1 Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment option is that he/she will likely become more isolated, will see their quality of life continue to deteriorate, and may even experience serious depression and a variety of other problems that can be viewed as symptoms of their hearing loss. Kochkin, 2000 Rationale > All of the fitting targets and settings are based on the Hearing Levels. > Hearing levels are converted to SPL to Coupler to Hearing Aid using average correction factors. > Hearing aid gain to match target is based on average residual ear canal volume. > Only the patients know if the results meet their needs. Topics > In-Situ Audiometry Customizing the gain target > Integrated t Real-Ear measures Customizing the gain target match > Verify Comfort In-Situ functional measures In-Situ Audiometry > In Situ means that the measurement is made while the hearing aid is in the patient s ear > Goal of hearing instrument fitting is the performance of the hearing aid in the patient s ear In-Situ Audiometry Allows for calculating gain targets that represent your patient s actual fitting. Instead of using data for an average fitting. 1

2 In-Situ Audiometry > In Situ Audiometry takes into account residual ear canal volume >depth of the instrument in the ear canal seal in the ear canal >effects of venting specific receiver in that instrument Residual Ear Canal Volume When the hearing aid fits deeper in the canal the residual volume decreases As the residual volume decreases the sound pressure increases (for the same intensity of input). Effect of Insertion Depth The curves represent the mathematical variation one could expect if the depth of hearing-aid insertion were increased and decreased one and two standard deviations. re: Normal REIR db Difference SD 28 mm -1 SD 30 mm Normal 32 mm +1 SD 34 mm +2 SD 36 mm Frequency in Hz -2 SD -1 SD +1 SD +2 SD (DeJonge, 1996). In-Situ Audiometry > The signal is produced by a sine wave generator within the circuit > Thresholds are measured with only the hearing aid in the patient s ear > Targets are recalculated l using in-situ it thresholds > Use in-situ UCLs to set the maximum output levels of the hearing aid In-Situ Audiometry > In-Situ thresholds are generally better than ER-3A thresholds > Targets calculated using in-situ thresholds are more accurate since they represent the actual hearing loss Targets are based on the specific fitting not average data In-Situ Audiometry > Better starting point for Best Fit or to troubleshoot > All other process work better feedback control noise reduction compression settings 2

3 In-Situ Audiometry > Using the In Situ thresholds to calculate the target results in a better initial starting point and in the long run can save time > Fewer troubleshooting ti complaints associated with high frequencies, occlusion, or tolerance issues In-Situ Thresholds Normal vs. In-Situ Thresholds Clinical Validation > In-situ-based fitting success Subjects were sent home with Insert-earphone-based fitting in one memory and in-situ-based fitting in the other Subjects did not know which setting was in a particular memory Clinical Validation > In-situ-based fitting success They were asked to rate various situations for >Clarity >Loudness >Naturalness >Overall Sound Quality Memory preference was recorded Clinical Validation > N = 18 > Age Range: yrs Mean = 66.4 yrs > All previous users of well-fit hearing aids based on probe-microphone verification 3

4 Clinical Validation average data Clinical Validation Frequency (Hz) IN SITU ER-3A Insert Ear Phone In-Situ Audiometry vs. ER-3A db HL B Change in db Frequency (Hz) CIC N=30 ears ITC N=10 ears (Ioannou, 2000) Fitting Preference Data of ts Number Subject UCL Measures > Predicted VS Measured > Software allows the use of UCL measures at individual frequencies. > Use in-situ UCLs to set the maximum output levels of the hearing aid 0 IN SITU ER3 NO PREFERENCE Type of Audiometry 1 (Ioannou, 2000) In-Situ UCL Normal vs. In-Situ UCL 4

5 In-Situ Audiometry - Summary > Corrects the Hearing Levels based on the actual insertion depth of the hearing aid. > The corrected Hearing levels represent the actual test conditions for your patient. > The customized data are used to compute the gain requirements (targets) for each hearing aid for that particular patient. Survey > How many use Real Ear Measures with almost all fittings? > How many use it 50% of the time? > How many people never do real-ear measurement? How often do we use REM? Aud HIS Never Occasionally Less than half the time Half the time Most of the Time Almost Always Kirkwood D. Survey: Dispensers fitted more hearing aids in 2005 at higher prices. Hear J.2006;59(4):48 REM Survey Data > 2006 HR Dispenser Survey: More than half (57%) of dispensing offices possess REM equipment, but Only about one quarter (23%) use it routinely during adult hearing aid fittings. Strom KE. The HR 2006 dispenser survey. Hearing Review. 2006;13(6):34 Variables Affecting Target-Matching Accuracy > Transducer variability > Hearing Levels In-Situ measures correct this > Hearing aid size > Residual canal space variables > Vent size/slit leak Why Don t We Do Real Ear Measures? > Too expensive > Too much space > Too cumbersome > Too time consuming > Doesn t really help much > One Solution is to use real-ear measures 5

