1 RESEARCH Validation of Resonance Frequency Analysis by Comparing Implant Stability Quotient Values With Histomorphometric Data Ashish Thomas Kunnekel, MDS 1 * K. Chandrasekharan Nair, MDS 2 E. Munirathnam Naidu, MDS 3 Gomathinayagam Sivagami, MDS 4 The study was designed to determine the relationship between implant stability quotient (ISQ) values measured using resonance frequency analysis (RFA) and implant-bone distance measured histomorphometrically. Ten identical implants were equally divided into 2 groups based on primary stability at placement. Osteotomies were prepared in harvested goat femurs. ISQ values were measured and compared with implant-bone distance determined by micrometry. Based on the results, it was concluded that RFA can be used to measure implant stability reliably. Key Words: implant stability quotient, resonance frequency analysis, implant-bone interface INTRODUCTION It is of interest to develop and evaluate assessment methods for implant stability, both for research purposes and for prognostic evaluation that led to the development of resonance frequency analysis (RFA) by Meredith in ,2 The use of RFA to assess implant stability has been validated by several authors. 3 9 Despite 1 Department of Prosthodontics, Maulana Azad Institute of Dental Sciences, New Delhi, India. 2 Department of Prosthodontics, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, India. 3 Meenakshi Ammal Dental College, Meenakshi University, Chennai, Tamil Nadu, India. 4 Department of Prosthodontics and Implantology, Meenakshi Ammal Dental College, Chennai, India. * Corresponding author, drkunnekel@gmail.com DOI: /AAID-JOI-D widespread acceptance of RFA, as reflected by its extensive application in current research, its reliability is still debated The objective of this study was to validate the accuracy of RFA by measuring the implant stability quotient (ISQ) values with the Osstell instrument (Osstell AB, Gothenburg, Sweden) and correlating it with the implant-bone distance measured histomorphometrically. MATERIALS AND METHODS Implant specifications Ten indigenously fabricated commercially pure titanium, single piece, root form, threaded implants with sandblasted surfaces were used (Figure 1). All implants were Journal of Oral Implantology 301

2 Validation of Resonance Frequency Analysis Using Histomorphometry 4.7 mm in diameter at the level of the finish margin and 12 mm in length. An access channel for the SmartPeg (Osstell AB) of the Osstell instrument was machined in the implant. The implants are developed under the brand name Idea. Study groups Implants were equally divided into 2 groups: 1. The tight fit or control group comprised implants placed in osteotomy sites prepared according to the manufacturer s recommendations to ensure primary stability. 2. The loose fit or test group comprised implants placed in oversized osteotomy sites in relation to the implant dimensions to artificially induce implant instability. Harvesting and storage of femurs Five femurs were harvested from 3 goats immediately post mortem. Osteotomy preparations were commenced immediately following harvesting. Selection of osteotomy sites Ten osteotomy sites (5 test and 5 control) were identified in 5 femurs. The medial epiphyses of the femurs were selected as sites for implant placement because of the presence of adequate trabecular bone. Two sites (1 test and 1 control) were selected in each epiphysis. Radiovisiographs (RVGs) of the proposed sites were taken and analyzed to confirm the presence of adequate bone. Sites were randomly assigned to each group (Figure 2). Osteotomy preparations All osteotomies were prepared by the same operator using a drilling speed of 2500 rpm, under copious irrigation with normal saline. Tight Fit (Control) Group Penetration of cortical bone was achieved with no. 6 round burs. A pilot drill 2.3 mm in diameter was used to prepare 12-mm deep osteotomies. The osteotomies were progressively enlarged using 2.8-mm and 3.4-mm diameter drills, until a depth of 12 mm was achieved. The osteotomies were widened to a depth of 8 mm from the crest using a 3.8- mm drill. The final size of the control osteotomies were 3.8 mm in diameter until a depth of 8 mm, and 3.3 mm in diameter from a depth of 8 mm to 12 mm. Loose Fit (Test) Groups Osteotomy preparations for the loose fit group were identical to those of the tight fit group until preparation with the 3.4-mm drill. Subsequently, 3.8 mm, 4.4 mm, and a final drill diameter of 4.8 mm were used to enlarge the osteotomies. The final size of the test osteotomies was 4.8 mm in diameter and 12 mm in depth. Implant insertion Tight Fit (Control) Implants Control implants were placed into designated osteotomies. The implants were rotated to the full depth of the osteotomies using a ratchet. Complete seating was visually verified by ensuring that the finish margins were at the level of the crest of the bone (Figure 3) and radiographically confirmed with a radiovisiograph (RVG) (Figure 4). Primary stability of the implants was verified by exerting lateral forces and confirming absence of motion. Loose Fit (Test) Implants Test implants were placed into osteotomies corresponding to the loose fit group. Implants were seated to their final positions in a manner identical to that described for the control implants. Lack of primary stability was verified by exerting lateral forces and noting motion. Measurement of implant stability quotient values using RFA Tight Fit (Control) Implants The SmartPeg of the Ostell instrument was screwed into the access channels in the 302 Vol. XXXVII/No. Three/2011

3 Kunnekel et al FIGURES 1 6. FIGURE 1. Implants used. FIGURE 2. Radiovisiograph (RVG) to identify osteotomy sites. FIGURE 3. Implant in final position. FIGURE 4. RVG of tight fit (control) and loose fit (test) implants in final position. FIGURE 5. Measurement of implant stability quotient (ISQ). FIGURE 6. ISQ value recorded in Ostell instrument. control implants according to the manufacturer s instructions. Three ISQ values were obtained from each of the following positions: 3, 6, 9, and 12 o clock (Figure 5). The data were automatically stored in the device and subsequently transferred to a personal computer (Figure 6). Loose Fit (Test) Implants Measurement of ISQ values for the loose fit (test) implants was done in a manner identical to that used for the tight fit (control) implants, as described above. Processing of specimens for histomorphometric analysis The tight fit (control) and loose fit (test) implants with 5 mm of surrounding bone were removed en bloc using a diamond disc mounted in a straight handpiece under copious irrigation with tap water (Figure 7). The specimens were fixed by immediate placement in 10% formalin solution, and dehydrated by immersion in 70% acetone, followed by 3 changes of absolute acetone, for durations of 2 hours, 90 minutes, 90 minutes, and 2 hours, respectively. Journal of Oral Implantology 303

