1 Fitting Decisions and their Impact on Hearing Aid User Benefit Mallory Maine, AuD Audiologist, GN ReSound
2 Agenda Common Fitting Oversights #1 Setting the coupler type in fitting software To set or not to set? #2 Taking fitting ranges at face value Where s the gain? #3 Reducing more than feedback Why calibrate? #4 Overprescribing directionality We can t get enough, but is it ever too much?
3 Oversight #1: Setting the Coupler Type in Fitting Software
4 Understanding the Display Open and closed configurations Vent effects
5 Understanding the Display Configured as Open with Dome Configured as Closed with No Vent
6 Investigation REMs of various RIE coupling options Hearing instruments on with low gain to visualize coupling effects on gain All hearing instrument features off and mics to omni 55dB speech noise
7 Real ear db SPL Effects on Gain open dome tulip dome power dome custom earmold Frequency (Hz)
8 LF Gain for Power Dome vs. Custom mold Muller & McBride, 2012
9 Findings We know that changing the physical coupling to the ear will effect primarily the low frequency gain.
10 Trivia Back to the fitting software What does it need to know? Question: Which will affect the hearing instrument gain settings? A. if it s an open or closed fitting B. vent size C. the exact coupling to the ear D. All of the above
11 Question: Which will affect the HI gain settings? Trick question because really it depends on the fitting software!
12 Investigation Coupler measurements of RIEs Compare gains of most open to most occluding vent settings available in software Set features off and mic to omni on all instruments Programmed with N3* audiogram Hz k 1.5k 2k 3k 4k 6k HTL *Bisgaard et al, 2010)
13 Gain for occluding option minus gain for open option (db) Changes in Closed to Open 20 Effect on hearing instrument gains of changing from occluding to open fitting: Approach Frequency (Hz) ReSound Manufacturer "Physical 1 Properties" Starkey Manufacturer "Acoustic 2 Options"
14 Gain for occluding option minus gain for open option (db) Changes in Open to Closed 20 Effect on hearing instrument gains of changing from open to occluding fitting: Approach Frequency (Hz) "ReSound Manufacturer reconfigure" 3
15 Gain for occluding option minus gain for open option (db) Changes in Open to Closed Effect on hearing instrument gains of changing from open to occluding fitting: Approach Frequency (Hz) Oticon Manufacturer "Acoustics" 4 Phonak Manufacturer "Acoustic 5 Parameters" "Siemens Manufacturer Acoustical 6 Parameters"
16 Findings Three approaches when change from open to closed: 1. Remains the same: gain was not affected 2. More gain for closed: prescribes more gain for closed across frequencies due to ability for usable LF gain and compensation for lost canal resonance 3. More low frequency gain for open- LF gain is boosted for open in an attempt to compensate for roll off and HF are reduced for canal resonance
17 Case Study Complaint FSW display: closed
18 Case Study Complaint It s very sharp sounding and I still don t feel like I m hearing well enough Problem: COUPLING The prescribed gains are different than what the patient is actually getting.
19 Conclusion: Configuration and Vent Effects Manufacturers are different in how these options are treated in fitting software so know your manufacturer No harm in setting them appropriately When in doubt, Verify!
20 Oversight #2: Taking Fitting Ranges at Face Value
21 How fitting ranges are calculated Ingredients: Electroacoustic specs in coupler Real-ear-to-coupler difference + MLE (CORFIG) Fitting prescription Directions: Add CORFIG to coupler FOG to get max insertion gain Plug audios into prescription Build in safety margin based on MSG and to allow VC use and fine-tuning Competitive environment what do other manufacturers publish? Point being.fitting ranges are not standardized
22 Investigation Compare MPO and fitting ranges of similar RIE products Fit with out of the box fitting Measure in coupler with 65dB input
23 Fitting ranges for similar RIE products FOG 46 db MPO 108 db SPL FOG 45 db MPO 108 db SPL FOG 45 db MPO 112 db SPL FOG 50 db MPO 113 db SPL
24 2cc gain (db) First Fit vs. NAL NL 2 40 Moderate audiogram, 65 db input = NAL NL 2
25 The Facts First fit matters more than the fitting range Six manufacturers fitting software did not meet NAL-NL1 or DSL i/o although their rationale was similar or the same. Even without the hearing aid being inserted in the ear, the prescriptive targets of the proprietary formulas are not matching NAL or DSL targets. Keidser, G., Brew, C., Peck, A. (2003). Proprietary fitting algorithms compared with one another and with generic formulas. The Hearing Journal (56). 3,
26 The Facts What does it matter if fitting rules are based on generic rules or not? Might get shorted on reserve gain Why such differences in first fit gain when the fitting rule is supposed to be based of generic rules? The fitting rule and non audiometric settings.