6 Too Expensive Takes too much Space $12,000 - $15,000 In a solo practice doing 5 units per week for 5 years $11 per unit Too Cumbersome Too Time Consuming > Boys Town National Research Hospital estimates 5 to 10 minutes per ear. > Then the data need to be uploaded to your hearing aid fitting system. > Estimate t about 30 minutes. If everything works smoothly maybe 20 minutes Doesn t really help much? Doesn t really help much? Conclusion: Initial fit may not tbe the best fit 9 Ears Identical Settings Variation: ~15dB Courtesy of Mike Valente, 2007 Yanz & Olson,

7 Using Real Ear Measures > Is Real Ear a good idea? > If you could obtain Real Ear Measures (REM) in the fitting software in seconds, without any additional equipment in the office, would you do it? Standard Real Ear Loudspeaker Computer and monitor Microphone housing Probe tube Ear Loop Connect to PC (Noah) via Ethernet > No Additional equipment > Uses existing clinical space > No Additional cost > Very little additional time > Benefits of REM More accurate target match More successful Best Fit More knowledgeable fine-tuning Probe tube Programming connections > Can we expect that a simple integrated system will work as well as a complicated stand-alone system? > The key measure is the Real-Ear-Coupler Difference Measuring > If the residual ear cavity is smaller than the 2cc coupler cavity, the amount of SPL in the ear cavity is: 2cc coupler REM Background > If the residual ear cavity is larger than the 2 cc coupler cavity, the amount of SPL in the ear cavity is: 2cc coupler REM re SPL Mor Less SPL 7

8 Measurement with Commercial Device Measurement with Commercial Device > Apply electrical signal to transducer > Measure SPL generated in 2 cc coupler Audioscan Verifit > Apply same electrical signal to transducer > Measure SPL generated in the ear Audioscan Verifit SPL in Ear Canal REUR SPL in Coupler 2cc and Simulated Real Ear Respo onses (db SPL) Derivation of > Apply electrical signal to transducer > Measure SPL in coupler and in ear > = Real-ear SPL - Coupler SPL Real Ear SPL Coupler SPL (db) Audioscan Verifit Uses for > Predict real-ear acoustic behavior from known coupler behavior > Pediatric application > Manufacturing software uses average s to simulate real-ear performance based on known coupler performance Average vs. Measured Fikret-Pasa, Revit,

9 Average vs. Measured Average vs. Measured Some s are smaller than average Other s are larger than average Bottom Line on Average > Average is sufficient for some ears > but not for others > You don t know if an ear is average unless you measure it > Integrated REM Sound generated through the hearing aid receiver Hearing aid microphone measures sound in the ear Measurement results integrated in Inspire OS and stored in the hearing aid No separate equipment or data entry required > Standard REM Loudspeaker generates sound sound Probe microphone measures sound in the ear Measurement results stored in database of fitting audiologist audiologist Separate equipment for measurement and data transfer > Real Ear system as part of the hearing aid Rear Ear Aided (REAR) >measurement of the hearing aid output (db SPL) while it is in the ear 2cc Coupler >measurement of the hearing aid output while it is in a 2cc coupler Real Ear to Coupler Difference () Average Measured 9

10 FFToMic BTE FFToMic ITE FFToMic ITC Two essential sets of data > Individual hearing aid coupler response Measured at the lab Stored in the Hearing Aid > Measured Measured in the ear Stored in the Hearing Aid From Real-Ear SPL to REAR > Add Microphone Location Effect (MLE) > Changes with Hearing Aid style > Quite consistent across subjects nge in db Relative Cha Microphone Location Effects Frequency in Hz FFToMic BTE FFToMic ITE FFToMic ITC Bentler and Pavlovic, 1989 Deriving REAR from Coupler Microphone Location Effects + = Relative Chang ge in db Frequency in Hz Mic Coupler Measured + + Location = Effects Individual REAR Integrated REM is used for all adjustments Integrated REM is used for all adjustments Change in Gain, Max Output, + = Compression Ratio, Compression Threshold + = Stored Coupler New Parameter Settings + = New Coupler New Coupler + = Stored New Real-Ear Aided 10

11 SPL) 2cc and Simulated Real Ear s (db Calculate Real Ear 2cc Coupler 70 db input (db) Real Ear Aided (REAR) minus 2cc Coupler =Real Ear to Coupler Difference Example Audiogram Target Matching Target Fit to REM Target: Average Ear matching ± 3 db Average Measured Fit to REM Target: non Average Ear Average Measured 11

12 Fit to REM Target: Average Ear modeling Average Measured Fit to REM Target: non Average Ear Average Measured BTE > Insert probe tube in ear canal > Start REM > Wait less than one minute > Replace RE earhook with standard earhook > is ready to be used ITE > Insert probe tube in ear canal > Start REM > Wait less than one minute > Remove probe tube from aid > is ready to be used Verify Comfort > Goal is to ensure that at a given frequency: Soft inputs are audible, yet perceived as soft Loud inputs are perceived as loud, but not uncomfortable Verify Comfort > Procedure to verify dynamic range: Patient is presented a series of soft and loud tones swept across frequencies Each side is verified individually Hearing Aid is muted during presentation >other side (if binaural fitting) is not muted Verify Comfort > Subjective measurement that verifies dynamic range Patient provides verbal response for the clinician Promotes interaction between patient and professional 12