4 Validation of Resonance Frequency Analysis Using Histomorphometry FIGURES FIGURE 7. Retrieved specimen en bloc. FIGURE 8. Specimen embedded in autopolymerizing resin. FIGURE 9. Sectioned specimen. FIGURE 10. Ground section of specimens. Resin embedding of the specimens in clear autopolymerizing resin was done by infiltration with methylmethacrylate monomer followed by inducing polymerization by adding polymer powder (Figure 8). Specimens were sectioned using a diamond disc mounted in a straight handpiece at a speed of 1500 rpm, under copious irrigation with normal saline. All implant abutments were severed at the level of the finish margins. Embedded specimens of both groups were sectioned longitudinally through the centers of the implants into 2mm thick slices (Figure 9). Fifty micron thick ground sections of the specimens were prepared by luting them to glass slides with cyanoacrylate followed by grinding with Emery paper of numbers 220, 280, 320, 400, and 600 sequentially under copious irrigation with normal saline. Adequate reduction of thickness was assessed by viewing the sections under a light microscope. 304 Vol. XXXVII/No. Three/2011 Specimens of both groups were stained with hematoxylin and eosin (Figure 10). Histomorphometric analysis and measurement of implant-bone distance Specimens of both groups were histomorphometrically analyzed using a light microscope. Implant-bone distance was measured using an ocular micrometer at 340 and 3100 magnifications, and were expressed in microns. Measurements were made at standardized locations for all specimens, namely at the first two threads on the right and left sides of the implants. Two readings were made in each region and the data were statistically analyzed (Figures 11 through 14). RESULTS Relationship between distance and ISQ values was assessed by Pearson s correlation analysis. A P value of.05 was considered as the level of significance.

5 Kunnekel et al FIGURES FIGURE 11. Histomorphometric image of interface of tight fit (control) specimen; implant and bone (340 magnification). FIGURE 12. Histomorphometric image of interface of loose fit (test) specimen; implant and bone (340 magnification). FIGURE 13. Histomorphometric image of interface of tight fit (control) specimen; implant and bone (3100 magnification). FIGURE 14. Histomorphometric image of interface of loose fit (test) specimen; implant and bone (3100 magnification). Tight Fit (Control) Group The highest and lowest mean ISQ values for the tight fit group were and 58.91, respectively. The means of the largest and smallest implant-bone distance were 183 mm and 129 mm, respectively (Table 1). The correlation between the mean values of implant-bone distance and ISQ values for the loose fit (test) group are presented in Table 2. Highest mean ISQ values of corresponded to the lowest implant-bone distance with a value of 147 mm. Lowest ISQ values of corresponded to the highest implantbone distance with a value of 180 mm. A negative relationship was found between the implant-bone distance and ISQ values. However, it was not statistically significant (r , P 5.19) (Table 3). Loose Fit (Test) Group The highest and lowest mean ISQ values were and 33.16, respectively (Table 2). The means of the largest and smallest gaps were 624 mm and 792 mm, respectively. Highest mean ISQ values of corresponded to the TABLE 1 Mean values of implant-bone distance (mm) and implant stability quotient (ISQ) for tight fit (control) group Specimen No. Mean Implant-Bone Distance (mm) SD ISQ Values SD Journal of Oral Implantology 305

6 Validation of Resonance Frequency Analysis Using Histomorphometry TABLE 2 Mean values of implant-bone distance (mm) and implant stability quotient (ISQ) for loose fit (test) group Specimen No. Mean Implant-Bone Distance (mm) SD ISQ Values SD least mean gap at the interface with a value of 624 mm. Least mean ISQ values of corresponded to the largest mean gap at the interface, with a value of 792 mm. A negative relationship was found between the implant-bone distance and ISQ values. However, it was not statistically significant (r , P 5.12) (Table 3). DISCUSSION While no gold standard exists to measure implant stability, 10 several methods have been proposed for the same, namely clinical assessment by exerting lateral forces with 2 opposing mirror handles, percussion, Dental Fine Tester (Kyocera), Periotest device, pullout and push-out tests, removal torque testing, and histologic and histomorphometric methods. 1,4 6,13 The accuracy and practicality of these methods has been widely challenged in the literature. One objective of this study was to quantify the stability of endosseous implants by measuring ISQ values using RFA. Undersized and oversized osteotomy preparations, in relation to implant dimensions were used TABLE 3 Correlation between implant-bone distance (mm) and implant stability quotient (ISQ) between tight fit (control) and loose fit (test) groups Group Correlation Coefficient P Value Tight fit (control) group Loose fit (test) group to simulate high and low primary stability conditions, according to a previous study. 6 The variable clinical stability of the implants was reflected by the resultant ISQ values, with higher values recorded for stable implants placed in undersized osteotomies and lower values recorded for mobile implants placed in oversized osteotomies. In a clinical situation, this would equate to higher values in case of clinically stable implants, and lower values in case of clinically mobile implants. These findings are in accordance with previous reports. 6 Mechanically, the resonance frequency (RF) of an object is strongly correlated to its boundary constraints. 13 ISQ values reflect the stiffness of the implant-bone interface The implant-bone interface was artificially designed to be stiffer in the tight fit or control group implants. The amount of surrounding bone in contact with the implants was greater as compared to the test group. These conditions were well reflected by the higher ISQ values for the control group, and are in accordance with findings of other studies. 6 8 Measurement of implant stability using RFA merely reflects on the implant-bone contact. It was hypothesized that ISQ values are related to the proximity of the bone to the implant and higher and lower ISQ values would correspond to lesser and greater implant-bone distances, respectively. The implant-bone distance was quantified by histomorphometric analysis of ground sections, which is a proven accurate method Vol. XXXVII/No. Three/2011

7 Kunnekel et al A negative correlation between ISQ values and the width of the interface within the tight fit and loose fit groups indicates that lower and higher ISQ values were found in cases of wider and narrower interfaces, respectively. In other words, ISQ values are inversely proportional to the implant-bone distance. These findings conform to the principles of basic vibrational theory, which states that the RF value of a structure is strongly affected by its effective vibrational length and the tightness of the boundary contact. The results are in accordance with other studies, where ISQ values were correlated to the density and cortical to the trabecular bone ratio at the interface. 1,5 Lower ISQ values have been measured following implantation in low-density bone compared to bone of high density. 6,14 This could be attributed to decreased stiffness of the interface in the former. 1 Bone of low density, containing a large volume of trabecular bone, will manifest with larger gaps at the interface as compared with densely packed compact bone. Results of this experiment are in accordance with these facts. A statistically significant correlation within and between the groups was not found. It could be speculated that this was due to the small sample size used in this study. Perhaps a significant correlation may be derived with further research with a larger sample size, which is not a realistic proposal when the constraints of animal experiments are considered. Types of tissue reactions to implantation depend on the size of the gaps between the implant and bone cavity surface. The critical threshold of micromotion above which fibrous encapsulation results instead of osseointegration is not 0 but seen to range between 50 and 150 mm. 16,17 The findings of this study indicate that high ISQ values showing clinically determined primary stability correspond to an interface of an average width of 158 mm. RFA presented higher ISQ values for deliberately stabilized implants. Results of this study indicate that RFA provides a good indication of clinical stability of an implant and the implant-bone distance. CONCLUSIONS The following conclusions were drawn from this study: 1. Stable implants present with higher ISQ values compared with mobile implants when subjected to RFA. 2. Higher ISQ values are indicative of lesser implant-bone distance and vice versa. 3. Clinical stability of implants is a function of the width of the implant-bone interface. ABBREVIATIONS ISQ: implant stability quotient RF: resonance frequency RFA: resonance frequency analysis RVG: radiovisiograph ACKNOWLEDGMENTS We extend our gratitude to Dr B. Sivapathasundharam, Dr Saraswathi, and the entire team at the Department of Oral and Maxillofacial Pathology, Meenakshi Ammal Dental College and Hospital for their expert guidance and help. REFERENCES 1. Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont. 1998:11; Aparacio C, Lang NP, Rangert B. Validity and clinical significance of biomechanical testing of implant/bone interface. Clin Oral Implants Res. 2006; 17(suppl 2): Barewal RM, Oates TW, Meredith N, Cochran DL. Resonance frequency measurement of implant stability in vivo on implants with a sandblasted and acid-etched surface. Int J Oral Maxillofac Implants. 2003;18: Nkenke E, Hahn M, Weinzierl K, Radespiel-Troger M, Neukam FW, Engelke K. Implant stability and histomorphometry: a correlation study in human cadavers. Clin Oral Implants Res. 2003;14: Journal of Oral Implantology 307

