Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss*
|
|
- Beverley Hicks
- 5 years ago
- Views:
Transcription
1 Anaesthesia 15, 7, doi:1.1111/anae.138 Original Article Continuous monitoring of haemoglobin concentration after in-vivo adjustment in patients undergoing surgery with blood loss* D. Frasca, 1 H. Mounios, B. Giraud, M. Boisson, B. Debaene 3 and O. Mimoz 1 Assistant Professor, Senior Anaesthetist, 3 Professor and Head of Department, Professor and Head of Surgical Intensive Care Unit, Department of Anaesthesia and Intensive Care, University Hospital of Poitiers, Poitiers, France Summary Non-invasive monitoring of haemoglobin concentration provides real-time measurement of haemoglobin concentration (SpHb) using multi-wavelength pulse co-oximetry. We hypothesised that in-vivo adjustment using the mean of three haemoglobinometer (HemoCue â ) measurements from an arterial blood sample at the first SpHb measurement (HCueART) would increase the accuracy of the monitor. The study included 1 adults for a total of 173 measurements of haemoglobin concentration. In-vivo adjusted SpHb was automatically calculated by the following formula: in-vivo adjusted SpHb = unadjusted SpHb (SpHb HCueART). The accuracy of in-vivo adjusted SpHb was compared with SpHb retrospectively adjusted using the same formula, except for haemoglobin level which was assessed at the central laboratory and then compared with all other available invasive methods of haemoglobin measurement (co-oximetry, HbSAT; arterial HemoCue, HCueART; capillary HemoCue, HCueCAP). Compared with laboratory measurement of haemoglobin concentration, bias (precision) for unadjusted SpHb, in-vivo adjusted SpHb, retrospectively adjusted SpHb, HbSAT, HCueART and HCueCAP were. (1.),.3 (1.1),.3 (1.1),. (.7),. (.) and.5 (1.) g.dl 1, respectively. In-vivo adjustment of SpHb values using the mean of three arterial HemoCue measurements improved the accuracy of the device similar to those observed after a retrospective adjustment using central laboratory haemoglobin level.... Correspondence to: O. Mimoz o.mimoz@chu-poitiers.fr *Presented in part to the Societe Francßaise d Anesthesie et de Reanimation, September 13, Paris, France. Accepted: January 15 Introduction Accuracy of haemoglobin measurement remains an important determinant for transfusion decisionmaking, and can be obtained in several ways. Haemoglobin measurement at the central laboratory (HbLAB) is the gold standard but requires a blood sample and involves a time delay in obtaining the results. Pointof-care devices such as co-oximeters and haemoglobinometers have been developed to reduce this time but they also require a blood sample, are time-consuming, and do not provide a real-time assessment of change during the acute phase of blood loss. A non-invasive haemoglobin device (SpHb monitor; Masimo TM, Irvine, CA, USA) which provides real-time measurement of haemoglobin concentration has been introduced recently. Its performance has been evaluated in emergency departments, intensive care units and in the operating theatre with conflicting results [1 1]. Thus, the utility of SpHb for transfusion decision-making remains unproven. Improved 15 The Association of Anaesthetists of Great Britain and Ireland 83
2 Anaesthesia 15, 7, Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment accuracy was observed when SpHb was adjusted retrospectively using a laboratory measurement, suggesting that part of the bias between SpHb and the reference value is related to the patient [11, 1]. An in-vivo method for adjusting SpHb has been proposed recently by the manufacturer, but there is as yet no published validation in clinical practice. We hypothesised that in-vivo adjustment using the mean of three haemoglobinometer (HemoCue â ; Hb1, Angelholm, Sweden; HCueART) measurements from arterial blood sampled at the first SpHb measurement would increase the accuracy of the monitor. We also hypothesised that the accuracy of SpHb in vivo, adjusted using arterial HemoCue, which is available within a few minutes, would be similar to those of SpHb retrospectively adjusted using haemoglobin concentration measured at the central laboratory, which requires 3 min before the result is available. Finally, we compared the accuracy of SpHb adjusted in vivo with available invasive methods of haemoglobin measurements. Methods This prospective, observational study was conducted in the operating theatre at the University Hospital of Poitiers, France. After obtaining approval from the local ethics committee and written informed consent, adult patients undergoing elective major surgery with significant blood loss expected were recruited. Anaesthesia was induced with propofol.5 mg.kg 1 and sufentanil.3 mg.kg 1. Tracheal intubation was facilitated with rocuronium. mg.kg 1 and additional rocuronium administration was guided by neuromuscular monitoring during surgery. After induction of anaesthesia, an arterial catheter was placed in the radial artery to monitor blood pressure. Anaesthesia was maintained with isoflurane or desflurane and sufentanil. If necessary, a noradrenaline infusion was titrated to obtain a mean arterial pressure above 5 mmhg. Subjects were ventilated using volume-controlled mechanical ventilation (tidal volume: 8 ml.kg 1 ) with a mixture of oxygen and air (inspired oxygen fraction =.5). Respiratory rate was adjusted to maintain normocapnia. No patient received epidural or regional anaesthesia. Patients wore rainbow adult ReSposable sensors (R 5, Revision G) connected to a Radical-7 â Pulse co-oximeter, software version (Masimo), for continuous and non-invasive measurement of total haemoglobin concentration (SpHb), S p O, pulse rate and perfusion index (as an indicator of local perfusion). Sensors were applied to the patient following the directions for use provided by the manufacturer. This included the adequate application of the adhesive portion of the sensor so that the emitter and detector were precisely aligned on the finger. Sensors were covered with opaque