1 British Journal of Anaesthesia 84 (1): (2000) Behaviour of near-infrared light in the adult human head: implications for clinical near-infrared spectroscopy A. E. R. Young 1 *, T. J. Germon 1, N. J. Barnett 2, A. R. Manara 1 and R. J. Nelson 1 1 Departments of Anaesthesia and Neurosurgery, Frenchay Hospital, Frenchay, Bristol BS16 1LE, UK. 2 Product Development Group, Johnson and Johnson Medical Ltd, Newport, UK *Corresponding author: Department of Anaesthesia, Frenchay Hospital, Frenchay, Bristol BS16 1LE, UK To test theoretical assumptions supporting the use of near-infrared spectroscopy (NIRS) in clinical practice, we examined the behaviour of NIR light transmission and attenuation in the human head. Sterile probes for emitting and detecting NIR light at a fixed separation of 40 mm were placed in turn on intact skin, skull, dura and cerebral cortex of 10 patients undergoing elective neurosurgery. In the first five patients, the detecting probe was moved through successive extracerebral layers with the emitter on the skin surface. In the second five patients, the process was reversed, with the emitting probe moved and the detector in the same place on the scalp. NIR intensity was measured at each tissue interface and compared with the intensity measured at the skin surface with all layers intact. Removal of bone and dura from the light path caused a significant reduction in detected intensity. The largest mean reduction in light intensity was a 14-fold decrease with removal of bone (unadjusted P ; paired t test). The assumptions that extracerebral tissues contribute little to attenuation of NIR light in the adult head and that most of this attenuation occurs superficially in the scalp are drawn into question by this study. We postulate that the skull and/or its interface with other layers may act as an optical channel, distorting the behaviour of NIR light in the human head. Br J Anaesth 2000; 84: Keywords: measurement techniques, near-infrared spectroscopy; brain, near-infrared spectroscopy; surgery, neurological Accepted for publication: July 30, 1999 Reliable non-invasive monitoring of cerebral oxygenation in the concentration of the chromophores absorbing that and metabolism is an important goal in anaesthetic and light. In an organ composed of heterogeneous tissues, intensive care practice. On theoretical grounds, transcutane- absorption of NIR light depends on the proportion of ous reflectance-mode near-infrared spectroscopy (NIRS) the total pathlength which passes through each tissue offers considerable promise in fulfilling these aims. compartment and the change in the concentration of chromophores The principles which underlie the detection of changes in those compartments. in the concentration of oxy- and deoxyhaemoglobin by It has been tempting to assume that there are two principal NIRS are beguilingly simple. Human tissues are partly compartments in the human head: extracerebral tissues and translucent to NIR light. As NIR light passes through tissue, cerebral tissue. Much of the early work in cerebral NIR a proportion of NIR photons are absorbed by chromophores spectroscopy was conducted on the basis that the NIR such as oxy- and deoxyhaemoglobin. The amount of absorp- pathlength in extracerebral tissues was constant and contributed tion depends on two predetermined factors (the wavelength relatively little to NIR light absorption, particularly if of NIR light and the characteristic absorption spectra the emitting source and the detector were widely separated, of oxy- and deoxyhaemoglobin) and two variables (the thus increasing the proportion of pathlength thought to pass concentration of the chromophores in the tissue and the through cerebral tissue. These assumptions may be valid in distance that the photons have to travel from their emitting neonates, but there is increasing evidence that they do not source to the detector the NIR pathlength). The greater apply in adults. 1 4 the pathlength the greater the chance of absorption occurring Schematic representations of NIR spectrometers using and vice versa. single or dual channel sensors often depict an elliptical Thus if the pathlength of NIR light passing through light path between the emitter and detector. 5 However, the a homogeneous tissue remains constant, changes in the presence of discrete layers significantly affects transmission, absorption of NIR light are directly proportional to changes scattering and attenuation of NIR light. 6 9 Little is known The Board of Management and Trustees of the British Journal of Anaesthesia 2000

2 In vivo behaviour of near-infrared light Fig 1 Apparatus for measuring baseline recording, with both the emitter and detector on the skin surface. Fig 2 Apparatus with the emitter on the skin and the detector on the dura. detector was maintained by the wire coupling at each layer of the transmission of NIR light at tissue interfaces in vivo (Fig. 2). A further 30-s recording of light intensity was such as those between scalp and skull, skull and dura or made at the bone, dura and cerebral cortex. Haemostasis dura and CSF. If, as a result of these interactions, most was meticulous and blood loss negligible during these of the NIR light path in the adult head passes through phases. In the first five studies (group 1), the detecting extracerebral tissues, the capacity of NIRS to detect changes probe was moved progressively through the anatomical in cerebral oxygenation will be reduced, particularly when layers while the emitting probe remained on the scalp. In perfusion and oxygenation of extracerebral tissues are the second five patients, the position of the probes was unstable. reversed (group 2). We have studied, under carefully controlled clinical For each patient, considering light of one wavelength conditions, whether the generally accepted models of NIR within the NIR range (776.5 nm), a mean intensity value light behaviour apply in the human head, and determined for the 30-s recording at each layer was obtained. The the relative effects of the extracerebral tissue layers on the relative light intensity, using the scalp to scalp recording passage of NIR photons. as the baseline measurement, was also calculated for each patient at each extracerebral layer. The ratios were logtransformed Patients and methods and geometric means and 95% confidence intervals (CI) were calculated for groups 1 and 2 and also We studied 10 adult patients (six females), median age for groups 1 and 2 combined. One-sample t tests on the 47 (range 34 65) yr, undergoing elective neurosurgical log-transformed data were used to test the hypothesis that procedures involving a craniotomy or burr hole under there was no change in light intensity at each layer. Paired general anaesthesia. Anaesthesia was induced with propofol, t tests were used to compare the reduction of log-transformed fentanyl and vecuronium and maintained with intermittent light intensity between successive layers. positive pressure ventilation and 1% isoflurane in nitrous oxide and oxygen. Physiological variables, including endtidal carbon dioxide concentration, mean arterial pressure Results and arterial oxygen saturation, were maintained constant Mean recordings of detected light intensity for one wavelength and within normal limits for all patients for the duration of (776.5 nm) at each layer for each patient are shown the study. The study was approved by the Hospital Ethics in Table 1. The geometric mean of light intensity relative Committee and informed consent was obtained from all to a baseline recording (both probes on the skin) are shown patients. for groups 1 and 2 and for all 10 subjects in Table 2. These A sterilizable probe emitting NIR light of four wavelengths data are also shown as averaged values for the three other (776.5, 819.0, and nm) and a sterilizable wavelengths (819, 871 and 909 nm) for completeness (Table detecting probe (Johnson and Johnson Medical, Cerebral 3). The results for wavelength nm are expressed Redox Monitor 2001) were placed on the shaved, cleaned graphically in Figure 3. scalp. The two probes were kept parallel with a stainless Mean light intensity relative to that recorded with both steel wire coupling of 40 mm in length (Fig. 1). NIR probes on the skin (baseline) for all patients was intensity was recorded at the detector using a computerized (95% CI 0.223, 1.060) at the bone surface, (0.016, data collection system. An initial baseline measurement 0.071) at the dural surface and (0.005, 0.052) at the with both probes on the scalp was recorded for 30 s. Then, cerebral surface. Although there was an average halving of with one probe remaining in the same position on the scalp, light intensity at the bone surface (with loss of the scalp) the other was placed first onto exposed skull after reflection this was not statistically significant. Significant reductions of the scalp and periosteum, then onto intact dura after were seen with loss of bone and loss of the dura. The craniotomy or burr hole and finally onto the cerebral cortex largest geometric mean reduction was a 14-fold decrease after opening the dura. The separation of emitter and between the bone and dural surfaces with loss of the 39

