of
2 1 Introduction The performance characteristics of LIPASE liquicolor have been tested and documented in order to verify the clinical usefulness and compliance with the essential requirements of directive 98/79/EC. 2 Imprecision The imprecision (within run and day-to-day) of LIPASE liquicolor ([BUF] lot and [SUB] , exp date ) was calculated from six determinations on five consecutive days. Control sera with low, medium and high lipase levels were employed as sample material. The tests were performed on AU 400 system. The mean, standard deviation and coefficient of variation were calculated. Analyte concentration Intra-assay Inter-assay (U/l) SD (U/l) %CV SD (U/l) %CV The LIPASE liquicolor reagent shows a good intra- and interassay precision in the low, middle, and high concentration range. 3 Linearity and Detection Limit 3.1 Linearity The linearity of LIPASE liquicolor (lot SA-115, exp date ) was controlled by employing a high concentrated pool serum successively diluted with phys. Saline. The analysed concentrations were compared with the theoretical concentrations obtained from a linear regression. Deviations of measured against theoretical concentrations are expressed in absolute and relative values. Linearity was assumed within an accepted deviation of ± 10%. As an example data measured on the AU 400 system are reported. High Pool Analytical Data Regressed Data Deviation from Regression Line Content (%) U/l U/l U/l (%) The LIPASE liquicolor test is linear up to an Lipase concentration of at least 300 U/l. Design Verification and Product Data for Lipase 2/9
3 3.2 Detection Limit The detection limit has been evaluated from a 20-fold repeat of a zero sample (saline). Lipase liquicolor reagent: [BUF] lot and [SUB] , exp date The mean value was U/l with a standard deviation of U/l. From these results the detection limit has been calculated acc. to the formula: abs. mean + 3 SD = detection limit. The detection limit resulted to: 0.32 U/l. 4 Recovery of Control Sera A number of commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. The determinations of each control serum have been performed with LIPASE liquicolor (lot SA-115, exp date ). The mean values have been calculated and compared with the target values of the respective control sera. CONTROL SERUM RECOVERY HUMAN fresh Control LOT Target Range Result Deviation to Fresh serum U/l U/l U/l (%) Precinorm # Precipath # C.fas # Autocal Chema # 9A Mean Comparison of Methods LIPASE liquicolor reagent (lot SA-115, esp date ) run on Olympus AU 400 has been compared against Lipase reagent Roche (lot , exp date ) measured on Olympus AU 400 analyzer. Results < 40 U/l have been excluded. Control sera as well patient samples have been employed in the comparison (N=39). The results have been evaluated by a non-parametric regression analysis acc. to Passing & Bablok. Acceptance criteria were r> 0.95 and a slope between 0.90 and 1.10, acceptance criteria ± 10%. The linear regression obtained can be described as follows: r = Y = X X mean = U/l Y mean = U/l Sample Mean Reference Mean Test No. Lipase (Roche) Lipase Design Verification and Product Data for Lipase 3/9
4 Mean Both methods showed a good agreement and no significant deviation could be observed with any specific sample. Design Verification and Product Data for Lipase 4/9
5 6 Stability, Accelerated Stress 6.1 Linearity The linearity of LIPASE liquicolor (lot SA-115, exp date ) was controlled according to the procedure already described in 3.1. The reagents were stressed 10 days at 37 C. As an example data measured on the AU 400 system are reported. High Pool Analytical Data Regressed Data Deviation from Regression Line Content (%) U/l U/l U/l (%) The test is specified to cover a linear range up to about 300 U/l as indicated in section 3.1. The above results clearly confirm that the test achieves its specification after accelerated stress. 6.2 Recovery of Control Sera A number of commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. The determinations of each control serum have been performed with LIPASE liquicolor (lot SA-115, exp date ). The mean values have been calculated and compared with the target values of the respective control sera. Deviation to Fresh acceptance criteria is ± 10%. Design Verification and Product Data for Lipase 5/9