27 Example of Non-audiometric Settings: Experience Level A. Proprietary Experienced Non-linear (reference gains) B. Proprietary Experienced Linear Slightly HIGHER target gains C. Proprietary First time User Slightly LOWER target gains D. Proprietary Comfort User Much LOWER target gains
28 Findings Fitting ranges are not standardized Fitting ranges are only a small part of the fitting First fit varies greatly between manufacturers Fitting rationales maybe based on generic rules but are not the same and may have an impact on reserve gain Non audiometric settings can have a big impact on the targets and therefore the fitting
29 Conclusion: Fitting Ranges Remember to select the desired fitting rule in the FSW When looking at a fitting range be sure to not cut it too close Consider non audiometric setting effects When in doubt, verify!
30 Oversight #3: Reducing More than Feedback
31 The Basics Static: the feedback path created with the hearing instrument in the wearer s ear. Includes leaks, vents, ear canal, and device components Dynamic: the feedback paths created when the static path is altered as in using the phone or placing something near the hearing aid We need to manage both.
32 Feedback paths Closed ear mold Vented ear mold 1/2
33 Filters Constrained adaptation Kates, 1999
34 Filters Clamping If the actual feedback path measurement is not available A reference filter is applied and constraints based on the reference At times the reference constraints become inadequate When instability is detected gain is reduced
35 The Goal Abate feedback while maintaining gain and without artifact But how do you know if you are achieving the goal?
36 Investigation REIG measures completed with comparable 6 RIEs (n=40ears) Static assessment: Compared the amount of gain with feedback management on and off Dynamic assessment: Compared the amount of gain with feedback management and with and without phone to the ear
37 Now you see it now you don t
38 D1 D2 D3 D4 D5 D6
39 Findings Dynamic feedback can vary +/- 10dB from the static path Gain reduction (clamping) is used to eliminate dynamic feedback that is outside the constrained filter Some gain reductions are greater than others
40 Possible Complaints I hear quick bursts of feedback when the car door dings Inaccurate calibration Over prescribing feedback management settings I can hear fine normally but now that the settings were changed to control feedback I have really difficulty on the phone Over prescribing feedback management settings Using a system with greater amount of gain reduction
41 Conclusion: Calibration Calibrate the instrument when required or recommended by the manufacturer to create a proper template for the feedback management system Set the parameters in the FSW with caution and do not over prescribe to avoid artifact Measure to find out what the system is really doing
42 Oversight #4: Overprescribing Directionality
43 Two ears send distinct information to the auditory cortex of the brain to create a complete, rich and accurate portrayal of the auditory environment.
44 How the Brain Processes Sounds in the DETECT Environment CHOOSE A new sound in the environment draws your attention. Example: The sound of a candy wrapper during a lecture. Filtering out sounds until you hear only what you want to hear. Example: Your child s laugh amongst other children at play in a park. Top down Bottom up
45 User Preference: Binaural Omni & Directional Modes with Manual Switching 22.3% Omnidirectional 77.7% Directional Cord M, Surr R, Walden B & Olson L. (2002). Performance of directional microphones in everyday life. Journal of the American Academy of Audiology, 13:
46
47 Bentler et al, 2004 Results Results No significant difference between directional and Monofit fittings (p<.05) Cord et al (JAAA, 2007)- 12 subj, lab trial, in agreement
48 Let s review a few analogies
49 Some Collaborative Directionality Systems cut out the front and back view to listen on the side
50 And some Collaborative Directionality Systems cut out the front and sides to listen to the rear
51 Now, awareness of the entire environment to make informed choices
52 Okay now back to the hearing instrument and directionality
53 Dinner with Family and Friends Secondary Signal Primary Signal of Interest
54 Example A: Maximizing Audibility and/or SNR through Directionality Goal: Prioritizes the ear with the better speech signal when there is loud noise to one side. If there is a difference in signal-to-noise ratio at the two ears, the hearing aids will exaggerate the difference in gain and noise reduction at the two ears to enhance the ear with the best SNR.