3 Young et al. Table 1 NIR intensity for each patient at each layer, averaged over 30 s. Group 1 emitter maintained at the skin and the detector moved through the extracerebral layers, group 2 detector maintained at the skin with the emitter moved through the extracerebral layers Skin to skin Skin to bone Skin to dura Skin to cerebral cortex Group 1 (mean value) Patient No Patient No Patient No Patient No Patient No Skin to skin Bone to skin Dura to skin Cerebral cortex to skin Group 2 (mean value) Patient A Patient B Patient C Patient D Patient E Table 2 Geometric mean (95% confidence intervals) values for NIR intensity relative to that obtained with both probes on the skin for groups 1 and 2 and for both groups combined (total), using NIR light of wavelength 776.5nm. 1 One-sample t test, H 0 : geometric mean 1; 2 paired t test Probe position Group 1 Group 2 Total 1 Reduction from previous layer 2 n Skin and bone (0.2517, ) (0.0659, ) (0.2225, ) P Skin and dura (0.0106, ) (0.0097, ) (0.0161, ) (0.0337, 0.144) P P Skin and cerebral cortex (0.0010, ) (0.0060, ) (0.0052, ) 0.48 (0.2345, ) P P Table 3 Geometric mean (95% confidence intervals) values for NIR intensity relative to that obtained with both probes on the skin using averaged data from NIR light of wavelengths 819.0, and nm. 1 One-sample t test, H 0 : geometric mean 1; 2 paired t test skull (unadjusted P ; paired t test). There was no significant change in measured physiological variables during the study. Probe position Total 1 Reduction from previous layer 2 Discussion Skin and bone (0.2370, ) In 1977, it was recognized that the differential absorption P of NIR light by haemoglobin, oxyhaemoglobin and Skin and dura (0.0130, ) (0.0192, ) cytochrome C oxidase could be used to estimate blood flow P P Skin and cerebral cortex (0.0101, ) (0.2846, ) and tissue oxygenation in vivo. 10 Based on these principles, P P cerebral NIR spectroscopy has been used for some years to monitor haemoglobin, oxyhaemoglobin and cytochrome C oxidase in neonates whose heads transmit sufficient NIR light. 11 In older children and adults, transmission NIRS is prevented by the thickness of the scalp and skull, and myelination of the brain. Reflectance-mode NIRS was developed to overcome these problems. Using this technique, NIR light is collected from the surface of the scalp ipsilateral to the light source so that the photon pathlength is significantly reduced. NIR light is thought to follow an arc whose depth is proportional to the distance between the emitter and detector and will, in theory, travel through scalp, skull, dura and finally brain before returning through the same tissue layers to be detected at the skin surface. 5 The path of NIR light through the human head and thus the volume of cerebral tissue interrogated by reflectance- Fig 3 Mean fractional change in detected light intensity relative to baseline (with both probes on the skin) after loss of successive layers of the human head. mode NIRS is unknown. The pathlength of the emitted NIR light must be known if absorption spectra are to be used to 40

4 In vivo behaviour of near-infrared light optical pathlength were estimated in adult subjects with and without a scalp tourniquet. 20 No change was observed in NIR pathlength or CBV after arrest of scalp blood flow, confirmed by laser Doppler velocimetry. The authors con- cluded that scalp blood flow changes have little or no effect on NIRS measurement of cerebral haemodynamics. Cerebral blood flow (CBF) in anaesthetized patients has also been calculated using NIR probes placed first on the scalp and then on the dura. 4 CBF measured through the intact scalp was less than that measured using xenon 133 and was one- third that of NIRS measurement of CBF with probes placed on the dura. The authors concluded that underestimation was a result of extracerebral attenuation of NIR light. They suggested that 60 70% of the total optical pathlength of NIR light in the human head was through the extracerebral tissues, but they did not comment on the relative contribu- tions of the scalp and skull or on whether the dura or CSF might have an effect. Another recent in vivo NIR study demonstrated that an increase in interoptode distance from 0.7 to 5.5 cm increased the sensitivity of the monitor to cerebral attenuation changes. However, at any interoptode distance, extracerebral changes were significant compared with cerebral changes, implying that extracerebral effects are important in the use of clinical NIR spectroscopy monitors, particularly at shorter interoptode distances. 21 We studied transmission of NIR light in two distinct ways. In the first, a detecting probe was moved sequentially through the extracerebral layers of the cranium so that transmitted light intensity was sampled at each tissue interface. If the traditional model of NIR light transmission is correct, that is light travels in an arc whose depth is dependent on the separation of the emitting and detecting probes and that the extracerebral tissue layers have little effect on NIR light attenuation and path length, we would have anticipated small increases in detected optical intensity with removal of each successive extracerebral layer from the NIR light path. Alternatively, in the second part of the study, as the emitter was moved through the extracerebral layers, a greater proportion of the NIR pathlength would pass through cerebral tissue with an increased attenuation and we would have anticipated successive reductions in optical intensity detected at the skin surface. In contrast with the anticipated findings, we observed a major decrease in detected optical intensity when bone was removed from the NIR path in both arms of the study. The changes observed when the scalp was removed from either the emitting or detecting pathway varied considerably from patient to patient and the mean reduction in intensity was not statistically significant. There was a small, statistically significant decrease in detected light intensity when the dura and CSF were removed. We postulate that if all the extracerebral layers are intact, there is a preferential path or channel for NIR light from emitter to detector. That is, the bone or bone dura interfaces, and to a lesser extent the dura and CSF, may be allowing calculate absolute quantities of physiological chromophores. The mean photon pathlength may be estimated using timeof-flight and other methods These use sophisticated technology which is not yet suitable for bedside monitoring. The effect of the extracerebral layers on the passage and pathlength of NIR light is poorly understood. There is evidence that extracerebral attenuation of NIR light influences the accuracy of all cerebral spectrometers. 1 4 It is likely that the biological and optical characteristics of the extracerebral layers vary from patient to patient and from time to time. Indeed, it is theoretically possible for NIR light to travel from the emitter to the detector without passing through cerebral tissue. It has still to be shown that the configuration of a single NIR source and one or more ipsilateral receivers is capable of detecting intracranial NIR light attenuation changes through the intact scalp, skull, dura and CSF of adult humans. In the development of clinical NIRS monitors, extracerebral attenuation of NIR light has been assumed to be small and/or predictable. In particular, bone has been thought to be translucent to NIR light. Despite concerns about the effect of extracerebral tissues on NIR light, there have been few in vivo studies and most published work concerns the use of mathematical and in vitro models to predict the behaviour of NIR light in the adult human head In these phantom models, the amount of NIR light penetrating cerebral tissue is dependent on the separation of the light emitter and detector and on the optical characteristics of the extracerebral layers. The effect of discrete anatomical layers has been modelled. Using a simple Monte Carlo simulation of the adult head comprising two concentric spheres of differing media, Hiraoka and colleagues calculated that the pathlength of NIR light in cerebral tissue was 40 55% of the total pathlength 17. Others have studied the effect of the CSF layer in a simple model with layers of different optical properties and in a more complex model simulating cerebral sulci. 19 Mean optical pathlength was calculated experimentally and compared with Monte Carlo predictions. The results of that study suggested that at an emitter detector separation of 5.0 cm, 55% of the NIR light pathlength was in the scalp and skull, 20% in the CSF and, in contrast with Hiraoka s work, only 15% of the NIR pathlength was in the cerebral cortex. Because CSF has low scattering and absorption coefficients, the authors postulated that the CSF is acting as a conduit or channel for the light. In their study, however, the scalp and skull were considered as one homogeneous layer. In practice, the scalp and skull have very different optical properties and thus have very different effects on NIR light behaviour and path shape. The authors stated that the fractional pathlength of scalp and skull is much greater (55%) than that of any other layer. This would support our findings; since the absorption coefficient of the skin is much greater than that of the skull, most light should therefore pass unattenuated through the skull. In another study, cerebral blood volume (CBV) and 41

