General about Calibration and Service on Audiometers and Impedance Instruments

Size: px
Start display at page:

Download "General about Calibration and Service on Audiometers and Impedance Instruments"

Transcription

1 General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev /JF

2 General about Calibration and Service on Audiometers Purpose General about service General description of an audiometer The intention with this paper is to give new technicians or others who are interested, a brief explanation about the methods and necessary equipment used by the calibration and service of audiometers. An audiometer, together with accessories, is a delicate instrument, and should be treated with great care. Otherwise you cannot ensure that the measured results are correct and comparable to earlier tests and investigations made with other instruments. Anyhow, the normal use will often cause a slight change of the mechanical and electrical properties, mostly caused by normal wear of the instruments. It is therefore necessary to make periodic service on the audiometers, where the electrical and mechanical properties are checked by a skilled technician, who also makes judgements concerning eventual exchange of vital parts as cords and headsets in order to avoid defects in the future. The recommended interval between periodic service will normally be one year or less, for instruments in daily use. The basic audiometer produces pure sine tones of fixed frequencies, which are presented to the patient through a set of specially approved transducers. The output of the transducers can be controlled by a calibrated attenuator in the audiometer and the tone can be interrupted or presented without introducing any switch noise. The international standards IEC 645-1, IEC 645-2, IEC and IEC specifies the minimum demands for different kinds of audiometers. Air Conduction Earphones Measuring equipment for calibration - general Artificial ear For audiometry purposes, only a few types of earphones are approved. This is due to very strict demands to the mechanical and electrical properties. It is very important that the cushion has the correct shape and that the fitting to the earphone is good. Bad fitting of the cushion to the earphone can cause errors in the measurements. To be able to make a comparable and repeatable calibration of the output sound level of the earphones, it will be necessary for the technician to have a standardised set of reliable measuring equipment. The most important part of equipment is a standardised coupler, which is a sort of artificial ear, where the earphone can be placed in order to measure the acoustical output. In this way it will be possible to use the internationally standardised calibration values and to compare the measured results made by different laboraties. Calibration and Service of Audiometers - page 2 of 14 Rev /JF

3 Microphone Measuring amplifier Acoustic calibrator Calibration of pure tone air conduction The earphones for audiometers used by Interacoustics, are Telephonics type TDH39, and they are checked and calibrated by means of a coupler made in accordance to IEC 303. This coupler has a volume of 6cc but can be converted to a 2cc coupler using a supplied adapter. During the measurements the earphone is placed on top of the coupler, carefully centred over the opening, with the recommended total weight of 4,5N (459g). This static force can be supplied by a special spring arrangement or a fixed weight. It is important that the earphone is applied to the coupler without acoustical leakage. The coupler is supplied with a 1 inch condenser microphone of the pressure type. This is a very delicate device, and should be treated with great care by disassembling and assembling. The microphone has a very high output impedance and requires a polarisation voltage of typically 200V in order to function. To avoid problems with noise, a preamplifier is normally integrated on the coupler and this is again connected to a special input on the connected sound level meter, having the facilities of supplying the appropriate supply voltages. The coupler is connected to a sound level meter. This is a measuring amplifier with selectable amplification levels and an indicating display to show the measured sound pressure level. Sound pressure level (SPL), is expressed in db above a reference sound level of 20 µpa, which is defined as the minimum audible sound level. It is an advantage to have a measuring amplifier with an output for monitor, where an oscilloscope and frequency counter can be connected. During all calibration, it is important that the signals are undisturbed, and this can be monitored by using the oscilloscope, while the frequency counter can be used for the check of the exact frequency of the tones from the audiometer. In order to ensure that the indicator of the measuring amplifier is showing the correct values, it is advisable to have an acoustical calibrator. This device produces a calibrated sound pressure level and fits directly over the 1 inch microphone. The indicator on the measuring amplifier should, with the calibrator connected, show the level printed on the label of calibrator. If this is not the case, the sensitivity of the measuring amplifier should be adjusted, or the deviation should be taken in account by the measurements. The output level of an audiometer is normally expressed in Hearing Level (db HL), and can at least be changed in steps of 5 db, typically ranging from -10dB to 100 db. The HL indicates, that the instrument is calibrated to have output levels other than SPL (Sound Pressure Level). The international standard called ISO is providing the lowest sound pressure levels in db re. 20 µpa (SPL) for each audiometer Calibration and Service of Audiometers - page 3 of 14 Rev /JF

4 Microphone corrections frequency, which just can be heard by a normal hearing person. These values corresponds to a HL-setting of 0 db on the audiometer. Unfortunately this level is so low, that normal measuring equipment will be unable to give any readable results due to noise. During the check or calibration of the audiometer, the attenuator therefore has to be increased in order to make the earphone produce a usable level (normally 70 or 80 db HL). This means that the reading on the indicator of the measuring amplifier will be the sum of ISO levels and the selected HL-value. Before the calibration can be carried out, it is necessary to add some corrections, due to the fact that the 1 inch microphone is not linear through the whole frequency range. When you buy a microphone, you also get a chart showing its frequency characteristic. This is used to find, for each audiometer frequency, the deviation from its 0 db-level. A positive deviation indicates that the output of the microphone is higher than expected with a constant reference level, while a negative shows that the output is lower. Typically, the output will be flat at lower frequencies and fall off at higher frequencies. Positive deviations are added, while negative are subtracted in order to obtain the correct indication on the measuring amplifier. The correct indication for each frequency will then be: db HL+ ISO Δmic. Calibration of narrow-band masking noise Masking noise is calibrated in the same manner as pure tone. Only the reference values are different. Narrow band masking is calibrated using the values from ISO plus the values from ISO taking the figures valid for one third octave bandwidth. Also here the correction values for the microphone should be taken in account. Bone conduction Bone conductors Artificial mastoid The bone conductor is intended to produce vibration instead of sound. ISO gives standard reference threshold vibration values for bone conductors with a specified circular tip of an area of 175mm² and a static pressure of 5,4N (551g) on the human mastoid bone. The Threshold values in ISO are expressed in db above 1 µn and as for the pure tone, they are valid for an attenuator-setting of 0 db HL on the audiometer. A bone conductor is a small and compact device and is very sensitive to shock, which can have bad influence on its properties. It is therefore very important that it is not dropped on a hard surface. For the calibration purpose, there must be used another type of coupler than for pure tone. This device is called an artificial mastoid and it Calibration and Service of Audiometers - page 4 of 14 Rev /JF