55 Example B: Maximizing Audibility and/or SNR through Directionality Goal: Find the strongest speech and increase audibility and/or SNR The hearing instrument detects the direction of the dominant speech signal (front, side, or back) and automatically adjusts the focus direction when the signal source direction changes Possible configurations are bilateral front-facing directional, bilateral back-facing directional, or omni to one side with streaming of the sound on that side to the other ear
56 Findings Listeners prefer omni mic mode in most listening situations The auditory system needs awareness to sounds other than those that are in the look direction in order for the listener to choose the signal of interest Mic settings other than bilateral directional result in comparable SNR improvements along with environmental awareness
57 Possible complaints I am hearing the person next to me more than those in front I hear those in front very well but not those surrounding me I can t hear when someone comes up behind me I can hear the hearing aid changing in noise Add an omni program or an adaptive setting with omni abilities Set directional switching to a less aggressive parameter Asymmetric mic settings
58 Conclusion: Directionality For a basic or everyday program consider a setting that provides omni directional option Consider the patient's listening situation and intent for the program when prescribing directionality
59 Oversights and Solutions! #1 Setting the coupler type in fitting software To set or not to set? It depends on the manufacturer but it is safer to set than not because they could affect gain settings. #2 Taking fitting ranges at face value Where s the gain? Fitting range calculations are not standardized. Once the fitting rules and non-audiometric settings are factored in you might come up short. When looking at the fitting range leave plenty of reserve gain. #3 Reducing more than feedback Why calibrate? Calibration creates a template for feedback cancelation which will limit artifacts and clamping. Know your feedback manager to ensure large gain reductions are not taking place. #4 Overprescribing directionality We can t get enough, but is it ever too much? Yes, in certain cases default directionality is not ideal. Be sure to talk with the patient to assist in the selection and provide an omni setting for day to day use.
DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015.
www.bestsound-technology.com DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. First fit is an essential stage of the hearing aid fitting process and is a cornerstone of the ultimate
More informationHow to use AutoFit (IMC2) How to use AutoFit (IMC2)
How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain
More informationImportance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial
Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment
More informationAURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04
AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by
More informationThe Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2
The Devil is in the Fitting Details Why all NAL (or DSL) targets are not created equal Mona Dworsack Dodge, Au.D. Senior Audiologist Otometrics, DK August, 2012 Audiology Online 20961 What they share in
More informationAccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range
AccuQuest Spotlight: Successful Fittings with Oasis When fitting a client with hearing instruments, the experience should be positive for both the client and you, the hearing care professional (HCP). The
More informationManufacturers NAL-NL2 Fittings Fail Real-ear Verification
hearingreview.com http://www.hearingreview.com/2015/02/manufacturers-nal-nl2-fittings-fail-real-ear-verification/ Manufacturers NAL-NL2 Fittings Fail Real-ear Verification Research March 2015 Hearing Review
More informationEvidence base for hearing aid features:
Evidence base for hearing aid features: { the ʹwhat, how and whyʹ of technology selection, fitting and assessment. Drew Dundas, PhD Director of Audiology, Clinical Assistant Professor of Otolaryngology
More informationTop 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound
HVEM er HVEM Top 10 ideer til en god høreapparat tilpasning Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM 314363 WRITE DOWN YOUR TOP THREE NUMBER 1 Performing appropriate hearing assessment Thorough
More informationVerifying the Lyric Audibility Advantage
Field Study News Spring 2014 Verifying the Lyric Audibility Advantage John Pumford, Au.D. Manager of Clinical Research Phonak Lyric Introduction Since 2008, Lyric has presented people with mild to moderate
More informationA comparison of manufacturer-specific prescriptive procedures for infants
A comparison of manufacturer-specific prescriptive procedures for infants By Richard Seewald, Jillian Mills, Marlene Bagatto, Susan Scollie, and Sheila Moodie Early hearing detection and communication
More informationSingle channel noise reduction in hearing aids
Single channel noise reduction in hearing aids Recordings for perceptual evaluation Inge Brons Rolph Houben Wouter Dreschler Introduction Hearing impaired have difficulty understanding speech in noise
More informationAudiology Today MarApr2011
26 Stay on Target? By Jason A. Galster The focus on gain and targets has obscured the most important part of the real-ear measurement audibility. The topic of audibility is familiar; after all, a primary
More informationBeltone Solus Pro 1.9 Fitting Guide
Beltone Solus Pro 1.9 Fitting Guide Table of Contents Table of Contents... 2 Getting started... 3 Start Screen... 3 Assigning Devices... 4 Connection Process... 5 MSG Calibration... 5 Gain Adjustment...
