Evaluation of the Clinitek 200 urinary test-strip reader in the analysis of dog and rat urines in pre-clinical toxicology studies
|
|
- Claire McKenzie
- 5 years ago
- Views:
Transcription
1 240 Laboratory Animals (1993) 27, Evaluation of the Clinitek 200 urinary test-strip reader in the analysis of dog and rat urines in pre-clinical toxicology studies A. PAQUIGNON, G. TRAN & J. P. PROVOST Department of Biochemistry, Clinical Chemistry Laboratory, Pfizer, and Centre de Recherches, BP 159, Amboise, France Summary The performance of a clinical urinary test-strip reader Clinitek 200 was evaluated for dog and rat urines, in the context of pre-clinical toxicology studies. No major discrepancies were found between data generated by visual estimation or automatic measurement. Analysis of spiked samples showed good agreement between actual concentrations and Clinitek 200 responses for ketone bodies and glucose although a lack of sensitivity was found for the latter. Results for proteins showed over- or underestimation in dog and rat urines respectively at low concentrations, and overestimation at high concentrations in both species. Reproducibility of responses was excellent for ketone bodies, glucose and proteins but was weaker for haemoglobin and bilirubin. High bilirubin concentrations were found to interfere with the haemoglobin reaction. The ph measurements were found to be accurate only around ph 7. Specific gravity measurements were unreliable. Overall, the Clinitek 200 as a screening tool proved sufficiently reliable in the measurement of all parameters tested, with the exception of specific gravity. Keywords: Laboratory animals; Urinalysis; Test strip; Automated analysis The clinical chemistry analyses performed in the course of pre-clinical toxicology studies, with Correspondence to: lp Provost, Department of Biochemistry, Clinical Chemistry Laboratory, Pfizer, Centre de Recherches, BP 159, Amboise, France. Received 27 April 1992; accepted 25 September 1992 novel drug entities, generally include the analysis of urine constil uents. These studies generate large numbers of samples; for example in the course of a carcinogenicity study using rats, over 500 samples of urine need to be analysed. To meet these demands, qualitative or semiquantitative examinations of urinary constituents with reagent strips are performed. When changes in colour are estimated visually, each reactive pad on such a test strip must be compared individually with a standard set of colours, and at various times after wetting the pad in order to optimize the conditions for each parameter. These analyses are, therefore, tedious and relatively timeconsuming, and because of the subjective nature of the readings they are of doubtful accuracy (Peele et al., 1977). The introduction of automatic reagent strip readers, which llse the reflectance principle, may overcome some of these drawbacks. Automated analysers increa;e sample throughput (300 reagent strips can be processed per hour), and, within the context of regulatory toxicology studies, the printout of data produced by these machines improves compliance with the requirements of Good Laboratory Practice. Finally, as these analysers are generally equipped with an RS 232 interface, data can be automatically transferred to a main frame computer for data processing and statistical analysis. However, these automatic strip readers and the accompanying reagent strips are intended for human use. Whilst a number of studies have been perform~:d to assess the reliability of these systems in a human clinical context (Peele et al.,,,1977; Grinstead et al., 1987), their suitability for use with dog and rat urine has to our knowledge yet to be investigated.
2 Evaluation of the Clinitek 200 urinary test-strip reader 241 In this study, we examined the use of the Clinitek 200* reader for the analysis of dog and rat urines. We have assessed the following points: recovery and within-run reproducibility of responses using spiked samples, comparison between visual and automatic reading for ketone bodies, glucose, proteins, haemoglobin and bilirubin, and we compared results with those obtained by direct measurement of ph and specific gravity. Materials and methods Animals and sampling conditions Urine samples were obtained from untreated male and female beagle dogs (Marshall Farms, New York, USA) and rats (Crl: COBS-VAF-CD (SO) BR, Charles River, France). No distinction was made between sexes in these experiments. Body weights ranged from 6 to 9 kg for dogs and from 200 to 600 g for rats. During the sampling period, the animals were placed in individual stainless steel metabolism cages; dogs were deprived of food and water, rats were deprived of food but had free access to tap water. Dogs were maintained at a room temperature of 19± 3 C and rats at 20 ± 2 C with a relative humidity of 60 ± The air was changed in the animal rooms 14 times per hour. Urine was collected overnight (for approximately 16h) in plastic containers without addition of a preservative, and analysed within 1h following collection. Measurements using the Clinitek 200 and test-strips Reagent strip analysis, using Multistix 10-50* reagent strips was performed according to the manufacturer's instructions. Visual estimations were performed by a single trained observer. Automatic readings were performed using a Clinitek 200 urine analyser; a reference strip, supplied by the