5 Artificial mastoid, correction values Measuring amplifier Bone conductor calibration has a surface with mechanical properties, comparable to the human mastoid bone. IEC 373 describes such a mechanical coupler and it has a built-in accelerometer which delivers a calibrated signal to be used for the calibration of the bone conductors. As for the microphone, the artificial mastoid is supplied with a frequency chart which should be used for the determination of the correction factors. 3 different charts will normally be delivered, but use the chart for force sensitivity for this purpose. The chart also has a field with information about the force sensitivity at 1000Hz. This value is expressed in mv/n, but has to be converted to db re. V/N before further calculations are made. Having done that, it is possible to calculate the output of the artificial mastoid at the desired frequencies, simply by adding the ISO threshold values, the 1000Hz force level, the correction factors and the HL setting for each frequency - all expressed in db. If desired, the calculated db - values can easily be converted to µv, as the values are expressed in db re. 1µV. The correct reading on the measuring amplifier for each frequency will then be: db HL + ISO Force Sensitivity + Δmast. The artificial mastoid uses no preamplifier and can therefore be connected directly to the voltage input of the measuring amplifier. Due to the high impedance of the artificial mastoid, it is important that it is connected to a measuring amplifier with a high impedance input, typical above 1 MΩ, otherwise there will be errors in the measured values. During the measurements, it is important that the bone conductor under test is correctly placed on the middle of the surface of the artificial mastoid, and that the static pressure is correctly adjusted. Due to temperature sensitivity of the artificial mastoid, it is important that the ambient temperature is remained constant. The most exact calibration values are obtained at an ambient temperature of 23 C. Service on Audiometers and impedance instruments General Although the instruments from Interacoustics are made to be very reliable devices, it will be necessary to carry out periodic controls by a specialist in order to ensure that the performances are kept. Calibration and Service of Audiometers - page 5 of 14 Rev /JF

6 Checkpoints Cleaning Periodic replacements This should be done at least once a year by a skilled person with the necessary equipment. The equipment needed is shown in the enclosed Set-up for Calibration of Audiometers and Impedance Instruments. By the control it is checked that the transducers are delivering the correct signals, levels and frequencies, as well as the different function modes are tested. Special attention should be made to proper functioning of parts which is exposed to wear due to normal handling. This concerns the following: control devices, e.g. knobs and push buttons, indicating devices, e.g. lamps and displays, cords, plugs, headbands, earphone cushions, hoses and tubes for impedance probes. The rubber of the earphone cushion is often very contaminated and should therefore be cleaned. If it is inflexible or damaged, it will be necessary to replace it. Probes for impedance instruments will very often be contaminated with earwax, both in the tip of the probe and in the tiny tubes. These has to be cleaned in water and by means of a cleaning wire. Note that only the detachable mechanical parts of the probe are allowed to be exposed to water. On impedance instruments, it is recommended, by each periodic control, to replace the short hoses, connecting the tubes to the transducers in the probe, as they very easily can be damaged and cause leaks. Use the same type of hose, cut in the same length as the original ones. Additional Information Calibration chart Related standards IEC IEC It is advisable that the technician makes a chart showing all audiometer frequencies together with the calculated calibration values which corresponds to the set of equipment he uses. Enclosed you will find 2 examples of records, showing how calibration charts can be designed. One is intended for the acoustical coupler, while the other is for the artificial mastoid. The records are included in the Interacoustics Quality Manual: Test Procedures. Audiometers part 1: Pure tone audiometers Audiometers part 2: Equipment for speech audiometry Calibration and Service of Audiometers - page 6 of 14 Rev /JF

7 IEC IEC IEC ISO ISO ISO ISO ISO ISO ISO ISO 6189 ISO ISO ISO Audiometers part 3: Auditory test signals of short duration for audiometric and neuro-otological purposes Audiometers part 4: Equipment for extended high-frequency audiometry Instruments for the measurement of aural acoustic impedance/admittance. (Former ISO 389) Acoustics - Reference zero for the calibration of audiometric equipment Part 1: Reference equivalent threshold sound pressure levels for pure tones and supra-aural earphones. Acoustics - Reference zero for the calibration of audiometric equipment Part 2: Reference equivalent threshold sound pressure levels for pure tones and insert earphones. (Former ISO 7566) Acoustics - Reference zero for the calibration of audiometric equipment Part 3: Reference equivalent threshold force levels for pure tones and bone vibrators. (Former ISO 8798) Acoustics - Reference zero for the calibration of audiometric equipment Part 4: Reference levels for narrow - band masking noise. Acoustics - Reference zero for the calibration of audiometric equipment Part 5: Reference equivalent threshold sound pressure levels for pure tones in the frequency range 8 khz to 16 khz. (Not yet issued) Acoustics - Reference zero for the calibration of audiometric equipment Part 6: Reference equivalent threshold sound pressure levels for acoustic test signals of short duration. Acoustics - Reference zero for the calibration of audiometric equipment Part 7: Reference threshold of hearing under free-field and diffuse field listening conditions. Acoustics Pure tone air conduction threshold audiometry for hearing conservation purposes. Acoustics Audiometric test methods Part 1: Basic pure tone air conduction threshold audiometry. Acoustics Audiometric test methods Part 2: Sound field audiometry with pure tone and narrow-band test signals. Acoustics Audiometric test methods Part 3: Speech audiometry. Enclosed Pure Tone Air conduction Calibration Instruction chart. Bone Conductor Calibration Instruction chart Set-up for Calibration of Audiometers and Impedance instruments. Survey of reference values for audiometer calibration at standard frequencies. Survey of speech reference values. Calibration and Service of Audiometers - page 7 of 14 Rev /JF