13 Verify Comfort >Verify Comfort can be done: At the initial fit (Delivery) or first follow-up check to customize compression settings >Audibility for soft sounds >Loud sounds are loud but tolerable Troubleshooting tool used to resolve patient complaints Verify Comfort Verify Comfort > Loud is always 90 db SPL regardless of preferences > Soft is always 50 db SPL regardless of preferences > Make adjustments directly from this screen Verify Comfort > Provides an in-situ measurement to verify the patient s dynamic range. > Promotes interaction between the professional and the patient to achieve a better fit. Summary > We have learned how our hearing aids Create an individual target by allowing Hearing Levels to be measured through the hearing aid. (In-Situ Audiometry) Enable a precise target match by providing actual Real-Ear Coupler Differences. () Offer a system to verify the fitting assuring that soft sounds are clearly audible and that loud sounds are not uncomfortable (Verify Comfort) Thanks for your attention Do you have any questions? 13

Technical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S.

Technical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S. Technical Topics Where did my gain go? Thin Tube Open Fit BTE Verification Summary of presentation by C. Staples and S. Aiken History of Thin Tubes: Growth of new BTE market: First quarter of 2006 market:

More information

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04 AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by

More information

ChildFit. Widex Baby. Compass quick guide

ChildFit. Widex Baby. Compass quick guide ChildFit When working with very young children, it is not always possible to obtain all of the information necessary for an accurate hearing aid fitting. To that end, Widex has developed the ChildFit procedure

More information

AccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range

AccuQuest Spotlight: Successful Fittings with Oasis. Fitting Range AccuQuest Spotlight: Successful Fittings with Oasis When fitting a client with hearing instruments, the experience should be positive for both the client and you, the hearing care professional (HCP). The

More information

Building Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees

Building Practice Success Through Speechmap Real Ear Measurements. Chris Stokes-Rees Building Practice Success Through Speechmap Real Ear Measurements Chris Stokes-Rees Speaker Disclosure Relevant financial relationships: Audioscan (Applications Specialist) Conestoga College, Kitchener

More information

Verification of soft speech amplification in hearing aid fitting: A comparison of methods

Verification of soft speech amplification in hearing aid fitting: A comparison of methods Verification of soft speech amplification in hearing aid fitting: A comparison of methods Sarah E. Dawkins, B.A. AuD Research Project April 5, 2007 University of Memphis Project Advisor Robyn M. Cox, PhD.

More information

Mistakes on Real Ear Measures Clinicians Often Make. Michael Valente

Mistakes on Real Ear Measures Clinicians Often Make. Michael Valente Mistakes on Real Ear Measures Clinicians Often Make Michael Valente There are two national guidelines (ASHA, 1998; AAA 2006) stating the need to use REM to verify the performance of hearing aids in order

More information

Fitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound

Fitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound Fitting Decisions and their Impact on Hearing Aid User Benefit Mallory Maine, AuD Audiologist, GN ReSound Agenda Common Fitting Oversights #1 Setting the coupler type in fitting software To set or not

More information

Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit

Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit Initial-Fit Algorithm vs. Probe- Microphone Verification: Comparing Self-Perceived Benefit Harvey B. Abrams 1,2 Theresa Hnath Chisolm,2,1 Megan McManus 2 Rachel A.McArdle 1,2 1 Bay Pines VA Healthcare

More information

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015.

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. www.bestsound-technology.com DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. First fit is an essential stage of the hearing aid fitting process and is a cornerstone of the ultimate

More information

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2

The Devil is in the Fitting Details. What they share in common 8/23/2012 NAL NL2 The Devil is in the Fitting Details Why all NAL (or DSL) targets are not created equal Mona Dworsack Dodge, Au.D. Senior Audiologist Otometrics, DK August, 2012 Audiology Online 20961 What they share in

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All

More information

Verifying the Lyric Audibility Advantage

Verifying the Lyric Audibility Advantage Field Study News Spring 2014 Verifying the Lyric Audibility Advantage John Pumford, Au.D. Manager of Clinical Research Phonak Lyric Introduction Since 2008, Lyric has presented people with mild to moderate

More information

The REM Cookbook How to Correctly Perform Real Ear Measurements

The REM Cookbook How to Correctly Perform Real Ear Measurements www.sonici.com The REM Cookbook How to Correctly Perform Real Ear Measurements The REM Cookbook Terminology Why do audiologists perform real ear measurements? The main reason is to verify that speech sounds

More information

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies

보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency

More information

Manufacturers NAL-NL2 Fittings Fail Real-ear Verification

Manufacturers NAL-NL2 Fittings Fail Real-ear Verification hearingreview.com http://www.hearingreview.com/2015/02/manufacturers-nal-nl2-fittings-fail-real-ear-verification/ Manufacturers NAL-NL2 Fittings Fail Real-ear Verification Research March 2015 Hearing Review