manufacturer, with pads of known reflectance was used to calibrate the instrument. Printed non-numeric messages from Clinitek and Multistix are registered trademarks of Ames Co., Division of Miles Laboratories, Inc., Elkhart, USA. Clinitek 200 and their significance were: Neg (Negative), T (Traces), Sm (Small), M (Moderate), Lg (Large). Additionally, specific gravity was measured using a UR-l hand-refractometer (Atago, Japan), and the ph with a digital ph meter (Orion, USA). For ph measurements (using a ph meter) and for densitometry, urine samples were centrifuged (10 min at approximately 2000 rpm). Preparation of 'spiked' urines Spiked samples of urine were prepared by addition of the test analyte to urine samples which were negative for that parameter by reagent strip analysis. Ranges of final concentrations were as follows: ketone bodies, mg/i; glucose, O 5-20 g/i; proteins, 0'075-3 g/i; haemoglobin, O' mg/i; bilirubin, 4-64 mg/1. A zero concentration grade was assumed when no analyte was added. Materials used to 'spike' these urines were as follows: acetoacetic acid, a-d ( + ) glucose and dog and rat haemoglobin (Sigma Chemical Company, St Louis, USA); standard solutions of bovine serum albumin (Preciset) and human bilirubin (precilib) (Boehringer Mannheim France, Meylan, France). Results Comparison of Clinitek 200 and visual reading of reagent strips The data, summarized in Table 1, was generated using (unspiked) urine samples from control animals; it should be noted that we were not able to include glucose-positive urine. In both species, there was agreement between the automated (Clinitek 200) and visual estimates of analyte; however, the automated reading overlapped the visual reading by up to 3 sections of the reagentstrip scale. Analysis of spiked urine samples from dogs and rats using the Clinitek 200 We made 10replicate measurements with different urine samples to test the reproducibility of the reflectance readings. Data for the analysis of urine samples which had been spiked with
3 242 Paquignon, Tran & Provost Table 1. Comparison of Clinitek 200 and visual readings of reagent strips Ketone bodies (mg/l) Proteins (gil) Haemoglobin Dog Rat Visual reading Clinitek Visual reading C/initek N Neg T N Neg T / Neg I Neg ]8 T 5 4 I T II N Neg T 0,3 N Neg T 0 3 >3 Neg Neg T T ' I I 12 I 4 7 > ]8 N Neg T Sm M N Neg T Sm M Neg Neg T 4 I 1 2 T Sm 1 1 Sm II 11 M 4 I 3 M 2 2 Lg 1 1 Bilirubin N Neg 8m N Neg Sm Neg Neg Sm Sm 10 9 I N: Number of readings made. different concentrations of one of 5 analytes (ketone bodies, glucose, protein, haemoglobin, and bilirubin) are shown in Table 2. Ketone bodies. In both species, there was good agreement between the actual concentrations of this analyte and those determined with the teststrip reader. Glucose. Over the glucose concentration range used, spiked urine from dogs and rats returned similar results. When urine was spiked with O 5 gil of glucose, 9/10 samples were reported as not containing glucose by the Clinitek. Samples containing 1gil glucose gave results as 'trace', indicating that this concentration is probably the lowest level detected by the Clinitek reader. Proteins. In the analysis of dog urine, O' 075 gil protein gave a 'trace' response whereas that spiked with O' 15 gil protein gave a value of '0, 3 gil' protein; with rat urine spiked with 0,075 or 0 15 gil protein the Clinitek readings obtained were for a 'trace' amount. With samples from both species spiked with O' 3 gil protein level, there was good agreement between the added protein value and Clinitek readings. However, at higher concentrations, the Clinitek overestimated the protein concentration; all the samples spiked with O' 5 gil gave values of 'I gil' and those samples spiked with I gil gave values of 'greater than 3 gil'. Haemoglobin. All dog urine samples spiked with O' 3 mg/l of haemoglobin gave 'trace' values; in simil"rly spiked rat urine, 6/10 samples gave 'trace' values and 4/10 samples returned 'negative' values. At all other concentrations, there appeared to be some agreement between the added haemoglobin level and the Clinitek readings. However, for a given concentration, the readings from the Clinitek system were split between 2 succl~ssivescale readings; for example,
4 Evaluation of the Clinitek 200 urinary test-strip reader 243 Table 2. Automated analysis of spiked urine samples from dogs and rats Added value Dog Rat Ketone bodies (mg/i) Neg Tr Neg Tr I I 9 80 Glucose (g/i) Neg Tr >10 Neg Tr 2'5 5 >10 0,5 9 I 9 I 1 2'5 I , Proteins (g/i) Neg Tr 0,3 >3 Neg Tr 0'3 >3 0' ' , ' I I Haemoglobin (mg/i) Neg Tr Sm M Lg Neg Tr Sm M Lg 0, , I 9 Bilirubin (mg/i) Neg Sm M Lg Neg Sm M Lg * * * 9* 64 10* 10* Values for which an interference with the reactive pad for haemoglobin was found. with urine spiked with 60 mgll haemoglobin, bilirubin, 5110 and 1/10 samples gave negative 7/10 and 6/10 Clinitek readings indicated a values; however, in similarly spiked rat urine, the 'medium' haemoglobin value and the remaining Clinitek readings were for 'small' amounts samples were reported as 'large' for dog and rat present. At the other concentrations of bilirubin urines respectively. added, the agreement between the spiked bilirubin concentration and the Clinitek readings was Bilirubin. In dog urine spiked with 4 or 8 mg/l satisfactory for both species. A surprising