8 Validation of Resonance Frequency Analysis Using Histomorphometry 5. Huang HM, Chiu CL, Yeh LC, Lin CT, Lin LH, Lee SY. Early detection of implant healing process using resonance frequency analysis. Clin Oral Implants Res. 2003;14: Huang HM, Lee SY, Yeh CY, Lin CT. Resonance frequency assessment of dental implant stability with various bone qualities: a numerical approach. Clin Oral Implants Res. 2002;13: Gedrange T, Heitschold V, Mai R, Wolf P, Nicklisch M, Harzer W. An evaluation of resonance frequency analysis for the determination of the primary stability of orthodontic palatal implants: a study in human cadavers. Clin Oral Implants Res. 2005;16: Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res. 1997;8: Friberg B, Sennerby L, Linden B, Grondahl K, Lekholm U. Stability measurements of one-stage Branemark implants during healing in mandibles: a clinical resonance frequency analysis study. Int J Oral Maxillofac Surg. 1999;28: Al-Nawas B, Wagner W, Grotz KA. Insertion torque and resonance frequency analysis of dental implant systems in an animal model with loaded implants. Int J Oral Maxillofac Implants. 2006;21: Acka K, Chang TL, Tekdemir I, Fanuscu MI. Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis. Clin Oral Implants Res. 2006;17: Friberg B, Sennerby L, Meredith N, Lekholm U. A comparison between cutting torque and resonance frequency measurements of maxillary implants: a 20 month clinical study. Int J Oral Maxillofac Surg. 1999;28: Isidor F. Influence of forces on peri-implant bone. Clin Oral Implants Res. 2006;17(suppl 2): Berglundh T, Abrahamsson I, Lang NP, Lindhe J. De novo alveolar bone formation adjacent to endosseous implants: a model study in the dog. Clin Oral Implants Res. 2003;14: Friberg B, Sennerby L, Linden B, Grondahl K, Lekholm U. Stability measurements of one-stage Branemark implants during healing in mandibles: A clinical resonance frequency analysis study. Int J Oral Maxillofac Surg. 1999;28: Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH. Timing of loading and effect of micromotion on bone-dental implant interface: review of experimental literature. J Biomed Mater Res. 1998;43: Bischof M, Nedir R, Szmukler-Moncler S, Bernard JP, Samson J. Implant stability measurement of delayed and immediately loaded implants during healing: a clinical resonance frequency-analysis study with sandblasted-and-etched ITI implants. Clin Oral Implants Res. 2004;15: Vol. XXXVII/No. Three/2011

Architecture and microenvironment

Architecture and microenvironment RESEARCH Comparison of Delayed Implant Placement vs Immediate Implant Placement Using Resonance Frequency Analysis: A Pilot Study on Rabbits Ashish T. Kunnekel, MDS 1 * Mohit T. Dudani, MDS 2 Chandrasekharan

More information

Hiron Andreaza da Cunha, MS 1 /Carlos Eduardo Francischone, DMD 2 /Hugo Nary Filho, DDS 3 / Rubelisa Cândido Gomes de Oliveira, BDS 4

Hiron Andreaza da Cunha, MS 1 /Carlos Eduardo Francischone, DMD 2 /Hugo Nary Filho, DDS 3 / Rubelisa Cândido Gomes de Oliveira, BDS 4 A Comparison Between Cutting Torque and Resonance Frequency in the Assessment of Primary Stability and Final Torque Capacity of and TiUnite Single-Tooth Implants Under Immediate Loading Hiron Andreaza

More information

DEVELOPMENT OF PREDICTABLE STABILITY TEST FOR ASSESSMENT OF OPTIMUM LOADING TIME IN DENTAL IMPLANT

DEVELOPMENT OF PREDICTABLE STABILITY TEST FOR ASSESSMENT OF OPTIMUM LOADING TIME IN DENTAL IMPLANT ORIGINAL ARTICLE DOI:10.4047/jkap.2008.46.6.628 DEVELOPMENT OF PREDICTABLE STABILITY TEST FOR ASSESSMENT OF OPTIMUM LOADING TIME IN DENTAL IMPLANT Seong-Kyun Kim 1, DDS, PhD, Seong-Joo Heo 2, DDS, PhD,

More information

Osseointegrated dental implant treatment generally

Osseointegrated dental implant treatment generally Placement of Dental Implants Without Flap Surgery: A Clinical Report Bader H. Al-Ansari, BDS, MScD*/Robert R. Morris, DMD** Traditionally, the procedure of implant placement requires a surgical periosteal

More information

Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible

Predictability of the resonance frequency analysis in the survival of dental implants placed in the anterior non-atrophied edentulous mandible Journal section: Oral Surgery Publication Types: Research doi:10.4317/medoral.16982 http://dx.doi.org/doi:10.4317/medoral.16982 Predictability of the resonance frequency analysis in the survival of dental

More information

Assessment of Implant Stability of Two-piece Zirconium Dioxide Implants using the Resonance Frequency Analysis: A Pilot Study

Assessment of Implant Stability of Two-piece Zirconium Dioxide Implants using the Resonance Frequency Analysis: A Pilot Study IJEDS Assessment of Implant Stability of Two-piece Zirconium Dioxide Implants using 10.5005/jp-journals-10029-1103 the Resonance Frequency Analysis RESEARCH ARTICLE Assessment of Implant Stability of Two-piece

More information

Enhanced Implant Stability with a Chemically Modified SLA Surface: A Randomized Pilot Study

Enhanced Implant Stability with a Chemically Modified SLA Surface: A Randomized Pilot Study Enhanced Implant Stability with a Chemically Modified SLA Surface: A Randomized Pilot Study Thomas W. Oates, DMD, PhD 1 /Pilar Valderrama, DDS, MS 2 /Mark Bischof, Dr Med Dent 3 / Rabah Nedir, Dr Med Dent