More informationAbstract.
Combining Phase Cancellation, Frequency Shifting and Acoustic Fingerprint for Improved Feedback Suppression Josef Chalupper, Thomas A. Powers, Andre Steinbuss www.siemens.com Abstract Acoustic feedback
More informationClinical Tips for Matching Real-Ear Prescriptive Targets
AudiologyNOW! 2010 San Diego, CA Clinical Tips for Matching Real-Ear Prescriptive Targets Gabrielle Filips, Au.D., FAAA August Hernandez, Au.D., FAAA 1 Discussion topics Who s performing probe-microphone
More informationChildFit. Widex Baby. Compass quick guide
ChildFit When working with very young children, it is not always possible to obtain all of the information necessary for an accurate hearing aid fitting. To that end, Widex has developed the ChildFit procedure
More informationWalkthrough
0 8. Walkthrough Simulate Product. Product selection: Same look as estore. Filter Options: Technology levels listed by descriptor words. Simulate: Once product is selected, shows info and feature set Order
More informationIntroduction. What s new in Genie 2? New Fitting features. Support for new hearing aids:
Guide 207. Introduction Genie 2 and Genie have been updated with a variety of new features and enhancements. Please find a quick overview of these below: What s new in Genie 2? Support for new hearing
More informationBuilding Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees
Building Practice Success Through Speechmap Real Ear Measurements Chris Stokes-Rees Speaker Disclosure Relevant financial relationships: Audioscan (Applications Specialist) Conestoga College, Kitchener
More informationValidation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method
DSL v5.: A Presentation for the Ontario Infant Hearing Program Associates The Desired Sensation Level (DSL) Method Early development.... 198 Goal: To develop a computer-assisted electroacoustic-based procedure
More informationUser groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound.
P R O D U C T I N F O R M A T I O N S W I F T 0 + Fitting range db HL -0 0 0 0 0 0 0 Hz 0 00 0 000 00 00 General information Swift + is a very reliable and powerful programmable BTE instrument for people
More informationEvidence based selection of hearing aids and features
Evidence based selection of hearing aids and features Mark Laureyns Thomas More University College Department of Audiology Antwerp CRS - Amplifon Centre for Research & Studies Milan Italy European Association
More informationMuse Wireless CROS System
Clinical Validation of the Muse Wireless CROS System Ashley Hughes, Au.D., & Chad Parrish Introduction Fitting individuals with unilateral hearing loss (UHL) or asymmetrical hearing loss can pose unique
More informationReality Audiology: Insights from the pediatric real world
Reality Audiology: Insights from the pediatric real world Stacey Rich, MClAud How can we see the realities of clinical choices and real life use cases? Use logs are collected on HI HI use logs are read
More informationSMART HEARING. resoundpro.com. ReSound Aventa 3.9 fitting guide
SMART HEARING resoundpro.com ReSound Aventa 3.9 fitting guide This fitting guide gives an overview of how to fit ReSound LiNX 2 wireless hearing instruments with ReSound Aventa 3.9. A ReSound LiNX 2 962
More informationHearing Solutions Catalog Winter Supplement January 2009
Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with
More information2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures
Fitting Current Amplification Technology on Infants and Children Cindy Hogan, Ph.D./Doug Sladen, Ph.D. Mayo Clinic Rochester, Minnesota hogan.cynthia@mayo.edu sladen.douglas@mayo.edu AAA Pediatric Amplification
More informationAudiogram+: The ReSound Proprietary Fitting Algorithm
Abstract Hearing instruments should provide end-users with access to undistorted acoustic information to the degree possible. The Resound compression system uses state-of-the art technology and carefully
More informationQuick Guide Binaural REM
Quick Guide Binaural REM The purpose of this document is to provide a quick guide for the Binaural REM feature found in the REM440 Real-Ear Measurement module in the Affinity 2.0 and Callisto Suites. This
More informationTurn up the volume on life. The powerful choice for your super power users
Turn up the volume on life The powerful choice for your super power users Less of what holds them back 1 More gain without feedback Having a severe or profound hearing loss presents significant challenges
More informationEvidence-based Design Leads to Remote Microphone Hearing Instrument Technology
Copyright 2013 AudiologyOnline - All Rights Reserved Evidence-based Design Leads to Remote Microphone Hearing Instrument Technology Steve Hallenbeck, AuD, Jenny Groth, MA December 12, 2011 Introduction
More informationresoundpro.com PARTENRSHIPS Aventa 3 User Guide GUIDE FOR PROFESSIONALS
Aventa User Guide GUIDE FOR PROFESSIONALS resoundpro.com PARTENRSHIPS In this guide you will find: HOW TO START Airlink.... Other programming interfaces.... GOOD TO KNOW Start screen.... 5 Product selection....
