1 ORIGINAL RESEARCH P. Mordasini M. Hiller C. Brekenfeld G. Schroth U. Fischer J. Slotboom M. Arnold J. Gralla In Vivo Evaluation of the Phenox CRC Mechanical Thrombectomy Device in a Swine Model of Acute Vessel Occlusion BACKGROUND AND PURPOSE: Mechanical thrombectomy in ischemic stroke is of increasing interest as it is a promising strategy for fast and efficient recanalization. Several thrombectomy devices have been introduced to the armentarium of mechanical thrombectomy. Currently, new devices are under development and are continuously added to the neurointerventional tool box. Each device advocated so far has a different design and mechanical properties in terms of thrombus-device interaction. Therefore, a systematic evaluation under standardized conditions in vivo of these new devices is needed. The purpose of this study was to evaluate the efficiency, thrombus-device interaction, and potential complications of the novel Phenox CRC for distal mechanical thrombectomy in vivo. MATERIALS AND METHODS: The device was evaluated in an established animal model in the swine. Recanalization rate, thromboembolic events, vasospasm, and complications were assessed. Radiopaque thrombi (2 cm length) were used for the visualization of thrombus-device interaction during retrieval. The Phenox CRC (4 mm diameter) was assessed in 15 vessel occlusions. For every occlusion a maximum of 3 retrieval attempts were performed. RESULTS: Complete recanalization (TICI 3/TIMI 3) was achieved in 86.7% of vessel occlusions. In 66.7% (10/15), the first retrieval attempt was successful, and in 20% (3/15), the second attempt led to complete recanalization of the parent artery. In 2 cases (13.3%) thrombus retrieval was not successful (TICI 0/TIMI 0). In 1 case (6.7%) a minor embolic event occurred in a small side branch. No distal thromboembolic event was observed during the study. Thrombus-device interaction illustrated the entrapment of the thrombus by the microfilaments and the proximal cage of the device. No significant thrombus compression was observed. No vessel perforation, dissection, or fracture of the device occurred. CONCLUSIONS: In this small animal study, the Phenox CRC was a safe and effective device for mechanical thrombectomy. The unique design with a combination of microfilaments and proximal cage reduces thrombus compression with a consequently high recanalization and low complication rate. ABBREVIATIONS: CRC clot retriever cage; pcr Phenox clot retriever; TICI Thrombolysis in Cerebral Infarction; TIMI Thrombolysis in Myocardial Infarction The goal of stroke treatment is fast and complete recanalization of the occluded brain-supplying arteries to restore cerebral blood flow. Mechanical thrombectomy in ischemic stroke treatment is of increasing interest. Recent studies examined whether mechanical approaches accelerate the time to recanalization, increase the recanalization rate, prolong the time window for treatment, and decrease the incidence of symptomatic hemorrhage compared with the established thrombolysis treatment protocols Furthermore, mechanical thrombectomy may be more effective in the disposing of large clot burdens in proximal vessels such as internal carotid artery or carotid-t occlusions. 11 All mechanical thrombectomy devices are delivered by endovascular access proximal to the occlusion site. The various devices can be divided into 2 major groups according to where they apply force on the thrombus: proximal devices apply Received July 10, 2009; accepted after revision September 15. From the Institute of Diagnostic and Interventional Neuroradiology (P.M., M.H., C.B., G.S., J.S., J.G.) and Department of Neurology (U.F., M.A.), University Hospital-Inselspital, Bern, Switzerland. Please address correspondence to Jan Gralla, MD, MSc, University of Bern-Inselspital, Institute of Interventional and Diagnostic Neuroradiology, Freiburgstr 10, CH-3010 Bern, Switzerland; jan.gralla@insel.ch DOI /ajnr.A1928 force to the proximal base of the thrombus (this group includes various aspiration catheters); distal devices approach the thrombus proximally, but then are advanced by guidewire and microcatheter to pass the thrombus and to be unsheathed behind it, where force is applied to the distal base of the thrombus. 12 Several distal thrombectomy devices have been introduced to the armamentarium of mechanical thrombectomy in neurointerventions and are clinically used in acute stroke treatment. Although clinically efficient, studies have shown a new spectrum of complications such as vessel dissection, hemorrhage, and thromboembolic events in previously unaffected territories. 2,5,9,10 One aspect might be the compression of the thrombus during retrieval due to the application of force to the distal end of the thrombus. 12 Each device advocated so far has a different design and mechanical properties in terms of thrombus-device interaction, 12,13 and systematic evaluation is therefore crucial. The Phenox CRC (Phenox, Bochum, Germany) device is a modification of the pcr device, 8,14,15 adding an additional self-expanding cage at the proximal end of a brushlike device. To date, no systematic evaluation of the device is available in the literature. The present study was performed in an established animal 972 Mordasini AJNR 31 May

2 Fig 1. Graphic illustration (A) and photograph (B) of the Phenox CRC device. The core wire compound holds radially oriented polyamide microfilaments trimmed in a conical, brushlike shape. The proximal end is additionally wired by a self-expanding nitinol basketlike cage. Radiodense markers are at the proximal and distal ends (asterisks). The Phenox CRC after successful clot retrieval (C). Note the encasement of the clot by the cage at the proximal end. model by using radiopaque thrombi to evaluate the efficiency, thrombus-device interaction, and potential complications of the Phenox CRC device for distal mechanical thrombectomy in vivo. Materials and Methods Thrombectomy Device The Phenox CRC is a modification of the pcr (European CE-marked since 2006). 8,14,15 It consists of a highly flexible nitinol/platinumalloy core wire compound. This wire compound holds radially oriented polyamide microfilaments and is marked by radiodense markers on the proximal and distal ends. The fibers have a preferential proximal direction, and the contour of this brushlike device is trimmed in a conical shape with an increasing diameter distally. The filament compound is resilient against unraveling. Its proximal end has been modified and is additionally wired by a self-expanding nitinol basketlike cage (Fig 1). Two diameters are available: 2 mm proximal increasing over a length of 2 cm to 4 mm distally (Phenox CRC ), and 3 mm increasing over a length of 2 cm to 5 mm distally (Phenox CRC ). The device is firmly molded to the body of a 3-m-long 0.4-mm pusher wire. The Phenox CRC and its pusher are positioned in an insertion tube and are delivered through a braided inch microcatheter (Rebar-027, 2.8F; ev3, Irvine, California, or Renagade Hi-Flo, 2.8F; Boston Scientific, Natick, Massachusetts, are recommended). The use of a 6F to 8F guide catheter is generally recommended. In this study a 7F guide catheter (Guider Softip, Boston Scientific) was used as a guide catheter. Equipped with a microwire (Silver Speed 0.014; ev3) the microcatheter (Rebar-027, 2.8F; ev3) was navigated into the occluded vessel and the thrombus was passed. After successful placement of the microcatheter 1 to 2 cm distally to the thrombus, the microwire was pulled out and the Phenox CRC was inserted by using a Y-shaped connector/valve and the insertion tube. Only the Phenox CRC with a maximal distal diameter of 4 mm was included in this study. The device was deployed 1 to 2 cm distal to the thrombus by retrieval of the microcatheter. Once deployed the Phenox CRC cannot be pulled back into the microcatheter. Finally, both the Phenox CRC and the microcatheter were pulled back simultaneously to contact the thrombus and were retrieved into the guide catheter. During retrieval, continuous aspiration was applied through the guide catheter by using a 60-mL syringe to avoid embolization of thrombus fragments. Animal Care All procedures were conducted according to international guidelines and were approved by the responsible local authorities. Five swine ranging in weight from 40 to 47 kg (Department of Clinical Research, University of Bern, Bern, Switzerland) were used in this study. Under general anesthesia a 65-cm-long 8F catheter sheath (Arrow International, Reading, Pennsylvania) was introduced in the common femoral artery and continuously flushed with heparinized saline (10 U/mL). No further heparin bolus or other thrombolytic drugs were administered. Further details are described in previous studies. 12,13,16 Thrombus Preparation Thrombus preparation and application were described in a previous study on this animal model. 12,13,16 In brief, thrombi were created by mixing 10 ml of autologous blood of the animal with 1gofbarium sulfate and 0.25 ml of bovine thrombin solution (Dade Behring, Newark, New Jersey). The material was incubated for 1 hour at room temperature and cut in thrombi of 20 mm length, which were injected into the target vessel. Due to the radio-opacity gained by the added barium sulfate, thrombus movement and thrombus-device interaction were visualized and evaluated during angiography and fluoroscopy. INTERVENTIONAL ORIGINAL RESEARCH AJNR Am J Neuroradiol 31: May