shields to prevent optical interference per the manufacturer s directions for use. The sensor position was checked before every reading and readjusted if the adhesive portion became misaligned. If perfusion index was < 1%, the sensor was repositioned and recalibrated by switching the monitor off and on. For invasive assessment of haemoglobin level, arterial blood was drawn through a radial arterial catheter placed in the wrist contralateral to the SpHb sensor. Blood was collected into standard appropriate blood collection tubes. Reference haemoglobin values (HbLAB) were obtained by analysing arterial blood samples at the central laboratory using a Sysmex TM XT-1i automated haematology analyser (Roche TM Diagnostics, Paris, France). The confidence limits provided by the manufacturer for the Sysmex analyser are. g.dl 1. The same samples were also analysed with a satellite co-oximeter (Siemens TM RapidPoint 5, Siemens, Munich, Germany; HbSAT) and a HemoCue. Concomitantly, the fourth drop of blood after skin puncture on the ear was taken for testing of capillary blood with the same HemoCue point-ofcare device (HCueCAP). The anaesthetist was blinded to all haemoglobin values except those of HCueART that was used for clinical care. The Radical-7 is self-calibrating. The Sysmex measures haemoglobin concentration by colorimetry using the cyanide-free, sodium lauryl sulphate method, and is calibrated daily according to the manufacturer s instructions and good laboratory practice. The Rapid- Point is calibrated daily under the control of the central laboratory. The HemoCue point-of-care device is factory calibrated using the cyano-methaemoglobin method and does not require recalibration. The first SpHb ( first SpHb) measurement was recorded after the device had been reporting SpHb 8 15 The Association of Anaesthetists of Great Britain and Ireland
3 Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment Anaesthesia 15, 7, values for at least 15 min. Concomitantly, a reference HbLAB value ( first HbLAB) and the mean of three HCue- ART ( first HCueART) were obtained from an arterial blood sample. In-vivo SpHb adjustment ( IVadjusted SpHb) was automatically done by the Radical-7 interface and the displayed SpHb value was obtained using the following formula: IVadjustedSpHb ¼ unadjusted SpHb ð first SpHb first HCueARTÞ where unadjusted SpHb is the unadjusted value obtained by the monitor at the time of reading. To validate our choice of using the mean of three HCueART as the reference value, the SpHb value was also retrospectively adjusted ( Radjusted SpHb) using the following formula: RadjustedSpHb ¼ unadjusted SpHb ð first SpHb first HbLABÞ Then, simultaneous recording of unadjustedsphb, IVadjustedSpHb, HbLAB, HbSAT, HCueART and HCue- CAP values were manually collected before surgical incision and when required by the anaesthesiologist. Measures ended after completion of the surgical procedure. The perfusion index and the use of vasopressors (noradrenaline and/or ephedrine) were also recorded. The primary objective of the study was to assess the influence of in-vivo adjustment using the mean of three arterial HemoCue on the accuracy of the SpHb monitor. Secondary objectives were to compare: (i) the accuracy of the monitor after in-vivo and retrospective adjustment using haemoglobin concentration measured at the central laboratory; and (ii) the accuracy of the monitor after in-vivo adjustment to those of several invasive methods of haemoglobin monitoring. The sample size was set to a minimum of 15 measurements ( patients each with 3 measurements) according to Bland and Altman [13 15] recommendations for a precision of.3 SD of the 95% CI of the limits of agreement. Agreement between HbLAB (reference method) and haemoglobin values provided by the test devices was assessed as described by Bland and Altman [13 15]. Multiple haemoglobin measurements per patient provided unequal numbers of replicated data in pairs. Therefore, Bland and Altman analysis for repeated measures per subject as reported earlier [1, 15] was performed to calculate test method bias (error), SD (precision) and limits of agreement with 95% CI. Multiple mixed effects regression analysis using a general linear model was performed to evaluate the association of patients characteristics to SpHb biases. Variables analysed were as follows: age; sex; blood loss; fluid expansion; and use of vasopressor (noradrenaline). In addition, outliers for all haemoglobin measurements were calculated as difference value with the reference method greater than 1 or g.dl 1. The differences in outliers were analysed by the chi-squared test. Statistical analysis was performed with R 3.. (R Foundation, Vienna, Austria) with a significance level for two-tailed tests at p <.5. Results The study enrolled patients, of whom one was excluded owing to the inability to obtain a SpHb signal with a PI 1%. The characteristics of the 1 remaining patients are shown in Table 1. There were no adverse events associated with the use of SpHb sensors and shields. At the time of SpHb adjustment, mean (SD) unadjustedsphb, HbLAB, HbSAT, HCueART and HCueCAP values were 1.7 (.9), 1. (.), 1. (.7), 1. (.5) and 1.9 (1.) g.dl 1, respectively. Mean (SD) difference between first HCueART and first HbLAB was. (.1) g.dl 1. A median (IQR [range]) of 3 ( Table 1 Characteristics of the 1 patients. Values are median (IQR [range]) or number (proportion). Age; years 5 (5 7 [3 83]) Male 9 (7%) Surgery Total hip or knee 1 (%) replacement (redux) Hepatectomy 9 (%) Spinal surgery 5 (1%) Aortic aneurysm repair 5 (1%) Nephrectomy 3 (7%) Prostatectomy (5%) Meningioma 1 (%) Blood loss; ml 5 (3 1 [1 3]) Red cell transfusion 1 (39%) Fluid expansion; ml.h 1 95 (75 15 [5 175]) Use of vasopressor 1 (%) during procedure 15 The Association of Anaesthetists of Great Britain and Ireland 85