More informationTesting Digital Hearing Aids
Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All
More informationAudiogram+: GN Resound proprietary fitting rule
Audiogram+: GN Resound proprietary fitting rule Ole Dyrlund GN ReSound Audiological Research Copenhagen Loudness normalization - Principle Background for Audiogram+! Audiogram+ is a loudness normalization
More informationAudiological Bulletin no. 46
Audiological Bulletin no. 46 Fitting Passion 115 with Compass V4 News from Audiological Research and Communication 9 502 1119 001 10-07 2 This bulletin describes the four main steps in fitting Passion
More informationAn assessment of the validity of a subjective hearing aid verification procedure
An assessment of the validity of a subjective hearing aid verification procedure Whitney M. Vineyard, B. A. AuD Research Project April, 211 University of Memphis Project Advisor Robyn M. Cox, PhD. Purpose
More informationENZO 3D First fitting with ReSound Smart Fit 1.1
ENZO 3D First fitting with ReSound Smart Fit 1.1 This fitting guide gives an overview of how to fit ReSound ENZO 3D wireless hearing instruments with ReSound Smart Fit 1.1. A ReSound ENZO 3D 998 wireless
More informationPrescribe hearing aids to:
Harvey Dillon Audiology NOW! Prescribing hearing aids for adults and children Prescribing hearing aids for adults and children Adult Measure hearing thresholds (db HL) Child Measure hearing thresholds
More informationPhonak Naída TM. Product information
Phonak Naída TM Product information V Phonak Naída V provides enhanced hearing performance for people with severe to profound hearing loss. With 4 performance levels (9, 7, 5, ) and 3 models (RIC, SP,
More informationPhonak Sky V is the newest and most comprehensive portfolio exclusively designed for children of all ages with mild to profound hearing loss.
Phonak Sky TM Product information V Phonak Sky V is the newest and most comprehensive portfolio exclusively designed for children of all ages with mild to profound hearing loss. Powered by the innovative
More informationBest Practice Protocols
Best Practice Protocols SoundRecover for children What is SoundRecover? SoundRecover (non-linear frequency compression) seeks to give greater audibility of high-frequency everyday sounds by compressing
More informationbinax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way
www.bestsound-technology.com binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way Richard Schultz-Amling Sivantos, 2015. 1 Introduction Since the release
More informationSpeechZone 2. Author Tina Howard, Au.D., CCC-A, FAAA Senior Validation Specialist Unitron Favorite sound: wind chimes
SpeechZone 2 Difficulty understanding speech in noisy environments is the biggest complaint for those with hearing loss. 1 In the real world, important speech doesn t always come from in front of the listener.