3 Fig 2. Angiogram of the lingual artery before embolization (A). Passing of the microcatheter and microwire between thrombus and the vessel wall (B). Unsheathing of the Phenox CRC distally to the thrombus (C; proximal and distal markers: arrows). Retrieval of the device (D F). The thrombus is mobilized in a stretched position (arrows) by the device during retrieval without compression or compaction. Angiography and Study Protocol Selective intra-arterial digital subtraction angiography was performed on a biplane high-resolution angiography system (Axiom Artis zee; Siemens, Erlangen, Germany). Iopamidol (Iopamiro 300; Bracco, Milan, Italy) was used for vessel contrast. After initial angiography of the cranial vessels, selective angiography of the cranial and supraaortic vessels was performed. These vessels reproduce the anatomic setting of an occlusion of the middle cerebral artery and the basilar artery in the human circulation, 16 which are the common localizations for interventional stroke therapy. Mean vessel diameter was mm (range, mm; median, 2.7 mm), corresponding well with the target vessels in humans as well as with the diameter of thrombi extracted from cerebral vessels of stroke patients. 17 For selective thromboembolization the preformed thrombus was injected into a 7F guide catheter (Guider Softip) positioned in the target vessel. After thrombus application the guide catheter was removed for 5 minutes to restore arterial flow and to allow thrombus embedding. The precise localization of the thrombus and complete occlusion of the vessel were assessed by follow-up angiography. The retrieval attempts were conducted according to the recommendations of the manufacturer as described previously. During retrieval only the microwire, microcatheter, and the Phenox CRC entered the target vessel. For each attempt a new device was used. The number of attempts to retrieve the thrombus was limited to 3 for each vessel occlusion. After the third attempt the clot retrieval was rated as failed. The application time for the initial attempt as well as for the following attempts was recorded. The device was evaluated in 15 vessel occlusions. All retrieval maneuvers were documented digitally, assessing thrombus fragmentation and peripheral embolization. Follow-up angiography was performed immediately and 1 hour after the procedure to evaluate recanalization rate, vessel dissection, vasospasm, or perforation. Analysis of Data Thrombus movement, fragmentation, and possible distal dislocation during the passing procedure with the microwire or microcatheter and during delivery of the device were analyzed. Thrombus fragmentation or loss with possible thromboembolization in other vessels during retrieval was assessed. Assessment of Vessel Morphology Recanalization rate of the parent vessel by using the TICI grading scale (0 to 3) 18 and the scale for TIMI (0 to 3) was evaluated. 19 For statistical evaluation, TICI/TIMI grades 0 and 1 were rated as insufficient and TICI/TIMI grades 2 and 3 were rated as sufficient recanalization. Grading of vasospasm in case of recanalization according to a scale introduced by Grandin et al 20 was used as follows: 0 no narrowing; 1 slight narrowing ( 25% reduction in lumen diameter); 2 moderate narrowing (25% 50% stenosis or 50% 75% stenosis af- 974 Mordasini AJNR 31 May

4 Fig 3. Deployment of the device distal to the thrombus (A; distal marker: arrow). Device retrieval (B; distal marker: arrow). Partial recanalization after first retrieval attempt with residual thrombus (C; thrombus: arrow). Passing of the residual thrombus in a second attempt (D and E; arrows) and retrieval of the entire residual thrombus (F; arrow). fecting only a short segment of vessel); and 3 severe narrowing (50% 75% stenosis affecting a long segment of vessel or any stenosis 75%). Retrieval Device The retrieval success was evaluated regarding recanalization rate and number of attempts to complete thrombus retrieval as well as the total time interval to achieve recanalization, loss of thrombotic material at the tip of the guide catheter, and feasibility and technical failure. Results Efficiency and Complications A total of 22 retrieval attempts for recanalization were performed on 15 vessel occlusions. Initial deployment of the device distal to the thrombus was feasible in all cases. Complete recanalization (TICI/TIMI 3) was achieved in 86.7% of vessel occlusions (13 of 15 occlusions). In 66.7% (10/15), the first retrieval attempt was successful (Fig 2), and in 20% (3/15) the second attempt led to complete recanalization of the parent artery (Fig 3). In 2 cases (13.3%) thrombus retrieval was not successful (TICI/TIMI 0) and no clot material could be retrieved. Location of vessel occlusion, number of attemps per vessel, and recanalization results are summarized in the Table. No vessel straightening, indicating high friction during retrieval, occurred in the failed attempts. Mean number of attempts for recanalization was 1.7. Mean time to recanalization was 7 minutes 20 seconds (range, 1 29 minutes). Vasospasm rate was grade 0 in 20.0% (3/15), grade 1 in 33.3% (5/15), grade 2 in 20.0% (3/15), and grade 3 in 26.7% (4/15). Mean vasospasm rating after recanalization was 1.5. At follow-up angiography after 1 hour, vasospasms usually had resolved, allowing assessment of the recanalization result. In 1 case (6.7%) the distal part of thrombus was lost in the carrying vessel leading to a minor embolic event in a small adjacent side branch, but no distal thromboembolic event in the target vessel or previously unaffected vessel territories was observed. Neither vessel perforation nor vessel dissection was AJNR Am J Neuroradiol 31: May

5 Recanalization scores Vessel Attempts Recanalization (TICI/TIMI) Pig 1 Axillary branch right 1 3 Axillary branch right 1 3 Pig 2 Axillary branch left 1 3 Axillary branch left 2 3 Pig 3 Lingual artery left 2 3 Lingual artery right 1 3 Lingual artery left 3 0 Axillary branch right 1 3 Pig 4 Axillary branch right 1 3 Axillary branch left 1 3 Axillary branch left 1 3 Axillary branch right 1 3 Axillary branch left 3 0 Pig 5 Lingual artery right 1 3 Lingual artery left 2 3 found during control angiography. There was no device fracture during the study. Thrombus-Device Interaction The passage of the thrombus with the microcatheter and microwire was possible in all cases and never led to distal dislocation or fragmentation of the thrombus. In all cases passing of the thrombus occurred between the thrombus and the vessel wall without penetrating the thrombus, consistent with the findings in other studies 12,13,21 and clinical experience. Unsheathing of the device distally to the thrombus was feasible without complications. However, deployment of the device can be complicated by restricted space distal to the thrombus due to branching or narrowing of the target vessel. Thrombus-device interaction illustrated the entrapment of the thrombus by the proximal cage and the microfilaments of the device. Typically, the thrombus was encased proximally by the cage and loaded lengthwise as a whole by the microfilaments depicted by a subtle radiolucent rim between the opacified thrombus and the core wire. During mobilization the thrombus remained in a stretched position, which avoids thrombus compaction during retrieval. Consecutively, no significant thrombus compression occurred (Figs 2 and 4). In 1 case pulling back of the device led to dislocation of the thrombus by pushing it in a proximally located side branch. No shearing off of the thrombus from the device into the side vessel occurred. The engagement of the proximal tip of the thrombus by the cage led to stable fixation of the thrombus and enabled retrieval from the side branch without embolization (Fig 5). Retrieval of the device into the guiding catheter was uneventful in all cases, and no loss of thrombotic material at the tip of the guiding catheter was observed by using additional aspiration. Discussion The interest in mechanical thrombectomy for acute ischemic stroke has significantly increased in recent years. 6,8,10,22 Today, several distal thrombectomy devices with variable designs and shapes and therefore different mechanical properties in terms of thrombus-device interaction have been advocated. Currently, new thrombectomy devices are under development Fig 4. Fluoroscopic magnification during retrieval showing encasement of the thrombus proximally by the cage (arrowhead) and lengthwise loading of the thrombus by the microfilaments depicted by a subtle radiolucent rim between the opacified thrombus and the core wire (arrow). and are continuously added to the armamentarium of interventional stroke treatment. However, mechanical manipulation of a thrombus in the setting of acute ischemic stroke might cause severe complications. Therefore, systematic evaluation of the mechanical properties under standardized conditions in vivo is needed, and animal studies provide an attractive tool for device evaluation and assessment before clinical use in stroke patients. The Phenox CRC device is a modification of the Phenox pcr device, 8,14,15 adding an additional self-expanding nitinol cage at the proximal end of a brushlike device. In the present study, the novel device was found to be effective for recanalization with an overall complete recanalization rate (TICI/ TIMI 3) of 86.7%. Remarkably, complete recanalization could be achieved in the first attempt in 66.7% of vessel occlusions. After passing of the occlusion site and deployment, distal thrombectomy devices apply force to the distal base of the thrombus. Previous studies on the evaluation of distal thrombectomy devices in vivo have illustrated a large amount of thrombus compression during retrieval. Compression of the thrombus while pulled back increases its diameter and thereby increases the friction between the thrombus and the vessel wall. This situation leads to an increased force necessary to mobilize the thrombus, a reduction in retrieval rate, and more complicated passage maneuvers in the case of additional recanalization attempts as shown in previous studies. 12,13,23 Although this observation is difficult to quantify, we have clearly seen in this study that a minimum of thrombus compression during retrieval is a typical feature of the Phenox CRC device. Thrombus-device interaction in our study illustrated the encasement of the thrombus by the proximal cage and loading of the thrombus by the microfilaments of the Phenox CRC device. In contrast to other distal thrombectomy devices, the thrombus was typically mobilized in a stretched position with- 976 Mordasini AJNR 31 May