4 Anaesthesia 15, 7, Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment [ 8]) measurements was carried out per patient, leading to a total of 173 haemoglobin measurements. During the surgical procedure, HbLAB values dropped below 1 g.dl 1 in % of measurements and below 8 g.dl 1 in < 1% of measurements. Of the 1 patients, 1 (%) received vasopressors during surgery for a total of 3 measurements (5% of all measurements). Compared to the HbLAB reference method, bias, precision and 95% limits of agreement for undajdustedsphb, IVadjustedSpHb, RajustedSpHb, HbSAT, HCueART and HCueCAP are given in Table and corresponding Bland and Altman plots are presented in Fig. 1. Using the first arterial HemoCue instead of the mean of three arterial HemoCue to adjust SpHb values resulted in a non-significant deterioration of bias (. g.dl 1 vs.3 g.dl 1 ) and precision (1. g.dl 1 vs 1.1 g.dl 1 ) of the device. In multivariate analysis, there was no influence of tested covariates (including perfusion index and use of vasopressors) on unadjustedsphb and IVadjusted SpHb bias. Percentages of outliers are indicated in Table 3. Discussion This study is the largest one to assess the influence of in-vivo adjustment using the mean of three arterial HemoCue measurements on the accuracy of SpHb measurements in patients requiring elective surgery with a high risk of bleeding. This method of in-vivo adjustment was associated with an improvement of SpHb accuracy similar to those obtained after retrospective adjustment using haemoglobin assessment at the central laboratory. In-vivo adjustment is a new feature available on the Radical-7 monitor. It allows the clinician to adjust the initial SpHb measurement manually to match a value of his/her preferred reference device. Thereafter, the SpHb value is automatically adjusted by the user-entered bias. In a study conducted in surgical patients requiring 9 haemoglobin measurements, SpHb values were adjusted by subtracting the difference between the first SpHb and the first laboratory value from all subsequent SpHb values. SpHb bias (precision) was. (1.5) g.dl 1 before and.7 (1.) g.dl 1 after retrospective adjustment [11]. In a second study including 19 adult surgical patients requiring 73 measurements, both the bias (.8 g.dl 1 vs.1 g.dl 1 ) and precision (1. g.dl 1 vs.77 g.dl 1 ) improved after in-vivo adjustment [1]. In this latter study, adjustment was made using haemoglobin concentration measured with an automated gas analyser. These devices may be subject to significant intra-device and inter-device variability, greater than those of central laboratory analyser. Furthermore, the observed improvement in terms of coefficient of linear correlation (from.8 to.95) was not statistically tested. In this study, in-vivo adjustment improved SpHb precision to.3 g.dl 1 without significant impact on SpHb bias, which remained close to g.dl 1. The absence of significant effect on bias may be explained by the random changing direction of the difference between SpHb and the reference value. To expedite calibration of the monitor, we demonstrated that using the average of three values of HCueART allows the equivalent accuracy to the use of the laboratory value in < min. The use of Table Bias, precision and 95% limits of agreement for tested methods compared to reference method (based on 173 measurements). Values are number (95% CI). Bias; g.dl 1 Precision; g.dl 1 95% limits of agreement*; g.dl 1 unadjustedsphb. (.5 to.3) 1. (1. 1.5) 3. ( 3. to 3.) to. (..5) IVadjustedSpHb.3 (.3 to.3) 1.1 (.9 1.). (.5 to.) to 1.8 ( ) RadjustedSpHb.3 (.3 to.3) 1.1 (.9 1.). (.5 to.) to 1.8 ( ) HbSAT. (. to.).7 (..8). (. to.) to.7 (.7.7) HCueART. (..). (.3.).8 (.8 to.8) to.8 (.8.8) HCueCAP.5 (. to.) 1. (1. 1.5).9 ( 3. to.8) to 1.9 (1.8.1) unadjustedsphb, unadjusted continuous non-invasive haemoglobin measurement; IVadjustedSpHb, SpHb adjusted in vivo; RadjustedSpHb, SpHb adjusted retrospectively; HbSAT, satellite co-oximeter; HCueART, haemoglobinometer (arterial); HCueCAP, haemoglobinometer (capillary). * Adjusted for multiple measurements per patient The Association of Anaesthetists of Great Britain and Ireland
5 Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment Anaesthesia 15, 7, (a) (b) HbLAB - unadjusted SpHb (g.dl 1 ).. 3. HbLAB - IVadjusted SpHb (g.dl 1 ) Mean of HbLAB and unadjusted SpHb (g.dl 1 ) Mean of HbLAB and IVadjusted SpHb (g.dl 1 ) (c) (d) HbLAB - Radjusted SpHb (g.dl 1 ) HbLAB - HbSAT (g.dl 1 ).7.. (e) Mean of HbLAB and Radjusted SpHb (g.dl 1 ) (f) Mean of HbLAB and HbSAT (g.dl 1 ) HbLAB - HCueART (g.dl 1 ).8..8 HbLAB - HCueCAP (g.dl 1 ) Mean of HbLAB and HCueART (g.dl 1 ) Mean of HbLAB and HCueCAP (g.dl 1 ) Figure 1 Bland and Altman plot for (a) unadjusted continuous non-invasive haemoglobin measurement ( unadjusted SpHb), (b) SpHb adjusted in vivo ( IVadjusted SpH) or (c) retrospectively ( Radjusted SpHb), (d) satellite co-oximeter (HbSAT), (e) arterial haemoglobinometer (HCueART), and (f) capillary haemoglobinometer (HCueCAP), vs haematology analyser (HbLAB). Limits of agreement are adjusted for repeated measures from the same subject. Each point stands for one measurement. Mean bias is represented by the solid line and 95% limits of agreement by dashed lines. One hundred and seventy-three measurements were obtained. 15 The Association of Anaesthetists of Great Britain and Ireland 87