More information

Step-by-Step RECD Guide

Step-by-Step RECD Guide Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform

More information

THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT. Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004

THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT. Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004 THE OCCLUSION EFFECT: PHYSICAL MEASUREMENT AND SELF-REPORT Karen Jarvis Advisor: Robyn Cox, Ph.D. April 2004 THE OCCLUSION EFFECT Results from trapped energy in the ear canal Unoccluded, the ear canal

More information

Clinical Tips for Matching Real-Ear Prescriptive Targets

Clinical Tips for Matching Real-Ear Prescriptive Targets AudiologyNOW! 2010 San Diego, CA Clinical Tips for Matching Real-Ear Prescriptive Targets Gabrielle Filips, Au.D., FAAA August Hernandez, Au.D., FAAA 1 Discussion topics Who s performing probe-microphone

More information

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.

More information

User groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound.

User groups Most sensorineural, conductive and mixed hearing losses in the categories of severe and profound. P R O D U C T I N F O R M A T I O N S W I F T 0 + Fitting range db HL -0 0 0 0 0 0 0 Hz 0 00 0 000 00 00 General information Swift + is a very reliable and powerful programmable BTE instrument for people

More information

Hearing Aids and Assistive Listening Devices

Hearing Aids and Assistive Listening Devices Hearing Aids and Assistive Listening Devices Audiological Rehabilitation Three Goals of Hearing Aid Fitting Increase strength of to make it audible Shape the signal to match the Protect the ear from loudness

More information

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC.

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC. T E S T I N G FM SYSTEMS with the FONIX 6500-CX Hearing Aid Analyzer (Requires software version 4.20 or above) FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209

More information

Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa

Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa J Am Acad Audiol 24:660 670 (2013) Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: 10.3766/jaaa.24.8.3 Kristi Oeding* Michael Valente* Abstract Background:

More information

Verification of In Situ Thresholds and Integrated Real-Ear Measurements DOI: /jaaa

Verification of In Situ Thresholds and Integrated Real-Ear Measurements DOI: /jaaa J Am Acad Audiol 21:663 670 (2010) Verification of In Situ Thresholds and Integrated Real-Ear Measurements DOI: 10.3766/jaaa.21.10.6 Jeffrey J. DiGiovanni* Ryan M. Pratt* Abstract Background: Accurate

More information

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method

Validation Studies. How well does this work??? Speech perception (e.g., Erber & Witt 1977) Early Development... History of the DSL Method DSL v5.: A Presentation for the Ontario Infant Hearing Program Associates The Desired Sensation Level (DSL) Method Early development.... 198 Goal: To develop a computer-assisted electroacoustic-based procedure

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX FP40 Hearing Aid Analyzer FRYE 7 Introduction The following is a quick guide for testing digital hearing aids. All digital aids can be tested, but some of the

More information

Re/Habilitation of the Hearing Impaired. Better Hearing Philippines Inc.

Re/Habilitation of the Hearing Impaired. Better Hearing Philippines Inc. Re/Habilitation of the Hearing Impaired Better Hearing Philippines Inc. Nature of Hearing Loss Decreased Audibility Decreased Dynamic Range Decreased Frequency Resolution Decreased Temporal Resolution

More information

How to use AutoFit (IMC2) How to use AutoFit (IMC2)

How to use AutoFit (IMC2) How to use AutoFit (IMC2) How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain

More information

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments

2.0. Desktop Fitting Guide getting started. Preparation of the hearing instruments Desktop Fitting Guide getting started This Guide provides you a detailed introduction to hearing instrument fitting with Phonak Target 2.0. Please also find the [News] in the Phonak Target start screen.

More information

Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer

Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 www.frye.com

More information

A comparison of manufacturer-specific prescriptive procedures for infants

A comparison of manufacturer-specific prescriptive procedures for infants A comparison of manufacturer-specific prescriptive procedures for infants By Richard Seewald, Jillian Mills, Marlene Bagatto, Susan Scollie, and Sheila Moodie Early hearing detection and communication

More information

Quick Guide Binaural REM

Quick Guide Binaural REM Quick Guide Binaural REM The purpose of this document is to provide a quick guide for the Binaural REM feature found in the REM440 Real-Ear Measurement module in the Affinity 2.0 and Callisto Suites. This

More information

DSM PRO. Software Training Manual. Copyright November 2003

DSM PRO. Software Training Manual. Copyright November 2003 DSM PRO Software Training Manual Copyright November 2003 DSM PRO Training Manual Rev 1 Effective November 2003 INTRODUCTION Congratulations! Your decision to incorporate the MedRx DSM PRO into your daily

More information

Knowledge Sharing: Pediatric Amplification Tips and Tricks. Implementing Best Practice

Knowledge Sharing: Pediatric Amplification Tips and Tricks. Implementing Best Practice Knowledge Sharing: Pediatric Amplification Tips and Tricks Implementing Best Practice KIPAT Knowledge Implementation in Pediatric Audiology Team Lisa Davidson, Ph.D., CID Research at Washington University

More information

RECD-Measurements with Connexx 6.3 and Unity 2

RECD-Measurements with Connexx 6.3 and Unity 2 RECD-Measurements with Connexx 6.3 and Unity 2 Vera Spitzlei, Corporate Audiology Siemens Audiologische Technik GmbH For internal use only / Copyright Siemens AG 2009. All rights reserved. Overview Why

More information

Industry Emphasis. Hearing Aids. Basic Design and Function. Hearing Aid Types. Hearing Aid Types. Hearing Aid Types.