6 CONTROL SERUM RECOVERY HUMAN fresh HUMAN 10d 37 C Control LOT Target Range Result Result Deviation to Fresh serum U/l U/l U/l U/l (%) Precinorm # Precipath # C.fas # Autocal Chema # 9A ,1 Mean Accelerated stress data are indicative for a real-time stability of 18 months at a storage temperature of C. 7 Real Time Stability 6.1 Recovery of Control Sera Commercially available control sera and patient sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. The determinations of each control and patient serum have been performed with LIPASE liquicolor test. The mean values obtained with fresh (lot # FA-081) and three different reagent lots (lot # SA-133, lot # SA-171, lot # SA-074) have been calculated and compared with the target values of the respective control sera and means established with fresh reagent lot. SERUM AND CONTROL RECOVERY / Lot# FA-081 Lot# SA-133 Lot# SA-115 Lot# SA-074 Real Time Fresh 18 months 22 months 24 months Target [U/l] Range [U/l] Result [U/l] Result [U/l] Result [U/l] Result [U/l] Precinorm # Precipath # Hum Asy Level # 858UN Hum Asy Level # 651UE Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient All results support the claimed stability of 18 months from production. Design Verification and Product Data for Lipase 6/9
7 8 Open Vial Stability LIPASE liquicolor open vial stability has been tested for 28 days on the AU 400 system. The calibration has been performed on the first day. 8.1 Linearity The linearity of LIPASE liquicolor was controlled according to the procedure already described in 3.1. with fresh reagent and reagent stored on AU 400 system for 28 days. Open vial stability was assumed by linearity testing within an accepted deviation from regression line of ± 10%. The results are summarised below: High Pool Analytical Data LIPASE liquicolor reagent fresh [BUF] lot and [SUB] , exp date Regressed Data Deviation from Regression Line Analytical Data LIPASE liquicolor reagent 28 d open vial [BUF] lot and [SUB] , exp date Regressed Data Deviation from Regression Content (%) U/l U/l U/l (%) U/l U/l U/l (%) Line 8.2 Recovery of Control Sera A number of commercially available control sera have been employed. The control sera have been reconstituted/prepared according to the manufacturer s instructions. The calibration has been done on the first day of measurements. The determinations of each control serum have been performed with LIPASE liquicolor fresh and after 28 days open vial ([BUF] lot and [SUB] , exp date ). The mean value has been calculated and compared with the fresh mean of the respective control sera (accepted deviation ± 10%). Calibration stability Open vial stability CONTROL SERUM RECOVERY LIPASE liquicolor LIPASE liquicolor LIPASE liquicolor Fresh 8 days opened 28 days opened Control Lot Target [U/l] Range [U/l] Result [U/l] Result [U/l] Deviation % Result [U/l] Deviation % SERODOS SERODOS Plus Precinorm Precipath TruN TruP Sum Mean The on board stability and calibration stability of LIPASE liquicolor is confirmed for a minimum of 28 days. The calibration has been performed on the first day. The calibration stability is confirmed for a minimum of 8 days. Design Verification and Product Data for Lipase 7/9
8 9 Interferences Interference by bilirubin. hemoglobin, triglycerides and ascorbic acid have been studied by adding known amounts of the potentially interfering substance to a known sample. Recoveries have been analysed according to the method of Glick et al. (Clin.Chem ). Criterion :Recovery within ± 10 % of initial value. The results are summarised in the following table. Lipemia Triglycerides mg/dl Mean Lipase U/l Recovery % Bilirubin Bilirubin mg/dl Mean Lipase U/l Recovery % Hemoglobin Haemoglobin mg/dl Mean Lipase U/l Recovery % Ascorbic Acid Ascorbic Acid mg/dl Mean Lipase U/l Recovery % Design Verification and Product Data for Lipase 8/9
9 Interferences < 10%: Hemolytic samples up to 400 mg/dl hemoglobin by Bilirubin no interference up to 50 mg/dl by Triglycerides no interference up to 1100 mg/dl by Ascorbic acid no interference up to 50 mg/dl 10 Traceability LIPASE liquicolor calibration is performed with AUTOCAL REF and traceable to an in-house calibrator. Design Verification and Product Data for Lipase 9/9
liquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationliquicolor (AMP Buffer, IFCC) Design Verification
Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...
More information1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...