5 Young et al. optical channelling of NIR photons from emitter to 4 Owen-Reece H, Elwell CE, Harkness W, et al. Use of neardetector. Mathematical models have suggested that the CSF infrared spectroscopy to estimate cerebral blood flow in conscious and anaesthetised adult subjects. Br J Anaesth 1996; 76: 43 8 layer may act as an optical channel. 19 We do not know if 5 Cope M, Delpy DT, Reynolds EOR, Wray S, Wyatt JS, Van der the decrease in light intensity that we observed was a result Zee P. Methods of quantitating cerebral near infrared spectroscopy of loss of the bone layer itself or loss of the bone dura data. Adv Exp Med Biol 1988; 222: interface. 6 Firbank M, Schweiger M, Delpy DT. Investigation of light piping We accept that it was impossible to avoid some distortion through clear regions of scattering objects. Proc Soc Photooptical of boundary conditions after craniotomy or burr hole. Instrumentation Eng 1995; 2389: We also accept that there are minor increases in the 7 Firbank M, Arridge SR, Schweiger M, Delpy DT. An investigation of light transport through scattering bodies with non-scattering emitter detector distance (of the order of a few millimetres) regions. Phys Med Biol 1996; 41: with movement of the probes from the scalp to the cerebral 8 Hielscher AH, Liu H, Wang L, Tittel FK, Chance B, Jacques SL. surface. However, the consistency and magnitude of the Time resolved reflectance measurements on layered tissues with results in both arms of the study suggest that optical strongly varying optical properties. Proc Soc Photooptical channelling may be a real phenomenon and not an artefact Instrumentation Eng 1995; 2323: arising from disturbance of light transmission at optical 9 Okada E, Firbank M, Schweiger M, Arridge SR, Cope M, Delpy DT. A theoretical and experimental investigation of the effect of boundaries. It is difficult to envisage another clinical experisulci on light propagation in brain tissue. Proc Soc Photooptical mental technique in which it is possible to maintain boundary Instrumentation Eng 1995; 2626: 2 8 conditions. Furthermore, if the performance of NIRS is so 10 Jobsis FF. Non-invasive infrared monitoring of cerebral and sensitive to boundary conditions, it will be unreliable myocardial oxygen sufficiency and circulatory parameters. Science in those conditions such as head injury, subarachnoid 1977; 198: haemorrhage and postoperative monitoring when the boundcerebral blood and tissue oxygenation on newborn infants by 11 Cope M, Delpy DT. System for long-term measurement of ary conditions are altered by the clinical condition itself. infrared transillumination. Med Biol Eng Comput 1988; 26: We have only considered three major extracerebral layers. 12 Delpy DT, Cope M, van der Zee P, Arridge S, Wray S, Wyatt J. We have not been able to study the CSF layer independently Estimation of optical pathlength through tissue from direct time of the other layers and it is this layer which is most likely of flight measurement. Phys Med Biol 1988; 33: to change with small alterations in intracranial pressure not 13 Lakowicz JR, Berndt K. Frequency-domain measurements of accompanied by major changes in cerebral oxygenation. photon migration in tissues. Chem Phys Lett 1990; 166: Cope M, Delpy DT, Wray S, Wyatt JS, Reynolds EOR. A CCD Even under carefully controlled conditions, it is clear that spectrometer to quantitate the concentration of chromophores extracerebral tissues have a profound influence on NIR in living tissue utilising the absorption peak of water at 975nm. light transmission, and the behaviour of NIR light in these Adv Exp Med Biol 1989; 248: anatomical layers must be understood fully if NIRS is to 15 Matcher SJ, Cope M, Delpy DT. Use of the water absorption be developed as a reliable clinical monitor. spectrum to quantify tissue chromophore concentration changes in near infrared spectroscopy. Phys Med Biol 1994; 39: Hielscher AH, Liu H, Wang L, Tittel FK, Chance B, Jacques SL. Acknowledgements Determination of blood oxygenation in the brain by time-resolved We thank Dr John Carlin (Clinical Epidemiology and Biostatistics Unit, reflectance spectroscopy: influence of the skin, skull and meninges. Royal Children s Hospital, Melbourne, Australia) for statistical advice and Proc Soc Photooptical Instrumentation Eng 1994; 2136: 1 27 Dr Paul Brown (Department of Medical Physics, Frenchay Hospital) for 17 Hiraoka M, Firbank M, Essenpreis M, et al. A Monte Carlo help with data collection. Dr A. E. R. Young was supported by a grant investigation of optical pathlength in homogenous tissue and its from Research and Development Directorate, NHS Executive, South and application to near-infrared spectroscopy. Phys Med Biol 1993; 38: West Region Okada E, Firbank M, Delpy DT. The effect of overlying tissue on the spatial sensitivity profile of near-infrared spectroscopy. Phys References Med Biol 1995; 40: Germon TJ, Young AER, Manara AR, Nelson RJ. Extracerebral 19 Cui W, Kumar C, Chance B. Experimental study of migration absorption of near-infrared light influences the detection of depth for the photons measured at sample surface. Proc Soc increased cerebral oxygenation monitored by near-infrared Photooptical Instrumentation Eng 1991; 1431: spectroscopy. J Neurol Neurosurg Psychiatry 1995; 58: Owen-Reece H, Elwell CE, Wyatt JS, Delpy DT. The effect of 2 Harris DNF, Bailey SM. Near infrared spectroscopy in adults. scalp ischaemia on measurement of cerebral blood volume by Anaesthesia 1993; 48: near-infrared spectroscopy. Physiol Meas 1996; 17: Hopton P, Walsh TS, Lee A. CBF in adults using near-infrared 21 Germon TJ, Evans PD, Barnett NJ, Wall P, Manara AR, Nelson RJ. spectroscopy: potential for bedside measurement. Br J Anaesth 1996; 77: 131 Cerebral near-infrared spectroscopy: emitter detector separation must be increased. Br J Anaesth 1999; 82:

Cerebral near infrared spectroscopy: emitter detector separation must be increased

Cerebral near infrared spectroscopy: emitter detector separation must be increased British Journal of Anaesthesia 82 (6): 831 7 (1999) Cerebral near infrared spectroscopy: emitter detector separation must be increased T. J. Germon 1 *, P. D. Evans 3, N. J. Barnett 3,P.Wall 3, A. R. Manara

More information

Muscle OxyGen monitor. The Science Behind Moxy

Muscle OxyGen monitor. The Science Behind Moxy Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the

More information

REVIEW ARTICLE Near infrared spectroscopy

REVIEW ARTICLE Near infrared spectroscopy British Journal of Anaesthesia 82 (3): 418 26 (1999) REVIEW ARTICLE Near infrared spectroscopy H. Owen-Reece 1, M. Smith 1 *, C. E. Elwell 2 and J. C. Goldstone 3 1 Department of Neuroanaesthesia, National

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Phillips, J. P., Langford, R. M., Chang, S. H., Maney, K., Kyriacou, P. A. & Jones, D. P. (2009). Measurements of Cerebral

More information

Cerebrovascular Responses to Carbon Dioxide as Detected by Near-Infrared Spectrophotometry: Comparison of Three Different Measures

Cerebrovascular Responses to Carbon Dioxide as Detected by Near-Infrared Spectrophotometry: Comparison of Three Different Measures 003 1-399819413601-0020$03.00/0 PEDIATRIC RESEARCH Copyright O 1994 International Pediatric Research Foundation, Inc. Vol. 36, No. 1, 1994 Printed in U.S.A. Cerebrovascular Responses to Carbon Dioxide

More information

Supporting Information

Supporting Information Supporting Information Moriguchi and Hiraki 10.1073/pnas.0809747106 SI Text Differences in Brain Activation Between Preswitch and Postswitch Phases. The paired t test was used to compare the brain activation

More information

Clinically Available Optical Topography System

Clinically Available Optical Topography System Clinically Available Optical Topography System Clinically Available Optical Topography System 18 Fumio Kawaguchi Noriyoshi Ichikawa Noriyuki Fujiwara Yûichi Yamashita Shingo Kawasaki OVERVIEW: Progress

More information

Current bedside monitors of brain blood flow and oxygen delivery

Current bedside monitors of brain blood flow and oxygen delivery 24. Brain Chemistry Current bedside monitors of brain blood flow and oxygen delivery Global monitors Cannot detect regional abnormalities Local monitors Sample only a small region of the brain and highly

More information

Near Infrared Spectroscopy in Neonatal Intensive Care

Near Infrared Spectroscopy in Neonatal Intensive Care Near Infrared Spectroscopy in Neonatal Intensive Care Adré J. du Plessis, MBChB Fetal-Neonatal Neurology, Children s Hospital, Boston, MA Children s Hospital and Harvard Medical School Disclosure Statement

More information

Monitoring Tumor Therapy

Monitoring Tumor Therapy Monitoring Tumor Therapy Zenalux Biomedical, Inc. The Zenascope was used to monitor the physiological response to a vascular disrupting agent in mouse tumors. Changes in hemoglobin concentration, hemoglobin

More information

Noninvasive Detection of Bilirubin in Discrete Vessels

Noninvasive Detection of Bilirubin in Discrete Vessels , July 2-4, 2014, London, U.K. Noninvasive Detection of Bilirubin in Discrete Vessels Mark McEwen, Karen J. Reynolds Abstract Bilirubin, at the levels encountered in hyperbilirubinaemia, makes a similar

More information

Biomedical Instrumentation D. The Photoplethysmogram

Biomedical Instrumentation D. The Photoplethysmogram Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease

More information

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements

More information

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Maria Angela Franceschini, Sergio Fantini Bioengineering Center, Department of Electrical Engineering and

More information

In vivo optical imaging : revealing endogeneous optical contrast at depth

In vivo optical imaging : revealing endogeneous optical contrast at depth In vivo optical imaging : revealing endogeneous optical contrast at depth Anne PLANAT-CHRÉTIEN, Jean-Marc DINTEN CEA-LETI, MINATEC, Grenoble Jérôme GATEAU Université Paris Descartes, Paris 1 Why using

More information

Effects of hypoxaemia and bradycardia on neonatal

Effects of hypoxaemia and bradycardia on neonatal 376 Archives ofdisease in Childhood 1991; 66: 376-38 ORIGINAL ARTICLES The Neonatal Intensive Care Unit, North Staffordshire Maternity Hospital, Hilton Road, Harpfields, Stoke on Trent ST4 6SD L Nicola

More information

SPECTRAL CHARACTERISTICS OF SPONTANEOUS OSCILLATIONS IN CEREBRAL HAEMODYNAMICS ARE POSTURE DEPENDENT

SPECTRAL CHARACTERISTICS OF SPONTANEOUS OSCILLATIONS IN CEREBRAL HAEMODYNAMICS ARE POSTURE DEPENDENT SPECTRAL CHARACTERISTICS OF SPONTANEOUS OSCILLATIONS IN CEREBRAL HAEMODYNAMICS ARE POSTURE DEPENDENT Ilias Tachtsidis #, Clare E. Elwell #, Chuen-Wai Lee #, Terence S. Leung #, Martin Smith *, David T.