Industry Emphasis. Hearing Aids. Basic Design and Function. Hearing Aid Types. Hearing Aid Types. Hearing Aid Types. Industry Emphasis Hearing Aids Since the first hearing aids were manufactured in the early 1800 s the industry has emphasized making instruments: Less cumbersome Cosmetically appealing Better? 1 2 Basic

More information

Software Training Manual

Software Training Manual Software Training Manual Copyright June 2004 Avant Training Manual Rev 1 Effective Date June 1, 2004 INTRODUCTION Congratulations! Your decision to incorporate the MedRx Avant into your daily practice

More information

Electro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook

Electro-Acoustic Stimulation (EAS) Naída CI Q90 EAS System. Naída CI Q90 EAS System Components. Naída CI Q90 Acoustic Earhook Electro-Acoustic Stimulation (EAS) EAS is a general term that refers to the simultaneous delivery of electric stimulation (cochlear implant) and acoustic amplification (hearing aid) to the same ear. Naída

More information

Insio: A new standard in custom instruments

Insio: A new standard in custom instruments Insio: A new standard in custom instruments www.siemens.com Abstract: Insio custom hearing instruments introduce novel features throughout the complete fitting process in order to obtain the best possible

More information

personalization meets innov ation

personalization meets innov ation personalization meets innov ation Three products. Three price points. Premium innovations all around. Why should a truly personalized fit be available only in a premium hearing instrument? And why is it

More information

Elements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline

Elements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline Elements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline To the hearing-impaired listener the fidelity of a hearing aid is not fidelity to the input sound but fidelity

More information

How to Use InSituGram

How to Use InSituGram How to Use InSituGram www.siemens.com Wendy Davies Abstract: Siemens InSituGram allows pure tone audiometry to be performed using the hearing instrument. The insitu audiogram will usually be very similar

More information

Hearing Solutions Catalog Winter Supplement January 2009

Hearing Solutions Catalog Winter Supplement January 2009 Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with

More information

Walkthrough

Walkthrough 0 8. Walkthrough Simulate Product. Product selection: Same look as estore. Filter Options: Technology levels listed by descriptor words. Simulate: Once product is selected, shows info and feature set Order

More information

Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management

Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management Using the FONIX 7000 to Verify Coupler and Real-Ear Performance to Adhere to the AAA Guidelines for the Audiologic Management of Adult Hearing Impairment Elizabeth Baum, B.A. Au.D. Student Program in Audiology

More information

Contents #

Contents # TRAINING MANUAL Contents Introduction... 3 AVANT REMsp Software Overview..... 3 AVANT REMsp Software Options.... 4 Preparing for Testing.. 8 Live Speech Mapping.. 11 Real Ear Measurements. 16 Additional

More information

64 AUDIOLOGY TODAY Nov/Dec 2018 Vol 30 No 6

64 AUDIOLOGY TODAY Nov/Dec 2018 Vol 30 No 6 64 AUDIOLOGY TODAY Nov/Dec 2018 Vol 30 No 6 Stop the * Title inspired by Roeser and Clark (2008) Madness Verify Hearing Aid Fittings! Hope for the Future?* BY LATA A. KRISHNAN AND JENNIFER M. SIMPSON One

More information

An assessment of the validity of a subjective hearing aid verification procedure

An assessment of the validity of a subjective hearing aid verification procedure An assessment of the validity of a subjective hearing aid verification procedure Whitney M. Vineyard, B. A. AuD Research Project April, 211 University of Memphis Project Advisor Robyn M. Cox, PhD. Purpose

More information

Audiological Bulletin no. 46

Audiological Bulletin no. 46 Audiological Bulletin no. 46 Fitting Passion 115 with Compass V4 News from Audiological Research and Communication 9 502 1119 001 10-07 2 This bulletin describes the four main steps in fitting Passion

More information

FONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech

FONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech FONIX FP35 Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Move your hearing aid fittings into the world of the touch screen! The new FP35 Touch adds a new interactive dimension to hearing

More information

Audiology Today MarApr2011

Audiology Today MarApr2011 26 Stay on Target? By Jason A. Galster The focus on gain and targets has obscured the most important part of the real-ear measurement audibility. The topic of audibility is familiar; after all, a primary

More information

Instructions for use. Genie 2 and Genie

Instructions for use. Genie 2 and Genie Instructions for use Genie and Genie Table of content Introduction Introduction... 3 Intended Use.... 4 Installation Guide... 5 Genie Overview... 7 Genie Overview...8 Warnings...8 Definition of Product

More information

The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50

The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50 The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February 2009 09:50 The HF microphone as a low-tech solution for performing musicians and "ultra-audiophiles" Of the

More information

This Page Intentionally Left Blank

This Page Intentionally Left Blank TRAINING MANUAL This Page Intentionally Left Blank 2 Contents Introduction... 4 AVANT REM Speech Software Overview. 5 AVANT REM Speech Software Options 6 Preparing for Testing.. 9 Live Speech Mapping..