More information

Primary implant stability has been identified as a major

Primary implant stability has been identified as a major CLINICAL Osstell Resonance Frequency Measurement Values as a Prognostic Factor in Implant Dentistry Tim Krafft, PD, Dr. med. DMD 1 Friedrich Graef, Dr. rer. nat. 2 Matthias Karl, PD, DMD 3 * Resonance

More information

Assessment of Implant Stability- A Critical Review

Assessment of Implant Stability- A Critical Review Assessment of Implant Stability- A Critical Review Dr. Ginnia Bhayana 1, Dr. Ferah Rehman 2, Dr. Vinod Khanna 3, Dr. Sonia 4, Dr. Taruna Arora 5 1 Demonstrator (MDS), Department of Prosthodontics, PGIDS

More information

Local alveolar bone quality and quantity,

Local alveolar bone quality and quantity, RESEARCH Use of Osteotomes for Implant Bed Preparation Effect on Material Properties of Bone and Primary Implant Stability Tim Krafft, PD Dr Dr Friedrich Graef, Dr Werner Winter, Dr-Ing Manfred Wichmann,

More information

Implant stability, whether immediately following surgery

Implant stability, whether immediately following surgery RESEARCH The Evaluation of the Reliability of Periotest for Implant Stability s: An In Vitro Study Hakan Bilhan, PhD, DDS 1 Altug Cilingir, PhD, DDS 2 Canan Bural, PhD, DDS 3 Caglar Bilmenoglu, PhD, DDS

More information

Osstell ISQ. The objective way to measure implant stability.

Osstell ISQ. The objective way to measure implant stability. Osstell ISQ. The objective way to measure implant stability. www.osstell.com Three reasons why you need Osstell ISQ. Dr Tiziano Testori, Private Practice, Como, Italy: I use the Osstell meter as a quality-assurance

More information

Pressure Necrosis And Osseointegration: An Editorial White Paper*

Pressure Necrosis And Osseointegration: An Editorial White Paper* Pressure Necrosis And Osseointegration: An Editorial White Paper* Alan Meltzer, DMD, MScD; Harold Baumgarten, DMD; Tiziano Testori, MD, DDS; Paolo Trisi, DDS, PhD Published by Pressure Necrosis And Osseointegration:

More information

EFFECT OF IMPLANT DESIGN ON INITIAL STABILITY

EFFECT OF IMPLANT DESIGN ON INITIAL STABILITY CLINICAL EFFECT OF IMPLANT DESIGN ON INITIAL STABILITY OF TAPERED IMPLANTS Linus Chong, DDS, MS; Ahmed Khocht, DDS, MSD; Jon B. Suzuki, DDS, PhD; John Gaughan, PhD Implant design is one of the parameters

More information

Enhancing implant stability with osseodensification a case report with 2-year follow-up

Enhancing implant stability with osseodensification a case report with 2-year follow-up Enhancing implant stability with osseodensification a case report with 2-year follow-up Dr. Salah Huwais discusses how osseodensification facilitates ridge expansion with enhanced implant stability Introduction

More information

Dental Research Journal

Dental Research Journal Dental Research Journal Original Article The effect of shape, length and diameter of implants on primary stability based on resonance frequency analysis Hamidreza Barikani, Shadab Rashtak, Soolmaz Akbari,

More information

Resonance Frequency Measurement of Implant Stability In Vivo on Implants with a Sandblasted and Acid-Etched Surface

Resonance Frequency Measurement of Implant Stability In Vivo on Implants with a Sandblasted and Acid-Etched Surface Resonance Frequency Measurement of Implant Stability In Vivo on Implants with a Sandblasted and Acid-Etched Surface Reva M. Barewal, DDS, MS 1 /Thomas W. Oates, DMD, PhD 2 /Neil Meredith, BDS, MSc, PhD

More information

OSSEODENSIFICATION: A Paradigm Shift In Implant Osteotomy Preparation. Introducing Densah Bur Technology

OSSEODENSIFICATION: A Paradigm Shift In Implant Osteotomy Preparation. Introducing Densah Bur Technology OSSEODENSIFICATION: A Paradigm Shift In Implant Osteotomy Preparation Introducing Densah Bur Technology Densah Bur Technology * The Innovation That Makes Osseodensification Possible A New Paradigm In Dental

More information

The Implant Stability Quotient Whitebook. The relationship between reliable diagnostics and safe, successful dental implant procedures 1 st edition

The Implant Stability Quotient Whitebook. The relationship between reliable diagnostics and safe, successful dental implant procedures 1 st edition The Implant Stability Quotient Whitebook The relationship between reliable diagnostics and safe, successful dental implant procedures 1 st edition Executive summary 4 Foreword by Professor Lars Sennerby,

More information

Influence of Osseointegration Degree and Pattern on Resonance Frequency in the Assessment of Dental Implant Stability Using Finite Element Analysis

Influence of Osseointegration Degree and Pattern on Resonance Frequency in the Assessment of Dental Implant Stability Using Finite Element Analysis Influence of Osseointegration Degree and Pattern on Resonance Frequency in the Assessment of Dental Implant Stability Using Finite Element Analysis Bin Deng, BDS, PhD 1 /Keson B. Tan, BDS (Hons), MSD 2

More information

Oral implants have been used

Oral implants have been used RESEARCH Early Detection of Alterations in the Resonance Frequency Assessment of Oral Implant Stability on Various Bone Types: A Clinical Study Metin Şençimen, DDS, PhD 1 Aydın Gülses, DDS 1 * Jülide Özen,

More information

An evaluation of TimPlant using Osstell: A device for non-invasive assessment of dental implants stability

An evaluation of TimPlant using Osstell: A device for non-invasive assessment of dental implants stability An evaluation of TimPlant using Osstell: A device for non-invasive assessment of dental implants stability Sebastian Soicu 1, Emanuel Bratu 2, Dorin Bratu 3, Ioan Borsanu 4 1 D.M.D. Department of Prosthodontics,

More information

4766 Research Dr. San Antonio, TX insightdentalsystems.com

4766 Research Dr. San Antonio, TX insightdentalsystems.com OVERVIEW OF THE INSIGHT DENTAL IMPLANT DELIVERY SYSTEM The IDS system comes in a unit dose implant system where its advantage provides sterile instrumentation in one single-use kit. It is organized to

More information

Comparison of Implant Stability Using Resonance Frequency Analysis: Osteotome Versus Conventional Drilling

Comparison of Implant Stability Using Resonance Frequency Analysis: Osteotome Versus Conventional Drilling Original Article Comparison of Implant Stability Using Resonance Frequency Analysis: Osteotome Versus Conventional Drilling Rokhsareh Sadeghi 1, Amir Reza Rokn 2, Asghar Miremadi 3 1 Assistant Professor,

More information

Using immediate implant placement to address aesthetic and biomechanical challenges: A clinical presentation