5 244 Paquignon, Tran & Provost ph-meier 9 Refractometry A ' 1.010~ 6 7 B Value from Clinilek B L -' ' ' ~' , : :025 1_030 Value from Clinitek Refractometry B ~--"----~- ' Value from Clinitek Fig. 1. Comparison between Clinitek 200 and ph-meter for dog (a) and rat urines (b) ::U~_~, Value fromclinilek Fig. 2. Comparison between Clinitek 200 and refractometry for dog (a) and rat urines (b). finding was that at bilirubin concentrations of 64 mg!l in dog urine and of 8 mg!l or greater in rat urine, the reagent strip was positive for haemoglobin (flagged figures) although no haemoglobin had been added to the urine. Comparison between the Clinitek 200 and direct methods of measuring ph and density Comparisons of ph values for canine and rat urine measured using the Clinitek 200 and a ph meter, are shown in Figs la and lb. With rat urine, measurements of ph with the Clinitek 200 gave values around ph 7 which were in agreement. Below ph 6, 5, the Clinitek system underestimated the ph, whilst for alkaline samples (ph 7, 5 to 9) it overestimated the ph. These errors in ph estimation were marked, e.g. at ph 7, 5, values from the Clinitek system could differ by a ph value of 1 5 from the value obtained with a ph meter. Similar trends to overestimate and underestimate the ph were observed for d.ogs. Data for specific gravity measurements using the Clinitek 200 versus refractometry are shown in Fig. 2. The refractometer measured specific gravity up to a value of in increments of O'005 whereas the reagent strip estimated density up to 1'030 in increments of 0,005. There was poor agreement between the 2 sets of measurements for rat and dog urines. In addition, data from the Clinitek system was heterogeneous, particularly for dog urine. Discussion Comparison of Clinitek 200 and visual readings using unspiked urine samples For urines collected during toxicology studies, there were no major discrepancies when data generated by visual examination of test-strips were compared with those generated by the
6 Evaluation of the Clinitek 200 urinary test-strip reader 245 Clinitek 200. However, it should be stressed that visual discrimination between two successive steps of colouring may be difficult, especially for 'border-line' responses. Also the Clinitek 200 may allow a greater discrimination by automatic measurements (and hence improve accuracy) than data generated by visual estimation of reagent strip reactions. Analysis of spiked urine samples from dogs and rats There was good agreement between added concentrations and Clinitek 200 readings for ketone bodies. For glucose, the correlation was acceptable over the range 1 to 20 gil; below 1 gil, the Clinitek 200 system is unlikely to detect glycosuria in either species. There were differences in protein measurements with dog and rat urine; with O' 15 g proteinll urine, the protein content was overestimated in dog urine but underestimated in rat urine. Conversely, at higher protein levels (~0'5 gil), protein content was over-estimated in urine from both species. Protein estimation using test-strips relates primarily to albumin (Honnadel & Koller, 1987), and by reaction with a ph indicator, tetrabromophenol blue. Since the protein composition of laboratory animal urine varies with age and sex (Alt et al., 1980), a potential source of error using these test-strips is a gross underestimation of (.\(2u-globulin, which is a major urinary protein in adult male rats (Borghoff et al., 1991). Moreover, highly alkaline urine can interfere with the buffering system of the protein reactive pad in test-strips and hence affect colour development. These pitfalls have previously been described for the analysis of rat urine using test-strips (Evans & Parsons, 1986). The data from this study show that the agreement between actual and measured values for bovine albumin in urine from either dogs or rats is poor at the concentrations used. Quantitative methods for analysis of urinary protein should be used whenever accurate monitoring of protein excretion is required. The reproducibility of 10 successive readings using spiked urine samples was excellent in most cases, with the exception of haemoglobin (dog and rat urine, all levels of concentration) and for bilirubin in dogs at the lowest tested (4 mg/l) level. However, as these parameters are estimated on a qualitative scale, it is possible that this arbitrary range corresponds to border-line levels of response for the Clinitek 200 system. At the same time (a small) electronic drift or differences in colour development could also account for these phenomena. It is important to point out that at a high concentration of bilirubin, positive interferences were seen on the haemoglobin reaction pads. This is a surprising finding since haemoglobin is detected by its peroxidase-like activity; an explanation for this interference may be the presence of small quantities of interfering substances in the bilirubin standards used in this study. The reasons for the species differences are unclear. ph and density Our results are in agreement with and extend the findings of Grinstead et 01. (1987) for ph measurement in human urine. That ph values in the range 6 to 7 were optimal with the Clinitek 200 system was expected from data generated with freshly obtained human urine (Schumann, 1984). However, since the ph of urine from dogs and rats often falls outside this range (for example, the 5 and 95% percentile values for rat urine in our laboratory are ph 6' 3 and 8, 3; n == 260), then the use of the Clinitek 200 system to monitor urinary ph accurately in these 2 species is questionable. Since the Clinitek 200 system reads ph values by increments of o 5 ph units within its working range, the accuracy does not compare with that of a ph meter which measures to the nearest 0'1 ph unit or less. However, we consider that when only an approximate estimation of ph is required (as in screening tests for example), then the Clinitek 200 system is sufficiently reliable. The total lack of agreement between the Clinitek 200 system and refractometry for measurement of specific gravity can be explained by the fact that the density test pads on reagent strips react with cationic substances, and are