6 Anaesthesia 15, 7, Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment Table 3 Percentages of outliers defined as difference values between haemoglobin obtained by tested methods and the main laboratory equal to or higher than 1 g.dl 1 or g.dl 1 (n = 173 measurements). g.dl 1 unadjustedsphb 18 IVadjustedSpHb 8* 31* RadjustedSpHb 8* 3* HbSAT 1 HCueART < 1 HCueCAP 1 1 g.dl 1 unadjustedsphb, unadjusted continuous non-invasive haemoglobin measurement; IVadjustedSpHb, SpHb adjusted in vivo; RadjustedSpHb, SpHb adjusted retrospectively; HbSAT, satellite co-oximeter; HCueART, haemoglobinometer (arterial); HCueCAP, haemoglobinometer (capillary). *p =.1 (vs unadjusted SpHb). p =.1 (vs IVadjusted SpHb). only one arterial HemoCue moderately decreased the precision, but the difference was not clinically significant. The use of in-vivo adjustment when SpHb monitoring has the potential to improve patient care. Percentage of outliers between estimated and HbLAB haemoglobin values decreased from % with unadjustedsphb to 31% with IVadjusted SpHb for a difference over 1 g.dl 1, and from 18% to 8% for a difference over g.dl 1. These percentages were lower than those observed with capillary HemoCue, a method which is currently widely used in patients undergoing surgery to help with transfusion decisionmaking. These findings correlated well with the results of previous studies [11, 1] and may have clinical implications for the management of blood transfusion. In neurosurgery with SpHb monitoring [1], clinicians were able to initiate transfusions faster compared with physicians not using the technology, because they did not have to wait for a laboratory haemoglobin value. In-vivo adjusted SpHb may be even more relevant in such clinical settings, although large studies are required to confirm these interesting findings. Unfortunately, the accuracy of the SpHb device, even after in-vivo adjustment, remained lower than those observed with the use of arterial HemoCue or satellite co-oximeter. However, these techniques have their own limitations, including their invasiveness and inherent time delay for results. Moreover, their ability to detect a decrease in haemoglobin concentration earlier is lower than techniques with continuous monitoring of haemoglobin concentration. In our study, differences between in-vivo adjusted SpHb and actual haemoglobin values as high as 3 g.dl 1 were noted, although this observation was uncommon. Manufacturers should be encouraged to continue to improve the SpHb technology to make it even more accurate in the future, as they did with pulse oxygen saturation monitoring [17]. In conclusion, in-vivo adjustment using the mean of three arterial HemoCue measurements improves SpHb accuracy and facilitates faster transfusion decision-making. Prospective randomised trials should be considered to investigate further the clinical potential of in-vivo adjusted SpHb monitoring to reduce blood transfusions during surgery at high risk of bleeding. Competing interests The study was funded by University Hospital of Poitiers. Masimo Corp. supplied the Radical-7 monitors and the sensors used for measurements. The manufacturer had no input into the design or conduct of this study or in the decision to submit the manuscript for publication. DF and OM have received lecture fees and travel expenses from Masimo Corp. References 1. Gayat E, Bodin A, Sportiello C, et al. Performance evaluation of a noninvasive hemoglobin monitoring device. Annals of Emergency Medicine 11; 57: Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Critical Care Medicine 11; 39: Mimoz O, Frasca D, Medard A, Soubiron L, Debaene B, Dahyot-Fizelier C. Reliability of the HemoCue â hemoglobinometer in critically ill patients: a prospective observational study. Minerva Anestesiologica 11; 77: Coquin J, Dewitte A, Manach YL, et al. Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds. Critical Care Medicine 1; : Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. American Journal of Surgery 11; 1: Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue â with automated laboratory hemoglobin measurement. Anesthesiology 11; 115: Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesthesia and Analgesia 11; 11: The Association of Anaesthetists of Great Britain and Ireland
7 Frasca et al. Continuous monitoring of haemoglobin concentration using in-vivo adjustment Anaesthesia 15, 7, Applegate RL nd, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology 1; 11: Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. British Journal of Anaesthesia 13; 111: Vos JJ, Kalmar AF, Struys MMRF, et al. Accuracy of non-invasive measurement of haemoglobin concentration by pulse cooximetry during steady-state and dynamic conditions in liver surgery. British Journal of Anaesthesia 1; 19: Isosu T, Obara S, Hosono A, et al. Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients. Journal of Clinical Monitoring and Computing 13; 7: Miyashita R, Hirata N, Sugino S, Mimura M, Yamakage M. Improved non-invasive total haemoglobin measurements after in-vivo adjustment. Anaesthesia 1; 9: Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. Journal of Biopharmaceutical Statistics 7; 17: Bland JM, Altman DG. Agreed statistics: measurement method comparison. Anesthesiology 1; 11: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 198; 1: Lindner G, Exadaktylos AK. How noninvasive haemoglobin measurement with pulse co-oximetry can change your practice: an expert review. Emergency Medicine International 13; 13: Levrat Q, Petitpas F, Bouche G, Debaene B, Mimoz O. Usefulness of pulse oximetry using the SET technology in critically ill adult patients. Annales Francßaises d Anesthesie et de Reanimation 9; 8:. 15 The Association of Anaesthetists of Great Britain and Ireland 89
Comparison of haemoglobin measurement methods in the operating theatre
British Journal of Anaesthesia 111 (6): 946 54 (13) Advance Access publication 17 July 13. doi:.93/bja/aet252 Comparison of haemoglobin measurement methods in the operating theatre B. Giraud 1, D. Frasca
More informationMinimally invasive monitoring in patients under general anesthesia Vos, Jaap Jan
University of Groningen Minimally invasive monitoring in patients under general anesthesia Vos, Jaap Jan IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish
More informationThe specific transfusion trigger that is the threshold
A Comparison of Three Methods of Hemoglobin Monitoring in Patients Undergoing Spine Surgery Ronald D. Miller, MD, MS,* Theresa A. Ward, BSN, RN,* Stephen C. Shiboski, PhD, and Neal H. Cohen, MD, MPH, MS*
More informationAccuracy of the Masimo Pronto-7 system in patients with left ventricular assist device
Bruells et al. Journal of Cardiothoracic Surgery 2013, 8:159 RESEARCH ARTICLE Open Access Accuracy of the Masimo Pronto-7 system in patients with left ventricular assist device Christian S Bruells 1,3,