Using immediate implant placement to address aesthetic and biomechanical challenges: A clinical presentation Clinical Using immediate implant placement to address aesthetic and biomechanical challenges: A clinical presentation Pär-Olov Östman 1 Immediate implant loading has become an attractive option for meeting

More information

Resonance Frequency Analysis Measurements of Implants at Placement Surgery

Resonance Frequency Analysis Measurements of Implants at Placement Surgery Resonance Frequency Analysis Measurements of Implants at Placement Surgery Pär-Olov Östman, DDS a /Mats Hellman, DDS a /Inger Wendelhag, PhD b /Lars Sennerby, DDS, PhD c Purpose: The knowledge of what

More information

MANAGEMENT OF ATROPHIC ANTERIOR MAXILLA USING RIDGE SPLIT TECHNIQUE, IMMEDIATE IMPLANTATION AND TEMPORIZATION

MANAGEMENT OF ATROPHIC ANTERIOR MAXILLA USING RIDGE SPLIT TECHNIQUE, IMMEDIATE IMPLANTATION AND TEMPORIZATION Case Report International Journal of Dental and Health Sciences Volume 02, Issue 06 MANAGEMENT OF ATROPHIC ANTERIOR MAXILLA USING RIDGE SPLIT TECHNIQUE, IMMEDIATE IMPLANTATION AND TEMPORIZATION Rakshith

More information

The International Journal of Periodontics & Restorative Dentistry

The International Journal of Periodontics & Restorative Dentistry The nternational Journal of Periodontics & Restorative Dentistry 2009 BY QUNTESSENCE PUBLSHNG CO, NC. PRNTNG OF THS DOCUMENT S RESTRCTED TO PERSONAL USE ONLY. NO PART OF THS ARTCLE 549 Time Sequence of

More information

Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis.

Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis. Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis. Atieh MA, Ibrahim HM, Atieh AH. J Periodontol. 2010 Oct;81(10):1350-66. BACKGROUND: Platform

More information

Case study 2. A Retrospective Multi-Center Study on the Spiral Implant

Case study 2. A Retrospective Multi-Center Study on the Spiral Implant Case study 2 A Retrospective Multi-Center Study on the Spiral Implant Benny Karmon DMD, Jerry Kohen DMD, Ariel Lor DMD, Yiftach Gratciany DMD, Zvi Laster DMD, Gideon Hallel DMD MPA, Tsvia Karmon A Retrospective

More information

planning Primary stability and initial bone-to-implant contact: The effects on immediate placement and restoration of dental implants

planning Primary stability and initial bone-to-implant contact: The effects on immediate placement and restoration of dental implants planning Primary stability and initial bone-to-implant contact: The effects on immediate placement and restoration of dental implants Alan M. Meltzer, DMD, MScD With the growing popularity of immediate

More information

Evaluation of INNO TM dental Implants placed with high insertion torques (Up to 135 Ncm)

Evaluation of INNO TM dental Implants placed with high insertion torques (Up to 135 Ncm) Dental implant Journal: Vol. 3, May, 2014 Evaluation of INNO TM dental Implants placed with high insertion torques (Up to 135 Ncm) Daehee Lee DDS PhD ABSTRACT Purpose: This study evaluated on the clinical

More information

P rimary or mechanical implant

P rimary or mechanical implant IMPLANT DENTISTRY / VOLUME 0, NUMBER 0 2013 1 Effect of Location on Primary Stability and Healing of Dental Implants Alberto Monje, DDS,* Fernando Suarez, DDS,* Carlos Andrés Garaicoa, DDS, Florencio Monje,

More information

A STUDY ON THE CORRELATION BETWEEN IMPLANT STABILITY VALUES AND INITIAL INSERTION TORQUE

A STUDY ON THE CORRELATION BETWEEN IMPLANT STABILITY VALUES AND INITIAL INSERTION TORQUE J Korean Acad Prosthodont : Volume 44, Number 3, 2006 A STUDY ON THE CORRELATION BETWEEN IMPLANT STABILITY VALUES AND INITIAL INSERTION TORQUE Jong-Hyuk Lee 1,2, D.D.S., M.S.D., Jae-Ho Yang 1, D.D.S.,

More information

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

NARROW DIAMETER implant

NARROW DIAMETER implant ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball

More information

Evaluation of implant stability and crestal bone loss around the implant prior to prosthetic loading: A six month study

Evaluation of implant stability and crestal bone loss around the implant prior to prosthetic loading: A six month study Original Article Evaluation of implant stability and crestal bone loss around the implant prior to prosthetic loading: A six month study P. Singh, H. G. Garge*, V. S. Parmar, M. Viswambaran*, M. M. Goswami*

More information

A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis

A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis ORIGINAL ARTICLE http://dx.doi.org/10.4047/jap.2011.3.4.190 A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis In-Ho Kang,

More information

In 1981, Dr. Albrektsson, a member of

In 1981, Dr. Albrektsson, a member of Osseodensification facilitates ridge expansion with enhanced implant stability in the maxilla: part II case report with 2-year follow-up Drs. Ann Marie Hofbauer and Salah Huwais offer another case study

More information

Case Study. Case # 1 Author: Dr. Suheil Boutros (USA) 2013 Zimmer Dental, Inc. All rights reserved. 6557, Rev. 03/13.

Case Study. Case # 1 Author: Dr. Suheil Boutros (USA) 2013 Zimmer Dental, Inc. All rights reserved. 6557, Rev. 03/13. Placement of a Zimmer Trabecular Metal Dental Implant with Simultaneous Ridge Augmentation and Immediate Non-Functional Loading Following Tooth Extraction and Orthodontic Treatment for Implant Site Development

More information

Clinical Perspectives

Clinical Perspectives Clinical Perspectives Inside This Issue: Revised Drilling Guidelines For Parallel Walled Implants Case Presentation By: Pär-Olov Östman, DDS, PhD, MD Volume 8, Issue 1 Recommended Drilling Guidelines For

More information

Relation between insertion torque and bone implant contact percentage: an artificial bone study

Relation between insertion torque and bone implant contact percentage: an artificial bone study Clin Oral Invest (2012) 16:1679 1684 DOI 10.1007/s00784-011-0658-0 ORIGINAL ARTICLE Relation between insertion torque and bone implant contact percentage: an artificial bone study Cheng Liu & Ming-Tzu

More information

Objectives: The purpose of this investigation was to evaluate the effects of different

Objectives: The purpose of this investigation was to evaluate the effects of different www.scielo.br/jaos Comparison of implant stability after different implant surface treatments in dog bone Sun-Jong Kim 1, Myung-Rae Kim 2, Jae-Suk RIM 3, Sung-Min CHUNG 4, Sang-Wan Shin 5 1- DDS, MSD,

More information

Insertion Torque and Resonance Frequency Analysis of Dental Implant Systems in an Animal Model with Loaded Implants