More information

Testing FM Systems on the 7000 Hearing Aid Test System

Testing FM Systems on the 7000 Hearing Aid Test System Testing FM Systems on the 7000 Hearing Aid Test System Introduction Testing FM Systems on the 7000 Hearing Aid Test System This workbook describes how to test FM systems with the 7000 Hearing Aid Test

More information

Audiogram+: GN Resound proprietary fitting rule

Audiogram+: GN Resound proprietary fitting rule Audiogram+: GN Resound proprietary fitting rule Ole Dyrlund GN ReSound Audiological Research Copenhagen Loudness normalization - Principle Background for Audiogram+! Audiogram+ is a loudness normalization

More information

Testing FM Systems with FONIX FP40 Analyzers

Testing FM Systems with FONIX FP40 Analyzers Testing FM Systems with FONIX FP40 Analyzers FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 USA (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 http://www.frye.com e-mail: general@frye.com

More information

Lyric3 Programming Features

Lyric3 Programming Features Lyric3 Programming Features Introduction To help your patients experience the most optimal Lyric3 fitting, our clinical training staff has been working with providers around the country to understand their

More information

Paediatric Amplification

Paediatric Amplification Paediatric Amplification The paediatric technical advisory group (PTAG) of the NZAS recommends following the protocols found in UNHSEIP Diagnostic and Amplification Protocols (January, 2016). These guidelines

More information

Top 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound

Top 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM Top 10 ideer til en god høreapparat tilpasning Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM 314363 WRITE DOWN YOUR TOP THREE NUMBER 1 Performing appropriate hearing assessment Thorough

More information

Phonak Target 4.3. Desktop Fitting Guide. Content. March 2016

Phonak Target 4.3. Desktop Fitting Guide. Content. March 2016 Phonak Target 4.3 March 2016 Desktop Fitting Guide This guide provides you with a detailed introduction to latest hearing instrument fitting with Phonak Target. www.phonakpro.com/target_guide Please also

More information

pep Step up with pep Pep Specification Guide _Pep_Specification_Guide_UK.indd 1

pep Step up with pep Pep Specification Guide _Pep_Specification_Guide_UK.indd 1 pep Step up with pep Pep Specification Guide 124843_Pep_Specification_Guide_UK.indd 1 20.06.12 14:14 What is pep? It s that energetic feeling you get when life is great and you are ready to take on the

More information

Presenters: Sheila Moodie, University of Western Ontario; Eileen Rall, The Children's Hospital of Philadelphia; George Lindley, Oticon Pediatrics;

Presenters: Sheila Moodie, University of Western Ontario; Eileen Rall, The Children's Hospital of Philadelphia; George Lindley, Oticon Pediatrics; Presenters: Sheila Moodie, University of Western Ontario; Eileen Rall, The Children's Hospital of Philadelphia; George Lindley, Oticon Pediatrics; Contributors: Leisha Eiten, Boys Town National Research

More information

General about Calibration and Service on Audiometers and Impedance Instruments

General about Calibration and Service on Audiometers and Impedance Instruments General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev.4 2008-05-21/JF General about Calibration and Service on Audiometers

More information

2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures

2/16/2012. Fitting Current Amplification Technology on Infants and Children. Preselection Issues & Procedures Fitting Current Amplification Technology on Infants and Children Cindy Hogan, Ph.D./Doug Sladen, Ph.D. Mayo Clinic Rochester, Minnesota hogan.cynthia@mayo.edu sladen.douglas@mayo.edu AAA Pediatric Amplification

More information

Two Modified IEC Ear Simulators for Extended Dynamic Range

Two Modified IEC Ear Simulators for Extended Dynamic Range Two Modified IEC 60318-4 Ear Simulators for Extended Dynamic Range Peter Wulf-Andersen & Morten Wille The international standard IEC 60318-4 specifies an occluded ear simulator, often referred to as a

More information

Optimising Outcomes for Speech Mapping

Optimising Outcomes for Speech Mapping Optimising Outcomes for Speech Mapping Jan Pollard Chief Audiologist Sonic Craig Lett Audiologist ECS Back to Basics What is the point of all of this? Optimising the fitting for best patient outcomes and

More information

Instructions for use. Genie 2 and Genie

Instructions for use. Genie 2 and Genie Instructions for use Genie and Genie Table of Contents Introduction Introduction... Intended Use... Use Environment... Getting Ready... 4 Installation... 4 System Requirements... 4 Optional Tools... 4