More informationAvailable online at ScienceDirect. Procedia Engineering 132 (2015 )
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 160 166 The Manufacturing Engineering Society International Conference, MESIC 2015 Continuous hemoglobin measurement
More informationREPORT DOCUMENTATION PAGE
REPORT DOCUMENTATION PAGE Form Approved OMB No. 0704-0188 Public reporting burden for this collection of information is estimated to average 1 hour per response, including the time for reviewing instructions,
More informationEfficacy of Pulse Co-oximeter in Hemoglobin Estimation: A Non Invasive Method
Original Article DOI: 10.21276/APALM.1558 Efficacy of Pulse Co-oximeter in Hemoglobin Estimation: A Non Invasive Method Jayalaxmi Yadav K 1 *, Bharat C 1, Shridevi SH 1 and Udayshankar 2 1 Department of
More informationAbstract. Research Article. The Journal of Medical Research 2018; 4(1): 10-15
The Journal of Medical Research 2018; 4(1): 10-15 Research Article JMR 2018; 4(1): 10-15 January- February ISSN: 2395-7565 2018, All rights reserved www.medicinearticle.com Received: 08-01-2018 Accepted:
More informationEmpowering the RT with New Noninvasive Monitoring Capabilities
Empowering the RT with New Noninvasive Monitoring Capabilities Thomas Lamphere BS, RRT, RPFT Executive Director, Pennsylvania Society for Respiratory Care Adjunct Instructor, Respiratory Care Program,
More informationCurrent Use of Noninvasive Hemoglobin Monitoring in Anesthesia
Curr Anesthesiol Rep (2014) 4:233 241 DOI 10.1007/s40140-014-0070-9 ADVANCES IN MONITORING FOR ANESTHESIA (TM HEMMERLING, SECTION EDITOR) Current Use of Noninvasive Hemoglobin Monitoring in Anesthesia
More informationPermanent City Research Online URL:
Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement
More informationNon-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis
Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Gal S, Riskin A, Chistyakov I, Shifman N, Srugo I, and Kugelman A Pediatric Department and Pediatric Pulmonary Unit Bnai Zion
More informationThe accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems
doi:10.1111/j.1365-2044.2007.04987.x APPARATUS The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems D. Bolliger, 1 L. A. Steiner, 2 J. Kasper, 1 O.
More informationCLINICAL INVESTIGATIONS INTRA-OPERATIVE POINT OF CARE HAEMOGLOBIN ESTIMATION: A COMPARISON OF THREE METHODS
Sri Lankan Journal of Anaesthesiology 18(1) : 15-19 (20) CLINICAL INVESTIGATIONS INTRA-OPERATIVE POINT OF CARE HAEMOGLOBIN ESTIMATION: A COMPARISON OF THREE METHODS Nataraj Madagondapalli Srinivasan 1,
More informationTesting Masimo rainbow Technology with vpad-a1 1. Background
1 Testing Masimo rainbow Technology with vpad-a1 1. Background 1.1 Pulse Oximetry Pulse oximetery measures oxygen saturation or the ratio of oxyhemoglobin to the total amount of hemoglobin in the arterial
More informationAssessing Global Patient Status with the Halo Index
Whitepaper Halo: Assessing Global Patient Status with the Halo Index Summary Physiologic deterioration often occurs long before a patient crisis and manifests through subtle and often undetected changes
More informationDetermination of hemoglobin is one of the most commonly
ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning
More informationComparison of blood pressure measured at the arm, ankle and calf
doi:10.1111/j.1365-2044.2008.05633.x Comparison of blood pressure measured at the arm, ankle and calf C. Moore, 1 A. Dobson, 2 M. Kinagi 2 and B. Dillon 3 1 SpR Anaesthetics, 2 Consultant Anaesthetist,
More informationChandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology
Should NIRS be Standard Care for Pediatric CPB Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University Director of Pediatric Cardiac Anesthesiology Stanford Children s Hospital
More informationSetting The setting was tertiary care. The economic study appears to have been performed in Heidelberg, Germany.
Comparative analysis of costs of total intravenous anaesthesia with propofol and remifentanil vs. balanced anaesthesia with isoflurane and fentanyl Epple J, Kubitz J, Schmidt H, Motsch J, Bottiger B W,
More informationPATIENTS undergoing major abdominal or pelvic surgery
Evaluation of Pulse Cooximetry in Patients Undergoing Abdominal or Pelvic Surgery Richard L. Applegate II, M.D.,* Steven J. Barr, M.D., Carl E. Collier, D.O., James L. Rook, D.O., M.P.H., Dustin B. Mangus,
More informationCover Page. The handle holds various files of this Leiden University dissertation
Cover Page The handle http://hdl.handle.net/1887/46692 holds various files of this Leiden University dissertation Author: Zanten, Henriëtte van Title: Oxygen titration and compliance with targeting oxygen
More informationPOSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO. MD (ANAESTHESIOLOGY) FINAL EXAMINATION AUGUST 2011 Time : 1.00 p.m p.m.
POSTGRADUATE INSTITUTE OF MEDICINE UNIVERSITY OF COLOMBO MD (ANAESTHESIOLOGY) FINAL EXAMINATION AUGUST 2011 Date : 5 th August 2011 Time : 1.00 p.m. 4.00 p.m. Answer any three questions. Answer each question
More informationNon-invasive blood haemoglobin and plethysmographic variability index during brachial plexus block
British Journal of Anaesthesia 114 (5): 812 17 (2015) Advance Access publication 20 January 2015. doi:10.1093/bja/aeu484 Non-invasive blood haemoglobin and plethysmographic variability index during brachial
More informationAccuracy of Noninvasive Carboxyhemoglobin Measurements from the Rad-57: Analysis of a Recent Study by Touger et al
Whitepaper Accuracy of Noninvasive Carboxyhemoglobin Measurements from the Rad-57: Analysis of a Recent Study by Touger et al Summary The Rad-57 is a safe and effective device for noninvasive carboxyhemoglobin
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationEdwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor
Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation
More informationType of intervention Anaesthesia. Economic study type Cost-effectiveness analysis.