Insertion Torque and Resonance Frequency Analysis of Dental Implant Systems in an Animal Model with Loaded Implants Insertion Torque and Resonance Frequency Analysis of Dental Implant Systems in an Animal Model with Loaded Implants Bilal Al-Nawas, Priv-Doz, Dr Med, Dr Med Dent 1 /Wilfried Wagner, Univ-Prof, Dr Med,

More information

Assessment of Osstellä and Periotest Ò systems in measuring dental implant stability (in vitro study)

Assessment of Osstellä and Periotest Ò systems in measuring dental implant stability (in vitro study) The Saudi Dental Journal (2011) 23, 17 21 King Saud University The Saudi Dental Journal www.ksu.edu.sa www.sciencedirect.com ORIGINAL ARTICLE Assessment of Osstellä and Periotest Ò systems in measuring

More information

The development of modern Oral

The development of modern Oral RESEARCH In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants João César Zielak, DDS, MSc, PhD* Murilo Rorbacker, DDS Rodrigo Gomes, DDS Celso Yamashita, DDS, MSc, PhD Carla

More information

Australian Dental Journal

Australian Dental Journal Australian Dental Journal The official journal of the Australian Dental Association Australian Dental Journal 2017; 0: 1 7 doi: 10.1111/adj.12536 Can resonance frequency analysis detect narrow marginal

More information

Evaluation of Stability And Crestal Bone Loss of Dental Implants Subjected to two Loading Protocols in the Mandible.

Evaluation of Stability And Crestal Bone Loss of Dental Implants Subjected to two Loading Protocols in the Mandible. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 79-0853, p-issn: 79-0861.Volume 16, Issue 10 Ver. XV (Oct. 017), PP 01-08 www.iosrjournals.org Evaluation of Stability And Crestal Bone Loss

More information

New dental implant designs

New dental implant designs RESEARCH Mineralized Tissue Formation Associated With 2 Different Dental Implant Designs: Histomorphometric Analyses Performed in Dogs Sergio Olate, DDS, MS, PhD 1,2 * Henrique Duque de Miranda Chaves

More information

Implant stability is considered one of the most important

Implant stability is considered one of the most important Development of Implant Stability During Early Healing of Immediately Loaded Implants Antonin Simunek, MD, PhD 1 /Dana Kopecka, MD, PhD 2 /Tomas Brazda, MD 2 / Jakub Strnad, PhD 3 /Lukas Capek, PhD 4 /Radovan

More information

TOPICS. T O P I C S Day 1. Implant Locations. Implant Placement in the Posterior Maxilla. Anatomy and risk factors Option 1: Short implants

TOPICS. T O P I C S Day 1. Implant Locations. Implant Placement in the Posterior Maxilla. Anatomy and risk factors Option 1: Short implants T O P I C S Day 1 Factors influencing the long-term stability of dental implants Surgical procedures in posterior sites: Standard implant placement with or without flap elevation Surgical procedures in

More information

The success rate for osseointegration of

The success rate for osseointegration of RESEARCH CLINICAL EVALUATION OF THE NOBELACTIVE IMPLANT SYSTEM: ACASE SERIES OF 107 CONSECUTIVELY PLACED IMPLANTS AND A REVIEW OF THE IMPLANT FEATURES Tassos Irinakis, DDS, MSc; Colin Wiebe, DDS, MSc The

More information

Ilser Turkyilmaz Lars Sennerby Celal Tumer Muhittin Yenigul Mehmet Avci

Ilser Turkyilmaz Lars Sennerby Celal Tumer Muhittin Yenigul Mehmet Avci Ilser Turkyilmaz Lars Sennerby Celal Tumer Muhittin Yenigul Mehmet Avci Stability and marginal bone level measurements of unsplinted implants used for mandibular overdentures: a 1-year randomized prospective

More information

GMI FRONTIER implant system. Surgical procedures guide

GMI FRONTIER implant system. Surgical procedures guide GMI implant system Surgical procedures guide ABOUT THIS MANUAL This surgical procedures guide or surgical manual for the GMI implant system is designed solely to provide instructions for using GMI products,

More information

High-end solution STEP-BY-STEP USER GUIDE

High-end solution STEP-BY-STEP USER GUIDE High-end solution STEP-BY-STEP USER GUIDE System Contents REAL TIME Restoration Guided Conical drill Multi-Unit abutment, Conical connection Guided kit for conical connection implant procedure C1 & V3

More information

A retrospective study on separate single-tooth implant restorations to replace two or more consecutive. maxillary posterior teeth up to 6 years.

A retrospective study on separate single-tooth implant restorations to replace two or more consecutive. maxillary posterior teeth up to 6 years. Original Article A retrospective study on separate single-tooth implant restorations to replace two or more consecutive maxillary posterior teeth up to 6 years follow up Myat Nyan Department of Prosthodontics,

More information

Jie Han Martina Lulic Niklaus P. Lang

Jie Han Martina Lulic Niklaus P. Lang Jie Han Martina Lulic Niklaus P. Lang Factors influencing resonance frequency analysis assessed by Osstellt mentor during implant tissue integration: II. Implant surface modifications and implant diameter

More information

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and.

Non-osseointegrated. What type of mini-implants? 3/27/2008. Require a tight fit to be effective Stability depends on the quality and. Non-osseointegrated What type of mini-implants? Require a tight fit to be effective Stability depends on the quality and quantity of cortical and trabecular bone. Osseointegrated Non-osseointegrated AbsoAnchor

More information

Comparion of stability in titanium implants with different surface topographies in dogs

Comparion of stability in titanium implants with different surface topographies in dogs ORIGINAL ARTICLE DOI:10.4047/jap.2009.1.1.47 Comparion of stability in titanium implants with different surface topographies in dogs Nam-Sook Kim 1, DDS, Mong-Sook Vang 2, DDS, MSD, PhD, Hong-So Yang 2

More information

Areview of recent studies concerning molar. Single Molar Replacement with a Progressive Thread Design Implant System: A Retrospective Clinical Report

Areview of recent studies concerning molar. Single Molar Replacement with a Progressive Thread Design Implant System: A Retrospective Clinical Report Single Molar Replacement with a Progressive Thread Design Implant System: A Retrospective Clinical Report George E. Romanos, Dr med dent 1 /Georg H. Nentwig, Prof Dr med dent 2 Many clinical studies have

More information

The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog

The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog Carolina Manresa Manel Bosch José J. Echeverrıa The comparison between implant stability quotient and bone-implant contact revisited: an experiment in Beagle dog Authors affiliations: Carolina Manresa,

More information

Annals of Dental Research

Annals of Dental Research P a g e 1 ADR 7 (1), 1-6, 2017 Journal Homepage: www.hgpub.com/ojs2/index.php/adr Direct Sinus Lift with Putty Alloplastic Bone Substitute for Simultaneous Implant Placement A Case Report Chander Prakash*,

More information

Implant Placement in Maxillary Anterior Region Along with Soft and Hard Tissue Grafting- A Case Report.