8 Calibration and Service of Audiometers - page 8 of 14 Rev /JF

9 Pure Tone Air conduction Calibration Instruction Calibration Stand no.: Sign.: Measuring Amplifier type: sn.: Artificial Ear type: sn.: Microphone type: sn.: Cathode Follower type: sn.: Audiometer settings ISO for TDH 39 Mic. correction values Calibration values Max. allowed deviation [Hz] [db HTL] [db SPL] [db] [db SPL] [± db] The microphone correction values are found by reading the B&K calibration chart for the respective microphone. Read the output level of the microphone on the Y-axis of the calibration chart, at the frequency of the calibrator.(1000 Hz). Read the deviation from this level at each audiometer frequency and write it in the column for "mic. correction values". When the calibration curve is above this level, the correction value is positive, while negative when below. The values in the column for "calibration value" are found as an addition of attenuator, threshold and microphone correction levels. Calibration and Service of Audiometers - page 9 of 14 Rev /JF

10 Bone Conductor Calibration Instruction The thresholds for bone conductor are given in force values. These values has to be converted into voltage output for a given artificial mastoid. Output of artificial mastoid = T F + S FDV + C F + Att. setting T F S FDV C F : Threshold force (BC standard) [db re. 1µN] : Force sensitivity of the artificial mastoid at 1000 Hz. This value is given in mv/n on the calibration chart and has to be converted to db re. 1V/N. (= 1µV/µN) = 20 log (Force Sensitivity [V]) ex.: 20 log 0,16 V = -15,9 db : Correction factor for each frequency, read on the "constant dynamic force" curve of the artificial mastoid, giving the deviations from the 0 db line. Artificial mastoid no. Force sensitivity : mv/n S FDV = 20 log V = db Audiometer settings Art. mast. values ISO Freq. Att. S FDV C F T F Req. reading [Hz] [db] [db] [db] [db] [dbv] [mv] ,0 58,0 48, ,5 36,5 31,0 30,0 35,5 40,0 40,0 Calibration and Service of Audiometers - page 10 of 14 Rev /JF

11 Measuring Equipment Recommended set-up for Calibration of Audiometers and Impedance Instruments Oscilloscope Oscillator Multimeter Measuring amplifier Frequency counter Artificial mastoid 1 inch mic. with pre-amp Art. Ear 4152 as IEC 303 coupler (6cc) Art. Ear 4152 as IEC 126 coupler (2cc) Manometer MHC1 Bone conductor Loudspeaker Air conduction headset Insert Earphone Contra Earphone Ipsi Pressure Audiometer Impedance Audiometer The drawing shows the different types of equipment needed to make a complete calibration of Clinical Audiometers and Impedance Instruments. The measuring amplifier and Microphone, Artificial Ear and Artificial mastoid, are normally supplied from Brüel & Kjär although there will be other trade marks with similar equipment. The Brüel & Kjär equipment can be purchased through Interacoustics. A portable sound level meter can be used instead of the stationary measuring amplifier, but will have some disadvantages as it normally lacks the monitor output for oscilloscope and frequency counter as well as direct signal voltage measurement facilities. Calibration and Service of Audiometers - page 11 of 14 Rev /JF

12 The oscilloscope, frequency counter, tone generator and multimeter are common equipment which can be bought in a wide range of quality and prices. Calibration and Service of Audiometers - page 12 of 14 Rev /JF

13 Survey of reference values for audiometer calibration at standard frequencies Freq. ISO ANSI S3.6 (TDH39) ISO 389-4/ ANSI S3.6 (NB) ISO ANSI S3.6 (BC) ISO 389-2/ ANSI S3.6 (Insert) ISO Tone (FF) ISO Diffuse field (FF) (TDH39) (BC) [Hz] [db re. 20 µpa] [db re. 20 µpa] [db re. 20 µpa] [db re. 1µN] [db re. 1µN] [db re. 20 µpa] [db re. 20 µpa] [db re. 20 µpa] , ,0 22, , ,0 18, , ,5 14, ,5 25, ,0 11, ,5 8, ,0 5, ,5 11, ,5 4,0 3, ,5 5 52,5 52,5 4 2,5 1, , ,5 48,5 2 2,0 1, ,5 2,0 1, ,5 42,5 0 2,0 0, , ,5 0, ,5 6,5 6 36,5 36,5 2 0,5-1, ,5 35,5 2 0,0-1, ,5-1, ,5 6 29,5 29,5 5-4,0-3, ,5-6,0-4, ,5-4, ,5 9,5 5 35,5 35,5 5,5-6,5-5, ,0-3, ,5 15, ,5-0, ,0 0, ,5 5,5 TDH39 is calibrated using a 6cc acoustic coupler made in accordance to IEC 303 BC is calibrated using a mechanical coupler made in accordance to IEC 373 (1990-1) Insert phones are calibrated using a 2cc acoustic coupler made in accordance to IEC 126 Reference values for calibration of high frequency Freq. Interacoustics Standard (Koss/pro) ISO/TR HDA200 [khz] [db re. 20 µpa] [db re. 20 µpa] , , Koss/pro is calibrated using an acoustic coupler made in accordance to IEC 318 Calibration and Service of Audiometers - page 13 of 14 Rev /JF

14 Survey of speech reference values IEC 645-2, type A or B IEC 645-2, type A-E or B-E ANSI S3.6 type A, B or C ANSI S3.6 type A-E, B-E or C-E [db] [db] [db] [db] Earphones Calibration signal 20 0 (3,8)* 19,5 19,5 Speech weighted masking 20 0 (3,8)* 19,5 19,5 Bone conductor Calibration signal 55 NS(55) 55 NS(55) Speech weighted masking 55 NS(55) 55 NS(55) Loudspeaker Calibration signal ,5 14,5 Speech weighted masking ,5 14,5 Insert phones Calibration signal 20 NS(20) 12,5 NS(12,5) Speech weighted masking 20 NS(20) 12,5 NS(12,5) Insert masking phones Speech weighted masking 20 NS(20) 12,5 NS(12,5) NS = Not Standardised yet. Values in brackets are used by factory calibration. * Values in brackets are valid for Germany Calibration and Service of Audiometers - page 14 of 14 Rev /JF

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs.

INTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs. INTRODUCTION TO PURE TONE AUDIOMETERY (AUDIOMETER & TESTING ENVIRONMENT) By Mrs. Wedad Alhudaib with many thanks to Mrs. Tahani Alothman Topics : This lecture will incorporate both theoretical information

More information

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R

More information

Organisation Internationale de Métrologie Légale

Organisation Internationale de Métrologie Légale Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3

More information

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds

Signals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds Signals, systems, acoustics and the ear Week 1 Laboratory session: Measuring thresholds What s the most commonly used piece of electronic equipment in the audiological clinic? The Audiometer And what is

More information

Diagnostic Audiometer

Diagnostic Audiometer AD629 Technical Specificatons Page 0 Technical Specifications AD629 Diagnostic Audiometer Item No.8107429 3/2015 AD629 Technical Specificatons Page 1 License Overview International configuration AD629

More information

COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO :1998

COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO :1998 INTERNATIONAL STANDARD IS0 8253-l First edition 1989-11-15 COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO 8253-1:1998 Acoustics - Audiometric test methods - Part I: Basic

More information

AC40 Technical Specificatons Page 1. Technical Specifications AC40. Clinical Audiometer

AC40 Technical Specificatons Page 1. Technical Specifications AC40. Clinical Audiometer AC40 Technical Specificatons Page 1 Technical Specifications AC40 Clinical Audiometer D-0106990-B 05/2015 AC40 Technical Specificatons Page 2 License Overview International configuration AC40 Basic license

More information

AC40 Technical Specificatons. Technical Specifications AC40. Clinical Audiometer D F 2017/11

AC40 Technical Specificatons. Technical Specifications AC40. Clinical Audiometer D F 2017/11 AC40 Technical Specificatons Page 1 Technical Specifications AC40 Clinical Audiometer D-0106990-F 2017/11 AC40 Technical Specificatons Page 2 License Overview International configuration AC40 Basic license

More information

AA-M1C1. Audiometer. Space saving is achieved with a width of about 350 mm. Audiogram can be printed by built-in printer

AA-M1C1. Audiometer. Space saving is achieved with a width of about 350 mm. Audiogram can be printed by built-in printer Audiometer AA-MC Ideal for clinical practice in otolaryngology clinic Space saving is achieved with a width of about 350 mm Easy operation with touch panel Audiogram can be printed by built-in printer

More information

AD226 Diagnostic Audiometer

AD226 Diagnostic Audiometer AD226 Technical Specificatons Page 0 Technical Specifications AD226 Diagnostic Audiometer D-0106126-C 2017/11 AD226 Technical Specificatons Page 1 License Overview International configuration AD226 Basic

More information

MA 40 Operating Instructions

MA 40 Operating Instructions MA 40 Operating Instructions MAICO Diagnostics 10393 West 70 th Street Eden Prairie, MN 55344, USA Toll Free 888.941.4201 TABLE OF CONTENTS Warranty... 1 1.0 SPECIFICATIONS... 2 2.0 INTRODUCTION... 4 2.1

More information

HEARING CONSERVATION PROGRAM

HEARING CONSERVATION PROGRAM CALIFORNIA STATE UNIVERSITY, CHICO HEARING CONSERVATION PROGRAM PREPARED BY THE OFFICE OF ENVIRONMENTAL HEALTH AND SAFETY REVISED June 2008 TABLE OF CONTENTS Section Page 1.0 Introduction... 1-1 2.0 Exposure

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 389-7 Second edition 2005-11-01 Acoustics Reference zero for the calibration of audiometric equipment Part 7: Reference threshold of hearing under free-field and diffuse-field

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 389-8 First edition 2004-05-15 Acoustics Reference zero for the calibration of audiometric equipment Part 8: Reference equivalent threshold sound pressure levels for pure tones

More information

Two Modified IEC Ear Simulators for Extended Dynamic Range

Two Modified IEC Ear Simulators for Extended Dynamic Range Two Modified IEC 60318-4 Ear Simulators for Extended Dynamic Range Peter Wulf-Andersen & Morten Wille The international standard IEC 60318-4 specifies an occluded ear simulator, often referred to as a

More information

Recommended Procedure. Industrial Audiometry

Recommended Procedure. Industrial Audiometry Recommended Procedure Date of version: September Date for review: September 2018 Recommended Procedure General foreword This document presents a Recommended Procedure by the British Society of Audiology

More information

Recommended Procedure. Industrial Audiometry

Recommended Procedure. Industrial Audiometry Recommended Procedure Industrial Audiometry Date: November 2012 Date for review: November 2018 General foreword This document presents a Recommended Procedure by the British Society of Audiology (BSA).

More information

CITY OF FORT BRAGG HEARING CONSERVATION PROGRAM

CITY OF FORT BRAGG HEARING CONSERVATION PROGRAM CITY OF FORT BRAGG HEARING CONSERVATION PROGRAM A. PURPOSE It is the City s policy to protect the health and safety of employees through the establishment and enforcement of this Hearing Conservation Program

More information

INSERT EARPHONES. User Guide

INSERT EARPHONES. User Guide TM INSERT EARPHONES User Guide INTRODUCTION AND INTENDED USE ER-3C earphones are the next generation of insert earphones by Etymotic Research, designed to replace Etymotic s ER-3A Tubephone earphones,

More information

Operating Instructions MA 25

Operating Instructions MA 25 Operating Instructions MA 25 Table of Contents Page 1 Introduction... 2 2. Description... 3 2.1 Important safety note... 4 2.2 Unpacking and checking the MA 25... 5 2.3 Accessories... 5 2.5 Connecting

More information

The registration testing for audiometric device in CFDA

The registration testing for audiometric device in CFDA The registration testing for audiometric device in CFDA Beijing Center for Medical Devicee Quality Supervision and Testing of China Food and Drug Administration Beijing Institute of Medical Device Testing

More information

Supplementary Online Content

Supplementary Online Content Supplementary Online Content Lin FR, Ferrucci L. Hearing loss and falls among older adults in the United States. Arch Intern Med. 2012;172(4):369-371. eappendix. Audiometric Assessment and Other Study

More information

TWO CHANNEL CLINICAL AUDIOMETER AUDIOSTAR PRO

TWO CHANNEL CLINICAL AUDIOMETER AUDIOSTAR PRO TWO CHANNEL CLINICAL AUDIOMETER AUDIOSTAR PRO THE AUDIOMETER PERFECT FOR EVERY PATIENT POPULATION GSI AUDIOSTAR PRO FAST AND EFFICIENT The GSI AudioStar Pro continues the tradition of excellence in clinical

More information

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system

UNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.