More information

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Takafumi Nagaoka, Kaoru Sakatani, Takayuki Awano, Noriaki Yokose, Tatsuya Hoshino, Yoshihiro

More information

Role of Frequency Domain Optical Spectroscopy in the Detection of Neonatal Brain Hemorrhage A Newborn Piglet Study

Role of Frequency Domain Optical Spectroscopy in the Detection of Neonatal Brain Hemorrhage A Newborn Piglet Study The Journal of Maternal-Fetal Medicine 9:142 149 (2000) NEW TECHNOLOGY Role of Frequency Domain Optical Spectroscopy in the Detection of Neonatal Brain Hemorrhage A Newborn Piglet Study Miljan R. Stankovic,

More information

Oxygen Saturation Measurement of Calf Muscle During Exercise in Intermittent Claudication

Oxygen Saturation Measurement of Calf Muscle During Exercise in Intermittent Claudication Eur J Vasc Endovasc Surg 23, 388±392 (2002) doi:10.1006/ejvs.2002.1645, available online at http://www.idealibrary.com on Oxygen Saturation Measurement of Calf Muscle During Exercise in Intermittent Claudication

More information

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 OBJECTIVES 1. Understand how cerebral oximetry works 2. Understand how cerebral oximetry may guide intervention 3. Learn

More information

Illuminating the Black Box: Investigating Prefrontal Cortical Hemodynamics During Exercise With Near-Infrared Spectroscopy

Illuminating the Black Box: Investigating Prefrontal Cortical Hemodynamics During Exercise With Near-Infrared Spectroscopy Journal of Sport and Exercise Psychology, 2009, 31, 505-553 2009 Human Kinetics, Inc. Illuminating the Black Box: Investigating Prefrontal Cortical Hemodynamics During Exercise With Near-Infrared Spectroscopy

More information

About OMICS International

About OMICS International About OMICS International OMICS International through its Open Access Initiative is committed to make genuine and reliable contributions to the scientific community. OMICS International hosts over 700

More information

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract Progress Report No. 2-3, March 31, 1999 The Home Automation and Healthcare Consortium Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy Prof. Kamal Youcef-Toumi Principal Investigator

More information

Probing Brain Oxygenation with Near Infrared Spectroscopy

Probing Brain Oxygenation with Near Infrared Spectroscopy 258 ARTICLE Probing Brain Oxygenation with Near Infrared Spectroscopy Alexander Gersten, Jacqueline Perle*, Amir Raz and Robert Fried* Abstract The fundamentals of near infrared spectroscopy (NIRS) are

More information

Unchanged Muscle Deoxygenation Heterogeneity During Bicycle Exercise After 6 Weeks of Endurance Training

Unchanged Muscle Deoxygenation Heterogeneity During Bicycle Exercise After 6 Weeks of Endurance Training Unchanged Muscle Deoxygenation Heterogeneity During Bicycle Exercise After 6 Weeks of Endurance Training Ryotaro Kime, Masatsugu Niwayama, Masako Fujioka, Kiyoshi Shiroishi, Takuya Osawa, Kousuke Shimomura,

More information

FALSE POSITIVES IN FUNCTIONAL NEAR- INFRARED TOPOGRAPHY

FALSE POSITIVES IN FUNCTIONAL NEAR- INFRARED TOPOGRAPHY FALSE POSITIVES IN FUNCTIONAL NEAR- INFRARED TOPOGRAPHY Ilias Tachtsidis 1, Terence S. Leung 1, Anchal Chopra 1, Peck H. Koh 1, Caroline B. Reid 1, and Clare E. Elwell 1 Abstract: Functional cranial near-infrared

More information

ESAT/SCD, Department of Electrical Engineering & IBBT Future Health Department, Katholieke Universiteit Leuven, Belgium

ESAT/SCD, Department of Electrical Engineering & IBBT Future Health Department, Katholieke Universiteit Leuven, Belgium Canonical Correlation Analysis in the Study of Cerebral and Peripheral Haemodynamics Interrelations with Systemic Variables in Neonates Supported on ECMO. Alexander Caicedo 1, Maria D. Papademetriou 2,

More information

Breast Cancer Detection by Determination of Optical Properties of Non- Malignant and Malignant Breast Tissues

Breast Cancer Detection by Determination of Optical Properties of Non- Malignant and Malignant Breast Tissues Research Article Breast Cancer Detection by Determination of Optical Properties of Non- Malignant and Malignant Breast Tissues El-Sharkawy YH * Department of Biomedical Engineering, Military Technical

More information

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification by Larissa Christina Schudlo A thesis submitted in conformity with the requirements for the degree

More information

Oxygen dependency and precision of cytochrome oxidase signal from full spectral NIRS of the piglet brain

Oxygen dependency and precision of cytochrome oxidase signal from full spectral NIRS of the piglet brain Am J Physiol Heart Circ Physiol 279: H2202 H2209, 2000. Oxygen dependency and precision of cytochrome oxidase signal from full spectral NIRS of the piglet brain R. SPRINGETT, 1 J. NEWMAN, 2 M. COPE, 1

More information

Optimal Differentiation of Tissue Types Using Combined Mid and Near Infrared Spectroscopy

Optimal Differentiation of Tissue Types Using Combined Mid and Near Infrared Spectroscopy Optimal Differentiation of Tissue Types Using Combined Mid and Near Infrared Spectroscopy Mugdha V. Padalkar, M.S. 1, Cushla M. McGoverin, Ph.D. 1, Uday P. Palukuru, M.S. 1, Nicholas J. Caccese 1, Padraig

More information

Optical non-invasive methods for characterization of the human health status

Optical non-invasive methods for characterization of the human health status Optical non-invasive methods for characterization of the human health status Abstract Jens Kraitl and Hartmut Ewald Institute of General Electrical Engineering, University of Rostock, Rostock, Germany

More information

ExperimentalPhysiology

ExperimentalPhysiology Exp Physiol 97.5 (2012) pp 557 561 557 Editorial ExperimentalPhysiology Categorized or continuous? Strength of an association and linear regression Gordon B. Drummond 1 and Sarah L. Vowler 2 1 Department

More information

Non-invasive sun protection factor determination using LED light

Non-invasive sun protection factor determination using LED light Non-invasive sun protection factor determination using LED light J. Lademann, S. Schanzer, S. Kobylinski, C. Reble, G. Khazaka, G. Wiora, H. Karrer, M. C. Meinke Outline 1. Introduction UV light and its

More information

Photoacoustic characterisation of vascular tissue at NIR wavelengths.