More informationStep-by-Step RECD Guide
Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform
More informationInstructions for use. Genie 2 and Genie
Instructions for use Genie and Genie Table of content Introduction Introduction... 3 Intended Use.... 4 Installation Guide... 5 Genie Overview... 7 Genie Overview...8 Warnings...8 Definition of Product
More informationCan you hear me now? Amanda Wolfe, Au.D. Examining speech intelligibility differences between bilateral and unilateral telephone listening conditions
Examining speech intelligibility differences between bilateral and unilateral telephone listening conditions Amanda Wolfe, Au.D. Communicating on the telephone continues to be a point of difficulty for
More informationELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT
ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless
More informationElectro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook
Electro-Acoustic Stimulation (EAS) EAS is a general term that refers to the simultaneous delivery of electric stimulation (cochlear implant) and acoustic amplification (hearing aid) to the same ear. Naída
More informationWhy? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction
Directional Mics & Digital Noise Reduction Speech in Noise + Hearing Aids = Problems in Noise Why? Ted Venema PhD Recall: Two things we must do for hearing loss: 1. Improve audibility a gain issue achieved
More informationMistakes on Real Ear Measures Clinicians Often Make. Michael Valente
Mistakes on Real Ear Measures Clinicians Often Make Michael Valente There are two national guidelines (ASHA, 1998; AAA 2006) stating the need to use REM to verify the performance of hearing aids in order
More informationWIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38
THE WIDEX FITTING RATIONALE FOR EVOKE ERIK SCHMIDT, TECHNICAL AUDIOLOGY SPECIALIST WIDEX GLOBAL RESEARCH & DEVELOPMENT Introduction As part of the development of the EVOKE hearing aid, the Widex Fitting
More informationCounseling Patients on Fitting Options. Michael Valente
Counseling Patients on Fitting Options Michael Valente Review the Audiogram Computerized Audiogram Using AudBase into EMR Using SII- Counsel on Impact of Hearing Loss on Speech Recognition www.ssi.to
More informationTodd A. Ricketts, PhD
1 Todd A. Ricketts, PhD Associate Professor Vanderbilt Medical Center; and, Director, Dan Maddox Memorial Hearing Aid Research Laboratory Vanderbilt Bill Wilkerson Center For Otolaryngology and Communication
More informationReSound ENZO 3D fitting guide. A guide for professionals
ReSound ENZO 3D fitting guide A guide for professionals First fitting with ReSound Smart Fit TM 1.1 This fitting guide gives an overview of how to fit ReSound ENZO 3D wireless hearing instruments with
More informationDifferences in added stable gain between manufacturers, audiometric configurations, earmold styles, and frequency bands
Washington University School of Medicine Digital Commons@Becker Independent Studies and Capstones Program in Audiology and Communication Sciences 2008 Differences in added stable gain between manufacturers,
More informationProblem: Hearing in Noise. Array microphones in hearing aids. Microphone Arrays and their Applications 9/6/2012. Recorded September 12, 2012
Array microphones in hearing aids Dennis Van Vliet, AuD Senior Director of Professional Relations Starkey Hearing Technologies array [əˈreɪ] n 1. an impressive display or collection 2. (Military) an orderly
More informationpersonalization meets innov ation
personalization meets innov ation Three products. Three price points. Premium innovations all around. Why should a truly personalized fit be available only in a premium hearing instrument? And why is it
More informationBeltone Electronics 2601 Patriot Boulevard Glenview, IL U.S.A. (800)
Beltone Electronics 261 Patriot Boulevard Glenview, IL 626 U.S.A. (8) 621-1275 Beltone Canada 31 Supertest Road Toronto, Ontario M3J 2M4 Canada (416) 736-4444 www.beltone.com 21, Beltone Electronics 521993
More informationRECD-Measurements with Connexx 6.3 and Unity 2
RECD-Measurements with Connexx 6.3 and Unity 2 Vera Spitzlei, Corporate Audiology Siemens Audiologische Technik GmbH For internal use only / Copyright Siemens AG 2009. All rights reserved. Overview Why
More information2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments
Desktop Fitting Guide getting started This Guide provides you a detailed introduction to hearing instrument fitting with Phonak Target 2.0. Please also find the [News] in the Phonak Target start screen.
More informationHow to Use Connexx 8.3 Hearing Profile
How to Use Connexx 8.3 Hearing Profile Rebecca Herbig Au.D. Rebecca Herbig Au.D. As digital hearing aids become increasingly advanced, the spotlight often shines on signal processing features, software
More informationTopics in Amplification DYNAMIC NOISE MANAGEMENT TM A WINNING TEAM
July 2017 Topics in Amplification DYNAMIC NOISE MANAGEMENT TM A WINNING TEAM Speech in noise continues to be the most difficult situation reported by hearing aid users. The good news is that the development
More informationInsio: A new standard in custom instruments
Insio: A new standard in custom instruments www.siemens.com Abstract: Insio custom hearing instruments introduce novel features throughout the complete fitting process in order to obtain the best possible
More informationSonic Spotlight. SmartCompress. Advancing compression technology into the future
Sonic Spotlight SmartCompress Advancing compression technology into the future Speech Variable Processing (SVP) is the unique digital signal processing strategy that gives Sonic hearing aids their signature
More informationSimpler. There are enough mid-market hearing aids. So why make another one? you might ask...