More information

Equinox 2.0. PC-based audiometer. Time saving solutions. Interacoustics. leading diagnostic solutions

Equinox 2.0. PC-based audiometer. Time saving solutions. Interacoustics. leading diagnostic solutions Equinox 2.0 PC-based audiometer Time saving solutions Equinox 2.0 PC-based audiometer The best of two worlds Equinox 2.0 is a state of the art two channel clinical audiometer controlled from an external

More information

THESIS PRESENTATION. The Reference Equivalent Threshold Sound Pressure Level (RETSPL) values of the Creare headphones

THESIS PRESENTATION. The Reference Equivalent Threshold Sound Pressure Level (RETSPL) values of the Creare headphones THESIS PRESENTATION The Reference Equivalent Threshold Sound Pressure Level (RETSPL) values of the Creare headphones by Lee Shi Yuan (A0100011U / E0012334) Purpose Determine the RETSPL values of the Creare

More information

VALIDITY OF DIAGNOSTIC COMPUTER-BASED AIR AND FOREHEAD BONE CONDUCTION AUDIOMETRY

VALIDITY OF DIAGNOSTIC COMPUTER-BASED AIR AND FOREHEAD BONE CONDUCTION AUDIOMETRY VALIDITY OF DIAGNOSTIC COMPUTER-BASED AIR AND FOREHEAD BONE CONDUCTION AUDIOMETRY De Wet Swanepoel 1, 2 Leigh Biagio 1 1. Department of Communication Pathology, University of Pretoria, South Africa 2.

More information

AA222 / AA220. Audio Traveller. Save time on routine. Focus on results. Interacoustics. leading diagnostic solutions

AA222 / AA220. Audio Traveller. Save time on routine. Focus on results. Interacoustics. leading diagnostic solutions AA222 / AA220 Audio Traveller Save time on routine. Focus on results. AA222 / AA220 Audio Traveller Middle ear analyzer The AA222/AA220 is the most comprehensive combination of middle ear analyzer and

More information

TOWN OF FAIRFIELD PUBLIC HEALTH NURSING. MANUAL: School Health APPROVED BY: Board of Health School Medical Advisor

TOWN OF FAIRFIELD PUBLIC HEALTH NURSING. MANUAL: School Health APPROVED BY: Board of Health School Medical Advisor TOWN OF FAIRFIELD PUBLIC HEALTH NURSING MANUAL: School Health APPROVED BY: Board of Health School Medical Advisor POLICY: Audiometric Screening DATE REVISED AND APPROVED: 8-14-95, 3-13-06, 7-2015 I. Purpose

More information

Oscilla - excellence since 1960

Oscilla - excellence since 1960 Oscilla - excellence since 1960 Oscilla USB310 / 330 PC-based Screening Audiometer 11 frequencies: 125-8000 Hz (selectable in USB330) Left, right and binaural Pulse tone Warble tone (USB330) 4 automatic

More information

MA 39 Operating Instructions

MA 39 Operating Instructions MA 39 Operating Instructions MAICO Diagnostics 10393 West 70 th Street Eden Prairie, MN 55344, USA Toll Free 888.941.4201 MA 39 Operating Instructions TABLE OF CONTENTS Warranty... 1 1.0 SPECIFICATIONS...

More information

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA

Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA 2017-10 Presenter Peter Wulf-Andersen Engineering degree in Acoustics Co-founder

More information

Week 2 Systems (& a bit more about db)

Week 2 Systems (& a bit more about db) AUDL Signals & Systems for Speech & Hearing Reminder: signals as waveforms A graph of the instantaneousvalue of amplitude over time x-axis is always time (s, ms, µs) y-axis always a linear instantaneousamplitude

More information

MA 25 Operating Manual

MA 25 Operating Manual Table of Contents Page 1. Introduction... 1 2. Description... 2 2.1 Important safety note... 3 2.2 Unpacking and checking the MA 25... 5 2.3 Standard accessories... 5 2.4 Optional accessories... 5 2.5

More information

Diagnostic Impedance & Audiometry in one solution

Diagnostic Impedance & Audiometry in one solution AA222/AA222h Audio Traveller Diagnostic Impedance & Audiometry in one solution Interacoustics l e a d i n g d i a g n o s t i c s o l u t i o n s AA222/AA222h Audio Traveller Middle ear analyzer The AA222

More information

IEC/TC 29 Electroacoustics. Status: September 2010

IEC/TC 29 Electroacoustics. Status: September 2010 IEC/TC 29 Electroacoustics Status: September 2010 Secretary: Leif Nielsen Danish Standards Association Kollegievej 6 DK-2920 Charlottenlund Denmark Email address: len@ds.dk Chairman: Susan Dowson National

More information

Air conduction hearing thresholds of young and older Japanese adults for pure tones from 125 Hz to 16 khz

Air conduction hearing thresholds of young and older Japanese adults for pure tones from 125 Hz to 16 khz TECHNICAL REPORT #2011 The Acoustical Society of Japan Air conduction hearing thresholds of young and older Japanese adults for pure tones from 125 Hz to 16 khz Kenji Kurakata 1;, Tazu Mizunami 1, Kazuma

More information

EPIC-PLUS System for auditory evoked potentials and neuro-otology. PEA - VEMPs ASSR - E-ABR FACIAL NERVE

EPIC-PLUS System for auditory evoked potentials and neuro-otology. PEA - VEMPs ASSR - E-ABR FACIAL NERVE EPIC-PLUS System for auditory evoked potentials and neuro-otology PEA - VEMPs ASSR - E-ABR FACIAL NERVE EPIC-PLUS System for auditory evoked potentials and neuro-otology Epic-plus, a unique solution. With

More information

ER75 Electro-Acoustic Ear Simulator. Operating Manual

ER75 Electro-Acoustic Ear Simulator. Operating Manual ER75 Electro-Acoustic Ear Simulator Operating Manual ABOUT THIS MANUAL READ THIS OPERATING MANUAL BEFORE ATTEMPTING TO USE THE INSTRUMENT. Amplivox Ltd. 6 Oasis Park, Eynsham Oxfordshire, OX29 4TP United

More information

GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER

GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Setting The Clinical Standard GSI AUDIOSTAR PRO CLINICAL TWO-CHANNEL AUDIOMETER Tradition of Excellence The GSI AudioStar Pro continues the tradition of