Photoacoustic characterisation of vascular tissue at NIR wavelengths. Photoacoustic characterisation of vascular tissue at NIR wavelengths. Thomas J. Allen and Paul C. Beard Department of Medical Physics and Bioengineering, University College London, Gower Street, London,

More information

Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State

Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State T. Matsumoto, Y. Fuchita, K. Ichikawa, Y. Fukuda, N. Takemura, and K. Sakatani Abstract According to the

More information

Blood Volume and Haemoglobin Oxygen Content Changes in Human Bone Marrow during Orthostatic Stress

Blood Volume and Haemoglobin Oxygen Content Changes in Human Bone Marrow during Orthostatic Stress Blood Volume and Haemoglobin Oxygen Content Changes in Human Bone Marrow during Orthostatic Stress Tiziano Binzoni 1),2), Terence S. Leung 3), Celine Courvoisier 4), Remo Giust 4), Gilbert Tribillon 4),

More information

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Pulse oximetry revisited Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Topics that will be discussed and dissected: Revisiting physiology

More information

Organometallic molecules such as hemoglobin

Organometallic molecules such as hemoglobin 596 Regional Cerebrovascular Oxygen Saturation Measured by Optical Spectroscopy in Humans Patrick W. McCormick, MD; Melville Stewart, MS; Mark G. Goetting, MD; and Guruswamy Balakrishnan, MD Regional cerebrovascular

More information

(Paper JBO/NIR-04 received Apr. 29, 1996; revised manuscript received Nov. 20, 1996; accepted for publication Dec. 26, 1996.)

(Paper JBO/NIR-04 received Apr. 29, 1996; revised manuscript received Nov. 20, 1996; accepted for publication Dec. 26, 1996.) JURNAL F BIMEDICAL PTICS 2(2), 147 153 (APRIL 1997) QUANTITATIVE SPECTRSCPIC DETERMINATIN F HEMGLBIN CNCENTRATIN AND SATURATIN IN A TURBID MEDIUM: ANALYSIS F THE EFFECT F WATER ABSRPTIN Maria Angela Franceschini,

More information

Introduction. Device Description. Dual-Band Spectral Output

Introduction. Device Description. Dual-Band Spectral Output Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction

More information

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Takahiro Igarashi, Kaoru Sakatani, Norio Fujiwara, Yoshihiro

More information

Brain tissue oxygen saturation increases during the night in adolescents

Brain tissue oxygen saturation increases during the night in adolescents Zurich Open Repository and Archive University of Zurich Main Library Strickhofstrasse 39 CH-8057 Zurich www.zora.uzh.ch Year: 2013 Brain tissue oxygen saturation increases during the night in adolescents

More information

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Pulse oximeter Masimo pulse CO-oximeter http://www.masimo.com/produc ts/continuous/radical-7/

More information

Kaoru Sakatani, Yoichi Katayama, Takamitsu Yamamoto, Susumu Suzuki

Kaoru Sakatani, Yoichi Katayama, Takamitsu Yamamoto, Susumu Suzuki J Neurol Neurosurg Psychiatry 999;67:769 77 769 Neurosurgery, Nihon University School of Medicine, Japan K Sakatani Y Katayama T Yamamoto Neurosurgery, China-Japan Friendship Hospital, China K Sakatani

More information

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19

More information

Near-infrared spectroscopy for monitoring muscle oxygenation

Near-infrared spectroscopy for monitoring muscle oxygenation Acta Physiol Scand 2000, 168, 615±622 Near-infrared spectroscopy for monitoring muscle oxygenation R. BOUSHEL 1 and C.A. PIANTADOSI 2 1 Sports Medicine Research Unit, Department of Rheumatology, Bispebjerg

More information

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2):209-217. 2011 Oxford University Press Effect of Phenylephrine and Ephedrine Bolus Treatment on Cerebral Oxygenation in Anaesthetized

More information

Near-infrared spectroscopy measurement of the pulsatile component of cerebral blood flow and volume from arterial oscillations

Near-infrared spectroscopy measurement of the pulsatile component of cerebral blood flow and volume from arterial oscillations Journal of Biomedical Optics 121, 014033 January/February 2007 Near-infrared spectroscopy measurement of the pulsatile component of cerebral blood flow and volume from arterial oscillations George Themelis

More information

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound

Sound in medicine. CH.12. Dr.Rajaa أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان. General Properties of Sound CH.12. Dr.Rajaa Sound in medicine أ.م.د. رجاء سهيل جنم جامعة تكريت كلية Sound : It is the audible waves of frequency between 20 Hz and 20 khz. Infrasound : refers to the sound of frequency below the normal

More information

Measurements of oxygen saturation of brain, liver and heart areas in the supine and sitting position using near infrared spectrophotometry

Measurements of oxygen saturation of brain, liver and heart areas in the supine and sitting position using near infrared spectrophotometry ORIGINAL ARTICLE in different positions and organs Romanian Journal of Anaesthesia and Intensive Care 2017 Vol 24 No 2, 101-106 DOI: http://dx.doi.org/10.21454/rjaic.7518.242.fss Measurements of oxygen

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

NIRS measurement of peripheral fractional oxygen extraction (FOE) after cardiopulmonary bypass 1

NIRS measurement of peripheral fractional oxygen extraction (FOE) after cardiopulmonary bypass 1 Spectroscopy 16 (2002) 191 197 191 IOS Press NIRS measurement of peripheral fractional oxygen extraction (FOE) after cardiopulmonary bypass 1 Charles William Yoxall a, Kusum Menon b, Andrew John Macnab

More information

Yuko Kondo, Kaoru Sakatani, Noriya Hirose, Takeshi Maeda, Jitsu Kato, Setsuro Ogawa, and Yoichi Katayama

Yuko Kondo, Kaoru Sakatani, Noriya Hirose, Takeshi Maeda, Jitsu Kato, Setsuro Ogawa, and Yoichi Katayama Chapter 16 Effect of Spinal Anesthesia for Elective Cesarean Section on Cerebral Blood Oxygenation Changes: Comparison of Hyperbaric and Isobaric Bupivacaine Yuko Kondo, Kaoru Sakatani, Noriya Hirose,

More information

Evidence of ventricular contamination of the optical signal in preterm neonates with post hemorrhagic ventricle dilation

Evidence of ventricular contamination of the optical signal in preterm neonates with post hemorrhagic ventricle dilation Western University Scholarship@Western Medical Biophysics Publications Medical Biophysics Department 3-12-2015 Evidence of ventricular contamination of the optical signal in preterm neonates with post

More information

and Neurosurgery, Queen Square, London WC1N 3BG, UK 3 Department of Biological Sciences, University of Essex, Wivenhoe Park, Colchester Essex,

and Neurosurgery, Queen Square, London WC1N 3BG, UK 3 Department of Biological Sciences, University of Essex, Wivenhoe Park, Colchester Essex, INSTITUTE OF PHYSICS PUBLISHING Physiol. Meas. 28 (2007) 199 211 PHYSIOLOGICAL MEASUREMENT doi:10.1088/0967-3334/28/2/008 Investigation of in vivo measurement of cerebral cytochrome-c-oxidase redox changes

More information

Photoacoustic imaging of lipid rich plaques in human aorta.