More information

Introduction. What s new in Genie 2? New Fitting features. Support for new hearing aids:

Introduction. What s new in Genie 2? New Fitting features. Support for new hearing aids: Guide 207. Introduction Genie 2 and Genie have been updated with a variety of new features and enhancements. Please find a quick overview of these below: What s new in Genie 2? Support for new hearing

More information

bliss 1 Sonic Bliss Happiness sounds just like this Happiness sounds just like this. bliss Fitting Guide

bliss 1 Sonic Bliss Happiness sounds just like this Happiness sounds just like this. bliss Fitting Guide bliss Sonic Bliss Happiness sounds just like this Happiness sounds just like this. bliss Fitting Guide Sonic Bliss Happiness sounds just like this 3 Table of Contents Introduction Bliss BTE Overview 4

More information

resoundpro.com PARTENRSHIPS Aventa 3 User Guide GUIDE FOR PROFESSIONALS

resoundpro.com PARTENRSHIPS Aventa 3 User Guide GUIDE FOR PROFESSIONALS Aventa User Guide GUIDE FOR PROFESSIONALS resoundpro.com PARTENRSHIPS In this guide you will find: HOW TO START Airlink.... Other programming interfaces.... GOOD TO KNOW Start screen.... 5 Product selection....

More information

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way

binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way www.bestsound-technology.com binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way Richard Schultz-Amling Sivantos, 2015. 1 Introduction Since the release

More information

Counseling Patients on Fitting Options. Michael Valente

Counseling Patients on Fitting Options. Michael Valente Counseling Patients on Fitting Options Michael Valente Review the Audiogram Computerized Audiogram Using AudBase into EMR Using SII- Counsel on Impact of Hearing Loss on Speech Recognition www.ssi.to

More information

Sonic Spotlight. The Acoustic Flexibility of Sonic BTEs and minibtes: A Size, Sound and Set-up for Every Patient

Sonic Spotlight. The Acoustic Flexibility of Sonic BTEs and minibtes: A Size, Sound and Set-up for Every Patient 1 Sonic Spotlight Sonic Spotlight The Acoustic Flexibility of Sonic BTEs and minibtes: A Size, Sound and Set-up for Every Patient According to data from the Hearing Instruments Association (HIA), more

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Tomblin JB, Oleson JJ, Ambrose SE, Walker E, Moeller MP. The influence of hearing aids on the speech and language development of children with hearing loss. JAMA Otolaryngol

More information

Otolens Invisible In Canal Hearing Aids

Otolens Invisible In Canal Hearing Aids Otolens Invisible In Canal Hearing Aids Size matters DID YOU KNOW?... Hearing aid sales increased dramatically in the mid 1980s as a result. Deep canal hearing aid fitting Staab, WJ: The peritympanic instrument:

More information

Desktop Fitting Guide for Phonak Brio 3

Desktop Fitting Guide for Phonak Brio 3 Phonak Target 5.3.3 Desktop Fitting Guide for Phonak Brio 3 The Phonak Target fitting software is intended to be used by qualified hearing care professionals to configure, program, and fit hearing aids

More information

of the Real Ear Unaided Response

of the Real Ear Unaided Response Amplification and Aural Rehabilitation Reliability and lntersubject Variability of the Real Ear Unaided Response Michael Valente, PhD; Maureen Valente, MA; Joel Goebel, MD The Washington University School

More information

GSI TYMPSTAR PRO CLINICAL MIDDLE-EAR ANALYZER. Setting The Clinical Standard

GSI TYMPSTAR PRO CLINICAL MIDDLE-EAR ANALYZER. Setting The Clinical Standard GSI TYMPSTAR PRO CLINICAL MIDDLE-EAR ANALYZER GSI TYMPSTAR PRO CLINICAL MIDDLE-EAR ANALYZER New Standard for Clinical Impedance The GSI TympStar Pro is setting the clinical standard for performing a full

More information

Phonak Target. SoundRecover2 adult fitting guide. Content. The Connecting the hearing instruments. February 2018

Phonak Target. SoundRecover2 adult fitting guide. Content. The Connecting the hearing instruments. February 2018 Phonak Target February 2018 SoundRecover2 adult fitting guide The following fitting guide is intended for adults. For Pediatric fittings please see the separate Pediatric fitting guide. SoundRecover2 is

More information

Hearing Loss and Hearing Aids for Adults

Hearing Loss and Hearing Aids for Adults Consumer FaCts Hearing Loss and Hearing Aids for Adults 1J SPEECH-LANGUAGE PATHOLOGY AND AUDIOLOGY AND HEARING AID DISPENSERS BOARD Speech-Language Pathology and Audiology and Hearing Aid Dispensers Board

More information

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs.