More information

AC40. Clinical Audiometer. Save time Focus on results. Interacoustics. l e a d i n g d i a g n o s t i c s o l u t i o n s

AC40. Clinical Audiometer. Save time Focus on results. Interacoustics. l e a d i n g d i a g n o s t i c s o l u t i o n s AC40 Clinical Audiometer Save time Focus on results Interacoustics l e a d i n g d i a g n o s t i c s o l u t i o n s AC40 Clinical Audiometer Two channel audiometer The AC40 is an advanced clinical audiometer

More information

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04

AURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04 AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by

More information

FONIX FA-12 Hearing Evaluator Digital Audiometer

FONIX FA-12 Hearing Evaluator Digital Audiometer FONIX FA-12 Hearing Evaluator Digital Audiometer Operator s Manual version 1.22 2011, Frye Electronics, Inc. All Rights Reserved 6/09 Revised 4/27/11 Note on this Manual We have organized this manual

More information

AT235/AT235h Middle Ear Analyzers. Diagnostic & clinical tympanometry with basic audiometry.

AT235/AT235h Middle Ear Analyzers. Diagnostic & clinical tympanometry with basic audiometry. AT235/AT235h Middle Ear Analyzers Diagnostic & clinical tympanometry with basic audiometry. Fast & reliable middle ear analyzer The AT235 is an automatic middle ear analyzer ideal for diagnostic and screening

More information

3M Center for Hearing Conservation

3M Center for Hearing Conservation 3M Center for Hearing Conservation Key Terms in Occupational Hearing Conservation Absorption A noise control method featuring sound-absorbing materials that are placed in an area to reduce the reflection

More information

Health, Safety, Security and Environment

Health, Safety, Security and Environment Document owner and change code Document Owner Aaron Perronne Title HSSE Manager Mark X Change Code Description X N/A First Issue A Typographical/Grammatical correction; formatting change; text clarification-no

More information

Provisional guidelines for using Auditory Steady State Responses (ASSR) in babies. July 2009

Provisional guidelines for using Auditory Steady State Responses (ASSR) in babies. July 2009 5 NEWBORN HEARING SCREENING AND ASSESSMENT Provisional guidelines for using Auditory Steady State Responses (ASSR) in babies 10 Suggested parameter values for recording frequency-specific ASSR using Natus

More information

AMBCO 1000+P AUDIOMETER

AMBCO 1000+P AUDIOMETER Model 1000+ Printer User Manual AMBCO 1000+P AUDIOMETER AMBCO ELECTRONICS 15052 REDHILL AVE SUITE #D TUSTIN, CA 92780 (714) 259-7930 FAX (714) 259-1688 WWW.AMBCO.COM 10-1004, Rev. A DCO 17 008, 11 13 17

More information

Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms

Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms Reaffirmed by ANSI October 15, 2003 Reaffirmed by ANSI October 28, 2008 ANSI S3.1-1999 (Revision of ANSI S3.1-1991) American National Standard Maximum Permissible Ambient Noise Levels for Audiometric Test

More information

Contents. QuickSIN.. 15 Hearing Loss Simulator Master Hearing Aid Printing Limited Warranty.. 20 #0086.

Contents. QuickSIN.. 15 Hearing Loss Simulator Master Hearing Aid Printing Limited Warranty.. 20 #0086. TRAINING MANUAL Contents Introduction.. 3 AVANT A2D+ Software Overview... 4 AVANT A2D+ Software Options.. 4 Preparing for Testing. 7 Pure Tone Audiometry. 11 Speech Audiometry... 13 QuickSIN.. 15 Hearing

More information

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods

C HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods C HAPTER FOUR Audiometric Configurations in Children Andrea L. Pittman Introduction Recent studies suggest that the amplification needs of children and adults differ due to differences in perceptual ability.

More information

PC BASED AUDIOMETER GENERATING AUDIOGRAM TO ASSESS ACOUSTIC THRESHOLD

PC BASED AUDIOMETER GENERATING AUDIOGRAM TO ASSESS ACOUSTIC THRESHOLD Volume 119 No. 12 2018, 13939-13944 ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu PC BASED AUDIOMETER GENERATING AUDIOGRAM TO ASSESS ACOUSTIC THRESHOLD Mahalakshmi.A, Mohanavalli.M,

More information

User Manual. - Pro Edition -

User Manual. - Pro Edition - User Manual - Pro Edition - Table of Contents Introduction 3 Patient Home Screen 4 Managing Patients 5 Creating Patients 5 Editing Patient Details 5 Deleting Patients 5 Starting Activities 6 Viewing Results

More information

Title: Evaluation of the Kuduwave 5000 audiometer for compliance with standards for hearing conservation purposes

Title: Evaluation of the Kuduwave 5000 audiometer for compliance with standards for hearing conservation purposes Title: Evaluation of the Kuduwave 5000 audiometer for compliance with standards for hearing conservation purposes Author: AL Edwards Affiliation: Council for Scientific and Industrial Research, Centre

More information

AS5 AUDIOMETER USE AND MAINTENANCE BOOK

AS5 AUDIOMETER USE AND MAINTENANCE BOOK AS5 AUDIOMETER USE AND MAINTENANCE BOOK ATTENTION: The operators must carefully read and completely understand the present manual before using the product. M33610 G Rev.2 06.12 GIMA Spa Via Marconi, 1

More information

Model Safety Program

Model Safety Program Model Safety Program DATE: SUBJECT: Occupational Noise Exposure Program REGULATORY STATUTE: OSHA 29 CFR 1910.95 RESPONSIBILITY: The company Safety Officer is. He/she is solely responsible for all facets

More information

Audiology Equipment p r o d u c t s c AtA l o g u E

Audiology Equipment p r o d u c t s c AtA l o g u E products catalogue 2014 Audiology Equipment Orchestral perfection A full line of Audiology Equipment News for 2014 Our Bell, Harp and Piano audiometers now feature HyTech, Inventis Hybrid Technology. You

More information

Operation Manual. Screening Audiometer AS608/AS608e

Operation Manual. Screening Audiometer AS608/AS608e Operation Manual Screening Audiometer AS608/AS608e Valid from serial 744081 80696221 ver. 10/2009 Table of Contents Table of Contents... 1 Introduction... 1 Intended Use... 1 Extended Function... 1 Precautions...