6 Fig 5. Deployment of the device distally to the thrombus (A). Retrieval of the device leads to dislocation of the thrombus into a proximal side branch (B D; arrow). Fixation through engagement of the proximal tip of the thrombus by the cage enables retrieval from the side branch without embolization (E H). During retrieval no shearing off of the thrombus from the device into the side vessel occurred. out significant compression. Due to this stretched and encased in-line position of the thrombus with the device, it is less prone to shearing forces and flow during retrieval. This facilitates thrombectomy and decreases the risk of thrombus fragmentation and distal embolization. Unlike other distal thrombectomy devices, no shearing off of the thrombus from the device into the side vessel occurred. 12,13 The encasement of the proximal part of the thrombus by the cage seems to add significant additional fixation and stabilization of the thrombus-device complex (Fig 5). In only 1 case (6.7%) the distal part of a thrombus was lost in the carrying vessel leading to a minor embolic event into a small adjacent side branch, but no distal embolization into the target vessel was observed. This unique design allows effective thrombus mobilization and consecutively high recanalization rates, especially during the first retrieval attempt. The short time to recanalization (mean, 7 minutes 20 seconds) implies the possibility of fast recanalization in the clinical setting. The retention force of the microfilaments is either sufficient to mobilize the thrombus through transmission of the pulling force or they fold back, resulting in an unsuccessful, but less traumatic passage of the device through the thrombus as shown in an in vitro evaluation of the pcr. 15 Failed thrombus retrieval occurred in our study in 2 cases (TIMI 0 in 13.3%) despite repeated retrieval attempts. Firm clot adhesion to the vessel wall or jamming of the clot in the vessel due to vasospasm may be the cause of these failed retrieval attempts. Consistent with these findings, no thromboembolic complications were seen during repeated retrieval of the device through the thrombus in cases of partial recanalization or failed attempts. The attenuated array of the microfilaments may additionally be able to reduce distal emboli due to a filter effect. 15 Complications reported in clinical studies of distal thrombectomy devices such as vessel perforation, dissection, or device fracture did not occur in our study. This may be due to the limited number of attempts performed in the study or be biased by the different vessel morphology lacking atherosclerotic changes in the animal model. No vessel straightening occurred in failed attempts or in settings with increased resistance as observed for other distal devices 13 attributable to the flexibility of the backward-oriented microfilaments and the core wire. This flexibility allows a rather atraumatic retrieval process and may facilitate access to more distal branches such as the M2 or even M3 branches of the middle cerebral artery. 14 Limitations The device provoked a moderate degree of vasospasm during and after retrieval. However, repeat angiogram after restitution of vasospasm did not reveal any vessel wall damage. Vasospasm is a consistent finding as well in previous studies, 12,13,23 reflecting a limitation of the animal model with young vessels prone to vasospasm. However, the grade of vasospasm can be considered a measure for vessel wall irritation. The vessels in this animal model are also less tortuous, AJNR Am J Neuroradiol 31: May

7 less elongated, and have less atherosclerotic changes compared with the human vasculature encountered in stroke patients. Studies on thrombus morphology revealed a variety of different thrombi compositions found in human stroke. Although the predominant thrombus is a whole blood thrombus or red clot, the composition (eg, fibrin-rich thrombus or white clot) might have a significant influence on the mechanical stability and therefore the success rate of thrombectomy. The used whole blood thrombus conforms to a slightly firm clot. Such clot has different mechanical properties compared with hard clot. The generated whole blood thrombus advocated in this model might therefore not represent the whole spectrum of mechanical properties of thrombi found in acute stroke. Therefore, the success rate of the mechanical device might be overestimated in our model. Conclusions In this small animal study, the Phenox CRC clot retriever was a safe and effective device for mechanical thrombectomy. The unique design and mechanical concept with a combination of microfilaments and proximal cage seems to reduce thrombus compression with a consequently high recanalization. The uncomplicated applicability and the atraumatic design due to its flexibility lead to a short time to recanalization and a low complication rate. References 1. Lutsep HL, Clark WM, Nesbit GM, et al. Intraarterial suction thrombectomy in acute stroke. AJNR Am J Neuroradiol 2002;23: Gobin YP, Starkman S, Duckwiler GR, et al. MERCI 1: a phase 1 study of mechanical embolus removal in cerebral ischemia. Stroke 2004;35: Nedeltchev K, Remonda L, Do DD, et al. Acute stenting and thromboaspiration in basilar artery occlusions due to embolism from the dominating vertebral artery. Neuroradiology 2004;46: Greenberg RK, Ouriel K, Srivastava S, et al. Mechanical versus chemical thrombolysis: an in vitro differentiation of thrombolytic mechanisms. J Vasc Interv Radiol 2000;11: Smith WS, Sung G, Starkman S, et al. Safety and efficacy of mechanical embolectomy in acute ischemic stroke: results of the MERCI trial. Stroke 2005;36: Noser EA, Shaltoni HM, Hall CE, et al. Aggressive mechanical clot disruption: a safe adjunct to thrombolytic therapy in acute stroke? Stroke 2005;36: Versnick EJ, Do HM, Albers GW, et al. Mechanical thrombectomy for acute stroke. AJNR Am J Neuroradiol 2005;26: Liebig T, Reinartz J, Janker R, et al. Early clinical experience with a new thrombectomy device for the treatment of ischemic stroke. Stroke 2008;39: Smith WS. Safety of mechanical thrombectomy and intravenous tissue plasminogen activator in acute ischemic stroke: results of the Multi Mechanical Embolus Removal in Cerebral Ischemia (MERCI) trial, part I. AJNR Am J Neuroradiol 2006;27: Smith WS, Sung G, Saver J, et al. Mechanical thrombectomy for acute ischemic stroke: final results of the Multi MERCI trial. Stroke 2008;39: Nedeltchev K, Brekenfeld C, Remonda L, et al. Internal carotid artery stent implantation in 25 patients with acute stroke: preliminary results. Radiology 2005;237: Gralla J, Schroth G, Remonda L, et al. Mechanical thrombectomy for acute ischemic stroke: thrombus-device interaction, efficiency, and complications in vivo. Stroke 2006;37: Brekenfeld C, Schroth G, El Koussy M, et al. Mechanical thromboembolectomy for acute ischemic stroke: comparison of the catch thrombectomy device and the Merci Retriever in vivo. Stroke 2008;39: Henkes H, Reinartz J, Lowens S, et al. A device for fast mechanical clot retrieval from intracranial arteries (Phenox clot retriever). Neurocrit Care 2006;5: Liebig T, Reinartz J, Hannes R, et al. Comparative in vitro study of five mechanical embolectomy systems: effectiveness of clot removal and risk of distal embolization. Neuroradiology 2008;50: Gralla J, Schroth G, Remonda L, et al. A dedicated animal model for mechanical thrombectomy in acute stroke. AJNR Am J Neuroradiol 2006;27: Marder VJ, Chute DJ, Starkman S, et al. Analysis of thrombi retrieved from cerebral arteries of patients with acute ischemic stroke. Stroke 2006;37: Higashida RT, Furlan AJ, Roberts H, et al. Trial design and reporting standards for intra-arterial cerebral thrombolysis for acute ischemic stroke. Stroke 2003;34:e109 e The Thrombolysis in Myocardial Infarction (TIMI) trial, phase I findings: TIMI Study Group. N Engl J Med 1985;312: Grandin CB, Cosnard G, Hammer F, et al. Vasospasm after subarachnoid hemorrhage: diagnosis with MR angiography. AJNR Am J Neuroradiol 2000;21: Brekenfeld C, Tinguely P, Schroth G, et al. Percutaneous transluminal angioplasty and stent placement in acute vessel occlusion: evaluation of new methods for interventional stroke treatment. AJNR Am J Neuroradiol 2009;30: Devlin TG, Baxter BW, Feintuch TA, et al. The Merci Retrieval System for acute stroke: the Southeast Regional Stroke Center experience. Neurocrit Care 2007;6: Gralla J, Burkhardt M, Schroth G, et al. Occlusion length is a crucial determinant of efficiency and complication rate in thrombectomy for acute ischemic stroke. AJNR Am J Neuroradiol 2008;29: Mordasini AJNR 31 May