Photoacoustic imaging of lipid rich plaques in human aorta. Photoacoustic imaging of lipid rich plaques in human aorta. T. J. Allen a, A. Hall b, A. Dhillon b, J. S. Owen c and P. C. Beard a a Department of Medical Physics and Bioengineering, University College

More information

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Oxygen transport in blood Hemoglobin Oxygen O Deoxygen +O O Oximeters Arterial saturation

More information

Optical Sensor System for Hemoglobin Measurement

Optical Sensor System for Hemoglobin Measurement International Journal of Computational Engineering Research Vol, 03 Issue, 7 Optical Sensor System for Hemoglobin Measurement Rajashree Doshi 1, Anagha Panditrao 2 1 Department of Instrumentation and Control,

More information

Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves

Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves BY STEPHEN G. ROSENBAUM, B.A., LINNETTE D. ILIFF, B.SC, M. PHIL., PH.D., J. W. D. BULL, M.D., F.R.C.P., F.F.R., G. H. DU BOULAY,

More information

NIRS of the brain new diagnostic tool

NIRS of the brain new diagnostic tool NIRS of the brain new diagnostic tool Berndt Urlesberger Professor of Neonatology Head, Division of Neonatology, Department of Pediatrics Medical University Graz, Austria Email: berndt.urlesberger@medunigraz.at

More information

Special Section: Fractures

Special Section: Fractures 2609_c1_cover1.qxd 10/9/07 10:56 AM Page 1 THE LEADING EDGE September 2007, Vol. 26, No. 9 Special Section: Fractures September 2007 Vol. 26, No. 9 Pages 91081-1232 Fractures THE SOCIETY OF EXPLORATION

More information

PHOTOTHERAPY: WHAT DO WE KNOW AND HOW DOES IT WORK? Suite 3/36 O'Riordan Street Alexandria, Sydney NSW Australia

PHOTOTHERAPY: WHAT DO WE KNOW AND HOW DOES IT WORK? Suite 3/36 O'Riordan Street Alexandria, Sydney NSW Australia PHOTOTHERAPY: WHAT DO WE KNOW AND HOW DOES IT WORK? Suite 3/36 O'Riordan Street Alexandria, Sydney NSW Australia 2015 02 8331 8933 info@infinity-led.com.au DOCTORATE IN HEALTH SCIENCE DEAKIN UNIVERSITY

More information

Pulse oximetry in the accident and emergency department

Pulse oximetry in the accident and emergency department Archives of Emergency Medicine, 1989, 6, 137-142 Pulse oximetry in the accident and emergency department C. J. HOLBURN & M. J. ALLEN Accident and Emergency Department, Leicester Royal Infirmary, Leicester,

More information

Materials and Methods

Materials and Methods Anesthesiology 2007; 106:458 62 Copyright 2007, the American Society of Anesthesiologists, Inc. Lippincott Williams & Wilkins, Inc. Effects of Hemoglobin Concentration, Skull Thickness, and the Area of

More information

Near-infrared spectroscopic cerebral oxygenation reading in neonates and infants is associated with central venous oxygen saturation

Near-infrared spectroscopic cerebral oxygenation reading in neonates and infants is associated with central venous oxygen saturation Pediatric Anesthesia 2005 15: 102 109 doi:10.1111/j.14-9592.2005.014.x Near-infrared spectroscopic cerebral oxygenation reading in neonates and infants is associated with central venous oxygen saturation

More information

A Snapshot on Nuclear Cardiac Imaging

A Snapshot on Nuclear Cardiac Imaging Editorial A Snapshot on Nuclear Cardiac Imaging Khalil, M. Department of Physics, Faculty of Science, Helwan University. There is no doubt that nuclear medicine scanning devices are essential tool in the

More information

Pulse oximetry use in small animals

Pulse oximetry use in small animals Vet Times The website for the veterinary profession https://www.vettimes.co.uk Pulse oximetry use in small animals Author : Clara Rigotti, Marieke de Vries Categories : Vets Date : October 10, 2011 Clara

More information

Introduction to the Course and the Techniques. Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology

Introduction to the Course and the Techniques. Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology Introduction to the Course and the Techniques Jeffry R. Alger, PhD Ahmanson-Lovelace Brain Mapping Center Department of Neurology (jralger@ucla.edu) CTSI Neuroimaging April 2014 Rationale for the Course

More information

1. INTRODUCTION. National Hospital for Neurology & Neurosurgery, London

1. INTRODUCTION. National Hospital for Neurology & Neurosurgery, London RATE OF CHANGE IN CEREBRAL OXYGENATION AND BLOOD PRESSURE IN RESPONSE TO PASSIVE CHANGES IN POSTURE: A COMPARISON BETWEEN PURE AUTONOMIC FAILURE PATIENTS AND CONTROLS Ilias Tachtsidis, Clare E. Elwell,

More information

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Motor cortex activation fmri Near-infrared imaging Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Sergio Fantini s group, Department of Biomedical

More information

Cerebral Oximetry in Preterm Infants: Methods, Measurements and Evaluating Clinical Benefit

Cerebral Oximetry in Preterm Infants: Methods, Measurements and Evaluating Clinical Benefit Cerebral Oximetry in Preterm Infants: Methods, Measurements and Gorm Ole Greisen DrMedSci Professor and Consultant Neonatologist Department of Neonatology Rigshospitalet and Copenhagen University Copenhagen,

More information

BRITISH BIOMEDICAL BULLETIN

BRITISH BIOMEDICAL BULLETIN Journal Home Page www.bbbulletin.org BRITISH BIOMEDICAL BULLETIN Original Article Hemodynamic Changes after Infiltration of Local Anesthetic into Oral Mucosa during Oral Maxillofacial Surgery Kenichi Satoh*,

More information

(Paper JBO/NIR-O5 received Apr. 29, 1996; revised manuscript received Aug. 20, 1996; accepted for publication Sep. 5, 1996)

(Paper JBO/NIR-O5 received Apr. 29, 1996; revised manuscript received Aug. 20, 1996; accepted for publication Sep. 5, 1996) JOURNAL OF BIOMEDICAL OPTICS 1(4),387-398 (OCTOBER 1996) Monica Fabiani, t Gabriele Gratton, t and Paul M. Corballis* tuniversity of Missouri, Department of Psychology, McAlester Hall, Columbia, Missouri

More information

Up to date literature at your disposal to discover the benefits of laser dentistry

Up to date literature at your disposal to discover the benefits of laser dentistry Up to date literature at your disposal to discover the benefits of laser dentistry For further information contact info@doctor-smile.com +39-0444-349165 Dipartimento di Scienze e Tecnologie Biofisiche