Implant Placement in Maxillary Anterior Region Along with Soft and Hard Tissue Grafting- A Case Report. IOSR Journal of Dental and Medical Sciences (IOSR-JDMS) e-issn: 2279-0853, p-issn: 2279-0861.Volume 15, Issue 10 Ver. XII (October. 2016), PP 68-73 www.iosrjournals.org Implant Placement in Maxillary Anterior

More information

Identification of Stability Changes for Immediately Placed Dental Implants

Identification of Stability Changes for Immediately Placed Dental Implants Identification of Stability Changes for Immediately Placed Dental Implants Jason D. West, DDS, MS 1 /Thomas W. Oates, DMD, PhD 2 Purpose: To evaluate the changes in stability of immediately placed implants

More information

IQ IMPLANTS Smart Choice PRODUCT C ATA L O G

IQ IMPLANTS Smart Choice PRODUCT C ATA L O G IQ IMPLANTS Smart Choice PRODUCT C ATA L O G www.iqimplants.com info@iqimplants.com 2 TABLE OF CONTENTS GENERAL Company Profile 3 IMPLANTS 5 PROSTHETICS 13 TOOLS 27 3 COMPANY PROFILE IQ Implants Ltd. reflects

More information

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual Smarter Thinking. Simpler Design. Prima Plus Surgical Manual TABLE OF CONTENTS PRIMA PLUS IMPLANT SURGICAL MANUAL SURGERY Prima Plus Characteristics 4 Surgical Considerations 5 Prima Plus Surgical Sequence

More information

NobelActive. The NobelActive technical story

NobelActive. The NobelActive technical story NobelActive The NobelActive technical story The NobelActive technical story NobelActive origins The life cycle of NobelActive began following years of research and development on a self-drilling and bone-condensing

More information

scientific compendium

scientific compendium scientific compendium Zimmer Tapered Screw-Vent Implant Success Stories in Stability and Survival Rates Zimmer Tapered Screw-Vent Implant Scientific Compendium table of contents CLINICAL & PRE-CLINICAL

More information

Replacement of a missing posterior tooth in the mandible (isy by CAMLOG)

Replacement of a missing posterior tooth in the mandible (isy by CAMLOG) Replacement of a missing posterior tooth in the mandible (isy by CAMLOG) Implants are generally regarded as being a high-end therapy involving relatively high costs. In the quest for economical treatment

More information

The Brånemark osseointegration method, using titanium dental implants (fixtures)

The Brånemark osseointegration method, using titanium dental implants (fixtures) Early Failures in 4,641 Consecutively Placed Brånemark Dental Implants: A Study From Stage 1 Surgery to the Connection of Completed Prostheses Bertil Friberg, DDS/Torsten Jemt, DDS, PhD/Ulf Lekholm, DDS,

More information

10 Years of XiVE Implantology Unlimited

10 Years of XiVE Implantology Unlimited 10 Years of XiVE Implantology Unlimited Implantology Unlimited is obligate for XiVE and every limit gives always a push to surrender new challenges: Primary stability in low bone quality? No problem with

More information

Endosseous dental implants initially showed very

Endosseous dental implants initially showed very Five-mm-Diameter Implants without a Smooth Surface Collar: Report on 98 Consecutive Placements Franck Renouard, DDS*/Jean-Pierre Arnoux, DDS**/David P. Sarment, DDS*** In recent years, indications for

More information

The Use Of 6mm Long Implants In Cases With Limited Bone Height: A Preliminary 6-Month Clinical Study

The Use Of 6mm Long Implants In Cases With Limited Bone Height: A Preliminary 6-Month Clinical Study 26 News No.26 January 2011 The Use Of 6mm Long Implants In Cases With Limited Bone Height: A Preliminary 6-Month Clinical Study Make it Simple 2 The Use Of 6mm Long Implants In Cases With Limited Bone

More information

Cost-effective immediate restoration in the anterior region of the mandible (isy by CAMLOG)

Cost-effective immediate restoration in the anterior region of the mandible (isy by CAMLOG) Cost-effective immediate restoration in the anterior region of the mandible (isy by CAMLOG) Many patients wish to have permanent dentures, and often this wish can be fulfilled by one of the greatest developments

More information

Implant System MAKE IT SIMPLE

Implant System MAKE IT SIMPLE Implant System MAKE IT SIMPLE P. 4-5 P. 6-7 P. 8-9 P. 10-13 P. 14 P. 15-17 P. 18-19 P. 20-21 MIS Warranty: MIS exercises great care and effort in maintaining the superior quality of its products. All MIS

More information

immediate implantation and loading with Paltop Osteotomes for bone expansion Case Study

immediate implantation and loading with Paltop Osteotomes for bone expansion Case Study immediate implantation and loading with Paltop Osteotomes for bone expansion Case Study Osteotomes for bone expansion 2 This 55-year-old female patient presents with a failing maxillary cuspid. 3 A fistula

More information

Stability change of chemically modified sandblasted/acid-etched titanium palatal implants. A randomizedcontrolled

Stability change of chemically modified sandblasted/acid-etched titanium palatal implants. A randomizedcontrolled Marc Schätzle Roland Männchen Ulrike Balbach Christoph H. F. Hämmerle Helge Toutenburg Ronald E. Jung Stability change of chemically modified sandblasted/acid-etched titanium palatal implants. A randomizedcontrolled

More information

» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL MICROCONE. IPS Implant systems

» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL MICROCONE. IPS Implant systems » Guided Surgery by MEDENTiKA «IPS Implant systems MEDENTIGUIDE MICROCONE SURGERY MANUAL » MedentiGuide «MedentiGuide drill sleeves support the surgeon in preparing the implant bed for MEDENTiKA implants.

More information

Preliminary Data of a Prospective Clinical Study on the Osseotite NT Implant: 18-month Follow-up

Preliminary Data of a Prospective Clinical Study on the Osseotite NT Implant: 18-month Follow-up Preliminary Data of a Prospective Clinical Study on the Osseotite NT Implant: 18-month Follow-up Mithridade Davarpanah, MD, DDS 1 / Mihaela Caraman, DDS 2 /Serge Szmukler-Moncler, DDS, PhD 3 / Boris Jakubowicz-Kohen,

More information

More than pure esthetics. The natural and strong solution.