Intra-arterial thrombolysis has improved the prognosis of

Intra-arterial thrombolysis has improved the prognosis of Mechanical Thrombectomy for Acute Ischemic Stroke Thrombus Device Interaction, Efficiency, and Complications In Vivo Jan Gralla, MD; Gerhard Schroth, MD; Luca Remonda, MD; Krassen Nedeltchev, MD; Johannes

More information

Limitations of thrombolysis, including low recanalization

Limitations of thrombolysis, including low recanalization ORIGINAL RESEARCH Solitaire Flow-Restoration Device for Treatment of Acute Ischemic Stroke: Safety and Recanalization Efficacy Study in a Swine Vessel Occlusion Model R. Jahan BACKGROUND AND PURPOSE: The

More information

Early recanalization reduces mortality and improves outcome

Early recanalization reduces mortality and improves outcome ORIGINAL RESEARCH C. Brekenfeld G. Schroth P. Mordasini U. Fischer M.-L. Mono A. Weck M. Arnold M. El-Koussy J. Gralla Impact of Retrievable Stents on Acute Ischemic Stroke Treatment BACKGROUND AND PURPOSE:

More information

ENDOVASCULAR THERAPIES FOR ACUTE STROKE

ENDOVASCULAR THERAPIES FOR ACUTE STROKE ENDOVASCULAR THERAPIES FOR ACUTE STROKE Cerebral Arteriogram Cerebral Anatomy Cerebral Anatomy Brain Imaging Acute Ischemic Stroke (AIS) Therapy Main goal is to restore blood flow and improve perfusion

More information

Acute ischemic cerebrovascular stroke remains a severe disease

Acute ischemic cerebrovascular stroke remains a severe disease ORIGINAL RESEARCH J. Gralla G. Schroth L. Remonda A. Fleischmann J. Fandino J. Slotboom C. Brekenfeld A Dedicated Animal Model for Mechanical Thrombectomy in Acute Stroke BACKGROUND: Recent studies have

More information

New Stroke Interventions. Scott L. Zuckerman M.D. Vanderbilt Neurosurgery

New Stroke Interventions. Scott L. Zuckerman M.D. Vanderbilt Neurosurgery New Stroke Interventions Scott L. Zuckerman M.D. Vanderbilt Neurosurgery Agenda Clot Retrieval Devices Merci Penumbra Stent Retrievers Solitaire Trevo New Technology Funnel ReCover MERCI Retriever (2004)

More information

Department of Neuroradiology, Klinikum rechts der Isar, Technische Universität München, Ismaninger Straße 22, Munich, Germany 2

Department of Neuroradiology, Klinikum rechts der Isar, Technische Universität München, Ismaninger Straße 22, Munich, Germany 2 The Scientific World Journal Volume 212, Article ID 19763, 7 pages doi:1.11/212/19763 The cientificworldjournal Clinical Study The Phenox Clot Retriever as Part of a Multimodal Mechanical Thrombectomy

More information

Distal Mechanical Thrombectomy in Acute Ischemic Stroke Method and Benefit. Hans Henkes, Wiebke Kurre Stuttgart, Germany

Distal Mechanical Thrombectomy in Acute Ischemic Stroke Method and Benefit. Hans Henkes, Wiebke Kurre Stuttgart, Germany Distal Mechanical Thrombectomy in Acute Ischemic Stroke Method and Benefit Hans Henkes, Wiebke Kurre Stuttgart, Germany 1 Thrombectomy... with stent-retrievers is an evidence based therapy for intracranial

More information

Solitaire FR Revascularization Device

Solitaire FR Revascularization Device Solitaire FR Revascularization Device FEATURING PARAMETRIC DESIGN Restore. Retrieve. Revive. 5 Revolutionizing Mechanical Thrombectomy with Parametric Design Solitaire FR Device The Solitaire FR revascularization

More information

Mechanical recanalization in acute stroke treatment

Mechanical recanalization in acute stroke treatment Perspectives in Medicine (2012) 1, 54 58 Bartels E, Bartels S, Poppert H (Editors): New Trends in Neurosonology and Cerebral Hemodynamics an Update. Perspectives in Medicine (2012) 1, 54 58 journal homepage:

More information

The National Institutes of Health Stroke Scale (NIHSS)

The National Institutes of Health Stroke Scale (NIHSS) National Institutes of Health Stroke Scale Score and Vessel Occlusion in 252 Patients With Acute Ischemic Stroke Mirjam R. Heldner, MD; Christoph Zubler, MD; Heinrich P. Mattle, MD; Gerhard Schroth, MD;

More information

CAROTID ARTERY ANGIOPLASTY

CAROTID ARTERY ANGIOPLASTY CAROTID ARTERY ANGIOPLASTY Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This Medical Coverage Guideline

More information

Acute basilar artery occlusion (BAO) is associated with a very

Acute basilar artery occlusion (BAO) is associated with a very ORIGINAL RESEARCH INTERVENTIONAL Acute Basilar Artery Occlusion: Outcome of Mechanical Thrombectomy with Solitaire Stent within 8 Hours of Stroke Onset J.M. Baek, W. Yoon, S.K. Kim, M.Y. Jung, M.S. Park,

More information

Mechanical Thrombectomy of Large Vessel Occlusions Using Stent Retriever Devices

Mechanical Thrombectomy of Large Vessel Occlusions Using Stent Retriever Devices Mechanical Thrombectomy of Large Vessel Occlusions Using Stent Retriever Devices Joey English MD, PhD Medical Director, Neurointerventional Services California Pacific Medical Center Hospitals, San Francisco,

More information

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service M AY. 6. 2011 10:37 A M F D A - C D R H - O D E - P M O N O. 4147 P. 1 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control

More information

In acute ischemic stroke, recanalization of an occluded cerebral

In acute ischemic stroke, recanalization of an occluded cerebral ORIGINAL RESEARCH INTERVENTIONAL Double Solitaire Mechanical Thrombectomy in Acute Stroke: Effective Rescue Strategy for Refractory Artery Occlusions? J. Klisch, V. Sychra, C. Strasilla, C.A. Taschner,

More information

Recanalization is a powerful predictor of stroke outcome in

Recanalization is a powerful predictor of stroke outcome in ORIGINAL RESEARCH INTERVENTIONAL Histopathologic Evaluation of Arterial Wall Response to 5 Neurovascular Mechanical Thrombectomy Devices in a Swine Model B. Gory, D. Bresson, I. Kessler, M.L. Perrin, A.

More information

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER

1 Description. 2 Indications. 3 Warnings ASPIRATION CATHETER Page 1 of 5 ASPIRATION CATHETER Carefully read all instructions prior to use, observe all warnings and precautions noted throughout these instructions. Failure to do so may result in complications. STERILE.