More information

Neurovascular Physiology and Pathophysiology

Neurovascular Physiology and Pathophysiology Neurovascular Physiology and Pathophysiology The physiological questions aim at understanding the molecular and biochemical mechanisms, by which the brain adapts local blood flow to neuronal activity and

More information

Supporting Information

Supporting Information Supporting Information An efficient broadband and omnidirectional light-harvesting scheme employing the hierarchical structure based on ZnO nanorod/si 3 N 4 -coated Si microgroove on 5-inch single crystalline

More information

Chapter 27 PFC Blood Oxygenation Changes in Four Different Cognitive Tasks

Chapter 27 PFC Blood Oxygenation Changes in Four Different Cognitive Tasks Chapter 27 PFC Blood Oxygenation Changes in Four Different Cognitive Tasks Tomotaka Takeda, Yoshiaki Kawakami, Michiyo Konno, Yoshiaki Matsuda, Masayasu Nishino, Yoshihiro Suzuki, Yoshiaki Kawano, Kazunori

More information

Near-Infrared Spectroscopy of the Bladder: New Parameters for Noninvasive Evaluation of Detrusor Function

Near-Infrared Spectroscopy of the Bladder: New Parameters for Noninvasive Evaluation of Detrusor Function Near-Infrared Spectroscopy of the Bladder: New Parameters for Noninvasive Evaluation of Detrusor Function April 2013, Vancouver Babak Shadgan MD, MSc, PhD Postdoctoral Fellow Bladder Care Center Department

More information

The contribution of the external carotid artery to cerebral perfusion in carotid disease

The contribution of the external carotid artery to cerebral perfusion in carotid disease The contribution of the external carotid artery to cerebral perfusion in carotid disease Shirley J. Fearn, PhD, FRCS, Andrew J. Picton, BSc, Andrew J. Mortimer, MD, FRCA, Andrew D. Parry, MBChB, FRCS,

More information

Intrinsic Signal Optical Imaging

Intrinsic Signal Optical Imaging Intrinsic Signal Optical Imaging Introduction Intrinsic signal optical imaging (ISOI) is a technique used to map dynamics in single cells, brain slices and even and most importantly entire mammalian brains.

More information

Neurology and Neurosurgery, Queen Square, London WC1N 3BG, UK

Neurology and Neurosurgery, Queen Square, London WC1N 3BG, UK Investigation of in-vivo measurement of cerebral cytochrome-c-oxidase redox changes using near-infrared spectroscopy in patients with orthostatic hypotension. I Tachtsidis 1, M Tisdall 2, T S Leung 1,

More information

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood International Journal of Electronics Engineering Research. ISSN 0975-6450 Volume 9, Number 4 (2017) pp. 483-489 Research India Publications http://www.ripublication.com Development of an Optical Sensor

More information

Biomedical Research 2013; 24 (3): ISSN X

Biomedical Research 2013; 24 (3): ISSN X Biomedical Research 2013; 24 (3): 359-364 ISSN 0970-938X http://www.biomedres.info Investigating relative strengths and positions of electrical activity in the left and right hemispheres of the human brain

More information

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Evaluation of oesophageal reflectance pulse oximetry in major burns patients Burns 31 (2005) 337 341 www.elsevier.com/locate/burns Evaluation of oesophageal reflectance pulse oximetry in major burns patients S.K. Pal a, *, P.A. Kyriacou b, S. Kumaran c, S. Fadheel c, R. Emamdee

More information

Simulation study of optical transmission properties of ZnO thin film deposited on different substrates

Simulation study of optical transmission properties of ZnO thin film deposited on different substrates American Journal of Optics and Photonics 2013; 1(1) : 1-5 Published online February 20, 2013 (http://www.sciencepublishinggroup.com/j/ajop) doi: 10.11648/j.ajop.20130101.11 Simulation study of optical

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling

EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling EPSRC Centre for Doctoral Training in Industrially Focused Mathematical Modelling Uncertainty Quantification through multilevel Monte Carlo simulation in FEniCS Matteo Croci Contents 1. Introduction...

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

USE OF NEAR INFRARED SPECTROSCOPY TO IDENTIFY TRAUMATIC INTRACRANIAL HEMATOMAS

USE OF NEAR INFRARED SPECTROSCOPY TO IDENTIFY TRAUMATIC INTRACRANIAL HEMATOMAS JOURNAL OF BIOMEDICAL OPTICS 2(1), 31 41 (JANUARY 1997) USE OF NEAR INFRARED SPECTROSCOPY TO IDENTIFY TRAUMATIC INTRACRANIAL HEMATOMAS Claudia S. Robertson, Shankar P. Gopinath, and Britton Chance* Baylor

More information

Application of near-infrared spectroscopy for discrimination of mental workloads

Application of near-infrared spectroscopy for discrimination of mental workloads Application of near-infrared spectroscopy for discrimination of mental workloads A. Sassaroli a, F. Zheng a, M. Coutts a L. H. Hirshfield b, A. Girouard b, E. T. Solovey b, R. J. K. Jacob b, Y. Tong c,

More information

Contrast-enhanced diffuse optical tomography of brain perfusion in humans using ICG

Contrast-enhanced diffuse optical tomography of brain perfusion in humans using ICG Contrast-enhanced diffuse optical tomography of brain perfusion in humans using ICG Christina Habermehl* a, Christoph Schmitz a,b, Jens Steinbrink a,c a Berlin NeuroImaging Center, Charité University Hospital,

More information

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System

Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System Application of Phased Array Radar Theory to Ultrasonic Linear Array Medical Imaging System R. K. Saha, S. Karmakar, S. Saha, M. Roy, S. Sarkar and S.K. Sen Microelectronics Division, Saha Institute of

More information

(12) Patent Application Publication (10) Pub. No.: US 2008/ A1

(12) Patent Application Publication (10) Pub. No.: US 2008/ A1 US 20080081975A1 (19) United States (12) Patent Application Publication (10) Pub. No.: US 2008/0081975 A1 Agashe et al. (43) Pub. Date: (54) SYSTEM AND METHOD FOR DETECTION Publication Classification OF

More information

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama

More information

Developments in Ultrasonic Inspection II

Developments in Ultrasonic Inspection II Developments in Ultrasonic Inspection II An Ultrasonic Technique for the Testing of Plates Embedded in Concrete with Synthesis of Signals from a Multi-element Probe H. Ishida, Y. Kurozumi, Institute of

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

Stress Wave Focusing Transducers

Stress Wave Focusing Transducers UCRL-K-130697 PREPRINT Stress Wave Focusing Transducers Steven R. Visuri, Richard A. London, Luiz Da Silva This paper was prepared for submittal to Optical Society of America, Spring Topical Meetings Orlando,

More information

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert

University of Groningen. Quantitative CT myocardial perfusion Pelgrim, Gert University of Groningen Quantitative CT myocardial perfusion Pelgrim, Gert IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check

More information