More information

User Manual. - Pro Edition -

User Manual. - Pro Edition - User Manual - Pro Edition - Table of Contents Introduction 3 Patient Home Screen 4 Managing Patients 5 Creating Patients 5 Editing Patient Details 5 Deleting Patients 5 Starting Activities 6 Viewing Results

More information

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial

Importance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment

More information

Operating Instructions MA 52

Operating Instructions MA 52 MAICO Diagnostic GmbH Operating Instructions MA 52 MAICO Diagnostic GmbH, Salzufer 13/14, D-10587 Berlin, Tel. +49 30/70714650, Fax +49 30/70714699 Table of Contents... Page 1 Introduction... 4 2 Intended

More information

Procedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003

Procedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003 Procedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003 1. Purpose 1.1. The purpose of this procedure is to establish a TVA Hearing Conservation Program (HCP)

More information

Testing FM Systems on the 7000 Hearing Aid Test System

Testing FM Systems on the 7000 Hearing Aid Test System Testing FM Systems on the 7000 Hearing Aid Test System Introduction Testing FM Systems on the 7000 Hearing Aid Test System This workbook describes how to test FM systems with the 7000 Hearing Aid Test

More information

The true clinical hybrid audiometer. Interacoustics. Clinical Audiometer AC40. Stand alone & PC-based audiometry in one box.

The true clinical hybrid audiometer. Interacoustics. Clinical Audiometer AC40. Stand alone & PC-based audiometry in one box. AC40 Clinical Audiometer true hybrid The true clinical hybrid audiometer Stand alone & PC-based audiometry in one box Interacoustics leading diagnostic solutions AC40 Clinical Audiometer A true hybrid

More information

Neuro-Audio Version 2010

Neuro-Audio Version 2010 ABR PTA ASSR Multi-ASSR OAE TEOAE DPOAE SOAE ECochG MLR P300 Neuro-Audio Version 2010 one device for all audiological tests Auditory brainstem response (ABR)/Brainstem evoked response audiometry (BERA)

More information

Digital. hearing instruments have burst on the

Digital. hearing instruments have burst on the Testing Digital and Analog Hearing Instruments: Processing Time Delays and Phase Measurements A look at potential side effects and ways of measuring them by George J. Frye Digital. hearing instruments

More information

Step-by-Step RECD Guide

Step-by-Step RECD Guide Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform

More information

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids

Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Acoustical Measurements and Instrumentation: Paper ICA2016-388 Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Zargos Neves Masson (a),

More information

TRAINING MANUAL. AVANT The Perfect Screening Audiometer AIR+

TRAINING MANUAL. AVANT The Perfect Screening Audiometer AIR+ TRAINING MANUAL AVANT The Perfect Screening Audiometer AIR+ www.medrx-usa.com Contents Introduction. 3 AVANT Audiometer Software Overview.. 4 AVANT Audiometer Software Options.... 4 Preparing for Testing.

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX FP40 Hearing Aid Analyzer FRYE 7 Introduction The following is a quick guide for testing digital hearing aids. All digital aids can be tested, but some of the

More information

L G Davison*, BSc C Barlow, PhD, MIOA M Ashmore, RHAD, MIOA R Weinstein, RHAD Southampton Solent University Faculty of Maritime and Technology

L G Davison*, BSc C Barlow, PhD, MIOA M Ashmore, RHAD, MIOA R Weinstein, RHAD Southampton Solent University Faculty of Maritime and Technology Effectiveness of a New Tablet Based Hearing Screening System Compared to Traditional Audiometry, Among a Cohort of Individuals With a High Incidence of Hearing Loss August 7, 213 L G Davison*, BSc C Barlow,

More information

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC.

FM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC. T E S T I N G FM SYSTEMS with the FONIX 6500-CX Hearing Aid Analyzer (Requires software version 4.20 or above) FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209

More information

Hearing Conservation Services Specifications

Hearing Conservation Services Specifications Hearing Conservation Services Specifications FACILITIES AND EQUIPMENT The following equipment is used in assessing hearing, middle ear function and noise levels: - Maico MA40 audiometer - Maico MA41 audiometer

More information

Noise reduction in modern hearing aids long-term average gain measurements using speech

Noise reduction in modern hearing aids long-term average gain measurements using speech Noise reduction in modern hearing aids long-term average gain measurements using speech Karolina Smeds, Niklas Bergman, Sofia Hertzman, and Torbjörn Nyman Widex A/S, ORCA Europe, Maria Bangata 4, SE-118

More information

The true clinical hybrid audiometer Stand alone & PC-based audiometry in one box. Interacoustics. Clinical Audiometer AC40.

The true clinical hybrid audiometer Stand alone & PC-based audiometry in one box. Interacoustics. Clinical Audiometer AC40. AC40 Clinical Audiometer true hybrid The true clinical hybrid audiometer Stand alone & PC-based audiometry in one box Interacoustics leading diagnostic solutions AC40 Clinical Audiometer A true hybrid

More information

AUDIOLAB Two channel clinical audiometer. Tonal Audiometry Vocal Audiometry MASKING SISI ABLB Decay Test BEKESY

AUDIOLAB Two channel clinical audiometer. Tonal Audiometry Vocal Audiometry MASKING SISI ABLB Decay Test BEKESY AUDIOLAB Two channel clinical audiometer Tonal Audiometry Vocal Audiometry MASKING SISI ABLB Decay Test BEKESY AUDIOLAB Two channel clinical audiometer AudioLab, advanced technology in a small package.

More information

Operating Instructions MAICO ST 20

Operating Instructions MAICO ST 20 MAICO Diagnostic GmbH Operating Instructions MAICO ST 20 MAICO Diagnostic GmbH, Salzufer 13/14, D 10587 BERLIN, Tel. +49 30-70714650, Fax +49 30-70714699 Table of Contents Page 1 Introduction 2 2 Description

More information

Anatomy of the Ear. Hearing Screening and Audiometer Tutorial. The speaker has no conflict interest to declare

Anatomy of the Ear. Hearing Screening and Audiometer Tutorial. The speaker has no conflict interest to declare Hearing Screening and Audiometer Tutorial The speaker has no conflict interest to declare of This program did not receive any commercial support Anatomy of the Ear 1 Types of Hearing Loss Conductive Sensorineural

More information

TRAINING MANUAL. MedRx TINNOMETER. Revolutionary Tinnitus Assessment.