Comparison of the costs and recovery profiles of three anesthetic techniques for ambulatory anorectal surgery Li S T, Coloma M, White P F, Watcha M F, Chiu J W, Li H, Huber P J Record Status This is a
More informationDownloaded from UvA-DARE, the institutional repository of the University of Amsterdam (UvA)
Downloaded from UvA-DARE, the institutional repository of the University of Amsterdam (UvA) http://hdl.handle.net/11245/2.7621 File ID Filename Version uvapub:7621 77946n.pdf unknown SOURCE (OR PART OF
More informationPulse oximetry in the accident and emergency department
Archives of Emergency Medicine, 1989, 6, 137-142 Pulse oximetry in the accident and emergency department C. J. HOLBURN & M. J. ALLEN Accident and Emergency Department, Leicester Royal Infirmary, Leicester,
More informationOnline Supplementary Appendix
Online Supplementary Appendix This appendix has been provided by the authors to give readers additional information about their work. Supplement to: Lehman * LH, Saeed * M, Talmor D, Mark RG, and Malhotra
More informationDOCUMENT CONTROL PAGE
DOCUMENT CONTROL PAGE Title Title: UNDERGOING SPINAL DEFORMITY SURGERY Version: 2 Reference Number: Supersedes Supersedes: all other versions Description of Amendment(s): Revision of analgesia requirements
More informationNon Invasive methods of Hemoglobin
Non Invasive methods of Hemoglobin testing Dr. Ankit Mathur Consultant Transfusion Medicine & Transplant Immunology স প রভ ত 1 Donor selection First & the most important step in Blood Safety VNRBD Donor
More informationBiomedical Instrumentation D. The Photoplethysmogram
Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease
More informationPulse Oximeter Accuracy Study
Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709A Printed : 2014/5/26 This document is owned or controlled by Nihon Kohden and is protected by copyright law.
More information2 Philomeen Weijenborg, Moniek ter Kuile and Frank Willem Jansen.
Adapted from Fertil Steril 2007;87:373-80 Intraobserver and interobserver reliability of videotaped laparoscopy evaluations for endometriosis and adhesions 2 Philomeen Weijenborg, Moniek ter Kuile and
More informationCheck a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings
Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Mike McEvoy, PhD, RN, CCRN, NRP Professor Emeritus - Critical Care Medicine Albany Medical College Albany, New York Chair Resuscitation Committee
More informationANESTHESIA EXAM (four week rotation)
SPARROW HEALTH SYSTEM ANESTHESIA SERVICES ANESTHESIA EXAM (four week rotation) Circle the best answer 1. During spontaneous breathing, volatile anesthetics A. Increase tidal volume and decrease respiratory
More informationOxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood
Oxygen Saturation Monitors & Pulse Oximetry D. J. McMahon 141105 rev cewood 2017-11-14 2 3 Key Points Oxygen Saturation Monitors & Pulse Oximetry : Role of hemoglobin in respiration Basic principle of
More informationADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE
Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013
More informationPulse Oximeter Accuracy Study
Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24 This document is owned or controlled by Nihon Kohden and is protected by copyright law.
More informationClosed-loop Double-pump Automated System Manual Boluses
Closed-loop Double-pump Automated System versus Manual Boluses to treat Hypotension during Spinal Anaesthesia for Caesarean Section: randomised controlled trial Dr. Ban Leong SNG MBBS, MMED, FANZCA, FFPMANZCA,
More informationNational Emergency Laparotomy Audit. Help Box Text
National Emergency Laparotomy Audit Help Box Text Version Control Version 1.1 06/12/13 1.2 13/12/13 1.3 20/12/13 1.4 20/01/14 1.5 30/01/14 1.6 13/03/14 1.7 07/04/14 1.8 01/12/14 1.9 05/05/15 1.10 02/07/15
More informationpulse oximetry sensors 2007
pulse oximetry sensors 2007 Instruments and sensors containing Masimo SET technology are identifi ed with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure
More informationAssessing Agreement Between Methods Of Clinical Measurement
University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing
More informationSuggested items to be included in obstetric anaesthesia records
Suggested items to be included in obstetric anaesthesia records This list is intended as a guide to what fields could be included in an anaesthesia record used in obstetric practice. It is merely a suggested
More informationOriginal Article. Postoperative hypothermia and patient outcomes after major elective non-cardiac surgery. Abstract
Original Article doi:10.1111/anae.12129 Postoperative and patient outcomes after major elective non-cardiac surgery D. Karalapillai, 1 D. Story, 2 G. K. Hart, 3 M. Bailey, 4 D. Pilcher, 5 A. Schneider,
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO 2 ) 3 The Vigileo
More informationThe role of pulse oximetry in the accident and emergency department
Archives of Emergency Medicine, 1989, 6, 211-215 The role of pulse oximetry in the accident and emergency department M. A. LAMBERT & J. CRINNION Accident and Emergency Department, Leeds General Infirmary,
More information1. Influence of isoflurane exposure in pregnant rats on the learning and memory of offspring
1. Influence of isoflurane exposure in pregnant rats on the learning and memory of offspring Huang W, Dong Y, Zhao G, et al. BMC Anesthesiology 2018 18:5 Concerns remain about possible effects of general
More informationExercise Testing Interpretation in the Congenital Heart.