More information3 Linearity and Detection Limit Linearity Detection limit... 4
Design Verification liquicolor 1 Introduction... 2 2 Imprecision, Reproducibility... 2 2.1 Data from internal study... 2 2.2 Date from external study (Cholesterol Reference Method Laboratory Network)...
More informationDesign Verification. Form: /5 Rev. 007 valid of
Design Verification g 1 Determination of Diagnostic Sensitivity and Diagnostic Specificity... 2 Crossreactivity... 3 Potentially interfering substances... 3 2 Imprecision... 4 3 Standardisation... 4 4
More informationTotal Thyroxine ELISA (T4)
Design Verification Total Thyroxine ELISA (T4) Contents 1 Assay Principle... 2 2 Imprecision... 2 Within-run Imprecision... 2 Between run Imprecision... 2 3 Comparison of Methods, Accuracy... 2 4 Linearity...
More informationDesign Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...
Design Verification IMTEC-TSH RECEPTOR-ANTIBODIES CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 2.1 Determination of Diagnostic Sensitivity and Diagnostic Specificity...
More informationHexagon PSA. Design Verification. Contents
Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation
More informationIMTEC-ANA-LIA MAXX. Design Verification
Design Verification IMTEC-ANA-LIA MAXX CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Specificity... 2 3 Interferences... 4 4 Imprecision... 6 4.1 Within-Run Imprecision... 6 4.2 Between-Run
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationHumaTex CRP. Design Verification. Contents
Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...
More informationForum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007
Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationhc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine
Design Verification hc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine 1 Function... 2 2 Sensitivity and Dynamic Range... 2 2.1 Analytical Sensitivity... 2 3 Validation of Efficacy... 3 3.1
More informationSerodos and Serodos plus
Design Verification Serodos and Serodos plus Contents 1 Value Adjustment... 2 2 Target Determination... 2 3 Stability... 2 Real-Time Stability... 3 Stability after Reconstitution... 4 Stability after Reconstitution
More informationGentian Canine CRP Immunoassay Application Note for scil VitroVet*
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationStandard Operating Procedure
Subject Alkaline Phosphatase C111 Index Number Lab-1504 Section Laboratory Subsection Regional Clinic/Affiliate Hospital Laboratories Category Departmental Contact Kamprud, Elizabeth Last Revised 4/12/2017
More informationBÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:
More informationDesign Verification. Form:
Design Verification SYPHILIS TPHA liquid CONTENTS 1 Function... 2 2 Sensitivity... 2 Preparation of Serum Control Panel... 2 Kit Controls Positive and Negative... 3 3 Traceability... 3 4 Specificity and
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationEvaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser
Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser
More informationEvaluation Report. Eurolyser PT (INR) test kit (ST0180) on CUBE and smart analysers
Evaluation Report Eurolyser PT test kit (ST0180) on CUBE and smart analysers Locations Location 1: Eurolyser Diagnostica GmbH Operator: Simone Wieser Date: 30.06.2015 Specimens The specimens used for analysis
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationIDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats
IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,
More informationBias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University
Bias in clinical chemistry Elvar Theodorsson EFLM and Linköping University Bias a controversial subject Different perspectives Reseachers Users Regulatory bodies Standardisation organisations Metrologists
More informationIGG Diluent NaCl 9 % (50 ml) System-ID English System information For cobas c 311/501 analyzers:
Order information Analyzer(s) on which cobas c pack(s) can be used 03507432 190 150 tests System-ID 07 6787 5 cobas c 311, cobas c 501/502 11355279 216 Calibrator f.a.s. Proteins (5 1 ml) Code 656 03121305
More informationMETHOD VALIDATION CASE
METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,
More informationGUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS
GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS Ferruccio Ceriotti Servizio di Medicina di Laboratorio, Ospedale San Raffaele, Milano F. Ceriotti, Milano, 27-11-2015 2 Preamble Most of the
More informationEvaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser
Evaluation Report Eurolyser Phenobarbital test kit (VT0211) on solo Analyser Location Location: Eurolyser Diagnostica GmbH Operator: Bettina Schönauer Date: February 2016 - March 2016 Specimens The specimens
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationG. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626
Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationTurbidos. Design Verification. Contents
Design Verification Turbidos Contents 1 adjustment... 2 2 Target value determination... 2 3 Stability... 2 3.1 Accelerated stress data... 2 3.2 Open vial stability... 3 3.3 Real Time Stability... 4 4 Result
More information25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL
25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL I. PRINCIPLE The role of vitamin D in bone and mineral metabolism was recognized from its first identification as a factor that could cure rickets. However,
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,
More informationOriginal Article Interfering effect of bilirubin on the determination of alkaline phosphatase
Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationUrine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator
December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF
More informationHumaTex Febrile Antigens
Design Verification HumaTex Febrile Antigens Contents 1 Function... 2 2 Product Line... 2 3 Sensitivity... 2 Description of Control Materials... 2 Kit controls Positive and Negative... 3 4 Interferences...