7 246 Paquignon, Tran & Provost not sensitive, or only weakly so, to other urinary components such as proteins or organic acids. Our findings are in contrast with results obtained with human urine where there was good correlation between the two methods (Audoin et al., 1985). This difference is probably due to differing urinary composition in various species. Moreover, the narrow working range of I to for Clinitek 200 is unsuitable for high density urines obtained with dogs which were deprived of water during the sampling period. In conclusion, the Clinitek 200 is a sufficiently reliable instrument for the qualitative or semiquantitative estimation of ketone bodies, glucose, haemoglobin, bilirubin and ph in dog and rat urines. In contrast, the measurement of specific gravity :.s unreliable. References Alt JM, Hackbarth H, Deerberg F & Stolte H (1980) Proteinuria in rats in relation to age-dependent renal changes. Laboratory Animals 14, Audouin W, Couture G, Girard lc & Forest lc (1985) Evaluation de la densite urinaire par la bandelette reactive de la Cie Ames et comparaison avec la densite don nee par Ie refractometre et avec l'osmolalite urinaire. Clinical Biochemistry 18, Borghoff SJ, Miller AB, Bowen lp & Swenberg la (1991) Characteristics of chemical binding to Ci2u -globulin in vitro; evaluating structure-activity relationships. Toxicology and Applied Pharmacology 107, Evans GO & Parsons CE (1986) Potential errors in the measurement of total protein in male rat urine using test strips. Laboratory Animals 20, Grinstead GF, Scot: RE, Stevens BS, Ward VL & Wilson DM (1987) The Ames Clinitek 200/Multistix 9 urinalysis method compared with manual and microscopic methods. Clinical ChemistlY 33, Honnadel DC & K,)l1er A (1987) Urine protein, total. In: Methods in Clinical Chemistry (eds LA Kaplan, AJ Pesce), pp St Louis: CV Mosby Peele ld, Gadsden IlH & Crews R (1977) Evaluation of Ames "Clini-Tek". Clmical Chemistry 23, Peele JD, Gadsden RH & Crews R (1977) Semi-automated vs. visual reading of urinalysis dipsticks. Clinical Chemistry 23, Schumann GB (1984) Examination of urine. In: Clinical Chemistry, ThEOry, Analysis and Correlation (cds LA Kaplan, Al Pesce), pp St Louis: CV Mosby
Semi-Automatedvs. Visual Readingof UrinalysisDipsticks
CLIN. CHEM. 2/2, 222-22 () Semi-Automatedvs. Readingof UrinalysisDipsticks James D. Peeie, Jr., Richard H. Gadsden, and Rachel Crews Semi-automated reflectance readings of urinalysis dipsticks showed better
More informationVetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP
VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and
More informationClinical Test Report. of DUS10 (Urine Reagent Strips) Effective Date: April DFI Lab. Dong-Ai Hospital Medical Center: Clinical Pathology
Page 1 / 8 APPROVED Clinical Test Report of DUS10 (Urine Reagent Strips) Effective Date: April. 13. 2010 DFI Lab Dong-Ai Hospital Medical Center: Clinical Pathology Page 2 / 8 The evaluation followed Method
More informationTaking a dip into urinalysis
Vet Times The website for the veterinary profession https://www.vettimes.co.uk Taking a dip into urinalysis Author : Christine Jameison Categories : RVNs Date : July 1, 2009 Christine Jameison RVN, probes
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k063276 B. Purpose for Submission: New device C. Measurand: Urinary Glucose, Blood,
More informationCLINITEK Novus Automated Urine Chemistry Analyzer
CLINITEK, Atlas, Novus, Status+, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks are properties of their respective owners. Product
More informationEvaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing
Evaluation of VACUETTE Urine CCM Tubes for Urine Strip Testing Background The VACUETTE Urine is for the collection, transport and storage of urine samples for urine culture and urinalysis in the laboratory.
More informationPars Azmon (Automation) Pars Azmon (Manual) Biosystem (Automation) Biosystem (Manual) Ziest Chemi (Manual) 62
STANDARDIZATION OF URNIE PROTEIN MEASURMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty EQAP 7 : URINE PROTEIN Method No. Mean (mg/dl) CV% Pars Azmon (Automation) 45 25 33 Pars Azmon
More informationURINE PROTEIN In Normal State, Maximally There Is 150 mg/24h or 10 mg/dl Protein in Urine More Than 200 Protein Have Been Found in Urine Including 1/3
URNIE PROTEIN R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty URINE PROTEIN In Normal State, Maximally There Is 150 mg/24h or 10 mg/dl Protein in Urine More Than 200 Protein Have Been
More informationControls (positive and negative) for quality assessment-located in central laboratory Note: See central lab UA procedure for QC. Read strips visually.