Interpretation in the Congenital Heart. Stephen M. Paridon, MD Medical Director, Exercise Physiology Laboratory The Children s Hospital of Philadelphia Professor of Pediatrics The University of Pennsylvania
More information10. Severe traumatic brain injury also see flow chart Appendix 5
10. Severe traumatic brain injury also see flow chart Appendix 5 Introduction Severe traumatic brain injury (TBI) is the leading cause of death in children in the UK, accounting for 15% of deaths in 1-15
More informationFacilitating EndotracheaL Intubation by Laryngoscopy technique and Apneic Oxygenation Within the Intensive Care Unit (FELLOW)
Facilitating EndotracheaL Intubation by Laryngoscopy technique and Apneic Oxygenation Within the Intensive Data Analysis Plan: Apneic Oxygenation vs. No Apneic Oxygenation Background Critically ill patients
More information6. Endovascular aneurysm repair
Introduction The standard treatment for aortic aneurysm, open repair, involves a large abdominal incision and cross-clamping of the aorta. In recent years, a minimally invasive technique, endovascular
More informationRespiratory Physiology & Neurobiology
Respiratory Physiology & Neurobiology 182 (2012) 88 92 Contents lists available at SciVerse ScienceDirect Respiratory Physiology & Neurobiology j o ur nal homep age : www.elsevier.com/locate/resphysiol
More informationComparison of safety and cost of percutaneous versus surgical tracheostomy Bowen C P R, Whitney L R, Truwit J D, Durbin C G, Moore M M
Comparison of safety and cost of percutaneous versus surgical tracheostomy Bowen C P R, Whitney L R, Truwit J D, Durbin C G, Moore M M Record Status This is a critical abstract of an economic evaluation
More informationPilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit
Article ID: WMC001137 Pilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit Author(s):Dr. Jonathan S. Jahr, MD, Dr. Robert A. Gunther, PhD, Dr. Bernd Driessen,
More informationMuscle OxyGen monitor. The Science Behind Moxy
Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the
More informationFloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations
Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy
More informationThoracic epidural versus patient-controlled analgesia in elective bowel resections Paulsen E K, Porter M G, Helmer S D, Linhardt P W, Kliewer M L
Thoracic epidural versus patient-controlled analgesia in elective bowel resections Paulsen E K, Porter M G, Helmer S D, Linhardt P W, Kliewer M L Record Status This is a critical abstract of an economic
More informationDr. D.McG.Clarkson Determining the Accuracy of SpO2 Values
Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements
More informationRESEARCH ARTICLE. Helmut Hager & Goutham Mandadi & Debra Pulley & J. Chris Eagon & Edward Mascha & Benjamin Nutter & Andrea Kurz
OBES SURG (2009) 19:717 724 DOI 10.1007/s11695-008-9607-7 RESEARCH ARTICLE A Comparison of Noninvasive Blood Pressure Measurement on the Wrist with Invasive Arterial Blood Pressure Monitoring in Patients
More informationEileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2
CHARLS Pilot: Blood-based Biomarker Documentation Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 June 2011 1 University
More informationOptimising tracheal intubation success rate using the Airtraq laryngoscope
doi:10.1111/j.1365-2044.2008.05757.x APPARATUS Optimising tracheal intubation success rate using the Airtraq laryngoscope G. Dhonneur, 1 W. Abdi, 2 R. Amathieu, 2 S. Ndoko 2 and L. Tual 2 1 Professor,
More informationHow Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve
SpO2 Monitoring Contents 1 Introduction 1 What is SpO 2? How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve 5 How Do I Use SpO2? Choosing a Sensor
More informationAgreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog
Veterinary Anaesthesia and Analgesia, 2016, 43, 199 203 doi:10.1111/vaa.12312 SHORT COMMUNICATION Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure
More informationMonitoring oxygen saturation by using pulse EVALUATION OF A NEW PULSE OXIMETER SENSOR
EVALUATION OF A NEW PULSE OXIMETER SENSOR By Marco Fernandez, RN, MSN, Kathy Burns, RN, BSN, MNEd, CCRN, Beverly Calhoun, RN, BSN, Saramma George, RN, BA, Beverly Martin, RN, BSN, and Chris Weaver, RN,
More informationAnaesthetic considerations for laparoscopic surgery in canines
Vet Times The website for the veterinary profession https://www.vettimes.co.uk Anaesthetic considerations for laparoscopic surgery in canines Author : Chris Miller Categories : Canine, Companion animal,
More informationORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers
ISSN 1447-4999 ORIGINAL RESEARCH Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers Dr Gyorgyi Kamaras MD, FCEM 1,2, Dr Tamas Geller MD 1, Dr Csaba Dioszeghy MD
More informationComparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.
Title Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients Author(s) Greenland, KB; Ha, ID; Irwin, MG Citation Anaesthesia,
More informationStandard Operating Procedure (SOP) Management of intervention group patients SOP 001
` Standard Operating Procedure (SOP) Management of intervention group patients SOP 001 Authors: Mark Edwards & Rupert Pearse Authorisation: Rupert Pearse (Chief Investigator) Scope To provide guidance
More informationANAESTHESIA EDY SUWARSO
ANAESTHESIA EDY SUWARSO GENERAL REGIONAL LOCAL ANAESTHESIA WHAT DOES ANESTHESIA MEAN? The word anaesthesia is derived from the Greek: meaning insensible or without feeling. The adjective will be ANAESTHETIC.
More informationSCIENTIFIC EVIDENCE CNAP HD (sold as NIBP100D/100D-HD by BIOPAC)
SCIENTIFIC EVIDENCE CNAP HD (sold as NIBP100D/100D-HD by BIOPAC) Validation & Papers Overview 29.10.2015, v1.8 CNAP HD sold as NIBP100D/NIBP100D-HD 1 RESEARCH BIOPAC NIBP100D 021 Uttarakha nd, India Telles
More informationSUBJECT INFORMATION AND CONSENT FORM
Title of Study: Department of Anesthesiology British Columbia s Children Hospital 4480 Oak Street Vancouver V6H 3V4 Tel 604 875 2711 Fax 604 875 3221 SUBJECT INFORMATION AND CONSENT FORM Mobile Health:
More informationHemoCue Hb METHOD AND SAMPLE COLLECTION
HemoCue Hb 201 + METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing a Haemoglobin (Hb) test using the HemoCue Hb 201 + analyser.