Simpler There are enough mid-market hearing aids. So why make another one? you might ask... Because, until now, no-one could provide a single answer for the 7 out of 10 users who choose a mid-market device.
More informationTechnical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S.
Technical Topics Where did my gain go? Thin Tube Open Fit BTE Verification Summary of presentation by C. Staples and S. Aiken History of Thin Tubes: Growth of new BTE market: First quarter of 2006 market:
More informationC H A N N E L S A N D B A N D S A C T I V E N O I S E C O N T R O L 2
C H A N N E L S A N D B A N D S Audibel hearing aids offer between 4 and 16 truly independent channels and bands. Channels are sections of the frequency spectrum that are processed independently by the
More informationFor hearing aid noise reduction, babble is not just babble
For hearing aid noise reduction, babble is not just babble HELEN CONNOR SØRENSEN * AND CHARLOTTE T. JESPERSEN Global Audiology, GN ReSound A/S, Ballerup, Denmark Most modern hearing aids provide single-microphone
More informationBetter hearing is right here, right now.
Better hearing is right here, right now. Just as no two people are alike, no two patients hearing losses are the same. Now, with the addition of our full lineup of Wi Series custom styles to complement
More informationADHEAR The new bone-conduction hearing aid innovation
ADHEAR The new bone-conduction hearing aid innovation MED-EL has world-wide launched a new kind of hearing aid, ADHEAR, for people who have an hearing impairment and want to prevent surgery. This little
More informationReSound LiNX 3D The future is here a new dimension in hearing care
8PRODUCTS ReSound LiNX 3D The future is here a new dimension in hearing care In a series that began with the world s first Made for iphone hearing aid, ReSound LiNX 3D is the newest Smart Hearing aid from
More informationComparison of two newly introduced bone anchored hearing instruments in first-time users (Updated September 2010)
Comparison of two newly introduced bone anchored hearing instruments in first-time users (Updated September 2010) Steen Østergaard Olsen, Henrik Glad & Lars Holme Nielsen Research Laboratory Department
More informationProduct information. For more information visit
Product information Phonak Bolero V is the comprehensive BTE portfolio for mild to profound hearing losses. Phonak Bolero V includes three designs and four performance levels, including a noise generator
More informationInitial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit
Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit Harvey B. Abrams 1,2 Theresa Hnath Chisolm,2,1 Megan McManus 2 Rachel A.McArdle 1,2 1 Bay Pines VA Healthcare
More informationTOPICS IN AMPLIFICATION
August 2011 Directional modalities Directional Microphone Technology in Oasis 14.0 and Applications for Use Directional microphones are among the most important features found on hearing instruments today.
More information[Connexx 7.2 with RexFit 7.2]
[Connexx 7.2 with RexFit 7.2] Programming Guide An easy fit is a successful fit. [Connexx 7.2] Table of Contents First Fit Guide... 4 First Fit Adjustment... 5 Firmware Updates... 6 Follow-Up Fittings
More informationAre hearing aids the better rehabilitative choice when compared to PSAPs? On speech-intelligibility and soundquality,
Are hearing aids the better rehabilitative choice when compared to PSAPs? On speech-intelligibility and soundquality, the answer is yes. Filip M. Rønne 12 & Rikke Rossing 1 1 Eriksholm Research Centre,
More informationHow to select the optimum BTE fitting option
How to select the optimum BTE fitting option Page 2 I Rationales for fitting BTE instruments Page 2 II Wide range of BTE fitting options Life BTE Active BTE S-Model BTE Page 4 III Selecting the optimum
More informationIntroduction. What s new in Genie 2?
Guide 206.2 Introduction Genie 2 and Genie have been updated with a variety of new features and enhancements. Please find a quick overview of these below: What s new in Genie 2? Support for new hearing
More informationPhonak Brio 3 Winter 2018
Phonak Brio 3 Winter 2018 The Brio 3 comprehensive product portfolio for mild to profound hearing loss includes RIC, BTE and Custom products. Phonak Brio 3 hearing aids are ideal for members who demand
More informationPOWERFUL PERSPECTIVES
RUBRIK POWERFUL PERSPECTIVES SUPREMIA 7 3 1 RUBRIK 2 RUBRIK How we experience the world depends a lot on our perspective. Sometimes, changing our perspective changes our interpretation of the situation.