ACTION DATE INITIAL Adopted Reviewed 2/8/16 KLW Reviewed 12/12/16 KLW Reviewed Reviewed Revised 5/11/10 MAS Supersedes 2/6/08 name change from Bayer to Siemens TITLE: URINALYSIS TESTING USING Siemens MULTISTIX
More informationGIMA ŠLAPIMO ANALIZATORIAI
GIMA ŠLAPIMO ANALIZATORIAI Eil. Nr. 1. Aprašymas 24051 URILYZER 100 URINE ANALYZER - with printer Urilyzer 100 is a small urine test strip reader that offers a broad range of optimized features. The intuitive
More informationYear 1 MBChB Clinical Skills Session Urinalysis
Year 1 MBChB Clinical Skills Session Urinalysis Reviewed & ratified by: Dr V Taylor-Jones & Ms C Tierney. Urinalysis Aims and Objectives Aim: For the student to be able to safely conduct a urinalysis on
More informationURINE DIPSTICK AND SULPHOSALICYLIC ACID TEST. Špela Borštnar UREX 2015, Ljubljana, Slovenia
URINE DIPSTICK AND SULPHOSALICYLIC ACID TEST Špela Borštnar UREX 2015, Ljubljana, Slovenia KIDNEY DISEASE? severity of kidney disease = estimating GFR cause of kidney disease = urinalysis URINE EXAMINATION
More informationhc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine
Design Verification hc Step Serum/Urine, hc Step Detector and HumaPreg Serum/Urine 1 Function... 2 2 Sensitivity and Dynamic Range... 2 2.1 Analytical Sensitivity... 2 3 Validation of Efficacy... 3 3.1
More informationUrinalysis (Macroscopic Physical Tests) Urinalysis. Basic Urinalysis 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD
Urinalysis (Macroscopic Physical Tests) Mohammad Reza Bakhtiari DCLS, PhD Urinalysis Basic Urinalysis CLSI: U/A is the testing of urine with procedures commonly performed in an expeditious, reliable, accurate,
More informationDetection and Estimation of Some Abnormal Constituents. Amal Alamri
Detection and Estimation of Some Abnormal Constituents Amal Alamri Lecture Over view Abnormal constituent of urine Urine analysis Experiments Physical Chemical Micro/Macro 1-Detection of some abnormal
More informationcobas u 601 urine analyzer Providing precise and secure test results
Providing precise and secure test results 1 of 6 Providing precise and secure test results cobas u 601 urine analyzer is a modular, analytical system platform for urine strip testing and features a fully
More informationClinical Laboratory Science: Urinalysis
Clinical Laboratory Science: Urinalysis Urine is produced by the kidney to maintain constant plasma osmotic concentration; to regulate ph, electrolyte and fluid balances and to excrete some 50 grams of
More informationAcetoacetate Assay Kit (Colorimetric)
ab180875 Acetoacetate Assay Kit (Colorimetric) Instructions for Use For the rapid, sensitive and accurate measurement of endogenous Acetoacetate in various samples. This product is for research use only
More informationDesign Verification Function... 2 Analytical sensitivity... 2 Specificity, cross-reactivity and interferences... 3
Design Verification 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity and interferences...
More informationCreatinine (serum, plasma)
Creatinine (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Creatinine 1.2 Alternative names None 1.3 Description of analyte Creatinine is a heterocyclic nitrogenous compound (IUPAC
More informationIt s not just water! What is Urinalysis?
It s not just water! An introduction to Urinalysis What is Urinalysis? Urinalysis or the analysis of urine is one of the oldest laboratory procedures in the practice of medicine. It is a good test for
More informationHexagon OBTI. Design Verification. Contents
Design Verification Hexagon OBTI Contents 1 Function... 2 2 Analytical sensitivity... 2 2.1 Description of control materials... 2 2.2 Results with human haemoglobin... 2 3 Specificity, cross-reactivity
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationAssisting in the Analysis of Urine. Copyright 2011, 2007, 2003, 1999 by Saunders, an imprint of Elsevier Inc. All rights reserved.
Assisting in the Analysis of Urine Urinalysis Why is Urine is analyzed? To detect extrinsic conditions those in which the kidney is functioning normally, but abnormal end-products of metabolism are excreted
More informationURYXXON 300 Instrumental evaluation of urine test sticks
URYXXON 300 Instrumental evaluation of urine test sticks MACHEREY-NAGEL URYXXON 300 Instrumental evaluation Boost your lab performance immediate startup one measurement every 8 seconds (400 sticks/h) bar
More informationBCH472 [Practical] 1
BCH472 [Practical] 1 Physical Examination Chemical Examination 2 ph Color Specific Gravity Volume Odor Appearance Acidic: -Diabetic Ketoacidosis. -Starvation. -UTIs (E. coli). Alkaline: -UTIs (ureasplitting
More informationLABORATORY 5: The Complete Urinalysis
LABORATORY 5: The Complete Urinalysis Notes 1. This lab combines the objectives and activities of the macroscopic and microscopic lab activities. Students are expected to review those labs for reference.