More information

Endovascular stroke treatments are being increasingly used

Endovascular stroke treatments are being increasingly used Published March 18, 2010 as 10.3174/ajnr.A2050 ORIGINAL RESEARCH A.C. Flint S.P. Cullen B.S. Faigeles V.A. Rao Predicting Long-Term Outcome after Endovascular Stroke Treatment: The Totaled Health Risks

More information

An intravenous thrombolysis using recombinant tissue

An intravenous thrombolysis using recombinant tissue ORIGINAL RESEARCH I. Ikushima H. Ohta T. Hirai K. Yokogami D. Miyahara N. Maeda Y. Yamashita Balloon Catheter Disruption of Middle Cerebral Artery Thrombus in Conjunction with Thrombolysis for the Treatment

More information

Single Pass MCA Revascularization with Trevo

Single Pass MCA Revascularization with Trevo Single Pass MCA Revascularization with Trevo By Concentric Medical ProVueTM Retriever Case Reported By Sascha Prothmann, MD Klinikum rechts der Isar, Munich, Germany Images contained herein courtesy of

More information

Mechanical thrombectomy is an effective tool for recanalization

Mechanical thrombectomy is an effective tool for recanalization ORIGINAL RESEARCH INTERVENTIONAL Frequency and Relevance of Anterior Cerebral Artery Embolism Caused by Mechanical Thrombectomy of Middle Cerebral Artery Occlusion W. Kurre, K. Vorlaender, M. Aguilar-Pérez,

More information

Comparison of Five Major Recent Endovascular Treatment Trials

Comparison of Five Major Recent Endovascular Treatment Trials Comparison of Five Major Recent Endovascular Treatment Trials Sample size 500 # sites 70 (100 planned) 316 (500 planned) 196 (833 estimated) 206 (690 planned) 16 10 22 39 4 Treatment contrasts Baseline

More information

Spontaneous Recanalization after Complete Occlusion of the Common Carotid Artery with Subsequent Embolic Ischemic Stroke

Spontaneous Recanalization after Complete Occlusion of the Common Carotid Artery with Subsequent Embolic Ischemic Stroke Original Contribution Spontaneous Recanalization after Complete Occlusion of the Common Carotid Artery with Subsequent Embolic Ischemic Stroke Abstract Introduction: Acute carotid artery occlusion carries

More information

Institute for Science and Technology in Medicine, Keele University, Keele, Staffordshire ST6 6TP, UK 3

Institute for Science and Technology in Medicine, Keele University, Keele, Staffordshire ST6 6TP, UK 3 SAGE-Hindawi Access to Research Stroke Research and Treatment Volume 2011, Article ID 186424, 6 pages doi:10.4061/2011/186424 Research Article Comparison of the Effectiveness of Three Methods of Recanalization

More information

There have been numerous advances in cerebral revascularization

There have been numerous advances in cerebral revascularization ORIGINAL RESEARCH INTERVENTIONAL Comparison of Stent-Retriever Devices versus the Merci Retriever for Endovascular Treatment of Acute Stroke E. Broussalis, E. Trinka, W. Hitzl, A. Wallner, V. Chroust,

More information

Treatment with intravenous rtpa has proved successful in

Treatment with intravenous rtpa has proved successful in ORIGINAL RESEARCH INTERVENTIONAL Mechanical Embolectomy for Acute Ischemic Stroke in the Anterior Cerebral Circulation: The Gothenburg Experience during 2000 2011 A. Rentzos, C. Lundqvist, J.-E. Karlsson,

More information

Thromboembolic occlusion of major cerebral arteries is

Thromboembolic occlusion of major cerebral arteries is ORIGINAL RESEARCH Z. Kulcsár C. Bonvin V.M. Pereira S. Altrichter H. Yilmaz K.O. Lövblad R. Sztajzel D.A. Rüfenacht Penumbra System: A Novel Mechanical Thrombectomy Device for Large-Vessel Occlusions in

More information

Stroke Clinical Trials Update Transitioning to an Anatomic Diagnosis in Ischemic Stroke

Stroke Clinical Trials Update Transitioning to an Anatomic Diagnosis in Ischemic Stroke Stroke Clinical Trials Update Transitioning to an Anatomic Diagnosis in Ischemic Stroke Alexander A. Khalessi MD MS Director of Endovascular Neurosurgery Surgical Director of NeuroCritical Care University

More information

Current thrombolytic therapy in acute ischemic stroke is

Current thrombolytic therapy in acute ischemic stroke is Published May 22, 2008 as 10.3174/ajnr.A1110 ORIGINAL RESEARCH A. Bose H. Henkes K. Alfke W. Reith T.E. Mayer A. Berlis V. Branca S.P. Sit, for the Penumbra Phase 1 Stroke Trial Investigators The Penumbra

More information

The hyperdense vessel sign on CT predicts successful recanalization with the Merci device in acute ischemic stroke

The hyperdense vessel sign on CT predicts successful recanalization with the Merci device in acute ischemic stroke Editor s choice Scan to access more free content 1 Neuro Interventional Service, University of Iowa, Carver College of Medicine, Iowa City, Iowa, USA 2 Division of Interventional Neuroradiology, UCLA,

More information

The prognosis for acute basilar artery occlusion (BAO) is

The prognosis for acute basilar artery occlusion (BAO) is Published July 17, 2014 as 10.3174/ajnr.A4045 ORIGINAL RESEARCH INTERVENTIONAL Forced Arterial Suction Thrombectomy with the Penumbra Reperfusion Catheter in Acute Basilar Artery Occlusion: A Retrospective

More information

Since the National Institute of Neurologic Disorders and

Since the National Institute of Neurologic Disorders and ORIGINAL RESEARCH R.M. Sugg E.A. Noser H.M. Shaltoni N.R. Gonzales M.S. Campbell R. Weir E.D. Cacayorin J.C. Grotta Intra-Arterial Reteplase Compared to Urokinase for Thrombolytic Recanalization in Acute

More information

Place for Interventional Radiology in Acute Stroke

Place for Interventional Radiology in Acute Stroke Place for Interventional Radiology in Acute Stroke Dr Lakmalie Paranahewa MBBS, MD(Radiology), FRCR Consultant Interventional Radiologist Asiri Group of Hospitals Objectives Imaging in Stroke Neurovascular

More information

ORIGINAL RESEARCH. Gabriel A. Vidal, MD, 1,2 James M. Milburn, MD 3

ORIGINAL RESEARCH. Gabriel A. Vidal, MD, 1,2 James M. Milburn, MD 3 ORIGINAL RESEARCH Ochsner Journal 16:486 491, 2016 Ó Academic Division of Ochsner Clinic Foundation The Penumbra 5MAX ACE Catheter Is Safe, Efficient, and Cost Saving as a Primary Mechanical Thrombectomy

More information

Current treatment options for acute ischemic stroke include

Current treatment options for acute ischemic stroke include ORIGINAL RESEARCH M.-N. Psychogios A. Kreusch K. Wasser A. Mohr K. Gröschel M. Knauth Recanalization of Large Intracranial Vessels Using the Penumbra System: A Single-Center Experience BACKGROUND AND PURPOSE:

More information

Little is known about the proportion of acute ischemic strokes

Little is known about the proportion of acute ischemic strokes ORIGINAL RESEARCH ADULT BRAIN Whole-Brain Susceptibility-Weighted Thrombus Imaging in Stroke: Fragmented Thrombi Predict Worse Outcome P.P. Gratz, G. Schroth, J. Gralla, H.P. Mattle, U. Fischer, S. Jung,

More information

Mechanical Thrombectomy Using a Solitaire Stent in Acute Ischemic Stroke; Initial Experience in 40 Patients

Mechanical Thrombectomy Using a Solitaire Stent in Acute Ischemic Stroke; Initial Experience in 40 Patients Journal of Cerebrovascular and Endovascular Neurosurgery ISSN 2234-8565, EISSN 2287-3139, http://dx.doi.org/10.7461/jcen.2012.14.3.164 Original Article Mechanical Thrombectomy Using a Solitaire Stent in

More information

Ruby Coil. Large Volume Detachable Coils

Ruby Coil. Large Volume Detachable Coils Ruby Coil Ruby Case Examples 38 mm Hepatic Artery Aneurysm Pulmonary AVM 2 Coils Dr. James Benenati Miami Cardiac and Vascular Institute, FL Y90 Embolization Type 2 Endoleak Dr. J Moskovitz Florida Hospital,