More informationACCESS hstni SCIENTIFIC LITERATURE
ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval
More informationREAD HIGHLIGHTED CHANGES
BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)
More informationPHOS Precipath U plus (10 x 3 ml, for USA) Code Precipath U plus (10 x 3 ml) Code 301
Order information Analyzer(s) on which cobas c pack(s) can be used 03183793 122 250 tests System ID 07 6614 3 Roche/Hitachi cobas c 311, cobas c 501/502 10759350 190 Calibrator f.a.s. (12 x 3 ml) Code
More informationDesign Verification Function... 2 Analytical sensitivity... 2 Specificity, cross-reactivity and interferences... 3
Design Verification 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity and interferences...
More informationEvaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer
Hindenberg et al. BMC Veterinary Research (2017) 13:146 DOI 10.1186/s12917-017-1065-9 RESEARCH ARTICLE Open Access Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra
More informationQMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole
More information3. System Accuracy Evaluation LUX Hgb test 1. Samples
3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More informationFarouk AL Quadan lecture 3 Implentation of
How to implement a program? Establish written policies and procedures Assign responsibility for monitoring and reviewing Train staff Obtain control materials Collect data Set target values (mean, SD) Establish
More informationHexagon OBTI. Design Verification. Contents
Design Verification Hexagon OBTI Contents 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity
More informationDiagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.
2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationKinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase
Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,
More informationAIA-600II Assay Specifications TgAb Test Code 048
AIA-600II Assay Specifications TgAb Test Code 048 Screen Item Data Screen 1 Lot ***Enter current cal. lot no. Cal 1 0 IU/mL Cal 2 2 IU/mL (example) Cal 3 4 IU/mL (example) Cal 4 10 IU/mL (example) Cal
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationEvidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA
Evidence Based Commutability: Bias 2 Study Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Australian Bias Studies conducted by Gus Koerbin, ACT Pathology on behalf of AACB Harmonisation Committee
More informationINTERNAL QUALITY CONTROL ( Q I C) QC)
EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)
More informationThe Integrated Cardiovascular Clinical Network CHSA
The Integrated Cardiovascular Clinical Network CHSA Roche cobas b 101 HbA1c and Lipids Evaluation July 2013 Introduction Cardiovascular disease (CVD) is the leading cause of mortality in Australia, accounting
More informationTSH Receptor Autoantibody ELISA
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it TSH Receptor Autoantibody ELISA Product Data Sheet Cat.
More informationVerification and validation of diagnostic laboratory tests in clinical virology
Journal of Clinical Virology 40 (2007) 93 98 Review Verification and validation of diagnostic laboratory tests in clinical virology Holger F. Rabenau a,, Harald H. Kessler b, Marhild Kortenbusch a, Andreas
More informationApplicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.
Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required
More informationThe analytical phase
The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+
More informationUREA CE Insert. 01 English - Ref.: 27. Ref.:27
UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]
More informationFor in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date
AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2
More informationValidation Case studies: the good, the bad and the molecular
Validation Case studies: the good, the bad and the molecular Patti Jones PhD Professor of Pathology UT Southwestern Medical Center Director of Chemistry Children s Medical Center Dallas Presented by AACC
More informationMethod Comparison Report Semi-Annual 1/5/2018
Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community
More informationDKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum
DKK-1 ELISA, Cat.No. BI-20413 For the quantitative determination of DKK-1 in human serum ASSAY CHARACTERISTICS Method Standard range Conversion factor Sample volume / well Incubation time, temp. Sensitivity
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationApplicable To Employees of the Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics laboratories.