More informationEvaluation of frequency of abnormal Urine R.E tests in Pathology Laboratory
Original Article Evaluation of frequency of abnormal Urine R.E tests in Pathology Laboratory Shahina Yasmin Department of Pathology, Islamic International Medical College, Rawalpindi ABSTRACT Objective
More informationUrine Reagent Strip (Blood - Ketone - Glucose - Protein - ph - Bilirubin)
Urine Reagent Strip (Blood - Ketone - Glucose - Protein - ph - Bilirubin) Cat. No.:DTS411 Pkg.Size: Intended use CD Diagnostics Urine Reagent Strips (URS) for Urinalysis are firm plastic strips to which
More informationM.D.IPA, M.D.IPA Preferred, Optimum Choice and Optimum Choice Preferred STAT Laboratory List Revised Jan. 5, 2017
M.D.IPA, M.D.IPA Preferred, Optimum Choice and Optimum Choice Preferred STAT Laboratory List Revised Jan. 5, 2017 If laboratory results are required on a STAT basis, the designated commercial medical laboratory
More information2 Screen Islet Cell Autoantibody Human ELISA
2 Screen Islet Cell Autoantibody Human ELISA Cat. No.:DEIA4453 Pkg.Size:96T Intended use The 2 Screen Islet Cell autoantibody (2 Screen) ELISA kit is intended for use by professional persons only, for
More informationImproved methods of assessing proteinuria in hypertensive pregnancy
British Journal of Obstetrics and Gynaecology October 1997, Vol. 14, pp. 1159-1164 Improved methods of assessing proteinuria in hypertensive pregnancy *?Patrick J. Saudan MD, *?Mark A. Brown Associate
More informationDocumentation, Codebook, and Frequencies
Documentation, Codebook, and Frequencies Urinary Albumin and Urinary Creatinine Laboratory Survey Years: 2005 to 2006 SAS Transport File: ALB_CR_D.XPT November 2007 NHANES 2005 2006 Data Documentation
More informationDiabetic Nephropathy
Diabetic Nephropathy Outline Introduction of diabetic nephropathy Manifestations of diabetic nephropathy Staging of diabetic nephropathy Microalbuminuria Diagnosis of diabetic nephropathy Treatment of
More informationTSH Receptor Autoantibody ELISA
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it TSH Receptor Autoantibody ELISA Product Data Sheet Cat.
More informationInvestigating the recheck rules for urine analysis in children
Investigating the recheck rules for urine analysis in children Y.M. He*, S.W. Yao*, Y.J. Huang, B.S. Liang and H.Y. Liu Clinical Laboratory, Guangzhou Women and Children s Medical Center, Guangzhou Medical
More informationUrinalysis (Macroscopic( Chemical Tests) ) Background. Ishihara Color Blindness Tests 12/22/2012. Mohammad Reza Bakhtiari DCLS, PhD
Urinalysis (Macroscopic( Chemical Tests) ) Mohammad Reza Bakhtiari DCLS, PhD Background Routine chemical examination of urine has changed dramatically since the early days of urine testing, owing to the
More informationValidation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens
Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens This document describes the validation of a Neogen Fentanyl kit for the semi-quantitative analysis of
More informationCHEMSTRIP 7 - URINE TEST STRIP
University of California, San Francisco -- Department of Laboratory Medicine San Francisco General Hospital 1001 Potrero Avenue, San Francisco CA 94110 Clinical Laboratory- Eberhard Fiebig, MD, Director
More informationComparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity and Effect of Interferent
ORIGINAL ARTICLE Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method Comparison of Modified Jaffe s Kinetic Method and Enzymatic Method of Serum Creatinine Estimation for Precision, Linearity
More informationHuman Creatinine Urinary Detection Kit
Human Creatinine Urinary CATALOG NO: IRAAKT2509 Detection Kit LOT NO: SAMPLE INTENDED USE The Urinary Creatinine kit is designed to quantitatively measure creatinine present in urine samples. BACKGROUND
More informationCJ Shuster A&P2 Lab Addenum Urinanalysis 1. Urinanalysis
CJ Shuster A&P2 Lab Addenum Urinanalysis 1 Urinanalysis PLEASE NOTE: The actual urinanalysis goes rather quickly. You may want to skip directly to the Lab Exercise, collect data, and read the Introduction
More informationPHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE
PHYSICAL PROPERTIES AND DETECTION OF NORMAL CONSTITUENTS OF URINE - OBJECTIVES: 1- The simple examination of urine. 2- To detect some of the normal organic constituents of urine. 3- To detect some of the
More informationA. SAP is the D-Lab's name for a specific set of serum biochemical tests.
Understanding CBC, SAP, UA/Laura J. Steadman, DVM I. CBC - Complete Blood Count A. Three major types of cells are counted 1. Red Blood Cells 2. White Blood Cells 3. Platelets B. Cells are counted at the
More informationInspector's Accreditation Unit Activity Menu
01/12/20XX 15:58:57 Laboratory Accreditation Program Page 1 of 9 CHEMISTRY 1501 ALT, serum/plasma 1502 Albumin, serum/plasma 1504 Alkaline phosphatase, serum/plasma 1506 Amylase, serum/plasma 1508 Bilirubin,
More informationGentian Canine CRP Immunoassay Application Note for scil VitroVet*
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of
More informationSensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood
Sensatest 1 PARAMETER: BLOOD, GLUCOSE AND PROTEIN URINE REAGENT STRIPS (1 PARAMETER) Test for Urine Occult Blood INTENDED USE: Urine Reagent Strips-1B provide a semi-quantitative determination for the
More informationBasic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009
Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,
More informationPlasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana
Plasma Hemoglobin Determination: Two Manual Methods Compared An Honors Thesis (HONRS 499) by Anne M. Emerick Sharlene Strahl St. Francis Hospital and Health Centers Beech Grove, Indiana Dr. Nancy Behforouz
More informationDetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine
World s Only One Component Assay Simple & Easy to Use DetectX Urinary Creatinine Detection Kit Catalog Number K002-H1 Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine Please read this insert
More informationChymotrypsin ELISA Kit
Chymotrypsin ELISA Kit Cat. No.:DEIA10041 Pkg.Size:96T Intended use The Chymotrypsin ELISA Kit is a sandwich Enzyme Immuno Assay intended for the quantitative determination of Chymotrypsin in stool. General
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationautosamplers ASI-100 and ASI-100T Autosamplers
autosamplers ASI-100 and ASI-100T Autosamplers The ASI-100 and ASI-100T Automated Sample Injectors are highly flexible autosamplers for automating high performance liquid chromatography (HPLC) methods.