More information

Balloon Angioplasty for Intracranial Atherosclerotic Disease: a Multicenter Study

Balloon Angioplasty for Intracranial Atherosclerotic Disease: a Multicenter Study Balloon Angioplasty for Intracranial Atherosclerotic Disease: a Multicenter Study Lakshmi Sudha Prasanna Karanam 1, Mukesh Sharma 2, Anand Alurkar 3, Sridhar Reddy Baddam 1, Vijaya Pamidimukkala 1, and

More information

Endovascular Treatment for Acute Ischemic Stroke: Considerations from Recent Randomized Trials

Endovascular Treatment for Acute Ischemic Stroke: Considerations from Recent Randomized Trials Published online: March 13, 2015 1664 9737/15/0034 0115$39.50/0 Review Endovascular Treatment for Acute Ischemic Stroke: Considerations from Recent Randomized Trials Manabu Shirakawa a Shinichi Yoshimura

More information

1 Comparison of Warfarin and Aspirin for Symptomatic Intracranial Arterial Stenosis. N Engl J Med 352;13, March 31, 2005

1 Comparison of Warfarin and Aspirin for Symptomatic Intracranial Arterial Stenosis. N Engl J Med 352;13, March 31, 2005 The risk of ischemic stroke in patients with Intracranial Atherosclerotic Disease (ICAD) ranges from 7 to 24%. 1,2 Developed specifically for the treatment of ICAD, the Wingspan Stent System and Gateway

More information

Primary manual aspiration thrombectomy (MAT) for acute ischemic stroke: safety, feasibility and outcomes in 112 consecutive patients

Primary manual aspiration thrombectomy (MAT) for acute ischemic stroke: safety, feasibility and outcomes in 112 consecutive patients 1 Department of Neurosurgery, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA 2 Department of Neurology, Stroke Institute, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania,

More information

Comparative Analysis of Endovascular Stroke Therapy Using Urokinase, Penumbra System and Retrievable (Solitare) Stent

Comparative Analysis of Endovascular Stroke Therapy Using Urokinase, Penumbra System and Retrievable (Solitare) Stent www.jkns.or.kr http://dx.doi.org/10.3340/jkns.2015.57.5.342 J Korean Neurosurg Soc 57 (5) : 342-349, 2015 Print ISSN 2005-3711 On-line ISSN 1598-7876 Copyright 2015 The Korean Neurosurgical Society Clinical

More information

ORIGINAL PAPER. 8-F balloon guide catheter for embolization of anterior circulation aneurysms: an institutional experience in 152 patients

ORIGINAL PAPER. 8-F balloon guide catheter for embolization of anterior circulation aneurysms: an institutional experience in 152 patients Nagoya J. Med. Sci. 79. 435 ~ 441, 2017 doi:10.18999/nagjms.79.4.435 ORIGINAL PAPER 8-F balloon guide catheter for embolization of anterior circulation aneurysms: an institutional experience in 152 patients

More information

UNIVERSITY HOSPITAL UDINE/ITALY A SINGLE CENTRE EXPERIENCE IN STROKE TREATMET WITH EMBOTRAP II. TECHNOLOGY BASE ON CLOT RESEARCH

UNIVERSITY HOSPITAL UDINE/ITALY A SINGLE CENTRE EXPERIENCE IN STROKE TREATMET WITH EMBOTRAP II. TECHNOLOGY BASE ON CLOT RESEARCH UNIVERSITY HOSPITAL UDINE/ITALY A SINGLE CENTRE EXPERIENCE IN STROKE TREATMET WITH EMBOTRAP II. TECHNOLOGY BASE ON CLOT RESEARCH Massimo Sponza, Vladimir Gavrilović RIPERFUSION THERAPY Intraovenous thrombolysis

More information

Straub Endovascular System &

Straub Endovascular System & Straub Endovascular System & S t r a u b E n d o v a s c u l a r To o l s Straub Endovascular System Effective debulking in occluded arteries and veins Effective debulking in many indications Rotarex

More information

Stent Placement in Acute Cerebral Artery Occlusion Use of a Self-Expandable Intracranial Stent for Acute Stroke Treatment

Stent Placement in Acute Cerebral Artery Occlusion Use of a Self-Expandable Intracranial Stent for Acute Stroke Treatment Stent Placement in Acute Cerebral Artery Occlusion Use of a Self-Expandable Intracranial Stent for Acute Stroke Treatment Caspar Brekenfeld, MD; Gerhard Schroth, MD; Heinrich P. Mattle, MD; Do-Dai Do,

More information

Several new mechanical thrombectomy devices have been

Several new mechanical thrombectomy devices have been Published January 13, 2011 as 10.3174/ajnr.A2346 REVIEW ARTICLE I.Q. Grunwald A.K. Wakhloo S. Walter A.J. Molyneux J.V. Byrne S. Nagel A.L. Kühn M. Papadakis K. Fassbender J.S. Balami M. Roffi H. Sievert

More information

High-Flow, Small-Hole Arteriovenous Fistulas: Treatment with Electrodetachable Coils

High-Flow, Small-Hole Arteriovenous Fistulas: Treatment with Electrodetachable Coils High-Flow, Small-Hole Arteriovenous Fistulas: Treatment with Electrodetachable Coils Guido Guglielmi, Fernando Viñuela, Gary Duckwiler, Jacques Dion, and Alfredo Stocker Summary: We present one case of

More information

Distal arterial emboli may be sequelae of intravenous (IV)

Distal arterial emboli may be sequelae of intravenous (IV) ORIGINAL RESEARCH S. King P. Khatri J. Carrozella J. Spilker J. Broderick M. Hill T. Tomsick, for the IMS & IIMS II Investigators Anterior Cerebral Artery Emboli in Combined Intravenous and Intra-arterial

More information

System.

System. System www.penumbrainc.com POD System Case Examples 45 cm of POD Packing Coil in 1.5 to 3 mm diameter dilating vessel Bronchial Artery Embolization Dr. Amit Kakkar and Dr. Aksim Rivera, Bronx, NY Inflow:

More information

Evaluation of a new micromesh carotid stent with Optical Coherence Tomography.

Evaluation of a new micromesh carotid stent with Optical Coherence Tomography. Evaluation of a new micromesh carotid stent with Optical Coherence Tomography. Technical case report. J. Lemoine,S. Myla,Z. Chati,R.Aslam, M.Amor Clinic Louis Pasteur,Essey les Nancy.France Disclosure

More information

IMAGING IN ACUTE ISCHEMIC STROKE

IMAGING IN ACUTE ISCHEMIC STROKE IMAGING IN ACUTE ISCHEMIC STROKE Timo Krings MD, PhD, FRCP (C) Professor of Radiology & Surgery Braley Chair of Neuroradiology, Chief and Program Director of Diagnostic and Interventional Neuroradiology;

More information

The National Institute of Neurological Disorders and Stroke

The National Institute of Neurological Disorders and Stroke ORIGINAL RESEARCH A. Srinivasan M. Goyal P. Stys M. Sharma C. Lum Microcatheter Navigation and Thrombolysis in Acute Symptomatic Cervical Internal Carotid Occlusion BACKGROUND AND PURPOSE: The treatment

More information

Shallow aneurysms with wide necks pose a technical challenge

Shallow aneurysms with wide necks pose a technical challenge ORIGINAL RESEARCH INTERVENTIONAL Coil Protection Using Small Helical Coils for Wide-Neck Intracranial Aneurysms: A Novel Approach Y.D. Cho, J.Y. Lee, J.H. Seo, S.J. Lee, H.-S. Kang, J.E. Kim, O.-K. Kwon,

More information

Irretrievable unraveled coil remaining in the vascular lumen between the cerebral aneurysm and puncture site

Irretrievable unraveled coil remaining in the vascular lumen between the cerebral aneurysm and puncture site Technical Note JNET 3:42-46, 2009 Irretrievable unraveled coil remaining in the vascular lumen between the cerebral aneurysm and puncture site Kouhei NII 1) Masanari ONIZUK 1) Yoshirou KNEKO 2) Hiroshi

More information

Shallow aneurysms with wide necks pose a technical challenge

Shallow aneurysms with wide necks pose a technical challenge Published June 14, 2012 as 10.3174/ajnr.A3157 ORIGINAL RESEARCH Y.D. Cho J.Y. Lee J.H. Seo S.J. Lee H.-S. Kang J.E. Kim O.-K. Kwon Y.J. Son M.H. Han Coil Protection Using Small Helical Coils for Wide-Neck

More information

Extra- and intracranial tandem occlusions in the anterior circulation - clinical outcome of endovascular treatment in acute major stroke.