More informationFlip Specification Guide. flip your expectations
Flip Specification Guide flip your expectations flip 40 Flip provides all of the features patients want most: simple operation, wireless connectivity, and a size that s stylish and discreet all with the
More informationELECTROACOUSTIC EVALUATION OF THE RESOUND MULTI MIC WITH OTOMETRICS AURICAL HIT
ELECTROACOUSTIC EVALUATION OF THE RESOUND MULTI MIC WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless technology and
More informationYou regular readers of Page Ten
PAGE 0 You regular readers of Page Ten probably recall that last month we had to interrupt Robyn Cox, PhD, and her inquiring friend in the middle of their discussion on verification procedures for the
More informationHead & Neck Institute. Hearing Aid Guide. Achieve a world of better hearing.
Head & Neck Institute Hearing Aid Guide Achieve a world of better hearing. Good hearing is part of a full and active life. Let us help you achieve a world of better hearing. The Right Choice. Cleveland
More informationDEVELOPMENT OF NAL-NL2
DEVELOPMENT OF NAL-NL2 Harvey Dillon, Gitte Keidser, Teresa Ching, Matt Flax, Scott Brewer The HEARing CRC & The National Acoustic Laboratories creating sound value value TM TM www.hearingcrc.org Prescribe
More informationMax. Powered by Era. Hearing matters
Max Powered by Era Hearing matters Meet Max Max is a new Super Power hearing instrument from Unitron. Max believes in fighting over-amplification and protecting long term hearing health. Max believes that
More informationMake the world louder!
Make the world louder! Proven ultra power amplification for extreme audibility Power and dynamic Feedback Management Sumo DM can fulfil the output needs of even your most demanding clients - even those
More informationHyperFit Desktop Fitting Guide Supplemental Instructions for the HyperSound Tinnitus Module
HyperFit Desktop Fitting Guide Supplemental Instructions for the HyperSound Tinnitus Module Table of Contents Introduction to this Supplement... 2 Hearing Care Professionals... 2 Intended Use of the HyperSound
More informationProduct Portfolio Supplement
2016 Product Portfolio Supplement I N T R O D U C I N G Muse BTE 13 The latest release in the Muse family Made for At Starkey Hearing Technologies, we ve always focused on audibility and speech understanding
More informationVerification of soft speech amplification in hearing aid fitting: A comparison of methods
Verification of soft speech amplification in hearing aid fitting: A comparison of methods Sarah E. Dawkins, B.A. AuD Research Project April 5, 2007 University of Memphis Project Advisor Robyn M. Cox, PhD.
More informationHear Better With FM. Get more from everyday situations. Life is on
Hear Better With FM Get more from everyday situations Life is on We are sensitive to the needs of everyone who depends on our knowledge, ideas and care. And by creatively challenging the limits of technology,
More information보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies
보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency
More informationThe REM Cookbook How to Correctly Perform Real Ear Measurements
www.sonici.com The REM Cookbook How to Correctly Perform Real Ear Measurements The REM Cookbook Terminology Why do audiologists perform real ear measurements? The main reason is to verify that speech sounds
More informationBinaural Processing for Understanding Speech in Background Noise
www.siemens.com/hearing Binaural Processing for Understanding Speech in Background Noise How it works, why it doesn t always work, and how we can make it better with binaural beamforming algorithms Joel
More informationEFFECT OF DIRECTIONAL STRATEGY ON AUDIBILITY OF SOUNDS IN THE ENVIRONMENT. Charlotte Jespersen, MA Brent Kirkwood, PhD Jennifer Groth, MA
EFFECT OF DIRECTIONAL STRATEGY ON AUDIBILITY OF SOUNDS IN THE ENVIRONMENT Charlotte Jespersen, MA Brent Kirkwood, PhD Jennifer Groth, MA ABSTRACT Directional microphones in hearing aids have been well-documented
More informationClinical fitting guide
Fitting Super Power is now super easy A quick reference to the Super Power features of Max: Super Power pre-sets Frequency compression 2 Power Adaptation Manager (PAM) A Sonova brand Why are we making
More informationSonic Spotlight. Binaural Coordination: Making the Connection
Binaural Coordination: Making the Connection 1 Sonic Spotlight Binaural Coordination: Making the Connection Binaural Coordination is the global term that refers to the management of wireless technology
More information