Subject Calcium C311 Index Number Lab-8810 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Judy Dayton Last Revised 4/11/2017 References Required document for Laboratory
More informationCEDIA OPIATE APPLICATION BECKMAN COULTER AU5800
CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800 Catalog No. 100089, 100098, 1661248 Intended for the qualitative and semiquantitative determination of opiates in human urine For In Vitro Diagnostic and
More informationLDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings
LDL LD Insert Ref.:129 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. [For in vitro diagnostic use.] Test principle. The method
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationIDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats
IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction
More informationThe Virtues and Pitfalls of Implementing a New Test
The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations
More informationYY/T Translated English of Chinese Standard: YY/T PHARMACEUTICAL INDUSTRY STANDARD
Translated English of Chinese Standard: YY/T1518-2017 www.chinesestandard.net Buy True-PDF Auto-delivery. Sales@ChineseStandard.net PHARMACEUTICAL INDUSTRY STANDARD YY OF THE PEOPLE S REPUBLIC OF CHINA
More informationContents of the Kit. Instruction Manual B R A H M S anti-tg n LIA. B R A H M S Service
B R A H M S GmbH Neuendorfstr. 25 D-16761 Hennigsdorf Luminescence immunoassay (LIA) for the quantitative determination of autoantibodies to thyroglobulin (Tg) in human serum (Coated Tube System) Article
More informationA validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls
A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding
More informationInstructions for Use. Tg Antibody ELISA
Instructions for Use Tg Antibody ELISA Enzyme Immunoassay for the Quantitative Determination of Autoantibodies to Thyroglobulin (Tg) in Serum and Plasma I V D REF EA618/96 12 x 8 2 8 C DLD Gesellschaft
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationFor In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date
SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by
More informationCoping with Analytical Interferences
Coping with Analytical Interferences (Handling Icteric, Hemolytic and Lipaemic Samples) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements
More informationHbA1c (Human) ELISA Kit
HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationCommutability of Calibration and Control Materials for Serum Lipase
Clinical Chemistry 47:12 2108 2113 (2001) Enzymes and Protein Markers Commutability of Calibration and Control Materials for Serum Lipase Giampaolo Cattozzo, 1* Carlo Franzini, 2 and Gianvico Melzi d Eril
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationTechnical Bulletin No. 162
CPAL Central Pennsylvania Alliance Laboratory Technical Bulletin No. 162 cobas 6800 HCV Viral Load Assay - New Platform - June 1, 2017 Contact: Heather Habig, MLS (ASCP) CM, MB CM, 717-851-1422 Operations
More informationThis package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.
en CHOLESTEROL 7D62 304796/R02 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. Read Highlighted Changes Revised October 2012 Package insert
More information1. Reagent Store at 2-8 ºC.
LDL Insert Ref.:146 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. Professional use. [For in vitro diagnostic use.] Test principle.
More informationDelta Check Calculation Guide
Delta Check Calculation Guide National Technology 2017, All Rights Reserved By Senior Scientific Researcher, Asmaa Taher Table of Contents Definition... 2 Purpose... 2 Delta Check Research Studies... 2
More informationColorectal cancer - CEA
Colorectal cancer - CEA What is carcinoembryonic antigen? Human carcinoembryonic antigen (CEA) (MW = 175000 200000 g/mol) is a glycoprotein involved in cell adhesion. It is normally produced during fetal
More informationEvaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples
Evaluation of the EUROLYSER-SOLO Total T4 assay for veterinary samples Dr. rer.-nat. Volker Schlüter Zacherlweg 18 D-82061 Neuried Date: 2011 April 27 th Scope of the study Goal of this study was the validation
More informationGlutamine Kit. For the determination of glutamine and glutamate in human EDTA plasma and serum. For research use only K 7732.
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it Manual Kit For the determination of glutamine and glutamate
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationEXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids
DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More information