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs. INTRODUCTION TO PURE TONE AUDIOMETERY (AUDIOMETER & TESTING ENVIRONMENT) By Mrs. Wedad Alhudaib with many thanks to Mrs. Tahani Alothman Topics : This lecture will incorporate both theoretical information

More information

DEVELOPMENT OF NAL-NL2

DEVELOPMENT OF NAL-NL2 DEVELOPMENT OF NAL-NL2 Harvey Dillon, Gitte Keidser, Teresa Ching, Matt Flax, Scott Brewer The HEARing CRC & The National Acoustic Laboratories creating sound value value TM TM www.hearingcrc.org Prescribe

More information

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA 2017-10 Presenter Peter Wulf-Andersen Engineering degree in Acoustics Co-founder

More information

Evidence base for hearing aid features:

Evidence base for hearing aid features: Evidence base for hearing aid features: { the ʹwhat, how and whyʹ of technology selection, fitting and assessment. Drew Dundas, PhD Director of Audiology, Clinical Assistant Professor of Otolaryngology

More information

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods C HAPTER FOUR Audiometric Configurations in Children Andrea L. Pittman Introduction Recent studies suggest that the amplification needs of children and adults differ due to differences in perceptual ability.

More information

AND9020/D. Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION

AND9020/D. Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION This information note describes the feedback cancellation feature provided in ON Semiconductor s latest digital hearing

More information

Comparing Speech Perception Abilities of Children with Cochlear Implants and Digital Hearing Aids

Comparing Speech Perception Abilities of Children with Cochlear Implants and Digital Hearing Aids Comparing Speech Perception Abilities of Children with Cochlear Implants and Digital Hearing Aids Lisa S. Davidson, PhD CID at Washington University St.Louis, Missouri Acknowledgements Support for this

More information

WIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38

WIDEXPRESS THE WIDEX FITTING RATIONALE FOR EVOKE MARCH 2018 ISSUE NO. 38 THE WIDEX FITTING RATIONALE FOR EVOKE ERIK SCHMIDT, TECHNICAL AUDIOLOGY SPECIALIST WIDEX GLOBAL RESEARCH & DEVELOPMENT Introduction As part of the development of the EVOKE hearing aid, the Widex Fitting

More information

Why? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction

Why? Speech in Noise + Hearing Aids = Problems in Noise. Recall: Two things we must do for hearing loss: Directional Mics & Digital Noise Reduction Directional Mics & Digital Noise Reduction Speech in Noise + Hearing Aids = Problems in Noise Why? Ted Venema PhD Recall: Two things we must do for hearing loss: 1. Improve audibility a gain issue achieved

More information

Feelin The Squeeze? 10/17/2016. Any of these been on your mind lately? Key Part of NAS 12 Recommendations

Feelin The Squeeze? 10/17/2016. Any of these been on your mind lately? Key Part of NAS 12 Recommendations Any of these been on your mind lately? The Art of Independent Practice In Today s Td Corporate/Commodity Culture Presenter David J. Smriga, M.A. Senior Audiology Consultant Audioscan Education MAC Conference

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3

More information

Phonak Target 5.4. Phonak Target Fitting Guide. Content. August 2018

Phonak Target 5.4. Phonak Target Fitting Guide. Content. August 2018 Phonak Target 5.4 August 2018 Phonak Target Fitting Guide The Phonak Target fitting software is intended to be used by qualified hearing care professionals to configure, program, and fit hearing aids to

More information

Phonak Target 6.0. Phonak Target Fitting Guide. Content. November 2018

Phonak Target 6.0. Phonak Target Fitting Guide. Content. November 2018 Phonak Target 6.0 November 2018 Phonak Target Fitting Guide The Phonak Target fitting software is intended to be used by qualified hearing care professionals to configure, program, and fit hearing aids

More information

LEARNING TO USE YOUR FIRST HEARING AID

LEARNING TO USE YOUR FIRST HEARING AID Hearing Aid User Manual PHLHA46 LEARNING TO USE YOUR FIRST HEARING AID Your hearing aid cannot return your level of hearing to normal or halt further hearing deterioration, but proper use of your hearing

More information

Using digital hearing aids to visualize real-life effects of signal processing

Using digital hearing aids to visualize real-life effects of signal processing Using digital hearing aids to visualize real-life effects of signal processing By Francis Kuk, Anne Damsgaard, Maja Bulow, and Carl Ludvigsen Digital signal processing (DSP) hearing aids have made many

More information

Digital. hearing instruments have burst on the

Digital. hearing instruments have burst on the Testing Digital and Analog Hearing Instruments: Processing Time Delays and Phase Measurements A look at potential side effects and ways of measuring them by George J. Frye Digital. hearing instruments

More information

Oscilla - excellence since 1960

Oscilla - excellence since 1960 Oscilla - excellence since 1960 Oscilla USB310 / 330 PC-based Screening Audiometer 11 frequencies: 125-8000 Hz (selectable in USB330) Left, right and binaural Pulse tone Warble tone (USB330) 4 automatic

More information

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds Signals, systems, acoustics and the ear Week 1 Laboratory session: Measuring thresholds What s the most commonly used piece of electronic equipment in the audiological clinic? The Audiometer And what is

More information