More informationAvailable online at
Available online at www.annclinlabsci.org Annals of Clinical & Laboratory Science, vol. 45, no. 4, 2015 391 Influence of Vitamin C on Urine Dipstick Test Results Dae-Hyun Ko 1, Tae-Dong Jeong 1, Sollip
More informationGlucose Detection Kit
ab102517 Glucose Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Glucose levels in various samples This product is for research use only and is not intended for
More informationDRUG TESTING POLICY. Policy Number: ADMINISTRATIVE T0 Effective Date: October 1, Related Policies None
DRUG TESTING POLICY UnitedHealthcare Oxford Reimbursement Policy Policy Number: ADMINISTRATIVE 259.1 T0 Effective Date: October 1, 2017 Table of Contents Page INSTRUCTIONS FOR USE... 1 APPLICABLE LINES
More informationDiabetes and Hypertension
Diabetes and Hypertension M.Nakhjvani,M.D Tehran University of Medical Sciences 20-8-96 Hypertension Common DM comorbidity Prevalence depends on diabetes type, age, BMI, ethnicity Major risk factor for
More informationSAFETY ASPECTS OF MIDAZOLAM
Br. J. clin. Pharmac. (1983), 16, 37S-41S Biological Pharmaceutical Research Department, F. Hoffmann-La Roche & Co Ltd, CH-4002 Basle, Switzerland 1 The LD50 in the rat and the mouse is about 1600 mg/kg
More informationBovine Insulin ELISA
Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:
More informationINSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY
INTENDED USE INSTANT-VIEW H. pylori Rapid Test CLIA WAIVED Test for Whole Blood QUALITATIVE IN-VITRO DIAGNOSTIC TEST FOR EXTERNAL USE ONLY INSTANT-VIEW H. pylori Rapid Test is a rapid qualitative immunoassay
More informationGLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE
More informationCase Log Number(s) Veterinarian or VTS Accurately report test results, using appropriate units of measurement Quality Control/Assurance Date Mastered
AVCPT Skills List Candidate: Understanding of test methodology, techniques and ability to perform testing must be applied to each skill. The overall goal is to provide accurate and valid results to assist
More informationThe NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16
The NEPHROCHECK Test System Training Program The NEPHROCHECK Training Program US: Astute Medical, Inc. 2014 PN 700001 Rev D 2014/09/16 Learning Objectives At the conclusion of this training, participants
More informationELEGANCE Chlamydia pneumoniae IgG ELISA KIT
INTENDED USE The ELEGANCE Chlamydia pneumoniae IgG ELISA has been designed for the in vitro diagnostic measurement of anti- Chlamydia pneumoniae IgG in the screening of human serum. PRINCIPLES OF THE ELEGANCE
More informationDRUG TESTING POLICY. Policy Number: ADMINISTRATIVE T0 Effective Date: January 1, Related Policies None
DRUG TESTING POLICY UnitedHealthcare Oxford Reimbursement Policy Policy Number: ADMINISTRATIVE 259.2 T0 Effective Date: January 1, 2018 Table of Contents Page INSTRUCTIONS FOR USE... 1 APPLICABLE LINES
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationSAFETY DATA SHEET (SDS) SDS 066 Revision number: 1 Rev. 04/15. Uriflet S 9HA Part Numbers: For use with the AUTION HYBRID AU-4050
SAFETY DATA SHEET (SDS) Revision date: SDS 066 Revision number: 1 Rev. 04/15 Uriflet S 9HA Part Numbers: 73521 For use with the AUTION HYBRID AU-4050 ARKRAY USA, Inc. 5182 West 76 th Street Minneapolis,
More informationSummary of Recommendation Statements Kidney International Supplements (2013) 3, 5 14; doi: /kisup
http://www.kidney-international.org & 2013 DIGO Summary of Recommendation Statements idney International Supplements (2013) 3, 5 14; doi:10.1038/kisup.2012.77 Chapter 1: Definition and classification of
More informationLABORATORY 6: The Complete Urinalysis with Automation
LABORATORY 6: The Complete Urinalysis with Automation PreLab Preparation: 1. Review YouTube videos on Clinitek Status Principle & Operation http://www.youtube.com/watch?v=j-8loszd-u8&feature=related http://www.youtube.com/watch?v=s1tpqqh9m88&feature=endscreen&nr=1
More informationHexagon PSA. Design Verification. Contents
Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More information02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2
For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection
More informationBÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:
More informationUrinalysis and Body Fluids CRg. Unit Outline. Chemical Exam of Urine. Unit 3. Reagent strip manufactures
Urinalysis and Body Fluids CRg Unit 3 Chemical Examination of Urine Part 1 Overview of Procedure & ph Unit Outline Introduction to reagent strip testing ph Glucose & Carbohydrates Ketones Protein Bilirubin
More informationDRUG TESTING POLICY. Policy Number: ADMINISTRATIVE T0 Effective Date: March 1, Related Policies None
DRUG TESTING POLICY UnitedHealthcare Oxford Reimbursement Policy Policy Number: ADMINISTRATIVE 259.3 T0 Effective Date: March 1, 2018 Table of Contents Page INSTRUCTIONS FOR USE... 1 APPLICABLE LINES OF
More informationRoutine urine examination
Routine urine examination 尿常规检查 Huawei Liang, PhD E-mail: hwliang99@163.com Objectives To examine urine for the presence of normal and abnormal constituents by routine urine analysis Principles Urine formation
More informationbeta Hydroxybutyrate (beta HB) Assay Kit
ab83390 beta Hydroxybutyrate (beta HB) Assay Kit Instructions for Use For the rapid, sensitive and accurate measurement of beta Hydroxybutyrate in various samples. This product is for research use only
More informationA. History Urinalysis is the oldest lab test still being performed today
III. THE ROUTINE URINALYSIS A. History Urinalysis is the oldest lab test still being performed today 1. Cave man noted change in urine properties associated with disease 2. Babylonians and Egyptians noted
More informationThe Urinary System. Lab Exercise 38. Objectives. Introduction
Lab Exercise The Urinary System Objectives - Be able to identify the structures of the urinary system and give their function - Be able to recognize the gross anatomy of the kidney - Identify the components
More informationInsulin (Porcine/Canine) ELISA
Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01
More informationContents. Section 1: Basic Laboratory Principles. 1. Laboratory Equipments, Glassware, Laboratory Hazards and General Laboratory Rules...