Extra- and intracranial tandem occlusions in the anterior circulation - clinical outcome of endovascular treatment in acute major stroke. Extra- and intracranial tandem occlusions in the anterior circulation - clinical outcome of endovascular treatment in acute major stroke. Poster No.: C-1669 Congress: ECR 2014 Type: Scientific Exhibit

More information

Endovascular Neurointervention in Cerebral Ischemia

Endovascular Neurointervention in Cerebral Ischemia Endovascular Neurointervention in Cerebral Ischemia Beyond Thrombolytics Curtis A. Given II, MD Co-Director, Neurointerventional Services Baptist Physician Lexington 72 y/o female with a recent diagnosis

More information

New and Future Endovascular Treatment Strategies for Acute Ischemic Stroke

New and Future Endovascular Treatment Strategies for Acute Ischemic Stroke New and Future Endovascular Treatment Strategies for Acute Ischemic Stroke Gary M. Nesbit, MD, George Luh, MD, Raymond Tien, MD, PhD, and Stanley L. Barnwell, MD, PhD Cerebral revascularization strategies

More information

Acute basilar artery occlusion (BAO) carries a high morbidity

Acute basilar artery occlusion (BAO) carries a high morbidity ORIGINAL RESEARCH INTERVENTIONAL Combined Use of Mechanical Thrombectomy with Angioplasty and Stenting for Acute Basilar Occlusions with Underlying Severe Intracranial Vertebrobasilar Stenosis: Preliminary

More information

Acute vertebrobasilar occlusion is generally associated with an

Acute vertebrobasilar occlusion is generally associated with an ORIGINAL RESEARCH INTERVENTIONAL Treatment of Acute Vertebrobasilar Occlusion Using Thrombectomy with Stent Retrievers: Initial Experience with 18 Patients M. Espinosa de Rueda, G. Parrilla, J. Zamarro,

More information

Pipeline Embolization Device

Pipeline Embolization Device Pipeline Embolization Device The power to redefine aneurysm treatment. REDEFINE The Pipeline device redefines treatment for large or giant wide-necked aneurysms by reconstructing the parent artery and

More information

Use of EKOS Catheter in the management of Venous Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group

Use of EKOS Catheter in the management of Venous Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group Use of EKOS Catheter in the management of Venous Thromboembolism @ Mr. Manoj Niverthi, Mr. Sarang Pujari, and Ms. Nupur Dandavate, The GTF Group Introduction Georgia Thrombosis Forum (GTF, www.gtfonline.net)

More information

Advances in Neuro-Endovascular Care for Acute Stroke

Advances in Neuro-Endovascular Care for Acute Stroke Advances in Neuro-Endovascular Care for Acute Stroke Ciarán J. Powers, MD, PhD, FAANS Associate Professor Program Director Department of Neurological Surgery Surgical Director Comprehensive Stroke Center

More information

Mechanical thrombectomy in Plymouth. Will Adams. Will Adams

Mechanical thrombectomy in Plymouth. Will Adams. Will Adams Mechanical thrombectomy in Plymouth Will Adams Will Adams History Intra-arterial intervention 1995 (NINDS) iv tpa improved clinical outcome in patients treated within 3 hours of ictus but limited recanalisation

More information

Acute Management of Stroke due to Intracranial Steno-occlusion. Joon-Tae Kim, MD, PhD Department of Neurology Chonnam National University Hospital

Acute Management of Stroke due to Intracranial Steno-occlusion. Joon-Tae Kim, MD, PhD Department of Neurology Chonnam National University Hospital Acute Management of Stroke due to Intracranial Steno-occlusion Joon-Tae Kim, MD, PhD Department of Neurology Chonnam National University Hospital None Disclosure Contents Current status of acute management

More information

SEE IT. BELIEVE IT. THE CONFIDENCE OF CLARITY. Solitaire Platinum. Revascularization Device

SEE IT. BELIEVE IT. THE CONFIDENCE OF CLARITY. Solitaire Platinum. Revascularization Device SEE IT. BELIEVE IT. THE CONFIDENCE OF CLARITY. Revascularization THE CONFIDENCE OF CLARITY. The key features that make the device effective have been retained including our unique Parametric overlapping

More information

Historical. Medical Policy

Historical. Medical Policy Medical Policy Subject: Mechanical Embolectomy for Treatment of Acute Stroke Policy #: SURG.00098 Current Effective Date: 01/01/2016 Status: Revised Last Review Date: 08/06/2015 Description/Scope This

More information

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved. Interventional Radiology Coding Case Studies Prepared by Stacie L. Buck, RHIA, CCS-P, RCC, CIRCC, AAPC Fellow President & Senior Consultant Week of October 29, 2018 Mesenteric Arteriogram & Thrombectomy/Thrombolysis

More information

NIH Public Access Author Manuscript J Am Coll Radiol. Author manuscript; available in PMC 2013 June 24.

NIH Public Access Author Manuscript J Am Coll Radiol. Author manuscript; available in PMC 2013 June 24. NIH Public Access Author Manuscript Published in final edited form as: J Am Coll Radiol. 2010 January ; 7(1): 73 76. doi:10.1016/j.jacr.2009.06.015. Cerebral Aneurysms Janet C. Miller, DPhil, Joshua A.

More information

The principal goal in treating acute ischemic stroke is rapid

The principal goal in treating acute ischemic stroke is rapid ORIGINAL RESEARCH S. Sugiura K. Iwaisako S. Toyota H. Takimoto Simultaneous Treatment with Intravenous Recombinant Tissue Plasminogen Activator and Endovascular Therapy for Acute Ischemic Stroke Within

More information

Chronic Total Occlusion (CTO) Technologies

Chronic Total Occlusion (CTO) Technologies to receive our latest news and key activities. Chronic Total Occlusion (CTO) Technologies Re-open vital channels LinkedIn page Follow us on CORDIS EMEA OUTBACK LTD Re-Entry Catheter True Lumen Re-Entry

More information

Directions For Use. All directions should be read before use

Directions For Use. All directions should be read before use Directions For Use All directions should be read before use DEVICE DESCRIPTION: The CLEANER.XT Rotational Thrombectomy System is a percutaneous, 6Fr catheter based system (single piece construction) that

More information

Copyright HMP Communications

Copyright HMP Communications Ocelot With Wildcat in a Complicated Superficial Femoral Artery Chronic Total Occlusion Soundos K. Moualla, MD, FACC, FSCAI; Richard R. Heuser, MD, FACC, FACP, FESC, FSCAI From Phoenix Heart Center, Phoenix,

More information

Disclosures. Anesthesia for Endovascular Treatment of Acute Ischemic Stroke. Acute Ischemic Stroke. Acute Stroke = Medical Emergency!

Disclosures. Anesthesia for Endovascular Treatment of Acute Ischemic Stroke. Acute Ischemic Stroke. Acute Stroke = Medical Emergency! Disclosures Anesthesia for Endovascular Treatment of Acute Ischemic Stroke I have nothing to disclose. Chanhung Lee MD, PhD Associate Professor Anesthesia and perioperative Care Acute Ischemic Stroke 780,000

More information

Patient selection for i.v. thrombolysis and thrombectomy

Patient selection for i.v. thrombolysis and thrombectomy 3 rd Congress of the European Academy of Neurology Amsterdam, The Netherlands, June 24 27, 2017 Teaching Course 8 Acute treatment and early secondary prevention of stroke Level 2 Patient selection for

More information

North American Clinical Experience with the EKOS MicroLysUS Infusion Catheter for the Treatment of Embolic Stroke

North American Clinical Experience with the EKOS MicroLysUS Infusion Catheter for the Treatment of Embolic Stroke AJNR Am J Neuroradiol 24:534 538, March 2003 North American Clinical Experience with the EKOS MicroLysUS Infusion Catheter for the Treatment of Embolic Stroke Brian R. Mahon, Gary M. Nesbit, Stanley L.