More than pure esthetics. The natural and strong solution. More than pure esthetics. The natural and strong solution. Expand your patient pool with an innovative solution. With the Straumann PURE Ceramic Implant, clinicians have an additional option to convince

More information

BASIC Level 1 Practitioner Certification

BASIC Level 1 Practitioner Certification Chapter 5 Place the Prosthesis Because dental implants are now an important and distinct part of prosthodontic dentistry, general dentists must be well versed in all aspects of the treatment option. The

More information

Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis

Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis Journal section: Implantology Publication Types: Research doi:10.4317/jced.52873 http://dx.doi.org/10.4317/jced.52873 Biomechanical evaluation of oversized drilling technique on primary implant stability

More information

The REAL Mini Implant

The REAL Mini Implant by INTRA-LOCK The REAL Mini Implant IMMEDIATE LOAD STABILIZE DENTURES SIMPLIFIED PROCEDURE PROVISIONAL PROSTHETICS DIRECT PLACEMENT ENGINEERED EVOLUTION. Intra-Lock has elevated the engineering and delivery

More information

» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL QUATTROCONE QUATTROCONE30. IPS Implant systems

» Guided Surgery. by MEDENTiKA « MEDENTIGUIDE SURGERY MANUAL QUATTROCONE QUATTROCONE30. IPS Implant systems » Guided Surgery by MEDENTiKA «IPS Implant systems MEDENTIGUIDE QUATTROCONE QUATTROCONE30 SURGERY MANUAL » MedentiGuide «MedentiGuide drill sleeves support the surgeon in preparing the implant bed for

More information

Trefoil Procedure manual

Trefoil Procedure manual Trefoil Procedure manual This procedure manual does not replace attending a complete training program Note: In order to improve readability, Nobel Biocare does not use or in the running text. By doing

More information

Evaluation of peri-implant tissue response according to the presence of keratinized mucosa Abstract Purpose: Materials and methods Results:

Evaluation of peri-implant tissue response according to the presence of keratinized mucosa Abstract Purpose: Materials and methods Results: Evaluation of peri-implant tissue response according to the presence of keratinized mucosa Bum-Soo Kim 1, Young-Kyun Kim 1, Pil-Young Yun 1, Yang-Jin Lee 2, Hyo-Jeong Lee 3, Su-Gwan Kim 4 1Department of

More information

Stability of implants placed in fresh sockets versus healed alveolar sites: Early findings

Stability of implants placed in fresh sockets versus healed alveolar sites: Early findings Sergio Alexandre Gehrke Ulisses Tavares da Silva Neto Paulo Henrique Orlato Rossetti Sidney Eiji Watinaga Gabriela Giro Jamil Awad Shibli Stability of implants placed in fresh sockets versus healed alveolar

More information

TRINON. Index. Index. Overview Page 4. Catalogue

TRINON. Index. Index. Overview Page 4. Catalogue QK-Implant Information Components are updated periodically, check with your local Trinon partner on current models Please follow Instructions for Use for individual components Images are not to scale see

More information

Osteoconductive S.L.A. Surface. Fantastic surface, Bioseal, and Perfect achievement of CMI f ixation.

Osteoconductive S.L.A. Surface. Fantastic surface, Bioseal, and Perfect achievement of CMI f ixation. Fantastic surface, Bioseal, and Perfect achievement of CMI f ixation. Osteoconductive S.L.A. Surface No residual acid and cytotoxicity were observed in the surface examination. IS-II active Product Catalog

More information

Young-Jin Park, DDS,* and Sung-Am Cho, DDS, MS, PhD

Young-Jin Park, DDS,* and Sung-Am Cho, DDS, MS, PhD J Oral Maxillofac Surg 68:1338-1344, 2010 Retrospective Chart Analysis on Survival Rate of Fixtures Installed at the Tuberosity Bone for Cases With Missing Unilateral Upper Molars: A Study of 7 Cases Young-Jin

More information

A self-tapping internal hex. implant by MAKE IT SIMPLE

A self-tapping internal hex. implant by MAKE IT SIMPLE A self-tapping internal hex. implant by MAKE IT SIMPLE P. 4-5 P. 6-7 P. 8 P. 9 P.10 P.11 P.12 P.13 P.14 P.15 P.16-17 P. 18-19 P. 20-21 P. 22 P. 23 P. 24-25 MIS Warranty: MIS exercises great care and effort

More information

Unitek Temporary Anchorage Device (TAD) System

Unitek Temporary Anchorage Device (TAD) System Planning and Placement Guide Treatment Planning Panoramic or P.A. X-ray - visualize root proximity Curved explorer tip - outline roots intraorally Bone sounding - measure tissue thickness using probe with

More information

JASON C. GUERRERO B.S., Marquette University, 2001 D.D.S., Marquette University, 2005

JASON C. GUERRERO B.S., Marquette University, 2001 D.D.S., Marquette University, 2005 Clinical Assessment of Primary Implant Stability Parameters Using a Stereolithographic Surgical Guide BY JASON C. GUERRERO B.S., Marquette University, 2001 D.D.S., Marquette University, 2005 THESIS Submitted

More information

Updated secondary implant stability data of two dental implant systems. A retrospective cohort study

Updated secondary implant stability data of two dental implant systems. A retrospective cohort study Journal section: Periodontology Publication Types: Research doi:10.4317/jced.54146 http://dx.doi.org/10.4317/jced.54146 Updated secondary implant stability data of two dental implant systems. A retrospective

More information

Height* FC width Prep depth RC width 4.0mm+ 4.3mm N/A 6.3mm

Height* FC width Prep depth RC width 4.0mm+ 4.3mm N/A 6.3mm STERN ERA IMPLANT ABUTMENT Summary Resilient precision overdenture attachment. Universal hinge with vertical movement. Titanium abutment, nylon male. Manufactured for most popular screw and cylinder implants.

More information

THE CORRELATION BETWEEN BONE DENSITY AND IMPLANT STABILITY Youssef M 1 BDS, Shaaban A M 2 PhD, Eldibany R 2 PhD.

THE CORRELATION BETWEEN BONE DENSITY AND IMPLANT STABILITY Youssef M 1 BDS, Shaaban A M 2 PhD, Eldibany R 2 PhD. THE CORRELATION BETWEEN BONE DENSITY AND IMPLANT STABILITY Youssef M 1 BDS, Shaaban A M 2 PhD, Eldibany R 2 PhD. Abstract: Introduction: Long-term studies have documented the successful treatment of edentulous

More information

External Hexagon Hybrid Implants. External Hexagon Hybrid Implants. Versatility.

External Hexagon Hybrid Implants. External Hexagon Hybrid Implants. Versatility. External Hexagon Hybrid Implants Versatility. www.pibranemark.com HEX-S Solid Enhanced Macrogeometry Easy, Safe and Simplified installation (!). Special Conical Drills have the same geometry of Hybrid

More information

Long-term success for osteointegrated

Long-term success for osteointegrated CASE REPORT Symphyseal Bone Cylinders Tapping With the Dental Implant Into Insufficiency Bone Situated Esthetic Area at One-Stage Surgery: A Case Report and the Description of the New Technique Umut Tekin,

More information

Bone Level. Surgical Guideline. Dental Implant System

Bone Level. Surgical Guideline. Dental Implant System Bone Level BL Surgical Guideline Dental Implant System Bone Level This guideline was developed in close collaboration with experienced implant dentists and dental technicians. Experts from America, Europe

More information