More informationJMSCR Vol 04 Issue 01 Page January 2016
www.jmscr.igmpublication.org Impact Factor 3.79 Index Copernicus Value: 5.88 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: http://dx.doi.org/10.18535/jmscr/v4i1.04 Haemodynamic Effects during Induction in
More informationMeasuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices
Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices By Zhang Feng, Senior Medical Applications Engineer Marten Smith, Medical Marketing Manager Medical Products Group Microchip
More informationConflicts of Interest
Anesthesia for Major Abdominal Cancer Resection John E. Ellis MD Adjunct Professor University of Pennsylvania johnellis1700@gmail.com Conflicts of Interest 1 Upper Abdominal Surgery Focus on oncologic
More informationProposed presentation of data for ICU-ROX.
Proposed presentation of data for ICU-ROX. Version 1 was posted online on 21 November 2017 (prior to the interim analysis which occurred when the 500 th participant reached day 28). This version (version
More informationCardiovascular Effects of Anesthesia for Cesarean Delivery in the Cardiac Patient
Cardiovascular Effects of Anesthesia for Cesarean Delivery in the Cardiac Patient Katherine W. Arendt, M.D. Associate Professor of Anesthesiology Mayo Clinic, Rochester, Minnesota Cardiac Problems in Pregnancy
More informationSupplementary Online Content
Supplementary Online Content Dhabangi A, Ainomugisha B, Cserti-Gazdewich C, et al. Effect of transfusion of red blood cells with longer vs shorter storage duration on elevated blood lactate levels in children
More informationExclusion Criteria 1. Operator or supervisor feels specific intra- procedural laryngoscopy device will be required.
FELLOW Study Data Analysis Plan Direct Laryngoscopy vs Video Laryngoscopy Background Respiratory failure requiring endotracheal intubation occurs in as many as 40% of critically ill patients. Procedural
More informationAbsolute Cerebral Oximeters for Cardiovascular Surgical Cases
Absolute Cerebral Oximeters for Cardiovascular Surgical Cases Mary E. Arthur, MD, Associate Professor, Anesthesiology and Perioperative Medicine Medical College of Georgia at Georgia Regents University
More informationJMSCR Vol 07 Issue 04 Page April 2019
www.jmscr.igmpublication.org Index Copernicus Value: 79.54 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: https://dx.doi.org/10.18535/jmscr/v7i4.76 A study to compare the antiemetic efficacy of ondansetron
More informationEuropean Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery
European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery INTRODUCTION The European Board of Anaesthesiology regards it as essential that certain core
More informationPermanent City Research Online URL:
Pal, S. K., Kyriacou, P. A., Kumaran, S., Fadheel, S., Emamdee, R., Langford, R. M. & Jones, D. P. (2005). Evaluation of oesophageal reflectance pulse oximetry in major burns patients. Burns, 31(3), pp.
More informationA comparison of fentanyl, sufentanil, and remifentanil for fast-track cardiac anesthesia Engoren M, Luther G, Fenn-Buderer N
A comparison of fentanyl, sufentanil, and remifentanil for fast-track cardiac anesthesia Engoren M, Luther G, Fenn-Buderer N Record Status This is a critical abstract of an economic evaluation that meets
More informationEndovascular Aneurysm Repair (EVAR)
Endovascular Aneurysm Repair (EVAR) Exceptional healthcare, personally delivered You have been diagnosed with an abdominal aortic aneurysm (AAA). This means that the main artery in your abdomen (the aorta)
More informationRevision of 10/27/2017 Form #280 Page 1 of 12 PVDOMICS STUDY Clinical Center Right Heart Catheterization (RHC) Results Form #280
Revision of 10/27/2017 Form #280 Page 1 of 12 PVDOMICS STUDY Clinical Center Right Heart Catheterization (RHC) Results Form #280 Instructions: Review PVDOMICS MOP Chapter 100 prior to completing right
More informationSepsis Wave II Webinar Series. Sepsis Reassessment
Sepsis Wave II Webinar Series Sepsis Reassessment Presenters Nova Panebianco, MD Todd Slesinger, MD Fluid Reassessment in Sepsis Todd L. Slesinger, MD, FACEP, FCCM, FCCP, FAAEM Residency Program Director
More informationEvidence-Based. Management of Severe Sepsis. What is the BP Target?
Evidence-Based Management of Severe Sepsis Michael A. Gropper, MD, PhD Professor and Vice Chair of Anesthesia Director, Critical Care Medicine Chair, Quality Improvment University of California San Francisco
More informationPostoperative hypothermia and patient outcomes after elective cardiac surgery
doi:10.1111/j.1365-2044.2011.06784.x ORIGINAL ARTICLE Postoperative hypothermia and patient outcomes after elective cardiac surgery D. Karalapillai, 1 D. Story, 2 G. K. Hart, 3,4 M. Bailey, 5 D. Pilcher,
More informationInfinity MCable -Masimo SET
Infinity MCable -Masimo SET Bring the advantages of Masimo s Signal Extraction Technology (SET ) to your pulse oximetry monitoring The noninvasive, motion-tolerant Infinity MCable -Masimo SET works with
More informationJoint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service.
Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Royal College of Radiologists British Society of Interventional Radiology The Vascular Society of Great
More informationStudy population The study population comprised patients who had undergone major abdominal surgery in routine care.
Evaluation of costs and effects of epidural analgesia and patient-controlled intravenous analgesia after major abdominal surgery. Bartha E, Carlsson P, Kalman S Record Status This is a critical abstract
More informationCorrelation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin
Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.
More informationA Prospective Pilot Study Comparing Nasal Blood Sampling and Venipuncture for the Assessment of Hemoglobin Levels and INR
The Laryngoscope VC 2016 The American Laryngological, Rhinological and Otological Society, Inc. A Prospective Pilot Study Comparing Nasal Blood Sampling and Venipuncture for the Assessment of Hemoglobin
More informationIMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY.
Clinical Evidence Guide IMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY. With the INVOS cerebral/somatic oximeter An examination of controlled studies reveals that responding to cerebral desaturation
More information