Contents Section 1: Basic Laboratory Principles 1. Laboratory Equipments, Glassware, Laboratory Hazards and General Laboratory Rules...3 Laboratory Equipments 3 Laboratory Glassware 4 Laboratory Hazards
More informationMyeloperoxidase (MPO) ELISA Stool, Urine KIT
Myeloperoxidase (MPO) ELISA Stool, Urine KIT Cat. No.:DEIA6211 Pkg.Size:96T Intended use The Myeloperoxidase Assay is intended for the quantitative determination of Myeloperoxidase in urine and stool,
More informationThe analytical phase
The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+
More informationPorcine/Canine Insulin ELISA
Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research
More informationAuto HbA1C ANALYZER HbA1LC HPLC 3660
Turnkey Laboratories Solutions Auto HbA1C ANALYZER HbA1LC HPLC 3660 Viewable Accuracy Real Time Summarized Chromatogram Authentic Results EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES
More informationIn Vitro Diagnostic Glucose Test System
CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationURINE REAGENT STRIPS FOR URINALYSIS
CORTEZ DIAGNOSTICS, INC. 20521 Califa Street, Suite 102 and 116 Woodland Hills, CA, 91367 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external
More informationHow to interpret your urine sample results
How to interpret your urine sample results Chronic UTI Info Factsheet Series Once you have submitted your urine sample for analysis, it will be sent off to the local laboratory or hospital laboratory if
More informationReferences Required document for Laboratory Accreditation by the College of American Pathologists.
Subject Urinalysis Clinitek Advantus Regional Index Number Lab-1589 Section Laboratory Subsection Regional Clinic / Affiliate Hospital Laboratories Category Departmental Contact Elizabeth A Kamprud Last
More informationJournal of Chemical and Pharmaceutical Research
Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(2):770-775 Validation of Rapid Liquid Chromatographic
More informationepithelial cell excretion are expressed always as twelve hour rates." Withholding fluid during the afternoon and night except for 200 cc.
THE ADDIS SEDIMENT COUNT IN By JOHN D. LYTTLE NORMAL CHILDREN (From the Babies Hospit and the Department of Pediatrics, Colege of Physicians and Surgeons, Columbia University, New York City) (Received
More informationIntroduction. Study Objective. Methodology. Results. Data Analysis. Interpretation
WHITE PAPER COMPARISON STUDY BETWEEN THE HETTICH EBA 200S (HP) BLOOD TUBE PACKAGE 8 (FORMALLY HELMER QUICKSPIN PLUS) AND A TRADITIONAL 3 LITER CENTRIFUGE IN BLOOD CHEMISTRY DETERMINATIONS Introduction
More informationTABLE OF CONTENTS GENERAL INFORMATION... 1
BIOO RESEARCH PRODUCTS Glucose Assay Kit Manual Catalog #: 5611-01 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure Overview... 1 Required Materials
More informationBias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University
Bias in clinical chemistry Elvar Theodorsson EFLM and Linköping University Bias a controversial subject Different perspectives Reseachers Users Regulatory bodies Standardisation organisations Metrologists
More informationAn Improved Test for Phenylketonuria
An Improved Test for Phenylketonuria Chauncey 0. Rupe and Alfred H. Free HENYLKETONURIA (phenylpyruvic oligophrenia) is a metabolic disorder first recognized by Foiling (1) approximately 25 years ago.
More informationDiagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.
2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846
More informationEXERCISE 11: MACROSCOPIC AND CHEMICAL EXAMINATION OF URINE
EXERCISE 11: MACROSCOPIC AND CHEMICAL EXAMINATION OF URINE Skills: 30 points Objectives: 1. State the specimen of choice for chemical analysis of urine. 2. List two macroscopic observations of urine and
More information