More information

The National Institutes of Health Stroke Scale (NIHSS)

The National Institutes of Health Stroke Scale (NIHSS) National Institutes of Health Stroke Scale Score and Vessel Occlusion in 2152 Patients With Acute Ischemic Stroke Mirjam R. Heldner, MD; Christoph Zubler, MD; Heinrich P. Mattle, MD; Gerhard Schroth, MD;

More information

Stent-Assisted Recanalization for Acute Ischemic Stroke

Stent-Assisted Recanalization for Acute Ischemic Stroke Stent-Assisted Recanalization for Acute Ischemic Stroke Department of 1 Neurosurgery, 2 Diagnostic Radiology, Cerebrovascular Center, Yongdong Severance Hospital, Yonsei University College of Medicine,

More information

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER

NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER Page 1 of 5 NON-COMPLIANT PTCA RAPID EXCHANGE DILATATION CATHETER STERILE. SINGLE USE ONLY. Sterilized with ethylene oxide gas. Non pyrogenic. Do not resterilize. Do not use opened or damaged packages.

More information

Delayed treatment for a case of acute ischaemic stroke using mechanical embolectomy the L5 Merci Retriever

Delayed treatment for a case of acute ischaemic stroke using mechanical embolectomy the L5 Merci Retriever Case Study Australian Institute of Radiography The Radiographer 2009; 56 (2): 27 33 Delayed treatment for a case of acute ischaemic stroke using mechanical embolectomy the L5 Merci Retriever J Velkovic

More information

ANGIO Mentor Based Neurovascular Interventions Training Curriculum

ANGIO Mentor Based Neurovascular Interventions Training Curriculum SIMULATION BASED NEUROVASCULAR INTERVENTIONS TRAINING CURRICULUM i ANGIO Mentor Based Neurovascular Interventions Training Curriculum Description In the era of duty hour restrictions and increasing medico-legal

More information

PARADIGM SHIFT FOR THROMBOLYSIS IN PATIENTS WITH ACUTE ISCHAEMIC STROKE, FROM EXTENSION OF THE TIME WINDOW TO RAPID RECANALISATION AFTER SYMPTOM ONSET

PARADIGM SHIFT FOR THROMBOLYSIS IN PATIENTS WITH ACUTE ISCHAEMIC STROKE, FROM EXTENSION OF THE TIME WINDOW TO RAPID RECANALISATION AFTER SYMPTOM ONSET PARADIGM SHIFT FOR THROMBOLYSIS IN PATIENTS WITH ACUTE ISCHAEMIC STROKE, FROM EXTENSION OF THE TIME WINDOW TO RAPID RECANALISATION AFTER SYMPTOM ONSET Hye Seon Jeong, *Jei Kim Department of Neurology and

More information

Endovascular treatment for pseudoocclusion of the internal carotid artery

Endovascular treatment for pseudoocclusion of the internal carotid artery Endovascular treatment for pseudoocclusion of the internal carotid artery Daqiao Guo, Xiao Tang, Weiguo Fu Institute of Vascular Surgery, Fudan University, Department of Vascular Surgery, Zhongshan Hospital

More information

Acoustic Pulse Thrombolysis Treatment

Acoustic Pulse Thrombolysis Treatment Acoustic Pulse Thrombolysis Treatment BTGVascular.com SETTING THE STANDARD FOR VASCULAR THERAPIES Quickly & safely dissolve thrombus with the EKOS System. The Acoustic Pulse Difference Acoustic Pulse Thrombolysis

More information

Repair of Intracranial Vessel Perforation with Onyx-18 Using an Exovascular Retreating Catheter Technique

Repair of Intracranial Vessel Perforation with Onyx-18 Using an Exovascular Retreating Catheter Technique Repair of Intracranial Vessel Perforation with Onyx-18 Using an Exovascular Retreating Catheter Technique Michael Horowitz M.D. Pittsburgh, Pennsylvania Background Iatrogenic intraprocedural rupture rates

More information

Balloon-assisted guide catheter positioning to overcome extreme cervical carotid tortuosity: technique and case experience

Balloon-assisted guide catheter positioning to overcome extreme cervical carotid tortuosity: technique and case experience Department of Neurosurgery, Stony Brook University Medical Center, Cerebrovascular Center, Stony Brook, New York, USA Correspondence to Dr D Fiorella, Department of Neurological Surgery, Stony Brook University

More information

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved. Interventional Radiology Coding Case Studies Prepared by Stacie L. Buck, RHIA, CCS-P, RCC, CIRCC, AAPC Fellow President & Senior Consultant Week of June 4, 2018 Thrombolysis, Thrombectomy & Angioplasty

More information

Rheolytic Thrombectomy of the Occluded Internal Carotid Artery in the Setting of Acute Ischemic Stroke

Rheolytic Thrombectomy of the Occluded Internal Carotid Artery in the Setting of Acute Ischemic Stroke AJNR Am J Neuroradiol 22:526 530, March 2001 Rheolytic Thrombectomy of the Occluded Internal Carotid Artery in the Setting of Acute Ischemic Stroke Richard J. Bellon, Christopher M. Putman, Ronald F. Budzik,

More information

TIPS & TRICKS for optimal SILK placement

TIPS & TRICKS for optimal SILK placement I. Antiplatelet Premedication: The correctly chosen premedication is essential. PLAVIX (minimum 600mg) should ideally be administered during eight days prior to procedure (75mg per day). However, a loading

More information

ACUTE STROKE INTERVENTION: THE ROLE OF THROMBECTOMY AND IA LYSIS

ACUTE STROKE INTERVENTION: THE ROLE OF THROMBECTOMY AND IA LYSIS Associate Professor of Neurology Director of Neurointerventional Services University of Louisville School of Medicine ACUTE STROKE INTERVENTION: THE ROLE OF THROMBECTOMY AND IA LYSIS Conflict of Interest

More information

RECOMMENDED INSTRUCTIONS FOR USE

RECOMMENDED INSTRUCTIONS FOR USE Rapid Exchange PTCA Dilatation Catheter RECOMMENDED INSTRUCTIONS FOR USE Available in diameters 1.25mm to 4.5mm and in lengths 09mm to 40mm Caution: This device should be used only by physicians trained

More information

Abstract Aims: Large intracoronary thrombus in patients with acute coronary syndromes remains a challenge in percutaneous SPECIAL REPORT

Abstract Aims: Large intracoronary thrombus in patients with acute coronary syndromes remains a challenge in percutaneous SPECIAL REPORT SPECIAL REPORT First-in-man intracoronary use of the Trevo Pro 4 mechanical thrombectomy device for the retrieval of large intracoronary thrombus in patients with acute coronary syndromes Tim P. van de

More information

Basilar Thrombus and Occlusion of the Left Vertebral Artery

Basilar Thrombus and Occlusion of the Left Vertebral Artery WHAT WOULD YOU DO? Basilar Thrombus and Occlusion of the Left Vertebral Artery MODERATOR: GUILHERME DABUS, MD, FAHA PANEL: DONALD FREI, MD; ANSAAR T. RAI, MD, MBA; AND SUDHAKAR R. SATTI, MD CASE PRESENTATION

More information

Understanding aneurysms and flow diversion treatment

Understanding aneurysms and flow diversion treatment Surpass Streamline Flow Diverter See package insert for complete indications, contraindications, warnings and instructions for use. INTENDED USE / INDICATIONS FOR USE The Surpass Streamline Flow Diverter

More information

Acute Stroke Treatment: Current Trends 2010

Acute Stroke Treatment: Current Trends 2010 Acute Stroke Treatment: Current Trends 2010 Helmi L. Lutsep, MD Oregon Stroke Center Oregon Health & Science University Overview Ischemic Stroke Neuroprotectant trials to watch for IV tpa longer treatment

More information

Novel non-occlusive temporary endoluminal neck protection device to assist in the treatment of aneurysms in a canine model

Novel non-occlusive temporary endoluminal neck protection device to assist in the treatment of aneurysms in a canine model 1 Stroke and Cerebrovascular Center, Department of Neurosciences, Medical University of South Carolina, Charleston, South Carolina, USA 2 Stroke and Cerebrovascular Center, Department of Radiology, Medical

More information