TRAINING MANUAL. MedRx TINNOMETER. Revolutionary Tinnitus Assessment. TRAINING MANUAL Revolutionary Tinnitus Assessment MedRx TINNOMETER www.medrx-int.com Contents Introduction. 3 AVANT Audiometer Software Overview.. 4 AVANT Audiometer Software Options.... 4 Preparing for

More information

User Guide. For Jacoti Hearing Center Pro Version 1.0. Manufacture Year 2017

User Guide. For Jacoti Hearing Center Pro Version 1.0. Manufacture Year 2017 User Guide For Jacoti Hearing Center Pro Version 1.0 Manufacture Year 2017 Revision 26/10/2017 Table of contents 1_ Intended Use... 3 1.1_ Requirements... 3 1.2_ Headphones... 3 1.3_ Startup... 4 1.4_

More information

ACOUSTIC INFRASOUND AND LOW-FREQUENCY SOUND

ACOUSTIC INFRASOUND AND LOW-FREQUENCY SOUND 116 Acoustic ACOUSTIC INFRASOUND AND LOW-FREQUENCY SOUND These limits represent sound exposures to which it is believed nearly all workers may be repeatedly exposed without adverse effects that do not

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 11904-1 First edition 2002-10-01 Acoustics Determination of sound immission from sound sources placed close to the ear Part 1: Technique using a microphone in a real ear (MIRE

More information

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015.

DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. www.bestsound-technology.com DSL v5 in Connexx 7 Mikael Menard, Ph.D., Philippe Lantin Sivantos, 2015. First fit is an essential stage of the hearing aid fitting process and is a cornerstone of the ultimate

More information

AMBCO AUDIOMETER

AMBCO AUDIOMETER Model 1000+ User Manual AMBCO 1000+ AUDIOMETER AMBCO ELECTRONICS 15052 REDHILL AVE SUITE #D TUSTIN, CA 92780 (714) 259-7930 FAX (714) 259-1688 WWW.AMBCO.COM 10-1003, Rev. A DCO 17 008, 11 13 17 Model 1000+

More information

Instructions for Use. Oscilla TSM400 Screening Tympanometer. Specifications are subject to change without notice ID: 6795 / ver.

Instructions for Use. Oscilla TSM400 Screening Tympanometer. Specifications are subject to change without notice ID: 6795 / ver. Instructions for Use Oscilla TSM400 Screening Tympanometer Specifications are subject to change without notice 2017-02-09 1 Contents General description... 3 Installation... 4 Device overview... 5 Rear

More information

Aalborg Universitet. Published in: International Journal of Audiology. Publication date: Link to publication from Aalborg University

Aalborg Universitet. Published in: International Journal of Audiology. Publication date: Link to publication from Aalborg University Aalborg Universitet Preferred test conditions for determining hearing thresholds for standardization / ISO/TC 43/WG1 "Threshold of Hearing", convened by Henrik Møller Møller, Henrik Published in: International

More information

The following describes general use of ELIOS device

The following describes general use of ELIOS device General use of the device The following describes general use of ELIOS device 1. Introduction of graphical user interface Presence indicator of SD card Charging indicator General Informations Screen title

More information

Hearing Conservation Program

Hearing Conservation Program Hearing Conservation Program Table of Contents I. Program Goals and Objectives... 2 II. Scope and Application... 2 III. Regulatory Authority and Related Information... 2 IV. Definitions... 2 V. Responsibilities...

More information

Hearing Solutions Catalog Winter Supplement January 2009

Hearing Solutions Catalog Winter Supplement January 2009 Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with

More information

Santa Clarita Community College District HEARING CONSERVATION PROGRAM. Revised

Santa Clarita Community College District HEARING CONSERVATION PROGRAM. Revised Santa Clarita Community College District HEARING CONSERVATION PROGRAM Revised March 2018 TABLE OF CONTENTS INTRODUCTION...1 HEARING CONSERVATION PROGRAM...2 1.1 District Policy...2 1.2 Plan Review...2

More information

Knowledge Sharing: Pediatric Amplification Tips and Tricks. Implementing Best Practice

Knowledge Sharing: Pediatric Amplification Tips and Tricks. Implementing Best Practice Knowledge Sharing: Pediatric Amplification Tips and Tricks Implementing Best Practice KIPAT Knowledge Implementation in Pediatric Audiology Team Lisa Davidson, Ph.D., CID Research at Washington University

More information

Testing Digital Hearing Aids

Testing Digital Hearing Aids Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All

More information

Hearing Protection Systems

Hearing Protection Systems March 2018 ANR, AHNR, In-Ear How to Choose? Simon Smith Chief Engineer Gentex Europe ssmith@gentexcorp.eu.com +44 (0)1462 478000 Michelle Collier Product Manager Aircrew Hearing Protection & Communications

More information

Recommended Procedure. Pure-tone air-conduction and boneconduction threshold audiometry with and without masking

Recommended Procedure. Pure-tone air-conduction and boneconduction threshold audiometry with and without masking Pure-tone air-conduction and boneconduction threshold audiometry with and without masking Date: September [Minor amendments: February 2012, and December 2015] Due Date for of review: September Page2 General

More information

How to use AutoFit (IMC2) How to use AutoFit (IMC2)

How to use AutoFit (IMC2) How to use AutoFit (IMC2) How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain

More information

THE MECHANICS OF HEARING

THE MECHANICS OF HEARING CONTENTS The mechanics of hearing Hearing loss and the Noise at Work Regulations Loudness and the A weighting network Octave band analysis Hearing protection calculations Worked examples and self assessed

More information

Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids

Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids J Am Acad Audiol 12 : 281-285 (2001) Suitcase Lab Measurement of Digital Cellphone Interference Levels on Hearing Aids Mead C. Killion* Harry Teder' Russ Thoma* Abstract A low-cost, "real-life" method

More information

Operating Instructions MA 41

Operating Instructions MA 41 MAICO Diagnostic GmbH Operating Instructions MA 41 MAICO Diagnostic GmbH Salzufer 13/14 D-10587 Berlin Tel.: +49 30 707146-50 Fax: -99 PAGE INTENTIONALLY LEFT BLANK Table of Contents Page 1 Introduction...

More information